Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K200624

    Validate with FDA (Live)

    Date Cleared
    2020-04-27

    (48 days)

    Product Code
    Regulation Number
    868.5120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Percutaneous Introducer is intended to allow for the percutaneous placement of devices, needles or probes, in close proximity to nerves and around or into surgical wound or non-surgical wound sites. It may be used to inject or aspirate the introduction area via the luer hub of the needle/introducer.

    Device Description

    The Percutaneous Introducer is a single-use, disposable device consisting of a flexible polymer sheath with a molded plastic hub. The device incorporates a stainless steel needle inside the flexible cannula and is removed after the introducer is placed. The inserted stainless steel introducer needle has a sharp distal point. The flexible cannula acts as a conduit for percutaneous introduction of medical devices, such as catheters, needles, or probes. A stylet or porous plug (to prevent aspiration during insertion) comes seated in the standard 6 degree luer fitting of the proximal end of the needle hub. The Percutaneous Introducers are used for a short period of time and are supplied sterile and non-pyrogenic. The Percutaneous Introducer is available in four lengths and a 14 gauge.

    AI/ML Overview

    This document is a 510(k) summary for a "Percutaneous Introducer" device. It outlines the device description, intended use, technological characteristics, and non-clinical testing performed to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" with quantitative targets and corresponding "reported device performance" in the way a clinical study might. Instead, it describes various tests performed and states that the device "meets the requirements" or "performs comparably" or has "successful results."

    The tests and their reported outcomes are:

    Acceptance Criteria (Implicit from tests)Reported Device Performance
    Mechanical Performance:
    Penetration Force TestPerforms comparably to predicate device (K102460)
    Simulated Use Penetration/ Withdrawal/ Accordion/ Peelback TestPerforms comparably to predicate device (K102460); shown to verify resistance to collapsing and accordion/ peelback during tissue penetration.
    Tensile StrengthPerforms comparably to predicate device (K102460); proven to possess the ability to withstand a minimum acceptable tensile force.
    Material Biocompatibility:
    ISO 10993 compliant biocompatibilityMeets necessary biocompatibility requirements as per FDA guidance. Materials used support Substantial Equivalence.
    Cytotoxicity (ISO10993-5:2009)Tested with successful results.
    Sensitization (ISO10993-10:2010)Tested with successful results.
    Irritation/ Intracutaneous Reactivity (ISO10993-10:2010)Tested with successful results.
    EtO Residuals (I-CHM-2136 Rev. 0)Tested with successful results.
    Pyrogenicity (ISO10993-12:2017)Tested with successful results.
    Rabbit Blood Hemolysis (ISO10993-12:2017)Tested with successful results.
    Sterility & Packaging:
    Packaging- SterilityTested with successful results.
    Burst Testing (ASTM F 1140)Tested with successful results.
    Dye Penetration (ASTM F 1929)Tested with successful results.
    Functional Standards:
    ISO 594-1 (Luer Fittings)Tested with successful results.
    ISO 594-2 (Luer Fittings)Tested with successful results.
    ISO 10555-5 (Sterile, single-use intravascular catheters - introducers and dilators)Tested with successful results.

    The study concludes that based on these results, the Percutaneous Introducer "performs comparably to, and is substantially equivalent to predicate device (K102460)."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for any of the non-clinical tests (e.g., number of introducers tested for tensile strength, burst testing, etc.).
    The data provenance is from non-clinical bench and laboratory testing, not human or animal studies. The data would originate from the manufacturing and testing facilities of Epimed International, Inc. or their contracted labs. It is inherently prospective in the sense that the tests were conducted specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This section is not applicable in the traditional sense of clinical or diagnostic device ground truth. The "ground truth" for non-clinical performance and biocompatibility tests is established by adherence to recognized international standards (e.g., ISO, ASTM) and FDA guidance documents. The "experts" would be the qualified personnel within the testing laboratories who perform the tests according to these standards, generate the data, and interpret the results to ensure compliance. Their specific number or qualifications are not detailed here, as the focus is on adherence to standardized testing protocols.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving multiple human readers for diagnostic image interpretation or similar subjective assessments. The tests described are objective, standardized laboratory and bench tests where results are determined by measurement against specified criteria in the standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret results, and the AI's impact on their performance is being evaluated. This submission is for an interventional medical device (introducer) and focuses on its physical performance, material safety, and substantial equivalence, not diagnostic accuracy or human reader performance.

    6. Standalone (Algorithm Only) Performance

    No, a standalone (algorithm only) performance assessment was not done because the device is a physical medical instrument (a Percutaneous Introducer), not a software algorithm or AI-driven diagnostic tool. The concept of "algorithm only performance" does not apply to this type of device.

    7. Type of Ground Truth Used

    The "ground truth" for the non-clinical tests is based on established international standards and regulatory guidelines. For example:

    • Mechanical properties: Defined by performance criteria within ASTM and ISO standards (e.g., specific force values, absence of collapse/peelback).
    • Biocompatibility: Defined by the requirements and endpoints outlined in ISO 10993 series of standards (e.g., non-cytotoxic, non-sensitizing, non-irritating).
    • Sterility and Packaging: Defined by standards like ASTM F1140, ASTM F1929, and general sterility assurance principles.
    • Functional standards: Defined by the specific requirements of ISO 594-1, ISO 594-2, and ISO 10555-5.

    Essentially, the "ground truth" is compliance with these widely accepted benchmarks for medical device safety and performance.

    8. Sample Size for the Training Set

    Not applicable. This device is a physical medical instrument, not an AI or machine learning model. Therefore, there is no "training set" in the context of an algorithm. The development and testing process involves engineering design, material selection, and verification/validation testing, not data-driven machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191295

    Validate with FDA (Live)

    Date Cleared
    2019-08-06

    (84 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Introducer for Endoscopic/Laparoscopic procedures.

    Device Description

    The PMIPIN15 Percutaneous Introducer Needle is a single- use, disposable EO sterilized device made of 304 stainless steel with a luer hub made of polycarbonate. The introducer sheath is 5 Fr and manufactured from low density polyethylene. The device is manually inserted by the physician during endoscopic and/or laparoscopic procedures.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the PMIPIN15 Percutaneous Introducer Needle.

    The document is a 510(k) summary for a medical device, which primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a novel device with extensive new clinical performance data. Therefore, the "acceptance criteria" here are largely derived from the performance and characteristics of the predicate device and relevant international standards.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) submission, the "acceptance criteria" are essentially demonstrating equivalence to the predicate device and compliance with relevant standards. The "device performance" is presented as meeting these standards and being similar to the predicate.

    Acceptance Criteria CategorySpecific Acceptance Criterion (Implicitly derived or explicitly stated)Reported Device Performance
    BiocompatibilityCompliant with ISO 10993 series for medical devices.Passed Cytotoxicity Test (ISO10993-5), Dermal Sensitization Guinea Pig Maximization Tests (ISO10993-10), Intracutaneous Test (ISO10993-10).
    SterilizationValidated for Ethylene Oxide (EO) sterilization.Passed Ethylene Oxide Sterilization Validation (ISO 11135:2014).
    PackagingValidated for package integrity and shelf-life.Passed Package Validation (ASTM 2096-11, ASTM F1186/1886M-16, ASTM F88/F88M-15, ASTM D5276-98).
    Physical/Mechanical PerformanceEquivalent to predicate device in design and materials; compliant with ISO 11070:2014(E) for needle and hub strength, force at break, and penetration.Physical measurements and visual inspection confirmed similarity to predicate. Performance evaluation (strength of union, force at break) met ISO 11070:2014(E). Needle penetration testing met ISO 11070:2014(E).
    Accelerated Aging/Shelf LifeValidated for shelf life.Passed Accelerated aging testing per ASTM F1980-16. Device "met its design specifications" for a 2-year shelf life.
    Design SpecificationsMeet established design specifications (e.g., needle gauge, length, introducer diameter, materials).The device "met its design specifications" and shown to be similar to the predicate for all listed specifications.
    Indications for UseSame as predicate."Introducer for Endoscopic/Laparoscopic procedures." - Same as predicate.
    SterilityMust be sterile."Yes" (Sterility) - Same as predicate.
    Single UseMust be single use."Yes" (Single Use) - Same as predicate.
    BiocompatibleMust be biocompatible."Yes" (Biocompatible) - Same as predicate.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not provide specific sample sizes for the individual tests (biocompatibility, sterilization, package, mechanical performance). It states "A series of preclinical tests were performed" and refers to ISO and ASTM standards, which would dictate appropriate sample sizes for each test.

    • Test Set Sample Size: Not explicitly stated (implied by compliance with standards).
    • Data Provenance: The tests are non-clinical (laboratory-based, material-based). There's no human clinical data presented, so country of origin and retrospective/prospective distinctions are not applicable in the typical sense. Physical measurements and performance evaluations were done on samples of the new device and the predicate device. Needle penetration testing was conducted by an "independent lab."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. This device's evaluation is based on non-clinical engineering and materials testing, not on interpretation by clinical experts. The "ground truth" for these types of tests comes from the established scientific methods and standards (ISO, ASTM).


    4. Adjudication Method for the Test Set

    Not applicable. As described above, there is no clinical data or expert interpretation requiring adjudication. Performance is assessed against quantitative standards.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-powered diagnostic or assistive device. It is a physical percutaneous introducer needle.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device.


    7. The Type of Ground Truth Used

    For the non-clinical tests, the "ground truth" is based on:

    • International Standards: ISO 10993 (biocompatibility), ISO 11135 (sterilization), ISO 11070 (needles), ASTM standards (packaging, shelf-life). These standards define acceptable methodologies and performance thresholds.
    • Predicate Device Characteristics: For comparative aspects (design, materials, indications for use), the characteristics of the legally marketed predicate device (Ranfac Percutaneous Introducer PIN-15, K951090) serve as a de-facto "ground truth" for substantial equivalence.

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI device that requires a training set.


    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123974

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-02-21

    (57 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is to assist in the introduction of diagnostic or therapeutic devices into a vessel.

    Device Description

    Teleflex Medical OEM's Percutaneous Introducer and Kit is used to assist in the introduction of diagnostic or therapeutic devices into a vessel. These devices will be marketed in two configurations:

    • Percutaneous Introducer only, or
    • Percutaneous Introducer. Needle, Kit including Introducer Syringe and Guidewire

    A description of each individual component is provided below:

    Guidewire
    The guidewire is designed to allow ease of movement through the vessels of the bodv.

    18 Ga Needle
    The 18 Ga Needle design allows the syringe to be attached to it and for the guidewire to be inserted through it.

    Syringe
    The 10ml syringe design allows for insertion into the needle.

    Percutaneous Introducer
    The Percutaneous Introducer uses a secure locking luer collar hub design and also incorporates an ergonomic handle design. This protects against separation of the dilator and sheath assembly during insertion. The introducer is designed to allow the dilator to be advanced over the guidewire into the vessel thus facilitating the introduction of the sheath into the vessel. This device design is such that after advancement into the vessel the dilator can be removed to allow access through the sheath of the customer's chosen device. For easy removal, the split sheath can be removed by breaking the hub and removing the two halves of the sheath while still within the vessel. The sheath hub is designed to snap cleanly. The sheath peels easily and evenly, thus reducing complications during the procedure.

    AI/ML Overview

    The provided text describes a 510(k) summary for a medical device, the Teleflex Medical Percutaneous Introducer and Kit. It focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed study of the device meeting specific acceptance criteria in the manner typically associated with an AI/algorithm-driven device.

    Therefore, many of the requested elements (e.g., sample size for test/training sets, experts for ground truth, MRMC study, standalone performance) are not applicable as this submission pertains to a traditional medical device, not an AI medical device.

    Here's the information that can be extracted and a summary of why other points are not relevant based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The document states that the device was subjected to performance testing with pre-determined acceptance criteria, and it met all acceptance criteria. However, it does not provide an explicit table of these criteria or specific quantitative results for each. It lists general categories of tests.

    Acceptance Criteria CategoryReported Device Performance
    DimensionalMet all acceptance criteria
    TensileMet all acceptance criteria
    RadiopacityMet all acceptance criteria
    Corrosion ResistanceMet all acceptance criteria
    LeakageMet all acceptance criteria
    Other (not specified)Met all acceptance criteria

    Study Proving Device Meets Acceptance Criteria:

    The study consisted of "a full battery of performance testing." The document does not provide a specific name or detailed protocol for this study but indicates it was conducted by Teleflex Medical OEM.

    Non-Applicable Information for this Device (as per the provided text):

    Since this is a submission for a traditional medical device (percutaneous introducer kit) and not an AI/algorithm-driven device, the following points are not directly addressed or are not relevant to this type of 510(k) submission based on the provided document:

    • Sample size used for the test set and the data provenance: Not applicable. Performance testing for this type of device typically involves laboratory-based tests on device samples, not patient data in the context of an AI model.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI test set is not relevant here.
    • Adjudication method for the test set: Not applicable.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted diagnostic device.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI algorithm.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. The "ground truth" for this device's performance would be the specifications and standards it must meet (e.g., material strength, dimension tolerances).
    • The sample size for the training set: Not applicable. This device does not use an AI model that requires a training set.
    • How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K013120

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-01-24

    (128 days)

    Product Code
    Regulation Number
    882.5880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Percutaneous Introducer Set is intended for percutaneous introduction, manipulation, and removal of stimulation leads.

    Device Description

    The Percutaneous Introducer Set is a single use device allowing for introduction, manipulation, and removal of stimulation leads after percutaneous entry is gained with a needle. The sheath is made of HDPE (High Density Polyethylene) with a HDPE hub. It is slightly tapered at the distal end to make a smooth transition to the dilator. The dilator is made of 304 stainless steel tubing tapered at the distal end for smooth insertion. It has a molded HDPE hub. The guidewire directional guide is made of 304 stainless steel wire with radiuses at each end. It has depth markings etched into the surface along the distal end.

    AI/ML Overview

    This document, K013120, is a 510(k) summary for a Percutaneous Introducer Set. The device is intended for the introduction, manipulation, and removal of stimulation leads.

    The document indicates that the device's substantial equivalence is based on its similarity to a predicate device, the MedAmicus Epidural Introducer (K994097), in terms of intended use, basic design, manufacturing methods, and materials.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document
    does NOT list specific acceptance criteria with quantifiable metrics, nor does it provide a detailed table of reported device performance against such criteria.

    Instead, it offers a general statement: "Performance data indicates that the Percutaneous Introducer Set meets the functional requirements and specifications for this device."

    To infer potential "acceptance criteria" based on the provided information, we can consider the aspects of similarity to the predicate device and relevant material properties:

    Acceptance Criterion (Inferred from document)Reported Device Performance (as stated in document)
    Intended Use Equivalence: Percutaneous introduction, manipulation, and removal of stimulation leads."The Percutaneous Introducer Set is intended for percutaneous introduction, manipulation, and removal of stimulation leads." "The intended use... are similar to the predicate device..."
    Basic Design Equivalence: Physical structure and components (sheath, dilator, guidewire)."The basic design... are similar to the predicate device..." Detailed description of components: HDPE sheath, HDPE hub, 304 stainless steel dilator, 304 SS guidewire, platinum iridium marker band.
    Manufacturing Methods Equivalence: Processes used to produce the device."The methods of manufacturing... are similar to the predicate device..." "The sheath materials and manufacturing processes for the Percutaneous Introducer Set are equivalent to those used on the predicate MedAmicus device..."
    Material Equivalence: Composition of parts that come into contact with the patient or leads."The materials are similar to the predicate device..." "Materials used in the Percutaneous Introducer Set are identical to materials previously approved/cleared for market release." (Specific examples: HDPE identical to K990705, 304 SS identical to P970004, Platinum Iridium identical to P970004).
    Biocompatibility: Safety for contact with tissue."Applications referred to below are for Externally Communicating - Blood Path Indirect - Limited Exposure which includes all the testing required above." "Since the materials are identical to those previously approved/cleared, there are no new issues involving safety and effectiveness."
    Sterility: Assurance of being free from viable microorganisms."Present sterilization cycle assures a Sterility Assurance Level (SAL) of >10-6."
    Packaging Equivalence: Maintenance of sterility and physical integrity."The packaging materials for the Percutaneous Introducer Set are equivalent to the predicate MedAmicus Epidural Introducer. These materials consist of a double sterile barrier using Tyvek/Mylar pouches."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide information on a specific "test set" with a defined sample size or data provenance (e.g., country of origin, retrospective or prospective). The submission relies on establishing substantial equivalence to a predicate device, and the "performance data" mentioned is a general statement, not a detailed study report.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not provided in the document. The submission focuses on device equivalence rather than a study requiring expert-established ground truth for a test set.

    4. Adjudication Method

    This information is not provided in the document. As no specific test set or expert ground truth establishment is described, neither is an adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not performed or reported in this 510(k) submission. The document aims for substantial equivalence based on physical and functional characteristics, not on the impact of the device on human readers' performance with or without AI assistance.

    6. Standalone Performance Study

    A standalone (algorithm-only without human-in-the-loop performance) study was not performed or reported in this 510(k) submission. This is a medical device, not an AI/algorithm-based diagnostic tool. The "performance data" mentioned is a general statement about meeting functional requirements.

    7. Type of Ground Truth Used

    The concept of "ground truth" as typically used in the context of diagnostic performance studies (e.g., expert consensus, pathology, outcomes data) does not apply directly to this 510(k) submission. The "truth" for this submission is based on the demonstration of substantial equivalence to a legally marketed predicate device and the "identicality" of materials to those previously approved.

    8. Sample Size for the Training Set

    This information is not applicable and therefore not provided. This is not an AI/machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and therefore not provided. As there is no training set for an AI/machine learning model, there is no ground truth establishment for such a set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K000824

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-04-13

    (30 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K961466

    Validate with FDA (Live)

    Date Cleared
    1996-09-12

    (148 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K961219

    Validate with FDA (Live)

    Date Cleared
    1996-06-21

    (85 days)

    Product Code
    Regulation Number
    878.4350
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The proposed kit is indicated for suprapubic bladder drainage and percutancous access for the placement of theminsensor probes in patients undergoing ultrasound-guided urological cryosurgery procedures in accordance with the indications of legally marketed cryosurgery systems.

    Device Description

    The kit is comprised of guidewire(s), access needles(s), dilator(s), divinage catheter with connecting tube, scalpel and syringe.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a medical device (Percutaneous Introducer System) from 1996. It does not contain the type of detailed study information, acceptance criteria, or performance data that would be relevant to evaluating an AI/ML powered device. The document predates the widespread use of AI/ML in medical devices by decades.

    Therefore, I cannot extract the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies, as these concepts are not applicable to this 1996 device submission.

    The document primarily focuses on establishing "substantial equivalence" of the device and its components to already legally marketed devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1