Search Filters

Search Results

Found 15 results

510(k) Data Aggregation

    K Number
    K200893

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-05-04

    (396 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pronto Pump Tube Set is a part of the Pronto Pump System. This tube set provides peristaltic pump driven fluid transfer that facilities repeatable drug dosage distribution in healthcare environments. The Pronto Pump Tube set provides the fluid path way and pumping mechanism for the pump system.

    The device is for use with IV bags, syringes, elastomeric infusers, and other drug administration containers. Sets are sold sterile.

    The ProntoPump Sterile Tube Sets are fluid transfer tube sets used in conjunction with the ProntoPump pharmacy pump in hospital and compounding pharmacies to provide a pathway through which fluid is transferred from one source container into another suitable container.

    The device is for use with IV bags, syringes, elastomeric infusers, and other drug administration containers. Sets are sold sterile.

    Device Description

    The ProntoPump sterile tube set is a single channel tube set for use with the ProntoPump pharmacy pump. It consists of clear plastic tubing connected to a silicone tube that functions as a peristaltic pumping chamber. Peristaltic pumps function by a roller compressing the pump tubing and rolling along the length of the tube, squeezing the fluid inside the tube from one end to the other. It has a plastic cassette that holds the pumping chamber in place for insertion into the ProntoPump, which provides the rollers that the force the fluid to move. The proximal and distal ends of the PVC tubing have connectors for various devices and containers. Using the ProntoPump, the tube set is used to deliver specified volumes of liquid into a final dosing container in hospital pharmacies or compounding pharmacies. The device is for use with IV bags, syringes, elastomeric infusers, and other drug administration containers. Sets are sold sterile.

    The cassette is non-fluid-contact and helps hold the tubing in place for loading. It has a machine-readable barcode that can be read by the pharmacy pump. The barcode includes information about the particular tube set item number, lot number, and expiration date, which is used by pharmacy pump to facilitate set up of the tubing.

    The tube set is intended to be used by trained healthcare personnel, and is not intended to be used for direct patient contact. Tube sets are packaged individually in a peel pouch and sterilized by radiation.

    AI/ML Overview

    The provided text describes the regulatory clearance for the "ProntoPump Sterile Tube Set" (K200893) and its substantial equivalence to a predicate device (K062909). However, it does not explicitly state "acceptance criteria" for performance in a table format with specific quantitative goals and reported device performance against those goals for the K200893 device. Instead, it compares the technological characteristics and performance of the subject device to the predicate in a qualitative manner and mentions general ISO and ASTM standards to which the device was tested.

    The closest information relating to performance criteria and reported performance is about Volume Accuracy and Minimum Flow Rate, which are compared to the predicate device.

    Here's an attempt to create the requested table based on the limited quantitative performance data provided, and then address the other points.


    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided document, explicit "acceptance criteria" with pass/fail thresholds are not clearly stated for the K200893 device. Instead, the document compares its performance characteristics to the predicate device. The information below is extracted from the "Comparison of Technological Characteristics with The Predicate Device" table and the "Differences Compared to Predicate Device / Substantial Equivalence Discussion" section.

    Performance CharacteristicAcceptance Criteria (from predicate/discussion, if explicit)Reported Device Performance (ProntoPump Sterile Tube Set K200893)
    Volume AccuracyPredicate: +/- 0.02 mL (0.2-2.0 mL), +/- 1% (>2.0 mL)New Device: +/- 5% (1.0-5.0 mL), +/- 4% (>5 mL)
    Minimum Flow Rate at Top SpeedPredicate: At least 11 mL /sNew Device: At least 21 mL /s
    Tube Set Maximum UsagePredicate: Labelled for 200 L maximum useNew Device: Labelled for 40 L maximum use
    Sterilization Assurance Level (SAL)Sterilized by Gamma radiation (10^-6) (Implied standard)Sterilized by Gamma radiation (10^-6)
    BiocompatibilityISO 10993-1:2018Compliant with ISO 10993-1:2018
    Packaging IntegrityISO 11607-1 & ISO 11607-2Compliant with ISO 11607-1, ISO 11607-2
    Particulate MatterUSP 788Compliant with USP 788
    General Infusion Equipment StandardsISO 8536-4:2019Compliant with ISO 8536-4:2019
    Small-bore Connectors StandardsISO 80369-7:2016Compliant with ISO 80369-7:2016
    Shipping PerformanceASTM D4169-16Compliant with ASTM D4169-16

    Note: The document states "Although the accuracy of the ProntoPump tube sets is not as precise as the predicate device, in the context of doses prepared in Hospital and Compounding Pharmacies the accuracy of the volume delivered to the final container is substantially equivalent in terms of safety and effectiveness." This indicates that the "reported device performance" for volume accuracy, while different, was deemed acceptable for substantial equivalence.


    2. Sample Size Used for the Test Set and the Data Provenance

    The document refers to "Bench Testing" against various ISO and ASTM standards. However, it does not provide any specific sample sizes for these tests. It also does not specify the data provenance (e.g., country of origin, retrospective or prospective) for the performance testing results, as these are laboratory/bench tests, not clinical studies involving patient data.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of those Experts

    N/A - This submission is for a medical device (sterile tube set) and relies on bench testing against established engineering and safety standards (ISO, ASTM, USP). There are no human experts involved in establishing a "ground truth" in the context of clinical interpretation or diagnostic accuracy for this type of device. The ground truth is defined by the objective metrics and criteria outlined in the referenced standards.


    4. Adjudication Method for the Test Set

    N/A - As this is bench testing against objective engineering and safety standards, no adjudication method (like 2+1, 3+1 consensus) is applicable or mentioned. The results are typically measured against the specified tolerances within the standards.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A - This document describes a syringe pump accessory (sterile tube set), not an AI-powered diagnostic or interpretive device. Therefore, no MRMC comparative effectiveness study was conducted or is relevant.


    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    N/A - This device is a physical medical device (tube set) used in conjunction with a pump, not an algorithm. Therefore, no standalone algorithm performance study was conducted.


    7. The Type of Ground Truth Used

    The "ground truth" for the device's performance is established by the requirements and specifications within recognized international standards for medical devices, particularly those related to infusion equipment, sterilization, packaging, and biocompatibility. These include:

    • ISO 8536-4:2019 (Infusion equipment)
    • ASTM D4169-16 (Shipping Containers)
    • ISO 80369-7:2016 (Small-bore connectors)
    • ISO 10993-1:2018 (Biocompatibility)
    • USP 788 (Particulate Matter)
    • ISO 11607-1:2019 & ISO 11607-2:2019 (Packaging)
    • ISO 11137-1:2006 & ISO 11137-2:2013 (Radiation Sterilization)
    • The established performance characteristics of the legally marketed predicate device (K062909) also serve as a comparative ground truth for evaluating substantial equivalence.

    8. The Sample Size for the Training Set

    N/A - This is a physical medical device, not a machine learning or AI algorithm. Therefore, there is no concept of a "training set" for this submission.


    9. How the Ground Truth for the Training Set Was Established

    N/A - As there is no training set for a physical device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163430

    Validate with FDA (Live)

    Date Cleared
    2017-05-23

    (167 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OTC Use: Prontosan® Wound Gel is intended to cleanse and moisten the wound bed and for the management of minor cuts, abrasions, lacerations, and minor burns.

    Professional Use: Prontosan® Wound Gel is intended to cleanse and moisten the wound bed and for the management of ulcers, 1st and 2nd degree burns, partial and full thickness wounds, and surgical incisions. It can be used during wound dressing changes to soften encrusted wound dressings.

    Device Description

    Prontosan® Wound Gel is a clear, colorless and virtually odorless gel containing undecylenamidopropyl betaine, polyaminopropyl biguanide, glycerol, hydroxyethylcellulose and purified water. Prontosan Wound Gel is a nonpyrogenic solution used for wound management.

    Prontosan Wound Gel will be offered in an over-the-counter (OTC) version and a professional use (Rx only) version. Prontosan Wound Gel is used to moisten the wound bed and clean the wound surface, including those that are difficult to access. Prontosan Wound Gel is aseptically filled into a 30 mL low density polyethylene squeeze bottle with a screw cap.

    AI/ML Overview

    The provided text is a 510(k) Substantial Equivalence Determination letter from the FDA for a wound gel called Prontosan Wound Gel. This document states that the device is "substantially equivalent" to a legally marketed predicate device (Prontosan Wound Gel, K101882) and does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria.

    The 510(k) process for this device, in particular, focuses on the manufacturer's claim that no changes have been made to the device design, packaging, manufacturing process, indications for use or the intended use. The submission is purely for labeling content changes.

    Therefore, I cannot provide the requested information from this document. The document explicitly states:

    • No new studies were conducted or required to demonstrate performance. The substantial equivalence is based on the device being identical to a previously cleared predicate device.
    • No acceptance criteria are mentioned because the device's performance is assumed to be equivalent to the predicate, which would have already met its own performance criteria when it was initially cleared.

    In summary, the provided document does not contain the information needed to answer your questions regarding acceptance criteria and a study proving the device meets those criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161623

    Validate with FDA (Live)

    Date Cleared
    2016-11-10

    (150 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Prontosan Wound Irrigation Solution is intended for cleaning wounds and for moistening and lubricating absorbent wound dressings for ulcers, burns, post-surgical wounds and abrasions.

    Device Description

    Wound Irrigation Solution is a clear. clear. colorless liquid containing Prontosan undecylenamidopropyl betaine, polyaminopropyl biguanide, sodium hydroxide and purified water. The solution is aseptically filled using a blow fill seal process into low density polyethylene 40 mL ampoules; 350 mL squeeze bottles and 1000 mL bottles with screw caps. Prontosan Wound Irrigation Solution is used for cleaning wounds and for moistening and lubricating absorbent wound dressings for ulcers, burns, post- surgical wounds and abrasions. The mechanical action of moving across the wound provides for the mechanism of action and aids in the removal of foreign material such as dirt and debris. The Prontosan Wound Irrigation Solution Adaptor is a component that may be screwed on to the current Prontosan Wound Irrigation Solution containers (350 ml and 1000 ml bottles) and provides alternate spout geometry for which to access the solution inside the bottle. It is provided in sterile form, labeled for single patient use only.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called "Prontosan Wound Irrigation Solution." This document pertains to regulatory approval rather than a scientific study on the device's performance against detailed acceptance criteria in a clinical or standalone setting.

    Therefore, the document does not contain the information requested in the prompt regarding acceptance criteria and a study proving the device meets those criteria.

    Specifically, the document states:

    • "The subject device and the predicate device are exactly the same. No changes have been made to the device, manufacturing process, the packaging process or the intended use. The modification subject to this 510(k) is for labeling content only." (Page 4)
    • "No new non-clinical testing has been performed for this submission. The changes to the labeling are test results from testing previously performed on this product and submitted to the FDA in a prior submission." (Page 4)

    This indicates that the current submission is not based on new performance data or a new study, but rather a labeling change for a substantially equivalent device. Consequently, none of the requested study details (sample size, ground truth, expert qualifications, MRMC, standalone performance, etc.) are present in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130857

    Validate with FDA (Live)

    Date Cleared
    2013-10-08

    (194 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rx: Prontosan® Wound Gel X is indicated for the management of ulcers (including diabetic foot and leg ulcers and pressure ulcers), 14 and 2nd degree burns, partial and full thickness wounds, large surface area wounds and surgical incisions.
    OTC: Prontosan Wound Gel X is indicated for the management minor cuts, minor lacerations, minor burns (131 degree burns), and abrasions.

    Device Description

    Prontosan Wound Gel X is a ready to use, clear, odorless, amorphous hydrogel wound dressing that helps maintain a clean, moist wound environment. It is intended as a barrier to resist microbial colonization within the dressing and reduce microbial penetration through the dressing. The gel matrix includes the preservative, polyhexanide, a viscosity modifying agent and a betaine surfactant. GelX is supplied sterile in blind ended, heat sealed polyfoil 250g tubes fitted with PP screw caps.

    AI/ML Overview

    The document describes the Prontosan® Wound Gel X, a wound dressing, and its substantial equivalence to predicate devices. It mentions that performance testing was conducted to support this claim.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test NameAcceptance CriteriaReported Device Performance
    BiocompatibilityIn accordance with ISO 10993-1Test results met the acceptance criteria.
    USP <51>Not explicitly stated, implied to be passedTest results met the acceptance criteria.
    Strike Through Barrier TestNot explicitly stated, implied to be passedTest results met the acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide specific details on the sample sizes used for the biocompatibility, USP <51>, or Strike Through Barrier tests. It also does not specify the provenance of the data (e.g., country of origin, retrospective or prospective).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not provided. The testing described appears to be laboratory-based performance and biocompatibility testing, not studies requiring expert interpretation of results to establish ground truth in the context of medical imaging or diagnostic devices.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided. The testing described does not involve adjudication as it's not a study where human readers or algorithms interpret data against a ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is mentioned. The document focuses on demonstrating substantial equivalence through biocompatibility and performance testing, not on comparing human reader performance with and without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This is not applicable as Prontosan® Wound Gel X is a wound dressing, not an algorithm or AI device. The studies mentioned are traditional medical device performance tests.

    7. Type of Ground Truth Used

    The ground truth for the performance tests (Biocompatibility, USP <51>, Strike Through Barrier Test) would be established by the defined standards and protocols of these specific tests. For example, for biocompatibility, the ground truth is adherence to the requirements of ISO 10993-1. For USP <51>, it likely refers to antimicrobial effectiveness testing against predefined microorganisms and reduction criteria.

    8. Sample Size for the Training Set

    This is not applicable and not provided. Prontosan® Wound Gel X is a medical device (wound dressing), not a machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reason as above; there is no training set for this type of medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110744

    Validate with FDA (Live)

    Date Cleared
    2011-04-06

    (20 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Prontosan® Wound Irrigation Solution is intended for cleaning wounds and moistening absorbent wound dressings for the management of minor cuts, abrasions, lacerations and minor burns.

    Device Description

    The subject device of this submission is Prontosan® Wound Irrigation Solution indicated for over-the-counter use. Prontosan Wound Irrigation Solution is currently cleared in a prescription use only version. Prontosan Wound Irrigation Solution is a clear, colorless and nearly odorless liquid intended for the management of wounds. The mechanical action of fluid moving across the wound provides the mechanism of action and aids in the removal of foreign material such as dirt and debris. The subject device is offered in 40 ml ampoules and 350 ml squeeze bottles with screw caps. The solution contains polyhexanide, a preservative that inhibits microbial growth within the product.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Prontosan® Wound Irrigation Solution:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Microbial Growth Inhibition (USP <51> Category 2)Demonstrated effectiveness to inhibit the growth of microorganisms within the product. (Specific results not quantified, but stated as effective for all tested organisms.)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes an Antimicrobial Effectiveness Testing study.

    • Test Set Organisms: The study used a standard panel of Gram-positive and Gram-negative bacteria, and fungi.
      • Bacteria: Staphylococcus aureus, Pseudomonas aeruginosa, Escherichia coli, methicillin-resistant Staphylococcus aureus, Vancomycin-resistant enterococcus, Serratia marcescens, Acinetobacter baumannii, Proteus mirabilis, Staphylococcus epidermidis, Enterococcus faecalis, Enterobacter cloacae.
      • Fungi: Candida albicans, Aspergillus brasilienis (niger).
    • Data Provenance: Not explicitly stated, but it's an in-vitro laboratory test conducted according to USP <51> standards. This is generally considered a controlled laboratory environment rather than patient data from a specific country, and it's a prospective test designed to evaluate the product's preservative efficacy.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    Not applicable. The study is a laboratory-based antimicrobial effectiveness test, not one requiring expert human interpretation of medical images or clinical outcomes. The "ground truth" is defined by the objective measurement of microbial growth inhibition according to USP <51> protocol.

    4. Adjudication Method for the Test Set:

    Not applicable. This was a laboratory test with objective measurements based on USP <51> methodology, not requiring adjudication of human interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. This submission is for a wound irrigation solution, not for an AI-powered diagnostic or interpretive device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    No. This concept is not applicable to a wound irrigation solution. The "performance" here refers to the chemical and preservative efficacy of the solution itself, not an algorithm.

    7. The Type of Ground Truth Used:

    The ground truth used was objective laboratory measurements based on a validated standard (USP <51>) for antimicrobial effectiveness. This involves inoculating the product with specific microorganisms and observing their growth or inhibition over time, rather than expert consensus, pathology, or outcomes data in a clinical setting.

    8. The Sample Size for the Training Set:

    Not applicable. This device is a wound irrigation solution, not an AI or machine learning model that requires a training set. The "training set" concept is irrelevant here.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this type of device.


    Summary of the Study:

    The study described is an Antimicrobial Effectiveness Testing conducted according to USP <51> category 2. This non-clinical test evaluated the ability of the Prontosan® Wound Irrigation Solution (intended for over-the-counter use) to inhibit the growth of common bacterial and fungal microorganisms within the product itself. The listed organisms represent typical pathogens found in wound beds. The study concluded that the device was effective in inhibiting microbial growth, thereby demonstrating its stability and preservative efficacy. This testing was performed to support the substantial equivalence claim, particularly regarding the solution's identity (formulation and processing) with its prescription-use predicate and its similarity in active ingredients to Prontosan® Wound Gel.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103405

    Validate with FDA (Live)

    Date Cleared
    2010-12-22

    (33 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pronto catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.

    Device Description

    The Pronto V4 extraction catheter (Pronto) is a dual lumen rapid exchange catheter that has a working length of approximately 138 cm and is packaged with related accessories. The smaller wire lumen of the catheter is able to accommodate guidewires ≤0.014 inches in diameter. The larger extraction lumen allows for the removal of thrombus by use of the included 30 mL syringes through the extension line and stopcock. The catheter has a stiff proximal region and a flexible distal region. The catheter has a rounded distal tip with a protected, sloped opening of the extraction lumen to facilitate advancement of the catheter into the blood vessel and maximize extraction of thrombus through the extraction lumen. The catheter contains a radiopaque marker band located approximately 2 mm from the distal tip. The distal 18 cm of the catheter has a hydrophilic coating to enhance deliverability to the target vasculature. The device has positioning marks located at 95 cm and 105 cm from the distal tip, respectively. The proximal end of the catheter incorporates a standard luer adapter to facilitate the attachment of catheter to the included extension line, stopcock, and 30 mL syringes. A filter basket is included for assistance in filtering the blood removed during the procedure for laboratory analysis of any thrombus.

    AI/ML Overview

    The provided text does not describe a study involving "acceptance criteria" and "reported device performance" in the typical sense of a clinical or comparative effectiveness study. Instead, it describes a medical device, the Pronto V4 extraction catheter, that has undergone verification testing against specified acceptance criteria to demonstrate substantial equivalence to a predicate device for FDA clearance.

    Therefore, many of the requested fields cannot be directly answered from the provided text because they pertain to clinical study methodologies that are not present.

    However, I can extract information related to the device verification and the "acceptance criteria" as described in the document.

    Here's a breakdown based on the information available:


    1. Table of acceptance criteria and the reported device performance

    The document states: "Results of the verification testing and biomaterial assessments met the specified acceptance criteria and did not raise new safety or performance questions."

    While specific numerical acceptance criteria (e.g., "tensile strength > X N") and the exact reported values for each test are not provided, the document lists the categories of tests performed. The "reported device performance" is implicitly stated as "met the specified acceptance criteria" for all tests.

    Acceptance Criteria CategoryReported Device Performance
    Simulated anatomy/concomitant device useMet specified acceptance criteria
    Hydrophilic coating integrity and particulatesMet specified acceptance criteria
    Thrombus aspirationMet specified acceptance criteria
    Kink resistanceMet specified acceptance criteria
    Air aspirationMet specified acceptance criteria
    TensileMet specified acceptance criteria
    TorqueMet specified acceptance criteria
    Dimensional verificationsMet specified acceptance criteria
    RadiopacityMet specified acceptance criteria
    Biocompatibility (Cytotoxicity, Sensitization, Irritation/intracutaneous reactivity, Acute systemic toxicity, Material mediated pyrogens, Hemocompatibility (Hemolysis, Coagulation, Prothrombin time, Hematological parameters, Complement activation, Thrombogenicity))Met specified acceptance criteria; no new safety/performance questions raised

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified for any of the verification tests.
    • Data Provenance: Not specified. These are laboratory/bench tests, not clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. The "ground truth" for these engineering/biomaterial tests would be the established scientific and engineering principles and standards against which the device performance is measured, not expert clinical consensus as might be the case for diagnostic imaging.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. This concept applies to human expert review in diagnostic assessments, not to bench testing of a device's physical properties or biocompatibility. The tests would be performed and evaluated according to predefined protocols.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This document describes the clearance of a mechanical medical device (extraction catheter), not an AI-powered diagnostic tool. Therefore, an MRMC study related to human readers and AI assistance is not relevant or described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device is a physical catheter, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" in this context refers to established engineering standards, material science principles, and biological safety guidelines (e.g., ISO standards for biocompatibility) that define acceptable performance for each test. For example, for "Tensile," the ground truth would be a minimum tensile strength required for safe operation, likely derived from clinical use cases and engineering specifications.

    8. The sample size for the training set

    • Not Applicable. This document describes verification testing for a physical medical device, not a machine learning model. There is no "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. See response to #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101882

    Validate with FDA (Live)

    Date Cleared
    2010-11-03

    (120 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OTC Use: Prontosan Wound Gel is intended to cleanse and moisten the wound bed and for the management of minor cuts, abrasion, laceration, and minor burns. Professional Use: Prontosan Wound Gel is intended to cleanse and moisten the wound bed and for the management of ulcers, 1st and 2nd degree burns, partial and full thickness wounds, and surgical incisions. It can be used during wound dressing changes to soften encrusted wound dressings.

    Device Description

    Prontosan® Wound Gel is a clear, colorless and virtually odorless gel containing undecylenamidopropyl betaine, polyaminopropyl biguanide, glycerol, hydroxyethylcellulose and purified water. Prontosan Wound Gel is a nonpyrogenic solution used for wound management that is sterile by aseptic manufacturing until the product is first opened. PHMB is added to the product in the concentration of 0.1% w/w as a preservative to inhibit the growth of microorganisms within the product. Prontosan Wound Gel will be offered in an over-the-counter (OTC) version and a professional use (Rx only) version. Prontosan Wound Gel is used to moisten the wound bed and clean the wound surface, including those that are difficult to access. Prontosan Wound Gel is aseptically filled into a 30 mL low density polyethylene squeeze bottle with a screw cap.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Prontosan® Wound Gel. The submission focuses on establishing substantial equivalence to previously cleared predicate devices, primarily through in-vitro testing for preservative effectiveness and by referencing prior biocompatibility and functional testing. It is not a study about an AI-powered device, medical imaging, or any diagnostic/prognostic tool that would typically involve acceptance criteria for algorithms, human readers, or ground truth establishment as outlined in your request.

    Therefore, most of the information requested in your prompt (e.g., acceptance criteria for a device's performance, sample sizes for test sets, expert qualifications, MRMC studies, standalone performance, training set details) is not applicable to this submission. This document pertains to a medical device (wound gel) and its regulatory clearance process, which relies on demonstrating safety and effectiveness, often through comparison to existing products, rather than validating an algorithm's performance against clinical endpoints or expert interpretations.

    Here's a breakdown of why each point is not fully addressable from the given text:

    1. A table of acceptance criteria and the reported device performance:

      • Not Applicable. The document describes "in-vitro testing conducted to demonstrate the effectiveness of the preservative in Prontosan® Wound Gel and substantial equivalence with predicate device, Anasept® Antimicrobial Skin and Wound Gel." However, specific acceptance criteria (e.g., percentage reduction in microbial growth, specific performance metrics) and detailed results are not provided in this 510(k) summary. The summary concludes that "The non-clinical testing conducted with the proposed device demonstrates that the proposed device is as safe and effective as the predicate device," implying that any set criteria were met.
    2. Sample sized used for the test set and the data provenance:

      • Not Applicable. This information is not provided. The testing mentioned is "in-vitro testing" for preservative effectiveness, not a clinical trial or a test of an algorithm on a dataset.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. There is no "ground truth" in the context of expert interpretation, as this is not a diagnostic device or an AI algorithm relying on human labels. The testing is laboratory-based (in-vitro).
    4. Adjudication method for the test set:

      • Not Applicable. No human interpretation or adjudication is described.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not Applicable. This is not an AI device, and no MRMC study is mentioned.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not Applicable. This is not an algorithm.
    7. The type of ground truth used:

      • Not Applicable. The "truth" in this context would be laboratory measurements of preservative effectiveness and material properties, compared against established standards or predicate device performance. Not "expert consensus," "pathology," or "outcomes data" in the typical sense of AI/diagnostic device validation.
    8. The sample size for the training set:

      • Not Applicable. This is not an AI device, so there is no training set.
    9. How the ground truth for the training set was established:

      • Not Applicable. There is no training set or associated ground truth.

    Summary based on the provided text:

    The device, Prontosan® Wound Gel, gained 510(k) clearance (K101882) based on demonstrating substantial equivalence to predicate devices (Prontosan® Wound Gel, K090141, and Anasept® Antimicrobial Skin and Wound Gel, K073547). The primary evidence cited for this submission is:

    • In-vitro testing conducted to prove the effectiveness of the preservative (0.1% w/w PHMB) in Prontosan® Wound Gel.
    • The fact that Prontosan® Wound Gel (K101882) is the same product as in the original Prontosan Wound Gel submission (K090141), with the only difference being expanded indications for use.
    • The biocompatibility and functional performance testing conducted on Prontosan Wound Gel in K090141 is considered applicable to K101882.

    The conclusion is that this non-clinical testing demonstrates the proposed device is as safe and effective as the predicate device, thereby meeting the requirements for 510(k) clearance for its stated Indications for Use.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083784

    Validate with FDA (Live)

    Date Cleared
    2009-03-30

    (101 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PRONTO extraction catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.

    The SKYWAY support catheter is intended to be used in conjunction with steerable guidewires in order to access discreet regions of the arterial and or coronary vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices. The SKY WAY OTW also may be used to subselectively infuse/deliver therapeutic agents.

    The TWIN-PASS dual access catheter is intended to be used in conjunction with steerable guidewires in order to access discrete regions of the coronary and peripheral arterial vasculature, to facilitate placement and exchange of guidewires and other interventional devices, for use during two guidewire procedures and to subselectively infuse/deliver diagnostic or therapeutic agents.

    Device Description

    The Pronto V3 extraction catheter is a dual lumen, hydrophilically coated, rapid exchange catheter with related accessories. The smaller wire lumen of the catheter is able to accommodate guidewires that are <= 0.014"/0.36mm in diameter. The larger extraction lumen allows for the removal of thrombus by use of the included syringe through the extension line and stopcock. The catheter has a rounded distal tip with a protected, sloped opening of the extraction lumen to facilitate advancement of the catheter into the blood vessel and maximize extraction of thrombus through the extraction lumen. The proximal end of the catheter has a standard luer adapter that attaches to the included extension line, stopcock and syringes. The catheter has a working length of 140cm and is compatible with standard 6F guide catheters. A 70um filter basket is included for assistance in filtering the blood removed during the procedure for laboratory analysis of any thrombosis.

    The Pronto LP extraction catheter is a dual lumen, hydrophilically coated rapid exchange catheter with related accessories. The smaller wire lumen of the catheter is able to accommodate guidewires that are <0.014"/0.36mm in diameter. The larger extraction lumen comes pre-loaded with a stylet that resists kinking during delivery but is removed to allow for the removal of thrombus by aspiration. The catheter has a rounded distal tip with a protected, sloped opening of the extraction lumen to facilitate advancement of the catheter into the blood vessel and maximize extraction of thrombus the extraction lumen. The proximal end of the catheter has a Yadapter hub containing a self-sealing stylet port, as well as a side-port for the attachment of the included extension line. The catheter has a working length of 137cm and is compatible with standard 6F guide catheters. A 70um filter basket is included for assistance in filtering the blood removed during the procedure for laboratory analysis of any thrombosis.

    The Skyway OTW (over-the-wire) and Skyway RX (rapid exchange) support catheters are single lumen, hydrophilically coated catheters designed for use in the arterial vasculature. The catheters provide support for 0.014"/0.36mm guidewires during interventional procedures, and allow for the exchange of one distally located guidewire for another one while maintaining access to distal vasculature. The Skyway RX has a rapid exchange port that allows for the insertion of the catheter over a short guidewire; a stiffening mandrel provides support and pushability during catheter insertion.

    The Twin-Pass dual access catheter is a hydrophilically coated, dual lumen catheter designed for use in the arterial vasculature. The catheter provides support for 0.014"/0.36mm guidewires during interventional procedures, and the dual lumen design allows for the delivery of a second guidewire into distal vasculature while leaving the initial guidewire in place. A stiffening mandrel provides support and pushability during catheter insertion.

    AI/ML Overview

    I am sorry, but the provided text is a 510(k) premarket notification for several medical devices (catheters) and a subsequent administrative correction letter from the FDA. It does not contain information about explicit acceptance criteria or a study that proves a device meets such criteria.

    The document establishes that new versions of the Pronto V3 and LP extraction catheters, Skyway RX and OTW support catheters, and Twin-Pass dual access catheter are substantially equivalent to their predicate devices, primarily due to a hydrophilic coating modification.

    The "Summary of Studies" section explicitly states:

    "Pronto, Skyway and Twin-Pass catheters are substantially equivalent in intended use / indications and design to their predicate devices. Hydrophilic coating design differences have been qualified through biomaterial assessments and verification testing, the results of which did not raise any new safety or performance questions."

    This indicates that the submission relies on bench testing (biomaterial assessments and verification testing) to demonstrate that the modified devices perform similarly to their previously cleared predicate devices, rather than a clinical study with defined acceptance criteria for a specific performance metric.

    Therefore, I cannot provide the requested information, such as a table of acceptance criteria, sample sizes for test/training sets, expert ground truth establishment, or details about comparative effectiveness studies, as these are not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090141

    Validate with FDA (Live)

    Date Cleared
    2009-03-17

    (55 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Prontosan™ Wound Gel is intended for moistening and cleansing wounds by maceration of wound coatings and for moistening and lubricating absorbent wound dressings for ulcers, burns, post-surgical wounds, and abrasions.

    Device Description

    Prontosan™ Wound Gel is a clear, colourless liquid-gel containing undecylenamidopropyl betaine, polyaminopropyl biguanide, glycerol, hydroxyethylcellulose and purified water. Prontosan™ Wound Gel is used for managing and treating wounds by sustaining wound moisture between dressing changes and aiding in the manual removal of wound exudates and encrustations. The gel is aseptically filled using a blow fill seal process into low density polyethylene 30 mL squeeze bottles with screw caps.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device (Prontosan™ Wound Gel) and does not describe acceptance criteria or a study that rigorously tests the device against such criteria in the way a clinical performance study for an AI/ML device would. Instead, it demonstrates substantial equivalence to a predicate device based on similar intended use, technological characteristics, and safety testing.

    Therefore, most of the requested information regarding acceptance criteria, performance studies, sample sizes, expert ground truth, adjudication methods, and MRMC studies is not applicable to this submission.

    However, I can extract the relevant information regarding the stated safety and performance support for the device:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Test/DocumentationReported Performance
    BiocompatibilityBiocompatibility testingSupported
    Sterility/PurityUSP <51> Antimicrobial Effectiveness TestingSupported (Implied to meet USP standards)
    USP <85> Bacterial Endotoxin testingSupported (Implied to meet USP standards)
    Material/Product StabilityShelf life testingSupported
    Clinical Performance(Not applicable - substantial equivalence based on predicate, not clinical trials for performance metrics)(Not applicable)

    Notes on the "Reported Performance": The document states that these tests were "supported by" or "conducted with" Prontosan™ Wound Gel, implying successful completion and meeting of relevant standards to demonstrate safety and effectiveness.

    2. Sample size used for the test set and the data provenance:

    • Not applicable. This submission relies on substantial equivalence to a predicate device and direct safety testing, not a clinical "test set" in the context of an AI/ML or diagnostic performance study.
    • The provenance for the safety tests (biocompatibility, USP <51>, USP <85>, shelf life) is the product itself, as tested by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth in the context of expert review for a test set is not relevant to this type of device submission. The "ground truth" for safety aspects would be established by the respective test methodologies and their validated standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. There was no "test set" and no expert adjudication process described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a wound gel, not an AI/ML-powered diagnostic or assistive tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a wound gel, not an algorithm.

    7. The type of ground truth used:

    • For safety: The "ground truth" was established by the outcomes of standard, validated laboratory tests for biocompatibility, antimicrobial effectiveness (per USP <51>), bacterial endotoxin (per USP <85>), and product shelf life. These are objective measures with established pass/fail criteria.
    • For effectiveness/performance: The "ground truth" for effectiveness is primarily established by substantial equivalence to the predicate device, Prontosan™ Wound Irrigation Solution, which implies similar clinical utility for wound moistening and cleansing based on its known formulation and the presence of minor inactive ingredients (hydroxyethylcellulose and glycerol) that add viscosity and moisturizing properties. No new clinical performance studies proving efficacy were presented, as per the 510(k) pathway.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • Not applicable. Since there is no training set, this question is not relevant.

    Summary of Device Substantiation:

    The B. Braun Medical, Inc. Prontosan™ Wound Gel received 510(k) clearance based on demonstrating substantial equivalence to a legally marketed predicate device (Prontosan™ Wound Irrigation Solution). The primary evidence for this was:

    • Similar Intended Use: Both devices are for moistening and cleansing wounds.
    • Similar Technological Characteristics: The core active ingredients and mechanism of action are similar. The gel formulation adds inactive ingredients (hydroxyethylcellulose and glycerol) for viscosity and moisturizing, which are described as not raising new questions of safety or effectiveness.
    • Safety Testing: Biocompatibility testing, USP <51> Antimicrobial Effectiveness Testing, USP <85> Bacterial Endotoxin Testing, and shelf life testing were conducted to demonstrate the safety of the specific gel formulation. The document states these tests "supported" the safety and effectiveness, implying acceptable results.

    The 510(k) pathway (especially for Class II devices without special controls or Class I unclassified devices like this one) does not typically require extensive clinical trials or performance studies that would utilize concepts like "acceptance criteria" against clinical performance metrics, expert adjudication, MRMC studies, or AI training/test sets. The focus is on demonstrating that the new device is as safe and effective as a legally marketed predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072876

    Validate with FDA (Live)

    Date Cleared
    2008-06-19

    (254 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Prontosan Wound Irrigation Solution is intended for cleaning wounds and for moistening and lubricating absorbent wound dressings for ulcers, burns, post-surgical wounds and abrasions.

    Device Description

    Prontosan Wound Irrigation Solution is a clear, colorless containing liauid undecylenamidopropyl betaine. polyaminopropyl biguanide, sodium hydroxide and purified water. The solution is aseptically filled using a blow fill seal process into low density polyethylene 40 mL ampoules and 350 mL squeeze bottles with screw caps. Prontosan Wound Irrigation Solution is used for cleaning wounds and for moistening and lubricating absorbent wound dressings for ulcers, burns, post-surgical wounds and abrasions. The mechanical action of moving across the wound provides for the mechanism of action and aids in the removal of foreign material such as dirt and debris.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Prontosan™ Wound Irrigation Solution). It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain information about a study proving the device meets acceptance criteria in the context of performance metrics like sensitivity, specificity, accuracy, or expert agreement that would typically be found in an AI/software as a medical device (SaMD) submission.

    This document focuses on demonstrating substantial equivalence to an existing predicate device based on:

    1. Similar Indications for Use: The new device and the predicate device (Dermacyn® Wound Care solution) are both intended for cleaning and irrigating wounds and moistening wound dressings.
    2. Biocompatibility Testing: To ensure the materials used are safe for bodily contact.
    3. Functional Performance Testing: To verify the solution performs its intended physical functions.
    4. Shelf Life Testing: To ensure stability over time.
    5. Comparison Testing: Directly comparing the proposed device with the predicate device to show similarities and confirm no new safety/effectiveness issues.

    Therefore, for the type of information requested (acceptance criteria and study details related to device performance metrics, human-in-the-loop, ground truth, etc., typically found in SaMD studies), the provided text does not contain this information.

    Here's a breakdown of why each requested point cannot be addressed from the given text:

    1. A table of acceptance criteria and the reported device performance: This document doesn't provide specific performance metrics or acceptance criteria in the way an AI/software device would (e.g., sensitivity, specificity thresholds). It only states that comparison testing "demonstrated that Prontosan Wound Irrigation Solution, the proposed device is similar to Dermacyn Wound Care solution, the predicate device."
    2. Sample sized used for the test set and the data provenance: Not applicable or provided. The "testing" mentioned is likely laboratory or chemical analysis, not clinical trials with sample sizes in the context of disease detection or diagnosis.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a wound irrigation solution would relate to its chemical composition, sterility, physical properties, and safety, not expert interpretation of images or clinical data.
    4. Adjudication method: Not applicable.
    5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable. This type of study demonstrates improved human performance with AI assistance, which is irrelevant for a wound irrigation solution.
    6. Standalone (i.e. algorithm only without human-in-the-loop performance) study: Not applicable. This device is a solution, not an algorithm.
    7. Type of ground truth used: For this type of device, the "ground truth" would be established by rigorous chemical analysis, sterility testing, and biocompatibility standards, not expert consensus or pathology in a diagnostic sense.
    8. Sample size for the training set: Not applicable. There is no AI model being trained.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document is a 510(k) summary for a physical wound irrigation solution, not an AI/software medical device. As such, it relies on different types of evidence (biocompatibility, functional testing, comparison to predicate) to establish substantial equivalence, rather than the performance metrics and study designs typically associated with AI/SaMD devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2