Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K123549

    Validate with FDA (Live)

    Date Cleared
    2013-02-25

    (98 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft and/or allograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies). spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, or failed previous fusion.

    The Ballista/Ballista II instruments are intended to be used with Ballista II/Polaris 5.5mm implants. Cannulated screws and percutaneous rods may be used with the Ballista/Ballista II instruments to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and/or lordosis), tumor, stenosis, pseudoarthrosis, or failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft and/or allograft.

    The Polaris Spinal System may be used with the instruments in the AccuVision Minimally Invasive Spinal Exposure System to provide the surgeon with a minimally invasive approach for posterior spinal surgery.

    The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, the Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.

    Device Description

    The Polaris Spinal System is a non-cervical spinal fixation device. The system includes screws, various types and sizes of rods, locking nuts, hooks, lateral connectors, plugs, fixation washers, rod connectors/dominos and various cross connectors. Various instruments are also available for use by the surgeon to facilitate implantation of the device. This submission is a line extension to Polaris Spinal System to add other styles of multiaxial screw, including a screw with translating feature, and to provide the Polaris Spinal System sterile packed.

    AI/ML Overview

    The provided document describes a 510(k) summary for the Biomet Spine Polaris Spinal System - Translation Screw, which is a non-cervical spinal fixation device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Mechanical Testing:
    Static compression bending per ASTM 1717-11aMet all mechanical test requirements based on worst-case construct testing.
    Static torsion per ASTM 1717-11aMet all mechanical test requirements based on worst-case construct testing.
    Dynamic compression bending fatigue per ASTM 1717-11aMet all mechanical test requirements based on worst-case construct testing.
    Substantial Equivalence:Reported as substantially equivalent to predicate devices regarding material, design, operational principles, intended use, indications, and fundamental technology.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of clinical data for algorithmic performance. The testing performed was mechanical testing of the device hardware.

    • Test Set Sample Size: Not applicable in the context of clinical data or algorithm performance.
    • Data Provenance: Not applicable. The "data" comes from physical mechanical tests conducted on the device components.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. Ground truth, in the context of algorithm performance, typically refers to expert-validated clinical diagnoses or outcomes. This submission focuses on mechanical performance data of a physical medical device.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods are relevant for resolving discrepancies in expert opinions for clinical ground truth. Mechanical testing has well-defined pass/fail criteria per ASTM standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No, an MRMC comparative effectiveness study was not done. This device is a physical spinal fixation system, not an AI or imaging diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone algorithm performance study was not done. This submission is for a physical medical device.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The "ground truth" for the mechanical testing was the established pass/fail criteria defined by ASTM 1717-11a standard test methods for static and fatigue testing of spinal implant constructs in a vertebrectomy model.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical implant, not an AI algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1