Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K180099

    Validate with FDA (Live)

    Date Cleared
    2018-10-12

    (269 days)

    Product Code
    Regulation Number
    866.5270
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Optilite High Sensitivity C-Reactive Protein Kit is intended for the quantitative in vitro measurement of C-Reactive Protein in serum using the Binding Site Optilite analyser for evaluation of conditions thought to be associated with inflammation, in otherwise healthy individuals. This test should be used in conjunction with other laboratory and clinical findings.

    Device Description

    The Optilite High Sensitivity C-Reactive Protein Kit is comprised of latex reagent, single calibrator, controls (high and low) and reaction buffer in liquid form. The reagents contain 0.099% sodium azide as preservative.

    AI/ML Overview

    Acceptance Criteria and Study for Optilite High Sensitivity C-Reactive Protein Kit

    Here's a breakdown of the acceptance criteria and the study results for the Optilite High Sensitivity C-Reactive Protein Kit, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance CharacteristicAcceptance Criteria (Implicit from Study Design)Reported Device Performance (Optilite Kit)
    PrecisionTotal Precision: %CV < 10%Sample 1 (0.98 mg/L): 5.6%
    Within-Run Precision: %CV < 5%Sample 2 (1.55 mg/L): 5.0%
    Between-Run Precision: %CV < 8%Sample 3 (2.99 mg/L): 4.5%
    Between-Day Precision: %CV < 8%Sample 4 (5.41 mg/L): 4.4%
    Sample 5 (8.50 mg/L): 3.1%
    Individual Within-Run, Between-Run, and Between-Day %CVs were all within their respective implicit criteria (e.g., Within-Run %CVs ranged from 1.5% to 2.9%).
    Linearity/Reportable RangeDeviation from linearity ≤ 10%Linear over 0.44 - 12.16 mg/L with deviation from linearity ≤ 10%
    TraceabilityTraceable to an international reference standardERM-DA474
    Kit Stability (Shelf-life)Sufficient to support a claim (e.g., 12+ months)Currently supports a 13-month stability claim (on-going study)
    Open-vial StabilitySufficient for practical useUp to 3 months at 2-8°C
    On-board StabilitySufficient for practical useAt least 30 days on Optilite Analyser
    Detection Limit (LoQ)Clinically relevant lower limit of quantification0.5 mg/L
    Analytical Specificity (Interference)Mean results from spiked samples within 10% of control samplesNo significant interference from hemoglobin (5g/L), conjugated bilirubin (200mg/L), unconjugated bilirubin (200mg/L), rheumatoid factor (25lU/mL), triglyceride (1000mg/dL), intralipid (2000mg/dL), and 6 common therapeutic drugs.
    Method Comparison with PredicateRegression analysis (e.g., Passing & Bablok slope and intercept) indicating substantial agreement with predicate device (implicit acceptance of close to 1 for slope and close to 0 for intercept with narrow CIs)Passing & Bablok: y = 0.99x + 0.09 (95% CI Slope: 0.97 to 1.01, 95% CI Intercept: 0.05 to 0.11) - PASS
    Weighted Deming: y = 0.97x + 0.10 (95% CI Slope: 0.94 to 0.99, 95% CI Intercept: 0.07 to 0.13)
    Weighted Linear: y = 0.95x + 0.12 (95% CI Slope: 0.93 to 0.98, 95% CI Intercept: 0.10 to 0.15)
    Expected Values/Reference Range VerificationVerification of stated reference intervalVerified by testing 50 adult donor samples; Consensus reference interval for adults: <3 mg/L

    2. Sample Size Used for the Test Set and Data Provenance:

    • Precision Study: 5 serum samples (targeted at specific concentrations: 1mg/L, 1-3mg/L, 3mg/L, 5mg/L, and 9mg/L).
    • Linearity Study: Serially diluted serum samples (number not explicitly stated, but sufficient to cover 0.5 – 10.0 mg/L).
    • Detection Limit (LoB, LoD, LoQ) Studies:
      • LoB: 4 serum samples
      • LoD: 4 serum samples
      • LoQ: 4 serum samples
    • Analytical Specificity (Interference) Study: Serum samples with CRP concentrations targeted at <1mg/L, 1-3mg/L, 3mg/L, and >3mg/L.
    • Method Comparison with Predicate Device: 391 serum samples (302 reported results within the measuring range). These included 87 normal donors and 304 clinical samples.
    • Expected Values/Reference Range Verification: 50 adult donor samples.

    Data Provenance: The document does not explicitly state the country of origin for the data or whether it was retrospective or prospective. Given "The Binding Site Group, Ltd." is based in the UK, it's highly probable that the studies were conducted there. The nature of the studies (e.g., precision, linearity, method comparison) generally implies prospective testing of samples for analytical performance, though the 304 "clinical samples" in the method comparison could be retrospective or a mix.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. This device is an in vitro diagnostic (IVD) for quantitative measurement of a biomarker (C-Reactive Protein). The "ground truth" for its performance is established through comparisons with reference methods, calibrated materials, and statistical analysis of assay performance (precision, linearity, detection limits), rather than through expert consensus on qualitative interpretation.

    4. Adjudication Method for the Test Set:

    Not applicable, as this is a quantitative IVD device. Test results are numerical and assessed against predefined analytical performance criteria, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI-powered diagnostic imaging devices where human readers interpret images. The Optilite Kit is an automated immunoassay system, not an imaging device requiring human interpretation.

    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    Yes, a standalone study of the algorithm (the Optilite High Sensitivity C-Reactive Protein Kit and Optilite analyser) was performed. All the analytical performance characteristics (precision, linearity, detection limits, analytical specificity) and method comparison studies described are standalone evaluations of the device's performance without human intervention in the result generation process.

    7. The Type of Ground Truth Used:

    The ground truth for this device is established through:

    • International Reference Standards: The assay is traceable to the international reference standard ERM-DA474 for calibration.
    • Statistical Definitions: For precision (CV%), linearity (deviation from linearity), detection limits (LoB, LoD, LoQ), established statistical methodologies (CLSI EP5-A3, EP6-A, EP17-A2) define the acceptable "ground truth" or performance targets.
    • Reference Methods/Predicate Device: For method comparison, the results from the legally marketed predicate device (C-Reactive Protein High Sensitive Test System for Cobas Integra Instruments) serve as the comparative "ground truth" for demonstrating substantial equivalence.
    • Clinical Relevance: The reference interval (<3 mg/L) is derived from consensus and published literature (Macy et al., 1997), serving as a clinically accepted range.

    8. The Sample Size for the Training Set:

    Not applicable in the context of this device. This is a laboratory diagnostic kit, not an AI or machine learning algorithm that requires a "training set" in the conventional sense. The "training" of the device involves its design, optimization, and calibration using standardized materials and methods.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. As mentioned above, there isn't a "training set" for this type of device in the AI sense. The development and optimization of the reagent and instrument involved:

    • Chemical and Biological Principles: Based on well-established immunoturbidimetry principles and antibody-antigen reactions.
    • Calibration: Using a single calibrator diluted on the analyser, traceable to the international reference standard ERM-DA474, to establish the relationship between optical signal and CRP concentration.
    • Reagent Formulation and Instrument Design: Iterative development and testing of reagent components (latex reagent, calibrator, controls, buffer) and instrument parameters to achieve desired analytical performance characteristics.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1