Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K163402

    Validate with FDA (Live)

    Device Name
    NLZ Motor System
    Manufacturer
    Date Cleared
    2017-08-31

    (269 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NLZ Motor System is intended for use by dental professionals in the performance of dental restoration, prophylaxis and endodontic procedures.

    The NLZ Endo is intended for use by dental professionals in the performance of dental endodontic procedures.

    Device Description

    NLZ Motor Systems are available in two models;

    NLZ Motor Set, which consists of the Main Unit, the Control Unit and the Motor.

    NLZ E Motor Set, which consists of the Main Unit, the Control Unit, the Motor, and the Contra Angle Handpiece: NLZ Endo. The NLZ Endo is available only for NLZ E Motor Set and is not avialable for the NLZ Motor Set. The NLZ Endo is not equipped with any cooling system because it is a speed-reducing/low-speed handpiece.

    The Control Unit drives an electric micromotor, and turns on off or regulates the speed of an electric motor by its foot pedal of the dental unit. This product can be connected to dental units currently in use to add on a brushless electric micromotor with an LED light.

    AI/ML Overview

    The provided text describes Nakanishi Inc.'s NLZ Motor System and its substantial equivalence to predicate devices, focusing on various performance aspects. However, the document does not present acceptance criteria in a table format with specific quantitative targets and corresponding device performance measurements. Instead, it discusses compliance with general standards and the satisfactory outcome of various tests.

    Here's an analysis based on the available information, addressing as many of your points as possible:

    1. Table of acceptance criteria and the reported device performance:

    The document does not explicitly provide a table of acceptance criteria with numerical performance targets and reported values for the NLZ Motor System. Instead, it states that the device was compliant with various ISO and IEC standards. The "performance" is generally described as "met all requirements" or "do not raise concern."

    However, specific comparisons are made with predicate devices, which can infer performance expectations. For instance:

    FeatureAcceptance Criteria (Inferred from standards/predicates)Reported Device Performance (NLZ Motor System)
    BiocompatibilityConformity to ANSI/AAMI/ISO 10993-5:2009, ISO 10993-10:2010No biocompatibility concern raised
    Mechanical PerformanceConformity to ISO 14457:2012All samples compliant with ISO 14457:2012
    Electrical SafetyConformity to AAMI / ANSI/ES60601-1:2005/(R)2012 and A1:2012,, C1:2009/(R)2012 and A2:2010/(R)2012Test results indicate no concern
    Electromagnetic CompatibilityConformity to IEC 60601-1-2:2014Test results indicate no concern
    Overheating Risk ReductionProper Contra-Check judge and motor speed control for Contra-Restriction and Contra-SAFE. Reduced overheating/burn events.Verification test confirmed proper function and no new concern. Analyzed temperature increase for returned handpieces to confirm reduction of events.
    Reprocessing ValidationConformity to AAMI TIR 12:2010, AAMI TIR 30:2011, and FDA Reprocessing guidanceMet all requirements of the standards
    Range of Rotation Speed100 - 40,000 rpm (Similar to predicate)100 - 40,000 rpm
    Rotating DirectionForward and Reverse (Similar to predicate)Forward and Reverse
    Light Intensity>10,000 lx (for LED light)>10,000 lx (for LED light)
    NLZ Endo EccentricityNot explicitly stated for predicate in table, but performance compliant with ISO 14457.0.018 mm
    NLZ Endo Bur ExtractionNot explicitly stated for predicate in table, but performance compliant with ISO 14457.>45N
    NLZ Endo Max Bur Length25 mm (Predicate)46 mm

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not specified. The document mentions "All samples were compliant with ISO 14457:2012" and "This testing is verification test for software by putting excessive load to the Motor." However, the exact number of devices tested for each criterion is not provided.
    • Data provenance: Bench testing was performed internally by Nakanishi Inc. ("in-house standard," "bench testing was performed in order to verify conformity," "reprocessing validation was performed"). The document does not indicate the country of origin of the data beyond the company's location (Japan) and general compliance with international standards (ISO, IEC, AAMI, ANSI). The studies were retrospective as they describe testing conducted before submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The testing described is primarily technical bench testing and compliance with standards, not studies involving human experts establishing ground truth for diagnostic or clinical accuracy.

    4. Adjudication method for the test set:

    This information is not provided as the studies are technical performance and compliance tests rather than clinical evaluations requiring adjudication of subjective outcomes.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC comparative effectiveness study was done or is mentioned. This device is a dental motor system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device does not involve an algorithm for diagnostic or clinical interpretation that would operate in a standalone manner. The "overheating risk reduction functions" are automated features within the device, whose performance was verified through bench testing.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the performance testing, the "ground truth" was defined by:

    • International standards: ISO 14457, ANSI/AAMI/ISO 10993, IEC 60601-1-2, AAMI / ANSI/ES60601-1, AAMI TIR 12, AAMI TIR 30, and FDA Reprocessing guidance.
    • In-house standards: For the overheating risk reduction features.
    • Predicate devices: The performance characteristics of the NLZ Motor System were compared to those of the predicate devices for establishing substantial equivalence.

    8. The sample size for the training set:

    Not applicable. This device does not utilize a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established:

    Not applicable. As above, there is no mention of a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1