Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K170983

    Validate with FDA (Live)

    Device Name
    NIOX VERO
    Manufacturer
    Date Cleared
    2017-11-22

    (233 days)

    Product Code
    Regulation Number
    862.3080
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NIOX VERO® measures Nitric Oxide (NO) in human breath. Nitric Oxide is frequently increased in some airway inflammatory processes such as asthma. The fractional NO concentration in expired breath (FeNO), can be measured by NIOX VERO according to guidelines for NO measurement established by the American Thoracic Society.

    Measurement of FeNO by NIOX VERO is a quantitative, non-invasive, simple and safe method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy, as an indication of the therapeutic effect in patients with elevated FeNO levels. NIOX VERO is suitable for children, 7- 17 years, and adults 18 years and older.

    NIOX VERO 10 second test mode is for age 7 and up
    NIOX VERO 6 second test mode is for ages 7-10 only who cannot successfully complete a 10 second test.

    FeNO measurements provide the physician with means of evaluating an asthma patient's response to anti-inflammatory therapy, as an adjunct to the established clinical and laboratory assessments in asthma. The NIOX VERO is intended for prescription use and should only be used as directed in the NIOX VERO User Manual by trained healthcare professionals. NIOX VERO cannot be used with infants or by children under the age of 7 as measurement requires patient cooperation.

    NIOX VERO should not be used in critical care, emergency care or in anesthesiology.

    Device Description

    NIOX VERO is a portable system for the non-invasive, quantitative measurement of the fraction of exhaled nitric oxide (NO) in expired human breath (FeNO). NO levels increase during allergic airway inflammation. Measurement of changes in FeNO concentration is used in evaluating a patient's response to antiinflammatory therapy, as an adjunct to established clinical and laboratory assessments.

    The NIOX VERO system is comprised of the NIOX VERO unit with AC adapter, a rechargeable battery, an electrochemical NO sensor, disposable patient filters, and an exchangeable handle containing an internal NO scrubber filter. The NIOX Panel is an optional PC application for operation of the NIOX VERO from a PC and access to electronic medical record systems. The device can connect to the PC via a standard USB cable or wirelessly via Bluetooth.

    The patient empties their lungs, inhales deeply through the patient filter to total lung capacity and then slowly exhales for 10 seconds. A 6 second mode is available for children aged 7 – 10 who cannot perform a 10 second exhalation. In approximately one minute, the NO concentration is displayed in parts per billion (ppb).

    The NIOX VERO unit includes a sampling and gas conditioning system and a man-machine interface (MMI). The instrument controls the inhaled sample appropriately via the electronics and software program. Filtering of inhaled air elimination from ambient NO levels. A built-in flow control keeps exhalation standardized at 50 ml/s so that it is standardized for all patients. The sample enters an electromechanical sensor and interacts with an electrolyte. A chemical reaction takes place where electrons are generated proportional to the number of NO molecules.

    Results are processed using dedicated software. In order to verify the device's performance and reliability of measurements, there are built-in system control procedures and a Quality Control procedure to be performed on a daily basis.

    AI/ML Overview

    The Circassia NIOX VERO device measures the fractional nitric oxide concentration in expired breath (FeNO) to evaluate an asthma patient's response to anti-inflammatory therapy. The 510(k) submission (K170983) is specifically for the activation of an existing 6-second measurement mode for children aged 7-10, in addition to the previously cleared 10-second mode.

    Here's a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the NIOX VERO are compared against the predicate device (NIOX VERO, K150233), and for the 6-second mode, the performance was found to be identical to these established criteria.

    Acceptance CriteriaSpecified by Predicate (K150233)Reported Device Performance (NIOX VERO 6s mode)
    Analytical limits (low levels, limit of detection)5 ppbIdentical to predicate (5 ppb)
    Precision< 3 ppb for values < 30 ppb; < 10% for values ≥ 30 ppbIdentical to predicate
    Accuracy±5 ppb for values ≤ 30 ppb; 10% of measured value for values > 30 ppbIdentical to predicate
    Measurement Range5 - 300 ppbIdentical to predicate
    Linearity, reportable rangeSquared correlation coefficient $r^2$ ≥ 0.998, slope 0.95 - 1.05, intercept ±3 ppbIdentical to predicate
    Measurement Mode10s mode for patients 7 to 17 and adults 18+10s mode (7 to 17, adults 18+); 6s mode (7 to 10 only)

    2. Sample Size and Data Provenance

    • Test Set (Clinical Study AER-047):
      • Sample Size: 43 male and female subjects.
      • Data Provenance: Not explicitly stated, but the study was described as a "single-center" study. This suggests the data originated from one specific location. It was a "single-visit, point-of-care clinical validation study," indicating it was a prospective study.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    This information is not provided in the document. The study focuses on comparing the 6s mode against the 10s mode in children for their ability to successfully perform the test and maintain similar analytical performance characteristics, rather than expert-adjudicated ground truth for a diagnostic outcome.

    4. Adjudication Method (Test Set)

    This information is not provided. The clinical study described in the document is a validation study demonstrating the feasibility and performance equivalence of the 6s mode. It does not appear to involve a diagnostic outcome requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC study was not conducted as this device measures a quantitative physiological biomarker (FeNO). The study performed was a clinical validation to demonstrate the equivalence of the 6s measurement mode. The device does not involve human readers interpreting images, so the concept of "human readers improve with AI vs without AI" is not applicable.

    6. Standalone Performance (Algorithm Only)

    Yes, a standalone (algorithm only) performance assessment was done. The performance bench testing (precision, accuracy, linearity) of the NIOX VERO in 6s mode was conducted independently and demonstrated to be within the specifications of the original NIOX VERO performance. This refers to the device's inherent ability to accurately measure FeNO, irrespective of human interaction beyond operating the device as intended.

    7. Type of Ground Truth Used

    • Clinical Study (AER-047): The "ground truth" for the clinical study was the successful completion and measurement of FeNO for the 6-second mode in children aged 7-10, and the comparison of these measurements to the established performance characteristics of the 10-second mode. It was about patient cooperation and the device's ability to consistently provide measurements in this cohort.
    • Bench Testing: The ground truth for bench testing (precision, accuracy, linearity) was established by using certified calibration gas of known concentrations (5ppb, 25ppb, 75ppb, and 200ppb) as reference standards.

    8. Sample Size for the Training Set

    This information is not applicable and therefore not provided. The NIOX VERO is a measurement device, not an AI/ML-driven diagnostic algorithm that requires a training set in the conventional sense. The "6s measurement mode" was an existing functionality in the predicate device, not a newly developed algorithm requiring extensive training data.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no mention of a training set for an AI/ML algorithm. The device measures a physical parameter using electrochemical sensing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150233

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-02-26

    (24 days)

    Product Code
    Regulation Number
    862.3080
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NIOX VERO® measures Nitric Oxide (NO) in human breath. Nitric Oxide is frequently increased in some inflammatory processes such as asthma. The fractional NO concentration in expired breath (FeNO), can be measured by NIOX VERO according to guidelines for NO measurement established by the American Thoracic Society.

    Measurement of FeNO by NIOX VERO is a quantitative, non-invasive, simple and safe method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy, as an indication of the therapeutic effect in patients with elevated FeNO levels. NIOX VERO is suitable for children, approximately 7 - 17 years, and adults 18 years and older.

    FeNO measurements provide the physician with means of evaluating an asthma patient's response to anti-inflammatory therapy, as an adjunct to the established clinical and laboratory assessments in asthma. The NIOX VERO is intended for prescription use and should only be used as directed in the NIOX VERO User Manual by trained healthcare professionals. NIOX VERO cannot be used with infants or by children approximately under the age of 7, as measurement requires patient cooperation. NIOX VERO should not be used in critical care, emergency care or in anesthesiology.

    Device Description

    NIOX VERO is a portable system for the non-invasive, quantitative measurement of the fraction of exhaled nitric oxide (NO) in expired human breath (FeNO). NO levels increase during allergic airway inflammation. Measurement of changes in FeNO concentration is used in evaluating a patient's response to antiinflammatory therapy, as an adjunct to established clinical and laboratory assessments.

    The NIOX VERO system is comprised of the NIOX VERO unit with AC adapter, a rechargeable battery, an electrochemical NO sensor, disposable patient filters, and an exchangeable handle containing an internal NO scrubber filter. The NIOX Panel is an optional PC application for operation of the NIOX VERO from a PC and access to electronic medical record systems. The device can connect to the PC via a standard USB cable or wirelessly via Bluetooth.

    The patient empties their lungs, inhales deeply through the patient filter to total lung capacity and then slowly exhale for 10 seconds. In approximately one minute, the NO concentration is displayed in parts per billion (ppb).

    The NIOX VERO unit includes a sampling and gas conditioning system and a man-machine interface (MMI). The instrument controls the inhaled sample appropriately via the electronics and software program. Filtering of inhaled air elimination from ambient NO levels. A built-in flow control keeps exhalation standardized at 50 ml/s so that it is standardized for all patients. The sample enters an electromechanical sensor and interacts with an electrolyte. A chemical reaction takes place where electrons are generated proportional to the number of NO molecules.

    Results are processed using dedicated software. In order to verify the device's performance and reliability of measurements, there are built-in system control procedures and a Quality Control procedure to be performed on a daily basis.

    AI/ML Overview

    The provided text describes a Special 510(k) summary for the NIOX® VERO Airway Inflammation Monitor (K150233). This submission is specifically for the activation of existing Bluetooth technology on the predicate device (K133898). Therefore, the performance testing focuses on wireless functionality and its impact, not on the core diagnostic capabilities of measuring FeNO. The document states that the core performance characteristics (analytical limits, precision, accuracy, measurement range, linearity) are identical to the predicate device.

    Here's an analysis based on the provided text, addressing your points:

    1. A table of acceptance criteria and the reported device performance

    Since this 510(k) is for activating Bluetooth, the acceptance criteria and performance for the core function (FeNO measurement) are referenced as being identical to the predicate device (K133898). The key performance criteria for the Bluetooth activation itself relate to wireless coexistence and safety.

    Acceptance Criteria (from Predicate/Identical)Reported Device Performance (NIOX VERO)
    Analytical limits at low levels (limit of detection): 5 ppbIdentical to predicate: 5 ppb
    Precision (for values < 30 ppb): < 3 ppb of measured valueIdentical to predicate: < 3 ppb of measured value
    Precision (for values ≥ 30 ppb): < 10% of measured valueIdentical to predicate: < 10% of measured value
    Accuracy: +/- 5 ppb or max 10%Identical to predicate: +/- 5 ppb or max 10%
    Measurement Range: 5 - 300 ppbIdentical to predicate: 5 - 300 ppb
    Linearity, reportable range: Squared correlation coefficient r² ≥ 0.998, slope 0.95 - 1.05, intercept ±3ppbIdentical to predicate: Squared correlation coefficient r² ≥ 0.998, slope 0.95 - 1.05, intercept ±3ppb
    Wireless Coexistence/Safety: No risks from wireless interference in the clinical setting (as per FDA Radio Frequency Wireless Technology in Medical Devices guidance)Wireless coexistence testing performed; demonstrated no additional risks from wireless interference. FCC Part 15B testing repeated with Bluetooth active.

    2. Sample size used for the test set and the data provenance

    The document does not specify a separate "test set" sample size for the Bluetooth activation. It refers to
    "Verification and validation testing was previously performed for Bluetooth functionality prior to FDA clearance."
    and "FCC Part 15B testing was repeated with the Bluetooth module active in the NIOX VERO."

    The data provenance for the wireless testing is implied to be from the manufacturer's internal testing, conducted previously and then repeated for the US market activation. The country of origin for the internal testing is not explicitly stated, but the manufacturer is based in Sweden. Given the nature of Bluetooth standards and FCC regulations, such testing generally follows international or country-specific standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable for this type of submission. The testing described for Bluetooth activation does not involve expert-established ground truth in the way a clinical diagnostic study would. The ground truth for electrical and wireless performance is defined by industry standards and regulatory requirements (e.g., FCC regulations).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for this type of submission. Adjudication methods are typically used in studies where clinical interpretations or diagnoses are being evaluated by multiple experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a diagnostic tool for measuring FeNO, and the submission concerns the activation of wireless connectivity. It is not an AI-assisted diagnostic tool involving human reader interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device's core function (FeNO measurement) is standalone in the sense that it provides a quantitative measurement. The submission for Bluetooth activation itself is a re-validation of wireless functionality, which is a standalone technical performance. The text states:

    "Verification and validation were previously performed to demonstrate that the USB cable communications functions are equivalent to the Bluetooth wireless communication operation."

    This indicates that equivalence between wired and wireless communication, both standalone technical functions, was established.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the core FeNO measurement specified as "identical to predicate," the ground truth for analytical performance (accuracy, precision, linearity) typically relies on reference methods or calibrated gas standards. This document doesn't detail how the predicate's ground truth was established, but it would not be based on expert consensus, pathology, or outcomes data.

    For the Bluetooth activation, the "ground truth" for wireless coexistence and safety is compliance with regulatory standards and guidelines (e.g., FCC Part 15B, FDA Radio Frequency Wireless Technology in Medical Devices guidance).

    8. The sample size for the training set

    Not applicable for this type of device and submission. This is not a machine learning or AI-based device that would require training sets.

    9. How the ground truth for the training set was established

    Not applicable, as no training set is described or relevant for this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133898

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-11-06

    (321 days)

    Product Code
    Regulation Number
    862.3080
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NIOX VERO® measures Nitric Oxide (NO) in human breath. Nitric Oxide is frequently increased in some inflammatory processes such as asthma. The fractional NO concentration in expired breath (FeNO), can be measured by NIOX VERO according to guidelines for NO measurement established by the American Thoracic Society.

    Measurement of FeNO by NIOX VERO is a quantitative, non-invasive, simple and safe method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy, as an indication of the therapeutic effect in patients with elevated FeNO levels. NIOX VERO is suitable for children, approximately 7 - 17 years, and adults 18 years and older.

    FeNO measurements provide the physician with means of evaluating an asthma patient's response to anti-inflammatory therapy, as an adjunct to the established clinical and laboratory assessments in asthma. The NIOX VERO is intended for prescription use and should only be used as directed in the NIOX VERO User Manual by trained healthcare professionals. NIOX VERO cannot be used with infants or by children approximately under the age of 7, as measurement requires patient cooperation. NIOX VERO should not be used in critical care, emergency care or in anesthesiology.

    Device Description

    NIOX VERO is a portable system for the non-invasive, quantitative measurement of the fraction of exhaled nitric oxide (NO) in expired human breath (FeNO). NO levels increase during allergic airway inflammation. Measurement of changes in FeNO concentration is used in evaluating a patient's response to anti-inflammatory therapy, as an adjunct to established clinical and laboratory assessments.

    The NIOX VERO system is comprised of the NIOX VERO unit with AC adapter, a rechargeable battery, an electrochemical NO sensor, disposable patient filters, and an exchangeable handle containing an internal NO scrubber filter. The NIOX Panel is an optional PC application for operation of the NIOX VERO from a PC and access to electronic medical record systems.

    The patient empties their lungs, inhales deeply through the patient filter to total lung capacity and then slowly exhale for 10 seconds. In approximately one minute, the NO concentration is displayed in parts per billion (ppb).

    The NIOX VERO unit includes a sampling and gas conditioning system and a man-machine interface (MMI). The instrument controls the inhaled sample appropriately via the electronics and software program. Filtering of inhaled air elimination from ambient NO levels. A built-in flow control keeps exhalation standardized at 50 ml/s so that it is standardized for all patients. The sample enters an electromechanical sensor and interacts with an electrolyte. A chemical reaction takes place where electrons are generated proportional to the number of NO molecules.

    Results are processed using dedicated software. In order to verify the device's performance and reliability of measurements, there are built-in system control procedures and a Quality Control procedure to be performed on a daily basis.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (NIOX MINO predicate)Reported Device Performance (NIOX VERO)
    Analytical limits (LOD)5 ppb5 ppb
    Precision< 3 ppb for values < 30 ppb; < 10% for values ≥ 30 ppb< 3 ppb for values < 30 ppb; < 10% for values ≥ 30 ppb
    Accuracy+/- 5 ppb or max 10%+/- 5 ppb or max 10%
    Measurement Range5 - 300 ppb5 - 300 ppb
    LinearitySquared correlation coefficient r² ≥ 0.998, slope 0.95 - 1.05, intercept ±3ppbSquared correlation coefficient r² ≥ 0.998, slope 0.95 - 1.05, intercept ±3ppb

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "A comparison study was performed between the NIOX MINO and the NIOX VERO which demonstrate substantial equivalence in a clinical setting." However, the specific sample size for this clinical comparison study and the provenance (e.g., country of origin, retrospective or prospective) of the data are not provided in the supplied text.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not provided in the supplied text. The device measures Fractional exhaled Nitric Oxide (FeNO), and the ground truth for such measurements would typically be based on the device's analytical precision and accuracy using certified gas samples, rather than expert consensus on medical images or clinical judgment.

    4. Adjudication Method for the Test Set

    This information is not applicable to the type of device and study described and is not provided in the supplied text. Adjudication methods like 2+1 or 3+1 are typically used in studies where multiple human readers interpret data (e.g., images) and their interpretations need to be reconciled to establish a ground truth. For a device measuring a quantitative physiological parameter like FeNO, the "ground truth" is established by the accuracy and precision of the measurement itself against known standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement with AI vs. without AI Assistance

    This information is not applicable and not provided in the supplied text. The NIOX VERO is a standalone medical device that measures FeNO. It is not an AI-assisted diagnostic tool that would be used by human readers to interpret complex data. Therefore, an MRMC study and the concept of human readers improving with AI assistance are not relevant to this device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the performance characteristics (Analytical limits, Precision, Accuracy, Measurement Range, Linearity) listed in the table under "Reported Device Performance (NIOX VERO)" are indicative of standalone performance as they describe the intrinsic measurement capabilities of the device itself. These tests were conducted using "Nitric Oxide gas samples at certified concentrations," which evaluates the device's ability to accurately measure known quantities of NO independently.

    7. The Type of Ground Truth Used

    The ground truth for the performance testing was established using Nitric Oxide gas samples at certified concentrations. This allows for the direct comparison of the device's readings against a known, accurate reference.

    8. The Sample Size for the Training Set

    This information is not provided in the supplied text. The document describes a "510(k) premarket notification" for an established medical device type (FeNO measurement systems). While there may be internal development and calibration (training) processes, the FDA submission focuses on showing substantial equivalence through specific performance criteria rather than detailing training set sizes for machine learning models.

    9. How the Ground Truth for the Training Set Was Established

    This information is not explicitly provided in the supplied text. Similar to the test set, it is highly probable that the ground truth for any internal training or calibration of the device would also involve the use of Nitric Oxide gas samples at certified concentrations to ensure accurate and precise measurements. The document mentions "The sensor stabilization process has been modified. Instead of maintaining the sensor at a constant temperature, there is a compensation algorithm for temperature and humidity correction to the measurement result." This suggests calibration and potentially a training phase where data is collected and used to develop or refine such compensation algorithms, likely against known NO concentrations under varying environmental conditions.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1