Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K240387

    Validate with FDA (Live)

    Date Cleared
    2024-06-21

    (134 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PCI Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous coronary intervention (PCI) and percutaneous transluminal angioplasty (PTA).

    The PCI Guide Wires are not intended for use in the neurovasculature.

    Device Description

    The basic structure of the MINAMO blue / viola consists of a hybrid stainless steel and nitinol (Ni-Ti) core wire, with a stainless steel distal inner rope coil. Surrounding the inner rope coil and the distal core wire is a radiopaque platinum nickel / stainless steel outer coil. The outer and inner coils are soldered to the tapered core wire. A similar coil design is used with other ASAHI guide wires, such as the predicate MINAMO (K190176). In addition, coatings are applied on the surface of the product. The coil portion is coated with a silicone and hydrophilic coating. The proximal side of the core wire is coated with PFA, silicone and PTFE.

    The nominal outer diameter of the MINAMO is 0.36 mm. The devices are available in three lengths: 190 cm, 235 cm and 300 cm.

    The MINAMO blue / viola radiopaque distal tip enables the user to view the position of the tip under X-ray fluoroscopy.

    All sizes are available with a straight, pre-shape, and J-shape design.

    AI/ML Overview

    The provided text focuses on the 510(k) premarket notification for the MINAMO blue and MINAMO viola guide wires. It describes the device, compares it to predicate and reference devices, and outlines the non-clinical and biocompatibility testing performed to demonstrate substantial equivalence.

    However, the text does not contain the specific information requested regarding acceptance criteria and a study proving the device meets these criteria in the context of an AI/ML-enabled medical device.

    The document states that the MINAMO blue and MINAMO viola are catheter guide wires, which are physical medical devices, not AI/ML-enabled software devices. Therefore, the questions related to:

    • AI/ML performance metrics (e.g., Sensitivity, Specificity, AUC)
    • Sample sizes for test sets in an AI/ML context
    • Number of experts for ground truth establishment
    • Adjudication methods
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness studies
    • Standalone AI performance
    • Training set sample sizes and ground truth establishment for training data

    are not applicable to the submitted document as it describes a traditional medical device (guide wire) and not an AI/ML product.

    The document focuses on demonstrating the substantial equivalence of the guide wires to previously cleared predicate devices through:

    • Indications for Use: Same as the predicate.
    • Technological Characteristics: Similar materials, design, operating principle, and sterilization.
    • Non-Clinical Testing/Performance Data: Bench tests following FDA guidance, demonstrating performance similar to predicate devices and meeting acceptance criteria for physical and mechanical properties (e.g., tensile strength, torque strength, tip flexibility, coating integrity, lubricity, corrosion resistance, kink resistance, radiopacity, dimensional verification, simulated use).
    • Biocompatibility: In accordance with ISO 10993.

    Based on the provided text, here's what can be extracted and what cannot:


    1. A table of acceptance criteria and the reported device performance:

    The document states: "The in vitro bench tests demonstrated that the MINAMO blue and MINAMO viola met all acceptance criteria and performed similarly to the predicate devices."
    However, the specific numerical acceptance criteria and the quantified reported device performance values are NOT provided in the document. It only lists the types of tests performed.

    Types of Performance Tests (Implied Acceptance Criteria for these properties):

    • Tensile Strength
    • Torque Strength
    • Torqueability
    • Tip Flexibility
    • Coating Integrity
    • Lubricity
    • Visual Inspection
    • Corrosion Resistance
    • Kink Resistance
    • Radiopacity
    • Dimensional Verification
    • Simulated use in a clinically relevant mode
    • Coating Integrity/Particulate Evaluation in a clinically relevant mode

    Biocompatibility Tests (Implied Acceptance Criteria: Non-cytotoxic, Non-sensitizer, Non-irritant, Non-toxic, Non-pyrogenic, Non-hemolytic, Not an activator, Thromboresistant):

    TestConclusion
    Cytotoxicity - ISO 10993-5Non-cytotoxic
    Sensitization - ISO 10993-10Non-sensitizer
    Irritation - ISO 10993-23Non-irritant
    Systemic Toxicity - ISO 10993-11 (Acute)Non-toxic
    Systemic Toxicity - ISO 10993-11 (Pyrogen)Non-pyrogenic
    Hemocompatibility - ISO 10993-4 (Hemolysis)Non-hemolytic
    Hemocompatibility - ISO 10993-4 (PTT)Not an activator
    Hemocompatibility - ISO 10993-4 (Complement Activation)Not an Activator
    Hemocompatibility - ISO 10993-4 (Thrombogenicity)Thromboresistant

    For all other points (2-9), the information is NOT applicable to this document as it does not describe an AI/ML-enabled device or an AI/ML study.

    2. Sample sized used for the test set and the data provenance: Not applicable (not an AI/ML study). The document refers to "bench testing" and "in vitro" tests, implying physical samples of the guide wire.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable (not an AI/ML study requiring expert review of data).
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable (not an AI/ML study).
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): For the physical device, the "ground truth" would be established by standardized testing protocols and measurements of physical properties. For biocompatibility, it's based on ISO 10993 standards and the biological responses observed in the tests (e.g., no cellular reactivity, no delayed dermal contact sensitization).
    8. The sample size for the training set: Not applicable (no AI model).
    9. How the ground truth for the training set was established: Not applicable (no AI model).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1