    K254016

    Validate with FDA (Live)

    Device Name
    MI View&GO
    Date Cleared
    2026-02-10

    (57 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    21 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).

    MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.

    Device Description

    MI View&GO is a software-only medical device which will be delivered in conjunction with Siemens SPECT/CT and PET/CT scanners. MI View&GO software provides additional specific capabilities for handling of PET and SPECT as well as CT and MR data directly at the acquisition console.

    The MI View&GO software integrates molecular imaging more efficiently in the clinical environment by providing an interface for its users to review, post-process and read medical images immediately after acquisition. The purpose of the MI View&GO is to allow the technologist and reading physician to:

    • Review acquired and reconstructed images at the scanner console
    • Determine that the acquired data is of sufficient quality for reading, so the patient can be released.
    • Prepare images for reading
    • Perform a basic read
    AI/ML Overview

    Here's an analysis of the acceptance criteria and study detailed in the provided FDA 510(k) clearance letter for MI View&GO, structured according to your requested points:

    Acceptance Criteria and Device Performance Study for MI View&GO (K254016)

    1. Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Improved Lung Segmentation (Auto Lung 3D)For new organs (N/A for lung lobes, as they are existing organs with improved models)Not applicable, as lung lobes are "improved organs," not "new organs."
    For unchanged organs (other than lungs and lung lobes)Dice-score on other organs (not retrained) remained unchanged and was verified by recalculating the Dice score with the new algorithm.
    For improved organs (Lung Lobes): Average Dice coefficient per organ shall be greater than or equal to the average Dice coefficient per organ of the predicate algorithm.The average Dice coefficient for all 20 subjects was higher for each lobe in the subject device than in the predicate device. (Note: The document also states "although not greater than a +0.03 difference for all lobes," which clarifies that while improved, the improvement might not be substantial for every lobe.)
    Improved PERCIST Liver Algorithm (binary liver mask input)Average Dice coefficient > 0.8The liver met this criterion.
    Average Symmetric Surface Distance (ASSD) < 10 mmThe liver met this criterion.
    Improved PERCIST Liver Algorithm (Reference Region Placement)N/A (Comparative analysis, not a specific criterion for a single metric)Demonstrated to yield results in better agreement with semi-automatic evaluation by expert readers compared with the predicate method.
    Improved PERCIST Liver Algorithm (Intersection with Suspicious Uptake Masks)N/A (Comparative analysis, goal is fewer intersections)Subject device had fewer intersections (4 cases) compared to the predicate device (13 cases) out of 129 subjects.

    2. Sample Size Used for the Test Set and Data Provenance

    • Improved Lung Segmentation:
      • Sample Size: 20 patients.
      • Data Provenance:
        • Retrospective.
        • Half of the patients were new, and the other 50% were randomly selected from the predicate testing cohort.
        • 50% of patients were from the US.
        • All patients from Siemens Scanner.
    • Improved PERCIST Liver Algorithm (binary liver mask input):
      • Sample Size: 20 patients.
      • Data Provenance:
        • Patients obtained from clinical partners in Europe and USA.
        • Randomly selected with stratification.
        • All subjects from Siemens Scanner.
    • Improved PERCIST Liver Algorithm (Reference Region Placement & Intersection with Suspicious Uptake Masks):
      • Sample Size: 129 subjects for the "intersection with suspicious uptake masks" analysis.
      • Data Provenance: Not explicitly stated for the "reference region placement" analysis, but implied to be from the same or similar source as the 129 subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Improved Lung Segmentation: Not explicitly stated. The ground truth for segmentation metrics (Dice, ASSD) is typically established by manual segmentation performed by experts, but the number of experts and their qualifications are not detailed in this document.
    • Improved PERCIST Liver Algorithm (Reference Region Placement):
      • Number of Experts: Two expert readers.
      • Qualifications: "Expert readers" is mentioned, but specific qualifications (e.g., radiologist 10 years experience) are not provided.
    • Improved PERCIST Liver Algorithm (Intersection with Suspicious Uptake Masks):
      • Number of Experts: One expert reader.
      • Qualifications: "Expert reader" is mentioned; specific qualifications are not provided.

    4. Adjudication Method for the Test Set

    • Improved Lung Segmentation: Not explicitly mentioned. For segmentation ground truth derived from multiple experts, methods like consensus or averaging are common, but not specified here.
    • Improved PERCIST Liver Algorithm (Reference Region Placement): Semi-automatic evaluation by two expert readers. The document states the subject device algorithm was compared to this "reference standard," implying this semi-automatic output was considered the ground truth. No explicit adjudication method (like 2+1) is described for resolving differences between the two experts, if they occurred.
    • Improved PERCIST Liver Algorithm (Intersection with Suspicious Uptake Masks): Identified by "an expert reader." This implies a single expert's identification served as the ground truth. No adjudication mentioned.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No, a formal MRMC comparative effectiveness study involving human readers with and without AI assistance is not described in this document.
    • The studies conducted focus on the algorithm's performance against historical data, expert interpretations, or comparing an improved algorithm to a predicate algorithm.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done

    • Yes, standalone performance studies were conducted for specific features:
      • Improved Lung Segmentation: The Dice coefficient and ASSD evaluation was a standalone algorithmic performance assessment against presumed expert-derived ground truth.
      • Improved PERCIST Liver Algorithm (binary liver mask input): The Dice coefficient and ASSD evaluation for the liver mask was a standalone algorithmic performance assessment.
      • Improved PERCIST Liver Algorithm (Reference Region Placement): The comparison of the algorithm's results to the semi-automatic evaluation by two expert readers is a standalone algorithm assessment, where the expert input constitutes the ground truth.
      • Improved PERCIST Liver Algorithm (Intersection with Suspicious Uptake Masks): This was a standalone algorithmic evaluation of how often the algorithm's PERCIST VOIs intersected suspicious uptake areas identified by an expert.

    7. The Type of Ground Truth Used

    • Improved Lung Segmentation: Likely expert consensus/manual segmentation (implied by Dice coefficient and ASSD, which compare algorithm output to a gold standard segmentation).
    • Improved PERCIST Liver Algorithm (binary liver mask input): Likely expert consensus/manual segmentation (implied by Dice coefficient and ASSD for the liver mask).
    • Improved PERCIST Liver Algorithm (Reference Region Placement): Expert semi-automatic evaluation from two expert readers. These semi-automatic outputs were treated as the reference standard.
    • Improved PERCIST Liver Algorithm (Intersection with Suspicious Uptake Masks): Expert identification of suspicious tracer uptake masks by a single expert reader.

    8. The Sample Size for the Training Set

    • Not explicitly stated in the document. The document mentions that the lung lobe segmentation algorithm was "re-trained with additional data" and that there was "No overlap of patients between training, tuning, and test cohorts," but does not provide details on the training set's size.

    9. How the Ground Truth for the Training Set Was Established

    • Not explicitly stated in the document. For machine learning models, ground truth for training data is typically established through expert labeling (e.g., manual segmentation, disease annotation), but the specifics are not provided here.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242300

    Validate with FDA (Live)

    Device Name
    MI View&GO
    Date Cleared
    2024-08-30

    (28 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).

    MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.

    Device Description

    MI View&GO is a software-only medical device which will be delivered in conjunction with Siemens SPECT/CT and PET/CT scanners. MI View&GO software provides additional specific capabilities for handling of PET and SPECT as well as CT and MR data directly at the acquisition console.

    The MI View&GO software integrates molecular imaging more efficiently in the clinical environment by providing an interface for its users to review, post-process and read medical images immediately after acquisition. The purpose of the MI View&GO is to allow the technologist and reading physician to:

    • Review acquired and reconstructed images at the scanner console
    • Determine that the acquired data is of sufficient quality for reading, so the patient can be released.
    • Prepare images for reading
    • Perform a basic read
    AI/ML Overview

    The provided text is a 510(k) summary for the device MI View&GO VA30. It states that "The device under application, MI View&GO, did not conduct any additional performance testing" for determining substantial equivalence to the predicate device. Therefore, a study proving the device meets acceptance criteria as described in the prompt was not conducted and thus, the requested information (table of acceptance criteria, sample sizes, expert qualifications, etc.) is not available in the provided text.

    The substantial equivalence determination was based on:

    1. Software Verification and Validation: This involved demonstrating continued conformance with special controls for medical devices containing software. The testing supported that all software specifications met predetermined acceptance criteria and substantiated all requirements and functional specifications, including those related to device hazards.
    2. Risk Analysis: A risk analysis was completed, and risk control was implemented to mitigate identified hazards.
    3. Comparison to Predicate Device: The device was deemed substantially equivalent to the predicate device (MI View&GO VA20, K201202) because there are no differences in the Indications for Use, Intended Use, or Fundamental Technological Characteristics. The new features implemented (e.g., O Brain AC-PC, PERCIST, VOI Isocontour) did not raise any new issues of safety and effectiveness.

    In summary, the document indicates that the device's performance was not evaluated through a specific comparative study against acceptance criteria of the kind typically seen for novel diagnostic algorithms. Instead, its substantial equivalence was primarily established through software verification and validation, risk analysis, and a direct comparison of its technological characteristics and intended use to a previously cleared predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222172

    Validate with FDA (Live)

    Device Name
    MI View&GO
    Date Cleared
    2022-08-19

    (29 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as position emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).

    MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.

    Device Description

    MI View&GO is a software-only medical device which will be delivered in conjunction with Siemens SPECT/CT and PET/CT scanners. MI View&GO software provides additional specific capabilities for handling of PET and SPECT as well as CT and MR data directly at the acquisition console.

    The MI View&GO software integrates molecular imaging more efficiently in the clinical environment by providing an interface for its users to review, post-process and read medical images immediately after acquisition. The purpose of the MI View&GO is to allow the technologist and reading physician to:

    • Review acquired and reconstructed images at the scanner console
    • Determine that the acquired data is of sufficient quality for reading, so the patient can be released.
    • Prepare images for reading
    • Perform a basic read
    AI/ML Overview

    This FDA 510(k) summary for Siemens' MI View&GO VA20A software describes modifications to an existing medical image management and processing system. The summary does not include a detailed study proving the device meets specific acceptance criteria in the format requested.

    The document states that "Verification and Validation activities have been successfully performed on the software package, including assurance that functions work as designed, performance requirements and specifications have been met, and that all hazard mitigations have been fully implemented. All testing has met the predetermined acceptance values." However, it does not provide a table of acceptance criteria and reported device performance, nor does it specify the methodologies, sample sizes, or ground truth establishment for such testing.

    Therefore, I cannot provide the requested information from the provided text. The document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed performance study with quantitative acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201202

    Validate with FDA (Live)

    Device Name
    MI View&GO
    Date Cleared
    2020-06-02

    (29 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).

    MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.

    Device Description

    MI View&GO is a software-only medical device which will be delivered with the new generation of Siemens SPECT/CT and PET/CT scanners.

    MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR)

    MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study information for the MI View&GO VA10A device:

    Important Note: The provided document is a 510(k) summary for the MI View&GO VA10A device. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed de novo clinical study with specific acceptance criteria for novel performance claims. As such, many of the typical clinical study elements like "effect size," "sample size for test set," "number of experts," etc., are not explicitly described in the same way they would be for a new device with novel performance claims requiring rigorous clinical validation. The testing described is primarily for verification and validation of the software’s functions, performance requirements, and hazard mitigations against predetermined acceptance values, in line with established industry standards for medical device software.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of the submission (510(k) for a software update with substantial equivalence), the acceptance criteria are not in the form of clinical performance metrics like sensitivity/specificity for a diagnostic task. Instead, they relate to software engineering and regulatory compliance.

    Acceptance Criterion TypeReported Device PerformanceComments
    Software FunctionalityFunctions work as designed"Verification and Validation activities have been successfully performed on the software package, including assurance that functions work as designed."
    Performance RequirementsPerformance requirements and specifications have been met"performance requirements and specifications have been met"
    Hazard MitigationAll hazard mitigations have been fully implemented"all hazard mitigations have been fully implemented."
    Testing OutcomesAll testing has met the predetermined acceptance values"All testing has met the predetermined acceptance values."
    Risk ManagementRisk analyses in compliance with ISO 14971"Risk Management has been ensured via risk analyses in compliance with ISO 14971"
    Quality System ComplianceDesigned and manufactured in accordance with Quality System Regulations (21 CFR 820)"The device is designed and manufactured in accordance with Quality System Regulations as outlined in 21 CFR 820."
    CybersecuritySpecific cybersecurity controls implemented to prevent unauthorized access, modifications, misuse, or denial of use; controls for information storage and transfer"The software has specific cybersecurity controls to prevent unauthorized access, modifications, misuse or denial of use. Additionally, controls are enabled to prevent the unauthorized use of information that is stored, accessed or transferred between the software and external devices."
    TraceabilityTraceability of requirements ensured during component integration, software validation, and system testing"Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing."
    Compliance with StandardsAdherence to recognized and established industry standards (EN ISO 13485, IEC 62304, ISO 14971, IEC 62366-1, ISO 15223-1, NEMA PS 3.1 – 3.20)"Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards for development including EN ISO 13485 and IEC 62304." (Specific standards listed later)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated as a separate "test set" in the context of image data for diagnostic performance evaluation. The "test set" here refers to the software testing environment. The document states "Verification and Validation activities have been successfully performed on the software package." This implies internal testing with various inputs.
    • Data Provenance (e.g., country of origin of the data, retrospective or prospective): Not specified. Since this is primarily a software update and substantial equivalence claim, the focus is on functional verification rather than a de novo clinical study with specific patient data sets to prove diagnostic accuracy.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not specified. For this type of submission, "ground truth" refers more to established software requirements and specifications rather than a diagnostic standard established by human experts on a clinical dataset.
    • Qualifications of Experts: Not specified.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable or described. Adjudication is typically used in clinical studies when multiple human readers evaluate data, and their disagreements need to be resolved to establish a ground truth. Here, "ground truth" pertains to software functionality meeting its design specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, an MRMC comparative effectiveness study was not performed or described. The MI View&GO VA10A is described as an image processing software for viewing, manipulation, quantification, analysis, and comparison, not specifically an AI-driven diagnostic aid that would improve human reader performance in terms of diagnostic accuracy. Its features are largely related to workflow, display, and basic image processing.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable in the sense of a diagnostic algorithm's standalone performance (e.g., sensitivity/specificity for a disease). MI View&GO is a "medical diagnostic application" intended to be "utilized by appropriately trained health care professionals to aid in the management of diseases." It's a tool for professionals, not an autonomous diagnostic algorithm. Its performance is linked to the functionality described (viewing, manipulation, etc.), which were verified to meet specifications, not generate diagnostic outputs independently.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The "ground truth" for this submission are the predetermined acceptance values and functional specifications as defined by the software's design and requirements. It also includes compliance with recognized industry standards (ISO 14971, IEC 62304, etc.) and regulatory requirements (21 CFR 820, cybersecurity guidance). This is a software engineering "ground truth" rather than a clinical diagnostic ground truth like pathology or outcome data.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. MI View&GO VA10A is not described as an AI/Machine Learning device that undergoes a "training phase" on a dataset. It is image processing software.

    9. How the Ground Truth for the Training Set was Established

    • Ground Truth for Training Set: Not applicable. As it's not an AI/ML device requiring a training set, the establishment of ground truth for such a set is not relevant.

    Summary of the Study:

    The "study" or rather, the evidence presented in this 510(k) submission, is a demonstration of Verification and Validation (V&V) activities. This included ensuring that:

    1. The software's functions worked as designed.
    2. Performance requirements and specifications were met.
    3. All hazard mitigations were fully implemented according to risk analyses (ISO 14971).
    4. All testing met predetermined acceptance values.
    5. The device adheres to recognized industry standards (e.g., IEC 62304, ISO 13485) and Quality System Regulations (21 CFR 820).
    6. Cybersecurity controls are in place.

    The overall approach is a comparison to predicate devices (syngo.via MI Workflows, syngo.CT View&GO, syngo.CT Extended Functionality) to demonstrate substantial equivalence in terms of intended use, indications for use, and technological characteristics, especially for the new or modified features. The study proves that the device meets its acceptance criteria by stating that "All testing has met the predetermined acceptance values" and that V&V activities confirmed functional and performance requirements were met as per software development and regulatory standards.