Search Results
Found 8 results
510(k) Data Aggregation
(59 days)
The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging.
Endoscopic ICG System
Upon intravenous administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion in adults and pediatric patients ≥1 month of age, and at least one of the major extrahepatic bile ducts (cystic duct, common bile duct and common hepatic duct) in adults and pediatric patients ≥ 12 years of age, using near infrared imaging in accordance with the appropriately indicated endoscope.
Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.
Additionally, the KARL STORZ Endoscopic ICG System enables surgeon to perform minimally invasive cranial neurosurgery in adults and pediatrics and endonasal skull base surgery in adults and pediatrics > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging.
Upon interstitial administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
Exoscopic ICG System
VITOM ICG SYSTEM
The KARL STORZ VITOM ICG System is intended for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures in adults and pediatrics ≥ 1 month of age.
The VITOM ICG System is intended to provide a magnified view of the surgical field.
RUBINA Lens System
The RUBINA Lens System is intended for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures in adults and pediatrics ≥ 1 month of age.
Upon interstitial administration and use of the ICG consistent with its approved label, the RUBINA Lens System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
Upon intradermal administration and use of the ICG consistent with its approved label, the RUBINA Lens System is indicated for fluorescence imaging of lymph nodes and delineation of lymphatic vessels during lymphatic mapping in adults undergoing breast surgical procedures for which fluorescence imaging is a component of intraoperative management.
The RUBINA Lens System is intended to provide a wide-angle view of the surgical field.
The RUBINA® Lens supports anatomical visualization under white light and NIR/ICG fluorescence. Its native 16:9 full-screen image eliminates the need for digital zoom, enabling shorter working distances that help maintain fluorescence signal capture during procedures requiring continuous visualization of lymphatic or perfusion-related structures.
The device incorporates a wide-angle optical design and expanded focus range, supporting consistent imaging across variable working distances for applications such as:
- Fluorescence-guided assessment of tissue perfusion
- Visualization of lymphatic pathways and sentinel lymph nodes
The RUBINA® Lens features a 90° direction of view and may be used handheld or mounted to a compatible holding arm. Rotation of the attached camera head allows horizon adjustment to maintain anatomical orientation.
The system provides continuous white-light and NIR/ICG visualization for display on standard operating room monitors. The device achieves optimal illumination at approximately 18 cm and maintains fluorescence visualization across a broad working distance range, supporting use in applications such as perfusion assessment and lymphatic mapping.
The subject device RUBINA® Lens is compatible with the following components within the KARL STORZ ICG Imaging System:
- IMAGE1 S™ Rubina® camera head (TH121) previously cleared on K201399 and K202925.
- IMAGE1 S™ Camera Control Unit (CCU) (TC201US, TC304US) previously cleared on K212695, K201135, K233333, K232857.
- Fiber Light Cables (495NCSC, 495TIP): used to transmit visible and NIR light from the Power LED Rubina Light Source (TL400) to the RUBINA® Lens. The 495NCSC was previously cleared K201399, and K202925. The 495TIP Fiber Light Cable was most recently cleared in K233333.
- The Power LED Rubina® Light Source (TL400) previously cleared in K201399, K202925, K212695, K232857, and K233333. The TL400 is included as a subject device, as the KARL STORZ ICG Imaging System Indications for Use reflected in the TL400 labeling require revision within this 510(k). No modifications have been made to the TL400 with respect to materials, technological characteristics, performance, reprocessing, or any other essential design features.
- Footswitch (UF101): [Optional] Previously cleared in K201399, K202925, K212695, and K232857.
N/A
Ask a specific question about this device
KARL STORZ ICG Imaging System
The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging.
Endoscopic ICG System
Upon intravenous administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion in adults and pediatric patients ≥1 month of age, and at least one of the major extrahepatic bile duct, common bile duct and common hepatic duct) in adults and pediatric patients ≥ 12 years of age, using near infrared imaging in accordance with the appropriately indicated endoscope. Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.
Additionally, the KARL STORZ Endoscopic ICG System enables surgeon to perform minimally invasive cranial neurosurgery in adults and pediatrics and endonasal skull base surgery in adults and pediatrics > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging.
Upon interstitial administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
VITOM II ICG System
The KARL STORZ VITOM II ICG System is intended for capturing fluorescent mages for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures. The VITOM II ICG System is intended to provide a magnified view of the surgical field in standard white light.
KARL STORZ Image1 S CCU
The Imagel S camera control unit (CU) in combination with either a compatible camera head or an appropriately indicated video endoscope is intended for real-time visualization, image recording and documentation during general endoscopic and microscopic procedures in adults and pediatrics.
KARL STORZ ICG Imaging System
The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging for general surgical sites including the abdomen, bile ducts, brain/skull base, and the lymph nodes/lymphatic vessels. Components of the system include:
Scopes:
3D TIPCAM®1 Rubina videoendoscope
26006ACA/BCA, 26616ACA/BCA Rigid Endoscope
26003ACA/ARA/BCA/BRA/FCA/FRA/FCEA/FREA
26046ACA/ARA/BCA/BRA/FCA/FRA
28164AC/BC/FC VITOM II ICG/NIR Telescope
20916025AGA
Light Source:
Power LED Rubina (TL400) Foot Switch (UF101) Fiber Optic Light Cable (495TIP/NCSC/NAC)
Camera Head:
Image1 S 4U Rubina (TH121)
Camera Control Unit (CCU):
Image1 S Connect II (TC201US) Image1 S 4U-Link (TC304US)
KARL STORZ Image1 S CCU
The KARL STORZ IMAGE1 S Camera Control Unit (CCU) is a modular CCU that consists of Image1 S Connect and Connect II modules and the link modules. The Connect modules can be connected to minimum of one and a maximum of three links modules. The modularity enables customers to customize their Image1 S system to their specific video needs.
The Image1 S includes, but not limited to, the following features:
Brightness control Enhancement Control Automatic Light Source Control Shutter Control Image/Video Capture
Seven increments of zoom from 1-2.5x and adaptive zoom
Modules of the Image1 S CCU include: Image1 S Connect (TC200US) Image1 S Connect II (TC201US) Image1 S H3-Link (TC300US) Image1 S X-Link (TC301US) Image1 S D3-Link (TC302US) Image1 S 4U-Link (TC304US)
Accessories to the Image1 S CCU include: Microscope Footswitch (TC019) Image1 S Pilot (TC014) LINK Cable (TC011, TC012)
The provided text describes the regulatory submission for the KARL STORZ ICG Imaging System and KARL STORZ Image 1S Camera Control Unit.
However, the document explicitly states that "Clinical testing was not required to demonstrate substantial equivalence to the predicate devices." This means that the submission does not contain information about a study proving the device meets acceptance criteria based on clinical performance metrics (like sensitivity, specificity, accuracy, or human reader improvement with AI assistance).
The acceptance criteria and performance data mentioned in the document are non-clinical performance data, specifically related to electrical safety, electromagnetic compatibility, and software verification and validation. This type of information is usually presented as compliance with established standards rather than a clinical study with a test set, ground truth, or expert readers.
Therefore, most of the requested information regarding clinical study design (sample size, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance, training set details) cannot be extracted from this document because such a clinical study was not required or provided for this specific submission as per the FDA's determination of substantial equivalence to predicate devices (K212695 and K201135).
Here's what can be extracted and inferred from the document:
1. A table of acceptance criteria and the reported device performance
Based on the "Non-Clinical Performance Data" section, the acceptance criteria are compliance with relevant safety and software standards.
| Acceptance Criteria Category | Specific Standard/Requirement | Reported Device Performance/Compliance |
|---|---|---|
| Electrical Safety | IEC 60601-1:2005 + A1:2012 + A2:2021 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance | Electrical Safety testing was conducted in accordance with the specified standard. (Implies compliance, as it's part of a successful 510(k) submission). |
| Electromagnetic Compatibility (EMC) | IEC 60601-1-2: 2014 + A1:2020, Medical Electrical Equipment – Part 1-2: General requirements for basic safety and essential performance - Electromagnetic Compatibility | Electromagnetic Compatibility testing was conducted in accordance with the specified standard. (Implies compliance). |
| Software Verification and Validation | FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions" issued June 14, 2023. The software documentation level conforms to the Basic Level of documentation (no identified risks where a failure or flaw could present a hazardous situation with a probable risk of death or serious injury to a patient, user, or others). | Software verification and validation testing was conducted and documentation was provided as recommended by the FDA guidance. The software documentation level conforms to the Basic Level of documentation as there are no risks identified in which a failure or flaw of any device software function(s) could present a hazardous situation with a probable risk of death or serious injury, either to a patient, user of the device, or others in the environment of use. (Implies successful verification and validation according to the stated guidance and risk level). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not applicable / not provided. The document states "Clinical testing was not required." The "performance data" refers to non-clinical engineering and software testing, which does not involve a "test set" of clinical cases in the sense of imaging data for diagnostic performance evaluation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable / not provided. No clinical ground truth was established from experts as clinical testing was not required.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable / not provided. No clinical ground truth was established, therefore no adjudication method was used for clinical interpretation.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable / not provided. No MRMC study was performed as clinical testing was not required for this submission. The device is an imaging system, not an AI-based diagnostic aid that would assist human readers in interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable / not provided. The device is an imaging system, not a standalone diagnostic algorithm. No such performance study was conducted or required.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable / not provided for clinical performance. For the non-clinical performance data, the "ground truth" or standard was compliance with specified international and FDA-recognized standards for electrical safety, EMC, and software validation.
8. The sample size for the training set
- Not applicable / not provided. As this is an imaging system and not an AI/ML-based diagnostic algorithm, there is no mention of a "training set" of data in the context of machine learning model development. The software testing mentioned refers to standard software verification and validation, not machine learning model training.
9. How the ground truth for the training set was established
- Not applicable / not provided. See above.
Ask a specific question about this device
(52 days)
The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging.
Upon intravenous administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for biliary duct visualization.
Additionally, the KARL STORZ Endoscopic ICG System enables surgeon to perform minimally invasive cranial neurosurgery in adults and pediatrics and endonasal skull base surgery in adults and pediatrics > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging.
The KARL STORZ VITOM ICG System is intended for capturing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures in adults and pediatrics >Imonth of age. The VITOM ICG System is intended to provide a magnified view of the surgical field.
Upon interstitial administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
The subject device KARL STORZ ICG System includes the following components:
- VITOM EAGLE (TH201): a 3D video exoscope with 4K resolution used during open procedures for the evaluation of tissue perfusion, related tissue-transfer circulation in tissue and free flaps used in plastic, micro and reconstructive surgical procedures. The subject device VITOM EAGLE System is being indicated for use in in adults and pediatrics >1month of age.
- Fiber Light Cable (495VTE): used to transmit visible and NIR light from the Power LED Rubina light source to the VITOM Eagle.
- IMAGE1 Pilot (TC014): used to control the optical functions of the VITOM EAGLE.
- Microscope Footswitch (TC019): alternatively used control the optical functions of the VITOM EAGLE
- The Power LED Rubina light source (TL400) along with the footswitch (UF101): previously cleared in K201399, K202925 and K212695.
- Imagel S Camera Control Unit (TC201US, TC304US): previously cleared in K201399, K202925 and K212695.
The provided text describes the KARL STORZ ICG Imaging System and its acceptance criteria, along with a summary of the non-clinical performance data used to demonstrate substantial equivalence to a predicate device. However, it does not describe a study involving an AI algorithm. The device is an imaging system that uses Indocyanine Green (ICG) fluorescence for various surgical visualizations.
Here's a breakdown of the requested information based only on the provided text, heavily noting limitations due to the absence of AI-specific study details:
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a non-AI imaging system without specific AI performance metrics, the acceptance criteria are generally related to the technical performance of the imaging capabilities. The document states that the KARL STORZ ICG Imaging System (subject device) was compared to the predicate VITOM II ICG/NIR telescope of the KARL STORZ ICG Imaging System (K212695). The performance was demonstrated by testing for:
| Acceptance Criteria (Performance Metric) | Reported Device Performance (Subject Device vs. Predicate) |
|---|---|
| Spatial Resolution | Successfully demonstrated by comparison |
| Signal to Noise Ratio and Noise | Successfully demonstrated by comparison |
| Dynamic Range | Successfully demonstrated by comparison |
| Geometric Distortion | Successfully demonstrated by comparison |
| Depth of Field | Successfully demonstrated by comparison |
| Illumination Detection Uniformity | Successfully demonstrated by comparison |
| Latency | Successfully demonstrated by comparison |
| Penetration Depth | Successfully demonstrated by comparison |
| Simultaneous Color Contrast | Successfully demonstrated by comparison |
| Minimum Detectable Concentration of ICG | Successfully demonstrated by comparison |
| 3D Zoom and Rotation | Successfully demonstrated by comparison |
| 2D and 3D Mode Transition | Successfully demonstrated by comparison |
| Image Alignment | Successfully demonstrated by comparison |
| Photobiological Safety | Successfully demonstrated by comparison |
| Electrical Safety and EMC (IEC 60601-1, IEC 60601-1-2) | Follows FDA recognized consensus standards and tested accordingly |
2. Sample size used for the test set and the data provenance
The document does not specify a "test set" in the context of an AI algorithm or patient data. The performance evaluation was based on non-clinical bench testing comparing the subject device's imaging capabilities to a predicate device. Therefore, there's no mention of sample size in terms of patient data or data provenance (country of origin, retrospective/prospective). The "sample" here would refer to the physical devices and controlled test scenarios.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. As this is a non-AI imaging system being evaluated via bench testing, there are no "experts" establishing ground truth for a test set of images or patient data. The ground truth for the technical performance criteria would be established by validated measurement techniques and instrumentation during the bench tests.
4. Adjudication method for the test set
Not applicable, as there is no "test set" in the context of human expert review or an AI algorithm's output. The evaluation was based on objective technical measurements.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
No. The document explicitly states "Clinical testing was not required to demonstrate the substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to establish the substantial equivalence of the modifications." Furthermore, this is not an AI-assisted device, so MRMC studies on AI assistance would not be relevant.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. The device is an imaging system, not a standalone AI algorithm. It produces images for human surgeons to interpret.
7. The type of ground truth used
For the non-clinical performance data, the ground truth was based on objective measurements from bench testing (e.g., measuring spatial resolution, signal-to-noise ratio, etc.) against established technical specifications or a predicate device's performance.
8. The sample size for the training set
Not applicable. The device does not involve an AI algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable, as there is no AI algorithm or training set.
Summary of AI-related information (or lack thereof):
The provided text describes a submission for a KARL STORZ ICG Imaging System, which is a medical device for real-time visible and near-infrared fluorescence imaging during surgery. The entire document focuses on demonstrating the substantial equivalence of this updated imaging system to a previously cleared predicate device through non-clinical bench testing. There is no mention of any artificial intelligence (AI) component, machine learning model, or any studies related to AI performance, human-in-the-loop improvements with AI, or standalone algorithm performance. Therefore, most of the questions regarding AI-specific criteria cannot be answered from the provided text.
Ask a specific question about this device
(82 days)
The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging.
Upon intravenous administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization. Additionally, the KARL STORZ Endoscopic ICG System enables surgeon to perform minimally invasive cranial neurosurgery in adults and pediatrics and endonasal skull base surgery in adults and pediatrics > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging.
The KARL STORZ VITOM II ICG System is intended for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures. The VITOM II ICG System is intended to provide a magnified view of the surgical field in standard white light.
Upon interstitial administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
The modified KARL STORZ ICG Imaging System is identical to the KARL STORZ ICG Imaging System recently cleared under K202925. The modified KARL STORZ ICG Imaging System now includes the following components: TIPCAM®1 Rubina Videoscope (0°, 30°): a 3D image capable videoendoscope with 2D auto-leveling (auto-rotation) and 2D auto-switch display modes.
The 4mm, 5mm & 10mm Endoscopes and VITOM ICG telescopes connected to the optical coupler of the Image1 S 4U Rubina camera head, and the TIPCAM®1 Rubina Videoendoscope which connects to the Image1 S Camera Control Unit for image processing, as well as to the Power LED Rubina light source light source via compatible light cable as the source of illumination to allow visualization of internal anatomy. Visualization and navigation is performed initially using VIS imaging. NIR imaging is selected when visual assessment and/or confirmation of vessels, blood flow or tissue perfusion is desired.
For the NIR image, the user has three presentations of the ICG imagery to choose from:
a. Overlay: The white light image is overlaid with the NIR image. The NIR image could either by blue or green
b. Intensity Map: The white light image is overlaid with color transformed NIR image.
c. Monochromatic: The NIR image is indicated by the color white against a dark background.
The KARL STORZ ICG Imaging System can output a 4K image to the monitor 12G/3G-SDI, DVI-D and DisplayPort digital outputs and also offers 7 increments of zoom ranging from 1x to 2.5x.
The reference device, TIPCAM1 Rubina Video Endoscope System, is identical in design and materials of construction to the subject device. The only difference is that the ICG functionality was not included as part of K201526.
The provided text does not describe a study involving an AI/Machine Learning device. Instead, it is a 510(k) premarket notification for the KARL STORZ ICG Imaging System, which is an endoscopic imaging system used to provide real-time visible and near-infrared fluorescence imaging.
The document states:
- "Clinical testing was not required to demonstrate the substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to establish the substantial equivalence of the modifications." This explicitly indicates that no comparative clinical study (including MRMC or standalone performance with human-in-the-loop) was performed or needed for this specific submission (K212695).
- The acceptance criteria and performance data mentioned are related to non-clinical bench testing to ensure the device met its design specifications and conformed to various electrical safety, EMC, and ISO endoscopic standards, as well as software verification and validation. This is typical for optical/imaging hardware devices rather than AI algorithms.
- There is no mention of "ground truth" in the context of diagnostic accuracy, nor any experts establishing such ground truth.
Therefore, providing answers to the requested points about AI acceptance criteria, study sample sizes, expert involvement, and ground truth establishment would be inaccurate as this information is not present in the provided text, because the device described is not an AI device.
Ask a specific question about this device
(79 days)
The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging.
Upon intravenous administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.
Additionally, the KARL STORZ Endoscopic ICG System enables surgeon to perform minimally invasive cranial neurosurgery in adults and pediatrics and endonasal skull base surgery in adults and pediatrics > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging.
The KARL STORZ VITOM II ICG System is intended for capturing fluorescent mages for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures. The VITOM II ICG System is intended to provide a magnified view of the surgical field in standard white light.
Upon interstitial administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
The subject device, KARL STORZ ICG Imaging System includes the following components: 4mm, 5mm & 10mm HOPKINS ICG/NIR Endoscopes, VITOM II ICG Telescope, Image1 S Rubina Camera head, Power LED Rubina light source, footswitch, Fiber Light Cables, Image1 S CCU.
The expanded indication for visualization of the lymphatic vessel is only to the Endoscopic ICG system that includes the 4mm, 5mm & 10mm HOPKINS ICG/NIR Endoscopes.
The addition of the reprocessing modalities is to the 4mm, 5mm & 10mm HOPKINS ICG/NIR Endoscopes and the VITOM II ICG/NIR telescope.
The endoscopes/telescope are intended to be connected to the optical coupler of the camera head, which connects to the CCU for image processing, as well as to the light source via compatible light cable as the source of illumination to allow visualization of internal anatomy. The users can switch between WLi (standard white light) mode, for visualization of the endoscopic and microscopic procedures and NIR (Near Infrared) mode to detect ICG presence.
The KARL STORZ ICG Imaging System can be used with any medical grade monitor with a DVI-D or 12G/3G-SDI input.
The provided text describes a 510(k) submission for the KARL STORZ ICG Imaging System (K202925). It outlines the device's intended use and compares it to predicate devices to establish substantial equivalence. However, this document does not contain information about acceptance criteria, clinical study results, sample sizes for training or test sets, number/qualifications of experts, adjudication methods, MRMC studies, or standalone algorithm performance.
The document explicitly states: "Clinical testing was not required to demonstrate the substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to establish substantial equivalence of the expanded indications." This indicates that the regulatory clearance for this device was based on non-clinical data and comparison to existing predicate devices, rather than a clinical study with detailed performance metrics against predefined acceptance criteria for AI/ML performance.
Therefore, I cannot provide the requested information about acceptance criteria and study proving device performance as it is not present in the provided text.
To answer your request precisely, based on the provided document, the following information is NOT available:
- A table of acceptance criteria and the reported device performance: Not provided. The submission relies on substantial equivalence to predicates based on non-clinical data.
- Sample sizes used for the test set and the data provenance: Not provided, as clinical testing was not performed for this clearance.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not provided, as clinical testing and ground truth establishment for AI/ML performance are not detailed.
- Adjudication method for the test set: Not provided.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not provided. The submission focuses on the imaging system itself, not an AI component assisting human readers.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not provided. The device is an imaging system, not an AI algorithm with standalone performance metrics mentioned.
- The type of ground truth used: Not provided.
- The sample size for the training set: Not provided.
- How the ground truth for the training set was established: Not provided.
Ask a specific question about this device
(89 days)
The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging.
The KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, or at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.
Additionally, the KARL STORZ Endoscopic ICG System enables surgeon to perform minimally invasive cranial neurosurgery in adults and pediatrics and endonasal skull base surgery in adults and pediatrics > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging.
The KARL STORZ VITOM II ICG System is intended for capturing fluorescent mages for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures. The VITOM II ICG System is intended to provide a magnified view of the surgical field in standard white light.
The KARL STORZ ICG Imaging System is used to provide real-time high-definition (HD) endoscopic or telescopic images of visible (VIS) and near-infrared (NIR) indocyanine green (ICG) dye fluorescence during minimally invasive, neuro- and endonasal skull base surgery as well as plastic, micro- and reconstructive surgical procedures in general and pediatric populations.
The overall system includes a 4mm HOPKINS ICG/NIR Endoscope (0°, 30° or 45°) for use in neuro- and endonasal skull base surgery, a 5mm & 10mm HOPKINS ICG/NIR Endoscope (0° or 30°) for use in minimally invasive procedures and a VITOM II ICG/NIR Telescope for use in plastic, micro- and reconstructive surgical procedures for VIS and NIR illumination and imaging, a light source with foot switch for emission of VIS and NIR illumination, a color video camera head capable of capturing both VIS and NIR imaging, and a KARL STORZ ICG Kit. Additional accessories used with the KARL STORZ ICG Imaging System include two standards fiber-optic light cables for transmission of VIS and NIR light and the Image1 S Camera Control Unit (CCU). The KARL STORZ ICG Imaging System can be used with any medical grade HD monitor with a DVI-D or 3G-SDI input.
The provided text, K180146, describes the KARL STORZ ICG Imaging System, an endoscopic device for real-time visible and near-infrared fluorescence imaging, particularly for assessing vessels, blood flow, and tissue perfusion. The submission is a 510(k) for a modified device, primarily adding a 4mm endoscope for neuro- and endonasal skull base surgery in adults and pediatrics.
Based on the document, here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are primarily defined by adherence to established standards for medical devices and demonstration of equivalence to predicate devices through bench testing and literature review. The document does not provide specific quantitative acceptance criteria for clinical performance (e.g., a specific sensitivity or specificity for detecting blood flow issues). Instead, it relies on demonstrating that the device meets its design specifications and performs comparably to the predicate devices and that the expanded indications are supported by existing clinical literature.
| Parameter | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Safety | Compliance with Electrical Safety and EMC standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18). | "KARL STORZ ICG Imaging System follows the FDA recognized consensus standards and is tested according to the following standards: Electrical Safety and EMC IEC 60601-1 IEC 60601-1-2 IEC 60601-2-18" |
| Compliance with Biocompatibility standards (ISO 10993) including systemic toxicity, intracutaneous irritation, maximization sensitization, mucosal irritation. | "Biocompatibility (ISO 10993) – Systemic toxicity, Intracutaneous irritation, Maximization sensitization, Mucosal Irritation" (indicating it was tested for these) | |
| Reprocessing (Cleaning and Steam Sterilization) validation according to AAMI and ISO standards (AAMI TIR30:2011, AAMI TIR 12:2010, ANSI/AAMI/ISO 14937:2009, ANSI/AAMI ST9:2010/A4:2013, ISO TS 15883:2005). | "Reprocessing (Cleaning and Steam Sterilization) – AAMI TIR30:2011 AAMI TIR 12:2010 ANSI/AAMI/ISO 14937:2009 ANSI/AAMI ST9:2010/A4:2013 ISO TS 15883:2005 Processing/Reprocessing Medical Device in Health Care Settings: Validation Methods and Labeling" (indicating testing was performed) | |
| Performance | Meeting design specifications through performance testing for: Color Performance, Detection Linearity, Distortion, Dynamic Range, Illumination, Resolution, SNR & Sensitivity, Depth of Field, UV Exposure, Irradiance, Color Reproduction. | "Performance Testing – Color Performance, Detection Linearity, Distortion, Dynamic Range, Illumination, Resolution, SNR & Sensitivity, Depth of Field, UV Exposure, Irradiance, Color Reproduction" and "Additional bench testing was performed to ensure the device met its design specifications. The bench testing performed verified and validated that the KARL STORZ ICG Imaging System has met all its design specification..." |
| Substantial equivalence to predicate device (K171238) regarding intended use, design, technology, and performance specifications, specifically supporting the addition of 4mm endoscopes. Expected to have similar performance characteristics for parameters like Endoscope Type, Direction of View, Field of View, Imaging Type, Imaging Agent, Light Source Compatibility. Depth of Field could differ based on new endoscopes but must be acceptable for intended use. | "The KARL STORZ ICG Imaging System is a modification of and substantially equivalent to the primary predicate, KARL STORZ Imaging System (K171238), in regards to its intended use, design, technology, and performance specifications. The main difference between the subject and primary predicate device is the addition of the 4mm HOPKINS ICG/NIR Endoscopes to the KARL STORZ Endoscopic ICG System to be used in neuro- and endonasal skull base surgery in adults and pediatrics." Comparative Table: | |
| Endoscope Type | Rigid, rod lens | Rigid, rod lens |
| Endoscope Diameters | 4mm, 5mm & 10mm | 5mm & 10mm |
| Direction of View | 0°, 30°, 45° | 0°, 30° |
| Working Length | 18cm (4mm), 29cm (5mm), 31cm (10mm) | 29cm (5mm), 31cm (10mm) |
| Depth of Field | 8mm – 38mm | 30mm - 80mm (5mm), 30mm – 110mm (10mm) |
| Field of View | 80° | 74° |
| Imaging Type | Visible and near-infrared imaging | Visible and near-infrared imaging |
| Imaging Agent | ICG | ICG |
| Light Source Compatibility | Xenon | Xenon |
| Substantial equivalence to secondary predicate (K021050) regarding intended use (VIS imaging), design, technology, and performance, specifically regarding the addition of a filter for NIR imaging. Expected to have similar performance characteristics. | "The KARL STORZ ICG Imaging System is also substantial equivalent to the secondary predicate, KSEA Neuroendoscope & Accessories, in regards to its intended use (VIS imaging), design, technology, and performance specifications. The main difference between the subject and secondary predicate device is the addition of filter located at the eyepiece of the endoscope for the purpose of NIR imaging." Comparative Table: | |
| Endoscope Type | Rigid, rod lens | Rigid, rod lens |
| Endoscope Diameters | 4mm, 5mm & 10mm | 2.7mm, 4mm & 4.8mm |
| Direction of View | 0°, 30°, 45° | 0°, 30°, 45° |
| Working Length | 18cm (4mm), 29cm (5mm), 31cm (10mm) | 18 or 20 cm |
| Depth of Field | 8mm – 38mm | 8mm – 38mm |
| Field of View | 80° | 80° |
| Imaging Type | Visible and near-infrared imaging | Visible Imaging |
| Imaging Agent | ICG | N/A |
| Light Source Compatibility | Xenon | Xenon |
| Expanded Indications | Clinical support for NIR imaging in neuro- and endonasal skull base surgeries, and for use in pediatric populations. | "Clinical published literatures were provided to support the effectiveness of NIR imaging in the neuro- and endonasal skull base surgeries as well as the use of the KARL STORZ ICG Imaging System in pediatrics." (References provided in bibliography). |
2. Sample size used for the test set and the data provenance
The document explicitly states: "Clinical performance is not required to demonstrate substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to establish substantial equivalence."
Therefore, there appears to be no specific "test set" of patient data for evaluating algorithm performance in the way one would for an AI/ML device. The evaluation relies on:
- Bench testing: This refers to laboratory-based functional and safety testing of the device hardware and software. The sample size for this is not specified in terms of "patients" or "cases" but rather in terms of components or test conditions. The provenance is internal to the manufacturer's testing facility.
- Clinical Literature: Published studies from various sources were used to support the expanded indications for use in neuro- and endonasal skull base surgeries and in pediatric populations. The provenance of this data is from the respective published articles (e.g., Journal of Neurological Surgery Part B, Neurologia Medico-Chirurgica). These are retrospective studies conducted by various researchers, not specifically "prospective" studies for this particular 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Since no specific clinical "test set" for algorithm performance was used, there is no mention of experts for ground truth establishment in that context. The "ground truth" for the device's functional performance was established through engineering and quality control testing against design specifications and relevant standards. For the expanded indications, the "ground truth" is derived from the established scientific and clinical evidence presented in the published literature.
4. Adjudication method for the test set
Not applicable, as no clinical test set requiring expert adjudication for ground truth was generated for this submission.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was conducted or referenced in this submission. This device is an imaging system (hardware and associated software), not an AI-based diagnostic algorithm intended to assist human readers in image interpretation. Its purpose is to generate the imaging data (visible and NIR fluorescence), not to interpret it in an AI-assisted manner.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. The device itself is a tool for image acquisition, primarily used by surgeons for visual assessment during procedures. It does not perform an automated, standalone diagnostic algorithm.
7. The type of ground truth used
- For device performance validation: The ground truth was established by engineering design specifications and compliance with recognized industry standards (IEC, ISO, AAMI). This involved direct measurement and verification against established physical and electrical parameters.
- For expanded indications: The ground truth was established by published clinical literature related to the effectiveness and safety of NIR imaging using ICG in the specified surgical fields and pediatric populations. This relies on the expert consensus and evidence presented in those scientific publications.
8. The sample size for the training set
Not applicable. This device is not an AI/ML algorithm that requires a "training set" of data for learning. Its functionality is based on established optical and electronic principles.
9. How the ground truth for the training set was established
Not applicable. As there is no AI/ML training set, there is no ground truth establishment for it.
Ask a specific question about this device
(28 days)
The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging.
The KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.
The KARL STORZ VITOM II ICG System is intended for capturing fluorescent mages for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures. The VITOM II ICG System is intended to provide a magnified view of the surgical field in standard white light.
The KARL STORZ ICG Imaging System is used to provide real-time high definition (HD) endoscopic or telescope images of visible (VIS) and near-infrared (NIR) indocyanine green (ICG) dye fluorescence during minimally invasive surgery as well as plastic, micro- and reconstructive surgical procedures.
The overall system includes a 5mm &10mm HOPKINS ICG/NIR Endoscope for use in minimally invasive procedures and a VITOM II ICG/NIR Telescope for use in plastic, micro- and reconstructive surgical procedures for VIS and NIR illumination and imaging, a light source with foot switch for emission of VIS and NIR illumination, a color video camera head capable of capturing both VIS and NIR imaging, and KARL STORZ ICG Kit. Additional accessories used with the KARL STORZ ICG Imaging System include two standard fiber-optic light cables for transmission of VIS and NIR light and the Image1S Camera Control Unit (CCU).
The KARL STORZ ICG Imaging System can be used with any medical grade HD monitor with a DVI-D or 3G-SDI input.
The provided document is a 510(k) summary for the KARL STORZ ICG Imaging System. It describes the device, its intended use, and argues for its substantial equivalence to a predicate device. However, this document does not contain information about acceptance criteria or a study that proves the device meets specific performance acceptance criteria in the context of diagnostic accuracy or a similar measure typically associated with AI/algorithm performance.
The document states:
- "Clinical testing was not required to demonstrate the substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to assess safety and effectiveness and to establish the substantial equivalence of the modifications."
- "Additional bench testing was performed to ensure the device met its design specifications."
- "The bench testing performed verified and validated that the KARL STORZ ICG Imaging System has met all its design specification and is substantially equivalent to its predicate devices."
This indicates that the device's approval was based on demonstrating safety, effectiveness, and substantial equivalence to existing devices through non-clinical (bench) testing, primarily focusing on aspects like electrical safety, electromagnetic compatibility, biocompatibility, and mechanical properties, not on diagnostic performance against ground truth in a clinical setting.
Therefore, I cannot provide the requested information for acceptance criteria and a study proving device performance in the context of diagnostic accuracy, as this type of information is not present in the provided text.
Based on the nature of the device (an imaging system for real-time visible and near-infrared fluorescence imaging) and the approval process described, it is highly probable that the "acceptance criteria" referred to in the document pertain to technical specifications, safety standards, and functional performance benchmarks rather than clinical diagnostic accuracy metrics (like sensitivity, specificity, or AUC).
If this were a typical AI/algorithm-based diagnostic device, the requested information would be crucial. However, for a medical imaging system like this, the focus is often on its ability to produce clear, reliable images and its safety for patient use, with appropriate regulatory standards and bench testing serving as the "proof."
Ask a specific question about this device
(104 days)
The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging.
The KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.
The KARL STORZ VITOM II ICG System is intended for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures. The VITOM II ICG System is intended to provide a magnified view of the surgical field in standard white light.
The KARL STORZ ICG Imaging System is used to provide real-time high definition (HD) endoscopic or telescope images of visible (VIS) and near-infrared (NIR) indocyanine green (ICG) dye fluorescence during minimally invasive or open surgery. The system components are a HOPKINS ICG/NIR Endoscope and a VITOM II ICG/NIR Telescope for VIS and NIR illumination and imaging, a light source with foot switch for emission of VIS and NIR illumination, a color video camera head capable of capturing both VIS and NIR imaging, and KARL STORZ ICG Kit. Additional accessories used with the KARL STORZ ICG Imaging System include a standard fiber-optic light cable for transmission of VIS and NIR light and the Image1S Camera Control Unit (CCU). All components except for the VITOM II ICG/NIR Telescope were cleared in K152583. The KARL STORZ ICG Imaging System can be used with any medical grade HD monitor with a DVI-D or 3G-SDI input. The VITOM II ICG/NIR Telescope is a Class I device under 21 CFR 878.4700.
The provided text describes a 510(k) submission for the "KARL STORZ ICG Imaging System". This document focuses on demonstrating the substantial equivalence of the new device to previously cleared predicate devices, rather than proving that the device meets specific acceptance criteria through a standalone clinical or "algorithm only" study. Therefore, much of the requested information about acceptance criteria, detailed study design, ground truth establishment, and multi-reader multi-case studies is not present in the provided text.
Specifically, the document states: "Clinical testing was not required to demonstrate the substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to assess safety and effectiveness and to establish the substantial equivalence of the modifications."
This indicates that the submission relies on demonstrating comparable performance to predicate devices through non-clinical (bench and animal) testing, rather than establishing specific performance metrics against a predefined acceptance criterion in human clinical data.
However, I can extract information about the non-clinical testing performed and how it aims to demonstrate equivalence, which implicitly serves as the "study" proving the device's acceptable performance in the context of this 510(k).
Here's the information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state quantitative "acceptance criteria" in the format of a table with specific metrics (e.g., minimum sensitivity, maximum false positive rate). Instead, the performance is demonstrated by showing that the new system is "as safe as and as effective as" the predicate devices in various non-clinical tests. The reported 'performance' is qualitative: the system "successfully tested for its functions and performance" and "enable a selective visualization of the ICG fluorescence signal."
| Category | Acceptance Criteria (Implicit / Demonstrated Equivalence) | Reported Device Performance (Qualitative) |
|---|---|---|
| Device Function | The device should function as intended, providing real-time VIS and NIR fluorescence imaging, and selectively visualize ICG fluorescence signal. | "The KARL STORZ ICG Imaging System has been successfully tested for its functions and performance, including verification that the spectral characteristics of the ICG system illumination light source, light transmission system, and telescope enable a selective visualization of the ICG fluorescence signal as detected by the camera system." |
| Key Performance Aspects | Parameters such as UV Exposure, Color Reproduction, Irradiance Levels, Depth Dependency, and Visualization of ICG Signal should be comparable to predicate devices. | Comparative testing was conducted for: • UV Exposure • Color Reproduction • Irradiance Levels • Depth Dependency • Visualization of ICG Signal While specific quantitative results are not provided, the conclusion states that these tests "demonstrated that the subject device is as safe as and as effective as the primary predicate device," implying satisfactory performance on these aspects. |
| Biocompatibility | Should not change from the previously cleared system. | "The addition of the VITOM II ICG Telescope to the system does not change the biocompatibility... conducted for the KARL STORZ Endoscopic ICG System cleared in K152583." |
| Electromagnetic Compatibility (EMC) | Should not change from the previously cleared system. | "The addition of the VITOM II ICG Telescope to the system does not change the ... electromagnetic compatibility... conducted for the KARL STORZ Endoscopic ICG System cleared in K152583." |
| Electrical Safety | Should not change from the previously cleared system. | "The addition of the VITOM II ICG Telescope to the system does not change the ... electrical safety... conducted for the KARL STORZ Endoscopic ICG System cleared in K152583." |
| Software Validations | Should not change from the previously cleared system. | "The addition of the VITOM II ICG Telescope to the system does not change the ... software validations conducted for the KARL STORZ Endoscopic ICG System cleared in K152583." |
| Cleaning & Sterilization | Validation for the new component (VITOM II ICG Telescope) is required. | "Cleaning method validation and sterilization validation were performed on the VITOM II ICG Telescope." |
| Animal Study Performance | Performance in animal model should be adequate for evaluation. | "a GLP animal study was successfully performed by the NAMSA testing facility to evaluate the performance of the KARL STORZ ICG Imaging System with the VITOM II ICG Telescope in the porcine open surgery laparoscopy model." |
| Compliance with Standards | Device must meet relevant safety standards. | The device (and its predicates) are listed as meeting IEC 60601-1-2 and IEC 60601-1. The Visionsense predicate also met IEC 60825-1. "Bench testing was performed per IEC 62471." |
2. Sample size used for the test set and the data provenance
- Sample Size: Not explicitly stated with numerical values. The document refers to "comparative testing" and a "GLP animal study" without specifying the number of samples or animals used.
- Data Provenance:
- Bench Testing: Conducted in-house or by testing laboratories (e.g., "NAMSA testing facility"). The country of origin is not specified but implied to be within the scope of FDA regulations (likely USA or a country with recognized testing standards).
- Animal Study: Performed by "NAMSA testing facility" using a "porcine open surgery laparoscopy model."
- Retrospective/Prospective: All described tests are non-clinical and would be conducted prospectively as part of the device development and submission process.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This was a non-clinical study for substantial equivalence. The "ground truth" was based on the physical and optical properties of the device and its ability to visualize ICG fluorescence as measured by established testing methods and verified by the testing facility. There's no mention of human experts establishing "ground truth" in the diagnostic sense for a test set.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. This was a non-clinical study. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in clinical imaging studies involving multiple human readers.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. No MRMC study was performed as this was a 510(k) submission based on non-clinical substantial equivalence. The device is an imaging system, not an AI-powered diagnostic tool, and the submission explicitly states "Clinical testing was not required."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This concept doesn't directly apply here. The device itself is an integrated imaging system (camera, endoscope/telescope, light source). Its "standalone performance" is demonstrated through its physical and optical characteristics (e.g., light transmission, spectral characteristics, visualization of ICG signal) as measured in bench and animal studies, rather than an "algorithm."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the non-clinical tests would have been based on:
- Physical measurements and optical properties: For tests like "spectral characteristics," "irradiance levels," and "depth dependency," the ground truth is derived from established physics principles and calibrated measurement equipment.
- Known ICG fluorescence: For "visualization of ICG signal," the ground truth is the known presence and location of the fluorescent dye.
- Biological outcomes (animal model): In the GLP animal study, the "performance" would have been evaluated based on the ability of the system to correctly visualize blood flow/perfusion and biliary ducts as expected in the porcine model, potentially against direct observation or other validated methods if used in the study.
8. The sample size for the training set
Not applicable. This is not an AI/machine learning device that requires a "training set."
9. How the ground truth for the training set was established
Not applicable, as this is not an AI/machine learning device.
Ask a specific question about this device
Page 1 of 1