510(k) Data Aggregation

    K Number
    K211127

    Validate with FDA (Live)

    Date Cleared
    2021-09-08

    (145 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device uses the oscillometric method to automatically measure systolic and diastolic blood pressure as well as heart rate. The measurement position is at human being's arm. All values can be read out in the LCD panel. Measurement position is at human being's upper arm. The intended user of this over-the-counter device is adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) for home use. HL858DM features BP Category Indicator that will show the information with the readings on the user tracking their blood pressure level. HL858DM is equipped with an Advanced IHB detection feature to collect and analyze pulses. If the specific irregular heartbeats are detected and it may affect blood pressure reading with device will give the user a warning signal. The feature can inform the user that the measured blood pressure reading may be inaccurate once the specific irregular heartbeats are detected. Besides, the device features a built-in "Bluetooth Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth-enabled device. Also, users could simply synchronize the current date and time, and check the battery status of blood pressure montor by means of DailyChek® application software with the paired Bluetooth-enabled device.

    Device Description

    HL858DM automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use. The device has the Risk Category Indicator, Irregular Heartbeat Detector, Bluetooth data transmission functions. When the symbol Advanced IHB appears on screen indicates the specific heartbeat irregularity was detected during measurement. Additionally, after measurement, the BP Category Indicator feature will show the information with the readings on the screen for the user tracking their blood pressure level. Besides, HL858DM featured with an IHB feature can identify the specific irregular heartbeats that may cause deviated blood pressure reading.

    AI/ML Overview

    The device in question is the Full Automatic (NIBP) Blood Pressure Monitor, Model HL858DM.

    Here's an analysis of its acceptance criteria and the study proving its performance:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Feature/MetricAcceptance Criteria (Predicate Device)Reported Device Performance (Subject Device HL858DM)
    Accuracy - Pressure± 3mmHgSame as left (± 3mmHg)
    Accuracy - Pulse± 5%Same as left (± 5%)
    Method of MeasurementOscillometricOscillometric
    Measurement TypeDuring inflationDuring inflation
    Range of Measurement - Pressure0-300mmHgSame as left (0-300mmHg)
    Range of Measurement - Pulse40-199 Beats/minuteSame as left (40-199 Beats/minute)

    2. Sample Size for the Test Set and Data Provenance:

    • Sample Size: The document does not explicitly state the sample size used for the clinical validation. It only mentions that the device is compliant with ISO 81060-2:2018, which is the "Non-invasive sphygmomanometers- Part 2: Clinical validation of automated measurement type."
    • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. It only states that "All the relevant activities were performed by designate individual(s)."

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the clinical validation. ISO 81060-2 typically requires comparisons to expert observer measurements, but the specifics are not detailed here.

    4. Adjudication Method for the Test Set:

    • The document does not specify the adjudication method used for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was done. This device is a blood pressure monitor, not an AI-assisted diagnostic tool that would typically involve multiple human readers. The clinical validation focuses on the accuracy of the device's measurements against a reference method.

    6. Standalone Performance Study:

    • Yes, a standalone study was performed. The clinical validation was focused on the device's ability to accurately measure blood pressure and heart rate as a standalone system, in compliance with ISO 81060-2:2018. The document states, "The subject device HL858DM is compliant to the standard of ISO 81060-2:2018 Noninvasive sphygmomanometers- Part 2: Clinical validation of automated measurement type." This standard specifically outlines the requirements for clinical validation of automated blood pressure devices.

    7. Type of Ground Truth Used:

    • Based on compliance with ISO 81060-2:2018, the ground truth for blood pressure measurement is established through simultaneous auscultatory measurements by trained observers using a mercury sphygmomanometer or another validated reference oscillometric device. The standard dictates a rigorous protocol for comparison. The document does not explicitly detail the exact ground truth method used beyond referencing the standard.

    8. Sample Size for the Training Set:

    • This device is a hardware blood pressure monitor and does not explicitly use a "training set" in the context of machine learning algorithms for its core measurement function. Its measurement principle is based on oscillometry. The "training" in this context would refer to the calibration and internal algorithm development by the manufacturer, which is not detailed in the provided document.

    9. How Ground Truth for the Training Set Was Established:

    • As noted above, this device does not have a "training set" in the machine learning sense. The underlying algorithms for oscillometric blood pressure measurement are developed and validated through engineering and physiological principles, and then verified through clinical trials like the one referenced.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190507

    Validate with FDA (Live)

    Date Cleared
    2019-09-16

    (199 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HL858CL automatically measures human's Systolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 22 inches (approx.23 cm to 56 cm) and for home use.

    HL858CL detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.

    Besides, the device features a built-in "Bluetooth Data Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth-enabled device. Also, users could simply synchronize the current date and time, and check the battery status of blood pressure montor by means of DailyChek® application software with the paired Bluetooth-enabled device.

    Device Description

    HL858CL automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 22 inches (approx.23 cm to 56 cm) and for home use.

    The device will display a symbol O-+-, to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.

    Besides, when Triple Check mode is turned on by user, the symbol ( ) will display on the LCD. Then press Start/Stop button the device will take three consecutive measurements automatically at 1 minute intervals. After measurements are completed LCD will display the average values of the three measurements. And user can use the Bluetooth Data Transmission function, which provides users an optional choice to log, track and store their measurement data. Also synchronize the current date and time, and check the battery status of blood pressure monitor by means of DailyChek® application software with the paired Bluetooth-enabled device.

    AI/ML Overview

    The provided text describes the 510(k) submission for the Full Automatic (NIBP) Blood Pressure Monitor, Model HL858CL. The key study proving its performance is a clinical investigation conducted to demonstrate compliance with the standard ISO 81060-2: Second Edition 2013-05-01, "Non-invasive sphygmomanometers- Part 2: Clinical validation of automated measurement type."

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The core acceptance criteria for blood pressure monitors, as per ISO 81060-2, relate to the accuracy of BP measurements. While the exact numerical criteria for mean difference and standard deviation are not explicitly stated in the provided text, the document confirms that the device did meet these criteria. The text states: "The results of this clinical investigation show that the required limits for mean difference and standard deviation are fulfilled by the subject device HL858CL in the group of 90 subjects with qualified distribution."

    Acceptance Criteria (from ISO 81060-2)Reported Device Performance
    Mean Difference (Systolic)Fulfilled required limits
    Standard Deviation (Systolic)Fulfilled required limits
    Mean Difference (Diastolic)Fulfilled required limits
    Standard Deviation (Diastolic)Fulfilled required limits

    (Note: The exact numerical values for 'mean difference' and 'standard deviation' were not provided in the document, only the statement that the 'required limits' were fulfilled.)

    2. Sample Size and Data Provenance

    • Sample Size for Test Set: 90 subjects.
    • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. However, clinical validation studies for medical devices are typically prospective.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not specify the number or qualifications of experts used to establish ground truth. For clinical validation of blood pressure monitors according to ISO 81060-2, the ground truth is typically established by trained human observers using a reference sphygmomanometer, often with simultaneous auscultation and standardized procedures, but no details are provided on the specific experts involved.

    4. Adjudication Method for the Test Set

    The document does not explicitly detail an adjudication method. In ISO 81060-2 studies, blood pressure measurements are typically taken by multiple observers (often two) and averaged or compared against a standardized reference, but specific adjudication processes (e.g., 2+1, 3+1) are not mentioned in this summary.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This device is a blood pressure monitor, which is typically validated for its standalone accuracy against a reference standard, not for human reader improvement with AI assistance.

    6. Standalone (Algorithm Only) Performance

    Yes, the study described is essentially a standalone performance study. The device's accuracy in measuring blood pressure is assessed against a reference standard (as per ISO 81060-2), which means it's evaluating the algorithm and hardware without human interpretation of ambiguous readings.

    7. Type of Ground Truth Used

    The ground truth for the clinical validation was established by comparison to a reference standard as dictated by ISO 81060-2. This typically involves simultaneous measurements by the device under test and trained observers using a calibrated auscultatory method.

    8. Sample Size for the Training Set

    The document does not provide information about a training set size. For blood pressure monitors validated to ISO 81060-2, the primary clinical study is a validation study, not a deep learning model's training. The device's algorithm would have been developed and refined using internal data, but details of that process (including training set size) are not part of this 510(k) summary for clinical validation.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, information about a "training set" and its ground truth is not provided in this regulatory summary, as the primary focus is on the clinical validation against the ISO standard.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190133

    Validate with FDA (Live)

    Date Cleared
    2019-06-13

    (136 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HL158UB automatically measures human's Systolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with wrist circumference ranging from 5.3 inches to 7.7 inches (approx. 135 mm to 195 mm) and for home use.

    HL158UB detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the user tracking their blood pressure level.

    Device Description

    HL158UB automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with wrist circumference ranging from 5.3 inches to 7.7 inches (approx.13.5 cm to 19.5 cm) and for home use.

    The device will display a symbol ~, to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level. Besides, HL158UB has a built-in Wrist Position Guide. By holding START/STOP button, the display will illuminate with different icons that are designed to guide you move your wrist.

    AI/ML Overview

    The provided text describes the HL158UB Full Automatic (NIBP) Blood Pressure Monitor and its compliance with regulatory requirements, including clinical and non-clinical tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from ISO 81060-2:2013)Reported Device Performance (HL158UB)
    Accuracy: Pressure: $\pm$ 3 mmHgAccuracy: Pressure: $\pm$ 3 mmHg
    Accuracy: Pulse: $\pm$ 5%Accuracy: Pulse: $\pm$ 5%
    Clinical Study Compliance: Fulfillment of ISO 81060-2:2013 requirements for wrist-type blood pressure monitors for accuracy specification and performance."The clinical validation result demonstrates the subject blood pressure monitor fulfills the acceptance criteria therefore the accuracy specification and performance of the subject device is judged fully comply with the requirements of ISO 810601-2."
    Usability Goals and Design Considerations: Conformity to usability goals and design considerations and risk criteria of medical devices in risk management procedures."The HL158UB demonstrates conformity to the usability goals and design considerations and risk criteria of medical device in risk management procedure by alleviating potential use errors and risk derived from real world experience. This ensures HL158UB is not only safe to use but also offers usability."

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Study (Accuracy Test):
      • Sample Size: 100 valid subjects.
      • Data Provenance: Retrospective, conducted at Lohas Clinic in Taipei, Taiwan (R.O.C.), 2015-16.
    • Usability Validation (Human Factors Test):
      • Sample Size: 60 valid subjects (recruited across four target groups).
      • Data Provenance: Prospective, conducted at Health & Life Co., Ltd. in Taipei, Taiwan (R.O.C.), 2018.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number or qualifications of experts used to establish the ground truth for the clinical accuracy test. It only mentions that the clinical study was conducted at Lohas Clinic and assessed compliance with ISO 81060-2:2013, which typically involves a comparison to a reference standard (e.g., auscultatory method by trained observers). However, no details on these observers are provided.

    For the usability validation, the ground truth was based on observing user interactions and assessing if the device met usability goals and risk criteria as per IEC 62366-1, IEC 60601-1-6, and IEC 60601-1-11. This involves human factors engineers or usability specialists, but specific numbers and qualifications are not mentioned.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method (like 2+1 or 3+1 consensus) for the clinical accuracy study or the usability validation.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This device is a blood pressure monitor, not an AI-assisted diagnostic imaging tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, the primary evaluation for the HL158UB blood pressure monitor (an algorithm-driven device) is inherently a standalone performance assessment. The "clinical study" described evaluates the device's accuracy against a recognized standard (ISO 81060-2:2013), which is a measure of its standalone performance in blood pressure measurement. The usability study also assesses the device's performance in interaction with users, without explicitly comparing it to human readers.

    7. The Type of Ground Truth Used

    • For Clinical Accuracy: The ground truth for blood pressure measurement accuracy would typically be established by a reference standard, such as auscultation (manual sphygmomanometry) performed by trained observers, as specified by standards like ISO 81060-2. While the document mentions compliance with ISO 81060-2, it does not explicitly state the specific ground truth method used beyond "clinical validation result."
    • For Usability: The ground truth was based on observing user interactions and assessing the effectiveness, efficiency, ease of user learning, and user satisfaction against pre-defined usability goals and risk criteria. This is a form of empirical user performance data and expert evaluation.

    8. The Sample Size for the Training Set

    The document does not provide information about a separate training set. The HL158UB is compared to its predicate device, HL158UA, and states that it uses the "same... algorithm." This suggests that the core algorithm may have been developed and "trained" (in a broader sense, if it involves any machine learning components, which is not explicitly stated for a blood pressure monitor) prior to this submission, possibly in the development of the predicate device or a similar platform. No specific training set data is mentioned for the HL158UB itself in this document.

    9. How the Ground Truth for the Training Set Was Established

    As no specific training set is mentioned for the HL158UB, there is no information on how its ground truth would have been established. Given that it's a blood pressure monitor based on the oscillometric method, the "training" of such algorithms typically involves extensive physiological data, calibration, and validation against a gold standard for blood pressure measurement. However, this is not detailed in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190134

    Validate with FDA (Live)

    Date Cleared
    2019-05-20

    (112 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HL158TB automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with wrist circumference ranging from 5.3 inches to 7.7 inches (approx. 135 mm to 195 mm) and for home use.

    HL158TB detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the user tracking their blood pressure level.

    Device Description

    HL158TB automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with wrist circumference ranging from 5.3 inches to 7.7 inches (approx.13.5 cm to 19.5 cm) and for home use.

    The device will display a symbol ♥*~, to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.

    AI/ML Overview

    The provided text describes the submission for a medical device called the "Full Automatic (NIBP) Blood Pressure Monitor, Model HL158TB." This document focuses on demonstrating substantial equivalence to a previously cleared predicate device (HL158TA, K160336).

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of acceptance criteria and the reported device performance:

    The document states that the subject device (HL158TB) and the predicate device (HL158TA, K160336) have the "same method of measurement, measure type, critical components (ex. air pump, sensor...) and algorithm." Therefore, the acceptance criteria and performance are implicitly aligned with the predicate device. The performance specifications are given in comparison to the predicate device.

    ItemAcceptance Criteria (based on Predicate Device HL158TA)Reported Device Performance
    AccuracyPressure $\pm$ 3mmHgSame as predicate
    Pulse $\pm$ 5%Same as predicate
    Range of measurementCuff Pressure: 0~300mmHgSame as predicate
    Determination: 40~280mmHg (Blood Pressure)Same as predicate
    Pulse: 40~199 Beats/minuteSame as predicate
    InflationAutomatic Inflation (Air Pump)Same as predicate
    DeflationAutomatic Air Release Control ValveSame as predicate
    Pressure Changed Rate2~5mmHg/sec.Same as predicate
    DisplayLiquid Crystal DigitalSame as predicate
    Power Supply2 "AAA(LR03)(1.5V)" Alkaline BatteriesSame as predicate
    Storage/Transport-25°C~70°C, ≤93% R.H.Same as predicate
    Operating environment5°C ~ 40°C, 15% ~ 93% R.H., 700~1060 hPaSame as predicate
    MaterialABS housing and ABS keysSame as predicate
    Memory1*60, total 60 setsSame as predicate
    Cuff sizeWrist circumference approx. 135 ~ 195 mmSame as predicate
    Irregular HeartbeatYes (detection)Same as predicate

    Note: While the "Risk Category Indicator" and "Error icon" have been updated, the core accuracy and measurement parameters remain the same, implying the HL158TB meets the same performance acceptance criteria for blood pressure and pulse accuracy as the predicate.

    2. Sample size used for the test set and the data provenance:

    The document states: "In ISO 81060-2 Second Edition 2013-05-01 Non-invasive sphygmomanometers-Part 2: Clinical validation of automated measurement type, all the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met."

    However, this statement refers to the predicate device (K160336). The document explicitly says, "clinical study results can be transferred from the K160336 predicate to the subject device." Therefore, the clinical validation for the subject device relies on the predicate's study.

    • Sample size: Not specified for the predicate device's clinical validation study, only that it followed ISO 81060-2 which has specific patient number requirements (e.g., typically at least 85 subjects with specific age and blood pressure distributions).
    • Data provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not detailed in the provided text for the clinical study of the predicate device. The ISO 81060-2 standard typically involves trained observers for ground truth measurements (using auscultation), but their specific number and detailed qualifications are not provided here.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not detailed in the provided text. ISO 81060-2 outlines a specific methodology for collecting reference blood pressure measurements (often including multiple observers), but the specific adjudication method (e.g., 2+1) is not explicitly mentioned here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is a "Full Automatic Blood Pressure Monitor" which automatically measures blood pressure. It does not involve "human readers" or "AI assistance" in the context of interpreting medical images or complex diagnostic data. Its primary function is a direct measurement.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, the device is standalone. It is a "Full Automatic (NIBP) Blood Pressure Monitor." The clinical validation, transferred from the predicate device that followed ISO 81060-2, assessed the device's accuracy against reference measurements, which is an assessment of its standalone performance. The ISO 81060-2 standard is for the validation of automated sphygmomanometers.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For blood pressure monitors validated according to ISO 81060-2, the ground truth is typically established by cuff auscultation (e.g., by trained observers using a mercury sphygmomanometer), considered the gold standard for non-invasive blood pressure measurement in clinical validation studies.

    8. The sample size for the training set:

    This information is not provided. The document discusses clinical validation, which is an evaluation of the device's performance, not the training of an algorithm in the machine learning sense. While the device utilizes an oscillometric algorithm, details about its development or any training data used are not disclosed. The primary claim for HL158TB's performance relies on the transfer of clinical study results from its predicate, which shares the "same... algorithm."

    9. How the ground truth for the training set was established:

    This information is not provided. As mentioned above, details about algorithm development or training data are not available in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162338

    Validate with FDA (Live)

    Date Cleared
    2016-09-21

    (30 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HL158UA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with wrist circumference ranging from 5.3 inches to 7.7 inches (approx. 135 mm to 195 mm) and for home use.

    HL158UA detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.

    Device Description

    HL158UA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with wrist circumference ranging from 5.3 inches to 7.7 inches (approx.13.5 cm to 19.5 cm) and for home use.

    The device will display a symbol wor ma, to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.

    Besides, HL158UA has a built-in Wrist Position Guide. By holding START/STOP button, the display will illuminate with different icons that are designed to guide you move your wrist.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Full Automatic (NIBP) Blood Pressure Monitor model HL158UA. While it mentions clinical and non-clinical tests were performed and met acceptance criteria, it does not provide a detailed breakdown of the acceptance criteria nor the specific results of a study to prove the device meets these criteria in the format requested.

    The text focuses on demonstrating substantial equivalence to a predicate device (HL158VA K160557) by comparing specifications and listing standards the device complies with.

    Here's a breakdown of what can be extracted and what information is missing from the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that the device is compliant with ISO 81060-2: Second Edition 2013-05-01 "Non-invasive sphygmomanometers- Part 2: Clinical validation of automated measurement type." This standard defines the acceptance criteria for clinical accuracy of automated sphygmomanometers. However, the document does not explicitly list the acceptance criteria from this standard in a table, nor does it present the specific performance results from the clinical validation study against those criteria.

    It only states: "All the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met."

    Based on the ISO 81060-2 standard, typical acceptance criteria for blood pressure monitors involve:

    • Mean difference: The average difference between the device's readings and reference measurements (e.g., from a mercury sphygmomanometer).
    • Standard deviation: The variability of these differences.

    The typical acceptance criteria from ISO 81060-2 are:

    • Mean difference ± Standard Deviation: ≤ ±5 mmHg for both systolic and diastolic blood pressure.
    • Individual differences: Not more than 5 mmHg for a high percentage of measurements (e.g., 60% or 80%), and not more than 15 mmHg for nearly all measurements (e.g., 95%).

    Since the document does not provide the specific reported device performance, I cannot fill in the "Reported Device Performance" column with actual numbers.

    Metric (based on ISO 81060-2)Acceptance Criteria (from ISO 81060-2)Reported Device Performance (Not provided in text)
    Mean Difference (Systolic)≤ ±5 mmHgNot provided
    Standard Deviation (Systolic)≤ 8 mmHgNot provided
    Mean Difference (Diastolic)≤ ±5 mmHgNot provided
    Standard Deviation (Diastolic)≤ 8 mmHgNot provided
    Pulse Accuracy± 5%Same as left (from comparison table)

    2. Sample size used for the test set and the data provenance

    The document states compliance with ISO 81060-2 for clinical validation. This standard typically requires a minimum of 85 participants for clinical testing, covering a range of blood pressures and demographic characteristics.

    • Sample size for the test set: Not explicitly stated, but inferred to be compliant with the minimum requirements of ISO 81060-2 (typically 85 subjects).
    • Data provenance: Not specified (e.g., country of origin, retrospective or prospective). Clinical studies conforming to ISO standards are generally prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The ISO 81060-2 standard requires that reference blood pressure measurements, which serve as the ground truth, be taken by two trained observers (experts) using auscultation with a mercury sphygmomanometer or an equivalent validated device. These observers must be trained and blinded to each other's readings and the results of the automated device.

    • Number of experts: Not explicitly stated, but inferred to be at least two (as per ISO 81060-2).
    • Qualifications of experts: Not specified (e.g., "radiologist with 10 years of experience"), but implied to be trained observers in blood pressure measurement as required by the ISO standard.

    4. Adjudication method for the test set

    The ISO 81060-2 standard specifies a method for comparing the automated device's readings with reference measurements taken by two trained observers. It typically involves averaging the two observers' readings or comparing discrepancies. If there's a significant difference between the two observers, a third observer might be involved, or the measurement discarded.

    • Adjudication method: Not explicitly stated, but inferred to follow the methodology outlined in ISO 81060-2 for comparison against two independent observers' readings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is more relevant for diagnostic imaging devices where human readers interpret images with or without AI assistance. The device in question is a non-invasive blood pressure monitor, and its validation revolves around direct measurement accuracy, not interpretation by human readers.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the clinical validation described as compliant with ISO 81060-2 is inherently a standalone performance test for the device's algorithm. The standard assesses the device's ability to accurately measure blood pressure readings independently against a reference standard. While a human initiates the measurement, the accuracy evaluation focuses on the automated device's output.


    7. The type of ground truth used

    The ground truth for blood pressure monitors validated to ISO 81060-2 is established by expert auscultation measurements using a reference standard like a mercury sphygmomanometer. This involves trained observers listening to Korotkoff sounds.

    • Type of ground truth: Expert auscultation measurements (reference standard blood pressure readings).

    8. The sample size for the training set

    The document does not provide information on a training set sample size. This is typical for a medical device cleared via substantial equivalence, especially for a blood pressure monitor. While the device's development likely involved internal data for calibrating the oscillometric algorithm, the 510(k) summary only focuses on the validation against established standards. The ISO 81060-2 standard is for clinical validation, not algorithm training.


    9. How the ground truth for the training set was established

    As no training set information is provided in the document, this cannot be answered. If a training set were used for algorithm development, its ground truth would typically be established similarly to the validation set—i.e., through expert auscultation or other highly accurate reference methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160557

    Validate with FDA (Live)

    Date Cleared
    2016-07-06

    (128 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HL158VA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over the counter device is for adults aged 18 years and older with wrist circumference ranging from 5.3 inches to 7.7 inches (approx. 135 mm to 195 mm) and for home use.

    HLI S8VA detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.

    Besides, the device features a built-in "Bluetooth Data Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth-enabled device. Also, users could simply synchronize the current date and time, and check the battery status of blood pressure monitor by means of DailyChek® application software with the paired Bluetooth-enabled device.

    Device Description

    HL158VA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 5.3 inches to 7.7 inches (approx.13.5 cm to 19.5 cm) and for home use.

    The device will display a symbol wor or to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.

    Besides, HL158VA has a built-in Wrist Position Guide. By holding START/STOP button, the display will illuminate with different icons that are designed to guide you move your wrist. Furthermore, the user can use the Bluetooth Data Transmission function, which provides users an optional choice to log, track and store their measurement data. Also synchronize the current date and time, and check the battery status of blood pressure monitor by means of DailyChek® application software with the paired Bluetooth-enabled device.

    AI/ML Overview

    The document discusses the Food and Drug Administration's (FDA) review of the Full Automatic (NIBP) Blood Pressure Monitor, Model HL158VA. It determined the device to be substantially equivalent to previously marketed predicate devices.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The primary clinical performance acceptance criteria for the blood pressure monitor are related to its accuracy, as per ISO 81060-2.

    Acceptance Criteria (from ISO 81060-2:2013)Reported Device Performance (HL158VA)
    Mean difference limits for blood pressure measurementsFulfilled (results of clinical investigation)
    Standard deviation limits for blood pressure measurementsFulfilled (results of clinical investigation)
    Pressure Accuracy: ± 3mmHgAchieved
    Pulse Accuracy: ± 5%Achieved

    Additional Technical Specifications:

    FeaturePredicate Device (HL158BT, K093824)Subject Device (HL158VA)Acceptance/Performance
    Method of MeasurementOscillometricOscillometricSame as predicate, accepted
    Measurement TypeDuring deflationDuring inflationModified, but validated clinically
    Range of measurement (Pressure)0~300 mmHgRated Range of Cuff Pressure: 0300mmHg, Rated Range of Determination: 40280mmHgAccepted
    Range of measurement (Pulse)40~199 Beats/minute40~199 Beats/minuteSame as predicate, accepted
    InflationAutomaticAutomaticSame as predicate, accepted
    DeflationAutomatic (Rapid Exhaust Valve)Automatic (Passive Exhaust Valve)Modified, but presumably validated
    Pressure Changed Rate2-5 mmHg/sec. (from 150mmHg to 90mmHg)2-5 mmHg/sec. (from 90mmHg to 150mmHg)Modified, but acceptable.
    DisplayLiquid Crystal DigitalLiquid Crystal DigitalSame as predicate, accepted
    Memory2*60, total 1202*120, total 240Improved, accepted
    Number of Push Button7 keys6 keysModified, accepted
    Cuff size (Wrist circumference)135 ~ 195 mm (5.3~7.7 inches)135 ~ 195 mm (5.3~7.7 inches)Same as predicate, accepted
    Risk Category IndicatorYesYesSame as predicate, accepted
    Irregular Heartbeat DetectorYesYesSame as predicate, accepted
    Data Link FunctionYes (Via USB cable)Yes (Via Bluetooth)Modified, validated non-clinically
    Talking FunctionYesNoRemoved, accepted
    Wrist Position GuideNoYesAdded, validated

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: A "group of 100 subjects" was used for the clinical investigation.
    • Data Provenance: The document does not specify the country of origin. The study was a prospective clinical investigation, designed to evaluate the device against a known standard.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not provide details on the number or qualifications of experts used to establish the "ground truth" (reference blood pressure measurements). In ISO 81060-2 clinical validation, this typically involves trained observers taking auscultatory measurements using a mercury sphygmomanometer in a standardized manner.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not explicitly describe an adjudication method for the test set. For ISO 81060-2 clinical validation, measurements from multiple trained observers are often compared to minimize individual observer bias, but a formal "adjudication" in the sense of resolving discrepancies is not typically described in summary documents like this.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not performed, nor is it applicable to this device. This device is a standalone blood pressure monitor and does not involve human readers analyzing medical images with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance evaluation was conducted. The clinical investigation directly evaluated the accuracy of the device (HL158VA) against the ISO 81060-2 standard, which assesses the device's ability to measure blood pressure without human intervention influencing the measurement itself (though a human operates the device).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the clinical validation was based on reference blood pressure measurements taken by trained observers, as specified by the requirements of the ISO 81060-2 standard for non-invasive sphygmomanometers. This is a form of "expert measurement" or "reference method."

    8. The sample size for the training set

    The document does not mention a training set sample size. For medical devices like this, which measure physiological parameters using established principles (oscillometric method), a "training set" in the machine learning sense is not typically used or reported for regulatory submissions; rather, the device's algorithms are developed and then validated through clinical studies.

    9. How the ground truth for the training set was established

    As no training set (in the machine learning sense) is mentioned, there's no information on how its ground truth was established. The device's underlying algorithm is based on the well-established oscillometric method, and its accuracy is validated directly on a test set (the 100 subjects) rather than through a separate training/validation split typical of AI systems.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161156

    Validate with FDA (Live)

    Date Cleared
    2016-06-14

    (50 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HL858TA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.

    HL858TA detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.

    Device Description

    HL858TA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.

    The device will display a symbol ( heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.

    AI/ML Overview

    The provided document describes the K161156 premarket notification for the Full Automatic (NIBP) Blood Pressure Monitor, Model HL858TA. This device is a non-invasive blood pressure measurement system.

    Here's the breakdown of the acceptance criteria and study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The key performance acceptance criteria for non-invasive blood pressure monitors are typically defined by recognized standards like ISO 81060-2. The document explicitly states compliance with ISO 81060-2. This standard specifies limits for the mean difference and standard deviation between the device's measurements and reference measurements.

    While the exact numerical thresholds specified by ISO 81060-2 are not explicitly listed in this FDA summary document, the summary states:

    Acceptance Criteria (Based on ISO 81060-2)Reported Device Performance (HL858TA)
    Required limits for mean difference and standard deviation (between device & reference measurement) for blood pressure and pulse rate (as per ISO 81060-2)Fulfilled by the subject device HL858TA

    In addition to the clinical validation, the device also meets the following technical performance criteria, which are often derived from predicate device performance and relevant standards:

    Acceptance CriteriaReported Device Performance
    Pressure Measurement Range: 0-300mmHgSame (0-300mmHg) as predicate device
    Pulse Measurement Range: 40-199 Beats/minuteSame (40-199 Beats/minute) as predicate device
    Pressure Accuracy: ± 3mmHgSame (± 3mmHg) as predicate device
    Pulse Accuracy: ± 5%Same (± 5%) as predicate device

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: 105 subjects (used for the clinical investigation compliant with ISO 81060-2).
    • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. Clinical validation studies for these types of devices are typically prospective in nature, as they involve actively taking measurements on subjects under controlled conditions to compare against a reference method.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not explicitly state the number of experts or their qualifications used to establish the ground truth. However, for a clinical validation compliant with ISO 81060-2, the ground truth is established by simultaneous measurements using a reference standard, usually by trained operators (e.g., medical professionals using a mercury sphygmomanometer or an equivalent calibrated method). The standard specifies the requirements for these reference measurements and the personnel performing them.

    4. Adjudication method for the test set

    The document does not mention an adjudication method in the context of human readers or image interpretation, as this device measures physiological parameters directly. For blood pressure measurements, the "adjudication" is inherent in the direct comparison against a reference standard measurement, often performed simultaneously by a trained human operator.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic imaging devices where human interpretation is involved. The HL858TA is a direct measurement device and does not involve human interpretation of complex images aided by AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of the algorithm without human interaction. For this device, the "algorithm" is the oscillometric method and its processing of the pressure signals to determine BP and heart rate. The clinical validation described in the document (ISO 81060-2) directly assesses the performance of this algorithm in a standalone manner (albeit with a human applying the cuff and initiating the measurement), comparing its output to a reference standard. Therefore, a standalone performance assessment was done.

    7. The type of ground truth used

    The ground truth used was simultaneous measurements by a validated reference method (e.g., mercury sphygmomanometer) as per the requirements of ISO 81060-2. This is a form of direct physiological measurement ground truth, not pathology, outcomes data, or expert consensus (in the sense of interpreting subjective data).

    8. The sample size for the training set

    The document does not specify the sample size for a training set. This is because the device is not based on machine learning in the conventional sense (where a model is "trained" on a large dataset). It uses an established oscillometric method. Its calibration and internal algorithms are based on fundamental principles of blood pressure measurement, rather than a data-driven training process that requires a distinct "training set."

    9. How the ground truth for the training set was established

    As there is no "training set" in the machine learning sense, there is no ground truth established for it. The device's underlying principles are based on established biophysics and engineering, validated through rigorous clinical testing on a test set (as described in point 2).

    Ask a Question

    Ask a specific question about this device

    K Number
    K160336

    Validate with FDA (Live)

    Date Cleared
    2016-03-29

    (50 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HL158TA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with wrist circumference ranging from 5.3 inches to 7.7 inches (approx.135 mm to 195 mm) and for home use.

    HL158TA detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.

    Device Description

    HL158TA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with wrist circumference ranging from 5.3 inches to 7.7 inches (approx.135 mm to 195 mm) and for home use.

    The device will display a symbol ( heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Device: Full Automatic (NIBP) Blood Pressure Monitor, Model HL158TA
    Predicate Device: Full Automatic (NIBP) Blood Pressure Monitor, Model HL158BT (K093824)


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Accuracy (Pressure)± 3mmHg
    Accuracy (Pulse)± 5%
    Clinical ValidationMet required limits for mean difference and standard deviation in accordance with ISO 81060-2:2013
    EMC ComplianceCompliant with IEC 60601-1-2 Edition 3:2007-03
    Safety ComplianceCompliant with IEC 60601-1:2005, IEC 60601-1-11:2010
    FCC ComplianceCompliant with FCC 47 CFR Part 15, Subpart B
    BiocompatibilityCompliant with ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010
    ReliabilityCompliant with IEC 80601-2-30 Edition 1.1 2013-07
    Risk AssessmentCompliant with ISO 14971:2007
    Software V&VCompliant with IEC 62304 Ed.1.0 (2006), IEC 60601-1-4
    Usability ValidationCompliant with IEC 62366-1:2015, IEC 60601-1-6

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 85 subjects for the clinical investigation (ISO 81060-2).
    • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. It only states that the clinical investigation was performed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not provide information about the number of experts or their qualifications for establishing the ground truth in the clinical study. However, the ISO 81060-2 standard for non-invasive sphygmomanometers typically requires measurements by trained observers or reference devices to establish ground truth.


    4. Adjudication Method for the Test Set

    The document does not explicitly state the adjudication method used for the test set in the clinical study. ISO 81060-2 standards involve comparison against auscultatory blood pressure measurements, often performed by multiple trained observers to mitigate error.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The study described is a clinical validation study against a recognized standard (ISO 81060-2) for objective performance, not a study comparing human reader performance with and without AI assistance. The device is an automated blood pressure monitor, not an AI diagnostic tool that assists human readers.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the primary clinical study described for the HL158TA is a standalone performance study of the device's accuracy in measuring blood pressure, as required by ISO 81060-2. This standard evaluates the algorithm and hardware's ability to accurately measure blood pressure independently of human interpretation of readings, although human subjects are part of the measurement process. The device provides "Full Automatic" measurements.


    7. The Type of Ground Truth Used

    For the clinical study (ISO 81060-2), the ground truth for blood pressure measurements is typically established through auscultatory measurements performed by trained observers using a reference sphygmomanometer. This is the standard reference method against which automated devices are validated.


    8. The Sample Size for the Training Set

    The document does not provide information on the sample size used for the training set. It focuses on the clinical validation (test set) for regulatory submission.


    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established, as a training set is not explicitly mentioned or detailed in the regulatory submission information provided. This information is typically proprietary to the device manufacturer and not always included in 510(k) summaries, which focus on safety and effectiveness demonstration rather than development details.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160145

    Validate with FDA (Live)

    Date Cleared
    2016-03-13

    (52 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HL858CC automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.

    HL858CC detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.

    Besides, the device features a built-in "Bluetooth Data Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth-enabled device. Also, user could simply synchronize the current date and time, and check the battery status of blood pressure monitor by means of DailyChek® application software with the paired Bluetooth-enabled device.

    Device Description

    HL858CC automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.

    The device will display a symbol wor ma to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.

    Besides, When Triple Check mode is turned on by user, the symbol (& THE) will display on the LCD. Then press Start/Stop button the device will take three consecutive measurements automatically at 1 minute intervals. After measurements are completed, LCD will display the average values of the three measurements. And this device is designed with Rest Assure function as a countdown timer to help user in relax state for 5 minutes before taking measurement. Furthermore, the user can use the Bluetooth Data Transmission function, which provides users an optional choice to log, track and store their measurement data. Also synchronize the current date and time, and check the battery status of blood pressure monitor by means of DailyChek® application software with the paired Bluetooth-enabled device.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted for the Full Automatic (NIBP) Blood Pressure Monitor, Model HL858CC.

    1. Table of Acceptance Criteria and Reported Device Performance:

    CriterionAcceptance Criteria (Stated)Reported Device Performance
    Blood Pressure AccuracyPressure ± 3mmHg"The required limits for mean difference and standard deviation are fulfilled by the subject device HL858CC..."
    Pulse AccuracyPulse ± 5%"The required limits for mean difference and standard deviation are fulfilled by the subject device HL858CC..."
    Adherence to ISO StandardISO 81060-2: Second Edition 2013-05-01 compliant"HL858CC is compliant to the standard of ISO 81060-2: Second Edition 2013-05-01..."
    Functional PerformanceMeasures Systolic, Diastolic BP, Heart Rate via oscillometric method during inflation, detects irregular heartbeats, Risk Category Indicator, Bluetooth data transmission.The device performs these functions as described in the "Device Description" and "Intended Use" sections. The study confirms compliance with standards.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: 105 subjects.
    • Data Provenance: The document does not specify the country of origin of the data. The study is a clinical investigation, indicating it is likely prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    The document does not explicitly state the number of experts or their specific qualifications for establishing ground truth for the clinical validation. It mentions "all the relevant activities were performed by designate individual(s)", which is vague. In typical blood pressure monitor validation studies (like ISO 81060-2), ground truth is established by trained professionals (e.g., clinicians, nurses) using a reference sphygmomanometer.

    4. Adjudication Method for the Test Set:

    The document does not explicitly describe an adjudication method like 2+1 or 3+1. For blood pressure clinical validation according to ISO 81060-2, multiple measurements are typically taken by the reference method and the device, and the differences are calculated. The standard defines the statistical analysis (mean difference and standard deviation) to determine accuracy, which implicitly involves comparing the device's readings against highly accurate manual readings.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is a clinical validation of the device's accuracy against a reference standard, not a comparison of human reader performance with and without AI assistance.

    6. Standalone Performance:

    Yes, a standalone performance study was done. The clinical investigation directly assessed the accuracy of the HL858CC device itself against a clinical reference standard, without human-in-the-loop assistance. The results demonstrated that the device alone met the predetermined acceptance criteria for accuracy.

    7. Type of Ground Truth Used:

    The ground truth used for the clinical investigation (ISO 81060-2) would be reference blood pressure measurements, typically obtained using a mercury sphygmomanometer or another validated reference oscillometric device by trained clinical personnel. The document does not explicitly state "mercury sphygmomanometer" but refers to "Clinical validation of automated measurement type," which implies comparison against a gold standard method.

    8. Sample Size for the Training Set:

    The document does not provide any information regarding a training set or its sample size. This is a clinical validation study for a medical device (blood pressure monitor), not a machine learning model that typically requires a separate training set.

    9. How the Ground Truth for the Training Set Was Established:

    Since no training set is mentioned (as the device is not described as involving a machine learning model that would require one for development), this information is not applicable/provided. The study focuses on clinical validation of the device's accuracy.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153214

    Validate with FDA (Live)

    Date Cleared
    2016-01-07

    (63 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HL858DI automatically measures human's Systolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.

    HL858DI detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator with the readings on the screen for the user tracking their blood pressure level.

    Device Description

    HL858DI automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.

    The device will display a symbol ( ** or ma), to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.

    AI/ML Overview

    The provided document describes the Health & Life Co., Ltd. Full Automatic (NIBP) Blood Pressure Monitor, Model HL858DI, and its clinical validation according to ISO 81060-2.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for non-invasive sphygmomanometers are typically defined by standards like ISO 81060-2 for blood pressure measurement accuracy. The document states that the device is compliant with ISO 81060-2: Second Edition 2013-05-01. This standard sets specific requirements for mean difference and standard deviation between the device's readings and a reference measurement.

    Although the exact numerical acceptance criteria for mean difference and standard deviation from ISO 81060-2 are not explicitly listed in the document, it states that:

    Acceptance Criteria (Based on ISO 81060-2)Reported Device Performance
    Required limits for mean differenceFulfilled
    Required limits for standard deviationFulfilled

    Note: The specific numerical thresholds for "required limits" would be found within the ISO 81060-2 standard itself, which is referenced but not fully reproduced here.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 105 subjects.
    • Data Provenance: The document does not explicitly state the country of origin. It indicates it was a "clinical investigation," implying prospective data collection during the study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of experts or their specific qualifications for establishing the ground truth measurements during the clinical investigation. However, for a study compliant with ISO 81060-2, it is implicit that trained observers (often physicians or qualified medical personnel) follow a standardized protocol (e.g., using a mercury sphygmomanometer or validated auscultatory method) to provide the reference blood pressure readings against which the device's measurements are compared.

    4. Adjudication Method for the Test Set

    The document does not describe a specific adjudication method beyond stating that the relevant activities were performed by "designate individual(s)." In the context of ISO 81060-2, the ground truth is typically established by multiple trained observers taking readings independently, and then their measurements are averaged or assessed for agreement, but the specific details are not provided here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is a standalone clinical validation of the device's accuracy against a reference standard, not a comparison of human readers with and without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, a standalone study was done. The clinical investigation directly assessed the performance of the Full Automatic (NIBP) Blood Pressure Monitor, Model HL858DI, which operates as an automated device (algorithm only for measurement capture), against a human-obtained reference standard.

    7. The Type of Ground Truth Used

    The ground truth used was reference blood pressure measurements obtained from qualified personnel during the clinical investigation, following established protocols for accuracy in blood pressure measurement (as implied by compliance with ISO 81060-2). This is typically auscultatory measurements by trained observers using a mercury sphygmomanometer or another validated reference method.

    8. The Sample Size for the Training Set

    The document does not provide information regarding a separate training set or its sample size. The clinical investigation mentioned (with 105 subjects) appears to be the test set for validating the device's performance according to ISO 81060-2. As this is a blood pressure monitor, the "training" of its internal algorithm would likely occur during its development phase using various datasets, but these details are not part of this submission document.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is explicitly mentioned, the method for establishing its ground truth is also not provided in this document.