K Number
K160557
Date Cleared
2016-07-06

(128 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HL158VA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over the counter device is for adults aged 18 years and older with wrist circumference ranging from 5.3 inches to 7.7 inches (approx. 135 mm to 195 mm) and for home use.

HLI S8VA detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.

Besides, the device features a built-in "Bluetooth Data Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth-enabled device. Also, users could simply synchronize the current date and time, and check the battery status of blood pressure monitor by means of DailyChek® application software with the paired Bluetooth-enabled device.

Device Description

HL158VA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 5.3 inches to 7.7 inches (approx.13.5 cm to 19.5 cm) and for home use.

The device will display a symbol wor or to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.

Besides, HL158VA has a built-in Wrist Position Guide. By holding START/STOP button, the display will illuminate with different icons that are designed to guide you move your wrist. Furthermore, the user can use the Bluetooth Data Transmission function, which provides users an optional choice to log, track and store their measurement data. Also synchronize the current date and time, and check the battery status of blood pressure monitor by means of DailyChek® application software with the paired Bluetooth-enabled device.

AI/ML Overview

The document discusses the Food and Drug Administration's (FDA) review of the Full Automatic (NIBP) Blood Pressure Monitor, Model HL158VA. It determined the device to be substantially equivalent to previously marketed predicate devices.

Here's an analysis of the acceptance criteria and the study proving the device meets them:

1. A table of acceptance criteria and the reported device performance

The primary clinical performance acceptance criteria for the blood pressure monitor are related to its accuracy, as per ISO 81060-2.

Acceptance Criteria (from ISO 81060-2:2013)Reported Device Performance (HL158VA)
Mean difference limits for blood pressure measurementsFulfilled (results of clinical investigation)
Standard deviation limits for blood pressure measurementsFulfilled (results of clinical investigation)
Pressure Accuracy: ± 3mmHgAchieved
Pulse Accuracy: ± 5%Achieved

Additional Technical Specifications:

FeaturePredicate Device (HL158BT, K093824)Subject Device (HL158VA)Acceptance/Performance
Method of MeasurementOscillometricOscillometricSame as predicate, accepted
Measurement TypeDuring deflationDuring inflationModified, but validated clinically
Range of measurement (Pressure)0~300 mmHgRated Range of Cuff Pressure: 0300mmHg, Rated Range of Determination: 40280mmHgAccepted
Range of measurement (Pulse)40~199 Beats/minute40~199 Beats/minuteSame as predicate, accepted
InflationAutomaticAutomaticSame as predicate, accepted
DeflationAutomatic (Rapid Exhaust Valve)Automatic (Passive Exhaust Valve)Modified, but presumably validated
Pressure Changed Rate2-5 mmHg/sec. (from 150mmHg to 90mmHg)2-5 mmHg/sec. (from 90mmHg to 150mmHg)Modified, but acceptable.
DisplayLiquid Crystal DigitalLiquid Crystal DigitalSame as predicate, accepted
Memory2*60, total 1202*120, total 240Improved, accepted
Number of Push Button7 keys6 keysModified, accepted
Cuff size (Wrist circumference)135 ~ 195 mm (5.3~7.7 inches)135 ~ 195 mm (5.3~7.7 inches)Same as predicate, accepted
Risk Category IndicatorYesYesSame as predicate, accepted
Irregular Heartbeat DetectorYesYesSame as predicate, accepted
Data Link FunctionYes (Via USB cable)Yes (Via Bluetooth)Modified, validated non-clinically
Talking FunctionYesNoRemoved, accepted
Wrist Position GuideNoYesAdded, validated

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: A "group of 100 subjects" was used for the clinical investigation.
  • Data Provenance: The document does not specify the country of origin. The study was a prospective clinical investigation, designed to evaluate the device against a known standard.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

The document does not provide details on the number or qualifications of experts used to establish the "ground truth" (reference blood pressure measurements). In ISO 81060-2 clinical validation, this typically involves trained observers taking auscultatory measurements using a mercury sphygmomanometer in a standardized manner.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not explicitly describe an adjudication method for the test set. For ISO 81060-2 clinical validation, measurements from multiple trained observers are often compared to minimize individual observer bias, but a formal "adjudication" in the sense of resolving discrepancies is not typically described in summary documents like this.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A multi-reader multi-case (MRMC) comparative effectiveness study was not performed, nor is it applicable to this device. This device is a standalone blood pressure monitor and does not involve human readers analyzing medical images with or without AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance evaluation was conducted. The clinical investigation directly evaluated the accuracy of the device (HL158VA) against the ISO 81060-2 standard, which assesses the device's ability to measure blood pressure without human intervention influencing the measurement itself (though a human operates the device).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for the clinical validation was based on reference blood pressure measurements taken by trained observers, as specified by the requirements of the ISO 81060-2 standard for non-invasive sphygmomanometers. This is a form of "expert measurement" or "reference method."

8. The sample size for the training set

The document does not mention a training set sample size. For medical devices like this, which measure physiological parameters using established principles (oscillometric method), a "training set" in the machine learning sense is not typically used or reported for regulatory submissions; rather, the device's algorithms are developed and then validated through clinical studies.

9. How the ground truth for the training set was established

As no training set (in the machine learning sense) is mentioned, there's no information on how its ground truth was established. The device's underlying algorithm is based on the well-established oscillometric method, and its accuracy is validated directly on a test set (the 100 subjects) rather than through a separate training/validation split typical of AI systems.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern. Inside the circle is a stylized image of an eagle or bird-like figure, with its wings forming a flowing, abstract design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 6, 2016

Health & Life Co., Ltd. Sarah Su Director, OA & RA Division 9F, No.186, Jian Yi Road Zhonghe District, New Taipei City, TW 23553 Taiwan

Re: K160557

Trade/Device Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL158VA Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: June 7, 2016 Received: June 7, 2016

Dear Sarah Su:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mude Yellen
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K160557

Device Name

Full Automatic (NIBP) Blood Pressure Monitor, Model HL158VA

Indications for Use (Describe)

HL158VA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over the counter device is for adults aged 18 years and older with wrist circumference ranging from 5.3 inches to 7.7 inches (approx. 135 mm to 195 mm) and for home use.

HLI S8VA detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.

Besides, the device features a built-in "Bluetooth Data Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth-enabled device. Also, users could simply synchronize the current date and time, and check the battery status of blood pressure monitor by means of DailyChek® application software with the paired Bluetooth-enabled device.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

PREMARKET NOTIFICATION

510(k) SUMMARY

(As Required By 21 CFR 807.92)

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: Date:

1. Submitter:

Health & Life Co., Ltd. 9F. No.186, Jian Yi Road, Zhonghe District, New Taipei City, Taiwan, R.O.C TEL: +886-2-8227-1300 FAX: +886-2-8227-1301

Contact person: Sarah Su/ RA & QA Division. E-mail: sarah.su@hlmt.com.tw Tel: 886-2-8227-1300 ext.1201 Fax: 886-2-8227-1301

2. Name of the Device:

Trade Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL158VA Common Name: Blood Pressure Monitor Classification Name: Non-invasive Blood Pressure Measurement System Classification: Class II, 21 CFR 870.1130 Classification Panel: 74 Cardiovascular Product Code: DXN

3. Information for the 510(k) Cleared Device (Predicate Device):

A. Full Automatic (NIBP) Blood Pressure Monitor, Model: HL158BT (K093824)

B. Omron BP652N (HEM-6300-Z) with APS (K142917)

{4}------------------------------------------------

4. Device Description:

HL158VA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 5.3 inches to 7.7 inches (approx.13.5 cm to 19.5 cm) and for home use.

The device will display a symbol wor or to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.

Besides, HL158VA has a built-in Wrist Position Guide. By holding START/STOP button, the display will illuminate with different icons that are designed to guide you move your wrist. Furthermore, the user can use the Bluetooth Data Transmission function, which provides users an optional choice to log, track and store their measurement data. Also synchronize the current date and time, and check the battery status of blood pressure monitor by means of DailyChek® application software with the paired Bluetooth-enabled device.

5. Intended Use

HL158VA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 5.3 inches to 7.7 inches (approx.13.5 cm to 19.5 cm) and for home use.

HL158VA detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.

{5}------------------------------------------------

Besides, the device features a built-in "Bluetooth Data Transmission" function, which the device automatically transmit measuring results to paired enables Bluetooth-enabled device. Also, user could simply synchronize the current date and time, and check the battery status of blood pressure monitor by means of DailyChek application software with the paired Bluetooth-enabled device.

6. Comparison of device to predicate device: A. Product Specification Comparison Table of Subject Device HL158VA, and

Predicate Device HL158BT (K093824)
ItemPredicate DeviceHL158BT (K093824)Subject DeviceHL158VA
Method ofmeasurementOscillimetricSame as left
Measurement TypeDuring deflationDuring inflation
Range ofmeasurementPressure: 0~ 300mmHg,Pulse 40~199Beats/minuteRated Range of CuffPressure: 0~ 300mmHg,Rated Range ofDetermination:40280mmHg,Pulse 40199 Beats/minute
AccuracyPressure ± 3mmHgPulse ± 5%Same as left
InflationAutomaticSame as left
DeflationAutomatic(Rapid Exhaust Valve)Automatic(Passive Exhaust Valve)
PressureChanged Rate2-5 mmHg/sec.(from 150mmHg to 90mmHg)2-5 mmHg/sec.(from 90mmHg to 150mmHg)
DisplayLiquid Crystal DigitalSame as left
Power Supply2 "AAA (1.5V)"Alkaline Batteries2 "AAA (LR03)(1.5V)"Alkaline Batteries
Storage/- 20℃ ~ + 70℃,- 25°C ~ + 70°C
TransportationEnvironment(-4°F~ +158°F),≤90%R.H.(-13°F~ +158°F),≤ 93% R.H.
OperatingEnvironment10℃40℃,(50°F104°F),15%~90%R.H.5°C ~ 40°C(41°F104°F),15% ~ 93% R.H.7001060 hPa
MaterialABS housing andrubber keysABS housing andABS keys
Sets of memory2*60, total 1202*120, total 240
Number ofPush Button7 keys(Start/Stop,M(Memory),Date/Time ( $\Theta$ ),Set (+), User( $\frac{\text{\textcelsius}}{\text{\textcelsius}}$ ),Language,Talking Mode button )6 keys(Start/Stop,M(Memory),Date/Time ( $\Theta$ ), Set(+), User( $\frac{\text{\textcelsius}}{\text{\textcelsius}}$ ),Bluetoothbutton ( $\Theta$ ))
Storage pouchYes(Storage Case)Yes(Storage Pouch)
Cuff sizeWrist circumferenceapprox.135 ~ 195 mm(Approx. 5.3~7.7 inches)Same as left
Unit Weight(excluding batteries)Approx. 136 ± 5 gApprox. 128 ± 10 g
Risk CategoryIndicatorYesSame as left
Irregular HeartbeatDetectorYesSame as left
Data LinkFunctionYes(Via USB cable)Yes(Via Bluetooth)
Talking FunctionYesNo
Wrist Position GuideNoYes

Predicate Device HL158BT (K093824)

{6}------------------------------------------------

{7}------------------------------------------------

Changes from the predicate devices HL158BT (K093824):

    • Modifying the measurement type from "during deflation" into "during inflation".
    • Changing the memory of 120 (260, total 120) to 240 (2120, total 240).
    • Changing the 7 keys to the 6 keys.
    • Modifying the Data Link Function from USB into Bluetooth Data Transmission Function.
    • Removing the Talking Function.
    • Adding the Wrist Position Guide Function.

These additional features have been verified and validated and do not affect the safety and effectiveness of subject device HL158VA. For the product feature of Wrist Position Guide, was compared with the other predicate device OMRON BP652N. Please refer to Section 12. Substantial Equivalence Discussion for detail information.

7. Discussion of Clinical Tests Performed:

HL158VA is compliant to the standard of ISO 81060-2: Second Edition 2013-05-01 Non-invasive sphygmomanometers- Part 2: Clinical validation of automated measurement type. The results of this clinical investigation show that the required limits for mean difference and standard deviation are fulfilled by the subject device HL158VA in the group of 100 subjects with qualified distribution. Thus, all the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met.

8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:

The subject device was tested to evaluate its safety and effectiveness, including the followings:

  • a. EMC Test: IEC 60601-1-2 Edition 3:2007-03, Medical Electrical Equipment -Part 1-2: General requirements for safety - collateral standard: Electromagnetic compatibility - Requirements and Tests
  • b. Radio Frequency Wireless Test:

-ETSI EN 300 328. Electromagnetic compatibility and Radio spectrum Matters (ERM); Wideband transmission systems; Data transmission equipment operating in the 2.4 GHz ISM band and using wide band modulation techniques;

ર-ર

{8}------------------------------------------------

Harmonized EN covering the essential requirements of article 3.2 of the R&TTE Directive

-ETSI EN 301 489-1, Electromagnetic compatibility and Radio spectrum Matters (ERM); Electromagnetic Compatibility (EMC) standard for radio equipment and services; Part 1: Common technical requirements

-ETSI EN 301 489-17, Electromagnetic compatibility and Radio spectrum Matters (ERM); Electro Magnetic Compatibility (EMC) standard for radio equipment; Part 17: Specific conditions for Broadband Data Transmission Systems

Safety Test: c.

-IEC 60601-1:2005+A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

-IEC 60601-1-11:2010, Medical electrical equipment-Part 1-11: General Requirement for basic safety and essential performance- Collateral Standard: Requirements for medical electrical systems used in the home healthcare environment

d. FCC Test:

FCC 47 CFR Part 15, Subpart B & FCC 47 CFR Part 15, Subpart C

e. Biocompatibility Test:

-ISO 10993-1:2009, Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process

-ISO 10993-5:2009, Biological evaluation of medical devices-Part 5: Tests for In Vitro cytotoxicity

-ISO 10993-10:2010, Third Edition Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization

f. Reliability Test:

IEC 80601-2-30 Edition1.1 2013-07 Medical electrical equipment-Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers

  • Risk Assessment: ISO 14971:2007 Second Edition, Medical devices g. Application of risk management to medical devices
    ર-6

{9}------------------------------------------------

h. Software Verification and Validation:

-IEC 62304 Ed.1.1 (2015), Medical device software - Software life cycle processes,

-IEC 60601-1-4 Medical electrical equipment - Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems, edition 1.1.

Usability Validation: i.

-IEC 62366-1:2015 Medical devices - Part 1: Application of usability engineering to medical devices

-IEC 60601-1-6 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability

9. Conclusions:

The subject device was tested and fulfilled the requirements of those standards mentioned above, and it's concluded that the subject device is substantially equivalent to the predicate device.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).