K Number
K153214

Validate with FDA (Live)

Date Cleared
2016-01-07

(63 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HL858DI automatically measures human's Systolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.

HL858DI detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator with the readings on the screen for the user tracking their blood pressure level.

Device Description

HL858DI automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.

The device will display a symbol ( ** or ma), to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.

AI/ML Overview

The provided document describes the Health & Life Co., Ltd. Full Automatic (NIBP) Blood Pressure Monitor, Model HL858DI, and its clinical validation according to ISO 81060-2.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for non-invasive sphygmomanometers are typically defined by standards like ISO 81060-2 for blood pressure measurement accuracy. The document states that the device is compliant with ISO 81060-2: Second Edition 2013-05-01. This standard sets specific requirements for mean difference and standard deviation between the device's readings and a reference measurement.

Although the exact numerical acceptance criteria for mean difference and standard deviation from ISO 81060-2 are not explicitly listed in the document, it states that:

Acceptance Criteria (Based on ISO 81060-2)Reported Device Performance
Required limits for mean differenceFulfilled
Required limits for standard deviationFulfilled

Note: The specific numerical thresholds for "required limits" would be found within the ISO 81060-2 standard itself, which is referenced but not fully reproduced here.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 105 subjects.
  • Data Provenance: The document does not explicitly state the country of origin. It indicates it was a "clinical investigation," implying prospective data collection during the study.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the number of experts or their specific qualifications for establishing the ground truth measurements during the clinical investigation. However, for a study compliant with ISO 81060-2, it is implicit that trained observers (often physicians or qualified medical personnel) follow a standardized protocol (e.g., using a mercury sphygmomanometer or validated auscultatory method) to provide the reference blood pressure readings against which the device's measurements are compared.

4. Adjudication Method for the Test Set

The document does not describe a specific adjudication method beyond stating that the relevant activities were performed by "designate individual(s)." In the context of ISO 81060-2, the ground truth is typically established by multiple trained observers taking readings independently, and then their measurements are averaged or assessed for agreement, but the specific details are not provided here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is a standalone clinical validation of the device's accuracy against a reference standard, not a comparison of human readers with and without AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, a standalone study was done. The clinical investigation directly assessed the performance of the Full Automatic (NIBP) Blood Pressure Monitor, Model HL858DI, which operates as an automated device (algorithm only for measurement capture), against a human-obtained reference standard.

7. The Type of Ground Truth Used

The ground truth used was reference blood pressure measurements obtained from qualified personnel during the clinical investigation, following established protocols for accuracy in blood pressure measurement (as implied by compliance with ISO 81060-2). This is typically auscultatory measurements by trained observers using a mercury sphygmomanometer or another validated reference method.

8. The Sample Size for the Training Set

The document does not provide information regarding a separate training set or its sample size. The clinical investigation mentioned (with 105 subjects) appears to be the test set for validating the device's performance according to ISO 81060-2. As this is a blood pressure monitor, the "training" of its internal algorithm would likely occur during its development phase using various datasets, but these details are not part of this submission document.

9. How the Ground Truth for the Training Set Was Established

As no training set is explicitly mentioned, the method for establishing its ground truth is also not provided in this document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the caduceus symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 7, 2016

Health & Life Co., Ltd. Sarah Su Director 9F, No. 186, Jian Yi Road Zhonghe District, New Taipei City, TW 23553 Taiwan

Re: K153214

Trade/Device Name: Full Automatic (NIBP) Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 3, 2015 Received: November 5, 2015

Dear Sarah Su:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K153214

Device Name

Full Automatic (NIBP) Blood Pressure Monitor, Model HL858DI

Indications for Use (Describe)

HL858DI automatically measures human's Systolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.

HL858DI detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator with the readings on the screen for the user tracking their blood pressure level.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K153214 page 1 of 6

PREMARKET NOTIFICATION

510(k) SUMMARY

(As Required By 21 CFR 807.92)

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

NOV 0 3 2015 The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________ Date:

1. Submitter:

Health & Life Co., Ltd. 9F. No.186, Jian Yi Road, Zhonghe District, New Taipei City, Taiwan, R.O.C TEL: +886-2-8227-1300 FAX: +886-2-8227-1301

Contact person: Sarah Su/ Regulatory Affairs Dept. E-mail: sarah.su@hlmt.com.tw Tel: 886-2-8227-1300 ext.1201 Fax: 886-2-8227-1301

2. Name of the Device:

Trade Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL858DI Common Name: Blood Pressure Monitor Classification Name: Non-invasive Blood Pressure Measurement System Classification: Class II, 21 CFR 870.1130 Classification Panel: 74 Cardiovascular Product Code: DXN

3. Information for the 510(k) Cleared Device (Predicate Device):

Full Automatic (NIBP) Blood Pressure Monitor, Model: HL858CB (K142968)

{4}------------------------------------------------

4. Device Description:

HL858DI automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.

The device will display a symbol ( ** or ma), to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.

5. Intended Use

HL858DI automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.

HL858DI detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.

6. Comparison of device to predicate device:

Product Specification Comparison Table of Subject Device HL858DL, and Predicate Device HL858CB (K142968)

ItemPredicate DeviceHL858CB (K142968)Subject DeviceHL858DI
Method ofmeasurementOscillimetricSame as left
MeasurementDuring deflationDuring inflation

{5}------------------------------------------------

Type
Range ofmeasurementPressure 0- 300mmHg,Pulse 40-199 Beats/minuteSame as left
AccuracyPressure ± 3mmHgPulse ± 5%Same as left
PressureChanged Rate2~5mmHg/sec.Same as left
ExhaustAutomatic exhaust valveSame as left
DisplayLiquid Crystal DigitalSame as left
Power Supply6V 1A, 4 × AA/1.5V (LR6)Alkaline batteries, orAC adapter (optional)Same as left
Storage/TransportationEnvironment- 25℃ ~ +70℃ (approx. -13°F ~approx. +158°F),≤ 93% R.H.Same as left
OperatingEnvironment5℃ ~ 40℃ (approx. 41°F~approx. 104°F),15% ~ 93% R.H.Same as left
MaterialABS housing andABS keysSame as left
Sets ofmemory2*120, total 2402*60,total 120
Number ofPush Button7 +2 switch control(Triple-Check (Multi-Read),Rest Assure)5 key control
Storage pouchYesSame as left
Cuff sizeArm circumference approx.2343cm / 917 inch(Universal cuff)Same as left
Unit WeightApprox. 393 ± 10g(Excluding cuff and Batteries )Approx. 310 ± 10g(Excluding cuff and Batteries )
Risk CategoryIndicatorYesSame as left
IrregularHeartbeatDetectorYesSame as left
Triple-Check(Multi-Read)FunctionYesNo
Rest AssureFunctionYesNo
Data LinkfunctionYes(Via Bluetooth)No

{6}------------------------------------------------

Changes from the predicate devices HL858CB (K142968):

    • Modifying the measurement type from "during deflation" into "during inflation".
    • Changing the memory of 240 (2120, total 240) to 120 (260, total 120).
    • Changing the 7 key & 2 switch control to the 5 key control.
    • Changing the Unit Weight from Approx. 393 ± 10g to Approx. 310 ± 10g.
    • Removing the Triple-Check (Multi-Read) Function, Rest Assure Function and Bluetooth Data Transmission Function.

These features has been verified and validated and do not affect the safety and effectiveness of subject device HL858DI.Please refer to Section 12. Substantial Equivalence Discussion for detail information.

7. Discussion of Clinical Tests Performed:

HL858DI is compliant to the standard of ISO 81060-2: Second Edition 2013-05-01 Non-invasive sphygmomanometers- Part 2: Clinical validation of automated measurement type. The results of this clinical investigation show that the required limits for mean difference and standard deviation are fulfilled by the subject device HL858DI in the group of 105 subjects with qualified distribution. Thus, all the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met.

8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:

{7}------------------------------------------------

K153214 page 5 of 6

The subject device was tested to evaluate its safety and effectiveness, including the followings:

  • a. EMC Test: IEC 60601-1-2 Edition 3:2007-03, Medical Electrical Equipment -Part 1-2: General requirements for safety - collateral standard: Electromagnetic compatibility - Requirements and Tests
  • b. Safety Test:

-IEC 60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

-IEC 60601-1-11:2010, Medical electrical equipment-Part 1-11: General Requirement for basic safety and essential performance- Collateral Standard: Requirements for medical electrical systems used in the home healthcare environment

  • c. FCC Test:
    FCC 47 CFR Part 15, Subpart B

  • d. Biocompatibility Test:
    -ISO 10993-1:2009, Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process

-ISO 10993-5:2009, Biological evaluation of medical devices-Part 5: Tests for In Vitro cytotoxicity

-ISO 10993-10:2010, Third Edition Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization

  • e. Reliability Test:
    IEC 80601-2-30 Edition1.1 2013-07 Medical electricalequipment-Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers.

  • f. Risk Assessment: ISO 14971:2007 Second Edition, Medical devices -Application of risk management to medical devices

  • g. Software Verification and Validation:

-IEC 62304 Ed.1.0 (2006), Medical device software - Software life cycle processes.

-IEC 60601-1-4 Medical electrical equipment - Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems, edition 1.1

{8}------------------------------------------------

h. Usability Validation:

-IEC 62366-1:2015 Medical devices - Application of usability engineering to medical devices

-IEC 60601-1-6 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability

9. Conclusions:

The subject device was tested and fulfilled the requirements of those standards mentioned above, and it's concluded that the subject device is substantially equivalent to the predicate device.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).