Found 5 results

510(k) Data Aggregation

    K Number
    K242514

    Validate with FDA (Live)

    Date Cleared
    2025-04-24

    (244 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Endo Motor with E-LITE MAX, E-LITE PRO and ELITE-INO is cordless endodontic treatment motorized handpiece with root canal measurement capability. It can be used for preparation and enlargement of root canals, or measuring the canal length. And it can be used to enlarge the canals while monitoring the position of the file tip inside the canal.

    Device Description

    E-LITE MAX & ELITE-INO:

    The Endo Motor with E-LITE MAX and ELITE-INO is a low-speed rotating oral equipment mainly used for root canal preparation and root canal measurement. It is designed to enlarge the root canal, facilitate dentin filling, or measuring the root canal length. The product is a portable device consist of a cordless hand-piece and a charge base, it is powered by built-in lithium batteries and charged by the adapter. LCD displays parameters such as speed, torque, file system, Apex etc. Users can also set and modify by keys, and provide design of factory initialization, and calibration of the apex location. It can be used as a low-speed motorized handpiece and device for measuring canal length. The rotary speed and torque range are adjustable.

    There are three function features for E-LIT MAX and ELITE-INO:

    • Root Canal enlargement: Prepare the root canal with apex locator function.
    • Apex Location: Measure the length of the root canal.
    • Multi-function: Measuring the length while root canal preparation.

    The device can save 9 programs, each program can set different parameters according to the user.

    E-LITE PRO:

    The Endo Motor with E-LITE PRO is a low-speed rotating oral equipment mainly used for root canal preparation. It is designed to enlarge the root canal or facilitate dentin filling. The product is a portable device having a cordless hand-piece with a charging adapter, it is powered by built-in lithium batteries and charged by the adapter. LCD displays parameters such as speed, torque, file system, Apex etc. Users can also set and modify by keys, and provide design of factory initialization, and calibration of the apex location. It can be used as a low-speed motorized handpiece and device for measuring canal length. The rotary speed and torque range are adjustable.

    There is only one function features for E-LITE PRO:

    • Root Canal enlargement: Prepare the root canal, without apex locator function.

    The device can save 9 programs, each program can set different parameters according to the user.

    The Endo Motor must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

    The Endo Motor is intended to be sterilized prior to use.

    The compatible device can be used with Endo Motor is rotary instruments such as rotary root canal files, the working part of rotary instruments normally made of Nickel titanium (NiTi). These files are not included in the submission.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the "Endo Motor" (K242514) indicate that this is a dental handpiece and accessories, specifically an endodontic motor with root canal measurement capabilities. This type of device is not an AI/ML-enabled medical device and therefore the provided document does not contain the information typically found in an FDA submission for AI/ML devices regarding acceptance criteria, training data, ground truth establishment, or clinical study design (e.g., MRMC studies).

    The document states: "No clinical test data was used to support the decision of substantial equivalence." This further confirms that a study validating an AI/ML component was not part of this clearance.

    The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the performance of a traditional medical device (an endomotor) against established electrical safety, EMC, usability, and functional standards, as opposed to the performance of an AI/ML algorithm.

    Therefore, I cannot provide the requested information related to AI/ML device validation because it is not present in the provided document. The document describes compliance with recognized standards for traditional medical devices (e.g., IEC 60601-1 for electrical safety, ISO 10993-1 for biocompatibility).

    However, I can extract the general "acceptance criteria" and "device performance" as understood for this type of non-AI device from the provided tables, focusing on the comparative aspects:


    Acceptance Criteria and Device Performance (Based on traditional device comparison):

    The acceptance criteria for the Endo Motor (E-LITE MAX, E-LITE PRO, ELITE-INO) are primarily established through a direct comparison to a predicate device (Guilin Woodpecker Medical Instrument Co., Ltd.'s Endo Motor, K203320) and compliance with relevant international standards for medical devices. The objective is to demonstrate that the proposed device is "substantially equivalent" in terms of safety and effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/CriterionAcceptance Criteria (from Predicate Device or Standards)Reported Device Performance (Proposed Device)Judgment
    General
    Device ClassClass 1Class 1SE
    Prescription UsePrescription UsePrescription UseSE
    Anatomical SitesRoot canal, Softened dentinRoot canal, Softened dentinSE
    Target PopulationPatients in need of root canal surgeryPatients in need of root canal surgerySE
    Where UsedDental Clinic, University, Hospital, other clinical settingsDental Clinic, University, Hospital, other clinical settingsSE
    Intended UseCordless endodontic treatment motorized handpiece with root canal measurement capability. Can be used for preparation and enlargement of root canals, or measuring canal length, and enlarging while monitoring file tip position.Identical to predicate device.SE
    Safety & Performance
    Electrical SafetyCompliance with IEC 60601-1Compliance with IEC 60601-1SE
    EMCCompliance with IEC 60601-1-2Compliance with IEC 60601-1-2SE
    Usability EngineeringCompliance with IEC 62366-1Compliance with IEC 62366-1SE
    SoftwareCompliance with IEC 62304Compliance with IEC 62304SE
    BiocompatibilityCompliance with ISO 10993-1Compliance with ISO 10993-1SE
    Power SupplyLithium battery (DC3.7)Lithium battery (DC3.7)SE
    Charger Power SupplyAC100V-240VAC100V-240VSE
    Frequency of Supply Voltage50Hz ~ 60Hz50Hz ~ 60HzSE
    ComponentsMeasuring wire, File clip, Lip hook, Touch probe, Tester, Spray NozzleMeasuring wire, File clip, Lip hook, Touch probe, Tester, Spray NozzleSE
    Safety MechanismAutomatically stops or reversely rotates as soon as the file reaches the apical stop, to prevent perforation.Automatically stops or reversely rotates as soon as the file reaches the apical stop, to prevent perforation.SE
    Micro MotorBrushlessBrushlessSE
    USB PortsYesYesSE
    Operation ModeContinuous operationContinuous operationSE
    Key Performance Parameters (with noted differences)
    Dimension (Handpiece)Ai-Motor: 28x33.7x199.2mm; Endo Radar Plus: 27.8x29x212mmE-LITE MAX: φ32x150mm; E-LITE PRO: φ30x151mm; ELITE-INO: φ30x200mmNote 1 (ergonomic design doesn't compromise safety/performance)
    Weight (Handpiece)Ai-Motor: 150g; Endo Radar Plus: 151gE-LITE MAX: 110g; E-LITE PRO: 120g; ELITE-INO: 175gNote 1 (weight reduction doesn't compromise structural integrity/performance)
    Torque Range0.4 Ncm ~ 5 NcmE-LITE MAX: 0.63.9N·cm; ELITE-INO: 0.64.0N·cm; E-LITE PRO: 0.6~3.9N·cmNote 2 (adjustable by operator; device complies with IEC80601-2-60 and has auto-reverse/stop)
    Rotate Speed100 rpm~1200 rpmE-LITE MAX: 150800 rpm; ELITE-INO: 1201200 rpm; E-LITE PRO: 150~800 rpmNote 3 (within predicate range; does not affect safety/effectiveness)
    Primary Contra Angle Gear Ratio6:1E-LITE MAX: 16:1; ELITE-INO: 6:1; E-LITE PRO: 1:1Note 4 (tested; does not affect safety/effectiveness)

    2. Sample Size Used for the Test Set and Data Provenance

    This document does not describe a "test set" in the context of an AI/ML model's performance on a dataset of patient cases. Instead, "tests" refer to engineering and biocompatibility validations of the physical device. Therefore, typical sample size metrics for AI/ML validation are not applicable here. The data provenance is implied by the testing standards (e.g., in-house testing by the manufacturer based on international standards). The nature of this submission indicates a retrospective comparison to an already cleared predicate device, rather than a prospective clinical trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is not an AI/ML device and does not involve human expert adjudication of medical image data for ground truth establishment.

    4. Adjudication Method for the Test Set

    Not applicable for the reasons stated above.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. The document explicitly states: "No clinical test data was used to support the decision of substantial equivalence." This type of study is specifically relevant to AI/ML devices where the goal is to show the impact of the AI on human reviewer performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical endodontic handpiece, not a software algorithm.

    7. The type of ground truth used

    For this conventional medical device, "ground truth" refers to the established performance requirements defined by recognized standards (e.g., IEC, ISO) and the functional specifications of the predicate device. Compliance is demonstrated through laboratory testing of the physical device, rather than through expert consensus on medical data or pathology results.

    8. The sample size for the training set

    Not applicable, as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable, as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220829

    Validate with FDA (Live)

    Device Name
    Endo Motor
    Date Cleared
    2023-05-18

    (422 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    C-SMART-I PRO, C-SMART-I PILOT, C-SMART-MINI AP:
    The Endo Motor device is an endodontic treatment motorized handpiece with root canal measurement capability. It can be used to enlarge the canals while monitoring the position of the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

    C-Smart-mini:
    The Endo Motor device is an endodontic treatment motorized handpiece. It can be used to enlarge canals. It can be used as a low-speed motorized handpiece. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

    Device Description

    C-SMART-I PRO: The Endo Motor device C-SMART-I PRO is a low-speed rotating oral equipment mainly used for root canal preparation and root canal measurement. The product is a portable device powered by built-in lithium batteries and charged by the adapter. LCD displays parameters such as speed, torque, working mode, apex position, etc. Users can also set and modify by keys, and provide design of factory initialization, calibration and functional check of the apex locator. The Endo Motor device is an endodontic treatment motorized handpiece with root canal measurement capability. It can be used to enlarge the canals while monitoring the position of the file tip inside the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel. The Endo Motor device C-SMART-I PRO is intended to be sterilized prior to use. The material composition of root canal files that can be used with the Endo Motor device C-SMART-I PRO is Nickel titanium (NiTi). These files are not included in the submission. There are three working modes as: M1: Apex locator only. M2: Motor only. M3: (Dual mode) Motor with apex location function.

    C-SMART-I PILOT: The Endo Motor device C-SMART-I PILOT is a low-speed rotating oral equipment mainly used for root canal preparation and root canal measurement. The product is a portable device powered by built-in lithium batteries and charged by the adapter. LCD displays parameters such as speed, torque, working mode, apex position, etc. Users can also set and modify by keys, and provide design of function check, calibration, factory reset and can connect handpiece to control unit via Bluetooth. The Endo Motor device is an endodontic treatment motorized handpiece with root canal measurement capability. It can be used to enlarge the canals while monitoring the position of the file tip inside the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel. The Endo Motor device C-SMART-I PILOT is intended to be sterilized prior to use. The material composition of root canal files that can be used with the Endo Motor device C-SMART-I PILOT is Nickel titanium (NiTi). These files are not included in the submission. There are three working modes as: Endo motor mode: root canal preparation. Apex locator mode: root canal length measurement. Multi-function mode: undertake root canal measurement and preparation.

    C-SMART-MINI AP: The Endo Moto device C-SMART-MINI AP is a low-speed rotating oral equipment mainly used for root canal preparation and root canal measurement. The product is a portable device powered by built-in lithium batteries and charged by the adapter. LCD displays parameters such as speed, torque, apex position, etc. Users can also set and modify by keys, and provide design of functional check, calibration, set dominant hand and factory reset. The Endo Motor device is an endodontic treatment motorized handpiece with root canal measurement capability. It can be used to enlarge the canals while monitoring the position of the file tip inside the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel. The Endo Motor device C-SMART-MINI AP is intended to be sterilized prior to use. The material composition of root canal files that can be used with the Endo Motor device C-SMART-MINI AP is Nickel titanium (NiTi). These files are not included in the submission. There are three working modes as: Endo motor: Prepare the root canal, without apex locator function. Apex Locator: Measure the length of the root canal, without motor function. Multi-function: Measuring the length while root canal preparation.

    C-Smart-mini: The Endo Motors device C-Smart-mini is a low-speed rotating oral equipment mainly used for root canal preparation. The product is a portable device powered by built-in lithium batteries and charged by the adapter. LCD displays parameters such as speed, torque, rotary files program, etc. Users can also set and modify by keys, and provide design of calibration and screen display for right- or left-handed user. The Endo Motor device is an endodontic treatment motorized handpiece. It can be used to enlarge the canals. It can be used as a low-speed motorized handpiece. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel. The Endo Motor device C-Smart-mini is intended to be sterilized prior to use. The material composition of root canal files that can be used with the Endo Motor device C-Smart-mini is Nickel titanium (NiTi). These files are not included in the submission.

    AI/ML Overview

    The provided text does not describe the acceptance criteria and the study that proves the device meets the acceptance criteria. Instead, it is a 510(k) premarket notification for an "Endo Motor" device, outlining its features and comparing it to a predicate device to claim substantial equivalence.

    While the document references "Accuracy Testing (Internal test methods to demonstrate the apex position measurement accuracy of ±0.5mm)" among its non-clinical tests, and explicitly states "No clinical study is included in this submission," it does not provide the detailed information requested regarding acceptance criteria and a study proving those criteria are met for the device's performance.

    Therefore, I cannot extract the requested information from the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222548

    Validate with FDA (Live)

    Device Name
    Endo Motor
    Date Cleared
    2023-04-26

    (246 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Endo Motor, model: ZR-Rap, is a cordless endodontic treatment motorized handpiece with root canal measuring capability. It can be used to enlarge canals while monitoring the file tip inside the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length.

    This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

    Device Description

    The Endo Motor, model: ZR-Rap, is a battery-driven handpiece with a motor, equipped with a chuck for holding rotary instruments such as a dental file. The Endo Motor can be used for enlargement and preparation of root canals and can also be used as an apex locator.

    AI/ML Overview

    The provided text describes an FDA 510(k) premarket notification for an "Endo Motor" device. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than undergoing a full clinical efficacy study with specific performance acceptance criteria against a defined ground truth.

    Therefore, the document does not contain a detailed study proving the device meets specific acceptance criteria in the way you've outlined for diagnostic AI/ML devices. Instead, it focuses on non-clinical tests to show conformity to standards and comparability to a predicate device.

    However, I can extract the relevant information regarding performance claims and testing as best as possible from the provided text, especially concerning the "apex locator" function.

    Here's an attempt to answer your questions based on the provided document, acknowledging the limitations for a device of this type:


    Acceptance Criteria and Device Performance Study (Based on 510(k) Submission)

    The Endo Motor (Model: ZR-Rap) is a cordless endodontic treatment motorized handpiece with root canal measuring capability. The 510(k) submission aims to demonstrate substantial equivalence to a predicate device (Changzhou Sifary Medical Technology Co., Ltd. E-Connect S Endo Motor, K201993). The primary "acceptance criteria" in this context are conformity to relevant standards and comparable performance to the predicate, especially regarding the apex locator function.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Stated or Implied)Reported Device Performance (Proposed Device)Remarks
    Canal Enlargement Performance (Rotation Speed)150rpm-650rpmThis range is different from the predicate (120-1000 rpm) but the submission states: "The rotation speed of Endo Motor is within the speed range of the predicate device, and it has passed performance test, no affect on safety or efficacy." This implies the performance range is deemed acceptable and safe.
    Canal Enlargement Performance (Torque)0.6N.cm-4N.cmThis range is different from the predicate (0.5N.cm-4N.cm) but is implied to meet performance expectations and safety, similarly to rotation speed.
    Apex Locator Function Accuracy-0.5mm to +0.5mm for Apex position.This is explicitly stated as "Same" as the predicate device. This is the closest to a quantitative performance acceptance criterion mentioned for a functional aspect.
    BiocompatibilityMaterials conform to ISO10993.Verified through non-clinical testing against a recognized standard.
    Electromagnetic Compatibility (EMC)Conform to IEC60601-1-2:2014 & IEC60601-1-2Verified through non-clinical testing against recognized standards.
    Electrical, Mechanical, Thermal SafetyConform to IEC60601-1Verified through non-clinical testing against a recognized standard.
    General Dental Equipment Safety & PerformanceConform to IEC 80601-2-60:2019Verified through non-clinical testing against a recognized standard.
    Handpieces and Motors StandardsConform to ISO 14457:2017Verified through non-clinical testing against a recognized standard.
    Software Documentation (for medical devices)Moderate level of software documentation per FDA Guidance.Software documentation was provided and assessed.
    Reprocessing Validation (cleaning, disinfection, sterilization)Reprocessing validation performed per FDA Guidance Document.Validated to ensure the device components (Contra Angle, Lip Hook, File clip, Protective silicon cove) can be safely reprocessed, similar to the predicate's autoclavable components.
    Comparative Root Canal Measurement Performance (for apex locator function)Performed to evaluate the root canal length; results demonstrated substantial equivalence (implicitly met the predicate's -0.5mm to +0.5mm accuracy).This is a specific test for the critical apex locator function. While the exact methodology and pass/fail criteria beyond matching the predicate are not detailed, the conclusion is that it supports substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states "A Comparative Root Canal Measurement Performance Test to evaluate the root canal length" was performed. However, it does not specify the sample size (e.g., number of teeth, number of root canals, number of measurements) used for this test set nor the data provenance (e.g., country of origin, retrospective or prospective nature). Given the nature of a 510(k) for an existing technology, such tests are often conducted in-house on extracted teeth or phantoms.

    3. Number of Experts Used to Establish the Ground Truth and Qualifications

    The document does not specify the number of experts used, their qualifications, or if external experts were involved in establishing ground truth for the "Comparative Root Canal Measurement Performance Test." It implies that the test was conducted to compare its performance with existing methods or the predicate, but details on ground truth establishment are absent.

    4. Adjudication Method

    The document does not specify an adjudication method (e.g., 2+1, 3+1, none) for any test sets.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This type of study is typically associated with AI/ML diagnostic aids that assist human readers. The Endo Motor is a handpiece with an integrated apex locator, not a diagnostic AI/ML device that interacts with human readers for interpretation.

    6. Standalone Performance Study

    Yes, a standalone performance study was done for the apex locator function, as indicated by "A Comparative Root Canal Measurement Performance Test to evaluate the root canal length." The stated performance for this function is an accuracy of "-0.5mm to +0.5mm for Apex position." The study directly assessed the device's ability to measure root canal length.
    The other "standalone" performance studies were non-clinical tests demonstrating conformity to various safety and performance standards (e.g., ISO, IEC).

    7. Type of Ground Truth Used

    For the "Comparative Root Canal Measurement Performance Test," the document does not explicitly state the type of ground truth used. For an apex locator, common ground truths involve physical measurements taken after canal preparation (e.g., direct measurement with a caliper after sectioning a tooth or using a known reference point). It's likely an objective, physical measurement rather than expert consensus, pathology, or outcomes data, but this is an inference.

    8. Sample Size for the Training Set

    The concept of a "training set" is typically applicable to AI/ML models. Since this device is a traditional electromechanical medical instrument with integrated measurement capabilities rather than an AI/ML diagnostic tool, there is no mention or indication of a "training set" in the provided document.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a "training set," this question is not applicable to the information provided in the document.


    In summary: The submission relies on demonstrating conformity to international standards, comparing specifications and indications for use with a legally marketed predicate, and conducting specific non-clinical performance tests (like the root canal measurement performance test) to show substantial equivalence. It does not provide the detailed study design, sample sizes, expert qualifications, or ground truth establishment methods typically found in clinical trials for novel diagnostic devices, especially those incorporating AI/ML.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203320

    Validate with FDA (Live)

    Device Name
    Endo Motor
    Date Cleared
    2021-05-10

    (179 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Endo Motor, Ai-Motor and Endo Radar Plus are cordless endodontic treatment motorized handpiece with root canal measurement capability. It can be used for preparation and enlargement of root canals, or measuring the canal length. And it can be used to enlarge the canals while monitoring the file tip inside the canal.

    Device Description

    The Endo Motor Series includes a cordless hand-piece with charging adapter. The device has a chuck for holding rotary instruments such as dental files, and is mainly used in Endodontic surgeries. The device can enlarge the root canal by filing the dentin and measuring the root canal length, and may also be used for cutting and grinding for the preparation of root canal surgery. The rotary speed and the torque range can also be adjusted.

    AI/ML Overview

    The provided document, K203320, is a 510(k) premarket notification for an "Endo Motor" device. This device is a dental handpiece with root canal measurement capability. The 510(k) summary focuses on demonstrating "substantial equivalence" to a predicate device through non-clinical testing and comparison of technical specifications. It does not present a study proving the device meets specific acceptance criteria based on clinical performance metrics (e.g., accuracy, sensitivity, specificity) of an AI/algorithm-driven device.

    The document states that "Performance Comparison Testing of Root Canal Measurement Passed," but it does not provide details on the acceptance criteria or how this testing was conducted (e.g., sample size, ground truth, expert involvement).

    Therefore, I cannot fully answer your request for information regarding acceptance criteria and the study proving the device meets them, as the provided document lacks the empirical data and methodology typically found in studies for AI/algorithm-driven devices concerning clinical performance.

    However, I can extract the information provided about the non-clinical testing performed to support substantial equivalence:

    Acceptance Criteria & Device Performance (Based on Non-Clinical Testing for Substantial Equivalence)

    The document primarily relies on non-clinical testing and comparison to a predicate device to demonstrate substantial equivalence, rather than a clinical performance study with explicit acceptance criteria for an AI/algorithm.

    Acceptance Criterion (Non-Clinical)Reported Device PerformanceComments
    Cytotoxicity testing per ISO 10993-5Passed
    Sensitization testing per ISO 10993-10Passed
    Electrical safety testing per IEC 60601-1Passed
    Electromagnetic Disturbance (EMD) testing per IEC 60601-1-2Passed
    Software verification and validation per IEC 62304/FDA GuidanceMet RequirementsNo specific details on software performance are provided beyond meeting general requirements.
    Usability per IEC 62366-1Met specifications
    Shipping Containers and Systems Performance, per ASTM D4169Demonstrates package integrity maintained
    Dental Performance Requirements, per IEC 80601-2-60PassedGeneral statement, no specific metric given.
    Processing of Medical Device, per ISO 17664Passed
    Development, Validation and Routine Control of Sterilization Process, per ISO 17665-1Passed
    Moist Heat Sterilization, per ISO TS 17665-2Passed
    Dentistry - Handpieces and Motors, per ISO 14457PassedGeneral statement, no specific metric given.
    Performance Comparison Testing of Root Canal MeasurementPassedCrucially, no specific acceptance criteria (e.g., accuracy range in mm, error rate) for the root canal measurement capability are explicitly stated, nor are the details of this "Performance Comparison Testing" provided (e.g., sample size, methodology, ground truth establishment). This is the closest the document comes to a performance metric relevant to its "measurement capability."

    Detailed Answers to Specific Questions:

    1. A table of acceptance criteria and the reported device performance
    See table above. Note that for "Performance Comparison Testing of Root Canal Measurement," while it states "Passed," no quantitative acceptance criteria or specific performance metrics are provided in the document.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    Not applicable / Not provided. The document primarily reports non-clinical engineering and biological tests, and a comparison of technical specifications to a predicate device. There is no mention of a clinical "test set" with a specified sample size or data provenance in the context of root canal measurement performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    Not applicable / Not provided. There is no mention of expert involvement for establishing ground truth for any performance study in the document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable / Not provided. No test set or adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    No. The device is an "Endo Motor," a physical dental handpiece. It is not an AI/software device that would involve "human readers" or AI assistance in the way described for an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    The description implies that the device's "root canal measurement capability" is integral to the motor's function. While there's a "Performance Comparison Testing of Root Canal Measurement Passed," the document does not provide details on whether this was a true "standalone" algorithm performance evaluation with explicit metrics, or simply a functional test. Given the regulatory class (Class I, reserved) and the nature of the device (a physical handpiece), it is highly unlikely that a rigorous clinical standalone performance study, as typically seen for AI/ML devices, was conducted or required.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    Not provided. For the "Performance Comparison Testing of Root Canal Measurement," the document does not specify how the "ground truth" for root canal length was established.

    8. The sample size for the training set
    Not applicable / Not provided. This device is not described as an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established
    Not applicable / Not provided. This device is not described as an AI/ML device that requires a training set.

    In summary, the provided FDA 510(k) document (K203320) focuses on demonstrating "substantial equivalence" of a physical dental device (Endo Motor) through a comparison of design, intended use, and non-clinical testing results against a predicate device. It does not contain the kind of clinical performance study details (e.g., acceptance criteria for diagnostic accuracy, sample sizes of test/training sets, ground truth establishment by experts, MRMC studies) typically associated with AI/algorithm-driven medical devices. The "Performance Comparison Testing of Root Canal Measurement" is mentioned as "Passed," but without any specifics provided in this document regarding its methodology or quantitative results.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192649

    Validate with FDA (Live)

    Device Name
    Endo Motor
    Date Cleared
    2021-03-12

    (535 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Endo Motor T-Fine-II is indicated for use in standard endodontic procedures using rotary endodontic files for mechanical and rotary preparation of the root canal.

    Device Description

    Cordless endo motor is a professional equipment for root canal preparation. It is a drive for contra angle headpiece which is designed to assist dentists and dental surgeons perform standard endodontic procedures. The endo motor consists of Cordless main unit, Charger station and Adapter. It is intended to use with FDA cleared contra-angle handpiece (transmission ratio of contra angle handpiece: 16:1) which would hold the drill bit or file. However the device do not included the contra angle handpiece and files. The Endo smart is equipped with rechargeable Li-ion battery, which can be recharged using the provided charge station. By means of the different buttons on the main unit, the user could control various functions in the device, such as On/off, Speed, Torque and auto-reverse/auto forward mode. The Endo motor is not sterile and cannot be autoclaved.

    AI/ML Overview

    The provided document is a 510(k) summary for the Endo Motor Model: T-Fine-II. This type of submission is for demonstrating "substantial equivalence" to a legally marketed predicate device, rather than proving safety and efficacy through extensive clinical studies and setting specific acceptance criteria commonly seen for novel devices or highly complex AI/ML devices.

    Therefore, the document does not contain the detailed information typically associated with acceptance criteria and a study proving a device meets these criteria for an AI/ML-based medical device. Specifically, it lacks information on:

    • A table of acceptance criteria and reported device performance for an AI/ML system (as this is a mechanical dental device).
    • Sample sizes for test sets, data provenance, expert ground truth establishment, or adjudication methods (as no clinical or AI-specific validation studies were conducted for this device beyond bench testing for electrical safety and EMC, and software verification/validation).
    • MRMC studies, standalone algorithm performance, or specific ground truth types (pathology, outcomes data) relevant to AI/ML.
    • Training set details for an AI/ML model.

    What the document does provide in terms of "acceptance criteria" and "proof":

    For this specific device, the "acceptance criteria" are compliance with relevant safety and performance standards for dental electrical equipment and demonstrating substantial equivalence to a predicate device. The "proof" involves a combination of bench testing and comparison.

    Here's an interpretation based on the provided text, focusing on the available information:

    1. Acceptance Criteria and Reported Device Performance (as inferred from the document's purpose):

    Acceptance Criteria (Inferred from regulatory requirements for this device type)Reported Device Performance (from "Test Summary" and "Comparison to Predicate Device" sections)
    Electrical Safety: Comply with general requirements for medical electrical equipment.Complies with IEC 60601-1:2005+A1:2012, IEC 60601-1-2:2014, and IEC 80601-2-60:2012.
    Electromagnetic Compatibility (EMC): Ensure device functions without unacceptable electromagnetic interference.Complies with IEC 60601-1-2:2014.
    Cleaning and Disinfection: Compatibility with specified cleaning/disinfection procedures.Conducted according to AAMI TIR30 and AAMI TIR12.
    Software Functionality: Software operates as intended and is safe.Software verification and validation testing conducted. Determined to be of "moderate" level of concern. Documentation provided as recommended by FDA guidance. Display function validated in software evaluation report.
    Substantial Equivalence: Device is as safe and effective as a legally marketed predicate device.Compared to predicate device (Cordless ENDO-Handpiece "ENTRAN" K090931). Found very similar in design principle, intended use, indications for use, functions, material, and applicable standards. Minor differences (weight, displayed parameters, battery capacity, operating/storage temperature, torque range) do not raise new questions of safety or effectiveness as per the applicant's detailed notes and compliance with safety standards. Specifically, it provides auto reverse and stop function meeting IEC80601-2-60 requirements.
    Intended Use: Device performs its intended medical function."Endo Motor T-Fine-II is indicated for use in standard endodontic procedures using rotary endodontic files for mechanical and rotary preparation of the root canal." (Same intended use as predicate, despite minor wording differences).
    Performance (Mechanical/Functional): Basic mechanical parameters operate within acceptable ranges.Torque: 0.3-3.0 Ncm (Predicate: 0.5-4.0 Ncm) – Minor difference noted but considered safe due to adjustability and auto-reverse/stop functions. Transmission ratio of electric contra angle: 16:1 (Same as predicate). Auto-control mode: Yes (Same as predicate). Settable functions: Speed, Torque, Auto-control (Same as predicate).

    2. Sample Sizes and Data Provenance:

    • No specific sample sizes for test sets (patients/images) are mentioned because this device's validation is primarily through bench testing against engineering standards and comparison to a predicate, not clinical trials or AI/ML model validation with patient data.
    • The document implies manufacturing is in Foshan, Guangdong, China (manufacturer location).
    • The testing appears to be bench testing for engineering compliance, not retrospective or prospective clinical data collection.

    3. Number of Experts and Qualifications:

    • Not applicable. This information is relevant for studies establishing ground truth for AI/ML or clinical interpretation, which were not conducted or required for this device's 510(k) submission.

    4. Adjudication Method:

    • Not applicable. No clinical data interpretation requiring adjudication was performed or described.

    5. MRMC Comparative Effectiveness Study:

    • No. This type of study is for evaluating human reader performance with and without AI assistance, which is not relevant to this mechanical dental device.

    6. Standalone (Algorithm Only) Performance:

    • Not applicable. This device is a mechanical Endo Motor, not an AI algorithm. Its software is for controlling the device's functions, not for diagnostic or predictive tasks that would have standalone performance metrics.

    7. Type of Ground Truth Used:

    • No "ground truth" in the AI/ML sense (e.g., pathology, outcomes data, expert consensus) was used. The validation relies on:
      • Engineering Standards Compliance: The standards (IEC 60601 series, AAMI TIRs) themselves define the "ground truth" for safety and performance (e.g., passing electrical leakage tests, demonstrating specific torque output).
      • Predicate Device Comparison: The predicate device (K090931) serves as a benchmark for "equivalent" safety and effectiveness.

    8. Sample Size for Training Set:

    • Not applicable. There is no AI/ML model being trained with patient data for this device.

    9. How Ground Truth for Training Set Was Established:

    • Not applicable. As there is no AI/ML model training, this step is irrelevant.

    In summary, the provided document is a 510(k) summary for a Class I mechanical dental device. The "acceptance criteria" and "proof" presented align with the regulatory pathway for such devices, focusing on compliance with recognized electrical and mechanical safety standards and demonstrating substantial equivalence to a predicate device, rather than detailed clinical performance studies or AI/ML validation.