510(k) Data Aggregation
(124 days)
The ActiMed ENA C-T™ Total Cholesterol Test is a non-instrumented, enzymatic assay intended for the in vitro quantitative determination of total cholesterol in fingerstick whole blood. It is indicated for physician directed use (PDU) by the patient for screening and monitoring cholesterol.
In a PDU monitoring application, the patient's physician (or other medical professional under the direction of a physician) initially trains the patient to use the test properly and provides information on the proper interpretation of results and limitations of the test. With the physician directed indication for use, patients on cholesterol lowering therapy can test their own total cholesterol at home and provide the results to their physician or other healthcare organization which works with the physician to maintain compliance on the therapy.
In a PDU screening application, a physician (or other medical professional under the direction of a physician) initially trains a group of individuals to use the test properly and provides information on the proper interpretation of results and limitations of the test. The physician directed use version of the ENA.C.T™ Total Cholesterol Test kits are then distributed to the group of individuals who can use the test at home and then communicate their results to the physician or healthcare organization.
The ActiMed ENA-C-T™ Total Cholesterol Test is a noninstrumented, enzymatic assay intended for the in vitro quantitative determination of total cholesterol in fingerstick whole blood. It is based on stable, dry chemistry reagents embedded in a flow device with a factory calibrated scale. The device provides a direct visual read-out of test results. It has built-in controls to indicate that a sufficient volume of whole blood has been added as sample, to verify that the reagents are functioning, and to confirm when the test has been completed. The test device is self-actuating, requiring neither timing nor handling between addition of the sample and reading the result.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Device: ACTIMED LABORATORIES, INC. ENA-C.T™ Total Cholesterol Test Physicians Directed Use
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria (Stated Goal / Benchmark) | Reported Device Performance |
|---|---|
| Linear Range: | |
| 120 mg/dL to 360 mg/dL (Implied acceptance range, as this is the range it was designed and tested for) | 120 mg/dL to 360 mg/dL (Results from NCCLS Linearity Protocol testing and from testing 40 normal and abnormal clinical samples demonstrated this range) |
| Interference: | |
| No interference with: - Ascorbic acid (up to 10 mg/dL) - Uric acid (up to 11 mg/dL) - Hemoglobin (up to 250 mg/dL) - Triglycerides (up to 850 mg/dL) - Bilirubin (up to 20 mg/dL) | No interference found in samples containing up to 10 mg/dL ascorbic acid, 11 mg/dL uric acid, 250 mg/dL of hemoglobin, 850 mg/dL of triglycerides, or 20 mg/dL bilirubin. (Note: Acetaminophen interferes at 1 mg/dL, and high concentrations of ascorbic acid also interfere, which were identified limitations not part of "acceptance" but noted characteristics) |
| Precision (Reproducibility): | |
| Comparable to predicate device; generally less than 6.9% CV for paired fingerstick whole blood samples. (Implied acceptance based on comparison to predicate and overall average achieved) | CVs were 5.2%, 6.7%, and 8.8% for Sites 1, 2, and 3 respectively, with an average CV for all sites of 6.9%. This demonstrates the device was reproducible in the hands of trained patients. |
| Accuracy (Bias at key cholesterol levels): - Bias not to exceed 3% at 200 mg/dL compared to CRMLN reference method - Bias not to exceed 3% at 240 mg/dL compared to CRMLN reference method (Based on recommendation of the Laboratory Standardization Panel (LSP)) | Bias at 200 mg/dL was 2.86% (well within recommendation) Bias at 240 mg/dL was 2.69% (well within recommendation) |
| Correlation to CRMLN method: Within 10% (Implied acceptance based on stated equivalence) | Least squares regression equation: y = 1.018x + 2.13, with a correlation coefficient of r = 0.917. Correlations between the ENA-C-T™ device (PDU) and the CRMLN Method were within 10%. |
| Correlation to Predicate Device (ENA-C-T™ Total Cholesterol Test (POL)): Within 10% (Implied acceptance based on stated equivalence) | Least squares regression equation: y = 0.931x + 13.8, with a correlation coefficient of r = 0.899. Correlations between the ENA C T™ device (PDU) and the ENA C T™ device (POL) were within 10%. |
| Substantial Equivalence to Predicate Device: | |
| Intended use, design, materials, operational features, and performance comparable. | "Performance characteristics generated during analytical and clinical studies on the ENA-C-T™ Total Cholesterol Test (PDU) are comparable to those generated in the same study for the ENA-C T™ Total Cholesterol Test (POL)." "Thus, the ENA C T™ Total Cholesterol Test (PDU) is substantially equivalent to the ENA-C.T™ Total Cholesterol Test (POL) in terms of performance characteristics, as well as intended use, design, materials, and operational features." |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: 136 paired fingerstick whole blood and venous samples.
- Data Provenance: The studies were conducted at "three POL sites" (Physician Office Laboratory sites). While explicit country of origin is not stated, the submitter is in Burlington, NJ, USA, and the FDA approval is for the US market, strongly implying US-based data. The studies were prospective clinical studies to assess the performance of the ENA-C-T™ Total Cholesterol Test (PDU) in comparison to a reference method and the predicate device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- The text does not explicitly state the number of experts. The ground truth was established by laboratory testing at a "Cholesterol Reference Method Network Laboratory (CRMLN)" using the "Abell-Kendall serum reference method or enzymatic method." This implies that the ground truth analysis was performed by trained laboratory technicians/scientists at a certified reference lab, adhering to established high-standard methodologies (CRMLN). No specific individual "experts" or their years of experience are mentioned, but the CRMLN designation assures high qualification.
4. Adjudication Method for the Test Set:
- The document describes a direct comparison between the ENA-C-T™ PDU device and the CRMLN reference method, and also between the PDU device and the POL predicate device. This is a direct comparison method rather than an adjudication method involving multiple human readers interpreting the same device results. For precision, differences between self-test/self-read and professional-test/professional-read devices were assessed, but this isn't an adjudication method in the typical sense for interpretative disagreements.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done:
- No, an MRMC comparative effectiveness study was not done. The device provides a "direct visual read-out of test results" from a "factory-calibrated scale." The studies focused on the accuracy and precision of this visual read against laboratory reference methods and the predicate device, especially in the hands of trained patients. It was not a study comparing how human readers improve with AI assistance, as this is a non-instrumented, direct-read test, not an AI diagnostic tool.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done:
- Yes, a standalone performance assessment was conducted (algorithm-like, without human interpretation for the ground truth). The "ENA-C-T™ Total Cholesterol Test" is described as a "noninstrumented, enzymatic assay" with a "factory calibrated scale" that provides a "direct visual read-out." While a human reads the result from the scale, the core performance characteristics (linearity, interference, precision, bias, and correlation to reference methods) were evaluated based on the chemical reaction and scale design itself. The "self test/self read" component of the precision study involved human users, but the key performance metrics are inherent to the device's chemical and physical properties.
7. The Type of Ground Truth Used:
- Expert Consensus / Reference Method: The primary ground truth for accuracy and bias was established by a Cholesterol Reference Method Network Laboratory (CRMLN) using the Abell-Kendall serum reference method or an equivalent enzymatic method on venous blood samples. This is considered a gold standard and a form of expert-validated reference data.
8. The Sample Size for the Training Set:
- The document does not explicitly state a separate "training set" sample size for the device's development. The "40 normal and abnormal clinical samples" mentioned in the linearity section might have been part of early development/calibration, but it's not designated as a distinct training set. This is typical for older, non-AI based medical devices where empirical testing and calibration against known standards form the development process, rather than a machine learning training paradigm.
9. How the Ground Truth for the Training Set Was Established:
- As a distinct training set (in the machine learning sense) is not mentioned, the method for establishing ground truth for a training set is not applicable here. However, the device's inherent "factory calibrated scale" would have relied on a process similar to how the ground truth was established for the test set – comparison to high-accuracy laboratory reference methods to ensure the scale accurately reflects cholesterol concentrations.
Ask a specific question about this device
(97 days)
The ActiMed ENA C.T™ Total Cholesterol Test is a non-instrumented, enzymatic assay intended for the in vitro quantitative determination of total cholesterol in fingerstick whole blood. It is indicated for professional use in settings such as clinical laboratories and physician office laboratories (POLs) to screen for elevated cholesterol as a risk factor in coronary heart disease (CHD).
The ENA-C-T™ Total Cholesterol Test is based on stable, dry chemistry reagents embedded in a flow device with a factory calibrated scale. The device provides a direct visual read-out of test results. It has built-in controls to indicate that a sufficient volume of whole blood has been added as sample, to verify that the reagents are functioning, and to confirm when the test has been completed. The test device is self-actuating, requiring neither timing nor handling between addition of the sample and reading the result.
The ENA.C.T™ Total Cholesterol Test starts when a sufficient amount of whole blood has been introduced into the sample well. A self-actuating siphon, in turn, transfers the blood to an absorbent pad located immediately below the "START" window. The blood saturated pad shows through the clear window as a red indicator, signaling that enough sample has been added and that the analytical process has been initiated. The absorbent pad and a second filter separate red cells from plasma. The plasma subsequently flows into an enzyme pad containing cholesterol esterase and cholesterol oxidase which completely convert cholesterol and cholesterol esters into cholestenone and hydrogen peroxide. A flow delay pad restricts passage of the plasma to permit completion of the enzymatic reactions. Subsequently, the plasma flows into the measurement zone which acts as a capillary. In this zone, colorless dyes and the enzyme peroxidase are immobilized on a thin fabric layer. The hydrogen peroxide converted from cholesterol in the plasma converts the colorless dye layer into a blue color bar, the length of which is proportional to the amount of hydrogen peroxide, and in turn, cholesterol in the sample. The bar extends until a precise amount of plasma has completely filled the measurement zone and "END/QA" draw zone. This draw zone turns green at the end of the procedure when the cholesterol converting enzymes dissolved in the plasma react with control cholesterol contained in the draw zone. The appearance of a green color signals that the test is complete and that reagents were active. The total cholesterol concentration is directly read from the factory calibrated scale on the device. The entire process typically requires 12 to 15 minutes. The color bar formed is stable and may be read at any time within 48 hours.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
ENA·C·T™ Total Cholesterol Test: Acceptance Criteria and Study Details
The provided text details the 510(k) summary for the ENA·C·T™ Total Cholesterol Test, comparing it to the predicate device, the ChemTrak, Inc. AccuMeter® Cholesterol Test (K905405). The primary goal is to demonstrate substantial equivalence to the predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated in a dedicated section with specific numerical thresholds. Instead, the document discusses performance in relation to the predicate device and national recommendations (NRS/CHOL, LSP). I've inferred the "acceptance criteria" based on these comparisons and stated recommendations.
| Performance Characteristic | Acceptance Criteria (Inferred from comparisons & recommendations) | Reported ENA·C·T™ Performance |
|---|---|---|
| Linear Range | Comparable to predicate device (125-400 mg/dL) | 120 mg/dL to 360 mg/dL |
| Interfering Substances | No interference comparable to predicate device | No interference: up to 8 mg/dL ascorbic acid, 1 mg/dL acetaminophen, 14 mg/dL bilirubin, 200 mg/dL hemoglobin, 800 mg/dL triglycerides. (Comparable to AccuMeter's reported limits) |
| Precision (CV) | Generally less than 5% for controls at medical decision points (Comparable to predicate device) | Within-run CVs < 5% (Level 1 control), < 4% (Level 2 control). Total CVs ≤ 5% (Level 1), < 4% (Level 2). |
| Bias (vs. Abell-Kendall) | Not to exceed 3% (LSP recommendation) | 1.6% at 200 mg/dL, 0.9% at 240 mg/dL |
| Correlation (ENA·C·T™ vs. Abell-Kendall) | Within 10% (Compared to predicate device, though exact R-value for predicate is not given for this comparison) | r = 0.932 |
| Correlation (ENA·C·T™ vs. AccuMeter®) | Within 10% | r = 0.901 |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: 198 paired fingerstick whole blood and venous samples.
- Data Provenance: The studies were conducted at four Physician Office Laboratories (POLs) in the US. This implies a prospective collection of samples as part of the clinical studies.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The text does not specify the number of experts for interpreting the ENA·C·T™ or AccuMeter® test results at the POL sites. However, the ground truth (Abell-Kendall serum reference method) was performed at a Cholesterol Reference Method Network Laboratory (CRMLN). While the number of experts isn't given, CRMLN laboratories are specialized and highly qualified for cholesterol measurement. The Abell-Kendall method itself is a complex, gold-standard chemical assay, not typically interpreted by a single "expert" in the way a diagnostic image would be. The interpretation of the visual read-out of the ENA·C·T™ and AccuMeter® devices was likely performed by the POL staff who are the intended users, but their qualifications are not detailed beyond "potential users in POL settings."
4. Adjudication Method for the Test Set
The text indicates that the ENA·C·T™ and AccuMeter® results were "read and recorded by one operator" in the precision studies. For the clinical studies, where 198 samples were tested, it doesn't explicitly state an adjudication method for the device readings. Given the nature of a visual color bar reading from a factory-calibrated scale, a formal adjudication process (like 2+1 or 3+1) is typically not applicable as it would be for subjective interpretations (e.g., radiology). The Abell-Kendall method (ground truth) is an objective laboratory assay, not subject to individual interpretation in the same way.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done in the context of comparing human readers with AI vs. without AI assistance. The ENA·C·T™ device is a standalone diagnostic test, not an AI-assisted tool for human readers. The clinical studies involved human operators reading the visual results of the ENA·C·T™ and AccuMeter® devices, but not in a comparative "with AI vs. without AI" effectiveness study.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, the studies essentially assess the standalone performance of the device, as it is designed for a single operator to read the result visually. While a human is "in-the-loop" for reading the visual output, there is no separate "algorithm only" component that would perform the diagnosis without any human interaction (e.g., an AI system autonomously diagnosing from images). The device itself acts as the "algorithm" that produces a visual result for human interpretation.
7. The Type of Ground Truth Used
The primary ground truth used was the Abell-Kendall serum reference method, performed at a Cholesterol Reference Method Network Laboratory (CRMLN) on venous blood samples. This is considered a gold standard and a highly accurate reference method for total cholesterol measurement.
8. The Sample Size for the Training Set
The document does not explicitly state a separate "training set" for the ENA·C·T™ device in the context of machine learning. However, it mentions:
- "Each lot of devices is calibrated by using paired fingerstick blood and venous blood samples covering the dynamic range of the test." These samples are used to determine "calibration coefficients" and generate "measurement scales for each lot." This can be considered an internal calibration or "training" process for the physical device's scale. The sample size for this internal calibration is not specified.
- The clinical studies (198 samples) and analytical studies would be considered part of the validation or test set, not a training set.
9. How the Ground Truth for the Training Set Was Established
As noted above, if we consider the lot-specific calibration a form of "training," the ground truth for these samples was established using the Abell-Kendall serum reference method on venous blood samples. This ensures the device's fingerstick whole blood results are traceable to the National Reference System for Cholesterol (NRS/CHOL) and automatically converted to serum values.
Ask a specific question about this device