K Number
K981181

Validate with FDA (Live)

Date Cleared
1998-08-03

(124 days)

Product Code
Regulation Number
862.1175
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ActiMed ENA C-T™ Total Cholesterol Test is a non-instrumented, enzymatic assay intended for the in vitro quantitative determination of total cholesterol in fingerstick whole blood. It is indicated for physician directed use (PDU) by the patient for screening and monitoring cholesterol.

In a PDU monitoring application, the patient's physician (or other medical professional under the direction of a physician) initially trains the patient to use the test properly and provides information on the proper interpretation of results and limitations of the test. With the physician directed indication for use, patients on cholesterol lowering therapy can test their own total cholesterol at home and provide the results to their physician or other healthcare organization which works with the physician to maintain compliance on the therapy.

In a PDU screening application, a physician (or other medical professional under the direction of a physician) initially trains a group of individuals to use the test properly and provides information on the proper interpretation of results and limitations of the test. The physician directed use version of the ENA.C.T™ Total Cholesterol Test kits are then distributed to the group of individuals who can use the test at home and then communicate their results to the physician or healthcare organization.

Device Description

The ActiMed ENA-C-T™ Total Cholesterol Test is a noninstrumented, enzymatic assay intended for the in vitro quantitative determination of total cholesterol in fingerstick whole blood. It is based on stable, dry chemistry reagents embedded in a flow device with a factory calibrated scale. The device provides a direct visual read-out of test results. It has built-in controls to indicate that a sufficient volume of whole blood has been added as sample, to verify that the reagents are functioning, and to confirm when the test has been completed. The test device is self-actuating, requiring neither timing nor handling between addition of the sample and reading the result.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Device: ACTIMED LABORATORIES, INC. ENA-C.T™ Total Cholesterol Test Physicians Directed Use

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Stated Goal / Benchmark)Reported Device Performance
Linear Range:
120 mg/dL to 360 mg/dL (Implied acceptance range, as this is the range it was designed and tested for)120 mg/dL to 360 mg/dL (Results from NCCLS Linearity Protocol testing and from testing 40 normal and abnormal clinical samples demonstrated this range)
Interference:
No interference with: - Ascorbic acid (up to 10 mg/dL) - Uric acid (up to 11 mg/dL) - Hemoglobin (up to 250 mg/dL) - Triglycerides (up to 850 mg/dL) - Bilirubin (up to 20 mg/dL)No interference found in samples containing up to 10 mg/dL ascorbic acid, 11 mg/dL uric acid, 250 mg/dL of hemoglobin, 850 mg/dL of triglycerides, or 20 mg/dL bilirubin. (Note: Acetaminophen interferes at 1 mg/dL, and high concentrations of ascorbic acid also interfere, which were identified limitations not part of "acceptance" but noted characteristics)
Precision (Reproducibility):
Comparable to predicate device; generally less than 6.9% CV for paired fingerstick whole blood samples. (Implied acceptance based on comparison to predicate and overall average achieved)CVs were 5.2%, 6.7%, and 8.8% for Sites 1, 2, and 3 respectively, with an average CV for all sites of 6.9%. This demonstrates the device was reproducible in the hands of trained patients.
Accuracy (Bias at key cholesterol levels): - Bias not to exceed 3% at 200 mg/dL compared to CRMLN reference method - Bias not to exceed 3% at 240 mg/dL compared to CRMLN reference method (Based on recommendation of the Laboratory Standardization Panel (LSP))Bias at 200 mg/dL was 2.86% (well within recommendation) Bias at 240 mg/dL was 2.69% (well within recommendation)
Correlation to CRMLN method: Within 10% (Implied acceptance based on stated equivalence)Least squares regression equation: y = 1.018x + 2.13, with a correlation coefficient of r = 0.917. Correlations between the ENA-C-T™ device (PDU) and the CRMLN Method were within 10%.
Correlation to Predicate Device (ENA-C-T™ Total Cholesterol Test (POL)): Within 10% (Implied acceptance based on stated equivalence)Least squares regression equation: y = 0.931x + 13.8, with a correlation coefficient of r = 0.899. Correlations between the ENA C T™ device (PDU) and the ENA C T™ device (POL) were within 10%.
Substantial Equivalence to Predicate Device:
Intended use, design, materials, operational features, and performance comparable."Performance characteristics generated during analytical and clinical studies on the ENA-C-T™ Total Cholesterol Test (PDU) are comparable to those generated in the same study for the ENA-C T™ Total Cholesterol Test (POL)." "Thus, the ENA C T™ Total Cholesterol Test (PDU) is substantially equivalent to the ENA-C.T™ Total Cholesterol Test (POL) in terms of performance characteristics, as well as intended use, design, materials, and operational features."

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: 136 paired fingerstick whole blood and venous samples.
  • Data Provenance: The studies were conducted at "three POL sites" (Physician Office Laboratory sites). While explicit country of origin is not stated, the submitter is in Burlington, NJ, USA, and the FDA approval is for the US market, strongly implying US-based data. The studies were prospective clinical studies to assess the performance of the ENA-C-T™ Total Cholesterol Test (PDU) in comparison to a reference method and the predicate device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • The text does not explicitly state the number of experts. The ground truth was established by laboratory testing at a "Cholesterol Reference Method Network Laboratory (CRMLN)" using the "Abell-Kendall serum reference method or enzymatic method." This implies that the ground truth analysis was performed by trained laboratory technicians/scientists at a certified reference lab, adhering to established high-standard methodologies (CRMLN). No specific individual "experts" or their years of experience are mentioned, but the CRMLN designation assures high qualification.

4. Adjudication Method for the Test Set:

  • The document describes a direct comparison between the ENA-C-T™ PDU device and the CRMLN reference method, and also between the PDU device and the POL predicate device. This is a direct comparison method rather than an adjudication method involving multiple human readers interpreting the same device results. For precision, differences between self-test/self-read and professional-test/professional-read devices were assessed, but this isn't an adjudication method in the typical sense for interpretative disagreements.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done:

  • No, an MRMC comparative effectiveness study was not done. The device provides a "direct visual read-out of test results" from a "factory-calibrated scale." The studies focused on the accuracy and precision of this visual read against laboratory reference methods and the predicate device, especially in the hands of trained patients. It was not a study comparing how human readers improve with AI assistance, as this is a non-instrumented, direct-read test, not an AI diagnostic tool.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done:

  • Yes, a standalone performance assessment was conducted (algorithm-like, without human interpretation for the ground truth). The "ENA-C-T™ Total Cholesterol Test" is described as a "noninstrumented, enzymatic assay" with a "factory calibrated scale" that provides a "direct visual read-out." While a human reads the result from the scale, the core performance characteristics (linearity, interference, precision, bias, and correlation to reference methods) were evaluated based on the chemical reaction and scale design itself. The "self test/self read" component of the precision study involved human users, but the key performance metrics are inherent to the device's chemical and physical properties.

7. The Type of Ground Truth Used:

  • Expert Consensus / Reference Method: The primary ground truth for accuracy and bias was established by a Cholesterol Reference Method Network Laboratory (CRMLN) using the Abell-Kendall serum reference method or an equivalent enzymatic method on venous blood samples. This is considered a gold standard and a form of expert-validated reference data.

8. The Sample Size for the Training Set:

  • The document does not explicitly state a separate "training set" sample size for the device's development. The "40 normal and abnormal clinical samples" mentioned in the linearity section might have been part of early development/calibration, but it's not designated as a distinct training set. This is typical for older, non-AI based medical devices where empirical testing and calibration against known standards form the development process, rather than a machine learning training paradigm.

9. How the Ground Truth for the Training Set Was Established:

  • As a distinct training set (in the machine learning sense) is not mentioned, the method for establishing ground truth for a training set is not applicable here. However, the device's inherent "factory calibrated scale" would have relied on a process similar to how the ground truth was established for the test set – comparison to high-accuracy laboratory reference methods to ensure the scale accurately reflects cholesterol concentrations.

{0}------------------------------------------------

AUG 3 1998

198118

510(k) SUMMARY

ACTIMED LABORATORIES, INC. ENA-C.T™ Total Cholesterol Test Physicians Directed Use

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Susan McGeehan Manager, Quality Assurance ActiMed Laboratories, Inc. 5 Terri Lane Burlington, NJ 08016 (609) 387-7700, Ext. 119 Phone: (609) 387-2700 Facsimile:

Contact Person: same as above

March 27, 1998 Date Prepared:

Name of Device and Name/Address of Sponsor

ENA-C-T™ Total Cholesterol Test

ActiMed Laboratories, Inc. 5 Terri Lane Burlington, NJ 08016

Classification Name

Cholesterol (total) test system

Predicate Device

ActiMed Laboratories, Inc., ENA-C-T™ Total Cholesterol Test (POL) (K960377) and ChemTrak AccuMeter® Cholesterol Test (K905405).

{1}------------------------------------------------

Intended Use and Indications for Use

The ActiMed ENA-C'T™ Total Cholesterol Test is a noninstrumented, enzymatic assay intended for the in vitro quantitative determination of total cholesterol in fingerstick whole blood. It is indicated for physician directed use (PDU) by the patient for screening and monitoring cholesterol.

In a PDU monitoring application, the patient's physician (or other medical professional under the direction of a physician) initially trains the patient to use the test properly and provides information on the proper interpretation of results and limitations of the test. With the physician directed indication for use, patients on cholesterol lowering therapy can test their own total cholesterol at home and provide the results to their physician or other healthcare organization which works with the physician to maintain compliance on the therapy.

In a PDU screening application, a physician (or other medical professional under the direction of a physician) initially trains a group of individuals to use the test properly and provides information on the proper intervretation of results and limitations of the test. The physician directed use version of the ENA C.T™ Total Cholesterol Test kits are then distributed to the group of individuals who can use the test at home and then communicate their results to the physician or healthcare organization.

Technological Characteristics and Substantial Equivalence

The ENA-C-T™ Total Cholesterol Test is based on stable, dry chemistry reagents embedded in a flow device with a factory calibrated scale. The device provides a direct visual read-out of test results. It has built-in controls to indicate that a sufficient volume of whole blood has been added as sample, to verify that the reagents are functioning, and to confirm when the test has been completed. The test device is self-actuating, requiring neither timing nor handling between addition of the sample and reading the result.

The ENA·C·T™ Total Cholesterol Test - Physicians Directed Use (PDU) is substantially equivalent to the ENA-C-T™ Total Cholesterol Test (POL) in terms of intended use, design, materials, operational features, and performance. Both devices are based upon an enzymatic reaction that occurs when whole blood is added to dry chemistry reagents embedded in a flow device. Both devices are standardized to the National Reference System for Cholesterol (NRS/CHOL). Results from both devices are read visually within approximately 15-20 minutes of addition of sample to the device. For both

{2}------------------------------------------------

versions of the ENA.C.T™ Total Cholesterol Test, results in mg/dL are read directly from a lot-specific, factory-calibrated scale printed on the device.

Both the ENA.C.T™ Total Cholesterol Test (PDU) and the ENA.C.T™ Total Cholesterol Test (POL) have built-in controls. Both versions of the ENA·C·T™ Total Cholesterol test have controls to indicate that a sufficient volume of whole blood has been added as sample, to verify that the reagents are functioning, and to confirm when the test has been completed.

Performance characteristics generated during analytical and clinical studies on the ENA.C.T™ Total Cholesterol Test (PDU) are comparable to those generated in the same study for the ENA-C T™ Total Cholesterol Test (POL). The linear range of both versions of the ENA-C-T™ Total Cholesterol Test is 120 mg/dL to 360 mg/dL. The recovery of cholesterol in the presence of potentially interfering substances was determined for the ENA.C.T™ device. No interference was found in samples containing up to 10 mg/dL ascorbic acid. 11 mg/dL uric acid, 250 mg/dL of hemoglobin, 850 mg/dL of triglycerides, or 20 mg/dL bilirubin. Acetaminophen interferes at a level of 1 mg/dL. The test is affected by high concentrations of ascorbic acid.

Precision results for these two visually read tests are comparable, with percent coefficients of variation generally less than 6.9% for paired fingerstick whole blood samples.

In clinical studies at three POL sites, ENA-C-T™ Total Cholesterol Test (PDU), ENA C T™ Total Cholesterol Test (POL), and CRMLN results were generated on 136 paired fingerstick whole blood and venous samples. Correlation's between the ENA-CT™ Total Cholesterol Test (PDU) and the CRMLN methods and between the ENAC TTM Total Cholesterol Test (PDU)and the ENA-C T™ Total Cholesterol Test (POL) were acceptable. Thus, the ENA C T™ Total Cholesterol Test (PDU) is substantially equivalent to the ENA-C.T™ Total Cholesterol Test (POL) in terms of performance characteristics, as well as intended use, design, materials, and operational features.

Analytical and Clinical Testing

Linear Range

Results from NCCLS Linearity Protocol testing and from testing 40 normal and abnormal clinical samples demonstrated that the linear range of the ENA C TTM Total Cholesterol Test is 120 mg/dL to 360 mg/dL.

{3}------------------------------------------------

K981181

Interfering Substances

Studies of the recovery of cholesterol in the presence of potentially interfering substances were carried out. No interference was found in samples containing up to 10 mg/dL ascorbic acid, 11 mg/dL uric acid, 250 mg/dL of hemoglobin, 850 mg/dL of triglycerides, or 20 mg/dL bilirubin. Acetaminophen interferes at a level of 1 mg/dL.

Precision

The precision of the devices has been established by using data for the ENA C T™ device from the clinical studies. Precision was determined by obtaining the differences between self test/self read and professional test/professional read devices for each subject at each site. The CVs were 5.2%, 6.7%, and 8.8% for Site 1, 2, and 3 respectively, with an average CV for all sites of 6.9%. These results demonstrate that the visually read ENA-C-T™ Total Cholesterol Test was reproducible in the hands of patients trained in the use of the device.

Clinical Studies

A total of 136 paired fingerstick whole blood and venous samples were tested during clinical studies at three POL sites. Fingerstick whole blood samples were assayed at the sites by the ENA.C.T™ Total Cholesterol Test and venous samples were assayed at a Cholesterol Reference Method Network Laboratory (CRMLN) by the Abell-Kendall serum reference method or enzymatic method. For the comparison between the ENA-C-T™ Total Cholesterol Test (PDU) and the CRMLN method, the least squares regression equation for all sites was y = 1.018x + 2.13, with a correlation coefficient of r = 0.917. Bias at 200 mg/dL and at 240 mg/dL was 2.86% and 2.69%, respectively. For the comparison between the ENACT™ Total Cholesterol Test (PDU) and the ENA-C-T™ Total Cholesterol Test (POL), the least squares regression equation for all sites was y = 0.931x + 13.8, with a correlation coefficient of r = 0.899.

Precision studies conducted at three POL sites have demonstrated that the visually read ENA-C-T™ Total Cholesterol Test is reproducible, with an average CV of 6.9% for paired fingerstick whole blood samples from a total of 136 subjects. Clinical studies compared the ENA-C-T™ Total Cholesterol Test (PDU) to the CRMLN serum reference method and the ENA CT™ Total Cholesterol Test (PDU) to the ENA-C-T™ Total Cholesterol Test (POL) on samples from 136 patients. Bias of the ENA-C-T™ device (PDU), compared to the CRMLN serum reference method was 2.86% at the 200 mg/dL level and

{4}------------------------------------------------

K 981181

2.69% at the 240 mg/dL for the combined sites. These figures are well within the recommendation of the Laboratory Standardization Panel (LSP) on Blood Cholesterol Measurement that bias not exceed 3%. Correlations between the ENA-C-T™ device (PDU) and the CRMLN Method and between the ENA C T™ device (PDU) and the ENA C T™ device (POL) were within 10%. Thus, the ENA C TTM Total Cholesterol Test (PDU) has demonstrated acceptable precision, bias and accuracy, and has been shown to be substantially equivalent to the ENA.C.Tmm Total Cholesterol Test (POL).

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human figure, composed of three interconnected shapes.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

AUG 3 1998

Susan McGeehan · Manager - Quality Assurance ActiMed Laboratories, Inc. 5 Terri Lane Burlington, New Jersey 08016

Re : K981181 ENA-C-T™ Total Cholesterol Test Model F 20000, (Physician Directed Use) Regulatory Class: I Product Code: СНН Dated: March 27, 1998 Received: April 1, 1998

Dear Ms. McGeehan:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{6}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Litman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Page 1 of 1

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name:

Indications for Use:

The ActiMed ENA C-T™ Total Cholesterol Test is a non-instrumented, enzymatic assay intended for the in vitro quantitative determination of total cholesterol in fingerstick whole blood. It is indicated for physician directed use (PDU) by the patient for screening and monitoring cholesterol.

In a PDU monitoring application, the patient's physician (or other medical professional under the direction of a physician) initially trains the patient to use the test properly and provides information on the proper interpretation of results and limitations of the test. With the physician directed indication for use, patients on cholesterol lowering therapy can test their own total cholesterol at home and provide the results to their physician or other healthcare organization which works with the physician to maintain compliance on the therapy.

In a PDU screening application, a physician (or other medical professional under the direction of a physician) initially trains a group of individuals to use the test properly and provides information on the proper interpretation of results and limitations of the test. The physician directed use version of the ENA.C.T™ Total Cholesterol Test kits are then distributed to the group of individuals who can use the test at home and then communicate their results to the physician or healthcare organization.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation
----------------------------------------------------

(Division Sign-Off)

Division of Clinical Laboratory Devices

510(k) NumberK981181
------------------------
Prescription Use OROver-The Counter Use
Per 21 CFR 801.109

(Optional Format 1-2-96)

§ 862.1175 Cholesterol (total) test system.

(a)
Identification. A cholesterol (total) test system is a device intended to measure cholesterol in plasma and serum. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.