Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K233060

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-06-17

    (266 days)

    Product Code
    Regulation Number
    862.1295
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Binding assay for the in vitro quantitative determination of folate in erythrocytes (red blood cells, RBC). Folate measurements are used in the diagnosis and treatment of anemia. The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.

    Device Description

    Elecsys Folate III is a binding assay that makes use of a competitive test principle using a ruthenium labeled folate-binding assay.
    Elecsys Folate III is a binding assay for the in vitro quantitative determination of folate in erythrocytes (red blood cells, RBC). Folate measurements are used in the diagnosis and treatment of anemia. The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.
    Whole blood treated with anticoagulants (heparin or EDTA) is mixed with ascorbic acid solution and incubated for approximately 90 minutes at 20-25 °C. Lysis of the erythrocytes takes place, with liberation and stabilization of the intracellular folate. The resulting hemolysate sample is then used for subsequent measurement.
    Results are determined via a calibration curve, which is instrument-specifically generated by 2point calibration, and a master curve provided via the cobas link.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the Elecsys Folate III device, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implied / Stated)Reported Device Performance
    Precision"All predefined acceptance criteria was met for the precision experiments." (Specific numerical criteria not explicitly stated in this document but implied to be within acceptable limits as per CLSI guideline EP05-A3.)Repeatability (within-run precision) & Intermediate Precision (within-laboratory precision) (cobas e 801 analyzer)
    Analytical SensitivityBased on CLSI EP17-A2 guidelines for Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ).LoB: 45 ng/mL LoD: 70 ng/mL LoQ: 120 ng/mL
    LinearityMeasurements across the claimed measuring range (120 - 620 ng/mL) must be linear as assessed per CLSI EP06-Ed2.Linearity confirmed to support the measuring range of 120 - 620 ng/mL.
    DilutionData must support instruction for use for samples diluted 1:2.Data supports instruction for use.
    Endogenous Interferences"All predefined acceptance criteria were met" for various endogenous substances including Bilirubin, Intralipid, Biotin, Rheumatoid factors, IgG, IgA, IgM, at specified concentrations, confirming no significant interference. (Specific thresholds for non-interference not provided in text, but implied to be within acceptable limits).No significant interference for: Bilirubin: ≤ 29 mg/dL Intralipid: ≤ 1500 mg/dL Biotin: ≤ 1200 ng/mL Rheumatoid factors: ≤ 1000 IU/mL IgG: ≤ 1.6 g/dL IgA: ≤ 0.4 g/dL IgM: ≤ 1 g/dL
    Analytical Specificity/Cross-ReactivityExpected low cross-reactivity with specified compounds. (Specific thresholds for cross-reactivity not provided in text, but implied to be within acceptable limits).Low cross-reactivity: Amethopterin: (750 ng/mL) 1.7% Aminopterin: (750 ng/mL) 2.0% Folinic acid: (750 ng/mL) 2.6%
    Exogenous InterferencesNo interference from 17 commonly used pharmaceuticals and erythropoietin. (Specific thresholds for non-interference not provided in text, but implied to be within acceptable limits).No interference found from 17 commonly and 1 specially used pharmaceutical (erythropoietin) compounds.
    Sample Matrix ComparisonResults within specification, supporting the use of hemolysate prepared from whole blood treated with Na-heparin or K3-EDTA.Results were within specification and support the use of hemolysate prepared from whole blood and treated with Na-heparin or K3-EDTA.
    Method Comparison to PredicateHigh correlation and agreement with the predicate device (Elecsys Folate RBC).Number of samples: 119 (concentrations 132-618 ng/mL) Passing/Bablok: y = 1.04x - 14.6, τ = 0.913 Linear regression: y = 1.03x - 11.0, r = 0.991
    Reagent Stability (On-board)Reagent kits can be stored on-board for up to 16 weeks.Tested on one cobas e 801 analyzer; Elecsys Folate III reagent kits can be stored on-board for up to 16 weeks. (Note: The product comparison table states "2 weeks" for predicate; this indicates an improvement for the candidate device.)
    Calibration Stability (Lot)Calibration for a lot is recommended every 12 weeks; during this period, fresh kits of same lot can be used without re-calibration using the day 0 curve.Tested on one cobas e 801 analyzer. Calibration of an Elecsys Folate III reagent lot is recommended every 12 weeks.
    Calibration Stability (On-board)Reagent epacks can be stored on-board for up to 28 days without a new calibration.Tested on one cobas e 801 analyzer. Elecsys Folate III epacks can be stored on board of the analyzers for up to 28 days without a new calibration.

    2. Sample Sizes and Data Provenance

    • Precision (Repeatability & Intermediate Precision): Not explicitly stated, but typically involves multiple replicates over several days/runs with multiple instruments. Specific sample types are "Hemolysate 1" to "Hemolysate 5".
    • Lot-to-lot Reproducibility: "three reagent lots" were used.
    • Analytical Sensitivity (LoB, LoD, LoQ): Not explicitly stated, but determined according to CLSI EP17-A2, which involves specific numbers of blank and low-concentration samples.
    • Linearity and Dilution: "At least seven concentrations using hemolysate samples" for linearity. Dilution study used "high concentration hemolysate samples".
    • Endogenous Interferences: Not explicitly stated, but "various endogenous substances" were evaluated.
    • Analytical Specificity/Cross-Reactivity: Not explicitly stated.
    • Exogenous Interferences: "17 commonly and 1 specially used pharmaceutical" compounds.
    • Sample Matrix Comparison: "Whole blood samples were drawn into Na-Heparin and K3-EDTA tubes." (Number not specified).
    • Method Comparison to Predicate: 119 samples with concentrations between 132 and 618 ng/mL.
    • Reagent Stability (On-board), Lot Calibration Stability, On-board Calibration Stability: Tested on "one cobas e 801 analyzer".

    Data Provenance: The document states "NON-CLINICAL PERFORMANCE EVALUATION" and refers to CLSI (Clinical and Laboratory Standards Institute) guidelines, which are standard for in-vitro diagnostic device performance studies. The data is internal to Roche Diagnostics, a company with global operations. The specific country of origin for the studies is not stated, but given the submission is to the U.S. FDA, the studies are expected to meet international and U.S. regulatory standards. These are retrospective studies in the context of device development and validation.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    This device is an in-vitro diagnostic (IVD) assay that produces quantitative measurements of folate. It does not involve human interpretation of images or other subjective data. Therefore, the "ground truth" for its performance is established by reference methods, calibrated standards, and accurate measurement principles, rather than human expert consensus. No human experts were used to establish ground truth in the way one would for an AI imaging device.

    4. Adjudication Method (Test Set)

    Not applicable, as this is an IVD device producing quantitative results, not an AI imaging device requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is an in-vitro diagnostic device, not an AI device assisting human readers with interpreting cases.

    6. Standalone (Algorithm Only) Performance

    Yes, the studies described are standalone performance evaluations of the Elecsys Folate III assay (the "algorithm" in a broad sense for an IVD) without human intervention in the measurement process. The device performs the quantitative determination of folate using its defined binding assay and electrochemiluminescence immunoassay (ECLIA) method.

    7. Type of Ground Truth Used

    The ground truth for evaluating the Elecsys Folate III's performance is established through:

    • Reference materials/standards: For analytical sensitivity (LoB, LoD, LoQ) and linearity, where known concentrations are used.
    • Performance against a predicate method: For method comparison, the results are compared to a previously cleared, established method (Elecsys Folate RBC).
    • CLSI guidelines: Adherence to established scientific and statistical methodologies for validating assay performance, implying well-defined benchmarks for accuracy, precision, and interference.

    8. Sample Size for the Training Set

    Not explicitly stated. For an IVD like the Elecsys Folate III, a traditional "training set" as understood in machine learning is not directly applicable. The "training" for such a system would involve the development and optimization of the reagent formulations, assay parameters, and calibration curve algorithms, using various samples and experiments during the research and development phase. However, a specific training set size with corresponding ground truths is not documented in the same way as an AI algorithm.

    9. How the Ground Truth for the Training Set was Established

    As above, a formal "training set ground truth" isn't directly applicable in the machine learning sense. The "ground truth" during the development and optimization (analogous to training) would have been established through:

    • Known concentrations: Using purified folate standards or spiked samples.
    • Reference methods: Comparing early-stage assay performance against established, often more labor-intensive or gold-standard methods for folate determination.
    • Clinical correlation: (Though "clinical testing" is marked Not Applicable for this 510(k), early development might involve correlation with clinical status or outcomes).
    • Statistical optimization: Adjusting reagent ratios, incubation times, and instrument settings to achieve optimal analytical performance characteristics (sensitivity, specificity, precision, linearity) based on these known values and reference method comparisons.
    Ask a Question

    Ask a specific question about this device

    K Number
    K141426

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-10-17

    (140 days)

    Product Code
    Regulation Number
    862.1295
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Binding assay for the in vitro quantitative determination of folate in human serum. The binding assay is intended for use on Elecsys and cobas e immunoassay analyzers. Folic acid measurements are used in the diagnosis and treatment of anemias.

    Device Description

    The Folate III Assay employs a competitive test principle using natural folate binding protein (FBP) specific for folate. Folate in the sample competes with the added folate (labeled with biotin) for the binding sites on FBP (labeled with a ruthenium complex). Results are determined using a calibration curve that is generated specifically on each instrument by a 2 point calibration and a master curve (5-point-calibration) provided with the reagent bar code.

    AI/ML Overview

    Roche Diagnostics' Elecsys Folate III Assay is a quantitative in vitro diagnostic device designed for measuring folate levels in human serum, primarily for the diagnosis and treatment of anemias. The following outlines its acceptance criteria and the studies performed to demonstrate its performance.

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Elecsys Folate III Assay are primarily established through comparison to a predicate device, the Roche Elecsys Folate III (K082340). The device demonstrates comparable or improved analytical performance characteristics.

    FeaturePredicate Device: Roche Elecsys Folate III (K082340)Candidate Device: Elecsys Folate III AssayAcceptance Criteria (Implied by equivalence)
    Measuring Range1.50 - 20.0 ng/mL2.0 - 20.0 ng/mLSimilar or improved range
    Analytical Sensitivity
    Limit of Blank (LoB)0.640 ng/mL0.6 ng/mLEqual to or lower than predicate
    Limit of Detection (LoD)1.50 ng/mL1.2 ng/mLEqual to or lower than predicate
    Limit of Quantitation (LoQ)2.0 ng/mL2.0 ng/mLEqual to or lower than predicate
    Analytical Specificity (Cross-reactivity)
    Amethopterin2.3% at 750 ng/mL2.5% at 1500 ng/mLComparable or improved
    Aminopterin2.7% at 750 ng/mL4.4% at 1500 ng/mLComparable or improved
    Folonic acid2.3% at 750 ng/mL0.7% at 1500 ng/mLComparable or improved
    Precision (cobas e 411)
    Within-run (Repeatability)%CV range: 3.7% - 5.8%%CV range: 2.2% - 6.8%Comparable to predicate
    Total (Intermediate)%CV range: 6.1% - 9.4%%CV range: 3.7% - 10.8%Comparable to predicate
    Limitations (Interference)
    Bilirubinunaffected < 33 mg/dLunaffected < 29 mg/dLComparable
    Lipemiaunaffected < 1500 mg/dLunaffected < 1500 mg/dLComparable
    Biotinunaffected < 21 ng/mLunaffected < 21 ng/mLComparable
    Rheumatoid factorsunaffected < 1000 IU/mLunaffected < 1000 IU/mLComparable
    IgGunaffected < 1.6 g/dLunaffected < 1.6 g/dLComparable
    IgAunaffected < 0.4 g/dLunaffected < 0.4 g/dLComparable
    IgMNot specifiedunaffected < 1.0 g/dLComparable
    Reference Range Study (USA)
    Median31.8 nmol/L (14.0 ng/mL)26.8 nmol/L (11.8 ng/mL)Established reference range
    2.5th-97.5th percentile16.6-59.3 nmol/L (7.3-26.1 ng/mL)10.9-54.9 nmol/L (4.78-24.2 ng/mL)Established reference range
    Method Comparison (vs. Abbott Architect Folate K092740)Good correlation
    Slope-Passing/Bablok: 0.980, Linear regression: 0.976Near 1.0
    Intercept-Passing/Bablok: -0.095, Linear regression: 0.041Near 0.0
    Tau/r-Passing/Bablok: 0.924, Linear regression: 0.984High value, indicating strong correlation

    Study Details

    The provided document describes a premarket notification (510(k)) to demonstrate substantial equivalence to a legally marketed predicate device (Roche Elecsys Folate III, K082340). The studies conducted are analytical performance studies.

    1. Sample sizes used for the test set and the data provenance:

      • Precision (Repeatability/Intermediate): Human serum (HS) samples (5 levels) and PreciControl Varia (PCV) control levels (2 levels) were used. The document lists mean, SD, and CV values, which are derived from repeated measurements of these samples. The exact number of individual patient samples in the precision study is not explicitly stated, but typically these studies involve multiple runs over several days.
      • Analytical Specificity (Cross-reactivity): Specific cross-reactants (Amethopterin, Aminopterin, Folonic acid) were tested at concentrations of 1500 ng/mL.
      • Limitations (Interference): Bilirubin (< 29 mg/dL), Lipemia (< 1500 mg/dL), Biotin (< 21 ng/mL), Rheumatoid factors (< 1000 IU/mL), IgG (< 1.6 g/dL), IgM (< 1.0 g/dL), IgA (< 0.4 g/dL) were tested. Also, 16 commonly used pharmaceuticals and human erythropoietin were tested. The number of samples for each interference test is not specified.
      • Reference Range Study:
        • Predicate Device: N = 261 subjects from USA.
        • Candidate Device: N = 214 subjects from USA.
        • Provenance: All reference range data is from the USA. These appear to be prospective studies to establish healthy reference ranges.
      • Method Comparison: n = 106 patient samples were used, with folate concentrations ranging from 2.08 ng/mL to 19.6 ng/mL.
        • Provenance: Not explicitly stated, but typically these are retrospective or prospectively collected clinical samples.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
      This is an in vitro diagnostic (IVD) device for quantitative measurement of an analyte (folate). "Ground truth" in this context refers to the concentration of folate in the samples, determined by a reference method or the predicate device, rather than expert interpretation of images or clinical outcomes. Therefore, no "experts" in the sense of medical specialists reviewing results are mentioned as establishing ground truth. The ground truth for method comparison is the result obtained from the predicate device (Abbott Architect Folate assay, K092740).

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      Not applicable, as this is an IVD device for quantitative measurement, not an assessment of diagnostic performance based on human interpretation or consensus.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable. This is an IVD device, not an AI-assisted diagnostic imaging device that involves human readers.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Yes, the performance characteristics (e.g., precision, measuring range, analytical sensitivity, specificity) described are for the Elecsys Folate III Assay itself, operating as a standalone algorithm/device. The method comparison study directly compares the results of the candidate device with another established assay.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Analytical Performance: The ground truth for evaluating analytical performance parameters like LoB, LoD, LoQ, precision, and specificity is typically based on control materials, spiked samples, and established analytical methods.
      • Method Comparison: The ground truth for the method comparison study was the results obtained from the predicate device, the Abbott Architect Folate assay (K092740). The acceptance criteria for this comparison would be for the candidate device's results to closely agree with those of the predicate.
      • Reference Range: The reference range was established by testing samples from a healthy population in the USA (N=214).
    7. The sample size for the training set:
      The document does not specify a "training set" in the context of statistical modeling or machine learning. This is an immunoassay device, and its calibration curve is generated using a "2 point calibration and a master curve (5-point-calibration) provided with the reagent bar code" (page 3). The master curve is based on internal validation data, but a specific "training set sample size" for model development is not provided.

    8. How the ground truth for the training set was established:
      As above, the concept of a "training set" with established ground truth as in machine learning is not directly applicable to this immunoassay. The device's operational characteristics and performance are based on analytical validation following established guidelines (CLSI documents EP5-A2, EP17-A2, EP6-A, EP-09-A2-IR mentioned on page 12). The calibration curve, which is critical to the assay's function, is standardized with a master curve and a 2-point calibration, and the device is standardized against the WHO International Standard NIBSC code: 03/178 (page 7).

    Ask a Question

    Ask a specific question about this device

    K Number
    K094031

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-04-27

    (118 days)

    Product Code
    Regulation Number
    862.1660
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Elecsys Folate III CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Folate III reagent on the indicated Elecsys and cobas e immunoassay analyzers.

    Device Description

    The Elecsys Folate III CalCheck 5 is a lyophilized product consisting of Folate in human serum matrix. During manufacture, the analyte is spiked into the matrix at the desired concentration levels.

    AI/ML Overview

    The provided 510(k) summary (K094053) describes a device called "Elecsys Folate III CalCheck 5," which is an assayed control for use in calibration verification and verification of the assay range for the Elecsys Folate III reagent on immunoassay analyzers. This type of device is a quality control material and as such, the submission does not involve clinical studies with human participants, expert ground truth, or multi-reader multi-case studies, which are typically associated with diagnostic or AI-powered medical devices. The acceptance criteria and performance data for this device relate to its analytical accuracy and precision in controlling an in vitro diagnostic assay.

    Therefore, many of the requested categories are not applicable or cannot be extracted from this specific document, as they pertain to clinical efficacy studies rather than analytical performance of a control solution.

    Here's an analysis based on the available information:

    1. A table of acceptance criteria and the reported device performance

    The provided 510(k) summary does not contain a table of explicit acceptance criteria or reported device performance in the format typically seen for a new diagnostic device. For an assayed control, acceptance criteria would typically involve demonstrating that the control material consistently yields expected values within a defined range when run on the target assay. The submission primarily focuses on establishing substantial equivalence to a predicate device.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable directly in the context of a quality control material. The "test set" for an assayed control would typically involve rigorous analytical testing in a laboratory setting to confirm its assigned values and stability. The 510(k) summary does not detail the size or provenance of such analytical testing data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth, in the sense of expert clinical diagnosis, is not relevant for an assayed control material. The "ground truth" for a control material is its assigned concentration, which is determined through a robust analytical process by the manufacturer, not by clinical experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are used in clinical studies to resolve discrepancies among expert readers, which is not relevant for an assayed control.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a quality control material, not a diagnostic device involving human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an assayed control, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable in the clinical sense. For a quality control material, the "ground truth" is typically the assigned value of the analyte (Folate in this case) within the control, which is established by the manufacturer through precise analytical methods and verified against reference materials or methods. The 510(k) summary doesn't detail this process.

    8. The sample size for the training set

    Not applicable. This device is a quality control material, not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As above, this is not an algorithm requiring a training set.

    Summary of the K094053 Submission:

    The K094053 submission for the Elecsys Folate III CalCheck 5 is for an assayed control used to verify the performance of an in vitro diagnostic assay (Elecsys Folate III reagent). As such, the focus of the submission is on demonstrating its analytical performance and substantial equivalence to a predicate device (Elecsys HCG+β CalCheck 5 (K092168)). The 510(k) summary describes the device as a lyophilized product consisting of Folate in a human serum matrix, with analyte spiked in at desired concentrations.

    For devices of this nature, "acceptance criteria" and "study data" would typically refer to:

    • Accuracy: The ability of the control to yield results within a specified range relative to its assigned value.
    • Precision (Repeatability and Reproducibility): The consistency of results when the control is run multiple times, by different operators, or on different instruments.
    • Stability: The ability of the control to maintain its assigned values over its shelf life and during storage.
    • Matrix Equivalence: Ensuring the control matrix behaves similarly to patient samples in the assay.

    These analytical performance characteristics would be demonstrated through internal validation studies by Roche Diagnostics. However, the provided 510(k) summary document itself does not include the detailed analytical study report, raw data, or a explicit table of acceptance criteria and results for the Elecsys Folate III CalCheck 5. Such details are typically found in the full 510(k) submission, but are not always part of the publicly available summary. The summary focuses on the administrative details, device description, and intended use to establish substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1