K Number
K094031

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-04-27

(118 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Elecsys Folate III CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Folate III reagent on the indicated Elecsys and cobas e immunoassay analyzers.

Device Description

The Elecsys Folate III CalCheck 5 is a lyophilized product consisting of Folate in human serum matrix. During manufacture, the analyte is spiked into the matrix at the desired concentration levels.

AI/ML Overview

The provided 510(k) summary (K094053) describes a device called "Elecsys Folate III CalCheck 5," which is an assayed control for use in calibration verification and verification of the assay range for the Elecsys Folate III reagent on immunoassay analyzers. This type of device is a quality control material and as such, the submission does not involve clinical studies with human participants, expert ground truth, or multi-reader multi-case studies, which are typically associated with diagnostic or AI-powered medical devices. The acceptance criteria and performance data for this device relate to its analytical accuracy and precision in controlling an in vitro diagnostic assay.

Therefore, many of the requested categories are not applicable or cannot be extracted from this specific document, as they pertain to clinical efficacy studies rather than analytical performance of a control solution.

Here's an analysis based on the available information:

1. A table of acceptance criteria and the reported device performance

The provided 510(k) summary does not contain a table of explicit acceptance criteria or reported device performance in the format typically seen for a new diagnostic device. For an assayed control, acceptance criteria would typically involve demonstrating that the control material consistently yields expected values within a defined range when run on the target assay. The submission primarily focuses on establishing substantial equivalence to a predicate device.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable directly in the context of a quality control material. The "test set" for an assayed control would typically involve rigorous analytical testing in a laboratory setting to confirm its assigned values and stability. The 510(k) summary does not detail the size or provenance of such analytical testing data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. Ground truth, in the sense of expert clinical diagnosis, is not relevant for an assayed control material. The "ground truth" for a control material is its assigned concentration, which is determined through a robust analytical process by the manufacturer, not by clinical experts.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are used in clinical studies to resolve discrepancies among expert readers, which is not relevant for an assayed control.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a quality control material, not a diagnostic device involving human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is an assayed control, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable in the clinical sense. For a quality control material, the "ground truth" is typically the assigned value of the analyte (Folate in this case) within the control, which is established by the manufacturer through precise analytical methods and verified against reference materials or methods. The 510(k) summary doesn't detail this process.

8. The sample size for the training set

Not applicable. This device is a quality control material, not a machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable. As above, this is not an algorithm requiring a training set.

Summary of the K094053 Submission:

The K094053 submission for the Elecsys Folate III CalCheck 5 is for an assayed control used to verify the performance of an in vitro diagnostic assay (Elecsys Folate III reagent). As such, the focus of the submission is on demonstrating its analytical performance and substantial equivalence to a predicate device (Elecsys HCG+β CalCheck 5 (K092168)). The 510(k) summary describes the device as a lyophilized product consisting of Folate in a human serum matrix, with analyte spiked in at desired concentrations.

For devices of this nature, "acceptance criteria" and "study data" would typically refer to:

  • Accuracy: The ability of the control to yield results within a specified range relative to its assigned value.
  • Precision (Repeatability and Reproducibility): The consistency of results when the control is run multiple times, by different operators, or on different instruments.
  • Stability: The ability of the control to maintain its assigned values over its shelf life and during storage.
  • Matrix Equivalence: Ensuring the control matrix behaves similarly to patient samples in the assay.

These analytical performance characteristics would be demonstrated through internal validation studies by Roche Diagnostics. However, the provided 510(k) summary document itself does not include the detailed analytical study report, raw data, or a explicit table of acceptance criteria and results for the Elecsys Folate III CalCheck 5. Such details are typically found in the full 510(k) submission, but are not always part of the publicly available summary. The summary focuses on the administrative details, device description, and intended use to establish substantial equivalence.

{0}------------------------------------------------

K0940531

APR 2 7 2010

510(k) Summary

IntroductionAccording to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
Submittername, address,contactRoche Diagnostics9115 Hague Road, P.O. Box 50416Indianapolis, IN 46250-0416317-521-3577
Contact Person: Kelly Colleen O'Maine AdamsPhone: 317-521-3577Fax: 317-521-2324Email: colleen.adams@roche.com
Secondary Contact: Stephanie GreemanPhone: 317-521-2458Fax: 317-521-2324Email: stephanie.greeman@roche.com
Date Prepared: December 29, 2009
Device NameProprietary name: Elecsys Folate III CalCheck 5Common name: Folate CalCheck 5Classification name: Single (specified) analyte controls (assayed and unassayed)
PredicatedeviceThe Elecsys Folate III CalCheck 5 is substantially equivalent to other products in commercial distribution intended for similar use. We claim equivalency to the currently marketed Elecsys HCG+β CalCheck 5 (K092168).
DeviceDescriptionThe Elecsys Folate III CalCheck 5 is a lyophilized product consisting of Folate in human serum matrix. During manufacture, the analyte is spiked into the matrix at the desired concentration levels.
Intended useThe Elecsys Folate III CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Folate III reagent on the indicated Elecsys and cobas e immunoassay analyzers.
Continued on next page

Page 1 of 2

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a seal for the Department of Health & Human Services (HHS) in the USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002

Roche Diagnostics c/o Kelly Colleen O'Maine Adams Regulatory Affairs Consultant 9115 Hague Road P.O. Box 50416 Indianapolis, IN 46250-0416

APR 2 7 2010

K094031 Re:

Trade Name: Elecsys Folate III CalCheck 5 Regulation Number: 21 CFR 8862,1660 Regulation Name: Quality Control Material (assayed and unassayed) Regulatory Class: Class I. Reserved Product Codes: JJX Dated: March 10, 2010 Received: March 11, 2010

Dear Mrs. O'Maine Adams:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{2}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Signature

Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indication for Use

510(k) Number (if known):

Device Name: Elecsys Folate III CalCheck 5

Indication For Use:

The Elecsys Folate III CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Folate III reagent on the indicated Elecsys and cobas e immunoassay analyzers.

Prescription Use _ X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Carol Benson
Division Sign Off

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K094031

Page 1 of 1

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.