Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K212600

    Validate with FDA (Live)

    Date Cleared
    2021-11-18

    (93 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Medical Nitrile Examination Gloves (Non Sterile) are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    The subject device in this 510(k) Notification is Disposable Medical Nitrile Examination Gloves (Non sterile). The subject device is a patient examination glove made from nitrile compound, blue color, powder free and non-sterile (Per 21 CFR 880.6250, Class I) for single use only. The device meets the specifications in ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for Disposable Medical Nitrile Examination Gloves (Non-Sterile). This type of document is for a medical device that does not use AI/ML, and therefore, the requested information regarding AI/ML device performance and related study details is not applicable.

    However, I can extract the acceptance criteria and performance data for this device based on the non-clinical testing summary provided.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance/Result
    ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Dimensions TestTo determine the width, length, and thickness of the glovesWidth:80mm ±10mm (for small size)95mm ±10mm (for medium size)110 mm ±10mm (for large size)Length:220 mm (Minimum) (for small size)230 mm (Minimum)(for medium, large size)Thickness:Finger -0.05 mm (Minimum) (for small, medium, large size)Palm -0.05 mm (Minimum)(for small, medium, large size)Width:small size: 82 to 89mmmedium size: 91 to 96mmlarge size: 115 to 116mmLength:small size: 230mm (minimum)medium size: 240mm (minimum)large size: 238mm (minimum)Thickness:Finger:small size: 0.12mm (minimum)medium size: 0.13mm (minimum)large size: 0.12mm (minimum)Palm:small size: 0.08 mm (minimum)medium size: 0.08mm (minimum)large size: 0.09 mm (minimum)Passed
    ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Requirements TestTo determine the tensile strength and ultimate elongation before and after acceleration agingBefore Acceleration Aging:Tensile Strength (MPa): 14 (Minimum)Ultimate Elongation (%): 500 (Minimum)After Acceleration Aging:Tensile Strength (MPa): 14 (Minimum)Ultimate Elongation (%): 400 (Minimum)Before Acceleration Aging:Tensile Strength (MPa): 20 (Minimum)Ultimate Elongation (%): 521 (Minimum)After Acceleration Aging:Tensile Strength (MPa): 19 (Minimum)Ultimate Elongation (%): 505 (Minimum)Passed
    ASTM D6319-19 (ASTM D5151-11) Standard Test Method for Detection of Holes in Medical GlovesTo determine the holes in the gloves(Mean) AQL 2.5Passed
    ASMT D6319-19 (ASTM D6124-11) Standard Test Method for Residual Powder on Medical GlovesTo determine the residual powder in the gloves< 2.0 mg/glove0.33 mg/glove, Passed
    ISO 10993-10 Biological evaluation of medical devices — Part 10: Tests for skin irritation and skin sensitizationTo determine the potential of the material under test to produce skin irritation in rabbitsUnder the condition of study, testing articles are not an irritantUnder the conditions of study, the test article did not show significant evidence of causing skin sensitization in the guinea pig. (Note: Result listed under the "irritation" row in the document is the sensibility test result for the guinea pig).
    ISO 10993-10 Biological evaluation of medical devices — Part 10: Tests for skin irritation and skin sensitizationTo determine the skin sensitization potential of the material both in terms of induction and elicitation in guinea pigs.Under the conditions of the study, the testing articles are not a sensitizer.Under the conditions of the study the sample did not induce skin irritation. (Note: Result listed under the "sensitization" row in the document is the irritation test result).
    ISO 10993-5 Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicityTo evaluate the in vitro cytotoxic potential of the test item (both inner and outer surface) Extracts in L-929 mouse fibroblasts cells using elution methodUnder the conditions of study, no more than grade 2 cytotoxic reactionMild (Grade 2) cytotoxicity reaction observed. Passed.

    Since this document describes a Class I medical device (Disposable Medical Nitrile Examination Gloves), which is a low-risk device, the requirements for its acceptance parallel those of general product testing and biocompatibility, as opposed to complex AI/ML algorithms. Therefore, the following points regarding AI/ML specific study information are not applicable to this submission:

    1. Sample sizes used for the test set and the data provenance: Not applicable. Testing involves physical properties and biocompatibility, not data sets for an AI model.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for gloves (e.g., tensile strength, freedom from holes) is established through standardized physical and chemical testing, not expert interpretation of AI outputs.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The ground truth for this device is based on established material science and biological testing standards (e.g., ASTM and ISO standards) which define performance metrics for physical properties (e.g., length, width, thickness, tensile strength, elongation), chemical properties (e.g., powder residue), and biological responses (e.g., skin irritation, sensitization, cytotoxicity).
    7. The sample size for the training set: Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212029

    Validate with FDA (Live)

    Date Cleared
    2021-11-16

    (140 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Medical Nitrile Examination Glove is intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The subject device is powder free nitrile examination gloves. The subject device is blue. The subject device is non-sterile.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted for a medical device: "Disposable Medical Nitrile Examination Gloves (non-sterile)." This is a Class I medical device, and the study focused on non-clinical tests to demonstrate substantial equivalence to a predicate device, rather than the types of studies typically performed for AI/ML-based diagnostic devices. Therefore, many of the requested categories are not applicable.

    Here's the information extracted from the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestPurposeAcceptance CriteriaReported Device Performance
    ISO 10993-5 Biological evaluation of medical devices Test for in vitro cytotoxicityTo determine the potential cytotoxicity.Non-cytotoxicPass
    ISO 10993-11 Biological evaluation of medical devices Part 11: Tests for Systemic ToxicityTo determine the potential systemic toxicity.Does not cause systemic toxicity.Pass
    ISO 10993-10 Biological evaluation on medical device Part 10: Test for Irritation and Skin IrritationTo determine the potential for irritation and skin irritation.Non-irritant, and Non skin irritant.Pass
    ISO 10993-10 Biological evaluation on medical device Part 10: Test Skin SensitizationTo determine the potential skin sensitization.Does not cause skin sensitization.Pass
    ASTM D6319 Standard Specification for Nitrile Examination Gloves for Medical ApplicationTo test for: (1) freedom from holes (2) physical dimensions (3) Aging(1) Shall not leak (2) For size L (mm): Width: 110 ± 10, Length: ≥ 230, Finger and palm thickness: median value ≥ 0.05 (3) After Aging: Tensile Strength: ≥ 14 MPa, Ultimate elongation: ≥400%Pass

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in the context of data for an AI/ML device. The tests performed are for medical gloves and involve physical and biological property assessments, not analysis of data from patients or images. The provenance of the samples (e.g., country of origin for the gloves tested) is not specified beyond the manufacturer's location (China). These were non-clinical tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The ground truth for the physical and biological properties of the gloves is established by standardized testing protocols (ISO, ASTM) and measurements, not by expert interpretation of data relevant to AI/ML.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. As noted above, the tests are for physical and biological properties of gloves, not for AI/ML performance requiring adjudication of interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical glove, not an AI-based diagnostic tool. No MRMC study was conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a medical glove, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the non-clinical tests is based on objective measurements and established standards for material properties (e.g., tensile strength, elongation), dimensional specifications (e.g., length, width, thickness), and biocompatibility assays (e.g., cytotoxicity, irritation, sensitization, systemic toxicity).

    8. The sample size for the training set

    Not applicable. This device does not involve an AI/ML algorithm requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. This device does not involve an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211075

    Validate with FDA (Live)

    Date Cleared
    2021-08-05

    (115 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Medical Nitrile Examination Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    The subject device in this 510(k) Notification is Disposable Medical Nitrile Examination Gloves, Non-sterile. The subject device is a patient examination glove made from nitrile compound, blue color, powder free and non-sterile (Per 21 CFR 880.6250, Class I). The device meets the specifications in ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.

    AI/ML Overview

    The document describes the acceptance criteria and study proving the device meets them for Disposable Medical Nitrile Examination Gloves (Non-Sterile) (K211075).

    Here's the breakdown of the information requested:


    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodAcceptance CriteriaReported Device Performance
    ASMT D6319-19 - Physical Dimensions Test (Width)Small size: 80 ± 10 mm Medium size: 95 ± 10 mm Large size: 110 ± 10 mm Extra-large size: 120 ± 10 mmSmall size: 82mm (Minimum), 82 mm (Mean), 85 mm (Maximum) Medium size: 91 mm (Minimum), 92 mm (Mean), 93 mm (Maximum) Large size: 108 mm (Minimum), 109 mm (Mean), 110 mm (Maximum) Extra-large size: 121 mm (Minimum), 122 mm (Mean), 125 mm (Maximum) All Passed
    ASMT D6319-19 - Physical Dimensions Test (Length)Small size: 220 mm (Minimum) Medium size: 230 mm (Minimum) Large size: 230 mm (Minimum) Extra-large size: 230 mm (Minimum)Small size: 232 mm (Mean), 230 mm (Minimum) Medium size: 242 mm (Mean), 240 mm (Minimum) Large size: 239 mm (Mean), 237 mm (Minimum) Extra-large size: 243 mm (Mean), 240 mm (Minimum) All Passed
    ASMT D6319-19 - Physical Dimensions Test (Thickness)Small size: Finger – 0.05 mm (Minimum), Palm – 0.05 mm (Minimum) Medium size: Finger – 0.05 mm (Minimum), Palm – 0.05 mm (Minimum) Large size: Finger – 0.05 mm (Minimum), Palm – 0.05 mm (Minimum) Extra-large size: Finger – 0.05 mm (Minimum), Palm – 0.05 mm (Minimum)Small size: Finger – 0.13 mm (Mean), 0.12 mm (Minimum); Palm – 0.08 mm (Mean), 0.08 mm (Minimum) Medium size: Finger – 0.14 mm (Mean), 0.13 mm (Minimum); Palm – 0.09 mm (Mean), 0.08 mm (Minimum) Large size: Finger – 0.14 mm (Mean), 0.13 mm (Minimum); Palm – 0.09 mm (Mean), 0.09 mm (Minimum) Extra-large size: Finger – 0.14 mm (Mean), 0.13 mm (Minimum); Palm – 0.09 mm (Mean), 0.08 mm (Minimum) All Passed
    ASMT D6319-19 - Physical Requirements Test (Tensile Strength & Elongation)Before Aging: Tensile Strength 14 MPa (Minimum), Ultimate Elongation 500% (Minimum) After Acceleration Aging: Tensile Strength 14 MPa (Minimum), Ultimate Elongation 400% (Minimum)Before Acceleration Aging: Tensile Strength 31 MPa (Mean), 17 MPa (Minimum); Ultimate Elongation 563% (Mean), 520% (Minimum) After Acceleration Aging: Tensile Strength 41 MPa (Mean), 28 MPa (Minimum); Ultimate Elongation 555% (Mean), 524% (Minimum) Passed
    ASTM D6319-19 (ASTM D5151-11) - Freedom from Holes (Water Tight)AQL 2.5Passed
    ASMT D6319-19 (ASTM D6124-11) - Residual Powder< 2.0 mg/glove1.02 mg/glove Passed
    ISO 10993-10 - Skin IrritationUnder the condition of study not an irritantUnder the condition of study not an irritant
    ISO 10993-10 - Skin SensitizationUnder the conditions of the study not a sensitizer.Under the conditions of the study not a sensitizer.
    ISO 10993-5 - In Vitro CytotoxicityUnder the conditions of study non cytotoxicUnder the condition of the study, the test article showed no cytotoxicity to L-929 cells and cytotoxic level was 0.
    ISO 10993-11:2017 - Acute Systemic ToxicityUnder the conditions of the study no systemic toxicityUnder the conditions of the study no systemic toxicity

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact "sample size" for each test in terms of the number of individual gloves or test replicates. However, it indicates that the testing was performed according to ASTM D6319-19, which specifies sampling plans and test methods for nitrile examination gloves. For example, for "Freedom from Holes," it refers to an AQL (Acceptance Quality Limit) of 2.5, which is a statistical measure related to sampling.

    • Data Provenance: The testing was non-clinical. The specific country of origin for the data is not explicitly stated beyond the manufacturer being in China (Jiangsu Cherish Medical Technology Co., Ltd). The studies are retrospective in the sense that they were conducted on already manufactured gloves.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • This information is not applicable as the document describes non-clinical performance testing of a physical medical device (gloves), not a diagnostic algorithm requiring expert ground truth or consensus. The "ground truth" is established by the standardized test methods themselves (e.g., measuring dimensions, tensile strength, presence of holes).

    4. Adjudication Method for the Test Set

    • This information is not applicable for the reasons stated above. Adjudication methods are typically relevant for human interpretation or AI model performance evaluation against expert-derived ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices often involving human readers or AI algorithms for interpretation, not for examination gloves.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Not applicable. This document evaluates a physical medical device (gloves), not a software algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for this device's performance is established by standardized test methods and their defined specifications. These include:

    • ASTM D6319-19 (Standard Specification for Nitrile Examination Gloves for Medical Application)
    • ASTM D5151-11 (Standard Test Method for Detection of Holes in Medical Gloves)
    • ASTM D6124-11 (Standard Test Method for Residual Powder on Medical Gloves)
    • ISO 10993-10 (Biological Evaluation of Medical Devices – Tests for skin irritation and skin sensitization)
    • ISO 10993-5 (Biological Evaluation of Medical Devices – Tests for in vitro cytotoxicity)
    • ISO 10993-11:2017 (Biological Evaluation of Medical Devices – Tests for acute systemic toxicity)

    8. Sample Size for the Training Set

    • Not applicable. This device is a physical product (examination gloves), not an AI/ML model for which "training sets" are used.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1