Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K232624

    Validate with FDA (Live)

    Date Cleared
    2023-11-27

    (90 days)

    Product Code
    Regulation Number
    866.5270
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CardioPhase hsCRP is an in-vitro diagnostic reagent for the quantitative determination of C-reactive protein (CRP) in human serum, and heparin and EDTA plasma by means of particle enhanced immunonephelometry using the BN II and BN ProSpec® System. In acute phase response, increased levels of a number of plasma proteins, including C-reactive protein, is observed. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. High sensitivity CRP (hsCRP) measurements may be used as an independent risk marker for the identification of individuals at risk for future cardiovascular disease. Measurements of hsCRP, when used in conjunction with traditional clinical laboratory evaluation of acute coronary syndromes, may be useful as an independent marker of prognosis for recurrent events, in patients with stable coronary disease or acute coronary syndromes.

    Device Description

    The CardioPhase hsCRP assay is an in-vitro diagnostic reagent for the quantitative determination of Creactive protein (CRP) in human serum, and heparinized and EDTA plasma by means of particleenhanced immunoassay determination. The assay is traceable to the international standard ERM-DA474/IFCC. N Rheumatology Standard SL (cleared under K964527) is used for the establishment of reference curves for the immunonephelometric determination of C-reactive protein on the BN II and BN ProSpec® Systems. This calibrator consists of a mixture of human sera and elevated concentrations of CRP. The CardioPhase hsCRP reagent is a suspension of polystyrene (Latex) particles to which mouse monoclonal anti-human CRP antibodies (< 0.016 g/L) have been attached by covalent bonding. The reagent is a ready-to-use liquid containing preservatives. There are two product variants available. One variant (REF OQIY13) contains 3 x 2 mL vials / box, and the other variant (REF OQIY21) contains 5 x 5 mL vials / box. The assay's polystyrene particles coated with monoclonal antibodies specific to human CRP are aggregated when mixed with samples containing CRP. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the relevant protein in the sample. The result is evaluated by comparison with a standard of known concentration.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the method comparison study were that "Results from each lot of CardioPhase hsCRP met the predefined acceptance criteria." While the specific numerical acceptance criteria (e.g., maximum allowable bias) are not explicitly detailed in a table format within the provided text, the successful outcome is stated, and the resulting performance is presented as follows:

    Performance MetricLot 1 CardioPhase hsCRPLot 2 CardioPhase hsCRPLot 3 CardioPhase hsCRP
    Sample Size (N)119116113
    Range5.523 – 197.746 mg/L5.378 – 199.150 mg/L5.501 – 199.503 mg/L
    Regression Equation (y = mx + b)y = 0.959x + 0.932 mg/Ly = 0.955x + 0.584 mg/Ly = 1.032x - 0.070 mg/L
    Correlation Coefficient (r)0.9940.9960.994
    Coefficient of Determination (r²)0.9890.9910.989
    Observed Max Predicted Bias (for 10, 100, 150, 200 mg/L)5.2% (relative)Not explicitly stated per lot, but given as overall maximum.Not explicitly stated per lot, but given as overall maximum.
    Overall Max Predicted Bias5.2% (relative)5.2% (relative)5.2% (relative)

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Sample Size:
      • Lot 1: N = 119
      • Lot 2: N = 116
      • Lot 3: N = 113
      • The total number of samples used in the method comparison study is the sum of these, which is 348.
    • Data Provenance: The study was conducted at the "company site in Marburg, Germany." The samples used were "Native serum samples." The text does not explicitly state whether the samples were retrospective or prospective, but the phrasing "Native serum samples were measured" suggests they were existing samples at the time of the study rather than collected specifically for this study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the text. The study describes a method comparison between two quantitative laboratory assays (CardioPhase hsCRP and RCRP Flex reagent cartridge). For this type of in-vitro diagnostic device, the "ground truth" is typically defined by the reference method or the predicate device's measurement, not by human expert consensus or adjudication in the way it might be for image-based diagnostic AI.

    4. Adjudication Method for the Test Set

    Not applicable for this type of in-vitro diagnostic device and study design. The comparison is quantitative between two analytical methods, not involving human interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for image analysis or other diagnostic tools where human interpretation is a key component. The CardioPhase hsCRP is an in-vitro diagnostic reagent for quantitative measurement, which does not involve human readers in the same way.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the study described is a standalone performance study of the CardioPhase hsCRP assay compared to a predicate device. It evaluates the device's ability to quantitatively determine C-reactive protein concentrations independently. No human-in-the-loop component is mentioned for the performance evaluation itself.

    7. The Type of Ground Truth Used

    The "ground truth" for this method comparison study was established by the predicate device, RCRP Flex® reagent cartridge, which runs on the Dimension clinical chemistry system. Both the proposed device (CardioPhase hsCRP) and the predicate device are traceable to the international standard ERM-DA474/IFCC for C-reactive protein measurements. Therefore, the predicate device's measurements serve as the reference for comparison, and that reference itself is traceable to an international standard.

    8. The Sample Size for the Training Set

    The text does not specify a separate training set or its sample size. The described "method comparison study" is focused on verifying the performance of the device for regulatory submission, using a test set of samples. For in-vitro diagnostic devices, "training sets" are usually relevant for developing the assay itself (e.g., optimizing reagent concentrations, reaction conditions), but this information is not typically detailed in a 510(k) summary with respect to a "training set" of patient data for algorithm development.

    9. How the Ground Truth for the Training Set Was Established

    As no specific "training set" of patient samples is described in the provided text in the context of algorithm development, this information is not applicable. The assay itself relies on a biochemical principle and calibration traceable to an international standard (ERM-DA474/IFCC), and the calibration is established using N Rheumatology Standard SL, which is traceable to Siemens internal Master Calibrator, which is in turn directly traceable to ERM-DA474/IFCC.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212559

    Validate with FDA (Live)

    Date Cleared
    2022-12-16

    (490 days)

    Product Code
    Regulation Number
    866.5270
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CardioPhase® hsCRP is an in-vitro diagnostic reagent for the quantitative determination of C-reactive protein (CRP) in human serum, and heparin and EDTA plasma by means of particle enhanced immunonephelometry using the BN II and BN ProSpec® System. In acute phase response, increased levels of plasma proteins, including C-reactive protein, is observed. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. High sensitivity CRP (hsCRP) measurements may be used as an independent risk marker for the identification of individuals at risk for future cardiovascular disease. Measurements of hsCRP, when used in conjunction with traditional clinical laboratory evaluation of acute coronary syndromes, may be useful as an independent marker of prognosis for recurrent events, in patients with stable coronary disease or acute coronary syndromes.

    Device Description

    The CardioPhase hsCRP assay is an in vitro diagnostic reagent for the quantitative determination C-reactive protein, in human serum, and heparinized and EDTA plasma by means of particle-enhanced immunoassay determination. Polystyrene particles coated with monoclonal antibodies specific to human CRP are aggregated when mixed with samples containing CRP. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the relevant protein in the sample. The result is evaluated by comparison with a standard of known concentration.

    AI/ML Overview

    This document describes the CardioPhase® hsCRP device, a C-reactive protein immunological test system, and a Special 510(k) submission for a change in its reference standard material from ERM-DA470 to ERM-DA474/IFCC.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state numerical acceptance criteria for all performance characteristics, but rather describes the studies performed and their satisfactory outcomes which imply meeting internal acceptance limits. For instance, for linearity, the stated ranges confirm the measuring range, implying that the observed linearity falls within acceptable deviations. For matrix comparison, high correlation coefficients and slopes close to 1 with intercepts close to 0 indicate acceptable equivalence.

    Performance CharacteristicAcceptance Criteria (Implicit/Explicit)Reported Device Performance
    Detection CapabilitiesLimit of Blank (LoB): Values below the respective Limit of Quantitation (LoQ). Limit of Detection (LoD): Greater than LoB and equal to or below LoQ. Limit of Quantitation (LoQ): An imprecision goal of less than 20% CV. (Specific numerical LoB/LoD not provided as acceptance criteria, but derived from the study.)LoB: All results measured on blank samples yielded results below the respective LoQ. LoD: Calculated parametrically, greater than LoB and equal to or below LoQ. LoQ: Set to 0.094 mg/L based on the sample/instrument/reagent lot combination with the highest imprecision observed in the study (<11% CV), meeting the <20% CV goal.
    LinearityDeviation between mean measured value and predicted value of weighted linear regression compared to predefined acceptance criteria. (Specific numerical acceptance criteria not explicitly stated).CRP sensitive (CRP2): Linear range 0.151 - 10.89 mg/L (confirming measuring range of 0.16-10 mg/L). CRP (CRP1): Linear range 1.478 – 224 mg/L (confirming measuring range of 3.1-100 mg/L). This implies that the predefined acceptance criteria were met.
    Method ComparisonThe predicted bias between the predicate and candidate assay should be within acceptable limits for intended use. (Specific numerical acceptance criteria not explicitly stated but implied by the reported bias values).CRP2: Predicted bias of 6.9% at 1 mg/L and 7.5% at 3 mg/L. CRP1: Predicted bias of 8.17% at 10 mg/L. These values are presented as satisfactory for demonstrating substantial equivalence.
    Matrix ComparisonEquivalence for serum vs. EDTA plasma and serum vs. heparin plasma, indicated by slope close to 1, intercept close to 0, and high Pearson Correlation Coefficient (r). (Specific numerical acceptance criteria for slope, intercept, and r not explicitly stated but implied by the conclusion of equivalence).CRP1: Slopes (0.972-1.018), Intercepts (-0.118-0.091), Pearson r (0.995-0.998). CRP2: Slopes (0.962-1.061), Intercepts (-0.132-0.072), Pearson r (0.970-0.983). Conclusion: Confirmed equivalence for serum and EDTA/lithium heparin plasma.
    TraceabilityCalibrator N Rheumatology Standard SL must be traceable to the IFCC European Reference Material ERM-DA474/IFCC.N Rheumatology Standard SL is traceable to Siemens internal Master Calibrator which is directly traceable to ERM-DA474/IFCC.
    Substantial EquivalenceThe modified device's performance should be substantially equivalent to the predicate device in terms of intended use, design, basic scientific principle, and performance, and the change should not affect safety and efficacy.The presented performance data and the conclusion that "The modified device, CardioPhase hsCRP traceable to ERM-DA474/IFCC, is substantially equivalent to the predicate device... based on intended use design, and basic scientific principle and performance."

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Detection Capabilities (LoB, LoD, LoQ):

      • LoB: Five (5) independent analyte-free samples. Conducted with one (1) BN ProSpec System, one (1) BNII System, three (3) different reagent lots, one (1) calibrator lot, five (5) individual aliquots of each sample, and determination on three (3) days. This protocol resulted in 75 measurements per reagent lot, a total of 450 measurements (including the two workflows) on each of the two (2) analyzers, resulting in 900 measurements overall.
      • LoD/LoQ: Serum samples with concentrations ranging from approximately 0.05 mg/L to 0.26 mg/L CRP. Five replicates of six patient samples were run once per day for five days using three hsCRP Reagent lots on one BNProSpec and one BN II, totaling 1080 determinations.
      • Data Provenance: Not explicitly stated (e.g., country of origin). The samples are referred to as "analyte-free" and "patient samples" (serum), implying clinical relevance and likely of human origin. The study appears to be prospective as it was specifically designed for this evaluation.
    • Linearity:

      • Sample Size: A high CRP serum pool was mixed with a low serum pool to generate 13 concentrations. Each level was tested in four-fold determination.
      • Data Provenance: Not explicitly stated (e.g., country of origin). The samples are described as "serum pool," implying human origin. The study appears to be prospective.
    • Method Comparison:

      • Sample Size: Native samples in the range of 0.27 and 11.90 mg/L for CRP2 and native samples in the range of 3.87 and 61.40 mg/L for CRP1. The exact number of samples is not explicitly given, but it implies a sufficient number to perform linear regression and bias calculations across the specified ranges.
      • Data Provenance: Not explicitly stated (e.g., country of origin). Native samples imply human origin. The study appears to be prospective for this comparison.
    • Matrix Comparison:

      • Sample Size: For each sample type (EDTA plasma and lithium heparin plasma compared to serum), a total of 60 native samples spanning the measuring range were evaluated.
      • Data Provenance: Not explicitly stated (e.g., country of origin). Native samples imply human origin. The study appears to be prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For this in-vitro diagnostic device (quantitative determination of C-reactive protein), "ground truth" is typically established by reference methods or gravimetrically prepared standards with known concentrations, or by comparison to a reference standard material.

    • Traceability and Standardization: The ground truth for this device is primarily established by its traceability to the IFCC European Reference Material ERM-DA474/IFCC, which is "certified for C-reactive protein measurements." This reference material serves as the "expert" or definitive standard.
    • No human expert panel for ground truth: Unlike image-based diagnostic devices, this type of immunoassay does not involve human experts establishing a "ground truth" based on interpretations (e.g., radiologists labeling images). The "ground truth" is analytical and defined by international reference standards.

    4. Adjudication Method for the Test Set

    Not applicable in the conventional sense. For an in-vitro diagnostic assay that measures a quantitative analyte, the "ground truth" is the certified value assigned to reference materials or the result from a highly accurate reference method. There is no subjective interpretation by multiple human adjudicators in the way there would be for an imaging diagnosis.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This is an in-vitro diagnostic (IVD) device for quantitative measurement of a biomarker (CRP). MRMC studies are typically performed for devices that involve human interpretation, such as imaging AI applications, to assess how AI assistance impacts human reader performance. This device does not have a "human reader" component in that context.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, entirely. The entire performance evaluation (detection capabilities, linearity, method comparison, matrix comparison) demonstrates the standalone performance of the CardioPhase® hsCRP assay system (reagent and instrument) in measuring CRP levels. The device itself, once calibrated, provides a quantitative result without human-in-the-loop performance influencing the measurement. The human role is in operating the instrument and interpreting the numerical result in a clinical context, but not in directly influencing the measurement itself.

    7. Type of Ground Truth Used

    The ground truth used is primarily certified reference materials (ERM-DA474/IFCC) and comparative analysis against a previously cleared predicate device (which itself was traceable to ERM-DA470) which acts as a well-established reference. The "ground truth" for the calibrator N Rheumatology Standard SL is its traceability to the Siemens internal Master Calibrator, which in turn is directly traceable to ERM-DA474/IFCC.

    8. Sample Size for the Training Set

    Not applicable. This document describes the performance evaluation of a change to an existing in-vitro diagnostic reagent and its calibrator. This is not a machine learning or AI-based device that typically requires a "training set" in the context of model development. The characterization studies described are for analytical performance verification of the assay system, rather than training a computational algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of IVD device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1