Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K213834

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-03-18

    (99 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cardio Flow Peripheral Guide Wire is intended for temporary placement in peripheral vasculature the placement and exchange of diagnostic and therapeutic devices during percutaneous intravascular procedures. This guidewire device is intended for peripheral vascular use only.

    Device Description

    The Cardio Flow Peripheral Guide Wire (Guidewire), Model GW1001 is a 304V stainless steel mandrel guidewire with a fixed distal spring coil. The distal coil is radiopaque to fluoroscopy. The coil is medium-stiffness in load-force and the distal end is coated in hydrophobic silicone to facilitate tracking. The Guidewire has a maximum outer diameter of 0.36 mm (0.014-inch). The GW1001 Guidewire has overall nominal length of 325 cm.

    AI/ML Overview

    The Cardio Flow Peripheral Guide Wire (Model GW1001) underwent several non-clinical performance tests to demonstrate substantial equivalence to its predicate device, the Asahi Astato XS 20 Guide Wire (K103057).

    Here's the breakdown of the acceptance criteria and the study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (based on implied compliance with standards/guidance)Reported Device Performance
    Corrosion ResistanceCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
    Dimensional VerificationCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
    Kink ResistanceCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
    LubricityCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
    Particulate EvaluationCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
    RadiopacityCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
    Silicone Coating IntegrityCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
    Simulated Use (Compatibility)Compliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
    Tensile StrengthCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
    Tip FlexibilityCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
    Tip PullCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
    Torque StrengthCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
    Torqueability (Torque Response)Compliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
    Flex ResistanceResistant to damage by flexing in accordance with ISO 11070:2014 Annex G criterion, between two cylindrical mandrels with diameters equal to 20 times the guidewire diameter.Pass
    Fracture ResistanceShall not fracture or fail when tested in accordance with ISO 11070:2014 Annex F criterion, test article wrapped around a mandrel with a diameter 10 times larger than the guidewire outside diameter.Pass
    Tip LoadGuidewire tip load buckling force inducing buckling deformation with a 20 mm gauge length from the distal tip is documented. (Implied acceptance is demonstration of acceptable force for intended use)Pass
    SterilitySterilization validated according to ANSI/AAMI/ISO 11135:2014 with a minimum sterility assurance level (SAL) of 10-6.Pass
    Sterile barrier integrityPost terminal sterilization simulated distribution, seal peel (per ASTM F88), and bubble emission testing (per ASTM F2096) results demonstrate integrity.Pass
    BiocompatibilityCompliance with ISO 10993-4, -5, -10, -11 (for Cytotoxicity, Complement Activation, Intracutaneous Reactivity, Sensitization, Material Mediated-Pyrogenicity, Acute Systemic Toxicity, and Standard in Vivo Thrombogenicity in Canine), ASTM F756 (for Hemolysis), and ASTM F2888 (for Heparinized Platelet and Leukocyte Count).Pass
    Shelf life (2 years)Confirmation of device functional performance and sterile barrier pouch integrity (seal peel per ASTM F88 and bubble leak per ASTM F2096) with accelerated aging and simulated distribution per ASTM 4169-16.Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes bench evaluations and non-clinical performance tests. This implies that physical samples of the Cardio Flow Peripheral Guide Wire were tested in a laboratory setting.

    • Sample Size: Not explicitly stated for each individual test. However, it is standard practice for such tests to use a statistically significant number of samples to ensure robust results.
    • Data Provenance: The tests were conducted internally by or on behalf of Cardio Flow, Inc., likely at their facilities or a contracted testing lab, following established international and FDA-recognized standards. There is no mention of country of origin for test data, but the company is based in Mahtomedi, MN, USA. The data is prospective for the device under review.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. This was a physical device performance study, not a study requiring expert clinical assessment for ground truth. The "ground truth" for these tests is defined by the objective performance criteria specified in the referenced standards (e.g., ISO, ASTM, FDA guidance).

    4. Adjudication Method for the Test Set

    Not applicable, as this was a physical device performance study, not a clinical study involving human judgment that would require adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This device is a medical guidewire, not a diagnostic AI system that evaluates medical images or data. Therefore, an MRMC study comparing human readers with and without AI assistance is irrelevant.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This device is a physical medical guidewire, not an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for this type of device performance testing is established by objective engineering and performance specifications defined in recognized national and international standards (e.g., ISO, ASTM) and FDA guidance documents. These standards dictate acceptable ranges, thresholds, or qualitative outcomes (e.g., "Pass/Fail") for various physical and material properties.

    8. Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1