Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K252446

    Validate with FDA (Live)

    Device Name
    BruxZir® NOW
    Date Cleared
    2025-09-03

    (30 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is used for dental restorations using different CAD/CAM or manual milling machines. All blocks are processed through dental laboratories or by dental professionals.

    Contraindication: BruxZir® NOW Bridge Block design is limited to a maximum of four units with a maximum of two pontics between pillars.

    Device Description

    BruxZir® NOW is a pre-shaded, fully sintered zirconia CAD/CAM block used for fabricating single-unit or multiple-unit dental restorations. BruxZir® NOW is used for fabricating single-unit dental restorations, while BruxZir® NOW Bridge Block is used for fabricating multiple-unit dental restorations. BruxZir® NOW is available in the following shades: Bleach White, Bleach 1, Bleach 3, and 16 VITA Classical shades (A1, A2, A3, A3.5, A4, B1, B2, B3, B4, C1, C2, C3, C4, D2, D3, D4). BruxZir® NOW Bridge Block is available in the following 6 VITA Classical shades: A1, A2, A3, B1, C2, D2. The method of fabricating the restoration is with CAD/CAM milling systems.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K231940

    Validate with FDA (Live)

    Date Cleared
    2024-05-15

    (320 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BruxZir® NOW SRC is a pre-manufactured prosthetic component directly connected to an endosseous implant and it is intended for use in prosthetic rehabilitation. BruxZir® NOW SRC consists of two major parts. Specifically, the precemented titanium abutment and zirconia superstructure components make up a two-piece abutment. All digitally designed CAD/CAM customizations for BruxZZir® NOW SRC are only intended to be designed and manufactured according to digital dentistry workflow, which incorporates of the digital dentistry workflow including scan files from Intra-Oral Scanners, CAD software, milling machine and associated tooling and accessories.

    Device Description

    BruxZir® NOW SRC is a pre-manufactured two-piece hybrid abutment crown composed of a fully sintered zirconia superstructure and a pre-installed Ti-base (titanium base) that is used with the glidewell.io™ In-Office Solution. The zirconia and Ti-base components make up a two-piece abutment, which is used in conjunction with endosseous dental implants to restore chewing function. The CAD/CAM software is intended to design the final abutment shape including anatomic features and facilitate in-office milling.

    The Ti-base of BruxZir® NOW SRC is pre-cemented to a post-sintered zirconia superstructure, which is to be designed and fabricated to support screw-retained prosthetic restorations. After machining in the fastmill.io"M In-Office Mill, this two-piece abutment is mounted onto the implant and fixed with a screw. BruxZir® NOW SRC can be milled in-office by a dental professional and dental laboratories.

    The digital workflow requires the use of the following equipment:

    • Scanner: Intra-oral scanner
    • Design Software: CloudPoint FastDesign Chairside (K231529)
    • Restorative Material: Fully sintered zirconia; BruxZir® NOW(K220816)
    • Milling Unit: fastmill.io™ In-Office Mill

    The BruxZir® NOW SRC is designed and fabricated to be compatible with the Hahn™ Tapered Implant System (K143353) and the Inclusive® Tapered Implant System (K121406, K153099). BruxZir® NOW SRC is provided non-sterile and intended for single use and prescription use.

    AI/ML Overview

    The provided text describes the BruxZir® NOW SRC, a dental implant abutment, and its substantial equivalence to a predicate device. It primarily focuses on mechanical, sterilization, biocompatibility, shelf-life, and packaging data to support its claims. There is no information provided in the document about acceptance criteria, device performance, AI/algorithm studies, ground truth establishment, or expert involvement. The document primarily reports on non-clinical testing.

    Here's a breakdown of what CANNOT be answered based on the provided text:

    • Acceptance criteria and reported device performance (in the context of clinical or AI performance): The document doesn't define specific acceptance criteria for performance metrics like sensitivity, specificity, or accuracy, as it's not evaluating an AI/diagnostic device. It broadly states the device has "sufficient strength" and provides a "maximum angulation correction of 20 degrees" based on mechanical testing.
    • Sample size for test set and data provenance (for AI): Not applicable as there is no AI or clinical test set described.
    • Number of experts used to establish ground truth and qualifications: Not applicable.
    • Adjudication method: Not applicable.
    • Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable.
    • Standalone (algorithm-only) performance: Not applicable.
    • Type of ground truth used (for AI/clinical performance): Not applicable.
    • Sample size for the training set (for AI): Not applicable.
    • How the ground truth for the training set was established (for AI): Not applicable.

    However, I can extract information about the non-clinical studies performed and their general conclusions:

    Non-Clinical Studies and Findings:

    Study CategoryAcceptance Criteria / Standard FollowedReported Device Performance / Conclusion
    Mechanical PropertiesFDA Guidance Document, "Guidance for Industry and FDA Staff Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments" and ISO 14801:2016 standard. Testing performed with worst-case scenario.The fatigue limit data demonstrated that the subject device has sufficient strength and provides a maximum angulation correction of 20 degrees for its intended use. Used to address substantial equivalence.
    Sterilization ValidationISO 17665-1:2006 (for steam sterilization). Compared to reference device (Inclusive® Abutments, K160979) which had previous validation.The steam sterilization parameters previously validated for the reference device (K160979) can sufficiently sterilize the subject device for use in a clinical environment, indicating it meets the criteria.
    Biocompatibility EvaluationFDA Guidance Document, "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process'."Concluded that there is no biocompatibility concern.
    Shelf LifeBased on material stability (titanium alloy conforming to ASTM F136 and Zirconium Oxide).Shelf-life is not applicable due to the inherent stability of the materials, and no new safety or effectiveness concerns were raised due to aging.
    Packaging ValidationEvaluated suitability to withstand distribution environment. Compared to reference device (BruxZir® NOW, K220816).The results of previous packaging validation for the reference device supported the packaging for the subject device, indicating it can withstand distribution without damage.
    Use in MR EnvironmentFDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" and published literature.Non-clinical MR review performed using scientific rationale; indicates suitability or safety within an MR environment. (Specific parameters not detailed, but the review was conducted).

    Sample Sizes and Data Provenance for Non-Clinical Tests:

    • Mechanical Testing: The document refers to "worst-case scenario" testing but does not specify the number of samples or the origin of the data. It's implicitly laboratory-based data for device components.
    • Sterilization Validation: Refers to "testing sample used in the previous steam sterilization validation testing for the reference device (K160979)" which "represented the worst-case scenario." No specific number is given for the subject device itself.
    • Biocompatibility Evaluation: This is a biological evaluation following ISO 10993-1. Sample size not specified.
    • Shelf Life: Based on material science principles; no direct sample size for testing is implied.
    • Packaging Validation: "Previous packaging validation" for a reference device was used. No specific sample size for the subject device's packaging.
    • MR Environment: Based on scientific rationale and published literature; not a direct experimental study with a sample size for this submission.

    Further Notes:

    • The document states "No clinical data was included in this submission." This explicitly confirms that the information requested about clinical performance, expert ground truth, and reader studies is not present in this document.
    • The device is a "pre-manufactured prosthetic component" for dental implants, not an AI or diagnostic software. Therefore, the questions related to AI-specific performance metrics, training/test sets, and ground truth are not applicable to the scope of this FDA submission.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1