K Number
K121406

Validate with FDA (Live)

Date Cleared
2013-02-22

(288 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Inclusive® Tapered Implants are indicated for use in maxillary and mandibular partially or fully edentulous cases, to support single, multiple-unit, and overdenture restorations. The implants are to be used for immediate loading only in the presence of primary stability and appropriate occlusal loading.

Device Description

The Inclusive® Tapered Implants are manufactured from biocompatible Titanium alloy. The implant is designed with an internal hex connection and a tapered body to replace one or more missing teeth. The surface is blasted with Hydroxyl Apatite and acid etched to facilitate osseointegration. The Implants are available in the following diameters: 3.7mm. 4.7mm. and 5.2mm, and each diameter is available in the following lengths: 8mm, 10mm, 11.5mm, 13mm, and 16mm. The dental implants are provided sterile (gamma).

AI/ML Overview

The provided document describes the PRISMATIK DENTALCRAFT, INC. Inclusive® Tapered Implant System, a medical device. The document primarily focuses on demonstrating substantial equivalence to predicate devices rather than directly presenting the results of a standalone study against specific acceptance criteria in the format requested. However, it does reference performance data and biocompatibility tests conducted to meet acceptance criteria.

Based on the information provided, here's a breakdown of the requested elements:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Test / StandardReported Device Performance
Mechanical Performance (Implant to Abutment Compatibility)FDA's Guidance Document "Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" and ISO 14801: 2007 - Dentistry - Implants - Dynamic fatigue test for endosseous dental implants. This included fatigue testing on 30° Multi-Unit Abutments and evaluated the worst-case scenario."All testing conducted met the acceptance criteria and evaluated the worst-case scenario."
"performance testing data indicated the compatibility, and the safety and the effectiveness of the proposed device which meets the mechanical properties."
Biocompatibility1. Irritation (applying ISO 10993-10: 2010 Tests for Irritation and Skin Sensitization)"reports for the results indicated satisfactory and meeting the acceptance criteria"
2. Media (MEM) Elution test to determine the Cytotoxicity (applying ISO 10993-5: 2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity)"reports for the results indicated satisfactory and meeting the acceptance criteria"
3. Sensitization Tests for Irritation and Skin Sensitization (applying ISO 10993-10: 2010 Tests for Irritation and Skin Sensitization)"reports for the results indicated satisfactory and meeting the acceptance criteria"
SterilizationGamma sterilization process conforming to ISO 11137-1: Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices, and ANSI/AAMI/ISO 11137-2: Sterilized of Health Care Products - Radiation - Establishing the sterilized dose.The implant packages are "sterilized by a gamma sterilization process that conforms to ISO 11137-1... and ANSI/AAMI/ISO 11137-2."

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the sample sizes used for the mechanical performance testing (fatigue tests) or the biocompatibility tests (irritation, cytotoxicity, sensitization). It only states that worst-case scenarios were evaluated. There is no information regarding data provenance (e.g., country of origin, retrospective/prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

Not applicable. The tests described are laboratory-based mechanical and biological evaluations, not studies requiring human interpretation or expert ground truth for a test set in the context of diagnostic accuracy.

4. Adjudication Method for the Test Set:

Not applicable for the types of tests described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC comparative effectiveness study was not done. The document describes a 510(k) submission for substantial equivalence based on mechanical and biological performance, not a study involving human readers or AI assistance.

6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study:

No. This device is a physical dental implant, not an algorithm or AI system. The performance data relates to the physical properties and biocompatibility of the implant itself.

7. Type of Ground Truth Used:

For the mechanical performance, the "ground truth" or standard was established by the specified FDA guidance and ISO 14801: 2007 standard for dynamic fatigue testing. For biocompatibility, the "ground truth" was established by the relevant ISO 10993 standards. These are internationally recognized standards and guidance documents for medical device testing.

8. Sample Size for the Training Set:

Not applicable. This device is a physical implant and does not involve AI or machine learning algorithms that would require a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

PRISMATIK
DENTALCRAFT, INC.

FEB 2 2 2013

005-510 (k) Summary-807.92(c)

This 510 (K) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

121406

SUBMITTER INFORMATION A.

Company Name:Prismatik Dentalcraft, Inc.
Company Address:2212 Dupont Dr., Suite IJKIrvine, CA 92612
Company Phone:949-225-1269
Company FAX:949-553-0924
Primary Contact:Armin Zehtabchi,Senior RA/QA., /Project Manager(949) 225-1234
Secondary Contact:Marilyn PourazarSenior Director, RA/QA(949) 225-1269
Date Summary Prepared:2/20/13
DEVICE IDENTIFICATION
Trade/Proprietary Name:Inclusive® Tapered Implant System
21 CFR Reference:21 CFR 872.3640
21 CFR Common Name:Implant, endosseous, root-form
Classification:Class II

B.

Product Code:

Panel:

Dental

DZE

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for PRISMATIK DENTALCRAFT, INC. The logo consists of a black triangle on the left and the company name on the right. The triangle is divided into three smaller triangles, with the top triangle being the darkest and the bottom triangle being the lightest. The company name is written in a bold, sans-serif font.

C. IDENTIFICATION OF PREDICATE DEVICE

Trade/Proprietary Name:

Inclusive® Mini Implant (K100932), Sulzer's Screw-Vent Dental Implant System (K011028) and KIS-Krauser Implant System (K953235-Transffered to Glidewell Laboratories/Prismatik Dentalcraft, Inc., 2/2011)

D. DEVICE DESCRIPTION

The Inclusive® Tapered Implants are manufactured from biocompatible Titanium alloy. The implant is designed with an internal hex connection and a tapered body to replace one or more missing teeth. The surface is blasted with Hydroxyl Apatite and acid etched to facilitate osseointegration. The Implants are available in the following diameters: 3.7mm. 4.7mm. and 5.2mm, and each diameter is available in the following lengths: 8mm, 10mm, 11.5mm, 13mm, and 16mm. The dental implants are provided sterile (gamma).

E. INDICATIONS FOR USE

Inclusive® Tapered Implants are indicated for use in maxillary and mandibular partially or fully edentulous cases. to support single, multiple-unit, and overdenture restorations. The implants are to be used for immediate loading only in the presence of primary stability and appropriate occlusal loading.

F. SUBSTANTIAL EQUIVALENCE

The Inclusive® Tapered Implant System is substantially equivalent to Prismatik's Inclusive® Mini Implant (K100932), Sulzer's Screw-Vent Dental Implant System (K011028) and the KIS-Krauser Implant System (K953235-Transferred to Glidewell Laboratories/Prismatik Dentalcraft, Inc., 2/2011). These implants are substantially equivalent in intended use, indications for use, material, design and performance.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for PRISMATIK DENTALCRAFT, INC. The logo features a black triangle on the left side of the image. The text "PRISMATIK DENTALCRAFT, INC." is located to the right of the triangle. The text is in black and is in a sans-serif font.

Comparison of Predicate Devices

Elements ofComparisonPrismatik'sInclusive® TaperedImplant SystemSulzer's Screw-VentDental ImplantSystem(K011028)KIS-KrauserImplant System(K953235-Transferred toGlidewellLaboratories/PrismatikDentalcraft, Inc.,2/2011Prismatik's InclusiveMini Implant(K100932)
MaterialTitanium AlloyTitanium AlloyTitanium AlloyTitanium Alloy
IndicationsInclusive® TaperedImplants areindicated for use inmaxillary andmandibularpartially or fullyedentulous cases,to support single,multiple-unit, andoverdenturerestorations. Theimplants are to beused for immediateloading only in thepresence ofprimary stabilityand appropriateocclusal loading.Sulzer Dental'simplant systems aredesigned for use inedentulousmandibles ormaxillae forattachment ofcomplete dentureprostheses, or as aterminal orintermediaryabutment for fixedor removablebridgework, or as afreestanding singletooth replacement.The use of the4.7mm and 6.0mmimplant isrecommended whenthe quantity anddensity of bonewould dictate theuse of an implantlarger than 4.0mm.The KIS-KrauserImplant Systemimplants areindicated for singletooth replacements,as intermediateabutments onbridgework, as distalabutments for areasto be restored withfixed bridgework, tosupport overdenturesin the totallyedentulousmandible, and asabutmentssupporting a fullarch fixedprosthesis.Inclusive® MiniImplants are self-tapping threadedtitanium screwsindicated for long-term application.Inclusive® MiniImplants may alsobe used forprovisionalapplications. Thesedevices will allowimmediate loadingand long-termstabilization ofdentures andprovisionalstabilization ofdentures whilestandard implantsheal. To be used forimmediate loadingonly in the presenceof primary stabilityand appropriateocclusal.
DesignThreaded root-formimplantThreaded root-formimplantThreaded root-formimplantThreaded root-formimplant
Implant BodyGeometryScrew typeScrew typeScrew typeScrew type
Diameters (mm)3.7, 4.7 and 5.23.7, 4.7 and 6.04, 5 and 62.2, 2.5, 3.0
Lengths (mm)8, 10, 11.5, 13, and168, 10, 13 and 168, 10, 13, 15, 1810, 13, 15
DriverConnectionHexHexHexSquare
Prosthetic HeadInternalInternalExternalN/A
SterilityPackaged SterilePackaged SterilePackaged SterilePackaged Sterile

.

:

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a black triangle on the left and the company name on the right. The company name is written in a bold, sans-serif font.

్ర PERFORMANCE DATA

The following FDA's Guidance Document "Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments was used for the purpose of Implant to Abutment Compatibility. The Tapered Implant was used for the fatigue testing on the 30° Multi-Unit Abutments and by following the ISO 14801: 2007-Dentistry - Implants -Dynamic fatigue test for endosseous dental implants. All testing conducted met the acceptance criteria and evaluated the worst case scenario. In addition, performance testing data indicated the compatibility, and the safety and the effectiveness of the proposed device which meets the mechanical properties. In addition, to assure the biocompatibility, the following tests were performed, and the reports for the results indicated satisfactory and meeting the acceptance criteria:

    1. Irritation (applying ISO 10993-10: 2010 Tests for Irritation and Skin Sensitization).
    1. Media (MEM) Elution test to determine the Cytotoxicity (applying ISO 10993-5: 2009 Biological evaluation of medical devices -Part 5: Tests for in vitro cvtotoxicity
    1. Sensitization Tests for Irritation and Skin Sensitization (applying ISO 10993-10: 2010 Tests for Irritation and Skin Sensitization).

The implant packages are sterilized by a gamma sterilization process that conforms to ISO 11137-1: Sterilization of health care products -Radiation -Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices, and ANSI/AAMI/ISO 11137-2: Sterilized of Health Care Products-Radiation-Establishing the sterilized dose.

COMPARISON OF TECHNOLOGICAL DIFFERENCES H.

There are no known technological differences between the Inclusive® Tapered Implant and those of the predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the symbol. The logo is black and white and appears to be a simple, vectorized design.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 22, 2013

Mr. Armin Zehtabchi Senior Regulatory Affairs / Quality Assurance, Project Manager Prismatik Dentalcraft, Incorporated 2212 Dupont Drive, Suite IJK IRVINE CA 92612

Re: K121406

Trade/Device Name: Inclusive Tapered Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: February 11, 2013 Received: February 12, 2013

Dear Mr. Zehtabchi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Kwame O. Ulmer

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices

Office of Device Eyaluation

Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(K) Number: K121406

Device Name: Inclusive Tapered Implant System

Indications for Use:

Inclusive® Tapered Implants are indicated for use in maxillary and mandibular partially or fully edentulous cases, to support single, multiple-unit, and overdenture restorations. The implants are to be used for immediate loading only in the presence of primary stability and appropriate occlusal loading.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-the-Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.

Concurrence of CDRH, Office of Device Evaluation (ODE)

S. Runner 02.22 01 29:36 -05:00,

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

次121406 510(k) Number:

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.