510(k) Data Aggregation

    K Number
    K172456

    Validate with FDA (Live)

    Date Cleared
    2018-01-04

    (143 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMR Shoulder System is intended for partial or total, primary or revision shoulder joint replacement. The SMR Anatomic Shoulder System is indicated for partial or total, primary or revision shoulder joint replacement in patients suffering from disability due to:

    • non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
    • inflammatory degenerative joint disease such as rheumatoid arthritis;
    • treatment of acute fractures of the humeral head that cannot be treated with other fracture fixation methods;
    • revision of a failed primary implant;
    • cuff tear arthropathy (CTA Heads only);
    • glenoid arthrosis without excessive glenoid bone loss: A1, A2 and B1 according to Walch classification (SMR Hybrid Glenoid only).

    The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.

    The Hybrid Reverse Baseplate must not be used in cases of excessive glenoid bone loss and/or when bone graft is needed.

    The Modular SMR Shoulder System allows the assembly of components in various humeral and glenoid constructs. The constructs are intended for cemented and uncemented use as specified in the following table.

    In the Anatomic shoulder the humeral consists of the humeral stem, the humeral body, the adaptor taper and the humeral head. In the Reverse shoulder the humeral consists of the humeral stem, the reverse humeral body and the reverse liner. On the humeral side the fixation of the humeral stem determines if the construct is cemented or uncemented.

    The Anatomic glenoid construct consists of an all polyethylene glenoid, a polyethylene glenoid with metal peg or a metal back assembled with a liner; the Reverse glenoid consists of a metal back/connector/glenosphere construct or of a peg/baseplate/glenosphere construct.

    On the glenoid side, the fixation of the all polyethylene glenoid, the polyethylene glenoid with metal peg or the metal back determines if the construct is cemented or uncemented.

    The Delta TT Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

    • Non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and hip dysplasia;
    • Rheumatoid arthritis;
    • Post-traumatic arthritis,
    • Correction of functional deformity;
    • Fractures, dislocation of the hip and unsuccessful cup arthroplasty.
    • Revisions in cases of good remaining bone stock.
    Device Description

    Bone screws are used to provide further stability to the Delta TT acetabular cups and to the metal back glenoid of the SMR Shoulder System in reverse shoulder configuration.

    The Delta TT acetabular cups (K112898, K141395) are acetabular components intended for uncemented fixation; screw holes present in the Delta TT cups allow optional additional fixation of the cup to the bone using 6.5mm diameter bone screws.

    The SMR Shoulder System metal back glenoids (K110598, K113254, K133349, K161120, K163397) are components intended to be implanted into a hole drilled in the glenoid cavity. The metal back glenoid can be used in both anatomic and reverse shoulder configuration; when used in reverse shoulder configuration, the metal back glenoid are intended for uncemented use and are fixed to the glenoid bone through the use of 6.5mm bone screws.

    The current submission is for additional lengths of 6.5mm diameter bone screws:

    • the proposed range of bone screws used in combination with the SMR Shoulder System is 20mm to 55mm in length;
    • the proposed range of bone screws used in combination with the Delta TT Acetabular System is 15mm to 90mm in length.

    Design, material, manufacturing process and intended use of the new 6.5mm diameter bone screws are the same as those for the 6.5mm bone screws cleared in the previous 510(k)s.

    AI/ML Overview

    The provided document is a 510(k) premarket notification decision letter and accompanying information for Bone Screws 6.5mm by LimaCorporate S.p.A.

    This document describes a submission for a medical device (bone screws), not an AI/ML-enabled diagnostic or therapeutic device that would typically undergo the kind of acceptance criteria, standalone performance studies, and human-in-the-loop studies involving expert readers, ground truth establishment, and large image datasets as outlined in your request.

    Therefore, the information required to answer your prompt, such as acceptance criteria based on performance metrics (e.g., accuracy, sensitivity, specificity), sample sizes for test sets of images, data provenance, number and qualifications of experts, adjudication methods, MRMC studies, standalone performance, and training set details, is not present in this document.

    The document explicitly states under "Non-Clinical Testing" and "Clinical Testing":

    • Non-Clinical Testing: "The additional lengths of 6.5mm bone screws use the same fundamental scientific technology and device design concepts, are made from the same materials utilizing the same manufacturing processes, utilize the same packaging and sterilization processes, and have the same indications for use as the predicate devices; the new lengths of 6.5mm diameter bone screws therefore do not require any testing to determine substantial equivalency."
    • Clinical Testing: "Clinical testing was not necessary to demonstrate substantial equivalence of the new sizes of bone screws to the predicate devices."

    This indicates that the FDA's substantial equivalence determination for these bone screws was based on their similarity to existing predicate devices in terms of design, materials, manufacturing, intended use, and indications, rather than requiring performance studies against specific acceptance criteria for diagnostic accuracy or clinical outcomes that would involve the elements you've listed.

    In essence, the "study" proving the device meets acceptance criteria was a comparison to legally marketed predicate devices, demonstrating that the new bone screw lengths do not raise different questions of safety and effectiveness. The acceptance criteria for this type of device would primarily revolve around material compatibility, mechanical properties (which were not explicitly detailed as new testing data, given the stated substantial equivalence), manufacturing quality, and intended use as compared to the predicates, rather than the performance metrics of an AI/ML system.