510(k) Data Aggregation

    K Number
    K131301

    Validate with FDA (Live)

    Date Cleared
    2013-08-02

    (87 days)

    Product Code
    Regulation Number
    862.2750
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD FACS™ Sample Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCanto™ II and BD FACSCalibur™ flow cytometry systems.

    Pipetting blood, reagents, and lysing solution using the following previously cleared assays for flow cytometric analysis on BD FACSCanto™ II flow cytometry systems:

    • BD Multitest 6-Color TBNK Reagent with or without BD Trucount Tubes .
    • BD Multitest IMK Kit with or without BD Trucount Tubes .
    • BD Multitest CD3 FITC/CD16+CD56 PE/CD45 PerCP/CD19 APC with or . without BD Trucount Tubes
    • BD Multitest CD3 FITC/CD8 PE/CD45 PerCP/CD4 APC with or without BD . Trucount Tubes

    Pipetting blood, reagents, and lysing solution using the following previously cleared assays for flow cytometric analysis on BD FACSCalibur™ flow cytometry systems:

    • BD Multitest IMK Kit with or without BD Trucount Tubes .
    • BD Multitest CD3 FITC/CD16+CD56 PE/CD45 PerCP/CD19 APC with or . without BD Trucount Tubes
    • BD Multitest CD3 FITC/CD8 PE/CD45 PerCP/CD4 APC with or without BD . Trucount Tubes
    • BD Tritest CD3/CD16+56/CD45 with or without BD Trucount Tubes .
    • BD Tritest CD3/CD19/CD45 with or without BD Trucount Tubes .
    • BD Tritest CD3/CD4/CD45 with or without BD Trucount Tubes .
    • BD Tritest CD3/CD8/CD45 with or without Trucount Tubes .
    • BD Tritest CD4/CD8/CD3 with BD Trucount Tubes .

    For in vitro diagnostic use.

    Device Description

    The BD FACS™ Sample Prep Assistant III (SPA III) is a microprocessorcontrolled pipetting and diluting system which automatically prepares whole blood samples using the lyse / no-wash sample preparation method for flow cytometry. Used as an accessory to the BD FACSCalibur flow cytometer, the SPA III combines fluidic, optic, robotic, and electronic components to automatically prepare samples for acquisition and analysis.

    The SPA III pierces the sample tube cap to withdraw sample, aliquots blood and reagent into daughter tubes, and mixes the sample according to preprogrammed or custom protocols. The device also adds lysing solution and automates cleaning procedures. The unit consists of an enclosure, one robotic pipetting module moving in the X/Y/Z axes, a power supply, a central controller unit, fluid pumps, and a barcode reader.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the BD FACS™ Sample Prep Assistant III, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Study/CriterionAcceptance Criteria (Implied)Reported Device Performance
    AccuracyDemonstrate equivalent performance for sample preparation.The SPA III with the BD FACSCalibur system demonstrated equivalent performance in its ability to prepare human whole blood for flow cytometric analysis.
    PrecisionSystem precision within specification.The SPA III demonstrated system precision within specification.
    CarryoverSystem carryover within specification.The SPA III demonstrated system carryover within specification.

    Important Note: The document does not explicitly state numerical acceptance criteria for accuracy, precision, or carryover. It refers to established guidelines (CLSI document EP9-A2 for Accuracy, CLSI document EP5-A2 for Precision, and Class II Special Controls Guidance Document for Carryover) which likely define these specifications. The reported performance is qualitative, stating that the device "demonstrated equivalent performance" or was "within specification."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: Not explicitly stated. The "Accuracy" study mentions "Patient Samples," but the number of samples used is not provided.
    • Data Provenance: Not explicitly stated. The document describes the device preparing "human whole blood," implying human samples were used, but the country of origin is not mentioned. The studies appear to be prospective, laboratory-based evaluations.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The studies listed (Accuracy, Precision, Carryover) are performance assessments of the automated system itself, not studies relying on interpretation by experts to establish a "ground truth" for diagnostic purposes. The ground truth for these types of studies would likely be objective measurements or comparisons against established manual methods or reference instruments.

    4. Adjudication Method for the Test Set:

    This information is not provided. As noted above, the studies are performance assessments, not expert-opinion-based diagnostic studies.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    An MRMC comparative effectiveness study was not conducted. This type of study typically involves human readers interpreting diagnostic images or data, and the BD FACS™ Sample Prep Assistant III is an automated sample preparation device, not an interpretive diagnostic tool.

    6. Standalone (Algorithm Only) Performance:

    Yes, a standalone performance assessment was done. The studies listed (Accuracy, Precision, Carryover) are evaluations of the device's performance in preparing samples. The "results" section directly refers to the "SPA III" or "SPA III with the BD FACSCalibur system" demonstrating equivalent performance or being within specification, indicating that the device's automated functions were assessed directly.

    7. Type of Ground Truth Used:

    The type of ground truth used would be based on objective measurements and comparisons against established reference methods.

    • For Accuracy, the "Method Comparison and Bias Estimation" likely compared results from samples prepared by the SPA III to results from samples prepared by a predicate device or a manual, gold-standard method.
    • For Precision, the ground truth would be statistical measures of reproducibility and repeatability based on replicate measurements.
    • For Carryover, the ground truth would be the absence or minimal presence of analytes from a high-concentration sample in a subsequent low-concentration sample, measured objectively.

    8. Sample Size for the Training Set:

    This information is not applicable or not provided. The BD FACS™ Sample Prep Assistant III is an automated pipetting and diluting system. It is not an AI/machine learning algorithm that requires a "training set" in the conventional sense. Its functionality is based on pre-programmed protocols and mechanical precision, not learned patterns from data.

    9. How the Ground Truth for the Training Set Was Established:

    This question is not applicable as the device does not utilize a training set in the context of AI/machine learning. Its "ground truth" for operational accuracy and precision would be established through engineering specifications, calibration, and validation against known standards during its development and manufacturing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102064

    Validate with FDA (Live)

    Date Cleared
    2011-03-07

    (227 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD FACSTM Sample Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCanto II flow cytometry systems.

    Pipetting blood, reagents and lysing solution using the following previously cleared assays for flow cytometric analysis on BD FACSCanto II flow cytometry systems:

    • . BD Multitest 6-Color TBNK Reagent with or without BD Trucount Tubes
    • BD Multitest IMK Kit with or without BD Trucount Tubes .
    • BD Multitest CD3 FITC/CD16+CD56 PE/CD45 PerCP/CD19 APC with or . without BD Trucount Tubes
    • BD Multitest CD3 FITC/CD8 PE/CD45 PerCP/CD4 APC with or without BD . Trucount Tubes

    For in vitro diagnostic use.

    Device Description

    The BD FACS Sample Prep Assistant III is a microprocessor-controlled pipetting and diluting system which automatically prepares whole blood samples using the lyse / nowash sample preparation method for flow cytometry. Used as an accessory to the BD FACSCanto flow cytometer, the SPA III combines fluidic, optic, robotic, and electronic components to automatically prepare samples for acquisition and analysis.

    The SPA III pierces the sample tube cap to withdraw sample, aliquots blood and reagent into daughter tubes, and mixes the sample according to preprogrammed or custom protocols. The device also adds lysing solution and automates cleaning procedures. The unit consists of an enclosure, one robotic pipetting module moving in the X/Y/Z axes, a power supply, a central controller unit, fluid pumps, and a barcode reader.

    AI/ML Overview

    The provided text describes the BD FACS Sample Prep Assistant III (SPA III) device and its performance studies. However, it does not contain detailed acceptance criteria, specific reported device performance metrics with numerical values, sample sizes for test sets, data provenance, information about experts for ground truth, adjudication methods, details of comparative effectiveness studies (MRMC), or specifics about training sets.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (BD FACS Sample Prep Assistant II) through high-level study results.

    Here's an attempt to answer the questions based only on the provided text, with clear indications of what information is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Missing - Not explicitly stated in the document)

    The document states that the SPA III demonstrated performance "within specification" or "equivalent performance" in various studies, implying that there were pre-defined acceptance criteria. However, the specific numerical or qualitative criteria themselves are not provided.

    Reported Device Performance

    Study TypeGeneral Result of PerformanceSpecific Performance Metrics / Acceptance Criteria Met (Numerical/Qualitative)
    AccuracyThe SPA III with the BD FACSCanto system demonstrated equivalent performance.Missing (No numerical accuracy data or specific equivalence criteria provided)
    PrecisionThe SPA III with the BD FACSCanto system demonstrated system precision within specification.Missing (No numerical precision data or specific "specification" details provided)
    CarryoverThe SPA III with the BD FACSCanto system demonstrated system carryover within specification.Missing (No numerical carryover data or specific "specification" details provided)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. The "Accuracy" study mentions "Using Patient Samples," but the number of samples is not given.
    • Data Provenance (Country of Origin, Retrospective/Prospective): Not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Not applicable. This device is an automated sample preparation system for flow cytometry, not an AI/CADe device that interprets images or diagnoses conditions which would typically require expert ground truth.

    4. Adjudication Method for the Test Set

    • Not applicable for this type of device. The studies relate to the automated preparation process's accuracy, precision, and carryover, not diagnostic interpretation.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. This type of study is typically for diagnostic aids where human readers interpret cases. The SPA III is a sample preparation accessory.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, the performance studies (Accuracy, Precision, Carryover) evaluated the standalone device (SPA III with the BD FACSCanto system) without explicit human intervention in the preparation process itself. The "human-in-the-loop" concept is not directly applicable here as it's an automated lab instrument, not a diagnostic interpretation tool.

    7. The Type of Ground Truth Used

    • For the Accuracy study, the ground truth was based on "Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline, CLSI document EP9-A2." This implies comparison against an established reference method or the predicate device's performance. The specific nature of this "ground truth" (e.g., gold standard manual method, results from the predicate device) is not explicitly detailed beyond citing the CLSI guideline.
    • For Precision and Carryover, the ground truth implicitly refers to pre-defined "specifications" for these performance characteristics, as outlined in the cited CLSI and FDA guidelines.

    8. The Sample Size for the Training Set

    • Not specified. This device is a hardware and software system for automated sample preparation. It is unlikely to involve a "training set" in the machine learning sense for building a diagnostic algorithm. Its function relies on programmed protocols, not learned patterns from data.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no indication of a "training set" for an AI/ML algorithm.