Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K240613

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-06-20

    (107 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The APTUS Elbow Dorsal Olecranon Plates are indicated for fractures and osteotomies, in particular for the ulna.

    Device Description

    The subject device APTUS Elbow Dorsal Olecranon Plates are available in three (3) designs: Standard, Medium and Extended. The designs differ in the shape of the proximal ends of the Standard Plates are available in 6 lengths; The Medium Plates are available in 2 lenghts and the Extended Plates are available in 3 lenghts, each for left and right. All plates have appropriate for the olecranon. The subject device plates are compatible with screws and k-wires previously cleared in K051567, K09332, K112560, K232144 (for screws) and K092038 (for k-wires).

    The subject device plates are manufactured from titanium alloy, conforming to ASTM F136. All plates are provided sterile and selected plates are provided non-sterile aswell.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device (APTUS Elbow Dorsal Olecranon Plates), which focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific performance-based acceptance criteria through clinical studies or AI algorithm validation. Therefore, the information needed to answer many of your questions, particularly those related to AI performance, ground truth establishment, expert adjudication, and MRMC studies, is not present in this document.

    However, I can extract information regarding non-clinical testing and general regulatory context, and then explain why other requested details are absent.

    Here's a breakdown based on the provided text, and an explanation of what cannot be answered:

    1. A table of acceptance criteria and the reported device performance

    The document mentions non-clinical (mechanical) testing based on a specific ASTM standard and FDA guidance, but it does not provide a table of acceptance criteria and reported performance values. It only states a general conclusion:

    • Acceptance Criteria Mentioned: Minimum requirements pointed out in the "FDA guidance for Industry and Food and Drug Administration Staff, Orthopedic Fracture Fixation Plates – Performance Criteria for Safety and Performance Based Pathway (2022)" and ASTM F382.
    • Reported Device Performance: "The mechanical test results show that the bending structural stiffness of the Dorsal Olecranon Medartis APTUS Elbow Plates are higher than the acceptance values indicated in the relative FDA guidance..."
    • Specific Values: No specific numerical acceptance criteria or reported device performance values (e.g., in N/mm or similar units) are provided in this regulatory letter.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified for non-clinical (mechanical) tests. For clinical studies, the document explicitly states: "Clinical data were not provided in this submission." This means there was no clinical test set for patient data.
    • Data Provenance (e.g., country of origin of the data, retrospective or prospective): Not applicable as no clinical data or patient test sets were used or provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not applicable. This information would be relevant for studies involving human interpretation or AI algorithm validation using expert-labeled data, which were not part of this 510(k) submission. No clinical ground truth was established from patient data.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not applicable. Adjudication methods are used to establish ground truth in clinical or imaging studies, which were not conducted for this 510(k) submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The document explicitly states: "Clinical data were not provided in this submission." An MRMC study, especially one involving AI assistance, would fall under clinical data. This submission is for a traditional medical device (bone plates), not an AI/ML-driven diagnostic or therapeutic device that would typically undergo such a study.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No. This study type is for AI/ML algorithms. The device submitted is a mechanical orthopedic implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the non-clinical mechanical testing, the "ground truth" or reference for evaluating performance was defined by the stated FDA guidance and ASTM standards for mechanical properties (e.g., bending stiffness). There was no "ground truth" concerning patient outcomes, pathology, or expert consensus on clinical imaging, as no clinical data was used.

    8. The sample size for the training set

    • Not applicable. This applies to AI/ML algorithms that require training data. This submission is for a physical medical device.

    9. How the ground truth for the training set was established

    • Not applicable for the same reason as point 8.

    Summary of Device Acceptance & Study

    The acceptance of this device by the FDA (K240613) is based on its substantial equivalence to legally marketed predicate devices (K120070, K193633, K193639, K193554) and successful completion of non-clinical (mechanical) testing.

    • Acceptance Criteria for Substantial Equivalence: Similar intended use, similar technological characteristics (design, principle of operation, material), and comparable safety and effectiveness, supported by non-clinical testing.
    • Study Proving Acceptance:
      • Study Type: Non-clinical (mechanical) testing.
      • Objective: To demonstrate that the bending structural stiffness of the APTUS Elbow Dorsal Olecranon Plates is at least "higher than the acceptance values indicated in the relative FDA guidance for Industry and Food and Drug Administration Staff, Orthopedic Fracture Fixation Plates – Performance Criteria for Safety and Performance Based Pathway (2022)" and meets ASTM F382.
      • Results: The mechanical tests confirmed that the device's stiffness was higher than the specified acceptance values.
      • Clinical Data: No clinical data or studies were performed or submitted. The FDA 510(k) process for traditional devices like bone plates often relies on substantial equivalence and non-clinical data if the technological characteristics are well-understood.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1