Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K110947

    Validate with FDA (Live)

    Device Name
    APEX K1 HIP STEM
    Date Cleared
    2011-06-16

    (73 days)

    Product Code
    Regulation Number
    888.3358
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Apex Hip System K1™ Hip Stem is intended for use as the femoral component of a primary or revision total hip replacement when used with the Apex Interface™ Acetabular System. The Apex Interface™ Acetabular System articulates with the Apex Modular Femoral Head (Cobalt Chromium or Ceramic). The femoral hip stem is intended for uncemented fixation and single use implantation. These prosthesis may be used for hip arthroplasty to treat the following conditions, as appropriate:
    • Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
    • Rheumatoid arthritis;
    • Correction of functional deformity;
    • Congenital dislocation;
    • Revision procedures where other treatments or devices have failed;
    • Femoral neck and trochanteric fractures of the proximal femur.

    The Apex Hip System K1 Femoral Stem is also intended for use in hemiarthroplasty when used with the Apex Bipolar Head.

    The Apex Hip System Bipolar Head is intended for use in combination with an Apex Hip System femoral stem for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:
    • Femoral neck and trochanteric fractures of the proximal femur;
    • Osteonecrosis of the femoral head;
    • Revision procedures where other treatments or devices for these indications have failed.

    Device Description

    The Apex K1™ Hip Stems are rectangular in cross-section and tapered in the proximal-to-distal direction. The K1 Hip Stem is offered in 12 sizes and two neck offsets for each size. Standard necks have a 135 degree angle and lateralized (Lat) necks have a 130 degree angle. Sizes 7 to 12 also offer a Lat + offset with a 127 degree angle.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Apex K1™ Hip System, a medical device for hip replacement. It outlines the device's characteristics, intended use, and a comparison to a predicate device. However, it explicitly states: "No clinical studies were performed."

    Therefore, it is impossible to provide the requested information about acceptance criteria, device performance, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, or ground truth, as no clinical study was conducted for this device.

    The document indicates that the device's substantial equivalence to a legally marketed predicate device (K060072 - Apex K1™ Hip Stem) was established through non-clinical testing.

    Here's a summary based on the provided text, highlighting the absence of clinical study data:


    Absence of Clinical Study Data:

    The provided 510(k) Summary explicitly states: "No clinical studies were performed." Therefore, the device performance against acceptance criteria, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, and ground truth establishment, which are typically derived from clinical studies, cannot be detailed.

    Instead, the device's substantial equivalence was demonstrated through non-clinical testing and comparison to an existing predicate device.

    Non-Clinical Test Summary:

    The following non-clinical tests were conducted:

    • Fatigue Strength Testing: Performed according to ISO 7206-6, ISO-7206-4, ISO 7206-8, and ASTM 2068-09.
    • ROM (Range of Motion) evaluation: Performed per ISO 21535.

    Predicate Device Comparison:

    The Apex K1™ Hip Stem (subject device, sizes 0, 0 Lat, 1, 1 Lat, and 7 thru 12 Lat +) was found to be substantially equivalent to the predicate device (Apex K1™ Hip Stem, Sz 2 to 12, K060072) based on the following:

    CharacteristicApex K1 Hip Stem (Subject Device)Apex K1 Hip Stem (Predicate Device, K060072)
    Intended UseYes (for primary and revision total hip replacement)Yes (for primary and revision total hip replacement)
    Stem DesignIdentical to K060072Rectangular in cross-section and tapered in the proximal-to-distal direction
    Materials (K1 Stem)Identical to K060072Ti6Al4V per ASTMF136
    Plasma spray Titanium coatingIdentical to K060072Ti Coating per ASTM F1580
    SterilizationIdentical to K060072Ethylene oxide
    SAL (Sterility Assurance Level)Identical to K06007210^-6^
    PackagingIdentical to K060072Paper Board Box, Double Tyvek inner pouch

    Since no clinical studies were performed, the specific details requested for clinical performance acceptance criteria, study design, and ground truth establishment cannot be provided from this document. The basis for acceptance for this device was its substantial equivalence to a previously cleared predicate device, supported by the mentioned non-clinical engineering and material testing.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1