Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K113242

    Validate with FDA (Live)

    Date Cleared
    2012-01-05

    (64 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Apex ARC™ Hip Stem is intended for use as the femoral component of a primary, or revision total hip replacement when used with the Apex Interface™ Acetabular System. The Apex Interface™ Acetabular System articulates with the Apex Modular Femoral Head (Cobait Chromium or Ceramic). The femoral hip stem is intended for uncemented fixation and single use implantation. These prostheses may be used for hip arthroplasty to treat the following conditions, as appropriate:
    • Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
    • Rheumatoid arthritis;
    • Correction of functional deformity;
    • Congenital dislocation;
    • Revision procedures where other treatments or devices have failed;
    • Femoral neck and trochanteric fractures of the proximal femur.

    The Apex Hip System ARC™ Hip Stem is also intended for use in hemiarthroplasty when used with the Apex Bipolar Head.

    The Apex Hip System Bipolar Head is intended for use in combination with an Apex Hip System femoral stem for uncemented primary or revision hemiathroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:
    • Femoral neck and trochanteric fractures of the proximal femur;
    • Osteonecrosis of the femoral head;
    • Revision procedures where other treatments or devices for these indications have failed.

    Device Description

    The Apex ARC Hip Stem consists of a curved, rectangular tapered stem with a distal slot, and modular necks that connect to the tapered hole in the stem. The larger size stems have a lateral feature referred to as a lateral t-flange. The T-flange is reduced in size on the smaller stem sizes (size 1 and 2), and further reduced to no t-flange on the smallest (subject) size 0 stem. The femoral stems are manufactured from titanium alloy and the modular necks are manufactured from cobalt chromium alloy. The Apex ARC Hip Stem is available with and without HA coating.

    The necks are compatible with the Cobalt Chromium and Ceramic modular heads, and may be used with head diameters and offsets up to a maximum offset of +7 mm. The Apex ARC Hip Stem may be used in conjunction with the Apex Interface™ Acetabular System (Shells and Inserts) for total hip arthroplasty.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Apex ARC™ Hip Stem, focusing on acceptance criteria and the study proving conformance.

    Important Note: The provided document is a 510(k) Pre-market Notification for a medical device. This type of submission primarily focuses on demonstrating substantial equivalence to a previously legally marketed device (predicate device), rather than proving efficacy or meeting specific acceptance criteria through extensive clinical studies as one might find for a novel drug or a high-risk device requiring a PMA.


    Description of Acceptance Criteria and the Study Proving the Device Meets Acceptance Criteria

    The document states that the Apex ARC™ Hip Stem (specifically the new Size 0) is substantially equivalent to its predicate devices (Apex ARC Hip Stem K090845 and K111193). The primary method used to demonstrate this substantial equivalence, and thus "meet acceptance criteria," is through non-clinical testing and comparison of design features, materials, and intended use.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) for substantial equivalence based on a new size of an existing device, the "acceptance criteria" are not explicitly defined as pass/fail thresholds for clinical outcomes, but rather demonstrate that the new size performs comparably to the predicate device in relevant biomechanical tests.

    Acceptance Criteria (Implied by Substantial Equivalence to Predicate)Reported Device Performance
    Fatigue Strength: Must meet or exceed established standards for hip femoral components.Testing performed per ISO 7206-6, ISO-7206-4, ISO 7206-8, and ASTM 2068-09. (Specific numerical results not provided in this summary, but the conclusion is that the device is substantially equivalent, implying it met these standards.)
    Range of Motion (ROM): Must exhibit acceptable ROM.Evaluation performed per ISO 21535. (Specific numerical results not provided, but the conclusion implies acceptable performance.)
    Biocompatibility: (Implicit for titanium alloy and HA coating)Manufactured using equivalent materials to predicate(s). (Implies biocompatibility is assumed to be met based on predicate).
    Material Properties: (Implicit)Manufactured using equivalent materials to predicate(s).
    Design Features, Intended Use, Indications for Use: (Implicit)Compared to predicate device(s) and found to be substantially equivalent.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated as this is a non-clinical mechanical and material testing summary. For such tests, the sample size typically refers to the number of individual test articles (e.g., hip stems) subjected to each specific test. This information is usually detailed in the full test reports, which are not provided here.
    • Data Provenance: The testing was "conducted," implying laboratory testing. The country of origin for the data is implied to be within the scope of the submitter, OMNIlife science, Inc., based in East Taunton, MA, USA, and adherence to international standards (ISO, ASTM). The data is prospective in the sense that the new size 0 was manufactured and then tested.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not Applicable. For non-clinical, biomechanical testing of a medical device, "ground truth" is established by relevant engineering standards (ISO, ASTM) and the objective measurements obtained from properly calibrated testing equipment. There is no expert consensus involved in establishing the "ground truth" for the mechanical properties of a hip stem.

    4. Adjudication Method for the Test Set

    • Not Applicable. Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies or image-based diagnostic studies where human interpretation of data is involved. Here, the "study" is a series of non-clinical tests with objective, quantitative outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. The document explicitly states: "No clinical studies were performed." Therefore, no MRMC comparative effectiveness study was conducted.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done

    • Not Applicable. This device is a physical hip stem, not a software algorithm or AI-driven diagnostic tool. Therefore, the concept of "standalone algorithm performance" does not apply.

    7. The Type of Ground Truth Used

    • For the non-clinical tests (Fatigue Strength, ROM Evaluation), the "ground truth" is defined by established engineering standards and specifications (ISO 7206 series, ASTM 2068-09, ISO 21535) for hip implant performance. The device's measured performance against these standards constitutes the "ground truth" for its mechanical properties. There is no pathology, outcomes data, or expert consensus used for the "ground truth" in this context.

    8. The Sample Size for the Training Set

    • Not Applicable. This report does not involve machine learning or AI, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As there is no training set, there is no ground truth establishment for it.

    Summary of the Study Proving Device Acceptance Criteria:

    The "study" proving the Apex ARC™ Hip Stem (Size 0) meets its "acceptance criteria" (which in this 510(k) context means demonstrating substantial equivalence) is a non-clinical testing program.

    This program involved:

    • Fatigue Strength Testing: Performed according to international standards ISO 7206-6, ISO-7206-4, ISO 7206-8, and ASTM 2068-09. This testing assesses the material's ability to withstand repeated stress without failure, simulating the forces experienced during daily activity.
    • Range of Motion (ROM) Evaluation: Performed according to ISO 21535. This evaluates the articulation capabilities of the hip stem in conjunction with other components to ensure proper biomechanical function and prevent impingement.

    The conclusion drawn from these studies is that the "addition of one new stem size for the Apex ARC Hip Stem, in our opinion, is substantially equivalent to the predicate device." This implicitly means that the new size 0 performed comparably to the predicate devices and met the relevant performance standards for hip stems. No clinical studies were performed to support this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111193

    Validate with FDA (Live)

    Date Cleared
    2011-06-27

    (60 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Apex ARC™ Hip Stem is intended for use as the femoral component of a primary, or revision total hip replacement when used with the Apex Interface™ Acetabular System. The Apex Interface™ Acetabular System articulates with the Apex Modular Femoral Head (Cobalt Chromium or Ceramic). The femoral hip stem is intended for uncemented fixation and single use implantation. These prostheses may be used for hip arthroplasty to treat the following conditions, as appropriate:

    • Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
    • Rheumatoid arthritis;
    • · Correction of functional deformity;
    • · Congenital dislocation;
    • · Revision procedures where other treatments or devices have failed;
    • · Femoral neck and trochanteric fractures of the proximal femur.

    The Apex Hip System ARC™ Hip Stem is also intended for use in hemiarthroplasty when used with the Apex Bipolar Head.

    The Apex Hip System Bipolar Head is intended for use in combination with an Apex Hip System femoral stem for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:

    • · Femoral neck and trochanteric fractures of the proximal femur;
    • · Osteonecrosis of the femoral head;
    • Revision procedures where other treatments or devices for these indications have failed.
    Device Description

    The Apex ARC Hip Stem consists of a curved, rectangular tapered stem, and modular necks that connect to the tapered hole in the stem. The height of the anterior and posterior lateral protrusions (Lateral T-Flange) was reduced on Size 1 and 2 ARC Hip Stems. Three new neck sizes are offered, Long Neutral, Long 8 ° Varus/Valgus and Anteverted. The necks are compatible with the Cobalt Chromium and Ceramic modular heads, and may be used with head diameters and offsets up to a maximum offset of +7 mm. The Apex ARC Hip Stem may be used in conjunction with the Apex Interface™ Acetabular System (Shells and Inserts) for total hip arthroplasty.

    AI/ML Overview

    The Apex Hip System ARC™ Hip Stem is a medical device and the provided text describes its acceptance criteria and the study conducted to prove it meets them.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Apex Hip System ARC™ Hip Stem are based on demonstrating substantial equivalence to a predicate device (Apex ARC Hip Stem - K090845). This equivalence is primarily shown through non-clinical performance testing.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Demonstrated Equivalence to Predicate)Reported Device Performance
    Intended UsePrimary and revision total hip replacement (same as predicate)Met
    Design- T-Flange reduction (by half on Size 1 and 2 ARC Hip Stem)Met (Design change described and tested for performance)
    - Curved, rectangular tapered stem (same as predicate)Met
    - New neck designs: Long neutral, long 8° varus/valgus, antevertedMet (Compatibility and performance tested)
    Materials- Apex ARC Hip Stem (Ti6Al4V per ASTM F136) (identical to predicate)Met
    - Plasma spray Titanium coating (Unalloyed titanium plasma spray per ASTM F1580) (identical to predicate)Met
    - Hydroxyapatite Coating (Hydroxyapatite plasma spray coating per ASTM F1185) (identical to predicate)Met
    - Apex ARC Modular Neck (CoCr alloy per ASTM F1537) (identical to predicate)Met
    Packaging & Sterilization- Sterilization (Ethylene oxide) (identical to predicate)Met
    - SAL (10^-6^) (identical to predicate)Met
    - Packaging (Paper Board Box, Double Tyvek inner pouch) (identical to predicate)Met
    Non-Clinical Performance- Fatigue Strength Testing per ISO 7206-6, ISO-7206-4, ISO 7206-8 and ASTM 2068-09 (demonstrate equivalence/safety after design changes)Met (Tests conducted)
    - ROM evaluation per ISO 21535 (demonstrate equivalence/safety after design changes)Met (Tests conducted)

    2. Sample size used for the test set and the data provenance

    The document states, "Non-Clinical Test: The following tests were conducted." It then lists fatigue strength testing and ROM evaluation. However, the exact sample size (number of devices or components tested) for these non-clinical tests is not specified in the provided text.

    The data provenance is from non-clinical laboratory testing performed on the device itself (or its components) according to specified ISO and ASTM standards. No human or animal data is involved; thus, there is no country of origin or retrospective/prospective distinction in the typical sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a mechanical implant, and the acceptance criteria for performance are based on engineering standards and physical testing, not subjective expert assessment of medical images or patient outcomes as would be the case for an AI device.

    4. Adjudication method for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in AI studies where human experts disagree on interpretations. For mechanical performance testing of an implant, the results are objective measurements against defined standards, not subject to expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a hardware medical device (hip stem), not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and assessments of human reader improvement with AI assistance are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical implant, not an algorithm, so "standalone performance" in the context of AI does not apply.

    7. The type of ground truth used

    The "ground truth" for this device's performance is established by objective measurements against recognized industry standards (ISO and ASTM) for mechanical properties (fatigue strength) and range of motion. The predicate device's performance under these same standards effectively sets the benchmark for "ground truth" for demonstrating substantial equivalence.

    8. The sample size for the training set

    Not applicable. This section usually refers to the data used to train an AI model. Since this is a physical medical device, there is no "training set" in this context. The manufacturing and design process follows established engineering principles and quality systems; it's not "trained" on data.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for a physical device, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090845

    Validate with FDA (Live)

    Date Cleared
    2010-04-07

    (376 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Apex ARC Hip Stem is intended for use as the femoral component of a primary or revision total hip replacement. This femoral hip stem is intended for uncemented fixation and single use implantation. This prosthesis may be used for the following conditions, as appropriate:

    • Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
    • Rheumatoid arthritis;
    • Correction of functional deformity;
    • Congenital dislocation:
    • Revision procedures where other treatments or devices have failed;
    • Femoral neck and trochanteric fractures of the proximal femur.
    Device Description

    The Apex ARC Hip Stem consists of a curved, rectangular tapered stem, and modular necks that connect to the tapered hole in the stem. The femoral stems are manufactured from titanium alloy and the modular necks are manufactured from cobalt chromium alloy. Three neck sizes are offered, with a neutral, 8 degree, and 12 degree varus-valgus angle, respectively. The necks are compatible with the modular heads that are part of the Apex Modular and Apex K2 hip systems (K000788, K012918, and K073150) and may be used with head diameters and offsets up to a maximum offset of +7 mm. These configurations allow the user to choose a combination of stem, neck, and head components to appropriately fit the anatomy of the patient. The Apex ARC Hip Stem may be used in conjunction with the Apex Modular™ Acetabular Cup (K031110, K062489, and K073150) for total hip arthroplasty.

    AI/ML Overview

    The provided document describes the 510(k) Summary for the Apex ARC™ Hip Stem (K090845). This is a medical device submission to the FDA, and as such, it focuses on demonstrating "substantial equivalence" to predicate devices rather than proving performance against specific acceptance criteria through a traditional clinical study with defined endpoints as would be expected for a novel AI/software device.

    Therefore, the requested information elements (acceptance criteria, study details, sample sizes, expert ground truth, MRMC, standalone performance) are not directly applicable in the context of this device's submission type and the information provided. Traditional clinical trials with acceptance criteria and statistical analysis are typically not required for Class II devices demonstrating substantial equivalence through non-clinical testing.

    However, I can extract the relevant information regarding the non-clinical tests conducted which serve a similar purpose to demonstrating the device's functional integrity and safety, and compare them to the general characteristics of the predicate devices.

    Here's a breakdown based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) submission for a mechanical implant, "acceptance criteria" are not explicitly defined in terms the performance metrics commonly associated with AI/software (e.g., sensitivity, specificity, AUC). Instead, the device's "performance" is demonstrated through non-clinical mechanical and material testing based on recognized international and ASTM standards. The acceptance criterion for these tests is generally successful completion of the test without failure, meeting the requirements of the specified standard, and demonstrating equivalence to predicate devices.

    Acceptance Criterion (Implicit: Compliance with Standard and Predicate)Reported Device Performance
    Mechanical Strength & Durability:
    Distal fatigue strength per ISO 7206-4:1989, ISO 7206-8:1992, and ASTM 2068-09.Tests conducted and results deemed successful.
    Fretting potential per ISO 17853:2003.Tests conducted and results deemed successful.
    Disassembly strength after fatigue testing per ASTM F2009-00.Tests conducted and results deemed successful.
    Proximal fatigue strength per ISO 7206-6:1992 and ASTM 2068-09.Tests conducted and results deemed successful.
    Torsional strength of the modular neck.Tests conducted and results deemed successful.
    Burst test, fatigue test, post-fatigue burst test, pull-off, and rotational stability of the worst case modular neck-ceramic head combination per FDA Guidance for the Preparation of Premarket Notifications for Ceramic Ball Hip Systems, and ISO 7206-10:2003.Tests conducted and results deemed successful.
    Functional Compatibility:
    Range of motion per ISO 21535:2007.Tests conducted and results deemed successful.
    Biocompatibility/Sterilization:
    ETO Residuals per ANSI/AAMI/ISO 10993-7.Tests conducted and results deemed successful.
    Hydroxyapatite coating clearance.Previously cleared in K043123.
    Design and Material Equivalence to Predicates:
    Demonstrated through detailed comparison table with Apex K2™ Hip System and Aesculap Metha® Hip System.Features and materials largely align; differences (e.g., modular neck material and junction) addressed through specific additional testing where needed (e.g., torsional strength, burst/fatigue tests for neck combination).

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified in terms of number of physical devices tested for each non-clinical test. For mechanical tests, representative samples are typically used to meet the requirements of the standards. For material properties, it refers to the composition. The document doesn't detail the number of physical units tested.
    • Data Provenance: The tests are non-clinical, meaning they were performed in a laboratory setting, not on human subjects. No country of origin is specified for the testing, but the submitter (OMNI life science, Inc.) is based in E. Taunton, MA, USA. This is a prospective assessment for market clearance, meaning the tests were conducted specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to this 510(k) summary. For mechanical devices, "ground truth" typically relates to objective physical measurements or material properties as defined by engineering standards, not expert consensus as in diagnostic imaging. The "experts" involved would be the engineers and technicians conducting and validating the tests, adhering to the specified ISO and ASTM standards. Their qualifications are inherent in the rigorous nature of these testing facilities.

    4. Adjudication method for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for expert consensus on clinical findings, not for objective mechanical test results. The results of the non-clinical tests are binary (pass/fail compliance with the standard).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a total hip replacement device, not an AI/software diagnostic tool, and therefore no MRMC study was performed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this device's evaluation is primarily based on engineering standards (ISO, ASTM) for mechanical performance and material properties. This includes objective measurements of fatigue strength, fretting potential, disassembly strength, torsional strength, range of motion, and material composition. Biocompatibility (ETO residuals) is also assessed against established standards.

    8. The sample size for the training set

    Not applicable. This is a physical device, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.


    In summary: The Apex ARC™ Hip Stem, a Class II medical device, demonstrates its safety and effectiveness for market clearance by showing substantial equivalence to existing predicate devices (Apex K2™ Hip System and Aesculap Metha® Hip System) through a detailed comparison of design, materials, and intended use, and by successfully completing a series of non-clinical mechanical and material tests according to recognized international standards (ISO, ASTM) and FDA guidance documents. No clinical studies or AI-related evaluations were performed for this submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1