510(k) Data Aggregation

    K Number
    K221116

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-13

    (28 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XACT ACE Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of an instrument during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning and is intended to assist the physician in positioning of an instrument, such as a needle, where CT imaging is used for target trajectory planning and intraoperative tracking.

    Device Description

    The XACT ACE Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of instruments during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning based on CT images and is intended to assist the physician in positioning of an instrument, such as a needle, and reviewing instrument position during advancement to the target. The system quides (i.e., positions and steers) the instrument according to a predefined trajectory. The physician controls advancement of the instrument along the trajectory using a foot pedal. The system also allows for monitoring of motion associated with respiration during the procedure.

    The XACT ACE Robotic System comprises the following main components:

    • XACT ACE Robot which is placed on the patient and includes the robot positioning unit & the insertion module assembly
    • XACT ACE Console which includes a Control Unit, central computer (in the Control Unit) and monitor workstation for user trajectory planning, user interface and review of instrument position.
    AI/ML Overview

    The XACT ACE Robotic System is a modification of a previously cleared device (XACT Robotic System, ACE Model K201586) to expand the list of supported instruments to include ablation probes. The performance data provided focuses on verifying the compatibility and accuracy of the device when used with these new ablation probes.

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA clearance document does not explicitly present a table of acceptance criteria with numerical thresholds. Instead, it describes the verification and validation process in terms of meeting performance specifications and demonstrating similar performance characteristics to biopsy introducers. Based on the provided text, the key performance aspects evaluated were:

    Acceptance Criteria (Inferred from study description)Reported Device Performance
    Efficiently secures ablation probes.Instrument Insertion Kit efficiently secures the probes.
    Advances probes to target without physical damage.Probes advance to target without any physical damage.
    Robotic system configured for physical characteristics of probes.All relevant physical characteristics were adequately measured and configured in the robotic system.
    Meets performance specifications for steering to targets.Steering to targets met performance specifications.
    Meets target accuracy (tip-to-target average accuracy) comparable to biopsy introducers in a simulated clinical environment.Device meets all of its performance characteristics, with similar performance characteristics (tip-to-target average accuracy) compared to biopsy introducers.
    No changes to clinical workflow.Demonstrated no changes to the clinical workflow.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • For the "Needle Properties" protocol: The document states "each probe" was measured, implying a sample size of 4 ablation probes (corresponding to the four types of ablation probes introduced).
      • For the simulated clinical environment test: The document mentions steering "each of the probes to a variety of targets" and "steers the compatible listed ablation probes to a variety of targets." It doesn't specify an exact number of procedures or targets, but it implies a comprehensive evaluation across different scenarios for each of the 4 probe types.
    • Data Provenance: The studies were conducted internally as part of the device's verification and validation. The studies were performed in a simulated clinical environment (CT suite using a specific phantom), indicating a controlled, prospective experimental setup rather than retrospective use of human patient data. No country of origin for the data is explicitly mentioned beyond the submitter's address in Israel.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not provide details on the number or qualifications of experts used to establish ground truth for the test set. The ground truth for the simulated clinical environment test appears to be defined by the pre-planned "variety of targets" within the phantom. The evaluation of performance and meeting specifications would likely be overseen by engineers and potentially clinical specialists involved in the test.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1). The evaluation seems to rely on direct measurements and observations during the automated and user-controlled robotic procedures in the simulated environment, with the system's ability to reach predefined targets being the primary metric.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No multi-reader multi-case (MRMC) comparative effectiveness study was done. The study described focuses on the device's performance with new instruments rather than comparing human readers with and without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The studies described are for a user-controlled positioning system, and the "user advances the instrument along the trajectory using a foot pedal." While the robot autonomously steers, the overall procedure involves human input and control. The performance tests were conducted to demonstrate the robot's ability to steer and position instruments in conjunction with human control and CT imaging for planning and tracking, which is not a purely standalone algorithm-only performance assessment.

    7. Type of Ground Truth Used

    The ground truth used for the test set was based on predefined targets within a specific phantom in a simulated clinical environment. This is a form of objective physical ground truth established through the experimental setup rather than expert consensus, pathology, or outcomes data from human patients.

    8. Sample Size for the Training Set

    The document does not mention any separate training set or machine learning aspects for the expanded functionality. The system's operation appears to be based on pre-programmed algorithms and physical characteristics of the instruments, not on a machine learning model that would require a dedicated training set. The modifications involve supporting new instruments, which primarily required physical characterization and integration.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned in relation to the new functionality, this point is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201586

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-07-09

    (28 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XACT Robotic System, ACE Model is a user-controlled positioning system intended to assist in the planning and advancement of an instrument during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning and is intended to assist the physician in positioning of an instrument, such as a needle, where CT imaging is used for target trajectory planning and intraoperative tracking.

    Device Description

    The XACT Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of instruments during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning based on CT images and is intended to assist the physician in positioning of an instrument, such as a needle, and reviewing instrument position during advancement to the target. The system guides (i.e., positions and steers) the instrument according to a predefined trajectory. The physician controls advancement of the instrument along the trajectory using a foot pedal. The system also allows for monitoring of motion associated with respiration during the procedure.

    The XACT Robotic System, ACE Model comprises the following main components:

    • . XACT Robot which is placed on the patient and includes the robot positioning unit & the insertion module assembly
    • . XACT Console which includes a Control Unit, central computer (in the Control Unit) and monitor workstation for user trajectory planning, user interface and review of instrument position
    AI/ML Overview

    I am sorry, but the provided text does not contain the detailed information necessary to answer your request fully. The document is an FDA 510(k) clearance letter and a summary of the device's characteristics and performance tests.

    Here's what I can provide based on the given text, and what is missing:

    Information available from the text:

    • Device Name: XACT Robotic System, ACE Model
    • Predicate Device: XACT Robotic System (K191332)
    • Performance Tests Conducted: Mechanical Accuracy, Registration Precision, CT Phantom Accuracy, Robot Rigidity, Axes Mapping, Complete SW STD, Electrical safety (IEC 60601-1), Electromagnetic compatibility (IEC 60601-1-2).

    Missing Information:

    The text does not provide the following crucial details needed for your request:

    • A table of acceptance criteria and reported device performance: While performance tests are listed, the specific acceptance criteria for each test and the quantitative results showing the device met those criteria are not provided.
    • Sample sizes used for the test set and data provenance: The document mentions tests but not the number of cases or samples used, nor the origin (country, retrospective/prospective) of any data if clinical studies were performed.
    • Number of experts and their qualifications for establishing ground truth: This information is not present.
    • Adjudication method for the test set: Not mentioned.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, and the effect size: The text indicates performance and safety tests, but not comparative effectiveness studies with human readers.
    • If a standalone study (algorithm only) was done: The device is a "user-controlled positioning system intended to assist in the planning and advancement of an instrument." This suggests it's a human-in-the-loop system, and a standalone algorithm performance without human interaction isn't directly relevant or discussed.
    • The type of ground truth used: Not specified.
    • Sample size for the training set: The document describes performance tests and comparisons to a predicate device, but not a "training set" in the context of an algorithm's development. This device is a robotic system, not an AI diagnostic algorithm in the typical sense that would have a training set of data.
    • How the ground truth for the training set was established: Not applicable, as no training set for an algorithm is mentioned.

    Conclusion based on the provided text:

    The document describes a robotic system that assists in CT-guided procedures. It outlines several engineering and safety performance tests that were conducted to demonstrate substantial equivalence to a predicate device. However, it does not provide the specific quantitative acceptance criteria or results from these tests, nor does it delve into the details of clinical study design that would involve ground truth establishment, expert adjudication, or reader studies. It primarily focuses on the device's technological characteristics and safety and effectiveness from an engineering standpoint, rather than a diagnostic algorithm's performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191332

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-10-03

    (140 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XACT Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of an instrument during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning and is intended to assist the physician in positioning of an instrument, such as a needle, where CT imaging is used for target trajectory planning and intraoperative tracking.

    Device Description

    The XACT Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of instruments during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning based on CT images and is intended to assist the physician in positioning of an instrument, such as a needle, and reviewing instrument position during advancement to the target. The system quides (i.e., positions and steers) the instrument according to a predefined trajectory. The physician controls advancement of the instrument along the trajectory using a foot pedal. The system also allows for monitoring of motion associated with respiration during the procedure.

    The XACT Robotic System comprises the following main components:

    • . XACT Robot Positioning Unit & Insertion Module Assembly which is placed on the patient
    • XACT Control Unit (Central Computer) .
    • . XACT Computer Work Station which includes the user interface for trajectory planning and review of instrument position
    • XACT System Cart ●
    AI/ML Overview

    The provided text describes the XACT Robotic System, a user-controlled positioning system intended to assist in CT-guided percutaneous procedures. While the document outlines several performance tests and mentions preliminary clinical data, it does not explicitly define acceptance criteria as a table with numerical targets, nor does it detail a comparative effectiveness study (MRMC) or a standalone AI algorithm. It provides some performance figures from pre-clinical and preliminary clinical studies.

    Here's an attempt to extract the information based on the provided text, acknowledging the limitations regarding the completeness of "acceptance criteria" as you might expect in a formal table with specific thresholds:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't present "acceptance criteria" in a formal table with specific numerical thresholds for each test. Instead, it describes performance tests and their outcomes.

    Performance Test CategoryReported Device Performance (Proxy for Acceptance)
    Pre-clinical Animal Study (Porcine Model)- Target Reach: System reached the target in all 40 interventional procedures.
    - Average Accuracy: 1.41mm ± 0.94mm (Distance from target).
    - Adverse Events: No adverse events reported specific to the system (implied by "demonstrate the safety and effectiveness").
    Preliminary Clinical Data- Target Reach/Success Rate: Device successfully reached the target in 27 out of 31 cases (87.1%).
    - Procedure Completion (Manual Intervention): In 4 out of 31 subjects, the procedure was completed successfully with manual insertion after initial system use.
    - Specifications Met: "The system met its predefined specifications for use to guide and advance an instrument during CT-guided interventional procedures." (Specific values not provided).
    - Serious Adverse Events: No serious adverse events reported for any of the subjects.
    Bench/Laboratory Tests (General Statements)- Mechanical Accuracy: In accordance with ASTM F2554-10 (standard met).
    - Registration Precision: Assessed in a simulated clinical environment.
    - CT Phantom Accuracy: Evaluated device performance and system accuracy.
    - Rigidity: Tested under different weight loads.
    - Axes Mapping: Evaluated performance specifications of each axis.
    - Needle Properties: Evaluated mechanical strength of compatible needles.
    - Body Contour & Movement/Respiration Synchronization: Evaluated compatibility and synchronization.
    - System Lifetime Testing: Conducted.
    - Electrical Safety: In accordance with IEC 60601-1 (standard met).
    - Electromagnetic Compatibility (EMC): In accordance with IEC 60601-1-2 (standard met).
    - Biocompatibility: In accordance with ISO 10993 (standard met).

    2. Sample size used for the test set and the data provenance

    • Test Set (Pre-clinical animal study):
      • Sample Size: 40 interventional procedures.
      • Data Provenance: Porcine model (animal study). Specific country of origin is not mentioned, but it's a pre-clinical study.
    • Test Set (Preliminary clinical data):
      • Sample Size: 31 cases (interim set of subjects).
      • Data Provenance: Clinical data gathered in 2 clinical centers. Specific country of origin is not mentioned. It is prospective data collection since it's described as "preliminary clinical data, to supplement the bench and preclinical animal testing."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number of experts for "ground truth" establishment in the context of the animal or clinical studies. The results from the animal study (target reach, accuracy) and clinical study (successful target reach, adverse events) are presented as direct measurements. For the clinical data, the procedures were performed by physicians, as indicated by "physician defined trajectory path" and "manual insertion according to routine clinical procedure," implying the usual clinical practice standards would serve as the "truth" for procedure success and safety. However, the exact process of establishing ground truth by a panel of experts is not described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1 consensus) for establishing ground truth or evaluating outcomes in either the pre-clinical or preliminary clinical studies. Outcomes appear to be directly measured (e.g., distance from target, successful reach, reported adverse events).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study is mentioned. The study focuses on the device's ability to assist in procedures, not on evaluating human reader performance with or without AI assistance, as the system is a robotic positioning system, not an AI-based diagnostic tool for image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a "user-controlled positioning system" and "intended to assist in the planning and advancement of an instrument." It is designed for human-in-the-loop operation, where the physician controls advancement and defines the trajectory. Therefore, a standalone (algorithm only) performance evaluation without human interaction would not be applicable or described for this type of device. The accuracy results (1.41mm ± 0.94mm) from the animal study could be considered a measure of the system's standalone (robotic execution) accuracy once a trajectory is defined.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Pre-clinical Animal Study: The ground truth for accuracy was based on direct measurement of the "distance from the target" following robotic insertion in animal models. Successful target reach was also a direct outcome. Safety was assessed through reporting of adverse events.
    • Preliminary Clinical Data: The ground truth for success appears to be whether the device "successfully reached the target" and whether the "procedure was completed successfully" (even if later by manual intervention). Safety was assessed by "no serious adverse events reported." These are outcomes-based or procedural success-based ground truths, likely determined by the operating physician and recorded.

    8. The sample size for the training set

    The document does not provide information about a "training set" for an AI algorithm because the XACT Robotic System is described as a "user-controlled positioning system," not explicitly an AI-driven image analysis or diagnostic tool that would typically undergo a separate training phase with a large dataset. While it has software for planning, the core function described is mechanical assistance and guidance.

    9. How the ground truth for the training set was established

    As no training set for an AI algorithm is mentioned, the method for establishing its ground truth is also not applicable or described in the provided text.