Q-LF handpiece, Q-FX handpiece, Q-SC handpiece, and Q-BT handpiece in RF mode are intended for:
- Relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local blood circulation.
Q-EP handpiece in cooling mode is intended for:
- Use in dermatologic procedures for the removal of benign lesions of the skin and for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures.
Apollo Quattro is a combination stimulator used to relieve pain by applying radio frequency, low frequency, heating, and cooling energy to the affected area.
Apollo Quattro has a Q-EP handpiece, Q-LF handpiece, Q-FX handpiece, Q-SC handpiece, Q-BT handpiece, patch (cleared under K042301), patch cable, funnel hose, and power cable.
The Q-LF, Q-FX, Q-SC, and Q-BT handpieces are used for radio frequency or low frequency output, while the Q-EP handpiece is used for heating or cooling output, or for low frequency when used with the patch and patch cable.
The cooling function of Apollo Quattro is substantially equivalent to the device cleared in K213294. The Apollo Quattro is a cryosurgical device used to cool the skin, without the use of cryogenic gases or liquids, for the removal of benign skin lesions and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures. Surface contact cooling is achieved using a thermoelectric cooler (TEC), with an integrated aluminum plate, to lower the temperature of the skin.
N/A
Ask a specific question about this device
(379 days)
EMS is used for:
- Relaxation of muscle spasms
- Prevention or retardation of disuse atrophy
- Increasing local blood circulation
- Muscle re-education
- Maintaining or increasing range of motion
- Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
TENS is used for:
- Symptomatic relief and management of chronic, intractable pain
- Post-surgical acute pain
- Post-trauma acute pain
EVE Synergy has Synergy handpiece (1EA), Pro handpiece (2EA), Micro cable, Tip (Ball type, T type, COMB type) and Power adaptor & cable.
EVE Synergy has Synergy mode and Pro mode.
Synergy mode has radio frequency and low frequency functions using a synergy handpiece. The Synergy handpiece has a temperature sensor, so the output is automatically turned off when the skin surface temperature exceeds 43℃.
Pro mode has low frequency function that uses the Pro handpiece with tip.
It should be noted that the device has radio frequency (RF Mode) indicated for relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation. This RF mode was previously reviewed by the FDA under K212253 and was cleared under regulation 21 CFR 878.4400 (Product code PBX). In the subject submission, there are no changes being proposed related to the RF Mode, and, hence, the review of the RF Mode is not within the scope of the subject 510(k) submission.
The provided text is a 510(k) clearance letter and an accompanying 510(k) summary for the EVE Synergy (EVE-20M) device. This document focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than proving the device meets specific acceptance criteria through a comparative effectiveness study involving AI assistance or human-in-the-loop performance.
Therefore, many of the requested details regarding acceptance criteria for AI performance, specific study designs (MRMC, standalone), ground truth establishment for AI training/test sets, and sample sizes for these, are not present in this type of regulatory submission. This document primarily describes the physical and electrical characteristics of the device and compares them to predicate devices, along with standard electrical safety, EMC, and software validation testing.
Here's a breakdown of the requested information based on the provided text, highlighting what is present and what is not:
1. A table of acceptance criteria and the reported device performance
The document does not present "acceptance criteria" in the traditional sense of a specific performance metric (like accuracy, sensitivity, specificity) for a clinical outcome or an AI model's performance. Instead, it demonstrates compliance with safety and performance standards and similarity to predicate devices.
The "acceptance criteria" are implicitly met by showing:
- Compliance with various IEC standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 62304) for electrical safety, EMC, and software.
- The device's technical specifications (e.g., output voltage, current, pulse width, frequency, power density) are comparable to or within the safe limits established by the predicate devices.
Table of Performance (extracted from comparison tables):
| Characteristic | Acceptance Criteria (Implied by Predicate & Standards) | EVE Synergy (EVE-20M) Reported Performance |
|---|---|---|
| General Compliance | Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 14971, IEC 60601-1-6, IEC 62304 | Passed all testing in accordance with internal requirements, national standards, and international standards. |
| Device Configuration | Similar design, components, and method of current isolation to predicates. | Consists of AC/DC power supply, controller, LCD touch screen; Supports Synergy and Pro handpieces; Independent transformer isolation; Type BF Electrical Type. |
| Electrical Safety | Patient Leakage Current – Normal Condition: <100uA; Single Fault Condition: <300uA; Automatic Overload/No Load Trip; Automatic Shut Off. | Patient Leakage – Normal: <100uA; Single Fault: <300uA; Automatic Overload Trip: Yes; Automatic No Load Trip: Yes; Automatic Shut Off: Yes. |
| Biocompatibility | User-contacting materials (Stainless Steel 304, Polycarbonate) comparable to previously cleared devices. | Electrodes (SUS 304), LED Display part (PC) are the same as previously cleared K212253; Contact duration: Limited (< 24 hours). |
| EMS Mode Output | ||
| Waveform | Symmetrical Biphasic Waveform | Symmetrical Biphasic Waveform |
| Pulse Shape | Rectangular | Rectangular |
| Max Output Voltage | Comparable to predicates (e.g., Primary Predicate: 30V@500Ω, 54V@2kΩ, 54V@10kΩ; Secondary Predicate: 56V at all). Minor differences acceptable if safety demonstrated. | 34V@500Ω, 48V@2kΩ, 48V@10kΩ. "Similar to predicate devices". |
| Max Output Current | Comparable to predicates (e.g., Primary Predicate: 60mA@500Ω, 27mA@2kΩ, 5.4mA@10kΩ; Secondary Predicate: 92.86mA@500Ω, 26.7mA@2kΩ, 5.4mA@10kΩ). | 68mA@500Ω, 24mA@2kΩ, 4.8mA@10kΩ. "Similar to predicate devices". |
| Pulse Width | Within predicate range (20 to 400 μs). | 100 to 155 μs. "Within the range of the predicate devices". |
| Frequency | 3 to 200 Hz | 3 to 200 Hz |
| Net Charge @ 500 Ohms | 0 μC/pulse | 0 μC/pulse |
| Max Phase Charge | Within predicate range (e.g., Primary Predicate: 1.2 μC to 24 μC). | 10.54 μC@500Ω. "Within the range of the predicate device". |
| Max Current Density | ≤ 2 mA/cm² (IEC 60601-2-10). | 1.31 mA/cm² (Surface = 6 cm²). "Does not exceed the safety level 2 mA/cm²". |
| Max Power Density | Comparable to predicates (e.g., Primary Predicate: 22.2 mW/cm²; Secondary Predicate: 55 mW/cm²). | 47 mW/cm² @ 500Ω. "Similar to the predicate devices". |
| Burst Mode Parameters | Pulses/burst: 3-200; Bursts/sec: 1; Burst duration: 1-60 sec; Duty Cycle: Time on/off. | Pulses/burst: 3-200; Bursts/sec: 1; Burst duration: 1-60 sec; Duty Cycle: Time on/off. |
| On/Off Time | 1-60 seconds | 1-60 seconds |
| Treatment Time | Up to 60 minutes | Up to 60 minutes |
| TENS Mode Output | ||
| Waveform | Symmetrical Biphasic Waveform, Pulsed Monophasic (present in cleared predicate devices). | Symmetrical Biphasic Waveform, Pulsed Monophasic. "Does not raise new questions about safety and effectiveness." |
| Pulse Shape | Rectangular | Rectangular |
| Max Output Voltage | Comparable to predicates (e.g., Primary Predicate: 19V; Secondary Predicate: 36V). Minor differences acceptable if safety demonstrated. | 10V @500Ω, 10V @2kΩ, 10V @10kΩ. "Similar to those of the primary predicate device... does not raise new questions about safety and effectiveness." |
| Max Output Current | Comparable to predicates (e.g., Primary Predicate: 38mA@500Ω; Secondary Predicate: 67.8mA@500Ω). | 20mA@500Ω, 5.0mA@2kΩ, 1.0mA@10kΩ. (Note: Comparison comments for Voltage apply here as well). |
| Pulse Width | Within predicate range (20 to 400 μs). | 100 to 150 μs. "Within the range of the predicate devices". |
| Frequency | 3 to 200 Hz | 3 to 200 Hz |
| Net Charge @ 500 Ohms | 0 μC/pulse | 0 μC/pulse |
| Max Phase Charge | Within predicate range (e.g., Primary Predicate: 0.76 μC to 15.2 μC). | 3.0 μC@500Ω. "Within the range of the predicate device". |
| Max Current Density | ≤ 2 mA/cm² (IEC 60601-2-10). | 0.83 mA/cm² (Surface = 6 cm²). "Does not exceed the safety level 2 mA/cm²". |
| Max Power Density | Comparable to predicates (e.g., Primary Predicate: 8.9 mW/cm²; Secondary Predicate: 17.3 mW/cm²). | 8.3 mW/cm² @ 500Ω. "Similar to the primary predicate device". |
2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
This document does not describe a test set or data derived from human subjects. The testing described is non-clinical bench testing:
- Electrical safety tests
- Electromagnetic compatibility tests
- Performance tests (measuring output parameters against specifications)
- Software verification and validation
- Usability testing (presumably, for the device's user interface, not clinical user performance)
There is no mention of "country of origin for data" or "retrospective/prospective" as these are not relevant to bench testing for device parameters.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This is not applicable. "Ground truth" in the context of AI or clinical diagnostic studies (e.g., for radiologists interpreting images) is not established or referenced here because this is a physical medical device (muscle stimulator) undergoing bench testing for safety and performance specifications, not an AI diagnostic tool.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. There are no clinical or image-based test sets requiring adjudication by experts.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a muscle stimulator, not an AI-assisted diagnostic tool. No MRMC study was performed or required for this type of submission.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. There is no AI algorithm being cleared as a standalone device. The "software validation testing" mentioned (IEC 62304) relates to the safe and effective operation of the device's inherent control software, not an AI-driven clinical function.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable in the context of AI or clinical ground truth. For this submission, "ground truth" implicitly refers to the engineering specifications and established safety limits as defined by relevant IEC standards and comparison with predicate devices. The device's output parameters were measured against these defined standards and specifications.
8. The sample size for the training set
Not applicable. This is a medical device, not an AI model. There is no training set for an AI algorithm mentioned or required.
9. How the ground truth for the training set was established
Not applicable. As above, there is no AI training set.
Ask a specific question about this device
(182 days)
EMS is used for:
- · Relaxation of muscle spasms
- · Prevention or retardation of disuse atrophy
- · Increasing local blood circulation
- · Muscle re-education
- · Maintaining or increasing range of motion
- · Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
eMVFit has MULTI handpiece, EMS handpiece, patient switch, cradle, distilled water injection hose & funnel, key switch and power cable.
eMVFit has a user interface including an LCD touch screen and has MULTI mode and EMS mode. MULTI mode has RF and EMS functions that use the MULTI handpiece.
In the case of RF, the target temperature can be set and a vacuum can be used together, and a cooling function can be used as a safety system to prevent burns on the skin surface.
In addition, the MULTI handpiece has a temperature sensor, so the output is automatically turned off when the skin surface temperature exceeds 43 ℃.
EMS mode has an EMS function that uses the EMS handpiece.
The provided text is a 510(k) Summary for the eMVFit (MVF-10M) device, a powered muscle stimulator. It focuses on demonstrating substantial equivalence to predicate devices through various comparisons and non-clinical performance data.
However, the document does not contain information on acceptance criteria for a study demonstrating device performance in terms of clinical outcomes, nor does it describe a study involving human subjects or expert readers to prove the device meets specific acceptance criteria for diagnostic or clinical effectiveness.
The document primarily details:
- Device Description and Indications for Use: What the device is and what it's used for (muscle spasms, disuse atrophy, circulation, re-education, range of motion, DVT prevention).
- Predicate Device Comparison: A detailed table comparing the eMVFit's technical specifications (electrical parameters, dimensions, operation modes) with three predicate devices (INMODE LTD.'s Evolve System with T3 and Tone Applicators, and TimeWaver Production GmbH's TimeWaver Frequency).
- Substantial Equivalence Discussion: Explains that differences between the subject device and predicates do not raise new questions of safety or effectiveness, often citing compliance with IEC standards.
- Non-Clinical (Bench) Performance Data: Lists a series of standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 14971, IEC 60601-1-6, ISO 10993-10, ISO 10993-23, IEC 62304) to which the device was tested to demonstrate basic safety, essential performance, EMC, risk management, usability, biocompatibility, and software functions. It states the device passed all testing.
Therefore, I cannot fulfill your request for: a table of acceptance criteria and reported device performance (in the context of clinical outcomes), sample size, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, or ground truth establishment based on the provided text.
The document describes technical compliance and safety testing against engineering standards, not direct clinical performance or effectiveness studies with human data or expert evaluations. The "performance data" referred to is "Non-Clinical (Bench) Performance Data," meaning lab tests, not clinical trials.
The FDA's 510(k) pathway often relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than requiring new clinical efficacy studies, especially for devices with well-understood mechanisms of action like muscle stimulators. The information provided aligns with this regulatory pathway.
Ask a specific question about this device
(181 days)
The eMVFit (MVF-10M) Radio frequency is used for:
· Relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation.
· Temporary reduction in the appearance of cellulite.
eMVFit has MULTI handpiece, EMS handpiece, patient switch, cradle, distilled water injection hose & funnel, key switch and power cable.
eMVFit has a user interface including an LCD touch screen and has MULTI mode and EMS mode.
MULTI mode has RF and EMS functions that use the MULTI handpiece.
In the case of RF, the target temperature can be set and a vacuum can be used together, and a cooling function can be used as a safety system to prevent burns on the skin surface.
In addition, the MULTI handpiece has a temperature sensor, so the output is automatically turned off when the skin surface temperature exceeds 43°C.
EMS mode has an EMS function that uses the EMS handpiece.
The provided text describes the eMVFit (MVF-10M) device and its 510(k) summary for substantial equivalence to a predicate device (INMODE LTD. Embody System, K183450). This document primarily focuses on bench testing and compliance with standards to demonstrate safety and effectiveness for a medical device rather than a study with a human test set, expert readers, or ground truth in the context of clinical performance or diagnostic accuracy.
Therefore, many of the requested details, such as sample size, data provenance, number of experts, adjudication methods, MRMC studies, and ground truth types related to human performance or clinical outcomes, are not applicable in this regulatory submission. The acceptance criteria and "proof" in this context are for demonstrating the device's technical specifications, safety, and electromagnetic compatibility through non-clinical testing.
Here's a breakdown of the information that is available or NA (Not Applicable) based on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally established by compliance with international standards for medical electrical equipment. The "reported device performance" is that the device passed all these tests.
| Acceptance Criterion (Standard Compliance) | Reported Device Performance |
|---|---|
| Basic Safety and Essential Performance (IEC 60601-1:2005/A2:2020) | Passed all testing. |
| Electromagnetic Disturbances (IEC 60601-1-2:2014/A1:2020) | Passed all testing. |
| High Frequency Surgical Equipment (IEC 60601-2-2:2017/A1:2023) | Passed all testing. |
| Risk Management (ISO 14971:2019) | Documented by referring to the standard. |
| Usability (IEC 60601-1-6:2010/A2:2020) | Documented by referring to the standard. |
| Biocompatibility (ISO 10993-5:2009, ISO 10993-10:2021, ISO 10993-23:2021) | Passed testing; materials confirmed biocompatible. |
| Software Verification and Validation (IEC 62304:2015) | Tested and evaluated according to the standard. |
| Performance and Electrical Safety (Internal and FDA Guidance) | Met all requirements; output parameters measured. |
Study Proving Device Meets Acceptance Criteria
As noted, this is a non-clinical, bench testing-focused submission, not a human-in-the-loop or clinical study.
-
A table of acceptance criteria and the reported device performance:
(See table above) -
Sample size used for the test set and the data provenance:
- Sample Size: Not applicable. The "test set" here refers to the device and its components undergoing bench testing, not a dataset of patient cases.
- Data Provenance: Not applicable in terms the requested format. The data comes from internal laboratory testing conducted by the manufacturer (WEERO Co.,Ltd.). This is reported as non-clinical (bench) performance data.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable. Ground truth, in the sense of expert consensus on clinical findings, is not established for this type of non-clinical, hardware/software compliance testing. The "ground truth" for these tests is the established requirements and limits defined by the international standards themselves.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not Applicable. Adjudication methods are relevant for subjective interpretations of clinical data by multiple readers. This refers to objective measurements and compliance against specific technical standards.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This device is an electrosurgical cutting and coagulation device and accessories, intended for therapeutic use (e.g., muscle aches, cellulite), not a diagnostic device involving human readers or AI assistance in image interpretation. Therefore, an MRMC study is not relevant or performed for this submission.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is a physical device, not an AI algorithm for diagnostic interpretation that would have "standalone" performance metrics in the way described. The device's performance is assessed through its physical outputs and safety mechanisms.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Regulatory Standards and Design Specifications. The "ground truth" for the non-clinical testing is adherence to the specified technical parameters, safety limits, and performance requirements outlined in the referenced international standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2) and internal design specifications. Biocompatibility used data from ISO 10993 testing.
-
The sample size for the training set:
- Not Applicable. This device submission does not describe a machine learning model that requires a "training set" in the context of supervised learning for diagnostic accuracy. The "training" for this device would be its own design and manufacturing process to meet performance specifications.
-
How the ground truth for the training set was established:
- Not Applicable. (See point 8).
Ask a specific question about this device
(402 days)
Apollo duet(RF Mode) is intended for relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local blood circulation.
Apollo Duet has RF handpiece attached to the main body, and consists of an electrode cable and a power adapter. This equipment has RF mode, which is the main function, RF mode is a mode in which Radio-frequency is output in a bipolar method. An RF handpiece is used, and a high frequency is output through 4ea RF handpiece electrodes. In addition, the RF handpiece has a temperature sensor, so the output is automatically turned off when the skin surface temperature exceeds 45 C
The provided text describes the regulatory clearance of a medical device called "Apollo Duet" (Model: APD-4000) by the FDA. It does not contain information about a study that proves the device meets specific acceptance criteria related to its performance in terms of effectiveness for its intended use.
The document focuses on demonstrating substantial equivalence to a predicate device (InMode RF Pro System) based on technical, safety, and operational characteristics, rather than a clinical effectiveness study with defined acceptance criteria for efficacy.
Therefore, I cannot provide a table of acceptance criteria and reported device performance from this document. The information requested regarding sample sizes, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth for training sets is not present in the provided text.
Here's what can be extracted regarding testing and related safety:
Acceptance Criteria and Study Information (Based on provided text)
The document outlines safety and performance bench testing and adherence to international standards to demonstrate substantial equivalence and safety, rather than effectiveness acceptance criteria for its therapeutic claims (relief of muscle aches, spasm, and improved blood circulation).
1. A table of acceptance criteria and the reported device performance:
| Acceptance Criteria (Implied Safety/Technical) | Reported Device Performance (Compliance) |
|---|---|
| Electrical Safety (IEC 60601-1:2005, AMD1:2012) | Complies with standard |
| Electromagnetic Compatibility (IEC 60601-1-2:2014) | Complies with standard |
| High Frequency Surgical Equipment Safety (IEC 60601-2-2:2017) | Complies with standard |
| Nerve & Muscle Stimulators Safety (IEC 60601-2-10:2012/A1:2016) | Complies with standard |
| Risk Management (ISO 14971:2019) | Risk management recorded |
| Usability (IEC 60601-1-6:2010 + A1:2013) | Usability documented |
| Biocompatibility (ISO 10993-1) (Electrode: SUS 304, LED Display: PC (SR3108FM)) | Materials are biocompatible for intact skin, limited contact (<24 hours) |
| RF Ablation Thermal Effects (Ex-vivo testing) | Performed to compare thermal effects with predicate device |
| Skin Temperature Measurement | Measured with thermographic camera; output automatically turns off if skin temperature exceeds 45°C |
2. Sample size used for the test set and the data provenance:
- RF Ablation Thermal Effects (Ex-vivo testing): Not specified (type of tissue: porcine abdominal skin).
- Skin temperature measurement: 6 healthy volunteers (3 light skin, 3 dark skin). Data provenance: Not specified (implied to be from the manufacturer, WEERO Co., Republic of Korea). This appears to be a prospective safety test.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable/Not specified in the document. The tests seem to be objective measurements (e.g., temperature, compliance with standards) rather than expert-derived ground truth for performance.
4. Adjudication method for the test set:
Not applicable/Not specified.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC study was done, as this is not an AI-assisted diagnostic device, but an electrosurgical cutting and coagulation device and accessories.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not directly applicable in the context of typical AI standalone performance. The device itself is a standalone medical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For safety/technical testing, the "ground truth" is compliance with established international standards (e.g., IEC 60601 series, ISO 14971) and objective measurements (e.g., temperature thresholds). For ex-vivo testing, direct measurement of thermal effects on tissue.
8. The sample size for the training set:
Not applicable. This device does not appear to involve machine learning models that require a training set.
9. How the ground truth for the training set was established:
Not applicable.