K Number
K213294

Validate with FDA (Live)

Date Cleared
2022-11-10

(405 days)

Product Code
Regulation Number
878.4340
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dermal Cooling System is a cryosurgical instrument intended for use in dermatologic procedures for the removal of benign lesions of the skin and for use when cooling is intended for the temporary reduction of pain, swelling. inflammation, and hematoma from minor surgical procedures.

The Dermal Cooling System is further indicated to minimize pain, inflammation, and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief of injections.

When the Dermal Cooling System is used with the R2 Dermabrasion Tips, the intended use includes general dermabrasion, scar revision, acne scar revision, and tattoo removal.

The Dermal Cooling System is intended to be used by trained healthcare professionals.

Device Description

The Dermal Cooling System is a cryosurgical device used to cool the skin, without the use of cryogenic gases or liquids, for the removal of benign skin lesions and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures. The Dermal Cooling System is further indicated to minimize pain, inflammation, and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief of injections. Surface contact cooling is achieved using a thermoelectric cooler (TEC), with an integrated aluminum plate, to lower the temperature of the skin. For dermabrasion, the Dermal Cooling System handpiece is intended to serve as a handle to facilitate manual movement of the R2 Dermabrasion Tips.

The R2 Dermabrasion Tips are optional accessories for the Dermal Cooling System that may be attached to the distal end of the handpiece to facilitate manual dermabrasion. The R2 Dermabrasion Tips may be used with a commercially available topical or water to facilitate movement of the handpiece across the treatment area.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Dermal Cooling System:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
System Performance TestSystem is able to hold the internal coldplate temperature to -16.0°C ± 0.5°C for 20 minutesPassed

Detailed Information about the Study:

2. Sample Size Used for the Test Set and Data Provenance:

The document states, "No other performance testing was performed for the subject device for this Traditional 510(k)..." and "No preclinical or clinical testing was performed." The only reported test is a "System Performance Test" described as "Bench testing." This implies the "test set" for this specific performance criteria was likely a single or a small number of physical devices tested in a lab setting, rather than a clinical dataset. No information on data provenance (country of origin, retrospective/prospective) is applicable as it's not a clinical study involving patient data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

Not applicable. The reported test is a bench test, and the "ground truth" is established by the device's ability to maintain a specific coldplate temperature, not by expert interpretation of clinical data.

4. Adjudication Method for the Test Set:

Not applicable, as it's a bench test against an objective temperature criterion.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study was not done. The document explicitly states, "No preclinical or clinical testing was performed."

6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

This question is not applicable to the Dermal Cooling System. This device is a physical cryosurgical instrument, not an AI algorithm. Therefore, "standalone (algorithm only without human-in-the-loop performance)" is not relevant.

7. The Type of Ground Truth Used:

The ground truth for the "System Performance Test" was the pre-defined target temperature range for the internal coldplate (-16.0°C ± 0.5°C) maintained over a specific duration (20 minutes). This is a physical performance specification, not expert consensus, pathology, or outcomes data.

8. The Sample Size for the Training Set:

Not applicable. There is no mention of a training set as this is not an AI/ML device.

9. How the Ground Truth for the Training Set was Established:

Not applicable, as there is no training set mentioned or implied.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

R2 Technologies, Inc. Tarhan Kayihan Director of QA/RA 2603 Camino Ramon, Suite 200 San Ramon, California 94583

November 10, 2022

Re: K213294

Trade/Device Name: Dermal Cooling System Regulation Number: 21 CFR 878.4340 Regulation Name: Contact cooling system for aesthetic use Regulatory Class: Class II Product Code: QPZ, ILO, GED Dated: October 12, 2022 Received: October 14, 2022

Dear Tarhan Kayihan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Colin K. Chen -S

for

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213294

Device Name Dermal Cooling System

Indications for Use (Describe)

The Dermal Cooling System is a cryosurgical instrument intended for use in dermatologic procedures for the removal of benign lesions of the skin and for use when cooling is intended for the temporary reduction of pain, swelling. inflammation, and hematoma from minor surgical procedures.

The Dermal Cooling System is further indicated to minimize pain, inflammation, and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief of injections.

When the Dermal Cooling System is used with the R2 Dermabrasion Tips, the intended use includes general dermabrasion, scar revision, acne scar revision, and tattoo removal.

The Dermal Cooling System is intended to be used by trained healthcare professionals.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K213294

510(k) Summary

I. SUBMITTERR2 Technologies, Inc.2603 Camino Ramon, Suite 200Bishop Ranch 3San Ramon, CA 94583
Contact:Mr. Tarhan KayihanDirector of QA/RAR2 Technologies, Inc.Phone: 925-378-4400tkayihan@r2technologies.com
Date Prepared:October 12, 2022

II. DEVICE and ACCESSORY

Device Name:Dermal Cooling System
Device Common Name:Contact cooling for skin lesion pain relief
Accessory Name:R2 Dermabrasion Tips
Accessory Common Name:Brush, Dermabrasion, Manual
Regulatory Class:II
Product Code:QPZ, GED, ILO
Regulation:21 CFR 878.4340, 21 CFR 878.4800, 21 CFR 890.5720

III. PREDICATE

The primary predicate device is the Dermal Cooling System, K203006. The secondary predicate device is the COOLSKIN device, K083008.

IV. DEVICE DESCRIPTION

The Dermal Cooling System is identical to the device cleared in K203006. The Dermal Cooling System is a cryosurgical device used to cool the skin, without the use of cryogenic gases or liquids, for the removal of benign skin lesions and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures. The Dermal Cooling System is further indicated to minimize pain, inflammation, and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief of injections. Surface contact cooling is achieved using a thermoelectric cooler (TEC), with an integrated aluminum plate, to lower the temperature of the skin. For dermabrasion, the Dermal Cooling System handpiece is intended to serve as a handle to facilitate manual movement of the R2 Dermabrasion Tips.

The R2 Dermabrasion Tips are optional accessories for the Dermal Cooling System that may be attached to the distal end of the handpiece to facilitate manual dermabrasion. The R2 Dermabrasion Tips may be used with a commercially available topical or water to facilitate movement of the handpiece across the treatment area.

{4}------------------------------------------------

V. INDICATIONS FOR USE

The Dermal Cooling System is a cryosurgical instrument intended for use in dermatologic procedures for the removal of benign lesions of the skin and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures.

The Dermal Cooling System is further indicated to minimize pain, inflammation, and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief of injections.

When the Dermal Cooling System is used with the R2 Dermabrasion Tips, the intended use includes general dermabrasion, scar revision, acne scar revision, and tattoo removal.

The Dermal Cooling System is intended to be used by trained healthcare professionals.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE

The Dermal Cooling System ("subject device") and the predicate devices are Class II, thermoelectric contact cooling devices used to cool tissue during dermatologic procedures.

The primary predicate, the Dermal Cooling System, has not been modified for the proposed expanded indication, i.e., minimize pain, inflammation, and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief of injections.

The secondary predicate, the COOLSKIN device, is included to substantiate the proposed expanded indications for use as the cleared indication for the COOLSKIN device (K083008) includes "minimize pain and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief of injections." The Dermal Cooling System and the COOLSKIN device have these same fundamental scientific technological elements:

  • comprised of a cooling unit (chiller) with a cold plate/probe applicator used to cool the skin in the area intended for treatment
  • same cooling mechanism - thermoelectric cooling
  • . controlled contact surface cooling at a targeted treatment site on the skin
  • . metal cooling interface within an applicator

The key differences that exist between the Dermal Cooling System and the COOLSKIN device are:

  • size and shape of cold plate/probe applicator
    The differences do not raise new questions of safety and effectiveness with respect to the proposed expanded indication.

VII. PERFORMANCE DATA

Performance data is provided in support of the substantial equivalence determination, as summarized in the table below.

Bench testing was completed to demonstrate the ability of the Dermal Cooling System to meet performance specifications.

{5}------------------------------------------------

TestTest Method/RequirementAcceptance CriteriaResults
SystemVerificationSystem Performance Test:System performance atlowest expected clinicaltreatment temperature, atmaximum treatment timeSystem is able to holdthe internal coldplatetemperatureto -16.0°C ± 0.5°Cfor 20 minutesPassed

No other performance testing was performed for the subject device for this Traditional 510(k) since there was no change to the device design or methods for use required for the proposed expanded indication and the range of treatment parameters previously cleared for the Dermal Cooling System (K203006) allows the same cooling characteristics at the secondary predicate, COOLSKIN device.

No preclinical or clinical testing was performed.

VIII. CONCLUSIONS

The Dermal Cooling System that is the subject of this Traditional 510(k) is substantially equivalent to both the primary predicate (Dermal Cooling System) and the secondary predicate (COOLSKIN device). The subject device and both predicates are Class II devices intended for use to cool tissue in dermatologic procedures of the subject Dermal Cooling System and the primary predicate Dermal Cooling System are the same device; there are no changes required in the device design or the method by which the device is used for this expanded indication. The Dermal Cooling System can deliver the same treatment parameters using the same cooling mechanism (i.e., TECs) as the secondary predicate COOLSKIN device. The minor technological differences between the subject device and the secondary predicate are the size and shape of the cold plate/probe applicator; however, these differences do not raise new questions of safety and effectiveness for the proposed expanded indication.

§ 878.4340 Contact cooling system for aesthetic use.

(a)
Identification. A contact cooling system for aesthetic use is a device that is a combination of a cooling pad associated with a vacuum or mechanical massager intended for the disruption of adipocyte cells intended for non-invasive aesthetic use.(b)
Classification. Class II (special controls). The special controls for this device is FDA's “Guidance for Industry and FDA Staff; Class II Special Controls Guidance Document: Contact Cooling System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.