Found 3 results

510(k) Data Aggregation

    K Number
    K242536

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-10-15

    (50 days)

    Product Code
    Regulation Number
    884.1730
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TauTona Pneumoperitoneum Assist Device (TPAD) is intended for use with the Veress Needle for the establishment of a pneumoperitoneum during laparoscopic procedures.

    Device Description

    The TauTona Pneumoperitoneum Assist Device (TPAD) is a single use device used during laparoscopic surgical procedures. The device consists of a pad which contains an array of suction cups on one side; a port that attaches to a standard hospital vacuum line; a button to control vacuum to the suction cups; and a handle to allow user to manipulate the device. Suction is applied to the device to allow the user to manipulate (hold and pull on) the tissue around the Veress needle insertion site. The device is crescent in shape to allow removal of the device after use while the Veress needle remains in the patient.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria for a specific device. Instead, it is a 510(k) summary for the TauTona Pneumoperitoneum Assist Device (TPAD), which focuses on demonstrating substantial equivalence to a predicate device.

    The document highlights the clinical study performed to support this substantial equivalence. Below is an interpretation of the study and what can be inferred about "acceptance criteria" through the lens of equivalence:

    Interpretation of Acceptance Criteria and Study Performance

    The acceptance criteria, while not explicitly stated as numerical targets for performance metrics like sensitivity, specificity, or specific time thresholds, are implicitly based on demonstrating that the subject device is as safe and effective as the predicate device and does not introduce new questions of safety or effectiveness. This is a common approach for 510(k) submissions.

    Implicit Acceptance Criteria:

    • Non-inferiority in surgeon satisfaction: The average surgeon questionnaire scoring for TPAD should not be significantly worse than for Standard of Care (SOC).
    • Non-inferiority in patient satisfaction: The average patient questionnaire scoring for TPAD should not be significantly worse than for SOC.
    • Non-inferiority in procedural timing: The average time from incision to Veress needle insertion and from incision to insufflation for TPAD should not be significantly different from SOC.
    • Safety: No Severe or Moderate Adverse Events related to TPAD.

    Reported Device Performance (against implicit criteria):

    • Surgeon Satisfaction: "The primary endpoint of the average surgeon questionnaire scoring was not significantly different for SOC and TPAD across all responses."
    • Patient Satisfaction: "The average patient questionnaire scoring was not significantly different for SOC and TPAD across all responses."
    • Procedural Timing: "The average timing measurements from the incision to Veress Needle Insertion, as well as time from incision to insufflation, were not significantly different between SOC and TPAD for both timing durations."
    • Safety: "No Severe or Moderate Adverse Events related to the TPAD were observed."

    Study Details:

    Given the nature of the device (a physical assist device, not an AI/software device or diagnostic imaging tool), many of the specific questions related to AI performance metrics (like MRMC studies, standalone algorithm performance, number of experts for ground truth, adjudication methods) are not applicable or detailed in this 510(k) summary.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Criteria Type (Implicit)Acceptance Criteria (Demonstrate Non-inferiority to SOC)Reported Device Performance (vs. SOC)
    Performance - Surgeon SatisfactionSurgeon questionnaire scores for TPAD should not be significantly different from SOC."The primary endpoint of the average surgeon questionnaire scoring was not significantly different for SOC and TPAD across all responses."
    Performance - Patient SatisfactionPatient questionnaire scores for TPAD should not be significantly different from SOC."The average patient questionnaire scoring was not significantly different for SOC and TPAD across all responses."
    Performance - Procedural Timing (Incision to VN Insertion)Time from incision to Veress Needle Insertion for TPAD should not be significantly different from SOC."The average timing measurements from the incision to Veress Needle Insertion... were not significantly different between SOC and TPAD for both timing durations."
    Performance - Procedural Timing (Incision to Insufflation)Time from incision to insufflation for TPAD should not be significantly different from SOC."The average timing measurements... from incision to insufflation, were not significantly different between SOC and TPAD for both timing durations."
    SafetyNo Severe or Moderate Adverse Events related to TPAD should be observed."No Severe or Moderate Adverse Events related to the TPAD were observed."
    General EquivalenceDevice is as safe and effective as the predicate device across various metrics (satisfaction, time, safety, skin impacts)."Overall, across this wide range of metrics, the use of TPAD for Veress needle insertion and insufflation was not significantly different than the use of the standard of care for Veress needle insertion and insufflation. These clinical performance data therefore support that the subject device is substantially equivalent to the predicate."

    2. Sample Size and Data Provenance:

    • Test Set (Clinical Study) Sample Size: The clinical study was done in two phases, totaling 30 patients.
      • Phase 1: 20 patients
      • Phase 2: 10 patients
    • Data Provenance: The document does not explicitly state the country of origin. It is a clinical study (prospective), comparing the subject device (TPAD) to Standard of Care (SOC).

    3. Number of Experts and Qualifications for Ground Truth:

    • Not Applicable in the traditional sense for this type of device. The study assessed "surgeon satisfaction" and "patient satisfaction" via questionnaires and "timing measurements." These are direct measures from the users and procedures, not "ground truth" established by experts in the sense of image interpretation or diagnosis. The "experts" involved would be the surgeons using the device.

    4. Adjudication Method for the Test Set:

    • Not applicable. This study is not about subjective interpretation or diagnosis that would require adjudication. It's about direct measurements (time) and subjective feedback via questionnaires (satisfaction).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. An MRMC study is typically for evaluating diagnostic imaging devices where multiple readers interpret cases with and without AI assistance. This device is a physical assist device for a surgical procedure, not an imaging or diagnostic AI tool.

    6. Standalone (Algorithm Only) Performance:

    • Not applicable. The TPAD is a physical assist device, not an algorithm or software that operates in a standalone manner. The "performance" is inherently tied to its use by a human.

    7. Type of Ground Truth Used:

    • The "ground truth" for this study was primarily:
      • Direct procedural time measurements: Objective data like "time from incision to Veress Needle Insertion."
      • User (Surgeon) Reported Experience: Subjective feedback from surgeon questionnaires.
      • Patient Reported Experience: Subjective feedback from patient questionnaires.
      • Observed clinical outcomes: Lack of severe/moderate adverse events.

    8. Sample Size for the Training Set:

    • Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable. As above, no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233020

    Validate with FDA (Live)

    Date Cleared
    2024-01-12

    (112 days)

    Product Code
    Regulation Number
    884.1730
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TauTona Pneumoperitoneum Assist Device (TPAD) is intended for use in the upper left quadrant of the abdominal wall (i.e., Palmer's Point) with a Veress needle for the establishment of a pneumoperitoneum during laparoscopic procedures.

    Device Description

    The TauTona Pneumoperitoneum Assist Device (TPAD) is a single use device used during laparoscopic surgical procedures. The device consists of a pad which contains an array of suction cups on one side; a port that attaches to a standard hospital vacuum line; a button to control vacuum to the suction cups; and a handle to allow user to manipulate the device. Suction is applied to the device to allow the user to manipulate (hold and pull on) the tissue around the Veress needle insertion site. The device is crescent in shape to allow removal of the device after use while the Veress needle remains in the patient.

    AI/ML Overview

    The provided text is a 510(k) summary for the TauTona Pneumoperitoneum Assist Device (TPAD). It outlines the device's purpose, comparison to a predicate device, and the non-clinical and clinical performance testing performed.

    However, the 510(k) summary does not include a formal table of acceptance criteria for the device performance or detailed information about the specific "study that proves the device meets the acceptance criteria" in the format requested. It describes a clinical study to evaluate safety, effectiveness, and ease of use, but doesn't explicitly link specific metrics to predefined acceptance thresholds.

    Given the information available, I will extract and infer as much as possible to address your request, acknowledging where information is not explicitly stated.

    Here's an attempt to answer your questions based on the provided text:


    The provided 510(k) summary describes a clinical study to demonstrate the safety, effectiveness, and ease of use of the TauTona Pneumoperitoneum Assist Device (TPAD). While formal, quantitative "acceptance criteria" for metrics are not explicitly presented in a table format with corresponding reported performance, the summary describes the outcomes of the study.

    1. Table of Acceptance Criteria and Reported Device Performance

    As explicit acceptance criteria with numerical thresholds are not provided in the document, this table is constructed based on the described study outcomes and implied "acceptance" as demonstrated by the findings.

    Metric/Performance AreaAcceptance Criteria (Inferred from Study Design & Outcomes)Reported Device Performance
    Feasibility/Ease of Use (Primary Endpoint)Surgeon questionnaire median score not significantly different from a theoretical median of 4.0, indicating acceptable feasibility/ease of use.TPAD: Median score of 4.0 for all responses on surgeon questionnaire. Not significantly different from a theoretical median of 4.0 (Wilcoxon test).
    Time to Obtain Laparoscopic AccessNo statistically significant difference compared to Standard of Care (SOC).No statistically significant difference compared to SOC (unpaired t-test).
    Time to Complete InsufflationNo statistically significant difference compared to SOC.No statistically significant difference compared to SOC (unpaired t-test).
    Adverse Events (Safety)No device-related serious adverse events (SAEs). All other adverse events deemed mild and completely resolving. Bruising resolution comparable or better than SOC.No device-related SAEs. All other adverse events (bruising) deemed mild and related to insertion site/skin handling, with complete resolution. 6/10 TPAD subjects' bruising resolved within 3 days, 1/10 at 19 days. SOC (control) had 5/10 subjects with bruising still present at 7+ days.
    Patient Satisfaction (Skin Appearance/Bruising)Comparable or better patient satisfaction regarding skin appearance/bruising compared to SOC.Patient surveys and photographs were collected. Specific aggregate results for patient satisfaction are not detailed, but the bruising resolution data suggests a positive outcome for TPAD subjects compared to SOC.
    Veress Needle Placement SuccessAll surgeons able to place the Veress needle using the TPAD at the 5-20 mmHg pressure.All surgeons were able to place the Veress needle using the TPAD at the 5-20 mmHg pressure.

    Note: The "Acceptance Criteria" provided above are inferred from the study's discussion of "primary endpoint," "secondary endpoints," and "statistical significance" compared to standard of care or theoretical medians, rather than explicit pre-defined numerical thresholds in an acceptance criteria table.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Study): 20 individuals total, split into two arms:
      • TPAD arm: 10 patients
      • Standard of Care (SOC) arm: 10 patients
    • Data Provenance:
      • Country of Origin: United States (U.S.). The study was performed at an academic center in the U.S.
      • Retrospective or Prospective: Prospective (described as a "randomized clinical study").

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the "number of experts used to establish the ground truth" in the sense of independent expert review of cases for accuracy. Instead, the "ground truth" or device performance was assessed by:

    • Surgeons: who performed the procedures and completed questionnaires on device feasibility/ease of use. Their specific qualifications (e.g., years of experience, specialty) are not detailed beyond "surgeon."
    • Patients: who completed surveys to assess patient satisfaction.
    • Study Staff: who likely monitored adverse events, collected time metrics, and captured photographs.

    Therefore, the "experts" are the performing surgeons and the patients themselves providing direct feedback on their experience.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set in the context of expert review (e.g., 2+1, 3+1). The primary endpoint (surgeon questionnaire) was completed by the performing surgeon, not an independent panel, and secondary outcomes were objective measurements (time, adverse events) or patient self-reported.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No, an MRMC comparative effectiveness study was not performed. This device is not an AI-based diagnostic or assistive software for image interpretation, but rather a physical surgical assist device. Therefore, the concept of "human readers improving with AI assistance" is not applicable here. The clinical study compared the use of the TPAD device (a physical assist device) against the Standard of Care.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The device is a physical surgical assist device and does not involve AI algorithms or standalone performance. It is used with a human surgeon.

    7. The Type of Ground Truth Used

    The "ground truth" for the clinical study was established through:

    • Direct observation and measurement: Time metrics (time to obtain laparoscopic access, time to complete insufflation).
    • Surgeon feedback: Questionnaires on feasibility/ease of use, reflecting the practical aspects of the device during actual use.
    • Patient outcomes: Adverse event monitoring (bruising resolution) and patient reported satisfaction via surveys and photographs.

    This can be categorized as a combination of objective clinical measurements, direct user (surgeon) assessment, and patient-reported outcomes. It is not based on expert consensus of reviewed cases, pathology, or large-scale outcomes data in the traditional sense of diagnostic accuracy studies.

    8. The Sample Size for the Training Set

    This information is not applicable as the device is a physical medical device and contains no software/AI that would require a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as the device is a physical medical device and contains no software/AI that would require a "training set."

    Ask a Question

    Ask a specific question about this device

    K Number
    K131963

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-10-22

    (117 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TauTona Injector Device is a syringe assist device and is intended for use in the administration of sterile materials / autologous adipose tissue under aseptic conditions in accordance with the best judgment of the clinician. The TauTona Injector Device is intended for use with standard syringes / cannulas.

    Device Description

    The TauTona Injector (TTI) Device is a hand-held, sterile, single-use, disposable device and is designed to be used with, off-the-shelf syringes and injection cannulas. The device is used to assist the physician in injection fluids into the body. The clinician connects the TTI device to the pre-filled syringe and a standard re-injection cannula (Size 9, 14G). The device is activated by a trigger mechanism that enables fluid delivery in specific aliquot sizes. The TTI Device is a simple handle that consists of the following components:

    • . Pump with an enclosed motor
    • Battery pack (12V) .
    • . Controller (printed circuit board with software)
    • . Dispensing trigger
    • . On/off/aliquot size select switch
    • . Indicator light
    • . Syringe luer connector (female)
    • Cannula luer connector (male) .
    AI/ML Overview

    The provided text describes the TauTona Injector Device, a syringe assist device. However, it does not contain the specific details regarding acceptance criteria for device performance or a detailed study proving the device meets these criteria in the format requested.

    The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed performance study results against predefined acceptance criteria. While it mentions "Design verification testing" and "comparative testing," it does not present the outcomes of these tests in a quantifiable manner or against specific thresholds.

    Therefore, most of the requested information cannot be extracted directly from the provided text.

    Here's an attempt to answer the questions based on the available information, with clear indications where information is missing:


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (If stated)Reported Device Performance
    Not explicitly stated in the document for specific performance metrics. The overall acceptance criterion is likely "substantially equivalent" to predicate devices for intended use, operating principles, and safety/effectiveness profile."Design verification testing confirmed that the TauTona Injector Device performs according to the stated intended use." "Test data included in this 510(k) submission demonstrate similar performance of the TauTona Injector device as compared to the predicate devices." (No quantitative performance metrics or specific thresholds are provided.)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified.
    • Data Provenance: Not specified. The document refers to "Design verification testing," "comparative testing with the cited predicate devices," and "human factors testing," but no details on the origin or nature (retrospective/prospective) of the data are given.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts: Not applicable. This type of device (piston syringe/syringe assist device) typically relies on engineering and functional testing rather than expert-established ground truth for image interpretation or diagnosis.
    • Qualifications of Experts: Not applicable.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable to the type of testing described (functional, comparative, human factors).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret medical images, often with or without AI assistance. The TauTona Injector Device is a mechanical assist device, not an imaging or diagnostic AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable for this device. The TauTona Injector Device is a mechanical "human-in-the-loop" device by its very nature, designed to assist a physician in injecting fluids. It does not have an "algorithm-only" or "standalone" mode in the context of AI performance. The software validation mentioned refers to the internal control software, not a diagnostic algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Type of Ground Truth: Not explicitly stated in terms of expert consensus, pathology, or outcomes data, as these are typically associated with diagnostic devices. For this device, "ground truth" would likely be defined by engineering specifications and functional requirements (e.g., precise aliquot delivery, consistent force, biocompatibility compliance, electrical safety standards). These are verified through various tests like "functional testing," "biocompatibility testing," "sterilization validation," "electrical safety and electromagnetic compatibility testing," and "software validation."

    8. The sample size for the training set

    • Sample Size for Training Set: Not applicable. This device is not an AI/machine learning model that typically requires a "training set" for diagnostic performance.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable, as there is no mention of a training set for an AI/machine learning model.

    Summary of what the document does indicate about performance and testing:

    • Testing Types Mentioned:
      • Functional testing
      • Comparative testing with predicate devices
      • Human factors testing
      • Biocompatibility testing (ISO 10993)
      • Sterilization validation (ISO 11135)
      • Electrical safety and electromagnetic compatibility testing (IEC 60601)
      • Software validation (per FDA Guidance Document)
      • Packaging and shipping validation (ISTA 2A guidelines)
    • Conclusion: The device performs according to its intended use and demonstrates "similar performance" to predicate devices, thus establishing substantial equivalence without raising new safety or effectiveness questions.