K Number
K233020

Validate with FDA (Live)

Date Cleared
2024-01-12

(112 days)

Product Code
Regulation Number
884.1730
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TauTona Pneumoperitoneum Assist Device (TPAD) is intended for use in the upper left quadrant of the abdominal wall (i.e., Palmer's Point) with a Veress needle for the establishment of a pneumoperitoneum during laparoscopic procedures.

Device Description

The TauTona Pneumoperitoneum Assist Device (TPAD) is a single use device used during laparoscopic surgical procedures. The device consists of a pad which contains an array of suction cups on one side; a port that attaches to a standard hospital vacuum line; a button to control vacuum to the suction cups; and a handle to allow user to manipulate the device. Suction is applied to the device to allow the user to manipulate (hold and pull on) the tissue around the Veress needle insertion site. The device is crescent in shape to allow removal of the device after use while the Veress needle remains in the patient.

AI/ML Overview

The provided text is a 510(k) summary for the TauTona Pneumoperitoneum Assist Device (TPAD). It outlines the device's purpose, comparison to a predicate device, and the non-clinical and clinical performance testing performed.

However, the 510(k) summary does not include a formal table of acceptance criteria for the device performance or detailed information about the specific "study that proves the device meets the acceptance criteria" in the format requested. It describes a clinical study to evaluate safety, effectiveness, and ease of use, but doesn't explicitly link specific metrics to predefined acceptance thresholds.

Given the information available, I will extract and infer as much as possible to address your request, acknowledging where information is not explicitly stated.

Here's an attempt to answer your questions based on the provided text:


The provided 510(k) summary describes a clinical study to demonstrate the safety, effectiveness, and ease of use of the TauTona Pneumoperitoneum Assist Device (TPAD). While formal, quantitative "acceptance criteria" for metrics are not explicitly presented in a table format with corresponding reported performance, the summary describes the outcomes of the study.

1. Table of Acceptance Criteria and Reported Device Performance

As explicit acceptance criteria with numerical thresholds are not provided in the document, this table is constructed based on the described study outcomes and implied "acceptance" as demonstrated by the findings.

Metric/Performance AreaAcceptance Criteria (Inferred from Study Design & Outcomes)Reported Device Performance
Feasibility/Ease of Use (Primary Endpoint)Surgeon questionnaire median score not significantly different from a theoretical median of 4.0, indicating acceptable feasibility/ease of use.TPAD: Median score of 4.0 for all responses on surgeon questionnaire. Not significantly different from a theoretical median of 4.0 (Wilcoxon test).
Time to Obtain Laparoscopic AccessNo statistically significant difference compared to Standard of Care (SOC).No statistically significant difference compared to SOC (unpaired t-test).
Time to Complete InsufflationNo statistically significant difference compared to SOC.No statistically significant difference compared to SOC (unpaired t-test).
Adverse Events (Safety)No device-related serious adverse events (SAEs). All other adverse events deemed mild and completely resolving. Bruising resolution comparable or better than SOC.No device-related SAEs. All other adverse events (bruising) deemed mild and related to insertion site/skin handling, with complete resolution. 6/10 TPAD subjects' bruising resolved within 3 days, 1/10 at 19 days. SOC (control) had 5/10 subjects with bruising still present at 7+ days.
Patient Satisfaction (Skin Appearance/Bruising)Comparable or better patient satisfaction regarding skin appearance/bruising compared to SOC.Patient surveys and photographs were collected. Specific aggregate results for patient satisfaction are not detailed, but the bruising resolution data suggests a positive outcome for TPAD subjects compared to SOC.
Veress Needle Placement SuccessAll surgeons able to place the Veress needle using the TPAD at the 5-20 mmHg pressure.All surgeons were able to place the Veress needle using the TPAD at the 5-20 mmHg pressure.

Note: The "Acceptance Criteria" provided above are inferred from the study's discussion of "primary endpoint," "secondary endpoints," and "statistical significance" compared to standard of care or theoretical medians, rather than explicit pre-defined numerical thresholds in an acceptance criteria table.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Clinical Study): 20 individuals total, split into two arms:
    • TPAD arm: 10 patients
    • Standard of Care (SOC) arm: 10 patients
  • Data Provenance:
    • Country of Origin: United States (U.S.). The study was performed at an academic center in the U.S.
    • Retrospective or Prospective: Prospective (described as a "randomized clinical study").

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not explicitly state the "number of experts used to establish the ground truth" in the sense of independent expert review of cases for accuracy. Instead, the "ground truth" or device performance was assessed by:

  • Surgeons: who performed the procedures and completed questionnaires on device feasibility/ease of use. Their specific qualifications (e.g., years of experience, specialty) are not detailed beyond "surgeon."
  • Patients: who completed surveys to assess patient satisfaction.
  • Study Staff: who likely monitored adverse events, collected time metrics, and captured photographs.

Therefore, the "experts" are the performing surgeons and the patients themselves providing direct feedback on their experience.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method for the test set in the context of expert review (e.g., 2+1, 3+1). The primary endpoint (surgeon questionnaire) was completed by the performing surgeon, not an independent panel, and secondary outcomes were objective measurements (time, adverse events) or patient self-reported.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No, an MRMC comparative effectiveness study was not performed. This device is not an AI-based diagnostic or assistive software for image interpretation, but rather a physical surgical assist device. Therefore, the concept of "human readers improving with AI assistance" is not applicable here. The clinical study compared the use of the TPAD device (a physical assist device) against the Standard of Care.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable. The device is a physical surgical assist device and does not involve AI algorithms or standalone performance. It is used with a human surgeon.

7. The Type of Ground Truth Used

The "ground truth" for the clinical study was established through:

  • Direct observation and measurement: Time metrics (time to obtain laparoscopic access, time to complete insufflation).
  • Surgeon feedback: Questionnaires on feasibility/ease of use, reflecting the practical aspects of the device during actual use.
  • Patient outcomes: Adverse event monitoring (bruising resolution) and patient reported satisfaction via surveys and photographs.

This can be categorized as a combination of objective clinical measurements, direct user (surgeon) assessment, and patient-reported outcomes. It is not based on expert consensus of reviewed cases, pathology, or large-scale outcomes data in the traditional sense of diagnostic accuracy studies.

8. The Sample Size for the Training Set

This information is not applicable as the device is a physical medical device and contains no software/AI that would require a "training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as the device is a physical medical device and contains no software/AI that would require a "training set."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

January 12, 2024

TauTona Group Research & Development, Co. LLC Michael Blanchette Chief Operating Officer 604 Fifth Ave Suite D Redwood City, CA 94063

Re: K233020

Trade/Device Name: TauTona Pneumoperitoneum Assist Device (TPAD) Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF Dated: December 8, 2023 Received: December 11, 2023

Dear Michael Blanchette:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jason Roberts -S

Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K233020

Device Name

TauTona Pneumoperitoneum Assist Device (TPAD)

Indications for Use (Describe)

The TauTona Pneumoperitoneum Assist Device (TPAD) is intended for use in the upper left quadrant of the abdominal wall (i.e., Palmer's Point) with a Veress needle for the establishment of a pneumoperitoneum during laparoscopic procedures.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

SUBMITTER INFORMATION

Applicant:TauTona Research and Development Group, LLC
Contact:Mike Blanchette
Phone:650-503-8250
Email:mblanchette@tautonagroup.com
Address:604 Fifth Ave, Suite DRedwood City, CA 94063

CORRESPONDENT INFORMATION

Contact:Joshua Crist
Title:Regulatory Consultant- Medical Device
Firm:Biologics Consulting Group

DATE PREPARED: 01/09/2024

DEVICE INFORMATION

Device Name:TauTona Pneumoperitoneum Assist Device (TPAD)
Common Name:Laparoscopic insufflator
Regulation Number:884.1730
Regulation Name:Laparoscopic insufflator
Product Code:HIF
Regulatory Class:Class II

PREDICATE DEVICE INFORMATION

Device Name:Aragon Surgical LapCap
510(k) Number:K070651
Manufacturer:Aragon Surgical Inc.

The predicate device has not been subject to a design related recall.

{4}------------------------------------------------

510(K) SUMMARY

DEVICE DESCRIPTION

The TauTona Pneumoperitoneum Assist Device (TPAD) is a single use device used during laparoscopic surgical procedures. The device consists of a pad which contains an array of suction cups on one side; a port that attaches to a standard hospital vacuum line; a button to control vacuum to the suction cups; and a handle to allow user to manipulate the device. Suction is applied to the device to allow the user to manipulate (hold and pull on) the tissue around the Veress needle insertion site. The device is crescent in shape to allow removal of the device after use while the Veress needle remains in the patient.

INDICATIONS FOR USE

The TauTona Pneumoperitoneum Assist Device (TPAD) is intended for use in the upper left quadrant of the abdominal wall (i.e., Palmer's Point) with a Veress needle for the establishment of a pneumoperitoneum during laparoscopic procedures.

COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The table below compares the intended use and the technological characteristics of the subject device and predicate device.

Subject DeviceTauTonaPneumoperitoneum AssistDevice (TPAD)Predicate DeviceAragon Surgical LAPCAPK070651Comparison
Product CodeHIFHIFSame
Regulation Number884.1730884.1730Same
Regulatory ClassIIIISame
Indications for UseThe TauTonaPneumoperitoneum AssistDevice TPAD is intended foruse in the upper left quadrant ofthe abdominal wall (i.e.,Palmer's Point) with a Veressneedle for the establishment ofa pneumoperitoneum duringlaparoscopic procedures.The Aragon Surgical LapCap isintended for use in the peri-umbilical region of theabdominal wall with a Veressneedle for the establishment ofa pneumoperitoneum duringgynecologic (pelvic) andgeneral surgical(intraabdominal) laparoscopicprocedures.Similar indicationsfor use. TPADLocation for use isfor Palmers PointAccess.
Veress Needle InsertionUse of negative pressure toattach to skin to support Veressneedle insertion. Device can beUse of negative pressure toattach to skin to support Veressneedle insertion. Device isremoved with Veress Needle.Same operatingprinciple.Different removalrequirements.

Table 1: Comparator Table for Subject and Predicate Devices

{5}------------------------------------------------

Subject DeviceTauTonaPneumoperitoneum AssistDevice (TPAD)Predicate DeviceAragon Surgical LAPCAPK070651Comparison
removed while Veress Needleis in place.
Vacuum ApplicationArray of suction cupsLarge polycarbonate domeSimilar, the minordifferences aresupported by bothbench and clinicaldata.
Vacuum SourceExternal – Hospital VacuumLineExternal – Hospital VacuumLineSame
Software/ElectronicsNoneNoneSame
BiocompatibilityBiocompatibility testing inaccordance with ISO 10993-1Biocompatibility testing inaccordance with ISO 10993-1Same

510(K) SUMMARY

Both devices use negative pressure to attach to skin and support Veress Needle insertion. The subject device, however, applies the vacuum with a design that allows removal of the device after Veress Needle insertion.

SUMMARY OF NON-CLINICAL PERFORMANCE TESTING

Biocompatibility Testing

Studies were performed and confirmed that the Subject device is biocompatible for the intended patient contact profile in accordance with ISO 10993-1.

Electrical Safety

Not applicable. The subject device contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.

Electromagnetic Compatibility (EMC)

Not applicable. The subject device contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.

Software

Not applicable. The subject device contains no software.

Performance Testing

Other performance testing was conducted to show that the device meets its design requirements and performs as intended. The performance tests include:

  • Dimensional testing for vacuum line compatibility, device features
  • Vacuum control /holding forces. ●

{6}------------------------------------------------

  • . Tensile strength
  • Shelf-Life .

Clinical Data

A randomized clinical study consisting of two arms was performed at an academic center in the U.S. The purpose of this study was to evaluate the safety, effectiveness and ease of use of the TPAD for laparoscopic surgery when the Veress Needle was used to establish pneumoperitoneum. Study population included 20 individuals greater than 18 years old (age range 25 - 73 years old), with a BMI < 35 kg/m2 (range 21.6 - 34.8) requiring a laparoscopic surgery. Other demographics in the study included 55% white, 40% Hispanic/Latino and 5% Asian: 3 biological males and 7 biological females in the TPAD arm, while 5 biological males and 5 biological females were in the Standard of Care (SOC) arm. Ten (10) patients underwent laparoscopic surgery as otherwise planned, with their surgeon using the TPAD device to assist in placement of the Veress needle. Ten (10) additional patients underwent laparoscopic surgery following the SOC. Surgeons completed a questionnaire following the procedures to evaluate the device feasibility/ ease of use. All patients also completed surveys to assess patient satisfaction. The primary endpoint was device feasibility as measured by a surgeon questionnaire completed at the end of every surgery. Secondary endpoints included the time needed to obtain laparoscopic access using the device, time needed to complete insufflation, monitoring for adverse events, time to resolve bruising, and patient satisfaction on skin appearance/bruising as measured by patient surveys and photographs postoperatively. All surgeons were able to place the Veress needle using the TPAD at the 520-mmHg pressure. The surgeon questionnaire answers for both SOC (average range of 3.9 to 4.0) and TPAD (average range of 3.9 to 4.0) had a median of 4.0 for all responses. For all questionnaires, both SOC and TPAD were not significantly different from a theoretical median of 4.0 (Wilcoxon test). Also, the secondary endpoints did not have any statistically significant difference using unpaired t test.

There were no device related serious adverse events (SAEs). All other adverse events were deemed mild and were related to insertion site and skin handling bruising with complete resolution in the TPAD arm as follows: 6/10 subjects' bruising resolved within 3 days, 2/10 subjects between 4 - 6 days, 1/10 subject at 7 days and 1/10 subject at 19 days postoperatively. For the SOC arm bruising resolved within 3 days for 4/10 subjects, between 4 - 6 days for one subject and 5/10 subjects still had bruising at 7+ days postoperatively (one at 7 days, two at 7+ days, one at 8+ days, one at 9+ days and one at 16+ days).

CONCLUSION

The subject device has the same intended use as the predicate, with similar technological characteristics. The differences in technological characteristics do not raise different questions of safety and effectiveness. Performance testing conducted on subject device demonstrate that is as safe and effective as the predicate device to support substantial equivalence.

§ 884.1730 Laparoscopic insufflator.

(a)
Identification. A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.(b)
Classification. (1) Class II (performance standards).(2) Class I for tubing and tubing/filter kits which include accessory instruments that are not used to effect intra-abdominal insufflation (pneumoperitoneum). The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.