510(k) Data Aggregation

    K Number
    K252648

    Validate with FDA (Live)

    Date Cleared
    2026-03-13

    (204 days)

    Product Code
    Regulation Number
    876.5015
    Age Range
    22 - 77
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    8mm×20mm stents:
    The Niti-S SPAXUS™ Stent is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥6 cm in size, that are adherent to the gastric or bowel wall and are free of solid debris. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst resolution.

    10mm×20mm and 16mm×20mm stents:
    The Niti-S SPAXUS™ Stent is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size and walled-off necrosis ≥ 6 cm in size that are adherent to the gastric or bowel wall. Once placed, the Niti-S SPAXUS™ Stent functions as an access port allowing passage of standard and therapeutic endoscopes to facilitate debridement, irrigation and cystoscopy. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution. The Niti-S SPAXUS™ Stent is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of the gallbladder in patients with acute cholecystitis who are at high risk for surgery.

    Device Description

    The Niti-S SPAXUS™ Stent is a fully covered, self-expanding metallic stent. The stent is constructed from nickel-titanium alloy (nitinol), which provides flexibility, durability, and self-expansion properties. Its bi-flange design provides stable lumen apposition and helps mitigate migration risk. The complete silicone coating helps prevent leakage during drainage and facilitates the safe removal of the device. Radiopaque markers made of platinum/iridium and stainless steel enhance visibility under fluoroscopic guidance, aiding precise placement and monitoring.

    The Stent Delivery System (SDS) is a disposable, non-electrocautery system for the delivery and deployment of the stent at the target position. It is similar in design to the SDS used for the Esophageal TTS Stent (K240522).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251123

    Validate with FDA (Live)

    Date Cleared
    2025-12-29

    (262 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Niti-S Biliary Stent:
    The Niti-S Biliary Stent is indicated for the palliation of malignant strictures in the biliary tree.

    Niti-S Biliary Slim M Stent:
    The Niti-S Biliary Slim M Stent is indicated for the palliation of malignant strictures in the biliary tree.

    Device Description

    The Niti-S Biliary Stent consists of an implantable metallic stent and a disposable, flexible Stent Delivery System for placement of the stent. The stent is a flexible and expandable tubular device made of Nitinol wire. Two Nitinol wires are woven in a hook-type design. The stent design is identical to the predicate device cleared in K073667.

    The Stent Delivery System is a disposable system for the delivery and deployment of the stent at the target position.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243619

    Validate with FDA (Live)

    Date Cleared
    2025-10-30

    (342 days)

    Product Code
    Regulation Number
    878.3610
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Niti-S Esophageal Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas.

    The Esophageal TTS Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas.

    Device Description

    The Esophageal TTS Stents and Niti-S Esophageal Stents consist of an implantable metallic stent and a disposable, flexible introducer system for placement of the stent. The stent is a flexible and expandable tubular device made of Nitinol wire that is intended to be implanted to restore the structure and/or function of the esophagus. The introducer is a disposable system for delivery and deployment of the stent at the target position. Upon deployment, the stent imparts an outward radial force on the luminal surface of the esophagus to establish patency. A length of suture is attached to one end of the stent and can be used for stent repositioning during initial placement, in accordance with the instructions for use.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250663

    Validate with FDA (Live)

    Date Cleared
    2025-03-31

    (26 days)

    Product Code
    Regulation Number
    878.3610
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Niti-S Duodenal Comfort Stent is indicated for the palliative treatment of pyloric or duodenal obstructions caused by malignant neoplasms.

    Niti-S Colonic Comfort Stent is indicated for the palliative treatment of colorectal strictures produced by malignant neoplasms and to relieve large bowel obstruction prior to colectomy in patients with malignant strictures.

    Device Description

    The Niti-S Duodenal Comfort Stent and Niti-S Colonic Comfort Stent consist of an implantable metallic stent and an introducer system.

    The stent is made of nitinol wire and is designed as a flexible, fine mesh tubular prosthesis with several radiopaque markers. The Niti-S Duodenal Comfort Stent has a diameter of 22 mm, while the Niti-S Colonic Comfort Stent is available in diameters of 22 mm and 24 mm. Both stent types are available in lengths of 60 mm, 80 mm, 100 mm, and 120 mm.

    The stent delivery system accommodates a 0.035 in (0.89 mm) guidewire. It is advanced over the guidewire and passed through an endoscope into the duodenum or colon. Fluoroscopy is recommended to ensure accurate device placement.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the U.S. Food & Drug Administration (FDA) for medical devices. It primarily focuses on the substantial equivalence determination for the Niti-S Duodenal Comfort Stent and Niti-S Colonic Comfort Stent.

    Crucially, this document does not describe studies involving AI or software-based medical devices with performance metrics like sensitivity, specificity, or AUC. It is for a physical medical device (stent and delivery system) and the performance testing mentioned are mechanical and material integrity tests.

    Therefore, I cannot provide the information requested regarding acceptance criteria and studies proving the device meets those criteria in the context of an AI/software device. The questions about sample size for test sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set details are all irrelevant to the content of this document.

    The "Performance Testing" section (page 7) lists the following non-clinical tests:

    • Packaging strength and dye penetration
    • Sterility verification
    • Deployment accuracy test
    • Deployment force test
    • Dimensional test
    • Tensile strength test (SDS)
    • Trackability and visualization test
    • Repositioning force test

    These are standard engineering and material science tests for physical medical devices to ensure they function as intended and are safe. The document states that the results "demonstrate that the Niti-S Duodenal Comfort Stent and Niti-S Colonic Comfort Stent are as safe and effective as the predicate device and supports a determination of substantial equivalence."

    In summary, the provided document does not contain the information requested about AI device performance studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240522

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-03-22

    (28 days)

    Product Code
    Regulation Number
    878.3610
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Esophageal TTS Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas.

    Device Description

    The Esophageal TTS Stents that are the subject of this 510(k) are identical to the devices cleared in K221482, with the exception of the modification to its introducer system. The Esophageal TTS Stents consist of an implantable metallic stent and a disposable, flexible introducer system for placement of the stent. The stent is a flexible and expandable tubular device made of Nitinol wire that is intended to be implanted to restore the structure and/or function of the esophagus. The introducer is a disposable system for delivery and deployment of the stent at the target position. Upon deployment, the stent imparts an outward radial force on the luminal surface of the esophagus to establish patency.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "Esophageal TTS Stent." It explains that the new device is substantially equivalent to a previously cleared device (K221482) manufactured by the same company, Taewoong Medical.

    The core of the document focuses on demonstrating this substantial equivalence, noting that the stent itself is identical to the predicate device, with modifications only to the introducer system (the delivery mechanism). Therefore, much of the testing performed is related to the performance and biocompatibility of this modified introducer system.

    Based on the provided document, here's a breakdown of the requested information:

    This document is a marketing submission for a medical device (Esophageal Stent) that does not involve AI or software. Therefore, many of the requested criteria related to AI/software performance validation, such as "acceptance criteria for an AI model," "MRMC study," "standalone algorithm performance," "ground truth establishment for training set," are not applicable to this submission.

    The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the standard non-clinical performance and biocompatibility testing for a physical medical device, not an AI or software component.

    Here's a detailed response to the applicable sections of your request:


    1. A table of acceptance criteria and the reported device performance

    The document does not provide a formal table with specific numerical acceptance criteria and reported numerical performance for each test. Instead, it lists the types of performance tests conducted and then states a general conclusion that the device "meets its design requirements and performs as intended" and is "as safe and effective as the predicate device."

    However, we can infer the types of tests and the general outcome:

    Acceptance Criteria CategorySpecific Test Conducted (Implied Criteria: Meet Design Req. & Perform as Intended)Reported Device Performance
    BiocompatibilityCytotoxicity (ISO 10993-5:2009)Pass (Supports SE)
    Sensitization (ISO 10993-10:2021)Pass (Supports SE)
    Intracutaneous Reactivity (ISO 10993-23:2021)Pass (Supports SE)
    Acute Systemic Toxicity (ISO 10993-11:2017)Pass (Supports SE)
    Pyrogen Test (ISO 10993-11:2017, USP <151>)Pass (Supports SE)
    Device PerformanceDeployment TestMeets design requirements
    Deployment Force TestMeets design requirements
    Dimensional TestMeets design requirements
    Tensile Strength TestMeets design requirements

    Notes on Acceptance Criteria:

    • For biocompatibility tests, the acceptance criteria would be defined by the ISO standards, e.g., "non-cytotoxic," "not a sensitizer," etc.
    • For performance tests (Deployment, Deployment Force, Dimensional, Tensile Strength), the acceptance criteria would be specific to the device's engineering specifications, such as "deployment within X seconds," "deployment force less than Y Newtons," "dimensions within Z tolerance," "tensile strength greater than W," etc. These specific numerical criteria and results are not provided in this summary document. The document only states that the testing was conducted and found to demonstrate substantial equivalence.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size:
      • The document does not specify the sample sizes used for the non-clinical performance tests (Deployment, Deployment Force, Dimensional, Tensile Strength).
      • For biocompatibility, the sample sizes would be typical for the specific ISO test methods (e.g., a certain number of mice for acute systemic toxicity, guinea pigs for sensitization), but these are not explicitly detailed.
    • Data Provenance:
      • This is a submission for a medical device (stent and introducer), not data related to a clinical study or AI model.
      • The testing pertains to the physical characteristics and interaction of the device components.
      • The manufacturer is Taewoong Medical Co., Ltd., based in the Republic of Korea. It's highly probable the testing was conducted there or at subcontracted labs.
      • The tests are prospective in nature, as they are performed on newly manufactured devices or materials to demonstrate compliance for premarket submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This device is a physical medical device (stent and introducer system) and does not involve a diagnostic AI or imaging interpretation. Therefore, there is no "ground truth" derived from expert interpretation of images or clinical data in the context of an AI model.
    • The "ground truth" for this device's performance is established by engineering specifications, physical measurements, and standardized biological safety tests. The "experts" involved would be engineers, material scientists, and toxicologists/biologists conducting and interpreting the non-clinical tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies or for establishing ground truth in machine learning contexts (e.g., for radiology readings) where there might be disagreement among human readers. Since this is a submission for a physical device based on non-clinical performance and biocompatibility testing, such adjudication methods are not relevant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This submission is for a physical medical device (esophageal stent) and its introducer system. It does not involve AI or software, and therefore, no MRMC study or assessment of human reader improvement with AI assistance was conducted or would be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device does not have an "algorithm" component. It is a physical medical device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "ground truth" for this device is based on engineering specifications and established test methodologies.
      • For performance tests (Deployment, Deployment Force, Dimensional, Tensile Strength), the ground truth is defined by the pre-defined design requirements and specifications (e.g., a stent must deploy within a certain time, or a certain force is required for deployment).
      • For biocompatibility tests, the ground truth is defined by the results of standardized in vitro and in vivo biological tests as per ISO 10993 series, indicating the material's safety for contact with the body.
    • It is not based on expert consensus, pathology, or outcomes data in the way these terms are typically used for diagnostic or AI-related devices.

    8. The sample size for the training set

    • Not Applicable. This is not an AI or machine learning device. Therefore, there is no "training set." The device's design is based on engineering principles and knowledge gained from previous device iterations (the predicate device).

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set (being a physical medical device, not an AI), there is no ground truth establishment for a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223067

    Validate with FDA (Live)

    Date Cleared
    2023-06-14

    (264 days)

    Product Code
    Regulation Number
    878.3610
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Niti-S Duodenal Comfort Stent is indicated for the palliative treatment of pyloric or duodenal obstructions caused by malignant neoplasms.

    The Niti-S Colonic Comfort Stent is indicative treatment of colorectal strictures produced by malignant neoplasms and to relieve large bowel obstruction prior to colectomy in patients with malignant strictures.

    Device Description

    The Niti-S Duodenal Comfort Stent and Niti-S Colonic Comfort Stent consist of the implantable metallic stent and introducer system.

    The stent is made of nitinol wire. It is a flexible, fine mesh tubular prosthesis that has several radiopaque markers. The Niti-S Duodenal Comfort Stent has a diameter of 22 mm while the Niti-S Colonic Comfort Stent has diameters of 22 and 24 mm. They both have lengths of 60, 80, 100, and 120 mm.

    The introducer system accepts a 0.035 in (0.89 mm) guidewire. The introducer system is passed over the guidewire and through an endoscope into the duodenum or colon. Use of fluoroscopy is recommended to ensure correct placement of the device.

    AI/ML Overview

    This document is a 510(k) summary for the Niti-S Duodenal Comfort Stent and Niti-S Colonic Comfort Stent. It primarily focuses on demonstrating substantial equivalence to predicate devices through a comparison of technological characteristics and performance data from bench testing. It does not describe a study involving human subjects or AI.

    Here's an analysis of what is present and what is missing based on your request:

    1. Table of acceptance criteria and the reported device performance:

    The document mentions that "The results of the bench testing show that the subject devices meet their specifications" for various tests. However, it does not provide a specific table outlining the acceptance criteria and the numerical results for each test. Instead, it lists the types of bench tests conducted:

    Acceptance Criteria (Implied)Reported Device Performance
    Meet specifications for:- Deployment Accuracy
    - Deployment Force- Expansion Force
    - Compression Force- Dimensional
    - Tensile Strength- Shipping
    - Shelf-life- Corrosion
    - MR Compatibility- Axial Force
    - Trackability and Visualization- Repositioning Force
    - RadiopacityAll listed tests met specifications and demonstrate substantial equivalence.
    SterilityIn accordance with ISO 11135:2014 & AAMI/ANSI/ISO 11607-1:2019, AAMI/ANSI/ISO 11607-2:2019
    BiocompatibilityAcceptable biocompatibility profile (evaluated for cytotoxicity, sensitization, irritation/intracutaneous reactivity, acute systemic toxicity, material-mediated pyrogenicity, subacute/subchronic toxicity, chronic toxicity, and implantation effects)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: Not specified. The document only refers to "bench testing."
    • Data Provenance: Not specified. Bench testing is typically performed in a lab setting, so country of origin isn't directly relevant in the same way as clinical data. This is not a study on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    Not applicable. This document describes bench testing of a medical device (stent and introducer system), not an AI algorithm or a diagnostic tool requiring expert ground truth in a clinical context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. As this is bench testing, there is no expert adjudication of image interpretations or clinical diagnoses.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No. This is a 510(k) submission for a physical medical device (stent), not an AI-powered diagnostic or assistive tool. Therefore, no MRMC study or AI-related effectiveness analysis was conducted or reported here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No. This is for a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    Not applicable. For bench testing, the "ground truth" would be the engineering specifications and performance standards that the device is designed to meet. The document states the tests "meet their specifications."

    8. The sample size for the training set:

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223626

    Validate with FDA (Live)

    Date Cleared
    2023-05-26

    (172 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Niti-S Biliary Speed D Stent is indicated for the palliation of malignant strictures in the biliary tree.

    Device Description

    The Niti-S Biliary Speed D Stent consists of an implantable metallic stent and a disposable, flexible Stent Delivery System for placement of the stent is a flexible and expandable tubular device made of Nitinol wire. Two Nitinol wires are woven in a hook-type design. The stent design is identical to the predicate device cleared in K073667. The Stent Delivery System is a disposable system for the delivery and deployment of the stent at the target position.

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter and summary for a medical device called the "Niti-S Biliary Speed D Stent." This document focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than performing a de novo clinical study to establish performance against acceptance criteria for a novel AI/software medical device.

    Therefore, the information required to populate a table of acceptance criteria and a detailed study proving device performance (especially for an AI/software product, as implied by questions like sample size for training data, expert consensus, MRMC studies, etc.) is not present in this document.

    Specifically, the document states:

    • "Software Verification and Validation Testing: Not applicable. The subject device contains no software of any type." (Page 8)
    • "Animal Testing: Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device." (Page 9)
    • "Clinical Data: Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device." (Page 9)

    The 510(k) pathway for this device relies primarily on:

    • Identical Indications for Use
    • Highly similar technological characteristics to a predicate device (Niti-S Biliary Stent, K073667).
    • Non-clinical bench testing to confirm physical characteristics and performance (e.g., integrity, radial force, material properties) and biocompatibility testing for material changes in the delivery system.
    • Sterilization and Shelf Life testing.

    As such, I cannot provide the requested information about acceptance criteria and study details for an AI/software medical device based on this document. The document describes a physical medical device (stent and delivery system), not an AI or software product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223256

    Validate with FDA (Live)

    Date Cleared
    2023-04-05

    (166 days)

    Product Code
    Regulation Number
    876.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Optimos™ Cystotome is intended for use as an electrosurgical accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst that is visibly bulging into the gastrointestinal tract. This device is supplied sterile and is intended for single use only.

    Device Description

    Optimos™ Cystotome is an electrode which uses heat generated from the application of high frequency (HF) energy through the monopolar electrode to puncture a hole in tissue. Following a fine-needle aspiration (FNA) procedure, it is positioned over a guidewire, so the applied part is located at the area of interest. The Optimos™ Cystotome is for use to drain pancreatic pseudocysts. The Optimos™ Cystotome is a single use device provided sterile to the user. No reprocessing is performed by the user. The device is provided in three sizes, which have different diameters and tip lengths. The device itself consists of 3 parts: handle, sheath, and tip.

    AI/ML Overview

    The device is the Optimos™ Cystotome. The study described focuses on bench testing to demonstrate substantial equivalence to a predicate device, the Wilson-Cook Cystotome.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    1. The ability to puncture 2 type tissues (Porcine stomach and Duodenum).The test results showed that all test samples meet the required performance, puncture tissues. The Optimos™ Cystotome successfully punctured both porcine stomach and duodenum tissues.
    2. Substantial equivalence of thermal damage: There should be no significant difference in measurement of size (length, width, and depth) of the thermal damage zone between subject device and predicate device.The thermal effects were evaluated on both the subject device and the equivalent device (Wilson-Cook CST-10 Cystotome Cystoenterostomy Needle Knife). The performance of the Optimos™ Cystotome was comparable to the predicate device at 18, 100, and 120W on porcine stomach tissue and duodenum in the mucosa, submucosa, and muscularis propria. The damage zone was measured under a microscope and analyzed using histological strain and NADH. The test results met the acceptance criteria, indicating comparable thermal damage.
    Additional Bench Testing:
    Visual test, dimensional test, continuity test, endoscopic compatibility test, device operability test, high frequency dielectric strength test, detachable connector retention test, injection & leakage test, and tensile strength test (with corresponding acceptance criteria for each)All tests passed the set acceptance criteria. System tests confirmed compatibility with fine needle aspiration, guidewire, endoscope, and ESU.
    Biocompatibility Testing (for various test items)All biocompatibility tests (Cytotoxicity, Intracutaneous reactivity, Sensitization (GPMT), Acute systemic toxicity, Material mediated pyrogenicity test) passed. For example, Cytotoxicity showed a reactivity grade of 0, Intracutaneous reactivity showed no abnormal signs and a mean score of 0.00, and Sensitization showed a score of 0.0.
    Electrical Safety and Electromagnetic Compatibility (EMC) against standards: EN 60601-1 (2006) + A1:2013, IEC 60601-1:2005/AMD1:2012, IEC 60601-1-2, IEC 60601-2-2:2017, IEC 60601-2-18All listed electrical safety and EMC standards were passed (P).

    2. Sample size used for the test set and the data provenance

    • Sample Size for Thermal Effects Comparison:
      • Optimos™ Cystotome (Subject Device): 18 samples (Model number OCT1906)
      • Cystotome™ Cystoenterostomy Needle Knife (Predicate Device): 18 samples (Model number CST-10)
    • Data Provenance: This was a bench test conducted in a laboratory setting using porcine stomach and duodenum tissue. The document does not specify the country of origin of the data beyond the manufacturer (Taewoong Medical Co., Ltd. from Korea) performing the tests. It is retrospective in the sense that the data has already been collected and analyzed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • The document does not specify the use of experts to establish ground truth for the bench testing. The ground truth for thermal damage was established by quantitative measurements of thermal damage zones under a microscope, analyzed using histological strain and NADH.

    4. Adjudication method for the test set

    • An explicit adjudication method (e.g., 2+1, 3+1) is not mentioned or implied. The measurements and evaluations appear to be made against objective criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. This device is an electrosurgical accessory, not an AI-assisted diagnostic tool. Therefore, the concept of human readers improving with AI assistance is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A. This is a medical device (Cystotome), not an algorithm. The performance described is for the device itself (its ability to puncture tissue and the thermal effects it produces) in controlled bench testing.

    7. The type of ground truth used

    • For the thermal effects comparison, the ground truth was objective physical measurements derived from histological analysis of tissue samples (porcine stomach and duodenum). This involved measuring the length, width, and depth of the thermal damage zone under a microscope and analyzing it using histological strain and NADH.

    8. The sample size for the training set

    • N/A. This device does not involve machine learning or AI, so there is no concept of a training set.

    9. How the ground truth for the training set was established

    • N/A. As there is no training set, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221482

    Validate with FDA (Live)

    Date Cleared
    2022-06-22

    (30 days)

    Product Code
    Regulation Number
    878.3610
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Esophageal TTS Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas.

    Device Description

    The Esophageal TTS Stents consists of an implantable metallic stent and a disposable, flexible introducer system for placement of the stent is a flexible and expandable tubular device made of Nitinol wire that is intended to be implanted to restore the structure and/or function of the esophagus. The introducer is a disposable system for delivery and deployment of the stent at the target position. Upon deployment, the stent imparts an outward radial force on the luminal surface of the esophagus to establish patency.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Taewoong Medical Co., Ltd. Esophageal TTS Stent. It states that the device is "substantially equivalent" to previously cleared predicate devices (K211706, K123205, K113551) with the "exception of a modification to the Indications for Use statement."

    As such, this document does not contain acceptance criteria for this specific submission with new performance data to prove the device meets these criteria. Instead, it relies on the previous clearances and argues that the change in indication does not affect safety and efficacy, therefore no additional performance data is required for this 510(k).

    Therefore, I cannot provide the requested information regarding acceptance criteria and performance study details from this document. The document explicitly states:

    • No additional biocompatibility information is required.
    • Not applicable for Electrical safety and electromagnetic compatibility (EMC) as it contains no electrical components.
    • Not applicable for Software Verification and Validation Testing as it contains no software.
    • No additional performance data is required for Bench Testing because no design modifications were made.
    • Not applicable for Animal Testing.
    • Not applicable for Clinical Data.

    The 510(k) submission relies on the substantial equivalence principle, asserting that since the new device is identical in design and materials to previously cleared devices, and the change in indication does not introduce new risks, the prior performance data of the predicate devices is sufficient.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221071

    Validate with FDA (Live)

    Date Cleared
    2022-06-09

    (58 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Niti-S Biliary Slim M Stent is indicated for the palliation of malignant strictures in the biliary tree.

    Device Description

    The Niti-S Biliary Slim M Stent consists of the implantable metallic stent and introducer system. The stent is made of Nitinol wire. It is a flexible, fine mesh tubular prosthesis and it has 7 radiopaque markers; 3 in each end 1 in the center. It has a diameter of 6 or 8 mm and the length ranges from 40 to 120mm.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a medical device called the "Niti-S Biliary Slim M Stent." This type of submission aims to demonstrate substantial equivalence to a legally marketed predicate device, rather than proving safety and efficacy through extensive clinical trials for novel devices. As such, the document does not describe acceptance criteria for an AI/algorithm-driven device or study details for such a device, but rather for a physical medical device (a stent).

    Therefore, I cannot extract the information required by your prompt, which specifically asks for criteria and study details related to an AI/algorithm-driven device like:

    1. Acceptance criteria and device performance table: The document describes the device, its intended use, and compares its technical characteristics to predicate devices. It does not provide performance metrics like sensitivity, specificity, or AUC which would be relevant for an AI/algorithm.
    2. Sample size and data provenance: There is no mention of a test set, training set, or data provenance from patient images, as this is not an AI/algorithm.
    3. Experts for ground truth and qualifications: Not applicable, as there's no ground truth establishment for an AI.
    4. Adjudication method: Not applicable.
    5. MRMC study and effect size: No AI assistance or human reader performance is discussed.
    6. Standalone performance: No algorithm is being evaluated.
    7. Type of ground truth: Not applicable.
    8. Training set sample size: Not applicable.
    9. Ground truth for training set: Not applicable.

    Instead, the document focuses on:

    • Bench Testing: Describing physical tests performed on the stent (e.g., Pitting Corrosion Potential, Dimensional Verification, Radial Compression Force) and the delivery system (e.g., Delivery, Deployment, Withdrawal).
    • Biocompatibility Testing: Stating that the material is identical to a previously cleared predicate device.
    • Electrical Safety and EMC/Software Verification and Validation: Stating "Not applicable" as the device contains no electrical components or software.
    • Animal Testing / Clinical Data: Stating "Not applicable" as these were not deemed necessary to establish substantial equivalence for this device.

    In summary, the provided text describes the regulatory clearance process for a physical medical device (a stent) under the 510(k) pathway, and therefore does not contain the information requested about acceptance criteria and studies for an AI/algorithm-driven device.