K Number
K223067

Validate with FDA (Live)

Date Cleared
2023-06-14

(264 days)

Product Code
Regulation Number
878.3610
Age Range
0 - 21
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Niti-S Duodenal Comfort Stent is indicated for the palliative treatment of pyloric or duodenal obstructions caused by malignant neoplasms.

The Niti-S Colonic Comfort Stent is indicative treatment of colorectal strictures produced by malignant neoplasms and to relieve large bowel obstruction prior to colectomy in patients with malignant strictures.

Device Description

The Niti-S Duodenal Comfort Stent and Niti-S Colonic Comfort Stent consist of the implantable metallic stent and introducer system.

The stent is made of nitinol wire. It is a flexible, fine mesh tubular prosthesis that has several radiopaque markers. The Niti-S Duodenal Comfort Stent has a diameter of 22 mm while the Niti-S Colonic Comfort Stent has diameters of 22 and 24 mm. They both have lengths of 60, 80, 100, and 120 mm.

The introducer system accepts a 0.035 in (0.89 mm) guidewire. The introducer system is passed over the guidewire and through an endoscope into the duodenum or colon. Use of fluoroscopy is recommended to ensure correct placement of the device.

AI/ML Overview

This document is a 510(k) summary for the Niti-S Duodenal Comfort Stent and Niti-S Colonic Comfort Stent. It primarily focuses on demonstrating substantial equivalence to predicate devices through a comparison of technological characteristics and performance data from bench testing. It does not describe a study involving human subjects or AI.

Here's an analysis of what is present and what is missing based on your request:

1. Table of acceptance criteria and the reported device performance:

The document mentions that "The results of the bench testing show that the subject devices meet their specifications" for various tests. However, it does not provide a specific table outlining the acceptance criteria and the numerical results for each test. Instead, it lists the types of bench tests conducted:

Acceptance Criteria (Implied)Reported Device Performance
Meet specifications for:- Deployment Accuracy
- Deployment Force- Expansion Force
- Compression Force- Dimensional
- Tensile Strength- Shipping
- Shelf-life- Corrosion
- MR Compatibility- Axial Force
- Trackability and Visualization- Repositioning Force
- RadiopacityAll listed tests met specifications and demonstrate substantial equivalence.
SterilityIn accordance with ISO 11135:2014 & AAMI/ANSI/ISO 11607-1:2019, AAMI/ANSI/ISO 11607-2:2019
BiocompatibilityAcceptable biocompatibility profile (evaluated for cytotoxicity, sensitization, irritation/intracutaneous reactivity, acute systemic toxicity, material-mediated pyrogenicity, subacute/subchronic toxicity, chronic toxicity, and implantation effects)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Sample Size for Test Set: Not specified. The document only refers to "bench testing."
  • Data Provenance: Not specified. Bench testing is typically performed in a lab setting, so country of origin isn't directly relevant in the same way as clinical data. This is not a study on patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

Not applicable. This document describes bench testing of a medical device (stent and introducer system), not an AI algorithm or a diagnostic tool requiring expert ground truth in a clinical context.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable. As this is bench testing, there is no expert adjudication of image interpretations or clinical diagnoses.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No. This is a 510(k) submission for a physical medical device (stent), not an AI-powered diagnostic or assistive tool. Therefore, no MRMC study or AI-related effectiveness analysis was conducted or reported here.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

No. This is for a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

Not applicable. For bench testing, the "ground truth" would be the engineering specifications and performance standards that the device is designed to meet. The document states the tests "meet their specifications."

8. The sample size for the training set:

Not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established:

Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 14, 2023

Taewoong Medical Co., Ltd. % Daniel Dillon Senior Regulatory Specialist MED Institute 1330 Win Hentschel Boulevard Suite 100 West Lafayette, Indiana 47906

Re: K223067

Trade/Device Name: Niti-S Duodenal Comfort Stent Niti-S Colonic Comfort Stent Regulation Number: 21 CFR 878.3610 Regulation Name: Esophageal Prosthesis Regulatory Class: II Product Code: MUM (Niti-S Duodenal Comfort Stent) MQR (Niti-S Colonic Comfort Stent) Dated: May 9, 2023 Received: May 10, 2023

Dear Daniel Dillon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Nicholas Clay -S

for Glenn Bell, Ph.D. Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223067

Device Name

Niti-S Duodenal Comfort Stent; Niti-S Colonic Comfort Stent

Indications for Use (Describe)

The Niti-S Duodenal Comfort Stent is indicated for the palliative treatment of pyloric or duodenal obstructions caused by malignant neoplasms.

The Niti-S Colonic Comfort Stent is indicative treatment of colorectal strictures produced by malignant neoplasms and to relieve large bowel obstruction prior to colectomy in patients with malignant strictures.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SUBMITTER 1.

Applicant:Taewoong Medical Co., Ltd.14 Gojeong-ro, Wolgot-myeon, Gimpo-si, Gyeonggi-do,Korea, 10022Phone: +82-31-996-0641Fax: +82-31-996-0645
Contact:Yonjin Jeff KimRegulatory Affairs Manager / R&D and RA DepartmentPhone: +82-31-993-0641-4 (Ext.# 222)Fax: 82-31-996-0645-6E-mail: jinjeff@stent.net
Date Prepared:May 9, 2023
2.DEVICE
Device Trade Names:Device Common Names:Regulation Number:Classification Name:Regulatory Class:Product Codes:Niti-S Duodenal Comfort Stent;Niti-S Colonic Comfort StentDuodenal stent; colonic stent21 CFR 878.3610Esophageal ProsthesisClass IIMUM (Niti-S Duodenal Comfort Stent)MQR (Niti-S Colonic Comfort Stent)

PREDICATE DEVICES 3.

M.I. Tech Co., Ltd, HANAROSTENT LowAx™ Duodenum/Pylorus K183616 M.I. Tech Co., Ltd, HANAROSTENT LowAx™ Colon/Rectum K190141

DEVICE DESCRIPTION 4.

The Niti-S Duodenal Comfort Stent and Niti-S Colonic Comfort Stent consist of the implantable metallic stent and introducer system.

The stent is made of nitinol wire. It is a flexible, fine mesh tubular prosthesis that has several radiopaque markers. The Niti-S Duodenal Comfort Stent has a diameter of 22 mm while the Niti-S Colonic Comfort Stent has diameters of 22 and 24 mm. They both have lengths of 60, 80, 100, and 120 mm.

The introducer system accepts a 0.035 in (0.89 mm) guidewire. The introducer system is passed over the guidewire and through an endoscope into the duodenum or colon. Use of fluoroscopy is recommended to ensure correct placement of the device.

{4}------------------------------------------------

INDICATIONS FOR USE 5.

The Niti-S Duodenal Comfort Stent is indicated for the palliative treatment of pyloric or duodenal obstructions caused by malignant neoplasms.

The Niti-S Colonic Comfort Stent is indicated for the palliative treatment of colorectal strictures produced by malignant neoplasms and to relieve large bowel obstruction prior to colectomy in patients with malignant strictures.

SUBSTANTIAL EQUIVALENCE 6.

The table below compares the key technological features of the subject devices to the predicate devices (M. I. Tech Co., Ltd, HANAROSTENT LowAx™ Colon/Rectum and HANAROSTENT LowAx™ Duodenum/Pylorus).

Niti-S DuodenalComfort Stent(Subject Device)HANAROSTENTLowAX™Duodenum/PylorusStent (K183616)Niti-S ColonicComfort Stent(Subject Device)HANAROSTENTLowAx™Colon/Rectum Stent(K190141)
Stent
Main materialNitinol wireNitinol wireNitinol wireNitinol wire
DesignCross-and-hook weaveusing two wires;one head/flareCross-and-hook weaveusing two wires;one head/flareCross-and-hook weaveusing two wires;one head/flareCross-and-hook weaveusing two wires;one head/flare
BodyDiameters/HeadDiameters22 mm/27 mm22 mm/27 mm• 22 mm/27 mm• 24 mm/29 mm• 22 mm/27 mm• 25mm /30 mm
Lengths60, 80, 100, 120 mm60, 90, 120 mm60, 80, 100, 120 mm60, 90, 120 mm
Polymer cover?NoNoNoNo
RadiopaqueMarkers11 Pt/Ir, 2 STS 316L12 Gold11 Pt/Ir, 2 STS 316L12 Gold
Introducer
DesignCo-axial tubeCo-axial tubeCo-axial tubeCo-axial tube
Method ofplacement anddeploymentEndoscopic; releaseby pulling outer sheathEndoscopic; releaseby pulling outer sheathEndoscopic; releaseby pulling outer sheathEndoscopic; releaseby pulling outer sheath
Diameter10 Fr (3.3 mm)10.2 Fr (3.4 mm)10 Fr (3.3 mm)10.2 Fr (3.4 mm)
Usable Length220 cm230 cm220 cm230 cm
Compatibleguidewire0.035"0.035"0.035"0.035"

7. PERFORMANCE DATA

Sterility and Packaging Testing

The Niti-S Duodenal Comfort Stent and Niti-S Colonic Comfort Stent and introducer are provided as sterile devices. The sterilization method, validation method, and sterility assurance level are in accordance with ISO 11135:2014. "Sterilization of health-care products - Ethylene oxide - Requirement for the development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1(2018)]".

Packaging was designed, tested, and validated in accordance with AAMI/ANSI/ISO 11607-1:2019, "Packaging for terminally sterilized medical devices - Part 1: Requirements for

{5}------------------------------------------------

materials, sterile barrier systems and packaging" and AAMI/ANSI/ISO 11607-2:2019, "Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes".

Biocompatibility Testing

The Niti-S Duodenal Comfort Stent and Niti-S Colonic Comfort Stent were evaluated according to the FDA's guidance "Use of InternationalStandard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluating and testing within a risk management process". The stent and introducer were evaluated for cytotoxicity, sensitization, and irritation/ intracutaneous reactivity effects. The stent was also evaluated for acute systemic toxicity, material-mediated pyrogenicity, subacute/subchronic toxicity, chronic toxicity, and implantation effects. The evaluation determined that the devices have an acceptable biocompatibility profile.

Bench Testing

To establish the substantial equivalence of both Niti-S Duodenal Comfort Stent and Niti-S Colonic Comfort Stent, bench testing was conducted to validate the performance of both devices. The following tests were conducted:

  • Deployment Accuracy .
  • . Deployment Force
  • Expansion Force .
  • Compression Force .
  • Dimensional .
  • . Tensile Strength
  • . Shipping
  • . Shelf-life
  • Corrosion .
  • . MR Compatibility
  • . Axial Force
  • . Trackability and Visualization
  • . Repositioning Force
  • . Radiopacity

The results of the bench testing show that the subject devices meet their specifications and are substantially equivalent to the predicate devices.

CONCLUSION 8.

The subject devices have indications for use that are the same as the predicate devices and technological characteristics that are similar to the predicate devices. The performance data demonstrate that the subject devices meet their specification and do not raise new questions of

{6}------------------------------------------------

K223067
510(k) SummaryPage 4 of 4

safety or effectiveness. These results support a determination of substantial equivalence of the subject devices to the predicate devices.

§ 878.3610 Esophageal prosthesis.

(a)
Identification. An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.(b)
Classification. Class II. The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”