Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K053562

    Validate with FDA (Live)

    Date Cleared
    2006-06-23

    (184 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Collagen Repair Patch is intended for the reinforcement of soft tissues which are repaired by sutures or suture anchors, during rotator cuff repair surgery. Collagen Repair Patch is not intended to replace normal body structure or provide the full mechanical strength to repair the rotator cuff. Sutures used to repair the tear, and sutures or bone anchors used to attach the tissue to the bone, provide biomechanical strength for the rotator cuff repair. Collagen Repair Patch reinforces soft tissue and provides a scaffold that is eventually incorporated into the patients own soft tissue.

    Device Description

    Not Found

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Zimmer® Collagen Repair Patch, a surgical mesh intended for reinforcing soft tissues during rotator cuff repair surgery. However, the document does not contain specific acceptance criteria or detailed study results that would enable the construction of the requested table and the comprehensive answers to the questions about a device performance study.

    Here's a breakdown of what can be extracted and what information is missing:

    Information Present:

    • Device Name: Zimmer® Collagen Repair Patch
    • Intended Use: Reinforcement of soft tissues repaired by sutures or suture anchors, during rotator cuff repair surgery. It's not intended to replace normal body structure or provide full mechanical strength, but to reinforce soft tissue and provide a scaffold for incorporation into the patient's own tissue.
    • Predicate Devices: Permacol® Surgical Implant (K021056), Restore® Orthobiologic Soft Tissue Implant (K031969), CuffPatch™ (K042809).
    • Performance Data Mentioned: "Biocompatibility and bench studies have been completed and support the safety and effectiveness of Zimmer® Collagen Repair Patch for its intended use." It states that "biocompatibility test results show that the material... are non-toxic and non-sensitizing..." and "Test results demonstrate that the materials chosen and the design utilized... will meet the established specification necessary for consistent performance during its intended use."

    Missing Information Crucial for Your Request:

    The document does not provide:

    • Specific acceptance criteria (numerical or descriptive targets).
    • Reported device performance data (e.g., tensile strength, degradation rate, cell infiltration, etc.) quantified with actual values.
    • Sample sizes for any tests mentioned (e.g., number of specimens in bench studies, number of animals in biocompatibility).
    • Data provenance (country of origin, retrospective/prospective for any study). The studies mentioned are general "biocompatibility and bench studies."
    • Number of experts, their qualifications, or adjudication methods. These generally apply to clinical studies involving human interpretation or pathology, which are not detailed here.
    • Information about Multi-Reader Multi-Case (MRMC) comparative effectiveness studies. These are typically for diagnostic devices, not surgical meshes.
    • Information about standalone algorithm performance. This is irrelevant as the device is a physical surgical mesh, not a software algorithm.
    • Type of ground truth used. Similarly, this is generally for diagnostic or prognostic algorithms.
    • Sample size for training sets. Again, irrelevant for a physical medical device.
    • How ground truth for training data was established.

    Based on the provided text, I cannot complete the requested tables and answers comprehensively. The document focuses on regulatory submission (510(k) summary) rather than detailed study results.

    However, I can structure what is available and highlight the gaps:


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document does not explicitly state acceptance criteria or specific numerical performance data. The "Performance Data" section makes general statements about meeting specifications. Therefore, this table cannot be filled with concrete values.

    Acceptance Criteria (e.g., Tensile Strength > X MPa, Biocompatibility: Non-cytotoxic)Reported Device Performance (e.g., Tensile Strength = Y MPa, Biocompatibility: Meets ISO 10993)
    Not specified in the provided text.Biocompatibility: Non-toxic, non-sensitizing to biological tissues. Materials and design meet established specifications for consistent performance.
    (Likely would include mechanical properties, degradation rates, cellular response in vivo/vitro)(Actual quantitative results for the above are not provided.)

    2. Sample size used for the test set and the data provenance

    • Sample Size (Test Set): Not specified. The document mentions "biocompatibility and bench studies" but does not give sample sizes for these tests (e.g., number of samples tested for tensile strength, number of animals for in-vivo biocompatibility).
    • Data Provenance: Not specified. The studies mentioned ("biocompatibility and bench studies") are not described with details about their origin (e.g., country) or whether they were internal lab studies, external contract studies, etc. They are inherently prospective lab/bench studies in nature.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable/Not specified. This information is typically relevant for studies involving human interpretation of data (e.g., radiologists reading images for an AI device). For a physical medical device like a surgical mesh, "ground truth" would be established by scientific measurements and histological analyses, not by expert consensus in the same way. The document does not discuss any such expert panel for the given studies.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable/Not specified. As above, this applies to expert reviews, typically for AI or diagnostic device studies.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The Zimmer® Collagen Repair Patch is a physical surgical mesh, not an AI or diagnostic device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This question pertains to AI algorithms. The device is a physical surgical implant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For "biocompatibility and bench studies," the "ground truth" would implicitly be based on:

    • Standardized test methods: (e.g., ISO standards for cytotoxicity, sensitization, irritation).
    • Quantitative measurements: (e.g., tensile strength, elongation, burst strength for bench studies).
    • Histological assessment: (for tissue response in vivo in biocompatibility studies, typically interpreted by pathologists or toxicologists).

    The document does not detail which specific ground truth metrics were used, only that the material was found to be "non-toxic and non-sensitizing" and met "established specifications."

    8. The sample size for the training set

    Not applicable. There is no "training set" in the context of developing a physical surgical mesh. This term is relevant for machine learning algorithms.

    9. How the ground truth for the training set was established

    Not applicable, for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K050355

    Validate with FDA (Live)

    Date Cleared
    2005-03-09

    (23 days)

    Product Code
    Regulation Number
    884.5980
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Permacol® Surgical Implants are intended for use to support/reinforce soft tissue in surgical procedures. Permacol® Surgical Implant T-pieces are shaped for use in rectal intussusception repair and Permacol® Surgical Implant Rectocele-pieces are shaped for use in rectocele repair.

    Device Description

    Permacol® T-piece and Permacol® Rectocele-piece are substantially equivalent to the predicate devices because they have the same intended uses and very similar technological characteristics.

    AI/ML Overview

    The provided text is a 510(k) summary for the Permacol® Surgical Implant T-piece and Permacol® Surgical Implant Rectocele-piece. It primarily focuses on demonstrating substantial equivalence to predicate devices and does NOT contain information about acceptance criteria and a study proving a device meets those criteria, as typically found in comprehensive performance studies for AI/software devices.

    Therefore, I cannot fulfill the request as the necessary information is not present in the provided text. The document states that "Biocompatibility and bench studies have been completed and support the safety and effectiveness" but does not detail the acceptance criteria or specific performance results of these studies in a format that would allow for the requested table and study description. It also explicitly refers to these studies being incorporated by reference, meaning the details are not contained within this particular submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K043366

    Validate with FDA (Live)

    Date Cleared
    2004-12-17

    (9 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Permacol® Surgical Implant is intended for use as a soft tissue patch to reinforce soft tissue where weakness exists or ruptured soft tissue membranes. It is specifically indicated for the repair of abdominal wall defects and hernias, including but not limited to parastomal hernias.

    Device Description

    Permacol® Surgical Implant Crosslinked Porcine Dermal Collagen Mesh

    AI/ML Overview

    This 510(k) summary (K043366) is for Tissue Science Laboratories, plc's Permacol® Surgical Implant Crosslinked Porcine Dermal Collagen Mesh. The submission primarily focuses on establishing substantial equivalence to predicate devices based on intended use, technological characteristics, and performance data from biocompatibility and bench testing.

    Here's an analysis of the provided information against your requested criteria:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    BiocompatibilityNon-toxic to biological tissues"The biocompatibility test results show that the materials used in the design and manufacture of the device are non-toxic..."
    Non-sensitizing to biological tissues"...non-sensitizing to biological tissues."
    Material PerformanceMaterials consistent with intended use"...new techniques and results demonstrate that the materials are consistent with the intended use."
    Physical PerformanceMeet performance criteria during intended use"Laboratory tests demonstrate that the Permacol® Surgical Implant will meet chosen and the design utilized in manufacturing criteria of performance during their intended use."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify sample sizes for biocompatibility or bench testing. It only generally refers to "test results" and "laboratory tests."

    The data provenance is not explicitly stated as retrospective or prospective, nor does it mention a country of origin for the data. However, the manufacturer is in the UK, suggesting the tests were conducted under their guidance, possibly in the UK or a contract lab.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the given document. Biocompatibility and bench testing usually rely on standardized protocols and laboratory measurements rather than expert human assessment for ground truth.

    4. Adjudication Method for the Test Set:

    This information is not applicable and therefore not provided. Adjudication methods are typically used in clinical studies involving human interpretation or subjective assessment, which is not the case for the described biocompatibility and bench tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

    No MRMC comparative effectiveness study was mentioned or indicated. The submission focuses on substantial equivalence based on material and mechanical properties, not clinical efficacy comparisons with human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    Not applicable. This device is a surgical mesh, not a medical imaging or AI diagnostic algorithm. Therefore, "standalone algorithm performance" is not relevant.

    7. The Type of Ground Truth Used:

    For biocompatibility, the ground truth is established by standardized biological assays and observations for cellular response, toxicity, and sensitization, compared against established biological safety limits.

    For physical performance and material characterization (bench testing), the ground truth is established by objective measurements of properties like tensile strength, elasticity, pore size, degradation rate, etc., compared against pre-defined engineering specifications or comparisons to predicate devices' known properties.

    8. The Sample Size for the Training Set:

    Not applicable. This device is a physical medical implant, not an AI algorithm. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. As there is no training set for a physical device, there is no ground truth to establish for it.


    Summary of Device Performance Study:

    The provided 510(k) summary indicates that the substantial equivalence of the Permacol® Surgical Implant is supported by:

    • Biocompatibility testing: This demonstrated that the materials used are non-toxic and non-sensitizing. The "ground truth" here is the biological response observed in standardized tests, confirming the material's safety for biological contact.
    • Bench testing (laboratory tests): These tests confirmed that the device's materials are "consistent with the intended use" and that the product "will meet chosen and the design utilized in manufacturing criteria of performance during their intended use." The "ground truth" here is objective physical and mechanical measurements compared against predetermined specifications and predicate device characteristics.

    The submission does not include human clinical trial data, MRMC studies, or AI algorithm performance data, as these are not relevant to the type of device and the pathway chosen for regulatory clearance (510(k) substantial equivalence). The focus is on the device's physical and biological properties.

    Ask a Question

    Ask a specific question about this device

    K Number
    K021056

    Validate with FDA (Live)

    Device Name
    PERMACOL
    Date Cleared
    2002-10-31

    (213 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Permacol™ - Crosslinked Porcine Dermal Collagen Surgical Mesh ("Permacol™ ") is intended to be used for reinforcement of the soft tissues which are repaired by suture or suture anchors limited to the supraspinatus during rotator cuff repair surgery.

    Device Description

    Permacol™ is available in sheet form.

    AI/ML Overview

    The provided text describes a 510(k) submission for a surgical mesh, Permacol™ - Crosslinked Porcine Dermal Collagen Surgical Mesh. The submission focuses on demonstrating substantial equivalence to predicate devices rather than a study proving the device meets specific acceptance criteria based on detailed performance metrics.

    Here's an analysis based on the information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document states "Mechanical testing of Permacol™ has been performed in accordance with FDA's Surgical Mesh 510(k) Guidance. The results demonstrate that Permacol™ provides appropriate tensile strength, elasticity, stiffness, suture pullout strength, tear resistance, and puncturability, for use in soft tissue repair."

    However, specific quantitative acceptance criteria (e.g., minimum tensile strength of X MPa, elasticity range of Y-Z%) or the exact reported performance values are NOT provided in the submission summary. The FDA's 510(k) summary only states that the device "provides appropriate" values, implying that it met internal or guidance-recommended criteria without listing them.

    Acceptance Criteria (Quantitative)Reported Device Performance (Quantitative)
    Not explicitly stated in the 510(k) summary (e.g., specific thresholds for tensile strength, elasticity)Not explicitly stated in the 510(k) summary (e.g., actual measured tensile strength, elasticity values)
    Qualitative statements from the document:Qualitative statements from the document:
    Appropriate tensile strengthPermacol™ provides appropriate tensile strength
    Appropriate elasticityPermacol™ provides appropriate elasticity
    Appropriate stiffnessPermacol™ provides appropriate stiffness
    Appropriate suture pullout strengthPermacol™ provides appropriate suture pullout strength
    Appropriate tear resistancePermacol™ provides appropriate tear resistance
    Appropriate puncturabilityPermacol™ provides appropriate puncturability

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document does not specify the sample size used for the mechanical testing.
    • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. It only mentions "Mechanical testing of Permacol™ has been performed..."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable as the "test set" in this context refers to mechanical properties of a medical device, not a diagnostic or clinical performance study requiring expert ground truth for interpretation (e.g., radiology images or pathology slides). The assessment is based on physical material properties.

    4. Adjudication Method for the Test Set:

    This information is not applicable for the same reasons as #3. Mechanical testing results are objective measurements, not subject to expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret cases with and without AI assistance. This submission is for a surgical mesh, a physical implant.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    This information is not applicable. The device is a surgical mesh, not an algorithm. The performance described relates to the physical characteristics of the mesh itself.

    7. The Type of Ground Truth Used:

    The "ground truth" here would be the objective measurements of the physical and mechanical properties of the Permacol™ mesh. This would involve laboratory-based testing methods to determine:
    * Tensile strength
    * Elasticity
    * Stiffness
    * Suture pullout strength
    * Tear resistance
    * Puncturability

    These are intrinsic material properties and do not rely on expert consensus, pathology, or outcomes data in the way a diagnostic device would.

    8. The Sample Size for the Training Set:

    This information is not applicable. The Permacol™ mesh is a manufactured medical device, not an AI or algorithm that requires a "training set" in the computational sense. The document refers to its manufacturing process and testing of its final product properties.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013625

    Validate with FDA (Live)

    Device Name
    PERMACOL
    Date Cleared
    2002-01-17

    (73 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Permacol™ is indicated for use to provide soft tissue repair or reinforcement in plastic and reconstructive surgery of the face and head.

    Permacol™ tissue is intended for use as a soft tissue patch to reinforce soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue in plastic and reconstructive surgery of the face and head.

    Device Description

    Permacol™ consists of a sterile, tough, off-white, moist, flexible, fibrous flat sheet comprised of acellular crosslinked porcine dermal collagen and elastin. Each impermeable inner pack contains a small quantity of saline to maintain its moistness and flexibility. The product is available in thicknesses of 0.5-1.5 mm and dimensions of 0.8 x 0.8 to 4 x 6 inches.

    AI/ML Overview

    The provided 510(k) summary for Permacol™ Crosslinked Porcine Dermal Collagen Surgical Mesh does not contain the information requested regarding acceptance criteria or a study proving the device meets said criteria.

    The document is a premarket notification for a Class II medical device, essentially asserting its substantial equivalence to previously cleared predicate devices. It focuses on:

    • Device Identification and Classification: Naming the device, its sponsor, and its classification.
    • Predicate Devices: Listing similar devices already on the market.
    • Intended Use: Stating the purpose of the device (soft tissue repair/reinforcement in plastic and reconstructive surgery of the face and head).
    • Technological Characteristics: Describing the material, form, and how it is used.
    • Summary Basis for Substantial Equivalence: Arguing that the new device is sufficiently similar to existing ones and does not raise new questions of safety or effectiveness. This is the core of a 510(k) submission.

    The information about specific acceptance criteria, performance data from a study, sample sizes, expert involvement, ground truth, or comparative effectiveness studies (MRMC) is typically not included in a 510(k) summary. These details would be part of a more extensive submission to the FDA, which is not publicly available in this format, or would appear in a full study report.

    Therefore, for each of your requested points, the answer based solely on the provided text is:

    1. A table of acceptance criteria and the reported device performance: Not provided. The document states "these differences do not raise any new questions of safety or effectiveness, as confirmed by preclinical and clinical evaluation," but does not give specific criteria or performance metrics.
    2. Sample sizes used for the test set and the data provenance: Not provided. The document mentions "preclinical and clinical evaluation" in a general sense but gives no details on studies, test sets, or data provenance.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not provided.
    4. Adjudication method for the test set: Not provided.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document describes a surgical mesh, not an AI-assisted diagnostic or imaging device.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithmic device.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not provided.
    8. The sample size for the training set: Not applicable (not an AI/ML device) and not provided if referring to a general study.
    9. How the ground truth for the training set was established: Not applicable and not provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K992556

    Validate with FDA (Live)

    Device Name
    PERMACOL
    Date Cleared
    2000-02-17

    (202 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Permacol™ is intended for use as a soft tissue patch to reinforce soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue membranes. It is specifically indicated for use in the following types of soft tissue repair procedures: abdominal, inquinal, diaphragmatic, femoral, scrotal, . umbilical, and incisional hernias; . colon, rectal, urethral, and vaginal prolapse: muscle flap reinforcement; . . reconstruction of the pelvic floor; sacrocolposuspension; and . . urethral sling.

    Device Description

    Permacol™ is a sterile, tough. off-white. moist. flexible, fibrous flat sheet solely comprised of acellular crosslinked porcine dermal collagen and elastin

    AI/ML Overview

    The prompt requests information on acceptance criteria and study details for a device, but the provided text describes a surgical mesh (Permacol™) and its 510(k) premarket notification. This type of medical device submission typically focuses on demonstrating substantial equivalence to existing predicate devices rather than establishing specific quantitative acceptance criteria or conducting studies as would be done for a novel diagnostic algorithm.

    Therefore, many of the requested data points (such as sample sizes for test sets, number of experts, adjudication methods, MRMC studies, standalone algorithm performance, and training set details) are not applicable or available in this type of submission. The provided text outlines biocompatibility testing and clinical experience used to support substantial equivalence and safety/effectiveness, but not a controlled clinical trial with specific numerical outcome measures against acceptance criteria in the way a diagnostic AI would have.

    However, I can extract the information that is present and explain where the requested details are not provided.

    Here's a breakdown based on the provided text:


    Acceptance Criteria and Device Performance for Permacol™ Porcine Collagen Surgical Mesh

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria Category (as implied by FDA submission)Permacol™ Porcine Collagen Surgical Mesh Performance
    BiocompatibilityThe product passed all standard biocompatibility tests: Hemolysis, Cytotoxicity (elution method), Acute Systemic Toxicity, Implantation/Subchronic Toxicity, Genotoxicity (Ames Mutagenesis, Bone Marrow Micronucleus, Mouse Lymphoma Mutagenesis), Sensitization (Magnusson and Kligman), Pyrogenicity.
    Histological EvaluationHistological evaluation after 6 months implantation in rats demonstrated the material is safe and well tolerated.
    Clinical Safety & Tolerance"Clinical experience in several hundred patients in Europe also has demonstrated that Permacol™ is safe and well tolerated."
    Clinical Outcomes (Effectiveness)Case histories for 60 patients showed "successful clinical outcomes were achieved in nearly all cases."
    Adverse Effects"No product-related adverse effects were noted in any of the procedures."
    Ease of HandlingCase histories for 60 patients demonstrated Permacol™ is "easy to handle."
    Substantial Equivalence (Overall)"Permacol™ is substantially equivalent to the predicate devices in terms of its intended use, specific indications, and performance." Minor technological differences "do not raise any new concerns of either safety or effectiveness."

    2. Sample size used for the test set and the data provenance

    • Sample Size for Human Clinical Experience: "several hundred patients in Europe" for general clinical safety and tolerance.
    • Sample Size for Detailed Case Histories: 60 patients for detailed evaluation of handling and clinical outcomes.
    • Data Provenance: Clinical experience in Europe (specifically mentioned). The studies referenced appear to be retrospective collections of "case histories" rather than a prospective, randomized controlled trial.
    • Animal Study Sample Size: Not explicitly stated, but mentioned as "implantation in rats."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable / Not provided: This type of information is typically associated with diagnostic device evaluations, especially those involving expert interpretation of medical images or data. For a surgical mesh, "ground truth" would be established through direct clinical observation of patient outcomes and histological analysis, not through expert consensus on diagnostic interpretations. The document references "clinical experience" and "histological evaluation" which implies medical professionals (surgeons, pathologists) were involved, but the specific number or qualifications are not detailed as they would be for a diagnostic study.

    4. Adjudication method for the test set

    • Not Applicable / Not provided: Adjudication methods (e.g., 2+1, 3+1) are typically used in diagnostic studies where multiple readers interpret data and discrepancies need to be resolved. This is not relevant for the type of clinical experience and biocompatibility testing described for a surgical mesh.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: An MRMC comparative effectiveness study, especially concerning AI assistance, is not relevant to a surgical mesh like Permacol™. This device is not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable: Permacol™ is a physical surgical mesh, not an algorithm. Standalone algorithm performance is not relevant.

    7. The type of ground truth used

    • Biocompatibility Studies: Laboratory test results (e.g., cell viability, hemolytic index, mutagenicity assays) and histological findings from animal implantation.
    • Clinical Experience: Direct observation of patient safety (adverse events), tolerance, and surgical outcomes (e.g., successful repair, ease of handling). These would be derived from clinical records and surgeon's assessments.

    8. The sample size for the training set

    • Not Applicable: Permacol™ is a manufactured device, not an AI model that requires a "training set."

    9. How the ground truth for the training set was established

    • Not Applicable: As there is no AI model or training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1