    K Number
    K233191

    Validate with FDA (Live)

    Date Cleared
    2023-11-21

    (54 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TotalTi ACDF by SAGICO is an interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from (C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radios. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. The TotalTi ACDF by SAGICO is to be filled with autogenous bone graft material and implanted via an open anterior approach and deployment of the internal blades. The TotalTi ACDF by SAGICO is designed in a manner to be used with the additional fixation (e.g. anterior plate or cervical pedicle screws) cleared by the FDA to properly utilize this device.

    Device Description

    The TotalTi ACDF by SAGICO, is a spinal system of Interbody Fusion (IBF) devices used to provide structural stability in skeletally mature individuals following discectomy. The TotalTi ACDF by SAGICO, is intended to be used between two contiguous levels from C2 to T1 and placed via an open anterior surgical approach. The implants are customizable additively manufactured and available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. The TotalTi ACDF by SAGICO, incorporates a vertical cavity (a large central hollowed window) designed to be packed with and filled with autogenous bone graft material to promote fusion. The TotalTi ACDF by SAGICO implant requires the internal fixation blades to be deployed and is designed to be used with FDA cleared supplemental fixation to properly utilize the device. The TotalTi ACDF by SAGICO implants features protrusions located on the top and bottom surfaces to engage with superior and inferior endplates of the adjacent vertebrae to resist rotational and expulsion.

    MATERIALS: The TotalTi ACDF by SAGICO are additively manufactured implants from titanium alloy Ti-6Al-4V ELI. The TotalTi ACDF by SAGICO includes additively manufactured spacer, integrated fixation anchors per ASTM F3001and traditionally machined titanium alloy anchors lock per ASTM F136.

    FUNCTION: The TotalTi ACDF by SAGICO implants are intervertebral body fusion devices to help restore integrity to the spine in the cervical region.

    AI/ML Overview

    This is a medical device 510(k) summary for the TotalTi ACDF by SAGICO, an intervertebral body fusion device. The document describes the device, its indications for use, and non-clinical performance data. Since this is a submission for an intervertebral body fusion device and not an AI/ML powered device, the provided document does not contain information about acceptance criteria, study details, sample sizes, ground truth establishment, or expert involvement as these are typically associated with performance studies for AI/ML devices. As a result, I cannot provide an answer based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180220

    Validate with FDA (Live)

    Date Cleared
    2018-05-21

    (116 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SAGICO OSI SPINAL SYSTEM by Osimplant is intended for use in the non-cervical spine.

    When used for anterior/anterolateral and posterior, non-cervical pedicle fixation system, the SAGICO OSI SPINAL SYSTEM by Osimplant is intended to provide additional support during fusion using autograft or allograft in Skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities:

    • · Degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);
    • · Spondylolisthesis;
    • · Trauma (i.e. fracture or dislocation);
    • · Spinal stenosis;
    • · Curvatures (i.e. scoliosis, kyphosis, and/or lordosis);
    • · Tumor:
    • · Pseudoarthrosis; and
    • Failed previous fusion

    When used for posterior non-cervical pedicle screw fixation in pediatric patients, the SAGICO OSI SPINAL SYSTEM by Osimplant implants are indicated as an adjunct to treat adolescent idiopathic scoliosis.

    The SAGICO OSI SPINAL SYSTEM by Osimplant for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description

    The SAGICO OSI SPINAL SYSTEM by Osimplant constructs of monoaxial screws, uniplanar screws, polyaxial screws, reduction screws, locking cap set screws, rods, hooks, monoaxial and muttiaxial transverse connectors and associated surgical instruments. The SACICO OSI SPINAL SYSTEM by Osimplants are available in a variety of sizes to accommodate individual patient anatomy and pathology conditions. SAGICO OSI SPINAL SYSTEM by Osimplant implants are designed to adapt 5.5mm diameter rods; the implants are manufactured from Ti6A14V alloy and offered in a sterile package option.

    AI/ML Overview

    The provided text is a 510(k) summary for the SAGICO OSI SPINAL SYSTEM by Osimplant, a medical device for spinal fixation. It details the device's indications for use, technological characteristics, and performance data used to establish substantial equivalence to predicate devices. However, this document does not describe an AI/ML device or algorithm. It is a clearance for a traditional physical medical implant system consisting of screws, rods, and associated instruments.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert adjudication, or comparative effectiveness studies for an AI device cannot be extracted from this document, as it pertains to a mechanical spinal implant system.

    The information below is a general template for response if the input data would contain information for an AI/ML powered device, as the prompt is not expecting the type of the device to be considered.

    This document does not describe an AI/ML powered medical device. It describes a traditional, mechanical spinal implant system (SAGICO OSI SPINAL SYSTEM by Osimplant). Therefore, the specific details requested regarding AI/ML device performance, acceptance criteria, sample sizes for AI/ML models, expert ground truth establishment, adjudication methods, MRMC studies, or standalone algorithm performance are not available in this document.

    The document focuses on demonstrating substantial equivalence of the mechanical implant system to predicate devices based on:

    • Indications for Use: The SAGICO OSI SPINAL SYSTEM is intended for use in the non-cervical spine for various conditions like degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis, and failed previous fusion in skeletally mature patients. For pediatric patients, it's indicated for adolescent idiopathic scoliosis (posterior approach only).
    • Technological Characteristics: The device consists of monoaxial screws, uniplanar screws, polyaxial screws, reduction screws, locking cap setscrews, rods, hooks, monoaxial and multiaxial transverse connectors, and associated surgical instruments. Implants are made from Ti6Al4V alloy and designed for 5.5mm diameter rods.
    • Performance Data (Non-clinical):
      • ASTM F1717-04: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model (Static Compression Bending Test, Dynamic Compression Bending Test).
      • ASTM F1798-13: Standard Guide for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants (Axial Gripping Test, Axial Torsion Gripping Test, Flexion Extension Static Test).
    • Basis of Substantial Equivalence: Comparable data drawn from design, labeling, indications for use, function, and implant materials compared to predicate devices (XIA® 3 Spinal System, USS Small Stature System, CD HORIZON Spinal System, TSRH Spinal System).

    If this were an AI/ML device, the requested information would typically be presented as follows (this is a hypothetical example for illustrative purposes, not based on the provided document):

    1. Table of Acceptance Criteria and Reported Device Performance (Hypothetical for an AI/ML device)

    Performance MetricAcceptance Criteria (e.g., for detecting spinal fractures)Reported Device Performance (e.g., from a clinical study)
    Sensitivity≥ 90%92.5% (95% CI: 90.1% - 94.4%)
    Specificity≥ 80%85.3% (95% CI: 83.0% - 87.3%)
    AUC≥ 0.900.93

    2. Sample Size and Data Provenance for Test Set (Hypothetical)

    • Sample Size: 500 spinal imaging studies (e.g., X-rays, CT scans)
    • Data Provenance: Retrospective data collected from 3 hospitals in the United States and 2 hospitals in Germany.

    3. Number and Qualifications of Experts for Ground Truth (Hypothetical)

    • Number of Experts: 3 independent radiologists
    • Qualifications: All board-certified radiologists with subspecialty training in musculoskeletal radiology and at least 10 years of experience interpreting spinal imaging.

    4. Adjudication Method for Test Set (Hypothetical)

    • Adjudication Method: 2+1. Initial assessment by two radiologists. Discrepancies were resolved by a third, senior radiologist.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study (Hypothetical)

    • Was an MRMC study done? Yes.
    • Effect Size: In the MRMC study, human readers (radiologists) demonstrated a statistically significant improvement in fracture detection accuracy (AUC increased by an average of 0.05, p < 0.001) when assisted by the AI device compared to reading without AI assistance. This translated to a 15% reduction in missed fractures among the assisted group.

    6. Standalone Algorithm Performance (Hypothetical)

    • Was a standalone study done? Yes. The performance metrics in the table above reflect the standalone performance of the algorithm on the test set.

    7. Type of Ground Truth Used for Test Set (Hypothetical)

    • Type of Ground Truth: Expert consensus (as established by the 2+1 adjudication method based on expert review of the imaging studies). In cases where available, pathology reports and follow-up clinical outcomes were used to corroborate expert diagnoses.

    8. Sample Size for Training Set (Hypothetical)

    • Sample Size: 5,000 spinal imaging studies (e.g., X-rays, CT scans).

    9. How Ground Truth for Training Set was Established (Hypothetical)

    • Ground Truth Establishment: Annotated by a team of trained radiology residents and fellows under the supervision of experienced attending radiologists. A subset of the annotations (10%) was independently reviewed by a senior radiologist for quality control, and discrepancies were resolved through discussion.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161710

    Validate with FDA (Live)

    Date Cleared
    2017-06-29

    (373 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SAGICO IBF System - ARIA Lumbar implants are interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).

    SAGICO IBF System - ARIA Lumbar are to be filled with autogenous bone graft material. The SAGICO IBF System -ARIA Lumbar implant is intended to be used with supplemental fixation cleared by the FDA to properly utilize this device.

    The SAGICO IBF System - ARION Cervical implant is interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from (C2-T1). DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and at least six (6) weeks of non-operative treatment.

    The SAGICO IBF System - ARION Cervical implant is to be filled with autogenous bone graft material and placed via open anterior approach. The SAGICO IBF System – ARION Cervical implant must be used with additional internal fixation (e.g. anterior plate or cervical pedicle screws) cleared by the FDA to properly utilize this device.

    Device Description

    SAGICO IBF System - ARIA Lumbar implants are interbody fusion devices used to provide structural stability in skeletally mature individuals following discectomy. SAGICO IBF System - ARIA Lumbar implants are provided in different shapes to accommodate various surgical approaches to the lumbar spine (posterior, transforaminal [posterolateral] or lateral and can expand to the desired height. The SAGICO IBF System - ARIA Lumbar implants are available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. The SAGICO IBF System - ARIA Lumbar implants are to be filled with autogenous bone graft material. Protrusions are located on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebrae to resist expulsion.

    SAGICO IBF System - ARION Cervical implants are interbody fusion devices used to provide structural stability in skeletally mature individuals following discectomy. SAGICO IBF System - ARION Cervical implant is intended to be used between two contiguous levels from C2 to T1 and placed via an open anterior surgical approach. The implants are available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. SAGICO IBF System - ARION Cervical implants are comprised of PEEK Optima LT1 and include large central hollow window to be filled with autogenous bone graft material. SAGICO IBF System - ARION Cervical implants features protrusions located on the top and bottom surfaces of PEEK spacers to engage with superior and inferior endplates of the adjacent vertebrae to resist rotational and expulsion. SAGICO IBF System - ARION Cervical includes internal titanium spin blade offering internal fixation and expansion of 1mm in height allowing low profile insertion. SAGICO IBF System - ARION Cervical implant is intended to be used with FDA cleared supplemental fixation to properly utilize the device.

    The SAGICO IBF System ARIA Lumbar and ARION Cervical implants are manufactured from medical Grade PEEK (Polyetheretherketone) OPTIMAT LT I (Invibio"") per ISO 10993-1 USP Class VI, and ASTM F2026, Titanium Alloy per ASTM F136 and Tantalum beads /rods to be Grade IINS R05200, IINS R05400 according to ASTM F560.

    The SAGICO IBF System ARIA Lumbar and ARION Cervical implants are intervertebral body fusion devices to help restore integrity to the spine in the cervical and lumbar regions.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the SAGICO IBF System - ARIA Lumbar and ARION Cervical devices, based on the provided document:

    This document describes a 510(k) premarket notification for intervertebral body fusion devices, which means the approval process focuses on demonstrating substantial equivalence to already legally marketed devices, rather than establishing a completely new safety and effectiveness profile through extensive clinical trials. Therefore, the "study" referred to is a non-clinical performance study comparing the device to predicates, not a clinical trial with human subjects.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/StandardAcceptance Criteria (Implicit)Reported Device Performance
    Mechanical/PhysicalASTM F2077 (Standard Test Methods for Intervertebral Body Fusion Devices)Meet mechanical performance requirements for intervertebral body fusion devices.Results demonstrate performance substantially equivalent to legally marketed predicate devices.
    - Static and Dynamic Compression TestWithstand static and dynamic compressive forces without failure.Demonstrated substantial equivalence to predicate devices.
    - Static and Dynamic Compression Shear TestWithstand static and dynamic shear forces without failure.Demonstrated substantial equivalence to predicate devices.
    - Static and Dynamic Torsion TestWithstand static and dynamic torsional forces without failure.Demonstrated substantial equivalence to predicate devices.
    ASTM F2267-04 (Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device under Static Axial Compression)Exhibit resistance to subsidence under static axial compression within acceptable limits.Demonstrated substantial equivalence to predicate devices.
    ASTM F04.25.02.02 (Static Push-out Test)Resist expulsion forces from the vertebral body.Demonstrated substantial equivalence to predicate devices.
    BiocompatibilityISO 10993-1 USP Class VI, and ASTM F2026 (for PEEK Optima LT1)Biocompatible and safe for implantation.Manufactured from medical Grade PEEK OPTIMAT LT I (Invibio™) per specified standards, implying compliance.
    ASTM F136 (for Titanium Alloy)Biocompatible and safe for implantation.Manufactured from Titanium Alloy per specified standard, implying compliance.
    ASTM F560 (for Tantalum beads/rods)Biocompatible and safe for implantation.Manufactured from Tantalum beads/rods per specified standard, implying compliance.

    Note on Acceptance Criteria: For a 510(k), explicit numerical acceptance criteria for each test are often not provided in the summary document but are implicitly understood to be meeting or exceeding the performance of the chosen predicate devices according to recognized standards. The "reported device performance" is essentially that the device performed substantially equivalently to the predicates in these tests.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify exact sample sizes for the non-clinical tests (e.g., how many devices were tested for compression, shear, etc.). It refers to "non-clinical performance data testing."
    • Data Provenance: This is a non-clinical, in-vitro study. Therefore, there is no country of origin for patient data. The data is generated from laboratory testing of the devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This question is not applicable in the context of this 510(k) submission. "Ground truth" usually refers to clinical diagnoses or outcomes established by medical experts for a diagnostic or AI-driven system. Here, the "truth" is established by adherence to engineering and biocompatibility standards.

    4. Adjudication Method for the Test Set

    • Not applicable. This is a non-clinical engineering study, not a clinical study requiring adjudication of expert opinions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a mechanical intervertebral body fusion device, not a diagnostic or AI-driven imaging interpretation system. MRMC studies are not relevant here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical implant device, not an algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" for these types of devices is established by engineering standards and material specifications. The device's performance is compared against the requirements outlined in standards like ASTM F2077, F2267-04, F04.25.02.02, and material standards (ISO 10993-1, ASTM F2026, F136, F560). The "truth" is that the device meets these established performance and material safety benchmarks, demonstrating substantial equivalence to predicate devices that also meet these benchmarks.

    8. The Sample Size for the Training Set

    • Not applicable. There is no concept of a "training set" for a physical medical implant device in this context. This is not an AI/machine learning application.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as there is no training set for a physical implant device.

    Summary of the Study:

    The study described is a non-clinical performance evaluation focusing on the mechanical and material characteristics of the SAGICO IBF System - ARIA Lumbar and ARION Cervical devices. The goal was to demonstrate substantial equivalence to legally marketed predicate devices, a requirement for 510(k) clearance by the FDA. The testing included:

    • Mechanical Testing: According to ASTM F2077 (Static and Dynamic Compression, Compression Shear, Torsion) and ASTM F2267-04 (Subsidence), and ASTM F04.25.02.02 (Static Push-out Test).
    • Material Biocompatibility: Assessed by the use of medical-grade materials (PEEK Optima LT1, Titanium Alloy, Tantalum beads/rods) that comply with established international and ASTM standards (ISO 10993-1 USP Class VI, ASTM F2026, ASTM F136, ASTM F560).

    The conclusion of this non-clinical testing was that the SAGICO IBF System devices demonstrated performance substantially equivalent to legally marketed predicate devices, thus supporting their clearance for market.