510(k) Data Aggregation

    K Number
    K220716

    Validate with FDA (Live)

    Device Name
    Spartan SC 069
    Date Cleared
    2022-05-05

    (55 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spartan SC 069TM is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

    Device Description

    The Spartan SC 069™ is a support catheter designed for the introduction of interventional devices into the peripheral and neuro vasculature. The SC 069™ is a single lumen, flexible, variable stiffness composite catheter with a Nitinol structure. A radiopaque marker band on the distal tip of the device is used for visualization under fluoroscopy. The distal section of the catheter is coated with a hydrophilic coating to reduce the overall frictional force during intravascular use. The device is supplied sterile and intended for single use only.

    AI/ML Overview

    The provided text describes the performance data for the Spartan SC 069™ support catheter. This device appears to be a medical device, and the information presented is from a 510(k) premarket notification to the FDA, which focuses on demonstrating substantial equivalence to a predicate device, rather than a clinical study evaluating the device's diagnostic performance. Therefore, many of the requested fields regarding diagnostic acceptance criteria, multi-reader multi-case studies, and human-in-the-loop performance are not applicable.

    Here's the information extracted from the document:

    1. A table of acceptance criteria and the reported device performance

    The document lists various non-clinical bench tests and biocompatibility tests, with each test summary implying a pass/fail acceptance criterion. The results consistently state that the Spartan SC 069™ "met the acceptance criteria" or showed "acceptable" performance, or similar to the predicate device.

    TestTest Method Summary (Implied Criteria)Reported Device Performance
    Coating Lubricity (Friction Force)Acceptable friction force under simulated use conditions similar to predicate.The Spartan SC 069™ was found to have acceptable friction force under simulated use conditions similar to the predicate device.
    Hub Functional & DimensionalMeet ISO 594-1:1986-06-15 and ISO 594-2:1998-09-01 requirements.The Spartan SC 069™ met the acceptance criteria for hub functional and dimensional requirements.
    Torque StrengthAcceptable torsional strength similar to predicate in simulated path model.The Spartan SC 069™ exhibited acceptable torsional strength similar to the predicate device.
    TensileMeet ISO 10555-1:2013 Annex B.The Spartan SC 069™ met the acceptance criteria for tensile strength.
    Air AspirationMeet ISO 10555-1:2013(E) Annex D for air leakage.The Spartan SC 069™ met the acceptance criteria for air aspiration.
    Liquid LeakMeet ISO 10555-1:2013(E) Annex C.The Spartan SC 069™ met the acceptance criteria for liquid leakage.
    Particulate and Coating IntegrityAcceptable particulate generation and coating integrity, substantially equivalent to predicate.The Spartan SC 069™ met the acceptance criteria for particulate generation and coating integrity, and was found substantially equivalent to the predicate.
    StiffnessMeet acceptance criteria for catheter stiffness, substantially equivalent to predicate.The Spartan SC 069™ met the acceptance criteria for catheter stiffness and found substantially equivalent to the predicate.
    Kink ResistanceMeet acceptance criteria for kink resistance, substantially equivalent to predicate.The Spartan SC 069™ met the acceptance criteria for kink resistance and was found substantially equivalent to the predicate.
    RadiopacityMeet acceptance criteria for radiopacity, substantially equivalent to predicate.The Spartan SC 069™ met the acceptance criteria for radiopacity and was found substantially equivalent to the predicate.
    CorrosionMeet ISO 10555-1:2013(E) Annex A.The Spartan SC 069™ met the acceptance criteria for corrosion resistance.
    Static Burst PressureMeet ISO 10555-1:2013(E) Annex F for burst pressure.The Spartan SC 069™ met the acceptance criteria for static burst pressure.
    BucklingMeet acceptance criteria for tip deflection and buckling, substantially equivalent to predicate.The Spartan SC 069™ met the acceptance criteria for tip deflection and buckling and was found substantially equivalent to the predicate.
    Design ValidationMeet requirements for interventional device introduction, trackability, stability, guidewire interaction, and overall device integrity in a challenging use setting.The Spartan SC 069™ met the requirements for which it was designed and tested.
    Dimensional TestingMeet device specifications.The Spartan SC 069™ met the dimensional requirements.
    Visual InspectionMeet acceptance criteria.The Spartan SC 069™ met the visual inspection requirements.
    Dead Space VolumeMeasured and reported in labeling.The Spartan SC 069™ dead space volume was measured. The dead space volume is reported in the labeling.
    Pressure Flow Testing with Contrast MediaEvaluated flow rates and pressures of various solutions; reported in labeling.The Spartan SC 069™ pressure-flow performance was evaluated. Flow rates and pressures of various saline and contrast media solutions are reported in the labeling.
    Biocompatibility Tests
    Sensitization (Guinea Pig)No reaction at challenge; classified as non-sensitizer per ISO 10993-10.Non-sensitizer.
    Genotoxicity (Mouse Lymphoma)IMF of Test Article < GEF; considered non-mutagenic per ISO 10993-3.Non-mutagenic.
    Genotoxicity (Salmonella Typhimurium & Escherichia Coli)No statistically significant increase in revertant colonies compared to negative controls per ISO 10993-3.Non-genotoxic.
    Cytotoxicity (L929 MEM Elution)No cell lysis or reduction of cell growth at 24/48 hours; considered non-cytotoxic per ISO 10993-5.Non-cytotoxic.
    Irritation/Intracutaneous ReactivityNo significantly greater biological reaction than control per ISO 10993-10.Non-irritant.
    Acute Systemic ToxicityNo significantly greater biological reaction than control extracts per ISO 10993-11.No systemic toxicity.
    Pyrogenicity (Rabbit Pyrogen Test)Temperature increases did not exceed test limit; considered non-pyrogenic per ISO 10993-11.Non-pyrogenic.
    Hemocompatibility (Rabbit Blood Hemolysis)Hemolysis < 5% (direct and indirect contact) per ISO 10993-4.Non-hemolytic.
    Hemocompatibility (Partial Thromboplastin Time)Test article clotting time not significantly different from comparison article per ASTM International F2382-18 and ISO 10993-4.Hemocompatible.
    Hemocompatibility (SC5B-9 Complement Activation)No statistically significant increase in SC5b-9 concentrations compared to controls per ISO 10993-4.Hemocompatible.
    Hemocompatibility (Platelet and Leukocyte Binding)No effect on platelet and leukocyte concentrations per ISO 10993-4.Hemocompatible.
    Sterilization
    Sterilization ValidationAchieved a Sterility Assurance Level of 10-6 using overkill (half cycle) approach per ISO 11135:2014.Successful validation.
    Shelf Life
    Accelerated AgingMaintained safety and effectiveness after 2-year equivalent accelerated aging (55 °C, 75 days) and simulated shipping.Confirmed 2-year shelf life.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document relates to non-clinical bench testing and biocompatibility assessments of a medical device (catheter). It does not involve a "test set" in the context of diagnostic accuracy, human subjects, or data provenance like country of origin. The tests were performed in a laboratory setting for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the study involves bench testing and biocompatibility of a medical device, not diagnostic performance requiring expert ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as the study involves bench testing and biocompatibility of a medical device, not diagnostic performance requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This is a 510(k) submission for a physical medical device (catheter), not an AI-assisted diagnostic tool. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This is a 510(k) submission for a physical medical device (catheter), not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for these performance studies is established by the specifications and standards (e.g., ISO standards, internal design requirements) that the device must meet, as well as comparison to the predicate device. For biocompatibility, it's about not eliciting adverse biological responses.

    8. The sample size for the training set

    This information is not applicable. There is no "training set" as this is not an AI/machine learning study.

    9. How the ground truth for the training set was established

    This information is not applicable. There is no "training set" as this is not an AI/machine learning study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220075

    Validate with FDA (Live)

    Date Cleared
    2022-04-19

    (99 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spartan Center Wire™ is indicated for general intravascular use in the peripheral vasculature. It can be deployed into the vessel to facilitate the selective placement of diagnostic or therapeutic catheters.

    The device is not intended for use in the coronary or neuro vasculature.

    Device Description

    The Spartan Center Wire™ is a stabilizing exchange wire designed for temporary assistance in catheter exchange. The device is comprised of a tubular mesh tip attached to a pusher wire of adequate length to accommodate the exchanging of catheters commonly used for access, stabilization, or aspiration. The pusher wire diameter is 0.015" (0.38 mm) with length 350 cm to allow for the exchange of therapeutic devices without the use of extension wires or other exchange devices. The distal end is a permanently attached self-expanding mesh that is deployed out of an introducer sheath and into the vessel where it conforms to the vessel wall. There are 3 distal radiopaque markers and 1 proximal radiopaque marker visible under fluoroscopy. The deployed and stable mesh ensures that the wire is not displaced from the desired position and can be advantageously utilized in multiple exchanges or re-positioning of catheters. The Spartan Center Wire™ is offered in 2 available sizes to accommodate common vascular sizing. There is a 4 mm diameter basket with 15 mm useable length and a 6 mm basket with 30 mm useable length. Both sizes are compatible with catheters that have a minimum I.D. of 0.027".

    AI/ML Overview

    The provided text describes the performance data for the Spartan Center Wire™ medical device, primarily focusing on bench testing, biocompatibility, sterilization, pyrogenicity, shelf life, and animal studies. However, it does not contain information about acceptance criteria or studies related to an AI/ML-driven device's performance in a diagnostic context, MRMC studies, or multi-reader studies. The device in question is a physical catheter guidewire, not a medical AI software.

    Therefore, I cannot extract the requested information regarding acceptance criteria and performance of an AI/ML device from the provided text. The prompt asks for details that are not present in the document.

    The document explicitly states: "No clinical testing was conducted." and "No torque device is needed or supplied with the Spartan Center Wire™." This reinforces that the evaluation is for a physical device, not an AI/ML software.