510(k) Data Aggregation

    K Number
    K232223

    Validate with FDA (Live)

    Device Name
    PlaDuo System
    Manufacturer
    Date Cleared
    2023-10-24

    (90 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PlaDuo is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Argon handpiece is intended for the removal and destruction of skin lesions and the coagulation of tissue.

    Device Description

    The PlaDuo System is an electro-surgical device for use in dermatological applications. The device utilizes nitrogen and argon gas to be used separately to generate the desired power level to patients. The effect of the device is achieved by heating the outer layer of the skin so that part or all of it becomes non-viable and there is controlled damage to the underlying skin.

    The PlaDuo system consists of a system console with wells for 2 gas tanks, footswitch, 2 handpieces, and tip with 3 interchangeable guides.

    AI/ML Overview

    This document describes the FDA's 510(k) clearance for the ShenB Co Ltd's PlaDuo System. The system is an electrosurgical device that uses nitrogen and argon gas to generate plasma for dermatological and general surgical procedures.

    Based on the provided text, the device meets acceptance criteria through various performance tests, including:

    1. IEC 60601-1:2005 + A1:2012: Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
    2. IEC 60601-1-2:2014: Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests.
    3. IEC 60601-2-2:2017: Medical Electrical Equipment - Part 2-2: Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories.
    4. IEC 62304:2006+A1:2015: Medical Device Software Life Cycle Processes.
    5. EN ISO 14971:2012: Medical Devices - Application Of Risk Management To Medical Devices.
    6. Biocompatibility testing (cytotoxicity, irritation, and sensitization) per ISO 10993-1:2018 for patient-contacting materials (space guides).
    7. Software verification and validation testing in accordance with FDA's guidance.
    8. Thermal testing on ex vivo tissue specimens to verify the performance of the argon gas.

    Table of Acceptance Criteria and Reported Device Performance

    CategoryAcceptance Criteria (Implied)Reported Device Performance
    Electrical Safety & PerformanceCompliance with relevant IEC standards (60601-1, 60601-1-2, 60601-2-2) for basic safety and essential performance of electrosurgical equipment."Verification and validation activities were successfully completed and establish that the PlaDuo System control unit performs as intended. Testing included the following: IEC 60601-1:2005 + A1:2012; IEC 60601-1-2:2014; IEC 60601-2-2:2017;"
    Software Safety & PerformanceCompliance with IEC 62304 for medical device software life cycle processes and FDA's software guidance."IEC 62304:2006+A1:2015; Medical Device Software Life Cycle Processes;" and "Software verification and validation testing was conducted, and documentation provided in accordance with FDA's Guidance or the Content of Premarket Submissions for Software Contained in Medical Devices."
    Risk ManagementCompliance with ISO 14971 for application of risk management to medical devices."EN ISO 14971:2012; Medical Devices - Application Of Risk Management To Medical Devices"
    BiocompatibilityAll patient-contacting materials must be biocompatible (non-cytotoxic, non-irritating, non-sensitizing)."Biocompatibility –The only patient contacting material in the device are the space guides attached to each handpiece. Biocompatibility testing for cytotoxicity, irritation, and sensitization per ISO 10993-1:2018 are included with the submission and show no concern for patient contacting materials."
    Thermal Performance (Argon Handpiece)The argon handpiece should achieve consistent and desirable thermal damage profiles comparable to predicate/reference devices for the intended clinical effect."In addition, thermal testing on ex vivo tissue specimens was conducted to verify the performance of the argon gas used with the device. The Test device at stacked pulses produced measurable thermal damage in skin and in liver models and traceable thermal effect in muscle. The study demonstrated the ability of the device to achieve consistent thermal damage profiles in line with the target treatment and comparable to the profiles produced by the predicate and reference devices. Thus, it can be concluded that treatment by such a device at the appropriate testings will possess a desirable clinical treatment effect."
    Substantial Equivalence (Overall)The modified PlaDuo System (with Argon handpiece) must be substantially equivalent to the predicate device(s) for the proposed indications for use."The PlaDuo System that is the subject of this application is a next-generation of the currently cleared PlaDuo with Nitrogen only. Besides the addition of the gas and handpiece tip for use with Argon, there have been no changes to the device. Performance testing shows that the newly added argon feature produces consistent thermal damage profiles similar to predicate and reference devices. The subject device is substantially equivalent to the predicate device for the requested indications for use."

    Study Details:

    1. Sample sizes used for the test set and data provenance:

      • The document mentions "thermal testing on ex vivo tissue specimens." It specifies "skin and in liver models" and "muscle" as the tissue types. However, the exact sample size (number of specimens or tests) is not provided.
      • Data provenance: "ex vivo tissue specimens." The country of origin is not specified. The study's nature is likely prospective as it involves new testing for the modified device.
    2. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

      • The document does not specify the number of experts or their qualifications used to establish ground truth or interpret the thermal testing results. The conclusion about "desirable clinical treatment effect" based on thermal damage profiles implies expert assessment, but details are absent.
    3. Adjudication method for the test set:

      • Since the thermal testing is described as direct measurement of physical effects ("measurable thermal damage," "traceable thermal effect"), a human adjudication method in the sense of consensus reading (2+1, 3+1) is not applicable in the same way it would be for image interpretation. The method of assessing "consistent thermal damage profiles" and comparability to predicate devices would likely rely on quantitative measurements and engineering/biomedical expertise, but no specific adjudication method is mentioned.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was done. This device is an electrosurgical tool, not an AI interpreting medical images. The evaluation focuses on the safety and performance of the electrosurgical function itself, particularly its thermal effects, rather than how it assists human interpretation or decision-making in a diagnostic context.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This question is not applicable as the device is a piece of hardware with integrated software for control, not an AI algorithm to be evaluated in a standalone diagnostic capacity. The "software verification and validation testing" confirms the software's functional performance, which is a component of the device's overall standalone performance.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the thermal testing, the ground truth was observed thermal damage profiles on ex vivo tissue, measured directly. The comparison was made against profiles produced by predicate and reference devices, implying their established effectiveness provides a benchmark. It's a physical/biophysical ground truth.
    7. The sample size for the training set:

      • The document describes a 510(k) submission for a physical medical device, not an AI/machine learning model that typically has a "training set." Therefore, the concept of a training set sample size does not apply in this context.
    8. How the ground truth for the training set was established:

      • As this is not an AI/ML device, the question of establishing ground truth for a training set is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230968

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-07-06

    (92 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VYBE RF Electrosurgical System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis and the percutaneous treatment of facial wrinkles for use with Fitzpatrick Skin Type I to Skin Type V when using the 1MHz setting. The 2MHz setting has not been evaluated for use in the percutaneous treatment of facial wrinkles and is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The VYBE RF Electrosurgical System is comprised of the following components:

    • . The system main body, consisting of:
      • o LCD touch screen control panel
      • High-frequency generating output section (main P.C.B board or RF Generator) O
      • Power supply component Switch Mode Power Supply (SMPS) O
    • The accessories to the device include:
      • Handpiece with Handpiece Connector O
      • Disposable micro-needle cartridge (electrode) for insertion into Handpiece o
      • Foot switch O
        Radio frequency current (RF energy) is delivered from the RF Generator, through the handpiece and electrode tip into the target tissue. As RF energy passes through the skin, it generates an electro thermal reaction, which is capable of coagulating (causing minor dermal damage) the tissue. The bi-polar RF energy is delivered between independent adjacent electrode pairs (total 36 needle electrode, 6 x 6 array insertion). The RF generator and hand piece are not disposable. Each disposable micro-needle cartridge (electrode) is supplied sterile and is for single patient use only and cannot be re-sterilized.
    AI/ML Overview

    The document provided describes the VYBE RF Electrosurgical System and its substantial equivalence to a predicate device, the VIVACE Electrosurgical System. It includes performance data, but does not describe an acceptance criteria and study proving the device meets the acceptance criteria in the manner requested by the prompt for a typical AI/ML medical device submission.

    Specifically, the document focuses on regulatory compliance through electrical, biocompatibility, sterility, software verification and validation, and thermal testing. These are standard tests for medical devices to ensure safety and fundamental functionality, not necessarily performance benchmarks against clinical criteria in the same way an AI/ML diagnostic or prognostic tool would be evaluated.

    Therefore, many of the requested categories in the prompt regarding acceptance criteria, human reader studies, and ground truth establishment are not applicable or not explicitly detailed in the provided text.

    Here's a breakdown of what can and cannot be extracted from the document:

    1. A table of acceptance criteria and the reported device performance

    The document lists various tests and states that "All acceptance criteria were met." However, it does not provide a table with specific acceptance criteria values (e.g., a specific thermal threshold) or the quantitative reported performance for each of these tests. It only lists the standards followed.

    Acceptance Criteria CategoryReported Device Performance
    Electrical TestingMet IEC 60601-1, IEC 60601-1-2, IEC 60101-2-2, IEC 60601-1-6, IEC 62366 standards.
    Biocompatibility TestingMet ISO 10993-1, ISO 10993-5, ISO 10993-10 standards.
    Sterility Testing (micro-needle cartridge)Conforms to a Sterility Assurance Level (SAL) of 10^-6, meeting ISO 11135, ISO 11138-1, ISO 11138-2, ISO 11737-1, ISO 11737-2, ISO 10993-7 standards.
    Software Verification and Validation TestingConducted as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", issued May 11, 2005.
    Thermal TestingCompleted in accordance with FDA's Guidance for Industry and FDA Staff "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", issued March 9, 2020.

    2. Sample size used for the test set and the data provenance

    Not applicable for the types of tests conducted. These are primarily engineering, laboratory, and compliance tests, not clinical studies with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth as typically understood in AI/ML clinical studies (e.g., expert consensus on image interpretation) is not relevant to these engineering and compliance tests.

    4. Adjudication method for the test set

    Not applicable for the types of tests conducted.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an electrosurgical system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is an electrosurgical system.

    7. The type of ground truth used

    For the sterility testing, the ground truth is the established Sterility Assurance Level (SAL) of 10^-6 and adherence to the listed ISO standards.
    For other tests, the "ground truth" is adherence to the specified engineering and biocompatibility standards.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML algorithm that requires a training set in the conventional sense.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221363

    Validate with FDA (Live)

    Device Name
    AF Laser
    Manufacturer
    Date Cleared
    2022-07-20

    (70 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AF Laser device is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).

    Device Description

    The AF Laser is a medical laser used to treat nail fungus (onychomycosis) and promote the growth of healthy nails. The operating mechanism of this device is photochemical reaction that uses two wavelengths -- 635nm and 405nm. The AF laser is a compact, all-in-one device that is intended to be used on the floor and is operated by an intuitive 10.2 inch's wide LCD screen for ease of use. An LCD monitor incorporated into the deck of the device helps the operator check the exact location of treatment area during a procedure.

    AI/ML Overview

    This document describes the FDA's 510(k) clearance for the AF Laser, a device intended for the temporary increase of clear nail in patients with onychomycosis.

    Based on the provided text, the acceptance criteria and study proving the device meets these criteria can be summarized as follows:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document states that the AF Laser is "substantially equivalent" to its predicate device, the LunulaLaser (K153164). The acceptance criteria appear to be met by demonstrating this substantial equivalence, rather than through specific performance metrics outlined in a dedicated clinical study against a predefined threshold.

    Acceptance Criteria (Implied by Substantial Equivalence Basis)Reported Device Performance (Comparison to Predicate)
    Indications for Use: Must be identical or similarly effective for the intended treatment.Identical to LunulaLaser: "The AF Laser device is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.)."
    Laser Wavelength: Must be comparable.Identical to LunulaLaser: 405nm/635nm (±10%)
    Output Energy: Must be comparable.Similar to LunulaLaser: AF Laser: 405nm: 23mW ± 1.85mW, 635nm: 17mW ± 1.35mW LunulaLaser: 405nm: 23.00 ± 2.00mW, 635nm: 17.25 ± 1.25mW
    Output Area: Must be comparable.Identical to LunulaLaser: Line pattern electronically scanned over area of treatment.
    Output Type: Must be comparable.Identical to LunulaLaser: Constant Wave
    Operating Time: Must be comparable.Identical to LunulaLaser: 0-12 minutes (±5% with 1 minute increment for AF Laser, 0-12 minutes for LunulaLaser).
    Safety Standards Compliance: Must comply with relevant safety standards.Passed IEC 60601-1, IEC 60601-1-2, IEC 62471:2006, IEC 60825-1:2014. These are tests for basic safety, essential performance, electromagnetic compatibility, photobiological safety, and laser product safety.
    General Technological Characteristics: Must be similar.Similar in Dimension and Weight (minor differences noted). Identical in Screen type (LCD Touch Screen).

    2. Sample Size Used for the Test Set and the Data Provenance:

    The document explicitly states: "No Clinical testing was conducted as part of this submission." Therefore, there is no test set of clinical data from human subjects in the traditional sense for evaluating device performance against acceptance criteria for clinical efficacy. The submission relies on a comparison to a predicate device and engineering/bench testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    Since no clinical testing was performed, there was no clinical test set requiring expert ground truth establishment.

    4. Adjudication Method for the Test Set:

    Not applicable, as no clinical test set was used.

    5. If a Multi Reader Multi Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a laser for direct treatment, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable, as this is not a diagnostic algorithm. Performance was assessed through engineering and safety standards compliance, and comparison to the predicate device.

    7. The Type of Ground Truth Used:

    The "ground truth" for this submission appears to be the established safety and efficacy profile of the legally marketed predicate device (LunulaLaser K153164) and compliance with international safety and performance standards for laser devices.

    8. The Sample Size for the Training Set:

    Not applicable. As this is not a machine learning or AI device, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211562

    Validate with FDA (Live)

    Device Name
    Virtue RF
    Manufacturer
    Date Cleared
    2021-11-23

    (187 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Virtue RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Smart RF Handpiece:
    The Smart RF Handpiece with 1MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis, and the percutaneous treatment of facial wrinkles. The VirtueRF Electrosurgical System for facial wrinkles is intended for use with Skin Types I-V.
    The Smart RF Handpiece with 2MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The with 2MHz functionality is not intended to treat wrinkles.

    Exact RF Handpiece:
    The Exact RF Handpiece with 1MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Exact RF Handpiece is not intended to treat wrinkles.

    Deep RF Handpiece:
    The Deep RF Handpiece with 1MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Deep RF handpiece with 1MHz functionality is not intended to treat wrinkles.
    The Deep RF Handpiece with 2MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Deep RF handpiece with 2MHz functionality is not intended to treat wrinkles.
    The Deep RF Handpiece with 0.5MHz functionality is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Deep RF handpiece with 0.5MHz functionality is not intended to treat wrinkles.

    Device Description

    The VirtueRF device is a radio frequency output device using 0.5MHz, 1MHz & 2Mhz to deliver radio frequency energy to the human body with 3 applicators that attach to the connecting cable. The energy is applied to the human body using sterile disposable needle arrays consisting of 36 needles, 12 needles, o 1 needle. This product consists of main body, SmartRF Bipolar handpiece, DeepRF Bipolar handpiece, ExactRF Monopolar handpiece, connector cable, grounding plate for ExactRF, cooling plate for the DeepRF, foot switch and power cord.

    AI/ML Overview

    The provided text is a 510(k) summary for the Virtue RF device, an electrosurgical cutting and coagulation device. This type of FDA submission focuses on demonstrating substantial equivalence to a predicate device rather than undergoing a clinical trial to prove de novo effectiveness or meeting specific performance acceptance criteria for a novel device. As such, the document does not contain information typically found in a study demonstrating a device meets acceptance criteria involving human or animal subjects, ground truth, or statistical measures of performance like AUC, sensitivity, or specificity.

    Instead, the submission demonstrates substantial equivalence primarily through:

    1. Bench testing: Adherence to recognized electrical safety and electromagnetic compatibility standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2) and biological compatibility (ISO 10993-1).
    2. Thermal effect tests: Performed on four types of tissues (liver, kidney, muscle, and skin) as guided by FDA for electrosurgical devices.
    3. Direct comparison of technological characteristics: Side-by-side tables comparing the proposed device's various handpieces (Smart RF, Exact RF, Deep RF) with its own predicate device (Virtue RF K202415) and other reference devices (Primaeva Medical Miratone System K082391, Agnes K192728, and Agnes K160469, though K160469 is not explicitly listed as a reference device, it appears in a comparison table for Exact RF).

    Therefore, I cannot provide the detailed acceptance criteria and study information requested in the prompt, as these elements are not part of this specific type of regulatory submission. The document concludes that "Substantial equivalence is supported by the performance testing and there are no new questions of safety or effectiveness."

    To explicitly address your points:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria for Bench Testing: Adherence to the listed IEC standards (e.g., meeting specific safety thresholds for electrical current, electromagnetic emissions, and biological compatibility). The "reported device performance" is that it passed these tests.
    • Acceptance Criteria for Thermal Effect Tests: The criteria would be to demonstrate substantial equivalence to the predicate device in terms of thermal effects on tissues. The "reported device performance" is that the "Test shows that the VirtueRF is substantially equivalent to the predicate device."
    • Acceptance Criteria for Technological Characteristics: The criteria are for the proposed device's features (Source of energy/Modality, Radiofrequency, Output energy type, Electrode type, Max Power, RF Duration, Number of Needles, Grounding Mechanism, Cooling Modality) to be "Same" or "Similar" to the predicate/reference devices, with justifications where there are differences (e.g., "Similar; same as reference device" or "Same as reference; similar to predicate device"). The "reported device performance" is detailed in the comparison tables provided in the document (pages 6-8).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not applicable for this type of submission. The "tests" here are bench tests on the device's hardware and thermal effects on tissues, not clinical trials on human subjects. The tissue tests would likely use an ex vivo or in vivo animal model, but details on sample size or provenance are not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable. Ground truth from human experts is not used for this type of technical and bench testing-focused submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. No expert adjudication is involved.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is not an AI-assisted diagnostic device, and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not a software algorithm; it's an electrosurgical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the thermal effect tests, the "ground truth" would be objective measurements of tissue coagulation or heating, compared to measurements from the predicate device. For technical standards, the "ground truth" is adherence to the specified parameters in the IEC and ISO standards.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" as this is not a machine learning device.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193070

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-04-26

    (539 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VIVACE Electrosurgical System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis and the percutaneous treatment of facial wrinkles for use with Fitzpatrick Skin Type I to Skin Type V when using the 1MHz setting. The 2MHz setting has not been evaluated for use in the percutaneous treatment of facial wrinkles and is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The VIVACE Electrosurgical System is comprised of the following components:

    • . The system main body, consisting of:
      • LCD touch screen control panel O
      • High-frequency generating output section (main P.C.B board or RF Generator) O
      • Power supply component Switch Mode Power Supply (SMPS) O
    • The accessories to the device include: ●
      • Handpiece with Handpiece Connector o
      • Disposable micro-needle cartridge (electrode) for insertion into Handpiece O
      • Foot switch O
        Radio frequency current (RF energy) is delivered from the RF Generator, through the handpiece and electrode tip into the target tissue. As RF energy passes through the skin, it generates an electro thermal reaction, which is capable of coagulating (causing minor dermal damage) the tissue. The bi-polar RF energy is delivered between independent adjacent electrode pairs (total 36 needle electrode, 6 x 6 array insertion). The RF generator and hand piece are not disposable micro-needle cartridge (electrode) is supplied sterile and is for single patient use only and cannot be re-sterilized.
    AI/ML Overview

    The provided text is a 510(k) Summary for the VIVACE Electrosurgical System. It outlines the device's technical specifications and the performance data submitted to the FDA to establish its safety and efficacy and demonstrate substantial equivalence to predicate devices. However, it does not contain information about acceptance criteria or a study proving the device meets them in the context of an AI/ML medical device, which is what your request implies.

    The "Performance Data" section (Page 6 of the input) lists various tests that were conducted (electrical testing, biocompatibility testing, sterility testing, shelf life/packaging, software verification/validation, and thermal testing) and states that "All acceptance criteria were met." However, it does not define these acceptance criteria, nor does it provide detailed results or specific studies that you've requested (e.g., MRMC studies, sample sizes for test/training sets, expert qualifications, ground truth establishment methods for AI/ML performance studies).

    The "Thermal Testing" section briefly mentions:

    • An ex vivo animal study using a porcine tissue model.
    • Standardized H&E staining techniques to visualize and measure thermally damaged profile data through Nitro Blue Tetra Chlorine (NBTC) and H&E histological evaluation.
    • The finding that the vivace microneedling device showed consistent formation of measurable thermal lesions for highest energy level settings in the treatment porcine tissue at a temperature close to physiological.

    This is a performance test, not a study to establish clinical effectiveness or compare AI performance against a human or a human-assisted AI.

    Therefore, I cannot fulfill your request for the specific details of acceptance criteria and the study proving the device meets them as described in your prompt (which is structured for an AI/ML device performance study) because this regulatory document for the VIVACE Electrosurgical System does not contain such information. It focuses on demonstrating substantial equivalence through engineering and biological performance testing, not AI model performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202415

    Validate with FDA (Live)

    Device Name
    VirtueRF
    Manufacturer
    Date Cleared
    2021-01-22

    (151 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VirtueRF Electrosurgical System has separate indications for use for each of its functionalities: 1MHz and 2MHz.

    1MHz Indications for use: Intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis, and the percutaneous treatment of facial wrinkles. The VirtueRF Electrosurgical System is intended for use with Skin Types I-V.

    2MHz Indications for use: Intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The 2mHz frequency is not intended to treat wrinkles.

    Device Description

    The VirtueRF device is a radio frequency output device using 1MHz & 2Mhz to deliver radio frequency energy to the human body with an applicator attached to the connecting cable. The energy is applied to the human body using a sterile disposable needle array consisting of 36 needles. This product consists of main body, Needle Bipolar RF (Smart RF) handpiece cable, foot switch and power cord.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study information for the VirtueRF device:

    Based on the provided K202415 510(k) Summary for the VirtueRF device, the submission primarily focuses on establishing substantial equivalence to predicate devices through technical comparisons and standard performance testing. This type of submission generally does not include extensive clinical studies with specified acceptance criteria regarding diagnostic performance or human-AI comparative effectiveness, as it's not a diagnostic AI device.

    Therefore, many of your requested points regarding acceptance criteria for diagnostic performance, sample sizes for test sets, expert ground truth, MRMC studies, or standalone algorithm performance are not applicable in this context. The acceptance criteria here are related to meeting safety, electrical, biological, and thermal performance standards to demonstrate equivalence to existing devices.

    Nonetheless, I will extract what information is present or clearly implied.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implied/Direct)Reported Device Performance
    Electrical Safety & PerformanceCompliance with IEC 60601-1 (General Requirements for basic safety and essential performance).Testing was performed per IEC 60601-1.
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2 (General Requirements for basic safety and essential performance: electromagnetic compatibility).Testing was performed per IEC 60601-1-2.
    High Frequency Surgical EquipmentCompliance with IEC 60601-2-2 (Particular requirements for the basic safety and essential performance of high frequency surgical equipment).Testing was performed per IEC 60601-2-2.
    BiocompatibilityCompliance with ISO 10993-1 (Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process).Testing was performed per ISO 10993-1.
    Thermal EffectsThermal effects comparable to predicate devices, demonstrating safe and effective energy delivery to tissues."Thermal Effect Tests on 4 types of tissues (liver, kidney, muscle and skin) per FDA guidance... Test shows that the VirtueRF is substantially equivalent to the predicate device."
    Technical Equivalence (1MHz)Same source of energy/modality, radiofrequency, output energy type, electrode type, max power, and RF duration as Vivace Electrosurgical System (K150409).All characteristics (except MAX Power is 35.9W for VirtueRF vs 61W for Vivace) are "Same" or acceptable for equivalence, as stated.
    Technical Equivalence (2MHz)Same source of energy/modality, radiofrequency, output energy type, electrode type, max power, and similar RF duration as Secret RF (K182355).All characteristics are "Same" except RF Duration, which is "Similar" (100ms-800ms subset of 50ms-950ms).

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not present for the efficacy or diagnostic performance as those types of studies were not conducted or described. The "test set" in this context refers to the samples used for the engineering and bench testing (e.g., tissue samples for thermal effect tests, electrical testing setup), not a clinical trial test set for diagnostic performance. The document does not specify the number of tissue samples or the specifics of the electrical test configurations used.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable as the submission is not for a diagnostic device or an AI algorithm requiring expert-established ground truth for performance evaluation.

    4. Adjudication Method for the Test Set

    This information is not applicable for the reasons stated above.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices or AI-assisted diagnostic tools, which the VirtueRF electrosurgical system is not.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    No, a standalone algorithm performance study was not done. The VirtueRF is a medical device, not a standalone AI algorithm for diagnostic purposes. Its performance is evaluated through engineering, electrical, biocompatibility, and thermal tests, and comparison to predicates, not through AI-specific performance metrics.

    7. The Type of Ground Truth Used

    For the thermal effect tests, the "ground truth" would be established by direct physical measurements of temperature changes and tissue effects, validated against established scientific principles and comparison to the predicate device's known effects. This is a scientific and engineering ground truth rather than a clinical ground truth like pathology or expert consensus.

    8. The Sample Size for the Training Set

    This information is not applicable. The VirtueRF is a hardware device, not an AI model that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201735

    Validate with FDA (Live)

    Device Name
    PlaDuo System
    Manufacturer
    Date Cleared
    2020-12-14

    (172 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PlaDuo is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The ShenB Co Ltd PlaDuo is an electro-surgical device for use in dermatological applications. The effect of the device is achieved by heating the outer layer of the skin so that part or all of the epidermis becomes non-viable and there is controlled damage to the underlying dermis.

    The PlaDuo system consists of a system console, footswitch, handpiece, and tip with 3 interchangeable guides.

    AI/ML Overview

    The provided text is a 510(k) summary for the PlaDuo System, an electrosurgical device. It outlines non-clinical tests conducted to support its substantial equivalence to predicate devices, but no clinical studies or acceptance criteria based on clinical performance are mentioned. Therefore, I cannot provide details on acceptance criteria or studies that prove the device meets them in the context of clinical performance.

    However, based on the non-clinical tests performed, I can infer acceptance criteria for those tests and describe the study that proves the device met them.

    Here's a summary of the non-clinical acceptance criteria and the studies performed:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific TestAcceptance CriteriaReported Device Performance
    BiocompatibilityCytotoxicity (ISO 10993-05:2009)Device components (electrode tips) must show no cytotoxic effects in vitro.Pass
    Sensitization (ISO 10993-10:2010, Guinea Pig Maximization Test)Device components (electrode tips) must not induce sensitization (allergic reaction).Pass
    Irritation or Intracutaneous Reactivity (ISO 10993-10:2010, Animal Intracutaneous Test)Device components (electrode tips) must not cause irritation or intracutaneous reactivity.Pass
    Electrical Safety & EMCIEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2-12 (General requirements for basic safety and essential performance)Device must meet general requirements for basic safety and essential performance for medical electrical equipment.Not explicitly stated as "Pass" but implied by submission to standards.
    IEC 60101-2-2:2017 (Particular requirements for the basic safety and essential performance of high-frequency surgical equipment)Device must meet particular requirements for basic safety and essential performance of high-frequency surgical equipment.Not explicitly stated as "Pass" but implied by submission to standards.
    IEC 60601-1-2:2014 (General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances - Requirements and Tests)Device must meet electromagnetic compatibility (EMC) requirements and relevant tests for medical electrical equipment.Not explicitly stated as "Pass" but implied by submission to standards.
    Software Verification & ValidationFDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", May 11, 2005 (moderate level of concern)Software must be verified and validated according to FDA guidance for a moderate level of concern, ensuring proper functionality and safety.Documentation provided. Not explicitly stated as "Pass" but implied by submission and acceptance of documentation.
    Bench TestingOutput EnergyOutput energy must be within predefined design specifications.Operates safely and within predefined design specifications.
    Durability of the tipTip must demonstrate adequate durability according to predefined design specifications.Operates safely and within predefined design specifications.
    Thermal testing (in accordance with FDA's "Guidance for Industry and FDA Staff: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery”)Thermal effects must be within acceptable limits as outlined by FDA guidance to ensure patient safety and device efficacy.Operates safely and within predefined design specifications.

    Important Note: The document explicitly states "No Clinical study was conducted as part of this submission." Therefore, the following points address the questions in the context of the available non-clinical information.


    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Biocompatibility: The sample sizes for these tests are specific to the biological assay standards (e.g., number of cells for cytotoxicity, number of guinea pigs for sensitization, number of animals for intracutaneous reactivity). These specific numbers are not detailed in the provided summary, but the tests were performed in accordance with international ISO standards.
    • Electrical Safety & EMC: These are typically laboratory tests performed on a single or a few units of the device. The specific number of units tested is not stated.
    • Software Verification & Validation: This involves reviewing and testing the software, often through simulated scenarios or on development units. Sample size for actual "test cases" within software testing is not specified.
    • Bench Testing: This involved testing of the device and its components (e.g., tips). The specific number of devices or tips tested is not provided.

    The provenance of this non-clinical data is from the manufacturer, ShenB Co. Ltd., which is based in South Korea (Seongdong-Gu, Seoul). These were prospective tests conducted to demonstrate compliance with the relevant standards.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This question is not directly applicable as the studies were non-clinical, involving engineering, materials science, and software experts.

    • Biocompatibility: The "ground truth" (i.e., whether a material is biocompatible) is established by the results of standardized laboratory tests interpreted by trained toxicologists/biocompatibility specialists.
    • Electrical Safety & EMC: The "ground truth" (i.e., whether the device meets electrical safety and EMC standards) is established by measurements against predefined limits by qualified electrical engineers and testing laboratories.
    • Software Verification & Validation: The "ground truth" (i.e., whether the software functions as intended and is safe) is established by software engineers and quality assurance specialists comparing software behavior against its requirements and design specifications.
    • Bench Testing: The "ground truth" (i.e., performance against specifications) is established by engineers performing measurements and observations against design requirements.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This concept is primarily relevant for clinical studies involving human interpretation or subjective endpoints. For these non-clinical tests, adjudication methods like 2+1 or 3+1 are not typically used. The results of these tests are objectively measured against established standards and specifications. If there were any ambiguous results, internal quality processes or re-testing would be employed.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was performed as no clinical studies were conducted, and the device is an electrosurgical tool, not an AI-assisted diagnostic or interpretative system.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the PlaDuo System is an electrosurgical device, not an algorithm. The "standalone" performance relates to the device's functional integrity as an electrosurgical unit, which was assessed through bench testing and compliance with electrical safety standards.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests:

    • Biocompatibility: The ground truth is established by standardized laboratory test results (e.g., cell viability, immune response observed in animal models) compared against established safety thresholds defined by ISO standards.
    • Electrical Safety & EMC: The ground truth is established by objective measurements of electrical characteristics (e.g., leakage current, electromagnetic emissions) against the limits defined in IEC standards.
    • Software Verification & Validation: The ground truth is established by the software requirements specification and design documents, against which the functionality and performance of the software are tested.
    • Bench Testing: The ground truth is established by the device's predefined design specifications and engineering requirements for parameters like output energy, tip durability, and thermal effects.

    8. The sample size for the training set

    This question is not applicable as there is no mention of an algorithm or AI model in the submission that would require a training set.


    9. How the ground truth for the training set was established

    Not applicable, as no training set for an AI/algorithm was mentioned.