K Number
K193070

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-04-26

(539 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VIVACE Electrosurgical System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis and the percutaneous treatment of facial wrinkles for use with Fitzpatrick Skin Type I to Skin Type V when using the 1MHz setting. The 2MHz setting has not been evaluated for use in the percutaneous treatment of facial wrinkles and is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

Device Description

The VIVACE Electrosurgical System is comprised of the following components:

  • . The system main body, consisting of:
    • LCD touch screen control panel O
    • High-frequency generating output section (main P.C.B board or RF Generator) O
    • Power supply component Switch Mode Power Supply (SMPS) O
  • The accessories to the device include: ●
    • Handpiece with Handpiece Connector o
    • Disposable micro-needle cartridge (electrode) for insertion into Handpiece O
    • Foot switch O
      Radio frequency current (RF energy) is delivered from the RF Generator, through the handpiece and electrode tip into the target tissue. As RF energy passes through the skin, it generates an electro thermal reaction, which is capable of coagulating (causing minor dermal damage) the tissue. The bi-polar RF energy is delivered between independent adjacent electrode pairs (total 36 needle electrode, 6 x 6 array insertion). The RF generator and hand piece are not disposable micro-needle cartridge (electrode) is supplied sterile and is for single patient use only and cannot be re-sterilized.
AI/ML Overview

The provided text is a 510(k) Summary for the VIVACE Electrosurgical System. It outlines the device's technical specifications and the performance data submitted to the FDA to establish its safety and efficacy and demonstrate substantial equivalence to predicate devices. However, it does not contain information about acceptance criteria or a study proving the device meets them in the context of an AI/ML medical device, which is what your request implies.

The "Performance Data" section (Page 6 of the input) lists various tests that were conducted (electrical testing, biocompatibility testing, sterility testing, shelf life/packaging, software verification/validation, and thermal testing) and states that "All acceptance criteria were met." However, it does not define these acceptance criteria, nor does it provide detailed results or specific studies that you've requested (e.g., MRMC studies, sample sizes for test/training sets, expert qualifications, ground truth establishment methods for AI/ML performance studies).

The "Thermal Testing" section briefly mentions:

  • An ex vivo animal study using a porcine tissue model.
  • Standardized H&E staining techniques to visualize and measure thermally damaged profile data through Nitro Blue Tetra Chlorine (NBTC) and H&E histological evaluation.
  • The finding that the vivace microneedling device showed consistent formation of measurable thermal lesions for highest energy level settings in the treatment porcine tissue at a temperature close to physiological.

This is a performance test, not a study to establish clinical effectiveness or compare AI performance against a human or a human-assisted AI.

Therefore, I cannot fulfill your request for the specific details of acceptance criteria and the study proving the device meets them as described in your prompt (which is structured for an AI/ML device performance study) because this regulatory document for the VIVACE Electrosurgical System does not contain such information. It focuses on demonstrating substantial equivalence through engineering and biological performance testing, not AI model performance.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 26, 2021

ShenB Co Ltd. % Rachel Lord Director RA/OA Aesthetics Biomedical 4602 N 16th St Suite 300 Phoenix, Arizona 85016

Re: K193070

Trade/Device Name: VIVACE Electrosurgical Device Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: April 3, 2021 Received: April 5, 2021

Dear Rachel Lord:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm = identifies product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and

{1}------------------------------------------------

regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and. if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193070

Device Name Vivace Electrosurgical System

Indications for Use (Describe)

The VIVACE Electrosurgical System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis and the percutaneous treatment of facial wrinkles for use with Fitzpatrick Skin Type I to Skin Type V when using the 1MHz setting. The 2MHz setting has not been evaluated for use in the percutaneous treatment of facial wrinkles and is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 GER 801 Subpart D)Over-The-Counter Use (21 CER 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR Part 807.92.

A. SPONSOR INFORMATION

Name:ShenB Co., Ltd.
Address:Shenb Bldg 148, Seongsui-roSeongdong-Gu, Seoul, 04796Republic of Korea
Phone:82-2-466-0010
Fax:82-2-466-5473

Official Correspondent:

Name:Rachel Lord
Address:Aesthetics Biomedical4602 North 16th Street, Suite 300Phoenix, AZ 85016
Phone:(201) 841-7975
Email:rlord@aestheticsbiomed.com

Establishment Reg. No.: 3010226575

April 2, 2020 Date Prepared:

{4}------------------------------------------------

B. DEVICE NAME

Trade Name:VIVACE
Common Name:Electrosurgical System and Accessories
Classification Name:Electrosurgical cutting and coagulation device and accessories
Classification Number:21 CFR 878.4400 (Class II)
Product Code:OUH, GEI
Classification Panel:General and Plastic Surgery

C. PREDICATE DEVICE

The VIVACE Electrosurgical System is substantially equivalent to the existing VIVACE Electrosurgical System (primary predicate), K150409. The Secret RF Device K170325 is the reference device to support the 2MHz functionality.

D. INTENDED USE

The VIVACE Electrosurgical System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis and the percutaneous treatment of facial wrinkles for use with Fitzpatrick Skin Type V when using the 1MHz setting. The 2MHz setting has not been evaluated for use in the percutaneous treatment of facial wrinkles and is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

E. DEVICE DESCRIPTION

The VIVACE Electrosurgical System is comprised of the following components:

  • . The system main body, consisting of:
    • LCD touch screen control panel O
    • High-frequency generating output section (main P.C.B board or RF Generator) O
    • Power supply component Switch Mode Power Supply (SMPS) O
  • The accessories to the device include: ●
    • Handpiece with Handpiece Connector o
    • Disposable micro-needle cartridge (electrode) for insertion into Handpiece O
    • Foot switch O

Radio frequency current (RF energy) is delivered from the RF Generator, through the handpiece and electrode tip into the target tissue. As RF energy passes through the skin, it generates an electro thermal reaction, which is capable of coagulating (causing minor dermal damage) the tissue. The bi-polar RF energy is delivered between independent adjacent electrode pairs (total 36 needle electrode, 6 x 6 array insertion). The RF generator

{5}------------------------------------------------

and hand piece are not disposable micro-needle cartridge (electrode) is supplied sterile and is for single patient use only and cannot be re-sterilized.

F. TECHNOLOGICAL SPECIFICATIONS

The VIVACE Electrosurgical System is substantially equivalent to the primary predicate device, with respect to technological characteristics, such as intended use, principles of operation, target population, and energy source.

Electrical voltage and frequencyAC 120V 50/60 Hz
Power conception80 VA
Maximum Power36W ± 10% (load resistance $500\Omega$ )
Accuracy of the output1 MHz, 2 MHz
MicroneedleCartridgeElectrode36 each, 6 x 6 array
Exposed length0.5 ~ 3.5mm (0.1mm increments)
Outer diameterØ 0.3mm
Dimensions(W)380mm x (L)340mm x (H)1200mm

G. PERFORMANCE DATA

To establish safety and efficacy of the VIVACE Electrosurgical System, the following evaluations were completed following standards and FDA guidance documents. All acceptance criteria were met:

  • . Electrical Testing per:
    • o IEC 60601-1:2005/(R)2012, A1:2012
    • IEC 60601-1-2:2014, Ed. 4.0 o
    • IEC 60101-2-2:2017, Ed. 6.0 O
    • IEC 60601-1-6:2010 +A1 2013 O
    • IEC 62366:2007/AMD 1:2014 o
  • Biocompatibility Testing per: ●
    • ISO 10993-1:2009 о
    • o ISO 10993-5:2009
    • ISO 10993-10:2010 o
  • Sterility Testing: ●
    • o The micro-needle cartridge is supplied sterile and sterility conforms to a Sterility Assurance Level (SAL) of 10-6:
      • ISO 11135:2014 트
      • I ISO 11138-1:2017
      • l ISO 11138-2:2017
      • I ISO 11737-1:2018
      • I ISO 11737-2:2019
      • I ISO 10993-7:2008

{6}------------------------------------------------

  • Shelf Life and Packaging Testing per: ●
    • o ASTM F1980-16
    • o ASTM F1929-15
    • o ASTM F88/F88M-15
  • Software Verification and Validation Testing: ●
    • Software verification and validation testing were conducted as recommended by O FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", issued May 11, 2005.
  • Thermal Testing: ●
    • Testing of the thermal effects on tissue was completed in accordance with о FDA's Guidance for Industry and FDA Staff "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", issued March 9, 2020.
    • The ex vivo animal study utilized a porcine tissue model and standardized H&E o staining techniques to visualize and measure thermally damaged profile data through Nitro Blue Tetra Chlorine (NBTC) and H&E histological evaluation of VIVACE System treated tissue specimens. The VIVACE microneedling device showed consistent formation of measurable thermal lesions for highest energy level settings in the treatment porcine tissue at a temperature close to physiological.

{7}------------------------------------------------

H. SUMMARY OF SUBSTANTIAL EQUIVALENCE

The subject VIVACE Electrosurgical System is substantially equivalent to the primary predicate, the VIVACE Electrosurgical System (K150409). The Secret RF Device K 170325 is the reference device to support the 2MHz functionality. The below table VIVACE technological characteristics to the predicate devices.

Device nameVIVACE (Subject Device)PRIMARY PREDICATEVIVACE (K150409)REFERENCE PREDICATESecret RF (K170325)Difference/Risks/Notes
ManufacturerShenB Co., Ltd.ShenB Co., Ltd.ILOODA CO., LTDDifferent Manufacturer for ReferencePredicate (Secret RF)
Classification #& ProductCode878.4400OUH, GEI878.4400OUH878.4400GEI, OUHNO DIFFERENCE
Indication forUseThe VIVACE Electrosurgical Systemis intended for use in dermatologicand general surgical procedures forelectrocoagulation and hemostasis andthe percutaneous treatment facialwrinkles for use with Fitzpatrick SkinType I to Skin Type V when using the1MHz setting. The 2MHz setting hasnot been evaluated for use in thepercutaneous treatment of facialwrinkles and is intended for use indermatologic and general surgicalprocedures for electrocoagulation andhemostasis.Intended for use in dermatologicand general surgical procedures forelectrocoagulation and hemostasis,and the percutaneous treatment offacial wrinkles. This VivaceElectrosurgical System is intendedfor use with Skin Type I to SkinType V.Secret RF is intended for use indermatologic and general surgicalprocedures for electro-coagulationand hemostasis.NO DIFFERENCE
Source ofEnergyBipolar Fractional RFBipolar Fractional RFBipolar Fractional RFNO DIFFERENCE
Delivery systemBipolar Handpiece + Micro needleelectrodesBipolar Handpiece + Micro needleelectrodesBipolar Handpiece + Micro needleelectrodesNO DIFFERENCE
RF Frequency1MHz (±10%)1MHz (±10%)N/A• Additional (lower) frequency settingin comparison to existing VIVACEfrequency
2MHz (±10%)N/A2MHz (±10%)• NO DIFFERENCE from Secret RF2MHz frequency
Max power1MHz: 36 W (Load resistance 500Ω)1MHz: 61.5 W (Load resistance500Ω)N/A• VIVACE has a slightly lower outputas the predicate VIVACE, when using
Device nameVIVACE (Subject Device)PRIMARY PREDICATEVIVACE (K150409)REFERENCE PREDICATESecret RF (K170325)Difference/Risks/Notes
2MHz: 23.3 W (Load resistance$500Ω$ )N/A2 MHz: Max 25W± 10% (Loadresistance $500Ω$ )the 1 MHz setting.• VIVACE has a slightly lower outputthan the predicate Secret RF device,when using the 2 MHz setting.
Connectedelectrodes36 Electrodes36 ElectrodesMTR-AC-25 (25 Electrodes) orMTR-AC-64 (64 Electrodes)• NO DIFFERENCE from existingVivace device• Within Electrode Range of Secret RFdevice
User interfaceLCD touch screen control panelLCD touch screen control panelLCD touch screen control panelNO DIFFERENCE
ElectricalRequirements120VAC, 50/60Hz120VAC, 50/60Hz120VAC, 50/60Hz, 3-1ANO DIFFERENCE
Intensity1-10 Level (1 Step)1-10 Level (1 Step)0-10 Level (2/5/10 STEP)• NO DIFFERENCE from existingVivace device• NO DIFFERENCE in range ofintensity levels from existing Secret RFdevice
Weight(Without Arm)24kg24kg40kg• NO DIFFERENCE from existingVivace device• Lighter than Secret RF device
Needle insertdepth0.5~3.5mm (0.1mm step)0.5~3.5mm (0.1mm step)0.5~3.5mm (0.1 step)NO DIFFERENCE

{8}------------------------------------------------

CONCLUSION I.

The subject VIVACE device, when used with the 2 MHz setting, is identical to the predicate and reference devices (K170325 and K150409) with respect to the principles of operation, technological characteristics, as well as performance characteristics. Non-clinical testing was conducted to evaluate the performance of subject device in comparison to the predicate devices and results have demonstrated substantial equivalence of the predicates in terms of safety and effectiveness for the intended use.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.