Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K131792

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-12-16

    (181 days)

    Product Code
    Regulation Number
    882.5910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SyntheCel® Dura Onlay is intended for use as a dura replacement for the repair of dura mater in adults.

    Device Description

    SyntheCel® Dura Onlay is composed of biosynthesized cellulose and water with a unique construction of non-woven, interconnected cellulose fibers. SyntheCel® Dura Onlay functions as a mechanical layer which protects and repairs the dural defect while preventing further CSF leakage. SyntheCel® Dura Onlay is immunologically inert and has demonstrated minimal foreign body response. It is non-resorbable.

    AI/ML Overview

    The provided document is a 510(k) summary for the SyntheCel® Dura Onlay device. It primarily focuses on establishing substantial equivalence to a predicate device rather than detailing specific acceptance criteria and a standalone study proving those criteria.

    Therefore, many of the requested items cannot be definitively extracted from this document, as the submission relies on previous studies for the predicate device to establish safety and effectiveness.

    Here's what can be gathered:

    1. A table of acceptance criteria and the reported device performance

    The document does not list specific numeric acceptance criteria for performance such as sensitivity, specificity, or AUC, as it's not a performance study of a diagnostic AI device. Instead, the acceptance criteria are based on demonstrating substantial equivalence to a predicate device in terms of technological characteristics and safety/effectiveness data.

    Acceptance Criteria CategoryReported Device Performance (SyntheCel® Dura Onlay)
    MaterialBiosynthesized cellulose
    Dimensions1in x 1in (2.5cm x 2.5cm), 1in x 3in (2.5cm x 7.5cm), 2in x 2in (5.0cm x 5.0cm), 3in x 3in (7.5cm x 7.5cm), 4in x 5in (10.0cm x 12.0cm)
    FunctionConforms to contours of brain. Functions as a mechanical layer which protects and repairs the dural defect while preventing further CSF leakage. Immunologically inert and has demonstrated minimal foreign body response.
    ResorbabilityNon-resorbable
    SterilizationSterilized by irradiation
    BiocompatibilityDemonstrated to be non-irritating, non-sensitizing, non-mutagenic, non-cytotoxic, non-hemolytic, and non-pyrogenic (based on prior testing for predicate).
    Pyrogenicity< 0.06 EU/ml (Endotoxin Units / ml) using Limulus Amebocyte Lysate (LAL) test (based on prior testing for predicate).
    Mechanical StrengthTensile strength and burst strength demonstrated to be substantially equivalent to predicate devices (based on prior testing for predicate).
    Indications for UseIntended for use as a dura replacement for the repair of dura mater in adults.
    Clinical PerformanceNo new clinical data collected. Relies on data from predicate K113071, which evaluated SyntheCel® Dura Replacement Devices (including SyntheCel® Dura Repair and SyntheCel® Dura Onlay) for safety and effectiveness compared to a Control, demonstrating substantial equivalence to other legally marketed dura replacement products.

    2. Sample size used for the test set and the data provenance

    No new test set was used for this specific 510(k) submission. The document states: "No new clinical performance data were collected in support of this submission." and "No new non-clinical performance data were collected in support of this submission."

    The previous clinical data referenced was for the predicate device (K113071), but details about its sample size, country of origin, or retrospective/prospective nature are not provided in this document.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no new clinical or performance test set was developed for this submission. The ground truth for the predicate device's studies is not detailed here.

    4. Adjudication method for the test set

    Not applicable, as no new test set was used for this submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a dura substitute, not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical implant device, not an algorithm.

    7. The type of ground truth used

    For the predicate device's studies, the ground truth would typically be based on clinical outcomes relevant to dura repair (e.g., successful dural closure, lack of CSF leakage, absence of adverse reactions, foreign body response, inflammation) observed in patients. Mechanical testing used measurable physical properties (tensile strength, burst strength). Biocompatibility relied on established in-vitro and in-vivo tests aligned with ISO 10993 standards.

    8. The sample size for the training set

    Not applicable. This is a medical implant, not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a medical implant, not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122647

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-02-28

    (182 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Synthes Patient Specific Plates are intended for use in oral and maxillofacial surgery, trauma and reconstructive surgery.

    Specific Indications for Use:

    • Primary mandibular reconstruction with bone graft
    • Temporary bridging until delayed secondary reconstruction
    • Secondary mandibular reconstruction
    • Comminuted mandibular fractures
    • Fractures of edentulous and/or atrophic mandibles
    • Unstable mandibular fractures
    • Maxillary reconstruction with or without bone graft
    • Maxillary trauma
    Device Description

    Synthes Patient Specific Plates are a metallic bone plates used in conjunction with metallic bone screws for the internal fixation of maxillofacial bone, specifically in the areas of the mandible and maxilla. The design and dimensions (number and placement of holes, hole angulation, plate length, etc.) of each plate are based on of the patient's anatomic (CT scan) data, the surgical plan, and input from the surgeon. Synthes Patient Specific Plates are not intended to be further bent or contoured during surgery. Synthes Patient Specific Plates are manufactured from commercially pure titanium, are provided non-sterile, must be sterilized prior to use, and are intended for single use only.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Synthes Patient Specific Plates (K122647):

    This submission is a 510(k) for a Class II medical device, which typically relies on demonstrating substantial equivalence to a predicate device rather than full clinical trials. Therefore, the "study" described is primarily focused on non-clinical (bench) testing to support this equivalence.


    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryAcceptance Criteria (Implicit)Reported Device Performance
    Bending ResistanceProposed device should have equivalent or better in-plane resistance to bending compared to predicate devices."The bending resistance of the proposed plates was determined to be equivalent or better to the predicate devices."
    Cantilever Bend Fatigue LifeProposed device should have equivalent or better cantilever bend fatigue life compared to predicate devices."The fatigue life of the proposed plates was determined to be equivalent or better to the predicate devices."
    Safety & EffectivenessDifferences in technological characteristics should not raise any new questions of safety and effectiveness compared to predicate devices."Therefore, it is concluded that the proposed plates do not raise any new questions of safety or efficacy in terms of bending resistance." and "Therefore, it is concluded that the proposed plates do not raise any new questions of safety or efficacy in terms of cantilever bend fatigue life." The overall conclusion states: "The information submitted supports substantial equivalence."

    Study Information

    1. Sample Size used for the test set and the data provenance:

      • Sample Size: Not explicitly stated. The tests were "cross-sectional analysis" and "mechanical testing," suggesting multiple samples were tested for each device type (proposed and predicate). The exact number of plates or test specimens is not provided.
      • Data Provenance: Not explicitly stated, but as this is bench testing for a device manufactured by Synthes (USA), it's implied that the testing was performed in a controlled laboratory environment, likely within the US or a facility affiliated with the manufacturer. It is retrospective in the sense that the device design already existed and was being tested against established predicate devices.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. This was non-clinical bench testing, not a study involving human interpretation or clinical data where expert ground truth would be established. The "ground truth" here is the physical performance of the devices as measured by standardized mechanical tests.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. As above, this was non-clinical bench testing. Adjudication methods are relevant for studies involving human interpretation or clinical endpoints.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. An MRMC study was not conducted. This device is a custom-fit bone plate, not an AI-assisted diagnostic or therapeutic tool. The comparison was between the physical performance of the proposed plate and existing predicate plates.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No. This device is a physical medical implant, not a software algorithm. Its performance is inherent to its material and design, not an algorithm's output.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this non-clinical study was based on objective mechanical measurements of physical properties (bending resistance, fatigue life) compared against established performance characteristics of legally marketed predicate devices. There is no expert consensus, pathology, or outcomes data used to establish "ground truth" in this context.
    7. The sample size for the training set:

      • Not applicable. This device is not an AI/ML algorithm that requires a "training set." Its design is based on patient-specific CT data and surgical plans, which is a different concept from a machine learning training set.
    8. How the ground truth for the training set was established:

      • Not applicable. As there is no training set for an AI/ML algorithm involved, there is no ground truth for a training set to establish. The design of the individual patient-specific plate uses the patient's anatomical CT scan data and the surgeon's input as its 'blueprint'.
    Ask a Question

    Ask a specific question about this device

    K Number
    K113567

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-11-19

    (353 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Synthes MatrixMANDIBLE plate and screw system is intended for oral, maxillofacial surgery:

    • . Trauma
    • . Reconstructive surgery
    • . Orthognathic surgery (surgical correction of dentofacial deformities)
    Device Description

    The Synthes MatrixMANDIBLE Plate and Screw System consists of a variety of plates offered in multiple shapes and sizes and a variety of screws offered in multiple diameters and lengths to meet the anatomical needs of the patient.

    This submission pertains to 1.5 mm thick MatrixMANDIBLE Reconstruction Plates, which are available as left and right single angle reconstruction plates and in three sizes of double angle reconstruction plates. These plates are made from pure titanium and may be offered sterile (non-sterile implants must be sterilized prior to use).

    MatrixMANDIBLE 1.5 mm thick reconstruction plates are intended for single use only.

    AI/ML Overview

    The provided document is a 510(k) Summary for a medical device (MatrixMANDIBLE Plate and Screw System) and does not contain information about acceptance criteria and a study proving device performance in the context of an AI/ML powered device.

    Instead, this document describes a conventional medical device (bone plates and screws) and seeks to demonstrate substantial equivalence to previously cleared predicate devices based on:

    • Identical/similar Indications for Use.
    • Technological Similarities (principles of operation, configurations, material, screw compatibility).
    • Technological Differences (slight width difference, absence of a straight configuration, one more hole in single angle configurations compared to one predicate).
    • Non-Clinical Performance Data: An engineering assessment including cross-section analysis and finite element analysis was used to compare mechanical properties to predicate devices. This assessment determined the mechanical properties were comparable.
    • No Clinical Performance Data: The document explicitly states, "No clinical testing was performed to support this submission."

    Therefore, it is not possible to fill out the requested table and information as it pertains to an AI/ML device, as the provided input describes a different type of medical device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121502

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-08-23

    (94 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Synthes Curvilinear Distraction System is intended for use as a bone stabilizer and lengthening (and/or transport) device. The Synthes Curvilinear Distraction System is indicated for correction of congenital deficiencies or posttraumatic defects of the mandibular body and ramus where gradual bone distraction is required.

    The 2.0 mm Curvilinear Distractor is intended for use in adult and pediatric patients more than 1 year old.

    The 1.3 mm Curvilinear Distractor is intended for use in pediatric patients 4 years of age and younger.

    The Synthes Curvilinear Distraction System is intended for single use only.

    Device Description

    The Synthes Curvilinear Distraction System is a family of internal distraction osteogenesis devices that gradually advance the mandible along a specific trajectory of distraction. The system features various curved and straight distractors in two sizes; the 1.3mm Curvilinear Distractor and the 2.0mm Curvilinear Distractor. The distractors accept extension arms which move the point of activation to a location that is easily accessible with the activation instrument. Synthes Curvilinear Distraction System devices are manufactured from titanium alloy and chromium cobalt alloy. Devices are supplied non-sterile and must be sterilized prior to use.

    The distractor features a worm gear that is activated to move the distractor along a curved or straight track. The distractor consists of three main components: Track, Worm gear activation assembly, and Housing.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for Synthes Curvilinear Distraction System (K121502)

    This report details the acceptance criteria and the study that proves the Synthes Curvilinear Distraction System meets these criteria, based on the provided 510(k) summary (K121502).

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
    Mechanical Strength (Four Point Bend Testing)Withstand forces applied from mastication and resist permanent deformation within the range of predicate devices.The proposed 1.3 mm Curvilinear Distractor can withstand forces applied to it from mastication and can resist permanent deformation within the range of the predicate devices.
    Distraction Force (Torque-Force Testing)Generate sufficient force to overcome anatomical resistance to distraction (soft tissue resistance, callus stretching, etc.).The proposed 1.3 mm Curvilinear Distractor can generate sufficient force to overcome the anatomical resistance to distraction... by over three times the acceptance criterion.
    BiocompatibilityManufactured from titanium alloy and chromium cobalt alloy, meeting respective ASTM standards and having an established history of use as a surgical implant material.The proposed device is manufactured from titanium/titanium alloys and chromium cobalt, each of which meets the requirements of its respective ASTM standard.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify a distinct "test set" in the context of clinical data for performance evaluation. The performance data is derived from non-clinical mechanical testing.

    • Sample Size for Test Set: Not applicable as a clinical test set was not used. The mechanical testing involved specific test articles (Synthes Curvilinear Distraction System, specifically the 1.3 mm Curvilinear Distractor). The exact number of units tested for each mechanical test is not specified in this summary.
    • Data Provenance: The data is from non-clinical mechanical testing conducted by Synthes (USA), a company located in West Chester, PA, United States of America.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. Since clinical testing was not performed and performance was evaluated through mechanical testing, there was no ground truth established by human experts for a clinical test set. The "ground truth" for the mechanical tests would be the established engineering standards and physical properties for medical devices of this type, as measured by standard laboratory equipment and procedures.

    4. Adjudication Method for the Test Set

    Not applicable. As stated above, clinical testing with a human-in-the-loop component was not conducted, rendering an adjudication method unnecessary.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. The document explicitly states: "Clinical testing was not necessary for the determination of substantial equivalence." Therefore, no assessment of human readers' improvement with or without AI assistance was conducted.

    6. Standalone (Algorithm Only) Performance

    Not applicable. This device is a mechanical distraction system, not an algorithm or AI-powered system. Therefore, standalone algorithm performance is not relevant.

    7. Type of Ground Truth Used for Performance Evaluation

    The ground truth used for evaluating the performance of the Synthes Curvilinear Distraction System was based on:

    • Established engineering standards and benchmarks for mechanical strength (four-point bend testing) to demonstrate the ability to withstand mastication forces and resist permanent deformation.
    • Physiological requirements and existing clinical understanding of anatomical resistance to distraction, which served as the basis for the acceptance criterion for the torque-force testing.
    • ASTM standards for material composition and biocompatibility.

    8. Sample Size for the Training Set

    Not applicable. There was no "training set" as this device is a mechanical system and does not involve machine learning or AI that requires training data.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there was no training set, there was no ground truth established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121574

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-06-29

    (30 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Synthes MatrixMANDIBLE plate and screw system is intended for oral, maxillofacial surgery:

    • Trauma
    • Reconstructive surgery
    • Orthognathic surgery (surgical correction of dentofacial deformities)
    Device Description

    The Synthes MatrixMANDIBLE Plate and Screw System consists of a variety of plates offered in multiple shapes and sizes and a variety of screws offered in multiple diameters and lengths to meet the anatomical needs of the patient. System implants are manufactured in either titanium or titanium alloy and are intended for single use only.
    The MatrixMANDIBLE screws that are the subject of this premarket notification are made from titanium alloy (Ti-6Al-7Nb) and are available in a diameter of 2.0 mm and lengths ranging from 4 mm to 8 mm, and have a thread pitch of 0.5 mm. These screws work with all plates within the MatrixMANDIBLE Plate and Screw System.
    These devices are offered non-sterile and must be sterilized prior to use. MatrixMANDIBLE screws are intended for single use.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Synthes MatrixMANDIBLE Plate and Screw System, focusing on acceptance criteria and the supporting study:

    The provided document describes a 510(k) premarket notification for a medical device. This type of submission generally relies on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting extensive clinical efficacy trials. Therefore, the "acceptance criteria" and "study" described will primarily revolve around mechanical performance testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical "acceptance criteria" for the mechanical tests. Instead, it states that the performance of the proposed devices is "equivalent to or better than" the predicate devices. The reported device performance is presented as the results of the mechanical tests, which are implicitly compared against the predicate's performance.

    Test ParameterAcceptance Criteria (Implicit)Reported Device Performance
    Strip-out resistance (N·m)Equivalent to or better than predicate devicesDemonstrated equivalence or superiority to predicate devices (specific values not provided in summary)
    Pull-out strength (N·m)Equivalent to or better than predicate devicesDemonstrated equivalence or superiority to predicate devices (specific values not provided in summary)
    Yield Torque (N·m)Equivalent to or better than predicate devicesDemonstrated equivalence or superiority to predicate devices (specific values not provided in summary)
    Insertion Torque (N·m)Equivalent to or better than predicate devicesDemonstrated equivalence or superiority to predicate devices (specific values not provided in summary)
    Insertion Factor of SafetyEquivalent to or better than predicate devicesDemonstrated equivalence or superiority to predicate devices (specific values not provided in summary)

    Note: The summary document only states that the device's performance was "equivalent to or better" than the predicate. Specific numerical values for the predicate or the proposed device's performance are not provided in this 510(k) summary. For a full understanding of the acceptance criteria (e.g., specific thresholds or statistical equivalence margins) and the detailed results, one would typically need access to the full test report submitted to the FDA.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample sizes used for each mechanical test (e.g., how many screws were tested for pull-out strength).
    • Data Provenance: The testing was "Non-Clinical Testing" performed by the submitter, Synthes (USA), which is based in West Chester, PA, United States of America. The data is retrospective in the sense that it's generated for the submission, but it's new data specific to the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable as the study described is non-clinical mechanical testing, not a clinical study involving expert judgment of medical images or patient outcomes. The "ground truth" here is objective physical force measurements.

    4. Adjudication Method for the Test Set

    This section is not applicable for non-clinical mechanical testing. Adjudication methods like 2+1 or 3+1 are used for resolving discrepancies in expert interpretations in clinical or image-based studies.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission explicitly states: "No clinical testing was performed to support this submission."

    6. Standalone (Algorithm Only Without Human-in-the-loop Performance) Study

    This section is not applicable. This device is a bone plate and screw system, a physical medical device, not an AI algorithm. Therefore, "algorithm only" or "human-in-the-loop" performance are not relevant concepts here.

    7. Type of Ground Truth Used

    The ground truth used for the non-clinical testing was objective physical measurements of mechanical properties (e.g., force, torque) under controlled laboratory conditions, as defined by engineering standards. The "truth" is established by the direct measurement of these physical properties rather than by expert consensus, pathology, or outcomes data.

    8. Sample Size for the Training Set

    This section is not applicable. This is a physical medical device, not a machine learning model, so there is no "training set" in the computational sense.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113251

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-12-21

    (48 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Synthes MatrixMANDIBLE Preformed Reconstruction Plates are intended for use in oral and maxillofacial surgery, trauma and reconstructive surgery. This includes primary mandibular reconstruction, comminuted fractures and temporary bridging pending delayed secondary reconstruction, including fractures of edentulous and/or atrophic mandibles, as well as unstable fractures.

    Device Description

    The Synthes MatrixMANDIBLE Preformed Reconstruction Plates are anatomically contoured to match the body and angle regions of the mandible in most patients. These plates are designed for use with Synthes MatrixMANDIBLE screws that come in multiple diameters and lengths to meet the anatomical needs of the patient. System components are manufactured in either titanium or titanium alloy and are intended for single use only.

    AI/ML Overview

    The acceptance criteria and study proving device performance are described below based on the provided text.

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Dynamic and Static Fatigue Life: The MatrixMANDIBLE Preformed Reconstruction Plates must demonstrate a significantly longer fatigue life compared to the 2.5mm thick MatrixMANDIBLE plates under similar loading conditions.The study concluded that the MatrixMANDIBLE Preformed Reconstruction Plates maintain a significantly longer fatigue life compared to the 2.5mm thick MatrixMANDIBLE plates when tested under similar loading conditions.

    Study Details

    1. Sample Size and Data Provenance:

      • Test Set Sample Size: Not specified. The abstract mentions "dynamic and static testing" but does not give specific sample numbers for these tests.
      • Data Provenance: The testing was "non-clinical testing" conducted by Synthes. The country of origin for the data is not explicitly stated, but Synthes is based in West Chester, PA, USA, implying the testing likely occurred there. The data is retrospective in the sense that it's reported after the tests were conducted, but it's new data generated for this specific submission, not historical patient data.
    2. Number of Experts and Qualifications for Ground Truth (Test Set): Not applicable. This was non-clinical mechanical testing, not a study requiring expert interpretation of medical images or conditions.

    3. Adjudication Method (Test Set): Not applicable due to the nature of the non-clinical mechanical testing.

    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: No. This was a non-clinical study assessing mechanical properties, not a reader study.

    5. Standalone (Algorithm Only) Performance Study: Not applicable. This device is a bone plate, so there is no algorithm or AI component. The performance was assessed through physical mechanical testing.

    6. Type of Ground Truth Used: The ground truth was established by the physical measurements and results obtained from the dynamic and static mechanical testing under controlled laboratory conditions, comparing the new plates to a predicate device's performance.

    7. Training Set Sample Size: Not applicable. This was non-clinical mechanical testing; there was no training set in the context of machine learning or AI.

    8. How Ground Truth for Training Set was Established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111323

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-11-16

    (189 days)

    Product Code
    Regulation Number
    878.3500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Synthes SynPOR HD Porous Polyethylene Three Dimensional Implants are intended for the augmentation or reconstruction of the craniomaxillofacial skeleton.
    Specific indications (SynPOR HD Ocular Spheres):
    • Ocular reconstruction
    • Socket preservation
    Specific indications (SynPOR HD Facial Shapes):
    • Enhancement of the malar and chin
    • Correction of deficiencies of the malar and chin

    Device Description

    SynPOR HD Porous Polyethylene Three Dimensional Implants are devices intended for ocular reconstruction and socket preservation, enhancement of the midface and mandibular skeleton, and correction of deficiencies of the midface and mandible. The porous structure of the HDPE material allows for fibrovascular ingrowth.
    The implants include facial shapes (chin and malar implants for aesthetic and reconstructive surgery to augment the contours of the craniofacial skeleton) and ocular spheres (for restoring the volume of an eviscerated or enucleated eye globe).
    SynPOR HD Porous Polyethylene Implants are delivered sterile for single patient use and should not be resterilized.

    AI/ML Overview

    The requested information about the acceptance criteria and the study proving the device meets them is not available in the provided text. The document is a 510(k) summary for a medical device (Synthes SynPOR HD Porous Polyethylene Three Dimensional Implants), and it primarily focuses on establishing substantial equivalence to a predicate device rather than detailing specific performance acceptance criteria or a study designed to meet them.

    The text does not include:

    1. A table of acceptance criteria and reported device performance.
    2. Details about sample sizes, data provenance, number or qualifications of experts, or adjudication methods for any test set.
    3. Information on multi-reader multi-case (MRMC) comparative effectiveness studies or standalone algorithm performance.
    4. The type of ground truth used or the sample size for a training set.

    The document states that the new device is "substantially equivalent" to predicate devices based on "indications, materials, principles of operation, dimensions, and device design," and that "none of the modifications impact safety and effectiveness." This indicates a regulatory clearance pathway that relies on comparison to existing legally marketed devices, rather than a new study with explicit acceptance criteria for novel performance claims.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112689

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-10-26

    (41 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Synthes Sternal Fixation System is intended for use in primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion.

    The Synthes (USA) Titanium 2.4 mm Universal Locking Plates (12, 13 and 20 hole) are indicated for use in primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion.

    Device Description

    The Synthes (USA) Sternal Fixation System consists of machined titanium plates, a quick-release pin and 3.0 mm locking screws. The plates utilize screw fixation to create the construct.

    AI/ML Overview

    The provided text is a 510(k) summary for the Synthes Sternal Fixation System. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving performance against specific acceptance criteria. Therefore, most of the requested information regarding acceptance criteria, study design, and performance metrics for an AI/device study is not available in these documents.

    Here's a breakdown of what can be extracted and what is not present:

    1. Table of Acceptance Criteria and Reported Device Performance

    Not applicable. The document explicitly states: "Clinical data was not needed for this device." The submission is for a medical device (Sternal Fixation System), not an AI/software device that would typically have performance metrics like sensitivity, specificity, or AUC. The "performance" assessment for this type of device usually involves mechanical testing and comparison to predicate devices, which is not detailed here in terms of acceptance criteria and results.

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable. No test set for performance evaluation of an AI model is mentioned.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. No ground truth establishment for an AI model is mentioned.

    4. Adjudication Method

    Not applicable.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This is not an AI-assisted device, and no MRMC study is mentioned.

    6. Standalone (Algorithm Only) Performance Study

    No. This is a physical medical device (plates and screws), not an algorithm.

    7. Type of Ground Truth Used

    Not applicable.

    8. Sample Size for the Training Set

    Not applicable.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.

    Information Available from the Document:

    • Device Trade Name: Synthes Sternal Fixation System
    • Intended Use/Indications for Use: Primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion.
    • Comparison to Predicate Device: "The design features, material, and indications for use of the subject Sternal Fixation System are substantially equivalent to the predicate devices identified. Additionally, the safety and effectiveness of this system is adequately supported by documentation within this premarket notification."
    • Performance Data (Nonclinical and/or Clinical): "Clinical data was not needed for this device." This implies that non-clinical data (e.g., mechanical testing, material biocompatibility) was used to establish substantial equivalence, though the details of these tests and their criteria are not provided in this summary.

    In summary, the provided document is a 510(k) summary for a physical medical device aiming for substantial equivalence. It does not contain the type of AI-related performance data, acceptance criteria, or study details requested in the prompt.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1