K Number
K121502

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-08-23

(94 days)

Product Code
Regulation Number
872.4760
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Synthes Curvilinear Distraction System is intended for use as a bone stabilizer and lengthening (and/or transport) device. The Synthes Curvilinear Distraction System is indicated for correction of congenital deficiencies or posttraumatic defects of the mandibular body and ramus where gradual bone distraction is required.

The 2.0 mm Curvilinear Distractor is intended for use in adult and pediatric patients more than 1 year old.

The 1.3 mm Curvilinear Distractor is intended for use in pediatric patients 4 years of age and younger.

The Synthes Curvilinear Distraction System is intended for single use only.

Device Description

The Synthes Curvilinear Distraction System is a family of internal distraction osteogenesis devices that gradually advance the mandible along a specific trajectory of distraction. The system features various curved and straight distractors in two sizes; the 1.3mm Curvilinear Distractor and the 2.0mm Curvilinear Distractor. The distractors accept extension arms which move the point of activation to a location that is easily accessible with the activation instrument. Synthes Curvilinear Distraction System devices are manufactured from titanium alloy and chromium cobalt alloy. Devices are supplied non-sterile and must be sterilized prior to use.

The distractor features a worm gear that is activated to move the distractor along a curved or straight track. The distractor consists of three main components: Track, Worm gear activation assembly, and Housing.

AI/ML Overview

Acceptance Criteria and Device Performance Study for Synthes Curvilinear Distraction System (K121502)

This report details the acceptance criteria and the study that proves the Synthes Curvilinear Distraction System meets these criteria, based on the provided 510(k) summary (K121502).

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
Mechanical Strength (Four Point Bend Testing)Withstand forces applied from mastication and resist permanent deformation within the range of predicate devices.The proposed 1.3 mm Curvilinear Distractor can withstand forces applied to it from mastication and can resist permanent deformation within the range of the predicate devices.
Distraction Force (Torque-Force Testing)Generate sufficient force to overcome anatomical resistance to distraction (soft tissue resistance, callus stretching, etc.).The proposed 1.3 mm Curvilinear Distractor can generate sufficient force to overcome the anatomical resistance to distraction... by over three times the acceptance criterion.
BiocompatibilityManufactured from titanium alloy and chromium cobalt alloy, meeting respective ASTM standards and having an established history of use as a surgical implant material.The proposed device is manufactured from titanium/titanium alloys and chromium cobalt, each of which meets the requirements of its respective ASTM standard.

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not specify a distinct "test set" in the context of clinical data for performance evaluation. The performance data is derived from non-clinical mechanical testing.

  • Sample Size for Test Set: Not applicable as a clinical test set was not used. The mechanical testing involved specific test articles (Synthes Curvilinear Distraction System, specifically the 1.3 mm Curvilinear Distractor). The exact number of units tested for each mechanical test is not specified in this summary.
  • Data Provenance: The data is from non-clinical mechanical testing conducted by Synthes (USA), a company located in West Chester, PA, United States of America.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

Not applicable. Since clinical testing was not performed and performance was evaluated through mechanical testing, there was no ground truth established by human experts for a clinical test set. The "ground truth" for the mechanical tests would be the established engineering standards and physical properties for medical devices of this type, as measured by standard laboratory equipment and procedures.

4. Adjudication Method for the Test Set

Not applicable. As stated above, clinical testing with a human-in-the-loop component was not conducted, rendering an adjudication method unnecessary.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. The document explicitly states: "Clinical testing was not necessary for the determination of substantial equivalence." Therefore, no assessment of human readers' improvement with or without AI assistance was conducted.

6. Standalone (Algorithm Only) Performance

Not applicable. This device is a mechanical distraction system, not an algorithm or AI-powered system. Therefore, standalone algorithm performance is not relevant.

7. Type of Ground Truth Used for Performance Evaluation

The ground truth used for evaluating the performance of the Synthes Curvilinear Distraction System was based on:

  • Established engineering standards and benchmarks for mechanical strength (four-point bend testing) to demonstrate the ability to withstand mastication forces and resist permanent deformation.
  • Physiological requirements and existing clinical understanding of anatomical resistance to distraction, which served as the basis for the acceptance criterion for the torque-force testing.
  • ASTM standards for material composition and biocompatibility.

8. Sample Size for the Training Set

Not applicable. There was no "training set" as this device is a mechanical system and does not involve machine learning or AI that requires training data.

9. How the Ground Truth for the Training Set was Established

Not applicable. As there was no training set, there was no ground truth established for it.

{0}------------------------------------------------

K121502

Image /page/0/Picture/1 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a stylized image inside. A horizontal line runs underneath the word. A small registered trademark symbol is located to the right of the word.

5 510(k) Summary

AUG 2 3 2012

Page 1 of 3

Date PreparedMay 10, 2012
SubmitterSynthes (USA)1301 Goshen ParkwayWest Chester, PA 19380United States of America
ContactAlan T. Haleyhaley.alan@synthes.com(484) 356-9763
Device NameSynthes Curvilinear Distraction System(Mandibular distractor)
Classification NameClass II, 21 CFR 872.4760, Product Code MQN

Predicate Devices

  • Synthes Curvilinear Distraction System (K080153) ●
  • Synthes Craniomaxillofacial (CMF) Distraction System (K060138) ●
  • OsteoMed Intraoral Mandibular Distraction System (K013618) .
  • OsteoMed Pediatric Intraoral Mandibular Distraction System (K043434) .

Indications for Use

The Synthes Curvilinear Distraction System is intended for use as a bone stabilizer and lengthening (and/or transport) device. The Synthes Curvilinear Distraction System is indicated for correction of congenital deficiencies or posttraumatic defects of the mandibular body and ramus where gradual bone distraction is required.

The 2.0 mm Curvilinear Distractor is intended for use in adult and pediatric patients more than 1 year old.

The 1.3 mm Curvilinear Distractor is intended for use in pediatric patients 4 years of age and younger.

The Synthes Curvilinear Distraction System is intended for single use only.

Device Description

The Synthes Curvilinear Distraction System is a family of internal distraction osteogenesis devices that gradually advance the mandible along a specific trajectory of distraction. The system features various curved and straight distractors in two sizes; the 1.3mm Curvilinear Distractor and the 2.0mm Curvilinear Distractor. The distractors accept extension arms which move the point of activation to a location that is easily accessible with the activation instrument. Synthes Curvilinear Distraction System devices are manufactured from titanium alloy and chromium cobalt alloy. Devices are supplied non-sterile and must be sterilized prior to use.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a stylized image inside. A horizontal line runs underneath the word "SYNTHES" and the logo. The logo and text are likely part of a company's branding.

510(k) Summary 5

Device Description (continued)

The distractor features a worm gear that is activated to move the distractor along a curved or straight track. The distractor consists of three main components:

TrackThe track has grooves in 1 mm intervals which may be placed in astraight line (for a straight distractor) or on a centerline radius (for acurved distractor). The track is manufactured with a crimp that servesas a functional stop to prevent the distractor from separating at the endof the track.
The track is 35mm in length and can be cut to the desired length foreach particular patient by the surgeon. After cutting, the track iscrimped to re-establish the functional stop to prevent separation.
Worm gear activationassemblyThe worm activation assembly consists of the worm gear and auniversal joint activation hex. The universal joint is capable of + or -35° of angulation. The worm gear has a 1 mm pitch and rides alongthe grooves cut into the track. The worm gear activation assembly isinserted into the housing and the track with grooves is threadedthrough a slot in the side of the housing.
HousingThe housing includes a tab that lays on the activation assembly toprevent the distractor from reversing due to micromotion.

Comparison to Predicate Devices

Indications

The Indications statement for the proposed device is similar to the statements of all of the predicate devices in that they are to be used for distraction osteogenesis of the mandible. Although the wording for each specific Indications statement varies, they are all cleared for the same clinical application - bone stabilization and lengthening (and/or transport) of the mandible where gradual bone distraction is required.

Both the proposed device and the OsteoMed Pediatric Intraoral Mandibular Distraction System (K043434) predicate are intended for use in patients under 4 years of age. The proposed device is similar to the Synthes Craniomaxillofacial (CMF) Distraction System (K060138) predicate which includes devices for patients under the age of 12 months as well as devices for patients over 1 year of age.

The differences in the Indications statement for the proposed device in comparison to the predicates do not constitute a new intended use.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a design inside. A horizontal line is present below the word "SYNTHES". The registered trademark symbol is present to the right of the word.

5 510(k) Summary

Page 3 of 3

Comparison to Predicate Devices (continued)

Technological Similarities

  • The proposed device and the predicate devices consist of mesh footplates designed to attach . to the mandible with bone screws.
  • The proposed device and the predicate devices include an advancement mechanism designed . to move the mobilized segment(s) of the mandible to gencratc new bone as part of the distraction osteogenesis treatment process.
  • The proposed device uses the same worm drive mechanism as Synthes Curvilinear o Distraction System (K080153) predicate.
  • The proposed device and the predicate devices are manufactured from titanium/titanium . alloys and chromium cobalt, each of which meets the requirements of its respective ASTM standard. Titanium and titanium alloys have a long, established history of use as a surgical implant material.

Technological Differences

  • The advancement mechanism for the proposed device is a worm drive, whereas the Synthes . Craniomaxillofacial (CMF) Distraction System (K060138) the OsteoMed Intraoral Mandibular Distraction System (K013618), and OsteoMed Pediatric Intraoral Mandibular Distraction System (K043434) predicates employ a center lead screw.

Non-clinical performance data

Mechanical testing was used to demonstrate that any differences, where they do exist, do not negatively impact safety and effectiveness.

Four point bend testing was performed to demonstrate that the proposed 1.3 mm Curvilinear Distractor can withstand forces applied to it from mastication and can resist permanent deformation within the range of the predicate devices.

Torque-force testing was also conducted to show that the proposed 1.3 mm Curvilinear Distractor can generate sufficient force to overcome the anatomical resistance to distraction (soft tissue resistance, callus stretching, etc.) by over three times the acceptance criterion.

Clinical performance data

Clinical testing was not necessary for the determination of substantial equivalence.

Substantial Equivalence

The proposed device has the same intended use as the predicate devices. The mechanical testing included in this submission demonstrates that the slight differences in technological menaceristics do not raise any new questions of safety and effectiveness and that the proposed device is at least as safe and effective as the predicates. The information submitted supports substantial equivalence.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

· Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

AUG 2 3 2012

Synthes Incorporated Mr. Alan T. Haley CMF Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, Pennsylvania 19380

Re: K121502

Trade/Device Name: Synthes Curvilinear Distraction System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: MQN Dated: August 10, 2012 Received: August 13, 2012

Dear Mr. Haley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Mr. Haley

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

To. D. Stut

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices

Office of Device Evaluation

Center for Devices and

Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a stylized figure inside. A horizontal line runs beneath the word, adding a sense of completion to the design. The overall impression is one of a professional and established brand.

Indications for Use Statement 4

510(k) Number (if known): __

Device Name:

Synthes Curvilinear Distraction System

Indications for Use:

The Synthes Curvilinear Distraction System is intended for use as a bone stabilizer and lengthening (and/or transport) device. The Synthes Curvilinear Distraction System is indicated for correction of congenital deficiencies or posttraumatic defects of the mandibular body and ramus where gradual bone distraction is required.

K121502

The 2.0 mm Curvilinear Distractor is intended for use in adult and pediatric patients more than 1 year old.

The 1.3 mm Curvilinear Distractor is intended for use in pediatric patients 4 years of age and younger.

The Synthes Curvilinear Distraction System is intended for single use only.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Suver Runner

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K121502

§ 872.4760 Bone plate.

(a)
Identification. A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.(b)
Classification. Class II.