510(k) Data Aggregation

    K Number
    K233098

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-02

    (311 days)

    Product Code
    Regulation Number
    890.5850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MYOTouch Muscle Stimulator is indicated for use as an adjunct in the treatment of high and low anorectic malformations by helping with the identification of striated muscles to be used in anal reconstructions. The MYOTouch device is intended for use by trained personnel who are competent to apply the appropriate stimuli.

    Device Description

    MYOTouch is a hand-held device designed to be used for surgical correction of neonatal anorectal malformations. MYOTouch is intended for prescription use. The device is a 9V battery-powered, one-channel muscle stimulator. The device is supplied with a Sterile Stylus, which connects to the control unit by cable and plug. The MYOTouch muscle stimulator has been specially designed to meet the requirements of pediatric surgeons who need a reliable and easy to use muscle locating stimulator when carrying out the Pena, "Pull-through", or posterior sagittal anorectoplasty (PSARP) anorectal reconstruction technique. A trained professional or pediatric surgeon must use the MYOTouch device. The pediatric surgeon uses manual, push-button controls to control the electrical stimulation level. The unit is intended for prescription use and is designed with simplicity and ease of use. It has two preset modes available.

    AI/ML Overview

    I am sorry, but the provided text is a 510(k) premarket notification for a medical device called the MYOTouch Muscle Stimulator. It describes the device, its intended use, and compares it to a predicate device to establish substantial equivalence.

    However, the document explicitly states in Section VIII, "No clinical tests were submitted to support this 510(k)."

    Therefore, based on the information provided, I cannot answer your request regarding the acceptance criteria and the study that proves the device meets those criteria, as no such study is presented in this document. The submission relies on demonstrating substantial equivalence to a legally marketed predicate device through technical comparisons and compliance with recognized standards, rather than direct performance studies against acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121743

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-03-28

    (288 days)

    Product Code
    Regulation Number
    868.2775
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EasyMed InstrumentsCo., LTD. Peripheral Nerve Stimulator is a battery -powered device intended for monitoring the magnitude of neuromuscular blocks in general anesthesia; by delivering an electrical stimulus near a peripheral motor nerve.

    Device Description

    The EasyMed Peripheral Nerve Stimulator devices are battery powered peripheral nerve stimulators which provide low electrical direct current (DC) stimulation in order to determine the level of anesthetic nerve block. The Stimulus Amplitude control dial provides variable current control (0 to 70 mA into a 2K ohm load).

    The Output Stimulus Pulse Indicator flashes each time current passes through the patient. Functions include: Train-of-Four, Twitch, and 100Hz Tetanus. The device offers sufficient output to ensure supramaximal stimulation: from 0 to 70 mA.

    All finished products are tested and must meet all required release specifications before distribution. Testing includes physical testing (e.g. Pulse Amplitude, Pulse Width, Pulse Frequency and Low Battery Voltage Indicator) and visual examination.

    AI/ML Overview

    This document describes the validation of the EasyMed Instruments Co., LTD. Peripheral Nerve Stimulator.

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the EasyMed Peripheral Nerve Stimulator are based on compliance with specific electromagnetic compatibility (EMC) standards. The reported device performance indicates that the device passed all these tests.

    Test TypeTest RequirementReported Device Performance
    Radiated Emission (30 MHz to 1 GHz)EN 60601-1-2: 2007Pass
    ESD SusceptibilityEN 60601-1-2: 2007Pass
    Radiated Immunity (90 MHz to 2.5 GHz)EN 60601-1-2: 2007Pass
    Power-frequency magnetic field immunityEN 60601-1-2: 2007Pass

    Beyond these EMC specific tests, the document also mentions that "All finished products are tested and must meet all required release specifications before distribution. Testing includes physical testing (e.g. Pulse Amplitude, Pulse Width, Pulse Frequency and Low Battery Voltage Indicator) and visual examination." However, the specific acceptance criteria for these physical tests are not detailed in the provided text.

    2. Sample Size and Data Provenance

    The provided document does not specify the sample size used for the tests mentioned in the "Performance Data" section.
    The provenance of the data is not explicitly stated as "country of origin of the data" or "retrospective or prospective." However, given EasyMed Instruments Co., LTD. is located in China, it's reasonable to infer that the testing likely occurred in China. The testing described appears to be prospective testing of the device itself, rather than retrospective analysis of clinical data.

    3. Number and Qualifications of Experts for Ground Truth

    Not applicable. This document describes performance testing for an electrical medical device (peripheral nerve stimulator) against engineering standards (e.g., electromagnetic compatibility). The nature of this testing does not involve establishing ground truth based on expert review of medical images or clinical cases.

    4. Adjudication Method for the Test Set

    Not applicable. As the testing described is for engineering performance rather than clinical data interpretation, there is no adjudication method involved. The results are binary (Pass/Fail) based on predefined technical standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned or performed. This validation is for a medical device (a peripheral nerve stimulator), not an AI algorithm.

    6. Standalone (Algorithm Only) Performance

    Not applicable. This document describes the validation of a physical medical device, not an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for the tests described is the established technical requirements and limits defined by the international standards (e.g., EN 60601-1-2, EN 55011, EN 61000-4-2, EN 61000-4-3, EN 61000-4-8). The device's performance is measured against these objective, verifiable criteria.

    8. Sample Size for the Training Set

    Not applicable. This device is hardware, not a machine learning model, so there is no concept of a "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no training set, there is no ground truth for a training set to be established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081805

    Validate with FDA (Live)

    Device Name
    CUFFALERT
    Manufacturer
    Date Cleared
    2008-11-14

    (141 days)

    Product Code
    Regulation Number
    868.5750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To measure and monitor intra-cuff pressures of endotracheal, supraglottic airways, or tracheostomy tubes.
    The CuffAlert™ is not intended to replacement clinical judgment.

    Device Description

    The CuffAlert™ is a simple in-line pressure monitor which contains a calibrated diaphragm which distends with increases in cuff pressure.
    One can set the maximum pressure limit from 10 to 40 cm H2O by rotating and locking a knob. Once set, it is placed on the pilot ballowing it to be in-line with the cuff. If the cuff exceeds the set pressure limit on the CuffAlert™ the diaphragm distends making contact with the battery, thus activating the red LED.
    The accuracy of pressure measurement has been determined to be +/- 2 cm H2O @ 10 cm H2O; +/- 3 cm H2O @ 20 - 30 cm H2O; and +/- 4 cm H2O @ 40 cm H2O.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the CuffAlert™ device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Required Performance)Reported Device Performance (CuffAlert™)
    +/- 2 cm H₂O @ 10 cm H₂O+/- 2 cm H₂O @ 10 cm H₂O
    +/- 3 cm H₂O @ 20 - 30 cm H₂O+/- 3 cm H₂O @ 20 - 30 cm H₂O
    +/- 4 cm H₂O @ 40 cm H₂O+/- 4 cm H₂O @ 40 cm H₂O

    Note: The document describes the "accuracy of pressure measurement has been determined to be" these values. This language implies these are both the acceptance criteria and the (successful) reported performance.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the study). It only states that the accuracy "has been determined."

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The document does not provide information regarding the number of experts, their qualifications, or their involvement in establishing ground truth for any testing of the device.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method used for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it discuss human reader improvement with or without AI assistance. This device is a physical pressure monitor, not an AI-assisted diagnostic tool.

    6. Standalone (Algorithm Only) Performance Study

    The device is a physical pressure monitor and does not utilize an algorithm in the way a diagnostic AI would. Therefore, a "standalone (algorithm only)" performance study, as typically understood for AI, is not applicable and not mentioned. The stated accuracy is for the device's inherent mechanical function.

    7. Type of Ground Truth Used

    The ground truth for the accuracy assessment of a pressure monitor would typically be established by comparison against a calibrated reference standard pressure measurement device. While the document doesn't explicitly state "calibrated reference standard," in the context of pressure measurement accuracy, this is the implied method for "determining" the accuracy.

    8. Sample Size for the Training Set

    The document does not specify a training set sample size. This type of device (a mechanical pressure monitor) would not typically involve a "training set" in the context of machine learning or AI. Its performance is based on its design and physical calibration.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a "training set" in the context of AI/ML, this information is not applicable to the CuffAlert™ device. Its ground truth (accuracy) would be established through physical calibration against known pressure standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062112

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-06-04

    (315 days)

    Product Code
    Regulation Number
    876.5130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SunMed Foley Catheter is used to drain fluids to and from the urinary tract.

    Device Description

    The catheters are comprised of a 2 lumen shaft with proximal funnel, inflation valve and distal retaining balloon. One lumen is for draining fuleds to and from the urinary tract. The second lumen is to inflate and deflate the balloon. On models with a third lumen, it is used in conjunction with the first lumen for flushing the urinary tract. The balloon fill volumes in ml and shaft size in French Gauge (Fr), Charriere (Ch) or millimeter (mm) are indicated on the funnel of each individual cathter and the distal tip type can be indicated on the individual pack label and / or outer carton. Sizes range from 6 to 30 Ch / Fr.

    AI/ML Overview

    The provided text describes a 510(k) submission for the SunMed Foley Catheter, which is a medical device. This type of submission is for demonstrating substantial equivalence to a predicate device, not for clinical studies that establish acceptance criteria for an AI/algorithm-driven device.

    Therefore, the input does not contain the information necessary to answer the questions about acceptance criteria and a study proving device performance as typically understood for AI/algorithm devices.

    Specifically, the document discusses:

    • Intended Use: Draining fluids to and from the urinary tract.
    • Product Description: Two-lumen shaft with proximal funnel, inflation valve, and distal retaining balloon.
    • Predicate Device: Rusch Silicon Coated Foley Catheter (K980870).
    • Substantial Equivalence: Claimed due to equivalent intended use, manufacturing materials, operating principles, and physical, operational, and biological specifications.

    There is no mention of an AI/algorithm, performance metrics, acceptance criteria, test sets, ground truth establishment, expert adjudication, or MRMC comparative effectiveness studies.

    In summary, the provided document relates to a traditional medical device (Foley catheter) and does not include the details required to answer your specific questions, which are geared towards the evaluation of AI/algorithm-driven medical devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K955632

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-02-27

    (78 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K955631

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-02-27

    (78 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1