Search Filters

Search Results

Found 35 results

510(k) Data Aggregation

    K Number
    K252871

    Validate with FDA (Live)

    Date Cleared
    2026-02-06

    (149 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K243491

    Validate with FDA (Live)

    Date Cleared
    2024-12-11

    (29 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SternalPlate System is indicated for use in stabilization and fixation of fractures of the anterior chest wall including sternal fixation following sternotomy, sternal fracture(s), and sternal reconstructive surgical procedures to promote bony fusion.

    Device Description

    This Special 510(k) is submitted to evaluate the modifications to the Predicate Device plates resulting in the Subject Device plates. There have been no changes to the previously cleared plates and screws cleared in K183172, this submission is only for the addition of the four new Subject Device plates. The four Subject Device plates include T-Plates and Transverse Plates. Both Subject Device variants, the T-Plates and Transverse Plates, are geometrically derived from the following Predicate Device plates: the Ladder Plate Narrow (Reff/ 4740006) and the Straight 24-hole Plate (Ref# 4740012). The Subject Device plates therefore have the same features, the same tolerances, and are made from the same material as the Predicate Device plates. Furthermore, the Subject Device plates are manufactured with the same manufacturing process as the Predicate Device plates.

    AI/ML Overview

    The provided FDA 510(k) summary (K243491) for the Stryker SternalPlate Expansion device focuses on the mechanical performance of the device and its biocompatibility. The document does not contain information related to software, AI, or human-in-the-loop performance. Therefore, many of the requested criteria cannot be extracted from this document.

    Here's the information that can be extracted or deduced from the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document states that the "Subject Device plates met all acceptance criteria as compared to the Additional Predicate Devices." However, the specific quantitative acceptance criteria (e.g., minimum stiffness, maximum deflection) and the detailed reported performance values are not explicitly stated in the provided 510(k) summary. The summary only mentions "appropriate strength and stiffness."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document refers to "Performance Bench Testing" but does not specify the sample size for the mechanical tests. The tests compare the Subject Device to "Additional Predicate Devices," implying a comparative mechanical study rather than a clinical data-driven study. Data provenance (country of origin, retrospective/prospective) is not applicable as it's a bench test, not clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable as the document describes mechanical bench testing of an orthopedic implant, not a diagnostic device requiring expert interpretation of images or data to establish a ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable as the document describes mechanical bench testing. Adjudication methods are typically relevant for clinical studies or studies involving human readers/interpreters.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study. This document is for a physical orthopedic implant, not an AI-powered diagnostic tool. Therefore, the effect size of human readers with/without AI assistance is not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the device is a physical orthopedic implant, not a standalone algorithm or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the performance bench testing, the "ground truth" would be the engineering specifications and established mechanical testing standards against which the device's strength and stiffness are evaluated. The document states "The Subject Device plates met all acceptance criteria as compared to the Additional Predicate Devices," implying a comparison against established performance benchmarks of legally marketed devices.

    8. The sample size for the training set

    This question is not applicable as the device is a physical orthopedic implant. There is no concept of a "training set" in the context of mechanical bench testing for this type of device.

    9. How the ground truth for the training set was established

    This question is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241517

    Validate with FDA (Live)

    Date Cleared
    2024-08-27

    (90 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Q Guidance System, when used with the Spine Guidance 5.2 Software, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery in adult and pediatric (adolescent) patients.

    The system is indicated for any surgical procedure on the spine in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure such as the skull, pelvis or spine can be identified.

    The system assists in the positioning of instruments for procedures on the pelvis and spine, including:
    · Screw and Needle Placement in the spine or pelvis
    · Bone resection in the spine

    The Mako Spine System is intended to be used as an accessory to the Stryker Spine Guidance System. It is intended to be used to facilitate preparation and placement of screws of Stryker Spine implants in the non-cervical spine in adult and pediatric (adolescent) patients.

    The Mako Spine System is intended to be used as part of the Stryker Spine Guidance System, which is indicated for any surgical procedure on the spine in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure such as the spine or pelvis can be identified.

    The Stryker System 8 Dual Trigger Rotary Battery Powered Heavy Duty System is intended for use in the drilling, reaming, and decorticating of bone and other bone related tissue in a variety of surgical procedures. They are also usable in the placement of screws, wires, pins, and other fixation devices.

    The Stryker System 8 Dual Trigger Rotary Battery Powered Heavy Duty System, when used in conjunction with Stryker Spine Guidance Software, is intended to be used to facilitate preparation and placement of screws of Stryker Spine implant systems in adult and pediatric (adolescent) patients.

    Device Description

    The Spine Guidance System is a computer-assisted stereotaxic, image-guided, planning, and intraoperative guidance system intended to enable open or percutaneous computer-assisted surgery on the spine and pelvis. It assists the surgeon in precisely positioning manual and powered instruments and locating patient anatomy during spinal surgery. The system is comprised of the Spine Guidance 5.2 Software, Q Guidance System (computer platform), Mako Spine System, navigated accessories/ instruments (e.g., patient/ instrument trackers, pointers), and various system components (e.g., trackers, calibration instruments, etc.).

    The system provides intraoperative guidance to the surgeon using passive and active wireless optical tracking technologies as well as haptic trajectory guidance to facilitate pedicle preparation and screw placement using Mako Spine System.

    The Spine Guidance 5.2 Software displays the intraoperative location of navigated surgical instruments relative to imported patient medical images via wireless optical tracking technology. The software provides the functions to perform the indicated navigated spine surgical procedures. The software guides the user through the necessary preoperative and intraoperative steps required to set-up and perform the navigated spine surgical procedures.

    The Mako Spine System is an accessory to the subject Spine Guidance 5.2 (SG5.2) Software and extends the functionality of the SG5.2 Software to provide trajectory guidance to align the End Effector tube with a pre-planned screw trajectory to facilitate preparation and placement of screws in the non-cervical spine. The Mako Spine System includes a Mako 3.5 Robotic arm, new dedicated instrumentations (i.e. Spine Pedicle End Effector, Spine Pedicle Registration Tool, Pedicle Instruments, etc.), and previously cleared compatible instruments (i.e. Vizadisc, Rio System Quick Connect Base Array, End Effector Tracker Array, Calibration End-Effector, etc.). Mako 3.5 Robotic Arm is connected to the existing Q Guidance System (K233542) via Mako ethernet cable.

    The System 8 Dual Trigger Rotary Handpiece is used in the drilling, reaming, and decorticating of bone and other bone related tissue in a variety of surgical procedures. It is also used in the placement of screws, wires, pins, and other fixation devices. The handpiece, when used with a compatible instrument tracker, can be used with the Spine Guidance 5.2 Software to facilitate preparation and placement of screws using Stryker Spine implant systems.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study that proves the device meets specific performance metrics. Instead, the document is an FDA 510(k) clearance letter for the Stryker Q Guidance System, Mako Spine System, and System 8 Rotary Handpiece.

    A 510(k) submission generally focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than detailed performance studies with specific statistical acceptance criteria as would be found in a clinical trial or a robust standalone AI performance study.

    The document mentions "non-clinical performance testing" which includes:

    • Software testing as required by IEC 623304 and FDA Guidance on General Principles of Software Validation.
    • System accuracy verification per ASTM F2554.
    • Full system cadaver validation.
    • Biocompatibility verification according to ISO 10993-1:2018.
    • Sterilization validation of reusable devices per ISO 17665-1.
    • Shelf-life validated per ISO 11607-1:2019, ISO 11607-2:2019, ASTM F1980-21.
    • Electrical Safety and Electromagnetic Compatibility verification to IEC 60601-1 and IEC 60601-1-2.

    However, it does not provide details on:

    1. A table of acceptance criteria and reported device performance. While "System accuracy verification per ASTM F2554" is mentioned, the specific metrics, thresholds, and results are not provided.
    2. Sample size, data provenance, number of experts, adjudication methods for any test set, or details of any MRMC study. The document states "No clinical testing was required to support this submission," which means there would be no human reader study or corresponding ground truth establishment on a clinical test set as described in your prompt.
    3. Standalone (algorithm-only) performance results, as this device is a guidance system, not a diagnostic AI algorithm.
    4. Type of ground truth used in the context of an AI performance study, as such a study was not required or detailed.
    5. Sample size for the training set or how ground truth was established for it. This information pertains to AI model development, which is not the focus of this 510(k) submission document.

    In summary, based on the provided text, it is not possible to answer your questions regarding acceptance criteria, performance data, test set details, ground truth establishment, or AI-specific study methodologies, because the document is a 510(k) clearance that explicitly states "No clinical testing was required to support this submission" and focuses on demonstrating substantial equivalence through non-clinical performance and other regulatory compliance tests.

    The document does describe the Indications for Use for the Q Guidance System and Mako Spine System, which outlines the intended purpose and patient population. It also provides a Summary of Technological Characteristics comparing the subject devices to predicate devices. However, this is for demonstration of substantial equivalence, not a detailed performance study with quantifiable acceptance criteria and results as typically seen for AI/ML devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241171

    Validate with FDA (Live)

    Date Cleared
    2024-07-25

    (90 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K240651

    Validate with FDA (Live)

    Device Name
    MRI Universal
    Date Cleared
    2024-07-12

    (127 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Universal CMF System Intended use The Universal CMF System is intended for stabilization and rigid fixation of the craniomaxillofacial (CMF) skeleton. Indications for use Craniomaxillofacial Implants The Universal CMF System is Craniomaxillofacial (CMF) plate and screw system intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Mandible Implants The Universal CMF System (mandible modules) is a mandibular plate and screw system intended for stabilization and rigid fixation of mandibular fractures and mandibular reconstruction. MP Lefort I Plates Intended Use The Stryker Leibinger Universal CMF and Universal 2.0 Mini Plating Systems are Craniomaxillofacial (CMF) plate and screw systems intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Indications for Use The MP LeFort I Plates as part of the predicate subgroup Universal 2.0 Mini Plating System are implants indicated for osteotomy, stabilization, and rigid fixation of LeFort I fractures of the maxillofacial skeleton. Stryker Upper-Face AXS Screws And Mid-Face AXS Screws Indications for Use The Stryker Universal CMF System is a Cranio-maxillofacial (CMF) plate and screw system intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Stryker MP, Mandible, HMMF And MMF AXS Screws -Universal CMF System: Intended Use The Universal CMF System is intended for stabilization and rigid fixation of the craniomaxillofacial (CMF) skeleton. Indications for Use Craniomaxillofacial Implants The Universal CMF System is Craniomaxillofacial (CMF) plate and screw system intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Mandible Implants The Universal CMF System (mandible modules) is a mandibular plate and screw system intended for stabilization and rigid fixation of mandibular fractures and mandibular reconstruction. -Stryker Universal SMARTLock Hybrid MMF System: Intended Use The Stryker Universal SMARTLock Hybrid MMF System is intended to be used for temporary stabilization of mandibular and maxillary fractures in order to maintain proper occlusion during fracture healing. Indications for Use The Stryker Universal SMARTLock Hybrid MMF System is indicated for the treatment of mandibular and maxillary fractures in adults and adolescents (age 12 and higher) in whom permanent teeth have erupted. -Stryker MMF Screw Intended use The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion. Indications for use The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion. Stryker Universal Orbital Floor System Intended Use The Stryker Universal Floor System is intended to be used in the reconstruction of the floor and/or medial wall of the orbit. Indications for Use The Stryker Universal Orbital Floor System is indicated for the reconstructive treatment of orbital floor and/or medial wall trauma or bone excision in patients 15 years of age and older. Stryker Universal Smartlock Hybrid MMF System Intended use The Stryker Universal SMARTLock Hybrid MMF System is intended to be used for temporary stabilization of mandibular and maxillary fractures in order to maintain proper occlusion during fracture healing. Indications for use The Stryker Universal SMARTLock Hybrid MMF System is indicated for the treatment of mandibular and maxillary fractures in adults and adolescents (age 12 and higher) in whom permanent teeth have erupted. Stryker MMF Screw Intended use The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion. Indications for use The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion.

    Device Description

    This Bundled. Traditional 510(k) is for a single change the MR status that affects multiple devices and can be assessed during one review. There are no changes to the predicate devices themselves, just the addition to allow MR Conditional use of the implants in scope after MRI compatibility testing was completed. Associated labeling for each Predicate Device is also updated to reflect the MR conditional use of the implants.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Q Guidance System, when used with the Spine Guidance Software, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery in adult and pediatric (adolescent) patients.

    The system is indicated for any surgical procedure on the spine in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure such as the skull, pelvis or spine can be identified.

    The system assists in the positioning of instruments for procedures on the pelvis and spine, including:
    • Screw and Needle placement in the spine or pelvis.

    Device Description

    The Spine Guidance 5.0 system is an image guided stereotaxic, planning, and intraoperative guidance system intended to enable open or percutaneous computer-assisted surgery. It assists the surgeon in precisely positioning manual and powered instruments and locating patient anatomy during spinal surgery.

    The system is comprised of the Spine Guidance 5.0 Software, the Q Guidance System (computer platform), navigated accessories/ instruments (e.g., powered drills, pointers), and various system components (e.g., calibration devices, navigation adaptors, patient/ instrument trackers, etc.). The system provides intraoperative guidance to the surgeon using passive and active wireless optical tracking technologies. The computer platform consists of a computer, camera, big touchscreen monitor, and a small touchscreen monitor. The Spine Guidance 5.0 software functionality is described in terms of its capabilities that feature planning, registration, and navigation of medical devices.

    AI/ML Overview

    The provided document is an FDA 510(k) summary for the Stryker Spine Guidance Software (version 5.0) and associated instruments. It details the device's characteristics, intended use, and comparison to predicate devices, as well as the non-clinical testing performed to establish substantial equivalence.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    AccuracySystem demonstrated performance in 3D positional accuracy with a mean error ≤ 2.0 mm and in trajectory angle accuracy with a mean error < 2.0 degrees. This performance was determined using anatomically representative phantoms and utilizing a subset of system components and features that represent the worst-case combinations of all potential system components.
    SoftwareSoftware verification and validation testing was conducted as required by IEC 62304 and FDA Guidance on General Principles of Software Validation, January 11, 2002. All requirements were met, and no new issues of safety or effectiveness were raised.
    BiocompatibilityThe biocompatibility of all patient contact materials was verified according to ISO 10993-1:2018 and the FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, September 2023. No new issues of safety or effectiveness were raised.
    Electrical Safety and Electromagnetic CompatibilityVerified conformance to IEC 60601-1:2005. IEC 60601-1:2005/AMD1:2012. IEC 60601-1:2005/AMD2:2020. Verified conformance to IEC 60601-1-2:2014, IEC 60601-1-2:2014/AMD1:2020, CISPR 11 Group 1, Class A requirements as well as additional testing to verify compatibility with RFID devices operating in the 125-134 kHz and 13.56 MHz frequency band.
    SterilizationThe reusable subject devices underwent a steam sterilization to demonstrate that they can be expected to be sterile and have a sterility assurance level (SAL) of 10-6 or greater after processing. The Tracking Instrument Battery underwent gamma sterilization validation per ISO 11137-2: 2013/(R) 2019 to demonstrate that they can be expected to be sterile and have a sterility assurance level (SAL) of 10-6.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not specify the exact sample size for the test set used in the accuracy testing. It mentions "anatomically representative phantoms" and "a subset of system components and features that represent the worst-case combinations of all potential system components." The data provenance (country of origin, retrospective/prospective) is not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not mention the use of experts to establish a ground truth for the test set. The accuracy testing was performed using "anatomically representative phantoms," implying a technical or engineering validation rather than clinical or expert-based ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    No adjudication method for a test set is described, as the testing focused on technical accuracy using phantoms, not on human interpretation or clinical outcomes adjudicated by experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document explicitly states: "No clinical testing was required to support this submission." Therefore, no MRMC comparative effectiveness study was performed or reported. The device is a guidance system, not an AI diagnostic tool requiring human reader improvement comparison.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The document describes non-clinical testing for system accuracy and software validation. The reported accuracy of "mean error ≤ 2.0 mm for positional displacement and < 2.0 degrees for trajectory angle displacement" refers to the standalone performance of the system as measured against known phantom parameters. This suggests that the algorithm's performance in guiding instruments was assessed directly against a predefined ground truth in a controlled environment.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the accuracy testing, the ground truth was established by the precise measurements of "anatomically representative phantoms." This implies a technical or engineered ground truth rather than expert consensus, pathology, or outcomes data.

    8. The sample size for the training set

    The document does not provide information about a training set. This is a 510(k) summary focusing on a device that is substantially equivalent to a predicate, not necessarily a de novo submission for a novel AI algorithm requiring a specific training set size to establish performance. The software verification and validation would have involved testing against requirements, but not necessarily a "training set" in the machine learning sense.

    9. How the ground truth for the training set was established

    Since no training set is mentioned, the method for establishing its ground truth is not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233542

    Validate with FDA (Live)

    Date Cleared
    2024-03-14

    (132 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ortho Guidance Precision Knee Software:
    The Stryker guidance systems, with the Ortho Guidance Precision Knee Software, are intended as a planning and intraoperative quidance system to enable open computer-assisted surgerv. The systems can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified. The systems are indicated for conditions of the knee joint in which the use of computer-assisted surgery may be appropriate.

    Ortho Guidance Express Knee Software:
    The Stryker quidance systems, with the Ortho Guidance Express Knee Software, are intended as a planning and intraoperative guidance system to enable open computer-assisted surgery. The systems can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified. The systems are indicated for conditions of the knee joint in which the use of computer-assisted surgery may be appropriate.

    Ortho Guidance Versatile Hip Software:
    The Stryker guidance systems, with the Ortho Guidance Versatile Hip Software, are intended as a planning and intraoperative guidance system to enable open computer-assisted surgery.

    The systems can be used for intraoperative guidance where a reference to a rigid anatomical structure such as but not limited to the pelvis, or femur, can be identified.

    The systems are indicated for conditions of the hip joint in which the use of computer-assisted surgery may be appropriate.

    The system is indicated for the following surgical procedures:

    • Total Hip Arthroplasty (THA)

    • Precisely positioning instruments, implants, and bony tissue during orthopaedic hip surgery

    • Revisions

    Q Guidance System:
    The Q Guidance System is intended as an aid for precisely locating anatomical structures in open or percutaneous computer assisted surgery. When used with the Ortho Guidance Precision Knee, Express Knee, or Versatile Hip Software, the Q Guidance System is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate and where reference to a rigid anatomical structure such as the femur, tibia, or long bone can be identified.

    Device Description

    The system is intended as a planning and intraoperative guidance system to enable open computer-assisted surgery. It allows for the localization of surgical instruments, and visualization of their position relative to patient specific anatomical landmark information, assisting the surgeon in performing the intervention at a high level of precision. The system uses active optical tracking technology to display to the surgeon the intraoperative location of navigated surgical instruments relative to a computed anatomical model. The computed model is based on an intraoperative anatomy survey of the pelvis and/or leg.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Stryker Q Guidance System and its associated software, based on the provided document:


    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    System Accuracy (point)Mean accuracy of 2 mm point displacement within the registration zone.
    System Accuracy (angular)Mean accuracy of 2° angular axis displacement within the registration zone.
    Trueness for LED fiducial (mean Euclidean error)$\le 150 \mu m$
    Trueness for LED fiducial (95% percentile of Euclidean error)$< 0.35 mm$
    Trueness for reflective fiducial (mean Euclidean error)$\le 150 \mu m$
    Trueness for reflective fiducial (95% percentile of Euclidean error)$\le 0.35 mm$
    Trueness for single flashing LED fiducial$LME_{mean} \le 70 \mu m + 70 \mu m/m * L, L \le 1.5 m$
    Intended Use/User NeedsAll requirements were met, and no new issues of safety or effectiveness were raised through validation with intended users in cadaver labs or simulated use tests.
    SafetyEffectiveness of all risk controls determined in the device risk analysis was verified. No new questions of safety or effectiveness were raised.
    General Requirements and PerformanceAll components met their design specifications. No new questions of safety or effectiveness were raised.
    SoftwareSoftware verification and validation testing was conducted as required by IEC 62304 and FDA guidance on general principles of software validation, January 11, 2002. All requirements were met, and no new questions of safety or effectiveness were raised.
    BiocompatibilityNot applicable, as the subject devices are not intended to be patient contacting.
    Electrical SafetyConformance to IEC 60601-1 Ed. 3.2 (2005 + AMD1:2012 + AMD2:2020) was verified.
    Electromagnetic CompatibilityConformance to IEC 60601-1-2 Ed. 4.1 (2014 + AMD1:2020), CISPR 11 Group 1 Class A was verified.
    ShippingFunctionality of the devices after simulated shipping conditions was verified. No new questions of safety or effectiveness were raised.
    SterilizationNot applicable, as the subject devices are not intended to be sterilized.

    The study that proves the device meets the acceptance criteria is a non-clinical design verification and validation testing campaign.


    Study Details

    1. Sample size used for the test set and the data provenance:

      • The document does not specify a numerical sample size for the test set regarding the accuracy measurements. It mentions that testing was performed internally by the manufacturer.
      • Data Provenance: The data appears to be from internal testing conducted by Stryker Leibinger GmbH & Co. KG, located in Germany. The tests involved cadaver labs or simulated use, suggesting controlled laboratory environments rather than real-world patient data. The provenance is therefore prospective in a simulated environment.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not explicitly state the number or qualifications of experts used to establish ground truth for the numerical accuracy tests (e.g., fiducial trueness, system accuracy). These types of measurements are typically established using highly precise metrological equipment as the ground truth.
      • For "Intended Use/User Needs" testing, "intended users" were involved in cadaver labs or simulated use tests. However, the exact number and qualifications (e.g., specific surgical experience) of these "intended users" (e.g., surgeons) are not specified.
    3. Adjudication method for the test set:

      • The document does not describe a formal adjudication method for the test set results. The accuracy metrics are quantitative measurements against an established standard. The user needs validation was presumably based on observation and feedback from the intended users.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done or reported. The device is a surgical guidance system, not an AI diagnostic tool that assists human readers in interpreting medical images. The testing described focuses on the system's accuracy in tracking and guiding surgical instruments.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, standalone performance was assessed. The accuracy measurements for fiducial trueness and system point/angular accuracy represent the direct, standalone performance of the tracking and guidance algorithms. The "Intended Use/User Needs" validation also assessed the integration and utility with a human user.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the accuracy metrics (e.g., trueness for fiducials, system accuracy), the ground truth is implicitly defined by metrological standards and highly precise measurement systems, which assess the physical displacement and angulation of tracked objects.
      • For "Intended Use/User Needs," the ground truth involves user feedback and achievement of functional requirements observed in cadaver labs or simulated environments.
    7. The sample size for the training set:

      • The document does not specify a numerical sample size for the training set. This is a navigation system primarily based on optical tracking and software for workflow and display, rather than a machine learning model that requires a discrete "training set" of patient data in the conventional sense. The "training" of such a system would involve extensive engineering, calibration, and iterative development based on design parameters and physical principles.
    8. How the ground truth for the training set was established:

      • As this is not a machine learning-based device that would typically have a "training set" with explicit ground truth labels, the concept of ground truth for a training set in the AI sense does not directly apply here. Instead, during development and "training" of the underlying algorithms and hardware, ground truth is established through:
        • Physical calibration standards: For optical tracking, this involves precise optical benches and measurement devices to calibrate camera parameters, fiducial patterns, and spatial relationships.
        • Engineering specifications and simulations: Ground truth for software logic and mechanical tolerances is derived from engineering design requirements and validated through simulations.
        • Iterative testing and refinement: Early prototypes and software versions undergo extensive testing against known inputs and expected outputs to refine performance.
    Ask a Question

    Ask a specific question about this device

    K Number
    K231599

    Validate with FDA (Live)

    Date Cleared
    2023-08-24

    (84 days)

    Product Code
    Regulation Number
    872.4880
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    -Stryker Universal SMARTLock Hybrid MMF System: The Stryker Universal SMARTLock Hybrid MMF System is indicated for the treatment of mandibular and maxillary fractures in adults and adolescents (age 12 and higher) in whom permanent teeth have erupted.
    -Universal CMF System: Craniomaxillofacial Implants The Universal CMF System is Craniomaxillofacial (CMF) plate and screw system intended for osteotomy, stabilization, and rigid fixation of CMF fractures and reconstruction. Mandible Implants The Universal CMF System (mandible modules) is a mandibular plate and screw system intended for stabilization and rigid fixation of mandibular fractures and mandibular reconstruction.
    -Stryker MMF screw: The Stryker MMF Screw is intended for use as a bone screw in the temporary maxillomandibular fixation to provide indirect stabilization of fractures of the maxilla, mandible or both, where there is sufficient occlusion.

    Device Description

    The Subject Device screws are screw versions modified with the Stryker AXS Screw technology, where the abbreviation "AXS" stands for axial stability. This submission is adding the AXS Screw technology to the Subject Device. The Subject Device screws have modified geometry and features which will be implemented as AXS Screws in each of the existing Predicate/Reference Device systems. This includes screws with a length range of 4-20 mm and a diameter range from 2.0-2.7 mm. The materials (titanium and stainless steel) are the same as for the corresponding Predicate/Reference Devices. There are no other changes to the Predicate/Reference Device systems, only the implementation of the AXS Screw technology to each existing product line. The Subject Device screws will have the same clinical application, treat the same craniomaxillofacial anatomical locations, and are for the same patient population as the corresponding Predicate/Reference Devices.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to Stryker MP, Mandible, HMMF and MMF AXS Screws, which are intraosseous fixation screws. This submission is for a medical device (screws), not an AI/software as a medical device (SaMD). Therefore, the questions related to AI device performance, sample sizes for test/training sets, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance are not applicable to this document.

    The clearance is based on demonstrating substantial equivalence to predicate devices, primarily through engineering performance bench testing.

    Here's the relevant information that can be extracted from the document:

    1. A table of acceptance criteria and the reported device performance

    The document states that "The Subject Device met all pre-defined acceptance criteria, and the results of the V&V tests support the substantial equivalence of the Subject Device to the Predicate/Reference Devices." However, the specific numerical acceptance criteria for each test are not detailed in this summary document. The document lists the types of performance bench tests conducted:

    Test NameReported Device Performance (as stated in document summary)
    Pull-out Test per ASTM F543Met all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.
    Full Insertion and Removal Test per ASTM F543Met all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.
    Shaft Shear Test per ASTM F543Met all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.
    Head Shear TestMet all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.
    Locking Cantilever Bending TestMet all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.
    Locking Torque to Failure TestMet all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.
    Self-Retention TestMet all pre-defined acceptance criteria, supporting substantial equivalence to Predicate/Reference Devices.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the summary document. The document refers to "Verification and Validation (V&V) testing" and "test reports" being included within the full submission, but the summary itself does not detail specific sample sizes for these bench tests or the provenance of any data beyond the general statement that testing was performed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not Applicable. This is a medical device (screws) and does not involve AI or software requiring expert review for ground truth establishment. The "acceptance criteria" are engineering performance metrics.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not Applicable. This is a medical device (screws) and does not involve subjective human interpretation or adjudication for its performance evaluation in the way an AI diagnostic tool would. Performance is based on objective engineering tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable. This is a medical device (screws) and does not involve AI or human "readers."

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not Applicable. This is a medical device (screws) and does not involve an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For mechanical performance tests, the "ground truth" or reference standard would be the physical properties and behavior of the devices as measured by standardized testing methods (e.g., ASTM F543) and compared against pre-defined engineering specifications and the performance of the predicate device.

    8. The sample size for the training set

    Not Applicable. This is a medical device (screws) and does not involve a training set as would be used for AI/machine learning.

    9. How the ground truth for the training set was established

    Not Applicable. This is a medical device (screws) and does not involve a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231208

    Validate with FDA (Live)

    Date Cleared
    2023-08-14

    (109 days)

    Product Code
    Regulation Number
    882.5360
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Delta Resorbable Fixation System is intended for use in the fixation of bones of the cranial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients.

    The self-tapping screws and low profile emergency screws are intended for use in the fixation of bones of the cranial skeleton affected by trauma or for reconstruction, and can be used in pediatric patients older than 29 days and up to two (2) years of age (infants).

    Device Description

    The Stryker Resorbable Fixation System (also referred to as Delta System or Stryker Delta Resorbable Fixation System; marketed as DualStart) is a cranio-maxillofacial plating system intended for use in the fixation of bones of the craniofacial and midfacial skeleton affected by trauma or for reconstruction. The reference device can be used in both adult and pediatric patients but is not intended for use in the mandible and/or full load bearing procedures. It consists of resorbable bone fixation plates, meshes, and screws made of a copolymer of poly lactide and poly glycolide.

    Through this submission there is no change to the existing articles within the Delta System. The scope of this submission covers the addition of screws to the Delta System, which are shown as a modification of screws of the reference device to create the addition of self-tapping screws (STS) and low profile emergency screws (LPES). The subject STS and LPES, only, have a limited patient population of pediatric patients older than 29 days and up to two (2) years of age (infants).

    AI/ML Overview

    The provided text describes the acceptance criteria and supporting studies for the Stryker Resorbable Fixation System (K231208).

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Test)Reported Device Performance (Result)Standard
    Verification test (Initial)Insertion testPassedASTM F2502
    Verification test (After real-time degradation)Pull-out testPassedASTM F2502
    Verification test (After real-time degradation)Inherent viscosityPassedASTM F2502
    Usability validationTask performance (evaluated for use error)PassedN/A
    Usability validationKnowledge tasksPassedN/A
    Usability validationRating questions after task performancePassedN/A
    BioburdenBioburden tests (no new worst-case introduced with modified screw geometries)Passing resultN/A

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size for the test set for the performance bench testing (e.g., number of screws tested for insertion or pull-out). It generically states "Verification and validation testing protocols were constructed to ensure testing captured this specific limited patient population."

    • Data Provenance: The studies were conducted by Stryker Leibinger GmbH & Co. KG and are presumed to be internal company studies. The country of origin for the data is not specified, but the company address is in Freiburg, Germany. The studies are prospective in nature, as they are part of a premarket notification for a modified device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    For the usability validation, the document mentions "user validation testing" and "human factors and usability engineering," which would typically involve end-users (e.g., surgeons, medical staff). However, the number of experts used and their specific qualifications are not provided in this document.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method for the test sets (e.g., 2+1, 3+1, none). The "Passed" results imply direct measurement against established criteria.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was done. The device in question is a physical fixation system (screws), not a diagnostic imaging AI, so an MRMC study comparing human readers with and without AI assistance is not applicable. The studies focus on the physical and functional performance of the device and its usability.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Not Applicable. This device is a physical medical implant, not an algorithm or AI system. Therefore, standalone algorithm performance is not a relevant concept here.

    7. Type of Ground Truth Used

    For the performance bench testing:

    • The ground truth is established by objective engineering measurements against the performance requirements defined by the ASTM F2502 standard and internal specifications (e.g., successful insertion, specified pull-out strength, acceptable inherent viscosity).

    For the usability validation:

    • The ground truth is established by user validation through direct observation of task performance (evaluating for use error), assessment of knowledge (knowledge tasks), and subjective feedback (rating questions). This implies that the 'ground truth' for usability is defined by the absence of use errors, successful completion of knowledge tasks, and positive user ratings.

    8. Sample Size for the Training Set

    • Not applicable. This device is a physical medical device; there is no "training set" in the context of machine learning or AI models. The design and manufacturing processes are validated through traditional engineering methods and testing.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set for an AI/algorithm, there is no ground truth established for a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230733

    Validate with FDA (Live)

    Date Cleared
    2023-08-05

    (142 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dental
    The self-tapping screws and low profile emergency screws are intended for use in the fixation of bones of the maxillofacial and midfacial skeleton affected by trauma or for reconstruction, and can be used in pediatric patients older than 29 days and up to two (2) years of age (infants). The self-tapping screws and low profile emergency screws are designed to be compatible with the components (plates, meshes) of the Delta Resorbable Fixation System K213777.

    Device Description

    The subject devices, the Delta System self-tapping screw (STS) and low profile emergency screw (LPES), are intended for use in the fixation of bones of the maxillofacial and midfacial skeleton, affected by trauma or for reconstruction. The subject device can be used in pediatric patients older than 29 days and up to two (2) years of age (infants), but is not intended for use in the mandible and/or full load bearing procedures.

    The scope of this submission covers the addition of the subject device screws to the previously cleared Delta System. Through this submission there is no change to the existing articles within the Delta System. The subject device screws are designed to be compatible with the existing components of the Delta System (plates, meshes), which have been previously cleared through K213777. Compared to the original Delta System, the subject STS and LPES have a limited patient population of pediatric patients older than 29 days and up to two (2) years of age (infants).

    AI/ML Overview

    The document provided is a 510(k) summary for a medical device called the "Stryker Resorbable Fixation System". This summary outlines the device, its intended use, and comparative information to a predicate device to demonstrate substantial equivalence to a legally marketed device.

    However, the provided text does not contain information about an AI/ML device or a study to prove that a device meets acceptance criteria related to AI/ML performance. The document focuses on regulatory approval for an orthopaedic fixation system, specifically resorbable screws, based on traditional engineering performance testing (e.g., mechanical tests, material properties) and comparison to a predicate device.

    Therefore, I cannot provide the requested information regarding AI/ML acceptance criteria, study details, sample sizes, expert involvement, or MRMC studies, as these concepts are not applicable to the content of the provided document.

    The document discusses "acceptance criteria" in the context of:

    • Regulatory Acceptance: The FDA's determination of substantial equivalence.
    • Performance Testing: Bench test "acceptance criteria" are implied by the "Passed" results for tests like Insertion, Shear, and Pull-out, and "Passed" for Inherent Viscosity. These are mechanical and material property tests, not AI/ML performance metrics.
    • User Validation Testing: A validation study was done to confirm usability, comprehension of Instructions for Use, and that user needs are met, concluding a "passing result." This relates to human factors and usability, not AI/ML performance.

    In summary, the provided text does not contain the information needed to answer the prompt's specific questions about AI/ML device acceptance criteria and proving device performance related to AI/ML.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 4