K Number
K243491

Validate with FDA (Live)

Date Cleared
2024-12-11

(29 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SternalPlate System is indicated for use in stabilization and fixation of fractures of the anterior chest wall including sternal fixation following sternotomy, sternal fracture(s), and sternal reconstructive surgical procedures to promote bony fusion.

Device Description

This Special 510(k) is submitted to evaluate the modifications to the Predicate Device plates resulting in the Subject Device plates. There have been no changes to the previously cleared plates and screws cleared in K183172, this submission is only for the addition of the four new Subject Device plates. The four Subject Device plates include T-Plates and Transverse Plates. Both Subject Device variants, the T-Plates and Transverse Plates, are geometrically derived from the following Predicate Device plates: the Ladder Plate Narrow (Reff/ 4740006) and the Straight 24-hole Plate (Ref# 4740012). The Subject Device plates therefore have the same features, the same tolerances, and are made from the same material as the Predicate Device plates. Furthermore, the Subject Device plates are manufactured with the same manufacturing process as the Predicate Device plates.

AI/ML Overview

The provided FDA 510(k) summary (K243491) for the Stryker SternalPlate Expansion device focuses on the mechanical performance of the device and its biocompatibility. The document does not contain information related to software, AI, or human-in-the-loop performance. Therefore, many of the requested criteria cannot be extracted from this document.

Here's the information that can be extracted or deduced from the provided text:

1. A table of acceptance criteria and the reported device performance

The document states that the "Subject Device plates met all acceptance criteria as compared to the Additional Predicate Devices." However, the specific quantitative acceptance criteria (e.g., minimum stiffness, maximum deflection) and the detailed reported performance values are not explicitly stated in the provided 510(k) summary. The summary only mentions "appropriate strength and stiffness."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document refers to "Performance Bench Testing" but does not specify the sample size for the mechanical tests. The tests compare the Subject Device to "Additional Predicate Devices," implying a comparative mechanical study rather than a clinical data-driven study. Data provenance (country of origin, retrospective/prospective) is not applicable as it's a bench test, not clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This question is not applicable as the document describes mechanical bench testing of an orthopedic implant, not a diagnostic device requiring expert interpretation of images or data to establish a ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This question is not applicable as the document describes mechanical bench testing. Adjudication methods are typically relevant for clinical studies or studies involving human readers/interpreters.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study. This document is for a physical orthopedic implant, not an AI-powered diagnostic tool. Therefore, the effect size of human readers with/without AI assistance is not relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable as the device is a physical orthopedic implant, not a standalone algorithm or software.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the performance bench testing, the "ground truth" would be the engineering specifications and established mechanical testing standards against which the device's strength and stiffness are evaluated. The document states "The Subject Device plates met all acceptance criteria as compared to the Additional Predicate Devices," implying a comparison against established performance benchmarks of legally marketed devices.

8. The sample size for the training set

This question is not applicable as the device is a physical orthopedic implant. There is no concept of a "training set" in the context of mechanical bench testing for this type of device.

9. How the ground truth for the training set was established

This question is not applicable for the reasons stated above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 11, 2024

Stryker Leibinger GmbH & Co. KG Amelia Kesti Staff Regulatory Affairs Specialist Boetzinger Strasse 41 Freiburg, D-79111 Germany

Re: K243491

Trade/Device Name: SternalPlate Expansion Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: November 8, 2024 Received: November 12, 2024

Dear Amelia Kesti:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CHRISTOPHER FERREIRA -S

Christopher Ferreira, MS Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K243491

Device Name

SternalPlate Expansion

Indications for Use (Describe)

The SternalPlate System is indicated for use in stabilization of fractures of the anterior chest wall including sternal fixation following sternal fracture(s), and sternal reconstructive surgical procedures to promote bony fusion.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.

  • SUBMITTER [§807.92(a)(1)] I.
510(k) Owner:Stryker Leibinger GmbH & Co. KG
Boetzinger Strasse 41
D-79111 Freiburg, Germany
  • Submitter/ Contact Amelia Kesti Person: Staff Regulatory Affairs Specialist Stryker Craniomaxillofacial (CMF) 1941 Stryker Way Portage, MI 49002
    Phone: 269-330-5919

Date prepared: 11/08/2024

II. DEVICE [§807.92(a)(2)]

Trade Name:SternalPlate Expansion
Abbreviated Name:SternalPlate Expansion
Common or Usual Name:Plate, Fixation, Bone
Device:SternalPlate Expansion
Classification Name &Regulation Description:Plate, Fixation, Bone; per 21 CFR §888.3030
Regulation MedicalSpecialty & Review Panel:Office of Product Evaluation and Quality, Office ofOrthopedic Devices, Division of Restorative, Repair andTrauma, Devices, DHT6C
Product Code:HRS
Regulatory Device Class:Class II
*Note the company Stryker or legacy name Stryker Leibinger precedes the product/trade name andpredicate device in some documentation

{5}------------------------------------------------

III. PREDICATE DEVICE [§807.92(a)(3)]

  • A. Predicate Devices: The predicate devices for this Special 510(k) are:
      1. Primary Predicate: Stryker SternalPlate System K183172
      • i. Submission Branch of Predicate Device- Division of Orthopedic Devices, Office of Device Evaluation, Center for Devices and Radiological Health
      1. Additional Predicate: DePuy Synthes Sternal Fixation System K190963
      • Branch of Predicate Device- Division of i. Submission Restorative, Repair and Trauma Devices, Office of Orthopedic Devices, office of Product Evaluation and Quality, Center for Devices and Radiological Health
      1. Additional Predicate: KLS Martin Thoracic Plating System -K153482
      • i. Submission Branch of Predicate Device- Division of Orthopedic Devices, Office of Device Evaluation, Center for Devices and Radiological Health

IV. DEVICE DESCRIPTION [§807.92(a)(4)]

  • A. Submission Branch of Subject Device: Office of Orthopedic Devices, Office of Product Evaluation and Quality, Division of Stereotaxic, Trauma, and Restorative Devices, DHT6C
  • B. Subject Device: SternalPlate Expansion

This Special 510(k) is submitted to evaluate the modifications to the Predicate Device plates resulting in the Subject Device plates. There have been no changes to the previously cleared plates and screws cleared in K183172, this submission is only for the addition of the four new Subject Device plates.

INDICATIONS FOR USE [§807.92(a)(5)] V.

The Predicate and Subject devices are both intended for use in the anterior chest wall. The anterior chest wall includes both the sternum and the anterior ribs. The Subject Device does not change the existing indications for use for the Stryker SternalPlate System. The new plates fall within the scope of the existing indications in the anterior chest wall without alteration.

{6}------------------------------------------------

TABLE 5-1: COMPARISON OF INDICATIONS FOR USE

AspectSubject DevicePrimary Predicate Device (K183172)
IntendedUse/Indications forUseThe SternalPlate System is indicated for use in stabilizationand fixation of fractures of the anterior chest wall includingsternal fixation following sternotomy, sternal fracture(s), andsternal reconstructive surgical procedures to promote bonyfusion.The SternalPlate System is indicated for use in stabilization andfixation of fractures of the anterior chest wall including sternalfixation following sternotomy, sternal fracture(s), and sternalreconstructive surgical procedures to promote bony fusion.

Please note that this 510(k) lists several additional Predicate devices, however, these additional Predicate devices were used for testing and not verall substantial equivalence evaluation.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE [§807.92(a)(6)]

The Subject Device is compared to its Predicate Devices for substantial equivalence of technological characteristics based on the following criteria:

  • A. Principles of Operation
  • B. Technological Characteristics

A. Principles of Operation / Operating Principle

The basic operating principle of the Subject Device remains the same as the Predicate Device. The basic operating principle of the Subject Device, as well as the Predicate Device is fixation and stabilization of the anterior chest wall. The method of site preparation and fixation is the same for both the Subject and Predicate Device. The fundamental scientific technology of the Subject Device has not changed. This is because the principle of operation, the mechanism of action, the intended use, and the materials of construction have not changed as a result of the device modifications.

B. Technological Characteristics

There have been no changes to the previously cleared plates and screws cleared in K183172, this Special 510(k) is only for the addition of the four new Subject Device plates. The four Subject Device plates include T-Plates and Transverse Plates. Both Subject Device variants, the T-Plates and Transverse Plates, are geometrically derived from the following Predicate Device plates: the Ladder Plate Narrow (Reff/ 4740006) and the Straight 24-hole Plate (Ref# 4740012).

The Subject Device plates therefore have the same features, the same tolerances, and are made from the same material as the Predicate Device plates. Furthermore, the Subject

{7}------------------------------------------------

Device plates are manufactured with the same manufacturing process as the Predicate Device plates.

VI. PERFORMANCE DATA [§807.92(b)(7)]

Appropriate strength and stiffness of the new plates are confirmed by the Verification and Validation testing, and no new questions of safety or effectiveness arise as a result of the testing.

The subject devices have been evaluated for Biocompatibility per the requirements of ISO 10993-1:2018 and in alignment with corresponding FDA guidance. This evaluation has concluded that no new or increased biocompatibility risks are associated with the subject device.

Sterility testing was not required as a basis for substantial equivalence. There is no change in the subject device material, manufacturing process, duration or location of contact, or reprocessing methods.

Performance Bench Testing

As the Subject Device plates, unlike the original Predicate Device plates, can be fixated laterally to the sternum, additional Verification and Validation data was needed for the Subject Device. This submission uses well established test methods used in prior submissions including K183172 previously reviewed by the FDA. These tests are summarized within the Special 510(k).

Plate stiffness and strength were tested in comparison to the Additional Predicate Devices DePuy Synthes Sternal Fixation System (K190963) and KLS Martin Thoracic Plating System (K153482). The Subject Device plates met all acceptance criteria as compared to the Additional Predicate Devices. The results of this testing show that the Subject Device plates do not raise new questions in terms of safety and effectiveness when compared to the Predicate Device plates.

Animal Testing

Animal testing was not required as a basis for substantial equivalence.

Clinical Testing [§807.92(b)(2)]

Clinical testing was not required as a basis for substantial equivalence.

VII. CONCLUSIONS [§807.92(b)(3)]

In summary, the Subject Device is substantially equivalent to its Predicate Device. The Subject Device's features, tolerances, material, mechanism of action, intended use, and

{8}------------------------------------------------

manufacturing process are the same as the Predicate Device. The Subject Device's intended use is unchanged compared to the Predicate Device. Appropriate strength and stiffness of the new plates are confirmed by the Verification and Validation testing, and no new questions of safety or effectiveness arise as a result of the testing. There are no new questions of safety or effectiveness regarding the modifications to the Predicate Device plates resulting in the four Subject Device plates. According to the comparison based on the requirements of 21 CFR 807.87 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence to the predicate devices.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.