Search Filters

Search Results

Found 15 results

510(k) Data Aggregation

    K Number
    K070645

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-05-17

    (70 days)

    Product Code
    Regulation Number
    862.3350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Seradyn QMS® Topiramate assay is intended for the quantitative determination of topiramate in human serum or plasma on automated clinical chemistry analyzers.

    The results obtained are used in the diagnosis and treatment of topiramate overdose and in monitoring levels of topiramate to help ensure appropriate therapy.

    Device Description

    The Seradyn QMS® Topiramate assay is a homogeneous particle-enhanced turbidimetric immunoassay. The assay is based on competition between drug in the sample and drug coated onto a micronariticle for antibody binding sites of the topiramate antibody reagent. The topiramate-coated micropariticle peagent is rapidly agglutinated in the presence of the anti-topiramate antibody reagent and in the abserved of any competing drug in the sample. The rate of absorbance change is measured photometrically. When a smale containing topiramate is added, the agglutination reaction is partially inhibited, slowing down the rate of absorbance change. A concentration-dependent classic agglutination inhibition curve can be obtained with maximum rate of agglutination at the lowest topiramate concentration and the lowest agglutination rate at the highest topiramate concentration.

    The assay consists of reagents R1: anti-topiramate polyclonal antibody and R2: topiramate-ooated microparticles. A six-level set of Seradyn QMS® Topiramate Cali three-level set of Seradyn QMS® Topiramate Controls is used for quality control of the assay.

    AI/ML Overview

    Here's an analysis of the Seradyn QMS® Topiramate assay based on the provided 510(k) summary, structured to address your specific points:

    Seradyn QMS® Topiramate Assay Study Analysis

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance MetricAcceptance CriteriaReported Device Performance
    Accuracy (Recovery)100 ± 10%Mean Percent Recovery: 104.6%. Individual recoveries ranged from 101.5% to 109.9%. All concentrations (3.20 to 32.00 µg/mL) met the acceptance criteria.
    LinearityPercent Difference: ±10%All measured concentrations (1.5 to 35 µg/mL) showed a percent difference from the predicted result well within ±10% (ranging from -0.25% to 0.10%).
    Sensitivity (LOQ)≤20% CV; recovery ± 15%1.5 µg/mL (Claimed in package insert). Specific data for LOQ meeting these criteria is not directly presented in the table, but the claim is based on observed acceptable inter-assay precision and recovery.
    Assay RangeBased on Accuracy, Linearity, and Sensitivity (LOQ)1.5 to 32.0 µg/mL (Claimed reportable range).
    Precision (Total CV)< 10%Control 1 (2.94 µg/mL): 4.22% CV < 10%Control 2 (10.14 µg/mL): 3.37% CV < 10%Control 3 (25.69 µg/mL): 4.44% CV < 10% (All controls met the acceptance criteria).
    SpecificityUnlikely to be affected by topiramate metabolites"Metabolites of topiramate are found primarily in urine... They are not however seen at clinically significant levels in plasma or serum. The QMS topiramate assay serum and plasma results are unlikely to be affected by metabolism of topiramate drug." (Qualitative claim, no specific numerical criterion).
    Interferences (Endogenous & HAMA)≤10% error in detecting topiramateAll tested substances (Albumin, Bilirubin, Cholesterol, Gamma-Globulin, HAMA-1, HAMA-2, Hemoglobin, Heparin, Rheumatoid Factor, Triglycerides, Uric Acid) resulted in ≤10% error.
    Interferences (Co-Administered Drugs)≤10% error in detecting topiramateMost tested drugs (list provided) resulted in ≤10% error. Specific drugs identified as potentially cross-reacting with >10% error: Ibuprofen, Phenytoin, Tiagabine. This implies the device does not meet the criteria for these specific drugs.
    Interferences (Anticoagulants)No significant difference in recovery between serum and plasma samples"No significant difference between the recovery of topiramate in serum or plasma. The collection tubes evaluated show no adverse effects on the recovery of topiramate." (Qualitative claim).
    Calibration Curve StabilityN/A (implicit: stable for claimed period)Supported for a period of 27 days.
    Reagent On-Board StabilityN/A (implicit: stable for claimed period)Supported for 60 days.
    Method ComparisonExcellent correlation with predicateN = 148, Slope = 0.962, y-intercept = 0.228, R² = 0.986. The report states: "Results show excellent correlation between the two assays." (Qualitative interpretation of quantitative results).

    2. Sample Size for the Test Set and Data Provenance:

    • Accuracy: 12 concentrations for recovery study, each analyzed in triplicate (total of 36 measurements).
    • Linearity: 9 concentrations, number of replicates not specified.
    • Sensitivity (LOQ): Not specified directly, but implies multiple measurements around the LOQ value to determine CV and recovery.
    • Precision: 3 controls, N=80 for each (total 240 measurements for precision components).
    • Method Comparison: N = 148 patient samples.
    • Interference (Endogenous & HAMA): Not explicitly stated, but for each interferent, samples with two known topiramate levels (approx. 5 and 20 µg/mL) were assayed.
    • Interference (Co-Administered Drugs): Not explicitly stated, but for each compound, normal human serum with two known topiramate levels (approx. 5 and 20 µg/mL) was assayed.
    • Interference (Anticoagulants): Not explicitly stated, but implied comparison of serum and plasma.

    Data Provenance: The document does not specify the country of origin of the data. The studies are described as clinical testing, but it's common for these types of in vitro diagnostic studies to use banked or commercially sourced human serum/plasma samples, often without explicit geographical tags in 510(k) summaries. All studies appear to be prospective in the sense that they were designed experiments to evaluate the new device's performance against predefined criteria or a predicate device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This device is an in vitro diagnostic (IVD) assay for quantitative determination of a drug concentration. The "ground truth" here is the actual concentration of topiramate in the samples. This is typically established through:

    • Reference materials: For linearity and accuracy by recovery, known concentrations are prepared by precise dilution of a high calibrator.
    • Reference method/predicate device: For method comparison, the predicate Innofluor® Topiramate assay serves as the reference for comparison of patient sample results.

    Therefore, no human experts were used to establish the "ground truth" in the way they would be for image analysis or disease diagnosis. The "ground truth" is defined by laboratory standards, precise dilutions, and the established performance of a reference or predicate method.

    4. Adjudication Method for the Test Set:

    Not applicable, as the ground truth is quantitative (actual concentration) and not based on human interpretation requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging where multiple human readers interpret cases, and AI assistance might impact their performance. For a quantitative IVD assay, the performance is measured directly by analytical metrics (accuracy, precision, linearity, etc.) and comparison to a predicate device or reference method, not by human reader performance.

    6. Standalone Performance:

    Yes, the studies described (Accuracy, Linearity, Sensitivity, Precision, Specificity, Interferences, Stability) evaluate the algorithm's entire workflow (reagent interaction, measurement, and result calculation) standalone, without human-in-the-loop performance influencing the primary measurements. The assay quantifies topiramate concentration directly.

    7. Type of Ground Truth Used:

    • Known concentrations: For Accuracy by Recovery, Linearity, Sensitivity studies. These are prepared laboratory standards.
    • Predicate device results: For Method Comparison, the results from the Seradyn Innofluor® Topiramate assay (K970510) on patient samples serve as the comparison point.
    • Laboratory-spiked samples: For interference studies, known amounts of interferent and topiramate are added to serum samples.

    8. Sample Size for the Training Set:

    The document does not provide a sample size for a "training set." This assay is a homogeneous particle-enhanced turbidimetric immunoassay (PETIA), which is a chemical reaction-based method, not a machine learning or AI algorithm that typically requires a large "training set" of data in the common sense. The "development" or "optimization" of the assay would involve various experiments, but these are not usually referred to as a "training set" in the context of conventional IVD development. The calibration of the device uses a six-level set of Seradyn QMS® Topiramate Calibrators, but this is for operational calibration, not model training.

    9. How the Ground Truth for the Training Set Was Established:

    As noted above, the concept of a "training set" with established ground truth as in AI/ML is not directly applicable to this type of chemical immunoassay. The "ground truth" for calibrators and controls used in assay development and validation would be established through highly accurate reference methods, gravimetric/volumetric preparation, and traceability to established standards for the analyte (topiramate).

    Ask a Question

    Ask a specific question about this device

    K Number
    K062966

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-01-12

    (105 days)

    Product Code
    Regulation Number
    862.3350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QMS® Lamotrigine assay is intended for the quantitative determination of lamotrigine in human serum or plasma on automated clinical chemistry analyzers.

    Lamotrigine concentrations can be used as an aid in management of patients treated with lamotrigine.

    The QMS® Lamotrigine Calibrator set is intended for use in calibration of the QMS Lamotrigine assay.

    The QMS® Lamotrigine Control set is intended for use in quality control of the QMS Lamotrigine assay.

    Device Description

    The QMS Lamotrigine assay system is a homogeneous assay utilizing particle agglutination technology and is based on the competitive binding principle.

    The assay consists of reagents R1: anti-Lamotrigine sheep polyclonal antibody and R2: Lamotriginecoated microparticles. A six-level set of QMS Lamotrigine Calibrators (A through F) is used to calibrate the assay. A three-level set of QMS Lamotrigine Controls (1 thro the assay.

    AI/ML Overview

    The provided document describes the K062966 QMS® Lamotrigine assay, a homogeneous particle-enhanced turbidimetric immunoassay for the quantitative determination of lamotrigine in human serum or plasma.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" with specific pass/fail thresholds for each performance characteristic. Instead, it describes general methods used for testing and states that performance testing verified the device functions as intended and satisfied design specifications. However, we can infer some criteria and the reported performance from the "SUMMARY OF CLINICAL TESTING" section.

    Performance CharacteristicAcceptance Criteria (Inferred)Reported Device Performance
    Accuracy and LinearityDemonstrated linearity and accuracy across the reportable range.Determined by a study based on NCCLS guideline EP6.
    Sensitivity (LOQ)Establish the Limit of Quantitation.Functional Sensitivity (LOQ) determined to be 2.0 µg/mL.
    Assay RangeDefine the reportable range of the assay.Reportable range: 2.0 to 40.0 µg/mL.
    Method ComparisonShow correlation with a comparative method.Correlation studies conducted using NCCLS Guideline EP9.
    PrecisionDemonstrate acceptable precision.Performed using NCCLS guideline EP5.
    SpecificityMinimal or no significant cross-reactivity with metabolites and other drugs.N-2 oxide shows cross-reactivity but is in very minor concentrations. No significant cross-reactivity with other metabolites.
    InterferencesMinimal or no significant interference from common drugs.Of 26 drugs tested, none showed cross-reactivity.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes for the test sets used in each study (Accuracy and Linearity, Sensitivity, Method Comparison, Precision, Specificity, Interferences).

    • Data Provenance: The document does not explicitly state the country of origin or if the data was retrospective or prospective. However, given it's a clinical chemistry assay for human serum/plasma, it's highly likely the samples were human biological specimens.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. For an in vitro diagnostic (IVD) assay like this, "ground truth" is typically established by reference methods or validated techniques, not necessarily by "experts" in the same way it would be for imaging diagnostics requiring interpretation.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    This information is not applicable and therefore not provided. Adjudication methods are typically relevant for studies involving human interpretation of results (e.g., in imaging or pathology where multiple readers might disagree). For a quantitative chemical assay, the "truth" is determined by the output of a reference instrument or method.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    A Multi Reader Multi Case (MRMC) comparative effectiveness study was not performed or described. This type of study is relevant for evaluating the performance of human readers, often aided by AI, in complex diagnostic tasks (e.g., radiology). This device is a quantitative immunoassay, not an AI-assisted diagnostic imaging tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This device is a standalone assay kit that performs a quantitative measurement on automated clinical chemistry analyzers. The "algorithm" is the biochemical reaction and measurement process itself, not a separate computational algorithm that assists a human. Therefore, the performance described is the standalone performance of the assay system without human interpretation as a primary component of the diagnostic result. The results (lamotrigine concentrations) are then used by clinicians for patient management.

    7. Type of Ground Truth Used

    The ground truth for this type of quantitative assay would typically be established by:

    • Reference Methods: Highly accurate and precise analytical methods (e.g., Gas Chromatography-Mass Spectrometry (GC-MS) or Liquid Chromatography-Mass Spectrometry (LC-MS)) for determining the true concentration of lamotrigine in samples.
    • Spiked Samples: Known concentrations of lamotrigine added to a matrix (e.g., human serum or plasma that is negative for lamotrigine) to assess accuracy and linearity.
    • Patient Samples: Used for method comparison against a legally marketed predicate device or a well-established reference method.

    The document implicitly refers to these by mentioning "Accuracy and linearity were determined..." and "Correlation studies were conducted using patient samples..." indicating that the "ground truth" for those studies would have been the results from a reference method or the assigned values from spiked samples.

    8. Sample Size for the Training Set

    The document does not mention a training set because this is a traditional immunoassay, not a machine learning or AI-based device that requires "training." The "training" of such a system involves developing the reagents and optimizing the assay conditions.

    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" in the context of machine learning, this question is not applicable. The development of the assay's reagents and protocols is based on extensive biochemical research and optimization studies rather than "ground truth" established for a training dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062204

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-09-22

    (52 days)

    Product Code
    Regulation Number
    862.1205
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ARCHITECT® Cortisol is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of cortisol in human serum, plasma or urine on the ARCHITECT i System. The ARCHITECT Cortisol assay is intended for use as an aid in the diagnosis and treatment of adrenal disorders.

    The ARCHITECT Cortisol Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of cortisol in human serum, plasma or urine.

    Device Description

    The ARCHITECT Cortisol assay is a delayed one-step immunoassay for the quantitative determination of cortisol in human serum, plasma or urine using CMIA technology with flexible assay protocols, referred to as Chemiflex®.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the ARCHITECT Cortisol assay:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test CategoryAcceptance CriteriaReported Device Performance
    LinearityTarget +/- 20% Deviation (at 0% dilution for both serum pools)- 65 ug/dL Serum Pool: %DLP ranged from -8% to 4% for diluted samples. At the 0% dilution, the difference was-1.6 ug/dL, compared to the target value of N/A, which is acceptable relative to the ±20% deviation criteria. - 8 ug/dL Serum Pool: %DLP ranged from -12% to 3% for diluted samples. At the 0% dilution, the difference was-0.2 ug/dL, which is acceptable relative to the ±20% deviation criteria.
    Accuracy (Recovery)Serum: Target Recovery 100 ± 15% Urine: Target Recovery 100 ± 20%- Serum: % Recovery ranged from 86.1% to 98.5%. All values are within the acceptance criteria of 100 ± 15% (i.e., 85% to 115%). - Urine: % Recovery ranged from 84.6% to 100.9%. All values are within the acceptance criteria of 100 ± 20% (i.e., 80% to 120%).
    Sensitivity (LoD)Assay claim of LoD = 0.8 ug/dL is supported by data.- ARCHITECT i2000 LoD = 0.401 ug/dL - ARCHITECT i2000SR LoD = 0.255 ug/dL - Functional sensitivity (20% CV): Serum = 0.8 ug/dL, Urine = 1 ug/dL. All reported LoD values are less than or equal to the claimed 0.8 ug/dL for serum, and urine is close at 1 ug/dL, thus supporting the claim.
    Method ComparisonThe results of the Method comparison study met the design goals and acceptance criteria. (Specific numerical criteria not provided in the summary)The summary states that "The results of the Method comparison study met the design goals and acceptance criteria." No specific numerical values for agreement or bias are provided.
    PrecisionSerum: < 20% total CV Urine: < 20% total CV- Serum: Total CVs ranged from 2.5% to 6.2%. All reported total CVs are well below the <20% acceptance criterion. - Urine: Total CVs ranged from 3.8% to 6.4%. All reported total CVs are well below the <20% acceptance criterion.
    Interferences (Endogenous Substances)Not explicitly stated but inferred to be within acceptable clinical limits (e.g., typically ±10% or similar).- Serum: % Interference ranged from -7.8% to +13.2%. - Urine: % Interference ranged from -4.6% to 6.1%.
    Interferences (HAMA & RF)Acceptance Criteria: 100 ± 15% Recovery (for spiked samples)- HAMA: Grand Mean % Recovery = 100.1%, which is within 100 ± 15%. Individual % Recovery values ranged from 97.8% to 102.8%. - Rheumatoid Factor: Grand Mean % Recovery = 94.1%, which is within 100 ± 15%. Individual % Recovery values ranged from 89.7% to 102.1%.
    AnticoagulantsNo significant difference between serum and plasma recovery.The summary states: "The results indicate that there is no significant difference between the recovery of Cortisol in serum or plasma. The collection tubes evaluated show no adverse effects on the recovery of Cortisol, within the experimental error for the spiking study." This confirms the acceptance criterion.
    Specificity (Cross-Reactivity)Not explicitly stated but inferred to be low values to demonstrate specificity.Most cross-reactants showed very low % Cross Reactivity (0.0% to 2.8%). Noted exceptions were Fludrocortisone (36.8%) and Prednisolone (12.5%). The document doesn't explicitly state an acceptance criterion for cross-reactivity, but these values are reported.
    Calibration Curve Stability30 days stability.A 30-day calibration curve stability is supported by the data.
    Reagent On-Board Stability30 days stability.A 30-day on-board reagent stability claim is supported by the data.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Linearity: The study used a "regression analysis plot" and involved constructing "r regression and regression standard error (Reg SE)" for "each pool." The number of unique samples or data points for each dilution is not explicitly stated in the table, but it shows results for dilution range from 0% to 100% for two serum pools (65 ug/dL and 8 ug/dL). The guidance used was NCCLS guideline EP6-A. Provenance is not specified (e.g., country, retrospective/prospective).
    • Accuracy: Cortisol was spiked into human serum from 3 donors and human urine from 3 donors. The samples were analyzed in triplicate. Provenance is not specified.
    • Sensitivity:
      • LoB and LoD: 60 blank samples and 120 low-level samples were used to determine LoB and LoD.
      • Functional Sensitivity: Not explicitly stated, but based on CLSI guideline NCCLS EP17-A.
        Provenace of samples not specified.
    • Method Comparison: Patients samples were used (serum and urine). The number of samples is not provided. The study was conducted according to CLSI Guideline NCCLS EP9. Provenance not specified.
    • Precision:
      • 80 replicates (n=80) for each of the three MCC (Multi-constituent Control) levels and three serum panel levels on both the i2000 and i2000SR instruments.
      • 80 replicates (n=80) for each of the four urine panel levels on both the i2000 and i2000SR instruments.
        The study was performed using NCCLS guideline EP5-A2. Provenance not specified.
    • Interferences (Endogenous Substances):
      • Serum: 3 replicates (N=3) for each interferent concentration and Cortisol target concentration.
      • Urine: Data represents a comparison of "Unaltered Urine Control" vs. "Mock spike Control" and then individual interferents. Number of replicates not explicitly stated, but implies multiple measurements.
        The study was conducted using CLSI Guideline NCCLS EP7-A2. Provenance not specified.
    • Interferences (HAMA & RF):
      • HAMA: 10 samples (HAMA-1, HAMA-2 and 8 other samples) were tested.
      • Rheumatoid Factor (RF): 10 positive RF patient samples were assayed.
        Provenance not specified.
    • Specificity: Cortisol was spiked into cortisol-free human serum. Cross-reactant stock concentrates were prepared and spiked into aliquots of the 12 ug/dL cortisol serum. The total number of unique samples/aliquots is not precisely stated but involves a comprehensive list of 37 cross-reactants. Provenance not specified.
    • On-Board Stability: Details on the sample size used for stability studies (number of samples, replicates, etc.) are not provided in the summary. Provenance not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    The device is an in vitro diagnostic (IVD) assay for quantitative determination of cortisol. The "ground truth" for such devices is typically established through a reference method or known concentrations of analytes, not through human expert interpretation of images or clinical cases. Therefore, the concept of "experts" as in "radiologists with 10 years of experience" is not applicable here.

    • For studies like linearity, accuracy (recovery), sensitivity, and precision, the ground truth is based on:
      • Known concentrations: For spiked samples and controls.
      • Reference method values: For method comparison (in this case, the predicate device Abbott AxSYM Cortisol assay).
    • No information is provided about clinical expert adjudication or establishment of ground truth in the context of clinical expert review.

    4. Adjudication Method for the Test Set:

    Not applicable. As this is an IVD assay measuring an analyte concentration, the results are quantitative and compared against expected values, known concentrations, or results from a predicate device. There is no mention of a human expert adjudication process (e.g., 2+1, 3+1) because that process is typically associated with subjective interpretation tasks (like image review) where there can be disagreement among experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

    No. An MRMC study is not relevant for this type of IVD device. MRMC studies are typically used to evaluate the impact of AI on human reader performance, for instance, in diagnostic imaging where human readers interpret cases. This submission is for a standalone laboratory assay that quantifies a biomarker.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    Yes, the studies presented (linearity, accuracy, sensitivity, precision, interferences, specificity, stability) all represent the standalone performance of the ARCHITECT Cortisol assay without human-in-the-loop assistance in the interpretation of the assay result itself. The assay directly outputs a quantitative value. The human involvement is in performing the test and reviewing the quantitative result, but not in interpreting subjective data for diagnostical decision-making.

    7. The Type of Ground Truth Used:

    The ground truth used for these studies includes:

    • Known (prepared) concentrations: For linearity panels, spiked accuracy samples (serum and urine), and sensitivity studies (blank and low-level samples).
    • Reference method/predicate device comparison: For method comparison, the Abbott AxSYM Cortisol assay served as the comparative "truth" or reference method to demonstrate substantial equivalence.
    • Defined control material values: For precision studies (MCC samples).

    8. The Sample Size for the Training Set:

    This information is not applicable to this type of traditional in vitro diagnostic assay. These assays are developed through chemical and biological formulation and optimization, not through machine learning models that require distinct training sets. The development process involves extensive experimentation and validation, but not in the "training set" sense of AI/ML.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reason as above. There is no "training set" in the context of machine learning for this device. The development and optimization of the assay's reagents and protocols are based on established analytical chemistry principles and performance characteristics, validated against known standards and reference materials.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060998

    Validate with FDA (Live)

    Device Name
    QMS TOBRAMYCIN
    Manufacturer
    Date Cleared
    2006-07-21

    (101 days)

    Product Code
    Regulation Number
    862.3900
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QMS® Tobramycin assay is intended for the quantitative determination of tobramycin in human serum or plasma on automated clinical chemistry analyzers.

    The results obtained are used in the diagnosis and treatment of tobramycin overdose and in monitoring levels of tobramycin to help ensure appropriate therapy.

    Device Description

    The QMS® Tobramycin assay system is a homogeneous assay utilizing particle agglutination technology and is based on the competitive binding principle. The assay consists of reagents R1: anti- tobramycin monoclonal antibody and R2: tobramycin -coated microparticles. A six-level set of QMS® Tobramycin Calibrators (A through F) is used to calibrate the assay.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Seradyn QMS® Tobramycin assay, based on the provided 510(k) summary:

    Acceptance Criteria and Device Performance

    TestAcceptance CriteriaReported Device Performance
    Accuracy% Recovery: 100 ± 10%Mean Percent Recovery: 94.14%
    Linearity% Recovery: 100 ± 10%Correlation coefficient (R²): 0.9996
    PrecisionTotal CV: < 10%Low Control: 7.57%
    Mid Control: 4.23%
    High Control: 4.26%
    Interferences (Endogenous Substances)% Recovery: 100 ± 10%Albumin: 97.66%
    Bilirubin: 92.27%
    Cholesterol: 105.84%
    Gamma Globulins (IgG): 97.44%
    Hemoglobin (20 mg/dL): 97.16%
    Hemoglobin (500 mg/dL): 108.65%
    Uric Acid: 90.31%
    Rheumatoid Factor: 103.29%
    Triglyceride: 90.98%
    Interferences (HAMA)% Recovery: 100 ± 10%HAMA Type-1: 92.58%
    HAMA Type-2: 93.82%
    Sensitivity (LOQ)Lowest concentration reliably detected meeting accuracy requirements0.4 µg/mL
    Assay RangeBased on Accuracy, Linearity, Sensitivity0.4 to 10 µg/mL
    Method ComparisonExcellent correlation between predicate and new deviceN = 67, Slope = 0.979, y-intercept = -0.086, R = 0.992, R² = 0.984
    On-Board Stability (Calibration Curve)Sufficient stability for intended claim14 days
    On-Board Stability (Reagent)Sufficient stability for intended claim45 days

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Accuracy: The exact number of spiked samples is not explicitly stated beyond "across the range of the assay," but each sample was analyzed in triplicate. The samples used were human serum negative for the drug. Data provenance is not specified (e.g., country of origin), but it is a retrospective laboratory study using controlled samples.
      • Linearity: Similar to accuracy, a tobramycin in human serum pool was diluted, and samples were analyzed in triplicate. Data provenance is not specified.
      • Sensitivity (LOQ): Not specified.
      • Method Comparison: N = 67 patient samples. Data provenance is not specified, but it's a retrospective comparison using patient samples.
      • Precision: 80 measurements per control level (Low, Mid, High) over a period of time (e.g., 20 days with 4 replicates per day, which is a common NCCLS EP5-A2 setup, though not explicitly stated here). Data provenance not specified.
      • Interferences: 3 replicates per interfering substance. Data provenance not specified.
      • Cross-reactivity: Each cross-reactant drug was tested at a specific concentration (e.g., 30 µg/mL for 5-Fluorocytosine, 200 µg/mL for Amikacin).
      • Anticoagulants: Not explicitly stated but implies testing with different types of plasma samples and serum.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. This is a quantitative immunoassay for a drug, and the "ground truth" for the test set samples is typically established by precisely preparing known concentrations of tobramycin (for accuracy, linearity, sensitivity) or by using a a well-established comparative method (for method comparison). There are no human expert readers involved in establishing the ground truth for these types of in vitro diagnostic tests.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. As described above, this is laboratory testing of a quantitative assay, not an image interpretation or diagnostic performance study requiring expert adjudication. Recovered concentrations are compared to theoretical concentrations or to a predicate device's results.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI-assisted diagnostic device, nor does it involve human readers interpreting results in the way an MRMC study evaluates. It's an automated clinical chemistry analyzer.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the performance studies described (Accuracy, Linearity, Sensitivity, Precision, Method Comparison, Interferences, Stability) are all standalone performance evaluations of the assay system itself. This device is entirely automated; there is no "human-in-the-loop" once the sample is loaded and the assay initiated. The results are quantitative measurements produced directly by the instrument.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For Accuracy and Linearity: The ground truth was established by known, theoretical concentrations of tobramycin prepared by spiking or diluting human serum.
      • For Method Comparison: The "ground truth" was essentially the results obtained from the legally marketed predicate device (Abbott TDx/TDxFLx Tobramycin assay). The study aimed to show correlation and agreement with an already established method.
      • For Precision: The ground truth was the mean concentration of the control materials run multiple times.
      • For Interferences/Cross-reactivity: The ground truth involved known concentrations of tobramycin with the addition of known concentrations of interfering substances or cross-reactants.
    7. The sample size for the training set:

      • Not applicable. This is a traditional immunoassay, not a machine learning or AI-based device that requires a "training set" in the computational sense. The reagents and assay parameters are developed and optimized through standard biochemical and analytical chemistry processes, not through training on a dataset.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no "training set" in the context of this device. Development of the assay's reagents and methodologies would involve empirical testing and optimization to achieve desired analytical performance characteristics.
    Ask a Question

    Ask a specific question about this device

    K Number
    K060709

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-06-15

    (91 days)

    Product Code
    Regulation Number
    862.3450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Multigent® Gentamicin assay is intended for the quantitative determination of Gentamicin in human serum or plasma on the Architect C8000 System. The results obtained are used in the diagnosis and treatment of Gentamicin overdose and in monitoring levels of Gentamicin to ensure appropriate therapy.

    Device Description

    The Multigente Gentamicin assay system is a homogeneous assay utilizing particle agglutination technology and is based on the competitive binding principle. The assay consists of reagents R1: anti-gentamicin monoclonal antibody and R2: gentamicin-coated microparticles. A six-level set of Multigent" Gentamicin Calibrators (A through F) is used to calibrate the assay.

    AI/ML Overview

    This document describes the acceptance criteria and the studies performed to demonstrate the performance of the Multigent® Gentamicin assay.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance MetricAcceptance CriteriaReported Device Performance
    Accuracy (Recovery)100 ± 10% or 0.1 µg/mLConcentration (µg/mL)
    Linearity (Recovery)100 ± 10%Concentration (µg/mL)
    Sensitivity (LDD)Not explicitly stated in the "Acceptance Criteria" column, but the goal is to claim 0.1 µg/mL.The average LDD is 0.09 µg/mL. This supports a claim of 0.1 µg/mL.
    Interference (Bilirubin)100 ± 10%Mean Recovery: 3.42 µg/mL for a target of 3.44 µg/mL. (% Recovery for 20mg/dL Bilirubin: 99.42% (calculated from data). The table has an error in displaying this value.)
    Interference (Hemoglobin)100 ± 10%Mean Recovery: 3.38 µg/mL for a target of 3.44 µg/mL. (% Recovery for 2g/dL Hemoglobin: 98.26%)
    Interference (Triglyceride)100 ± 10%Mean Recovery: 3.30 µg/mL for a target of 3.44 µg/mL. (% Recovery for 1691 mg/dL Triglyceride: 95.83%)
    Interference (Total Protein)100 ± 10%Mean Recovery: 3.21 µg/mL for a target of 3.44 µg/mL. (% Recovery for 12 g/dL Total Protein: 93.41%)
    Interference (Rheumatoid Factor)100 ± 10%Mean Recovery: 3.26 µg/mL for a target of 2.46 µg/mL. (% Recovery for 582 IU Rheumatoid Factor: 132.34%) - This value exceeds the acceptance criteria of 100 ± 10%.
    Interference (HAMA Type-1)100 ± 10%Mean Recovery: 3.30 µg/mL. (% Recovery: 99.10%)
    Interference (HAMA Type-2)100 ± 10%Mean Recovery: 3.08 µg/mL. (% Recovery: 93.34%)
    Method Comparison (Correlation with Predicate)High correlation (e.g., R-squared close to 1)N = 55, Slope = 1.165, y-intercept = -0.719, R = 0.996, R² = 0.992. "Results show excellent correlation between the two assays."
    PrecisionCV (%) ranges from 1.07% to 5.69% for various control levels (details in document for within-run, between-day, between-run, and total precision).Low Control (2.68 µg/mL): Total CV 5.69%, SD 0.15Mid Control (6.47 µg/mL): Total CV 2.44%, SD 0.16High Control (9.41 µg/mL): Total CV 2.15%, SD 0.20(Full details for within-run, between-day, and between-run are available in the provided text, and all appear to be within acceptable limits for an immunoassay.)
    On-Board Stability (Calibration Curve)Data supports the stability period28 days
    On-Board Stability (Reagent)Data supports the stability period40 days
    AnticoagulantsNo significant difference in recovery with various anticoagulants"The results indicate that there is no significant difference between the recovery of Gentamicin in serum or plasma. The collection tubes evaluated show no adverse effects on the recovery of Gentamicin, within the experimental error for the spiking study."

    Note regarding Rheumatoid Factor interference: The reported % Recovery (132.34%) for Rheumatoid Factor at 582 IU exceeds the stated acceptance criteria of 100 ± 10%. This indicates potential interference from high levels of Rheumatoid Factor, which should be noted in the device labeling.

    2. Sample Size and Data Provenance:

    • Accuracy: 5 samples (0.25, 1.00, 2.25, 4.50, 8.00 µg/mL) each run in triplicate (total of 15 measurements in the table shown). Data provenance not specified (retrospective/prospective, country of origin).
    • Linearity: 4 serially diluted samples (6.88, 5.16, 3.44, 1.72 MG/ML) each run in triplicate (total of 12 measurements in the table shown). Data provenance not specified.
    • Precision: 3 control levels (low, mid, high) tested, with N=80 for each (number of replicates over time). Data provenance not specified.
    • Sensitivity: Calibrator A (0 µg/mL) run for a total of 20 replicates. Data provenance not specified.
    • Interferences (Endogenous Substances):
      • Bilirubin: 3 replicates for "N=3" (interpreted as the number of independent samples or measurement replicates, as the "N" column is inconsistent).
      • Hemoglobin: 2 replicates for "N=2".
      • Triglyceride, Total Protein: "N" is listed as a non-numeric character (presumably an error in transcription, but the text mentions "run in triplicate" for triglyceride and total protein).
      • Rheumatoid Factor: "N" is listed as a non-numeric character (text mentions "run in triplicate").
      • Data provenance not specified.
    • Interferences (HAMA): Duplicate HAMA samples (Type-1 and Type-2) and duplicate control samples. Data provenance not specified.
    • Interferences (Common Co-Administered Drugs): Test samples and control samples assayed in duplicate for each drug. Data provenance not specified.
    • Anticoagulants: Blood drawn from at least ten healthy donors for each of the 9 tube types. Samples were spiked with Gentamicin and run in duplicate. Data provenance not specified.
    • Method Comparison: 55 serum and plasma samples, ranging from 0.78 to 9.02 µg/mL Gentamicin. Data provenance not specified.

    3. Number of Experts used to establish the ground truth for the test set and the qualifications of those experts:

    This device is an in vitro diagnostic (IVD) assay for quantitative determination of a drug concentration. The "ground truth" for chemical concentration assays is typically established by reference methods, primary standards, or gravimetric/volumetric preparation of known concentrations. There is no mention of human "experts" establishing ground truth in the traditional sense of clinical opinion (e.g., radiologists, pathologists). The ground truth for performance studies like accuracy and linearity is based on the theoretical concentrations of prepared samples or reference materials. For method comparison, the predicate device (Abbott TDx®/TDxFLx® Gentamicin assay) serves as the reference method.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. Adjudication methods are typically used in studies involving subjective assessment (e.g., image interpretation) where multiple readers provide independent evaluations that might conflict. For quantitative chemical assays, the result is a numerical value, and "adjudication" is not a standard practice. Statistical methods are used to compare results to expected values or reference methods.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is an in vitro diagnostic assay, not an AI-assisted diagnostic tool for human readers. Therefore, an MRMC study or assessment of human reader improvement with AI assistance is irrelevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The performance studies described (Precision, Accuracy, Linearity, Sensitivity, etc.) represent the standalone performance of the Multigent® Gentamicin assay system. The device itself generates a quantitative result without direct human interpretation of a visual output. The human-in-the-loop would be the laboratory personnel operating the Architect C8000 System and interpreting the numerical result in a clinical context.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth for this device's performance relies on:

    • Theoretical Concentrations: For accuracy, linearity, and sensitivity studies, known concentrations of Gentamicin in control matrices serve as the ground truth. These are typically prepared using gravimetric or volumetric methods with highly pure reference standards.
    • Reference Method: For the method comparison study, the Abbott TDx®/TDxFLx® Gentamicin assay served as the reference (or comparative) method, with its results considered the "ground truth" for evaluating the new device's correlation.

    8. The sample size for the training set:

    Not applicable. This device is a chemical immunoassay, not a machine learning or artificial intelligence system that requires a "training set" in the computational sense. The assay works based on established biochemical principles and reagents rather than being "trained" on data.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of in vitro diagnostic device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052826

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-12-23

    (79 days)

    Product Code
    Regulation Number
    862.3320
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QMS® Quinidine assay is intended for the quantitative determination of quinidine in human serum or plasma on automated clinical chemistry analyzers.

    The results obtained are used in the diagnosis and treatment of quinidine overdose and in monitoring levels of quinidine to help ensure appropriate therapy.

    Device Description

    The QMS® Quinidine assay system is a homogeneous assay utilizing particle agglutination technology and is based on the competitive binding principle.

    In particle agglutination assays, the degree of agglutination is inversely proportional to the quantity of free drug in the reaction well. Hence, if no drug is present in the sample, the antibodies in the QMS® Quinidine Antibody Reagent (R1) will bind only to the bound drug on the particle which will cause it to agglutinate and will result in higher absorbance. If increased amount of competing drug is present in the sample, this will result in decreased binding of bound drug by the antibody, resulting in a relative decrease in particle agglutination. This in turn results in lower absorbance.

    The precise relationship between particle agglutination of the unlabeled drug in the sample is established by measuring the absorbance values of calibrators with known concentration of the The absorbance of unknown samples can be interpolated from the absorbance values of the drug, calibration curve and the concentration of the drug present in the sample can be calculated.

    The assay consists of reagents R1: anti-quinidine monoconal and R2: quinidine-oated microparticles. A six-level set of QMS® Quinidine Calibrators (A throu

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the QMS® Quinidine assay, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance CriteriaReported Device Performance
    Accuracy (% Recovery)100% ± 10%Mean Percent Recovery: 97.71% (for spiked samples)
    Linearity (R²)Not explicitly stated as a numerical AC, but implied good linearity0.9995 (correlation coefficient for linear regression)
    Sensitivity (LDD)Claim of 0.2 µg/mL supportedAverage LDD: 0.09 µg/mL
    Assay RangeBased on Accuracy, Linearity, Sensitivity0.2 to 8.0 µg/mL
    Precision (Total CV)< 10%Control 1: 9.09%Control 2: 6.37%Control 3: 5.83%
    Interference (% Recovery)100% ± 10% (for endogenous substances)Bilirubin: 103.0%Hemoglobin: 100.0%Triglyceride: 92.18%Total Protein: 99.68%HAMA Type-1: 91.12%HAMA Type-2: 90.65%
    Calibration Curve StabilityNot explicitly stated as a numerical ACSupported for 28 days
    Reagent On-Board StabilityNot explicitly stated as a numerical ACSupported for 25 days

    2. Sample Size Used for the Test Set and Data Provenance

    • Accuracy: Samples analyzed in duplicate. The number of samples for the accuracy study was 3 "theoretical concentrations" of quinidine (2.0, 4.0, 8.0 µg/mL), each tested in duplicate (total 6 measurements).
    • Linearity: Not explicitly stated as a number of individual samples, but the study was "based on the NCCLS guideline EP6: Evaluation of the Linearity of Quantitative Measurement." It involved testing 5 theoretical concentrations (0.25, 0.75, 1.5, 3.0, 6.0 µg/mL), each in duplicate.
    • Method Comparison: N = 50 (number of patient samples)
    • Precision: 80 measurements per control level (for 3 control levels, so 240 total measurements).
    • Specificity: N=3 for each cross-reactant.
    • Interferences (Endogenous Substances): N=2 for Bilirubin and Hemoglobin, N=3 for Triglyceride and Total Protein.
    • Interferences (HAMA): Samples analyzed in duplicate for Control, HAMA Type-1, HAMA Type-2.
    • Data Provenance: The document does not specify the country of origin for the human serum/plasma samples. It states that accuracy was determined by spiking USP traceable quinidine into "human serum negative for the drug." Method comparison used "patient samples." The studies appear to be retrospective in nature, as they are laboratory-based validation studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information (number and qualifications of experts) is not applicable for this device and study. The QMS® Quinidine assay is an in vitro diagnostic (IVD) device for quantitative determination of a drug in serum/plasma. Its performance is evaluated against established analytical methods and reference standards, not by human expert interpretation of images or clinical data for ground truth.

    4. Adjudication Method for the Test Set

    This is not applicable as the studies involve quantitative measurements and comparisons to reference values or other assays, not human interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically used for medical imaging devices where human readers interpret cases with and without AI assistance. This document describes the analytical performance of a quantitative immunoassay.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies presented are all standalone performance evaluations of the QMS® Quinidine assay. The device itself is an automated clinical chemistry analyzer for quantitative determination, meaning it operates without direct human interpretive input beyond sample loading and results reporting. The performance metrics (accuracy, linearity, precision, etc.) are inherent to the assay's ability to measure quinidine concentrations.

    7. The Type of Ground Truth Used

    The ground truth for the various studies includes:

    • Spiked Concentrations: For accuracy and linearity, USP traceable quinidine was spiked into human serum to create samples with known theoretical concentrations.
    • Reference Method Results: For method comparison, the Abbott TDx/TDxFLx Quinidine assay served as the reference method, and its results were compared to the QMS® Quinidine assay.
    • Known Concentrations/Materials: For sensitivity, specificity, and interference studies, samples with known concentrations of drug, cross-reactants, or interfering substances were used.

    8. The Sample Size for the Training Set

    The document describes the validation studies for the QMS® Quinidine assay. For an IVD device like this, the "training set" would typically refer to the data used during the development and optimization of the assay's reagents and methodology. This information is not provided in this 510(k) summary, as it generally focuses on the final validation results, not the developmental process.

    9. How the Ground Truth for the Training Set Was Established

    As the "training set" information is not provided, the method for establishing its ground truth is also not available in this document. During development, ground truth would likely be established using highly accurate reference methods, certified reference materials, and meticulously prepared known-concentration samples, similar to the validation methods but on a smaller scale and iteratively.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051211

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-11-18

    (191 days)

    Product Code
    Regulation Number
    862.3350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QMS Zonisamide assay is intended for the quantitative determination of zonisamide in human serum or plasma on automated clinical chemistry analyzers.

    Zonisamide concentrations can be used as an aid in management of patients treated with zonisamide.

    The QMS® Zonisamide Calibrator set is intended for use in calibration of the QMS Zonisamide assay.

    The QMS® Zonisamide Control set is intended for use in quality control of the QMS Zonisamide assay.

    Device Description

    The QMS Zonisamide assay system is a homogeneous assay utilizing particle agglutination technology and is based on the competitive binding principle.

    The assay consists of reagents R1: anti-Zonisamide rabbit polyclonal antibody and R2: Zonisamidecoated microparticles. A six-level set of QMS® Zonisamide Calibrators (A through F) is used to calibrate the assay. A three-level set of QMS® Zonisamide Controls (1 through 3) is used for quality control of the assay.

    AI/ML Overview

    The provided text describes the QMS® Zonisamide assay, a homogeneous particle-enhanced turbidimetric immunoassay for the quantitative determination of zonisamide in human serum or plasma. The information focuses on demonstrating its substantial equivalence to a legally marketed predicate device (Innofluor® Phenytoin) through performance testing.

    Here's an analysis of the acceptance criteria and the studies performed, formatted as requested:

    1. Table of Acceptance Criteria and the Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/Study GoalReported Device Performance
    Accuracy and LinearityTo evaluate the accuracy and linearity of the assayDemonstrated accuracy and linearity based on NCCLS guideline EP6.
    SensitivityAnalytical Sensitivity (Least Detectable Dose - LDD)1.0 µg/mL
    Functional Sensitivity (Limit of Quantitation - LOQ)3.0 µg/mL
    Assay RangeReportable range for the assay3.0 to 50.0 µg/mL (Package insert claim based on Accuracy, Linearity, and Sensitivity data)
    Method ComparisonCorrelation with another method (implicit: to show agreement with a recognized method)Correlation studies conducted using NCCLS Guideline EP9 (no specific correlation values provided in the summary).
    PrecisionTo evaluate the precision performance of the assayPrecision study performed using NCCLS guideline EP5 (no specific precision values provided in the summary).
    SpecificityNo significant cross-reactivity with major metabolites (NAZ and SMAP)No significant cross-reactivity for NAZ and SMAP.
    InterferencesNo significant cross-reactivity with common interfering substances/drugsOf 26 drugs tested, none showed cross-reactivity.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the sample sizes used for the test sets in the accuracy, linearity, method comparison, precision, specificity, or interference studies.

    Regarding data provenance:

    • The studies were conducted using well-established NCCLS guidelines, implying standardized laboratory testing.
    • The method comparison used "patient samples," suggesting a clinical context, but no specifics about country of origin or whether it was retrospective or prospective are provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided in the document. The studies described are primarily analytical performance studies of a quantitative assay, where the "ground truth" for assay accuracy and calibration would typically be established by reference methods or gravimetric preparation of standards, not by human expert interpretation.

    4. Adjudication Method for the Test Set

    This information is not applicable and therefore, not provided. Adjudication methods (like 2+1, 3+1) are typically used in studies involving human interpretation (e.g., image analysis, clinical diagnosis) where there might be inter-reader variability. For an automated quantitative assay, the "ground truth" is typically determined by objective reference methods or precise measurements, not by expert consensus adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This type of study is relevant for diagnostic devices that involve significant human interpretation (e.g., radiologists reading images) and the AI's impact on human performance. The QMS® Zonisamide assay is a quantitative determination assay performed on automated clinical chemistry analyzers, not an interpretive diagnostic tool requiring human readers in that sense.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies described are inherently standalone performance tests of the assay. The QMS® Zonisamide assay is a laboratory test system designed to provide quantitative results directly from automated analyzers. The reported performance metrics (accuracy, linearity, sensitivity, precision, specificity, interferences) evaluate the device's capability to accurately and reliably measure zonisamide concentrations without direct human interpretation influencing the measurement outcome.

    7. The Type of Ground Truth Used

    The ground truth for this type of quantitative assay would typically be established through:

    • Reference Methods / Gravimetric Standards: For accuracy and linearity, highly precise and accurate reference methods or gravimetrically prepared standards with known zonisamide concentrations would be used.
    • Known Concentrations: For sensitivity, specificity, and interference studies, samples with precisely known concentrations of zonisamide, metabolites, or interfering substances would be utilized.
    • Patient Samples: For method comparison studies, the "ground truth" would be the results obtained from an established comparative method on actual patient samples.

    The document does not explicitly detail the exact methods used for establishing ground truth for each study but implies standard laboratory practices using "patient samples" for method comparison and reference to NCCLS guidelines for other analytical performance.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore, not provided. The QMS® Zonisamide assay is a biochemical immunoassay, not a machine learning or AI algorithm that requires a separate "training set" in the context of supervised learning. The assay's performance is driven by its reagent formulation, reaction kinetics, and instrument calibration, not by an algorithm trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no "training set" in the context of an immunoassay. The chemical and biological principles of the assay itself, along with the manufacturing and quality control of the reagents and calibrators, establish its analytical performance capabilities.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052815

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-11-01

    (28 days)

    Product Code
    Regulation Number
    862.3035
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QMS® Amikacin assay is intended for the quantitative determination of amikacin in human serum or plasma on automated clinical chemistry analyzers.

    The results obtained are used in the diagnosis and treatment of amikacin overdose and in monitoring levels of amikacin to ensure appropriate therapy.

    Device Description

    The QMS® Amikacin assay system is a homogeneous assay utilizing particle agglutination technology and is based on the competitive binding principle.

    In particle agglutination assays, the degree of agglutination is inversely proportional to the quantity of free drug in the reaction well. Hence, if no drug is present in the sample, the antibodies in the QMS Amikacin Antibody Reagent (R1) will bind only to the bound drug on the particle which will cause it to agglutinate and will result in higher absorbance. If increased amount of competing drug is present in the sample, this will result in decreased binding of bound drug by the antibody, resulting in a relative decrease in particle agglutination. This in turn results in lower absorbance.

    The precise relationship between particle agglutination of the unlabeled drug in the sample is established by measuring the absorbance values of calibrators with known concentration of the drug. The absorbance of unknown samples can be interpolated from the absorbance values of the calibration curve and the concentration of the drug present in the sample can be calculated.

    The assay consists of reagents R1: anti-amikacin monoclonal antibody and R2: amikacin-coated microparticles. A six-level set of QMS® Amikacin Calibrators (A through F) is used to calibroothe assay.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the QMS® Amikacin assay, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance CriteriaReported Device Performance
    Accuracy% Recovery: 100 ± 10%Mean Percent Recovery: 94.02% (Specific recoveries: 95.87% for 9.2 µg/mL and 92.17% for 18.4 µg/mL) - Meets criteria
    LinearityCorrelation coefficient (R2) demonstrating linearityCorrelation Coefficient (R2): 0.9998 (Mean Percent Recovery: 100.41% over a range of 1.5 to 42.5 µg/mL; specific recoveries from 95.71% to 111.33%) - Meets criteria
    SensitivityClaimed LDD: 0.8 µg/mLAverage LDD: 0.54 µg/mL - Exceeds claimed performance (better sensitivity)
    Assay RangeNot explicitly stated as acceptance criteria, but derived from other dataReportable Range: 1.5 to 50 µg/mL (Based on Accuracy, Linearity, and Sensitivity data)
    Method ComparisonExcellent correlation with predicate deviceCorrelation to Abbott TDx/TDxFLx Amikacin: N = 56, Slope = 1.00, y-intercept = 0.25, R = 0.996, R2 = 0.992. Results show excellent correlation. - Meets criteria
    PrecisionTotal CV: < 10%Total CVs: Low Control: 9.94%, Mid Control: 6.22%, High Control: 6.32% - Meets criteria
    Interferences (Endogenous Substances)% Recovery: 100 ± 10%Bilirubin (15mg/dL): 96.40% Hemoglobin (10g/L): 93.42% Triglyceride (1691 mg/dL): 96.30% Total Protein (12 g/dL): 96.00% - All meet criteria
    Interferences (HAMA)% Recovery: 100 ± 10%HAMA Type-1: 100.5% HAMA Type-2: 98.04% - All meet criteria
    On-Board Stability (Calibration Curve)28 days stability supported by dataSupported: 28 days - Meets criteria
    On-Board Stability (Reagent)40 days stability supported by dataSupported: 40 days - Meets criteria

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Accuracy: Samples were spiked human serum. No specific number for "samples analyzed" is given, but it mentions "in duplicate." The specific source/country of the human serum is not mentioned. This appears to be a laboratory-controlled, prospective study.
    • Linearity: Not explicitly stated, but based on "a study based on the NCCLS guideline EP6." A set of 5 theoretical concentrations was tested, each in duplicate. USP Amikacin was used, indicating an in vitro, laboratory-controlled prospective study.
    • Sensitivity: Not explicitly stated, but related to the lowest detectable dose from laboratory measurements.
    • Method Comparison: N = 56 patient samples. The data provenance (country of origin, retrospective/prospective) is not explicitly stated, but the mention of "patient samples" suggests retrospective or prospective clinical samples. It does not state if these were collected specifically for this study or were archived.
    • Precision: 80 measurements for each of the three control levels over multiple runs/days. This refers to laboratory-controlled measurements of quality control materials.
    • Interferences:
      • Endogenous Substances: N=2 or N=3 for each interferent tested. These were likely laboratory-prepared samples with added interferents, a prospective study.
      • Human Anti-Mouse Antibodies (HAMA): Not explicitly stated, but given as "Rep 1" and "Rep 2" for HAMA Type-1 and Type-2, suggesting laboratory-spiked samples.
      • Cross-reactants: No 'N' is listed, but various cross-reactant drugs were tested at specified concentrations. This would be a laboratory-controlled study.
      • Anticoagulants: Not explicitly stated, but likely laboratory-controlled comparison of serum and plasma samples, a prospective study.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This document describes the performance characteristics of an in-vitro diagnostic (IVD) device for measuring a drug concentration. The "ground truth" in this context is established by:

    • Known Reference Values: For accuracy, linearity, and sensitivity, the ground truth is the spectroscopically pure USP traceable amikacin spiked into human serum at known concentrations.
    • Reference Method: For method comparison, the ground truth is considered the results obtained from the predicate device, the Abbott TDx/TDxFLx® Amikacin assay.

    Therefore, there were no clinical experts (e.g., radiologists) involved in establishing the "ground truth" in the way one would for image interpretation or diagnosis. The ground truth relies on analytical chemistry principles and a comparative reference method for this type of device.

    4. Adjudication Method for the Test Set

    Not applicable. This is an analytical device for quantitative measurement, not a human-interpreted diagnostic output requiring adjudication. The performance is assessed by comparing the device's output to known input concentrations or to measurements from a predicate device.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is a standalone analytical device, not a human-in-the-loop system where human reader performance is being evaluated or augmented by AI.

    6. Standalone (Algorithm Only) Performance

    Yes, the studies presented here are entirely standalone performance evaluations of the QMS® Amikacin assay. The device measures amikacin concentration in a sample without human interpretation or intervention in the measurement process itself. The "algorithm" here refers to the underlying chemical reaction and optical detection described in the "Description of Device," which is inherent to the device's operation.

    7. Type of Ground Truth Used

    • Known (Spiked) Concentrations: For accuracy, linearity, and sensitivity, the ground truth was established by precisely preparing samples with known concentrations of USP traceable amikacin.
    • Reference Standard (Predicate Device): For method comparison, the Abbott TDx/TDxFLx® Amikacin assay served as the reference standard to which the new device's results were compared.

    8. Sample Size for the Training Set

    No information about a "training set" is provided. This is a traditional IVD device, not a machine learning or AI-driven system that would typically have a distinct "training set" and "test set" in the computational sense. The "development" and "validation" of the assay are implied by the studies conducted, but not in the context of training a machine learning model.

    9. How Ground Truth for the Training Set Was Established

    Not applicable, as no dedicated "training set" is described for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K050419

    Validate with FDA (Live)

    Device Name
    QMS VANCOMYCIN
    Manufacturer
    Date Cleared
    2005-04-01

    (42 days)

    Product Code
    Regulation Number
    862.3950
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QMS® Vancomycin assay is intended for the quantitative determination of vancomycin in human serum or plasma on the Hitachi 717 analyzer.

    The results obtained are used in the diagnosis and treatment of vancomycin overdose and in monitoring levels of vancomycin to ensure appropriate therapy.

    Device Description

    The QMS® Vancomycin assay is a homogeneous particle-enhanced turbidimetric immunoassay. The assay is based on competition between drug in the sample and drug coated onto a microparticle for antibody binding sites of the vancomycin antibody reagent. The vancomycin-coated microparticle reagent is rapidly aqglutinated in the presence of the anti-vancomycin antibody reagent and in the absence of any competing drug in the sample. The rate of absorbance change is measured photometrically, and is directly proportional to the rate of agglutination of the particles. When a sample containing vancomycin is added, the agglutination reaction is partially inhibited, slowing down the rate of absorbance change. A concentrationdependent classic agglutination inhibition curve can be obtained, with maximum rate of agglutination at the lowest vancomycin concentration and the lowest agglutination rate at the highest vancomycin concentration.

    The assay consists of reagents R1: vancomycin monoclonal and R2: vancomycin-coated microparticles. A six-level set of QMS® Vancomycin Calibrators (A through F) i

    AI/ML Overview

    Here's a summary of the acceptance criteria and study findings for the QMS® Vancomycin assay, based on the provided 510(k) summary:

    Acceptance Criteria and Device Performance for QMS® Vancomycin Assay

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state pre-defined "acceptance criteria" with numerical thresholds for all tests. Instead, it presents study results and implies that the observed performance characteristics were deemed acceptable for substantial equivalence to the predicate device. For the purpose of this table, "Acceptance Criteria (Implied)" are derived from the overall goal of demonstrating equivalency or typical performance expectations for such assays, and "Reported Device Performance" are the results presented in the summary.

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    PrecisionAcceptable within-run, between-run, between-day, and total CVs for clinical use.Low Control (7.57 µg/mL): Total CV 8.84% Mid Control (20.79 µg/mL): Total CV 6.21% High Control (33.65 µg/mL): Total CV 5.12%
    Accuracy (Recovery)Mean Percent Recovery close to 100% across the assay range (e.g., 90-110%).Mean Percent Recovery: 99.61% (ranging from 91.11% to 110.61% across 9 theoretical concentrations from 5.00 to 100.00 µg/mL).
    Linearity (Dilution)Mean Percent Recovery close to 100% across the dilution range.Mean Percent Recovery: 100.17% (ranging from 95.71% to 107.20% across 5 theoretical concentrations from 2.50 to 75.00 µg/mL). R2= 0.9998 (from scatter plot).
    Sensitivity (LDD)Least Detectable Dose (LDD) must support the claimed lower limit of detection.LDD: 0.46 µg/mL, supporting a claim of 0.55 µg/mL.
    Specificity (CDP-I)Cross-reactivity with the vancomycin metabolite CDP-I should be low (e.g., <10%).CDP-I Cross-reactivity: <5% at 100 µg/mL CDP-I in serum containing 25 µg/mL vancomycin.
    Interferences (Endogenous)Less than 10% error in vancomycin detection with common endogenous substances at specified concentrations.All tested substances (Albumin, Bilirubin, Cholesterol, IgG, Hemoglobin, Heparin, Triglyceride, Rheumatoid Factor) resulted in less than 10% error at their respective tested concentrations (ranging from 91.64% to 100.07% recovery).
    Interferences (HAMA)HAMA interference should not lead to significantly falsely elevated results. The recovery for HAMA samples should be comparable to control samples.HAMA Type-1: Mean % Recovery 105.31% vs. Control 26.85 µg/mL. HAMA Type-2: Mean % Recovery 103.91% vs. Control 28.55 µg/mL. (Implied acceptable level of non-interference).
    Interferences (Co-Administered Drugs)Cross-reactivity with common co-administered drugs should be very low (e.g., <1%).All 44 listed co-administered drugs showed <0.3% cross-reactivity at 500 µg/mL in a vancomycin spiked serum pool at 25 µg/mL.
    Interferences (Structurally Related Drug)Cross-reactivity with structurally similar compounds like teicoplanin should be low.Teicoplanin Cross-reactivity: 0.03% to 0.68% across concentrations of 10-100 µg/mL teicoplanin.
    Method ComparisonExcellent correlation with a legally marketed predicate device (Abbott TDx® Vancomycin assay), demonstrating substantial equivalence. Slope-1, intercept-0, high R^2.N=146 patient serum samples. Slope: 1.031 y-intercept: 1.115 R^2: 0.970 (Results considered to show "excellent correlation").
    On-Board StabilityCalibration curve and reagents should demonstrate adequate stability for practical use.Calibration Curve Stability: 31 days supported by data. Reagent On-Board Stability: 62 days supported by data.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision: 80 replicates for each of the three control levels (Low, Mid, High) over multiple runs and days.
    • Accuracy: Triplicate measurements for 9 different theoretical concentrations.
    • Linearity: Triplicate measurements for 5 different theoretical concentrations.
    • Sensitivity: Determined using standard deviation of duplicate determinations of the zero calibrator and the 1st non-zero calibrator (Cal B, 5µg/mL).
    • Specificity (CDP-I): Not explicitly stated, but implies a sufficient number of replicates to determine cross-reactivity percentage at a specified concentration.
    • Interferences (Endogenous): Triplicate measurements for each of the 8 interfering substances tested.
    • Interferences (HAMA): Triplicate measurements for HAMA Type-1, HAMA Type-2, and their respective controls.
    • Interferences (Co-Administered Drugs): Not explicitly stated for each drug, but for all 44 drugs, measured at 500 µg/mL in a vancomycin spiked serum pool at 25 µg/mL, compared to a control serum.
    • Interferences (Structurally Related Drug - Teicoplanin): Not explicitly stated, but implies sufficient measurements at 4 different teicoplanin concentrations.
    • Method Comparison: 146 serum samples from patients.

    Data Provenance: The document does not explicitly state the country of origin for any data or whether the studies were retrospective or prospective, beyond stating "human serum or plasma" for the intended use and "patient serum samples" for method comparison. Given the context of a 510(k) submission in the US, it's likely conducted or overseen in the US.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This device is an in-vitro diagnostic assay for measuring vancomycin concentration. The "ground truth" for the test set (e.g., actual vancomycin concentrations, absence/presence of interfering substances) would be established by the precise formulation of controls, calibrators, spiked samples, and highly accurate reference methods or laboratory standards, rather than expert human interpretation. For the method comparison study, the "ground truth" for the patient samples was established by the predicate device, the Abbott TDx® Vancomycin assay. Therefore, the concept of "experts establishing ground truth" in the typical sense of radiologists or pathologists is not directly applicable here.

    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1 or 3+1) are typically used in studies involving subjective interpretation of medical images or clinical data where human readers might disagree. For an in-vitro diagnostic device that provides quantitative measurements, the "adjudication" is inherent in the analytical methods used to establish reference values (e.g., highly pure reference materials, established laboratory protocols, or a predicate device). No human expert adjudication was involved in generating the values used as "ground truth" for the performance studies presented.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned. This type of study is designed to evaluate the impact of a new diagnostic tool on decisions or performance of human readers, typically in image interpretation. This is not relevant for a quantitative laboratory assay like QMS® Vancomycin, which provides a numerical output for drug concentration.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, the studies presented are all standalone performance evaluations of the QMS® Vancomycin assay. The device directly measures vancomycin concentration in a sample, and the performance characteristics (precision, accuracy, sensitivity, specificity, interference, method comparison) are assessed based solely on its own analytical output. There is no "human-in-the-loop" interaction in the immediate result generation for a quantitative assay of this nature.

    7. Type of Ground Truth Used

    The ground truth used for various studies includes:

    • Precision: Defined concentrations of vancomycin in control materials.
    • Accuracy (Recovery): Known theoretical concentrations of USP traceable vancomycin spiked into human serum.
    • Linearity (Dilution): Known theoretical concentrations of vancomycin prepared by dilution.
    • Sensitivity: Relies on the statistical properties of measurements of a zero calibrator and a known low-concentration calibrator (Cal B at 5 µg/mL).
    • Specificity & Interferences: Known concentrations of vancomycin, the metabolite CDP-I, various endogenous substances, HAMA, co-administered drugs, and teicoplanin, either alone or spiked into vancomycin-containing serum.
    • Method Comparison: Results obtained from the predicate device, the Abbott TDx® Vancomycin assay, used as the reference method for patient serum samples.

    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of device development. For an immunoassay like QMS® Vancomycin, the development typically involves extensive characterization of reagents, optimization of assay parameters, and formulation of calibrators and controls. This process would involve numerous experiments and data points before the final performance studies (summarized here) are conducted to demonstrate suitability for market clearance. The data provided focuses solely on the validation/test performance of the finalized device.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, a formal "training set" as understood in machine learning (where an algorithm learns from data with established ground truth) is not directly applicable to the development of a homogeneous particle-enhanced turbidimetric immunoassay. Instead, the "ground truth" during development would be established through:

    • Chemical and biological characterization of reagents: Ensuring the quality and specificity of antibodies and microparticles.
    • Use of highly purified vancomycin standards (e.g., USP traceable): To prepare calibrators and to spike samples for recovery and linearity studies.
    • Reference laboratory methods: Potentially used during early development to confirm the accuracy of initial assay designs or to validate the content of control materials.
    • Industry standards and guidelines: Extensive experimentation guided by principles for developing accurate and precise quantitative immunoassays.
    Ask a Question

    Ask a specific question about this device

    K Number
    K042307

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-11-03

    (70 days)

    Product Code
    Regulation Number
    862.1150
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2