Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K143398

    Validate with FDA (Live)

    Date Cleared
    2015-06-26

    (212 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Plato MICROCATH 27B Microcatheter is intended for the introduction of interventional devices and infusion of diagnostic or therapeutic agents to the peripheral system. The cather is not intended for use in cither the coronary or neuro vasculature.

    Device Description

    The Plato MICROCATH® 27B Microcatheter is a single lumen microcatheter constructed with a flexible polymer shaft of varying stiffness to aide in accessing vasculature. The catheter is designed to be used with a guide catheter and a steerable guidewire for accessing the vasculature. The proximal end of the catheter has a diameter of 3.4F, tapering to a distal OD of 3.2F, which can be inserted into a 5F guide catheter. The inner diameter is constant throughout the shaft length and accommodates up to a 0.025" guidewire. The catheter is 150 cm in length with a straight tip configuration which can be steam-shaped to the doctor's preferred shape. A steam shaping mandrel is included in the packaging. The microcatheter has hydrophilic coating on the outer distal shaft to reduce friction during manipulation in vessels and has one radiopaque tip marker to facilitate fluoroscopic visualization.

    AI/ML Overview

    The provided document describes the FDA 510(k) summary for the Plato MICROCATH® 27B Microcatheter, focusing on demonstrating substantial equivalence to predicate devices, not on proving effectiveness for a new diagnostic or interventional AI device. Therefore, the traditional acceptance criteria, study design parameters, and concepts like 'human readers improve with AI' are not applicable to this submission.

    Instead, the document details performance data collected from non-clinical testing to show that the new microcatheter is as safe and effective as existing, legally marketed predicate devices. The acceptance criteria essentially derive from established standards for medical devices and comparisons to the predicate devices.

    Here's an adaptation of your requested table and information, tailored to the context of this medical device submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test (Acceptance Criteria Derived From)Acceptance Criteria (Implicit/Explicit from standards; "Comparable to predicate" suggests meeting predicate's performance)Reported Device Performance
    Biocompatibility:
    Cytotoxicity (ISO 10993-5)Non-cytotoxic (Scores = 0)Non-cytotoxic. Test article scores were 0 at 48 hours.
    Sensitization (ISO 10993-10)Negative for dermal sensitization (Scores = 0)Negative for dermal sensitization. Test articles sensitization scores were all 0.
    Irritation/Intracutaneous Reactivity (ISO 10993-10)Non-irritating (No significantly greater biological reaction than control)Non-irritating. Extracts of the test article did not show a significantly greater biological reaction than sites injected with control article.
    Acute Systemic Toxicity (ISO 10993-11)Non-toxic (No toxicity or animal weight loss)Non-toxic. Test articles showed no toxicity or animal weight loss for both cottonseed oil and saline extracts for 72 hour test period.
    Materials-Mediated Pyrogenicity (ISO 10993-11)Non-pyrogenic (No increases in individual temperatures)Non-pyrogenic. No increases in individual temperatures.
    Hemocompatibility – Hemolysis (ASTM 758-08)Non-hemolytic (Corrected hemolysis index below threshold)Non-hemolytic. Corrected hemolysis index was 0.15% by direct method, 0.23% by extract.
    Partial Thromboplastin Time (ASTM F2382)Similar activation profile to predicate (Minimal activators, minimal difference in clotting times)Both test article and predicate were minimal activators. Difference in clotting times between test article and predicate was 3 seconds.
    Complement Activation (ISO 10993-4)Less than or comparable to predicateComplement activation by the test article was less than that of the predicate device.
    Thrombogenicity in Dogs (ISO 10993-4)Comparable to predicate (Thrombosis grade, weight changes)Thrombosis grade comparable to predicate. Weight changes of the implants comparable between test and control articles.
    Functional Testing:
    Visual/Dimensional InspectionNo surface defects, visible droplets, or defects causing trauma; Meets dimensional specificationsNo surface defects or visible droplets of coating on catheters. All catheters met dimensional specifications.
    Air Ingress/Negative Collapse (ISO 10555)No air ingress, no lumen collapseHub fittings do not allow air ingress and no evidence of lumen collapse.
    Kink ResistanceResistant to kinking around small diameter turns per specificationDevice was resistant to kinking around small diameter turns per specification.
    Tensile Strength/Elongation (ISO 10555-1)Meets minimum force breakage based on tube diametersAll catheters met minimum force breakage based on tube diameters specified in ISO 10555-1.
    Liquid Leakage under Pressure/Leakage at Hub (ISO 10555-1)No leakage from hub or catheter body at specified pressuresNo leakage from hub or catheter body.
    Tip Stiffness (ASTM D747-10)Comparable to predicate devicesTip stiffness was comparable that of the predicate devices.
    Pressure vs. Flow CharacterizationFlow rates reported in Instructions for Use at 100 and 300 psiFlow rates reported in Instructions for Use at 1.00 and 300 psi. (Note: "1.00" might be a typo and likely meant 100 psi as per test description.)
    Static Burst Pressure (ISO 10555-1)Maximum peak pressures exceeded 300 psiMaximum peak pressures all exceeded 300 psi.
    Dynamic FlowNo leaks, breaks or occluded lumens at 750 psiNo leaks, breaks or occluded lumens at 750 psi.
    Flexibility Fatigue and ProfileNo signs of cracks or breakage post worst-case simulated useAll catheters showed no signs of cracks or breakage post worst case simulated use.
    Shape RetentionMaintain specified tip angle post shaping/conditioning/use; Tensile strength meets original specs after shapingAll catheters maintained specified tip angle after steam shaping, water bath conditioning and simulated use of guidewires. Tensile testing after tip shaping passed minimum tensile strength requirements.
    Torque to FailureNo signs of breakage, twists or collapsed lumens after specified number of torque turnsAll catheters showed no signs of breakage, twists or collapsed lumens after specified number of torque turns.
    Coating Lubricity and DurabilityMeets specified frictional forcesAll catheters met specified frictional forces.
    Coating IntegrityAcceptable coating coverage post simulated useAll samples showed acceptable coating coverage post simulated use.
    ParticulatesComparable numbers of particles to predicate in each size rangeThe test and predicate catheters had comparable numbers of particles in each size range.
    Chemical CompatibilityNo signs of degradation, corrosion, or physical decomposition after exposure to agentsAll catheters showed no signs of degradation, corrosion or physical decomposition.
    Latex Content (ASTM D6499-07)No detectable traces of latex foundNo detectable traces of latex found.
    Corrosion Resistance (ISO 10555-1)No signs of corrosion on metallic componentsNo signs of corrosion on metallic components of catheters.
    RadiopacityAcceptable radiopacity (evaluated by physicians)Both test and predicate catheters had acceptable radiopacity.
    MRI CompatibilityLabeled MRI Unsafe (as applicable for conducting/magnetic materials)Catheters are labeled MRI Unsafe on IFU.
    Simulated UseComparable Likert scores to predicate in Guidewire tracking and guide catheter movement; Interventional devices successfully deployedCatheters and predicate devices had comparable Likert scores in terms of Guidewire tracking and guide catheter movement. Interventional devices successfully deployed.
    Cadaver TestingComparable performance and access time to predicate in designated target; Ability to deploy/retrieve interventional devicesTest and predicate devices both exhibited comparable performance with similar access time to the designated target.
    Packaging Testing (ASTM F88-00, F2096-04)Minimum seal strength met, no bubbles under leak test conditionsAll sterile barrier pouches met minimum seal strength and showed no bubbles under leak test conditions.
    Shelf Life TestingContinued to meet specifications post accelerated aging and room temperature storageShelf life testing in progress. Expiration date will be advanced as aging data are available to demonstrate package and product continues to meet specifications.

    2. Sample Size Used for the Test Set and Data Provenance

    This document primarily reports on non-clinical bench and lab testing, and some limited biological/simulated use testing. The "test set" here refers to the number of devices or samples tested for each specific non-clinical performance attribute.

    • Sample Sizes: Not explicitly stated for each test, but implied by the nature of bench testing where multiple samples are typically used to ensure reliability. For example, "all catheters" met specifications implies multiple catheters were tested. For animal studies (thrombogenicity) and cadaver studies, the sample size would be small (e.g., number of dogs, cadavers). For Radiopacity and Simulated Use, "physicians," plural, suggests more than one device per observer.
    • Data Provenance: The data is primarily from bench-top laboratory testing and in vitro studies, in accordance with ISO and ASTM standards. Some tests involve animal models (e.g., thrombogenicity in dogs) and human cadavers. The context implies these studies were conducted by the manufacturer or their contracted labs. No specific country of origin for the data is mentioned, but the submission is to the U.S. FDA, and the company is US-based. This is all prospective testing, specifically designed to demonstrate compliance for the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Experts: The document mentions "physicians" evaluated radiopacity and participated in cadaver testing. For simulated use, "physicians" also provided Likert scores.
    • Number: Not explicitly stated how many physicians were involved in these evaluations. The plural "physicians" indicates more than one.
    • Qualifications: Not explicitly stated in terms of years of experience or specific subspecialty (e.g., interventional radiologist), but given the context of medical device evaluation, it is implied that these are qualified medical professionals with expertise relevant to microcatheters and their use.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not explicitly detailed. For tests involving "physicians" (Radiopacity, Simulated Use, Cadaver Testing), it is stated that comparisons were made or scores obtained. The document does not specify if there was a formal adjudication process (e.g., 2+1, 3+1), rather it presents the aggregate findings or an overall determination of "comparable" or "acceptable."

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. This type of study (often used for diagnostic imaging AI) is not applicable to a premarket notification for a physical medical device like a microcatheter. The comparison is primarily against established performance standards and a predicate device's existing characteristics. The "simulated use" and "cadaver testing" involve human readers/users (physicians) providing feedback, but this is to assess usability and performance in a simulated environment, not an MRMC study comparing AI with and without human assistance.
    • Effect Size of Human readers with/without AI: Not applicable, as this is not an AI device.

    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Not applicable. This submission is for a physical medical device (microcatheter), not an algorithm or AI product. The performance evaluations are intrinsic to the device itself and its interaction with biological and simulated environments.

    7. The Type of Ground Truth Used

    The concept of "ground truth" as it applies to diagnostic or AI studies is not directly applicable here. Instead, the basis for comparison and acceptance is:

    • Established Industry Standards: ISO 10993 (biocompatibility), ISO 10555 (intravascular catheters), ASTM standards.
    • FDA Guidance: FDA guidance for Short-Term and Long-Term Intravascular Catheters, FDA Special Controls Document for PTCA Catheters.
    • Predicate Device Performance: Direct comparison to the ev3 Marksman™ Catheter (K111490, K091559) and the Scientia Vascular, LLC Plato MICROCATH 27 (K121734). The goal is to demonstrate that the new device is "substantially equivalent" in safety and effectiveness.
    • Expert Observational Consensus: For aspects like radiopacity and simulated use feedback from physicians.

    8. The Sample Size for the Training Set

    • Not applicable. This is not a machine learning or AI device, so there is no "training set." The device design and materials are based on engineering principles and prior device knowledge.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set for an AI model, this question does not apply. The "ground truth" for the device's design and expected performance comes from established engineering principles, regulatory standards, and the performance characteristics of predicate devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K121734

    Validate with FDA (Live)

    Date Cleared
    2012-07-30

    (47 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Plato MICROCATH® 27 Microcatheters are intended to assist in the delivery of diagnostic and/or therapeutic agents to the peripheral and neurovascular systems.

    Device Description

    The Plato MICROCATH® 27 Microcatheters are single lumen catheters designed to be used with a guiding catheter and a steerable guidewire for accessing the vasculature. The microcatheter contains a metal hypotube reinforced shaft with a maximum diameter of 3.7F, tapering to a distal end of 3.3F, which can be inserted into a 6F guide catheter. The inner diameter is constant throughout the shaft length and accommodates up to a 0.025" guidewire. The catheter is 150 cm in length with two tip configurations; straight and 45°. All have steam shapeable tips, hydrophilic coating and a radiopaque marker at the distal tip. The devices are marketed with an accessory mandrel which allows the catheter tips to be steam-shaped to the doctor's preferred shape when desired.

    AI/ML Overview

    The provided text describes the "SCIENTIA VASCULAR LLC PLATO MICROCATH® 27 MICROCATHETER". It does not describe a study involving acceptance criteria related to an AI/ML device, but rather the non-clinical performance data for a microcatheter device. Therefore, I will extract information relevant to the device's performance against non-clinical acceptance criteria.

    Here's the breakdown of the information as per the request, adapted for a non-AI/ML medical device:

    1. A table of acceptance criteria and the reported device performance

    The document lists "Non-Clinical Performance Data" which served as the basis for demonstrating substantial equivalence. These can be considered the acceptance criteria for functional and safety aspects of the device.

    Acceptance Criteria (Test Performed)Reported Device Performance
    Visual/Dimensional InspectionConcluded to function as intended (implied pass)
    Air Ingress/Negative CollapseConcluded to function as intended (implied pass)
    Kink ResistanceConcluded to function as intended (implied pass)
    Tensile Strength/ElongationConcluded to function as intended (implied pass)
    Liquid Leakage under Pressure/Leakage at HubConcluded to function as intended (implied pass)
    Tip StiffnessConcluded to function as intended (implied pass)
    Pressure vs. Flow CharacterizationConcluded to function as intended (implied pass)
    Static and Dynamic Burst PressureConcluded to function as intended (implied pass)
    Flex FatigueConcluded to function as intended (implied pass)
    Shape RetentionConcluded to function as intended (implied pass)
    Coating Lubricity and DurabilityConcluded to function as intended (implied pass)
    Coating IntegrityConcluded to function as intended (implied pass)
    ParticulatesConcluded to function as intended (implied pass)
    Chemical compatibilityConcluded to function as intended (implied pass)
    LatexConcluded to function as intended (implied pass)
    Corrosion ResistanceConcluded to function as intended (implied pass)
    Biocompatibility (ISO 10993)Testing performed and results supported substantial equivalence
    Functional Testing (ISO 10555-1)Met the requirements of ISO 10555-1
    Simulated Use (Anatomical Model)Navigated successfully, guidewire tracking and movement tested (implied pass)
    Human Cadaver Model PerformanceComparable to the predicate for advancement to the intended vascular site

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size:
      • For the non-clinical bench testing, specific sample sizes are not provided in the document. The tests likely involved multiple units of the Plato MICROCATH 27 Microcatheter.
      • For the simulated use testing, an "anatomical model" was used. The number of models or tests conducted is not specified.
      • For the human cadaver model, the term "a human cadaver model" is used, implying at least one but the exact number of cadavers or trials is not specified.
    • Data Provenance: The tests are explicitly described as "Non-Clinical Performance Data" and include "bench testing," "simulated-use testing," and testing in a "human cadaver model." This indicates the data is from controlled laboratory and simulated environments. The country of origin of the testing is not specified, but the submission is to the FDA (United States). This testing would be considered prospective, as it was conducted specifically to evaluate the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable as the document describes non-clinical and simulated-use testing of a physical medical device, not an AI/ML system requiring expert ground truth for interpretation. The evaluation was based on engineering specifications, standardized test methods (ISO 10993, ISO 10555-1), and performance in physical models.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. The device's performance was evaluated against physical and functional criteria, not expert interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The document describes a physical medical device (microcatheter), not an AI-assisted diagnostic or therapeutic system that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This document is about a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance was established through:

    • Compliance with international standards (ISO 10993, ISO 10555-1).
    • Demonstration of expected functional performance in various bench tests (e.g., kink resistance, tensile strength, burst pressure, coating lubricity).
    • Comparative performance to a predicate device in simulated anatomical and human cadaver models for key operational aspects like maneuverability and advancement.

    8. The sample size for the training set

    This is not applicable. This document describes a physical medical device, not an AI/ML system that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable. This document describes a physical medical device, not an AI/ML system.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1