510(k) Data Aggregation

    K Number
    K983107

    Validate with FDA (Live)

    Device Name
    VFE
    Date Cleared
    1998-10-22

    (48 days)

    Product Code
    Regulation Number
    864.7290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VFE assay is an in vitro enzyme immunoassay for quantifying the concentration of von Willebrand Factor Antigen in human plasma to aid in differentiating patients with von Willebrand disease from those with Classical Hemophilia A.

    Device Description

    VFE is an enzyme immunoassay, in microplate format, intended for the in vitro quantification of von Willebrand Factor antigen in serum or plasma as an aid to differentiate patients with von Willebrand disease from those with classical Hemophilia A. This assay represents an updated technical format, with performance improvements, when compared to the predicate device.

    AI/ML Overview

    Here's an analysis of the provided text to extract the acceptance criteria and study information:

    Device: VFE (von Willebrand Factor Antigen Assay)

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" but rather presents a comparison to a predicate device (SPECTRO vWF) to demonstrate "substantial equivalence." The performance parameters of the VFE are compared to those of the predicate. We can infer that the VFE's performance should be comparable to or better than the predicate's for "substantial equivalence."

    ParameterAcceptance Criteria (Inferred from Predicate)Reported Device Performance (VFE)
    SensitivityComparable to 0.23% N0.78% N
    Intra-assay reproducibility - CVComparable to 6.2 - 8.05.0 - 5.8
    Inter-assay reproducibility - CVComparable to 6.3 - 12.87.0 - 11.5
    Accuracy - R (compared to predicate test results)N/A (predicate compared to Laurell Rocket)0.849 (n=118)

    Note on Sensitivity: The VFE's reported sensitivity (0.78% N) is higher than SPECTRO vWF's (0.23% N). In this context, a higher percentage might indicate a detection limit, where a lower percentage is generally better for sensitivity in an assay (meaning it can detect lower concentrations). However, without further context on the specific units and definition of "% N," it's difficult to definitively say if VFE is "better" or if it represents a different measure. The conclusion statement indicates "similar characteristics," suggesting this difference was not considered a deal-breaker for substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Accuracy (compared to predicate): 118 samples (n = 118)
    • Data Provenance: Not explicitly stated. The comparison is made to a predicate test, "SPECTRO vWF," which is also from Ramco Laboratories, Inc. No information regarding the country of origin or whether the data was retrospective or prospective is provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This information is not provided in the document. The accuracy of the VFE is compared to results from the predicate device, and the predicate's accuracy was "original comparison versus Laurell Rocket electrophoresis results." There is no mention of experts establishing a ground truth.

    4. Adjudication Method for the Test Set

    • This information is not provided. The comparison is statistical (R value) against a predicate device's results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • No. This device is an in vitro diagnostic (IVD) enzyme immunoassay, not an AI-powered diagnostic imaging device. Therefore, an MRMC study with human readers and AI assistance is not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, implicitly. The performance data presented (sensitivity, reproducibility, accuracy) represents the standalone performance of the VFE assay itself, without human interpretation in the sense of a "human-in-the-loop" for image or complex data analysis. The assay produces a quantitative result.

    7. The Type of Ground Truth Used

    • For the VFE's accuracy, the ground truth was the results obtained from the predicate device (SPECTRO vWF).
    • For the predicate device's original accuracy, the ground truth was derived from Laurell Rocket electrophoresis results.

    8. The Sample Size for the Training Set

    • This information is not applicable/not provided. The VFE is an enzyme immunoassay, not a machine learning model that requires a "training set" in the conventional sense. The "training" of an assay involves optimization during development, but specific sample sizes for such an activity are not documented in the same way as for AI models.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable/not provided. As mentioned above, this is an immunoassay, not an AI model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K981208

    Validate with FDA (Live)

    Device Name
    TFR
    Date Cleared
    1998-10-21

    (202 days)

    Product Code
    Regulation Number
    866.5880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TFR assay is an in vitro enzyme immunoassay for quantifying the concentration of transferrin receptor in human serum or plasma to aid in the diagnosis of iron deficiency anemia, particularly in the presence of other disease states.

    Device Description

    The TFR assay is an in vitro enzyme immunoassay based upon the double antibody sandwich method. Plasma or serum samples are diluted in buffer and pipetted into microwells pre-coated with polyclonal antibody to transferrin receptor. Horseradish peroxidase conjugated antibody specific for serum transferrin receptor (STR) is added to the wells and incubated. During this incubation, the STR binds to the polyclonal antibodies adsorbed to the wells and the HRP-conjugated second antibodies bind to the captured STR. Any unbound STR and excess HRP-conjugate are washed from the wells. Enzyme substrate is added to the wells and allowed to incubate, a stop solution is then added to stop the reaction and the intensity of the yellow product is measured in a microplate reader. The optical density of the resulting solution is directly proportional to the concentration of the STR in the standard samples. A standard curve is generated from the STR standards provided in the assay and the concentration of STR in the unknown sample is determined by comparing the unknown's optical density reading with the standard curve graph.

    AI/ML Overview

    The provided text describes a 510(k) submission for the TFR assay, focusing on its substantial equivalence to a predicate device (Quantikine IVD sTfR Immunoassay or QSI). The study primarily demonstrates agreement between the TFR assay and the predicate device, as well as agreement with clinical definitions of anemia.

    Here's an analysis of the acceptance criteria and study aspects based on the given information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in numerical thresholds, but rather implied by the comparison to the predicate device and clinical definitions. The 'agreement' percentages serve as the reported device performance against these implied criteria.

    Acceptance Criteria (Implied)Reported Device Performance (TfR vs. Clinical/Predicate)
    Agreement with clinical definition of IDA (TfR > 8.3 ug/ml)78.9% (30 out of 38 IDA samples)
    Agreement with clinical definition of ACD (TfR < 8.3 ug/ml)82.9% (145 out of 175 ACD samples)
    Agreement with individual QSI results (all 155 samples)90.3% (140 out of 155 times)
    Agreement with individual QSI results (IDA, QSI & ferritin concur)100.0% (17 out of 17 times)
    Agreement with individual QSI results (ACD, QSI & ferritin concur)98.8% (79 out of 80 times)
    Agreement with individual QSI results (Neither IDA nor ACD)92.3% (12 out of 13 times)
    Linear regression (TfR vs. QSI for all 155 samples)R² of 88.2%
    Overall agreement with QSI when QSI and ferritin concur98.97% (96 out of 97 times)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Size:
      • TFR results: 283 total samples (38 IDA, 175 ACD, 70 "Neither IDA nor ACD")
      • QSI results: 155 total samples (23 IDA, 119 ACD, 13 "Neither IDA nor ACD"). This subset of 155 samples had both TFR and QSI results.
    • Data Provenance:
      • Origin: Not explicitly stated, but implies human serum or plasma samples. No country of origin is mentioned.
      • Nature: Retrospective. The samples were collected "Over an 18 month period of time," suggesting they were existing samples at the time of the study's design.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • The ground truth (clinical definitions of IDA and ACD) was established based on specific laboratory values (Serum Iron, Iron Binding Capacity, Hemoglobin).
    • Number of Experts: Not specified. The clinical definitions were provided as existing criteria. It is implied that medical professionals established and utilized these definitions, but no specific number or qualifications are given for any "experts" in the context of the study.

    4. Adjudication Method for the Test Set

    • None directly specified. The clinical definitions were used as a direct "ground truth" for classifying samples into IDA, ACD, or "Neither IDA nor ACD." For comparisons between TFR and QSI, direct result comparisons were made, sometimes with the additional filter of ferritin results. This is not an adjudication process as typically understood in clinical trials with multiple human readers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    • No, an MRMC study was NOT done. This study is for an in vitro diagnostic (IVD) assay, not a medical imaging device involving human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, in essence, this is a standalone performance study. The TFR assay is an in vitro diagnostic (IVD) test that quantitatively measures transferrin receptor concentration. Its performance is evaluated based on its output (concentration values) compared to clinical definitions and the predicate device's output. There is no human-in-the-loop component for the assay's function itself.

    7. The Type of Ground Truth Used

    • The ground truth was established by clinical definitions based on specific laboratory parameters (Serum Iron, Iron Binding Capacity, Hemoglobin) for IDA and ACD. For some comparisons, concurrence with ferritin results (from another test) was also used to refine the population. The predicate device's results (QSI) also served as a comparative "ground truth" for assessing agreement.

    8. The Sample Size for the Training Set

    • Not applicable / not mentioned. This document describes a 510(k) submission for a diagnostic assay, not a machine learning or AI algorithm that typically involves a training set. The assay's performance is based on its chemical/enzymatic reactions.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable / not mentioned. As this is not an AI/ML algorithm, there is no training set in the conventional sense. The assay's "calibration" would be established through the use of standards provided within the assay kit, as mentioned ("A standard curve is generated from the STR standards provided in the assay").