K Number
K981208

Validate with FDA (Live)

Device Name
TFR
Date Cleared
1998-10-21

(202 days)

Product Code
Regulation Number
866.5880
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TFR assay is an in vitro enzyme immunoassay for quantifying the concentration of transferrin receptor in human serum or plasma to aid in the diagnosis of iron deficiency anemia, particularly in the presence of other disease states.

Device Description

The TFR assay is an in vitro enzyme immunoassay based upon the double antibody sandwich method. Plasma or serum samples are diluted in buffer and pipetted into microwells pre-coated with polyclonal antibody to transferrin receptor. Horseradish peroxidase conjugated antibody specific for serum transferrin receptor (STR) is added to the wells and incubated. During this incubation, the STR binds to the polyclonal antibodies adsorbed to the wells and the HRP-conjugated second antibodies bind to the captured STR. Any unbound STR and excess HRP-conjugate are washed from the wells. Enzyme substrate is added to the wells and allowed to incubate, a stop solution is then added to stop the reaction and the intensity of the yellow product is measured in a microplate reader. The optical density of the resulting solution is directly proportional to the concentration of the STR in the standard samples. A standard curve is generated from the STR standards provided in the assay and the concentration of STR in the unknown sample is determined by comparing the unknown's optical density reading with the standard curve graph.

AI/ML Overview

The provided text describes a 510(k) submission for the TFR assay, focusing on its substantial equivalence to a predicate device (Quantikine IVD sTfR Immunoassay or QSI). The study primarily demonstrates agreement between the TFR assay and the predicate device, as well as agreement with clinical definitions of anemia.

Here's an analysis of the acceptance criteria and study aspects based on the given information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated in numerical thresholds, but rather implied by the comparison to the predicate device and clinical definitions. The 'agreement' percentages serve as the reported device performance against these implied criteria.

Acceptance Criteria (Implied)Reported Device Performance (TfR vs. Clinical/Predicate)
Agreement with clinical definition of IDA (TfR > 8.3 ug/ml)78.9% (30 out of 38 IDA samples)
Agreement with clinical definition of ACD (TfR < 8.3 ug/ml)82.9% (145 out of 175 ACD samples)
Agreement with individual QSI results (all 155 samples)90.3% (140 out of 155 times)
Agreement with individual QSI results (IDA, QSI & ferritin concur)100.0% (17 out of 17 times)
Agreement with individual QSI results (ACD, QSI & ferritin concur)98.8% (79 out of 80 times)
Agreement with individual QSI results (Neither IDA nor ACD)92.3% (12 out of 13 times)
Linear regression (TfR vs. QSI for all 155 samples)R² of 88.2%
Overall agreement with QSI when QSI and ferritin concur98.97% (96 out of 97 times)

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Size:
    • TFR results: 283 total samples (38 IDA, 175 ACD, 70 "Neither IDA nor ACD")
    • QSI results: 155 total samples (23 IDA, 119 ACD, 13 "Neither IDA nor ACD"). This subset of 155 samples had both TFR and QSI results.
  • Data Provenance:
    • Origin: Not explicitly stated, but implies human serum or plasma samples. No country of origin is mentioned.
    • Nature: Retrospective. The samples were collected "Over an 18 month period of time," suggesting they were existing samples at the time of the study's design.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The ground truth (clinical definitions of IDA and ACD) was established based on specific laboratory values (Serum Iron, Iron Binding Capacity, Hemoglobin).
  • Number of Experts: Not specified. The clinical definitions were provided as existing criteria. It is implied that medical professionals established and utilized these definitions, but no specific number or qualifications are given for any "experts" in the context of the study.

4. Adjudication Method for the Test Set

  • None directly specified. The clinical definitions were used as a direct "ground truth" for classifying samples into IDA, ACD, or "Neither IDA nor ACD." For comparisons between TFR and QSI, direct result comparisons were made, sometimes with the additional filter of ferritin results. This is not an adjudication process as typically understood in clinical trials with multiple human readers.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

  • No, an MRMC study was NOT done. This study is for an in vitro diagnostic (IVD) assay, not a medical imaging device involving human readers or AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Yes, in essence, this is a standalone performance study. The TFR assay is an in vitro diagnostic (IVD) test that quantitatively measures transferrin receptor concentration. Its performance is evaluated based on its output (concentration values) compared to clinical definitions and the predicate device's output. There is no human-in-the-loop component for the assay's function itself.

7. The Type of Ground Truth Used

  • The ground truth was established by clinical definitions based on specific laboratory parameters (Serum Iron, Iron Binding Capacity, Hemoglobin) for IDA and ACD. For some comparisons, concurrence with ferritin results (from another test) was also used to refine the population. The predicate device's results (QSI) also served as a comparative "ground truth" for assessing agreement.

8. The Sample Size for the Training Set

  • Not applicable / not mentioned. This document describes a 510(k) submission for a diagnostic assay, not a machine learning or AI algorithm that typically involves a training set. The assay's performance is based on its chemical/enzymatic reactions.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable / not mentioned. As this is not an AI/ML algorithm, there is no training set in the conventional sense. The assay's "calibration" would be established through the use of standards provided within the assay kit, as mentioned ("A standard curve is generated from the STR standards provided in the assay").

{0}------------------------------------------------

10/21/98

Image /page/0/Picture/1 description: The image shows the logo and contact information for Ramco Laboratories Inc. The address is 4507 Mt. Vernon, Houston, Texas 77006. The phone number is (713) 526-9677, and the toll-free number is 1-800-231-6238. The fax number is 713-526-1528, and they are available 24 hours.

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

The assigned 510(k) number is: __ 981208

Submitter:Ramco Laboratories, Inc.4507 Mt. VernonHouston, TX 77006
Phone #:Fax #:(713) 526-9677(713) 526-1528
Contact Person:Jeffrey B. Grubb, President
Date Prepared:March 27, 1998
Device Name:Transferrin Receptor Assay
Trade Name:TfR
Common Name:TfR Assay
Classification Name:sTfR Immunological Test System
Predicate Device Name:Quantikine IVD sTfR ImmunoassayR&D Systems, Inc.614 McKinley Place N.E.Minneapolis, MN 55413510(k) #K970718

The TFR assay is an in vitro enzyme immunoassay based upon the double antibody sandwich method. Plasma or serum samples are diluted in buffer and pipetted into microwells pre-coated with polyclonal antibody to transferrin receptor. Horseradish peroxidase conjugated antibody specific for serum transferrin receptor (STR) is added to the wells and incubated. During this incubation, the STR binds to the polyclonal antibodies adsorbed to the wells and the HRP-conjugated second antibodies bind to the captured STR.

{1}------------------------------------------------

Any unbound STR and excess HRP-conjugate are washed from the wells. Enzyme substrate is added to the wells and allowed to incubate, a stop solution is then added to stop the reaction and the intensity of the yellow product is measured in a microplate reader. The optical density of the resulting solution is directly proportional to the concentration of the STR in the standard samples. A standard curve is generated from the STR standards provided in the assay and the concentration of STR in the unknown sample is determined by comparing the unknown's optical density reading with the standard curve graph.

The TFR assay is an in vitro enzyme immunoassay for quantifying the concentration of transferrin receptor in human serum or plasma to aid in the diagnosis of iron deficiency anemia, particularly in the presence of other disease states.

Ramco Laboratories, Inc. is claiming substantial equivalence of its TFR assay the R&D Systems (Minneapolis, Minnesota) Quantikine IVD sTfR Immunoassay ("QSI") which obtained 510(k) clearance on May 27, 1997 (K970718, Class II).

Both TFR and OSI's intended use is to quantitatively determine the level of transferrin receptor in human serum or plasma to aid in the diagnosis of iron deficiency anemia. Both are based on the microplate sandwich enzyme immunoassay technique, use calibrators to generate a log-log curve from which the quantity of an unknown sample is read, use a monoclonal transferrin antibody conjugated to horseradish peroxidase, use a microplate reader with a 450nm filter to read the results, and supply controls which, like the unknown samples, must be diluted prior to use. Finally, both assays offer a prescribed normal range, the upper limit of which is the indicator for the presence of iron deficiency anemia.

Both TFR and OSI use a monoclonal antibody in its conjugate. However, TFR uses a polyclonal transferrin receptor antibody to coat the microplate, making it a polyclonalmonoclonal assay, while QSI uses a monoclonal antibody to coat the plate making it a double monoclonal. TFR provides 7 standards consisting of TFR, isolated from human placenta, contained in a phosphate buffer with BSA and normal rabbit serum and calibrated in ne/ml while QSI provides 6 standards, consisting of human serum transferrin receptor, contained in buffered animal serum and calibrated in nmol/L. The assay results are reported in ug/ml in the TFR assay compared to nmol/L for OSI. Samples and controls are diluted 100:1 prior to use in the 7fR assay compared to 50:1 for QSI. Finally, TFR is a single-stage compared to QSI which is a two-stage assay.

Over an 18 month period of time, patient samples were collected. These samples were segregated into one of three categories: 1) Anemia of Chronic Disease ("ACD"), clinically defined as Serum Iron < 60, Iron Binding Capacity < 350 and Hemoglobin < 12; 2) Iron Deficiency Anemia ("IDA"), clinically defined as Serum Iron < 60, Iron Binding Capacity > 350 and Hemoglobin < 12; and 3) neither IDA nor ACD, samples that had either Serum Iron > 60 or Hemoglobin > 12.

Of the above samples, TFR results were obtained for 38 of the IDA, 175 of the ACD, and 70 of the "Neither IDA nor ACD" samples (283 total), while QSI results were obtained for 23 of the IDA, 119 of the IDA and 13 of the "Neither IDA nor ACD" samples (155 total).

{2}------------------------------------------------

Performing a linear regression of all values for the 155 samples for which a TFR and OSI value has been obtained yields an R2 of 88.2% (See Table 17).

Of the 38 IDA samples, QSI results were obtained for 23. IffR is in agreement (>8.3ug/ml) with the clinical definition of IDA 30 out of 38 times or 78.9% while QSI is in agreement with the clinical definition 18 out of 23 times or 78.3%. Of these 23 samples, TfR delivered the same individual result as OSI for 22 out of 23 samples.

The IDA population was further refined to those samples where QSI and ferritin results concurred (STR > 28.1 and ferritin < 30). This reduced the population to 17 samples. Of these 17 samples, TFR concurred with the individual QSI results 17 out of 17 times or 100.0% .

Of the 175 ACD samples, QSI results were obtained for 119. TfR is in agreement ( < 8.3 ug/ml) with the clinical definition of ACD 145 out of 175 times or 82.9% while OSI is in agreement with the clinical definition 81 out of 119 times or 68.1%. Of these 119 samples, TfR delivered the same individual result as OSI.

The ACD population was further refined to those samples where OSI and ferritin results concurred (STR < 28.1 and ferritin > 30). This reduced the population to 80 samples. Of these 80 samples, TFR concurred with the individual OSI results 79 out of 80 times or 98.8%.

Of the 70 "Neither IDA nor ACD" samples. OSI results were obtained for 13. Of these 13 samples, TFR delivered the same individual result as OSI 12 out of 13 times or 92.3%

When all samples results are taken as a whole, TFR delivered the same individual result as QSI 140 out of 155 times or 90.3%. When only those cases are used where OSI and ferritin are in agreement. TFR and OSI give the same result 96 out of 97 times or 98.97%.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

OCT 27 1998

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Jeffrey B. Grubb President Ramco Laboratories, Inc. 4507 Mt. Vernon 77006 Houston, Texas

  • Re: K981208/S1 Trade Name: ની ને જ II Requlatory Class: Product Code: JNM Dated: July 29, 1998 Received: July 31, 1998
    Dear Mr. Grubb:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

Image /page/3/Picture/9 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The eagle is composed of three curved lines that form the shape of the bird's body and wings. The text is in all capital letters and is evenly spaced around the circle.

{4}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial in your 510 (x) promodevice to a legally marketed predicate device equiration or your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro regaractic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and 394 1500. That would be please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, (301) 391 1009. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . be obtained from the Division of Small Manufacturers Assistance at its be extained more (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number: K981208

Device Name: IfR

The IJR assay is an in vitro enzyme immunoassay for quantifying Indications for Use: The 1918 assay is an in weeptor in human serum or plasina to aid in the diagnosis of iron deficiency anemia, particularly in the presence of other disease states.

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) NumberK981208
Prescription Use

OR

Over-The-Counter Use_____
-----------------------------

§ 866.5880 Transferrin immunological test system.

(a)
Identification. A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in serum, plasma, and other body fluids. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.(b)
Classification. Class II (performance standards).