Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K033726

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-01-16

    (49 days)

    Product Code
    Regulation Number
    870.1210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PVS "Pivot™" Steerable Delivery System® Microcatheter is intended to be used to access tortuous vasculature for sub-selective controlled infusion or delivery of diagnostic, embolic, and therapeutic agents into the distal, peripheral, coronary, and neurovasculature, and for guide wire exchange/support during diagnostic or interventional procedures.

    Device Description

    The PVS "Pivot™" Steerable Delivery System® Microcatheter is a 1.9F/2.4F (nominal distal/proximal) tubular device, ~150 cm in length, with a lumen to be used for delivery of contrast, drugs, or embolics. The lumen is constructed from a polymeric material and has an inside diameter of 0.0165". The device is coated on the outer diameter with a lubricious coating over the distal segment of the device. Two radiopaque markers are positioned at the distal tip of the device to aid visualization under fluoroscopy. The proximal end of the device has a standard luer adapter for attachment of accessories and can be used to flush the lumen. The subject device has the ability to access distal, tortuous vasculature over a guide wire, deliver embolics and agents, and has the ability to be steered like a guide wire as needed.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the PVS "Pivot™" Steerable Delivery System® Microcatheter:

    Acceptance Criteria and Device Performance

    The submission does not explicitly list quantitative "acceptance criteria" in the format of a requirement and a corresponding measured value for the individual performance tests. Instead, it states that the tests "yielded acceptable results" or were "verified in accordance with" standards. The acceptance criteria for this device are implicitly tied to demonstrating substantial equivalence to predicate devices and meeting established industry standards.

    Acceptance Criteria (Implied)Reported Device Performance
    Adherence to ISO 10993-1 for Biocompatibility"Biocompatibility of the PVS 'Pivot™' Steerable Delivery System® Microcatheter has been verified in accordance with ISO 10993-1... Test results confirmed biocompatibility of the subject device when tested as an external communicating, blood contact, short duration (<24 hours) devices."
    Adherence to ISO 10555-1 for Sterile, Single-Use Intravascular Catheters"Performance testing of the PVS 'Pivot™' Steerable Delivery System® Microcatheter was conducted in accordance with ISO 10555-1... Verification testing for the subject device included dimensional inspection, hub integrity, flow rate measurements, burst strength, tensile strength, guidewire compatibility testing and performance under simulated conditions. Subject product testing is believed to have yielded acceptable results."
    Acceptable Torsional Strength"In addition, torsional strength... tests also yielded acceptable results."
    Acceptable Torqueability"...torqueability... tests also yielded acceptable results."
    Acceptable Corrosion Resistance"...corrosion resistance tests also yielded acceptable results."
    Substantial Equivalence to Predicate Devices"Based on the indications for use, technological characteristics, and safety and performance testing, the subject PVS 'Pivot™' Steerable Delivery System® Microcatheter... is substantially equivalent in design, materials, sterilization, principles of operation and indications for use to current commercially available catheters/cited predicates."

    Study Details:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
      The document does not specify the exact sample sizes (number of devices) used for each individual performance test (e.g., dimensional inspection, burst strength, tensile strength). The data provenance is not mentioned, as these are in vitro engineering tests rather than clinical data from patients. The tests are prospective in the sense that they were conducted specifically for this submission.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
      This is not applicable as the studies described are in vitro performance and biocompatibility tests on the device itself, not clinical studies involving "ground truth" derived from expert consensus on patient cases.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      Not applicable, as there were no patient cases or expert readings requiring adjudication.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      No MRMC study was performed. This document describes a medical device (a microcatheter), not an AI algorithm.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Not applicable, as this is a physical medical device, not an algorithm.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
      Not applicable. The "ground truth" for the device's performance is derived from established engineering test methods and their defined pass/fail criteria according to ISO standards and internal specifications, rather than clinical outcomes or expert consensus on patient data.

    7. The sample size for the training set:
      Not applicable, as this is a physical medical device and does not involve AI or machine learning algorithms that require a "training set."

    8. How the ground truth for the training set was established:
      Not applicable, for the same reason as point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032494

    Validate with FDA (Live)

    Date Cleared
    2003-09-05

    (23 days)

    Product Code
    Regulation Number
    870.1210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PVS 1700 Steerable Delivery System®-10 Microcatheter is intended to be used to access tortuous vasculature for sub-selective controlled infusion or delivery of diagnostic, embolic, and therapeutic agents into the distal, peripheral, coronary, and neurovasculature, and for guide wire exchange/support during diagnostic or interventional procedures.

    Device Description

    The PVS 1700 Steerable Delivery System®-10 is a 1.9F/2.4F (nominal distal/proximal) tubular device, 155 cm in length, with a lumen to be used for delivery of contrast, drugs, or embolics. The lumen is constructed from a polymeric material and has an inside diameter of 0.0165". The device is coated on the outer diameter with a lubricious coating over the distal segment of the device. Two radiopaque markers are positioned at the distal tip of the device to aid visualization under fluoroscopy. The proximal end of the device has a standard luer adapter for attachment of accessories and can be used to flush the lumen. The subject device has the ability to access distal, tortuous vasculature over a guide wire, deliver embolics and agents, and has the ability to be steered like a guide wire as needed.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device (PVS 1700 Steerable Delivery System®-10 Microcatheter), which focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed clinical study results for novel acceptance criteria.

    Therefore, many of the requested details about specific acceptance criteria from a clinical study, sample size, expert ground truth establishment, and MRMC studies are not applicable or available in this type of regulatory submission. The device is primarily evaluated through bench testing and comparison to existing, cleared devices.

    Here's an analysis of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a quantitative, pass/fail manner for each test, nor does it provide detailed performance numbers in a table. Instead, it generally states that tests "yielded acceptable results" or "confirmed biocompatibility."

    Acceptance Criteria (Implied)Reported Device Performance
    Biocompatibility: Meet ISO 10993-1 standards for an external communicating, blood contact, short duration (<24 hours) device."Test results confirmed biocompatibility of the subject device when tested as an external communicating, blood contact, short duration (<24 hours) devices."
    Dimensional Inspection: Conform to specified dimensions."Verification testing for the subject device included dimensional inspection… Subject product testing is believed to have yielded acceptable results." (No specific dimensions or ranges reported, but implied conformance to design specifications).
    Hub Integrity: Maintain structural integrity of the hub."Verification testing for the subject device included… hub integrity… Subject product testing is believed to have yielded acceptable results." (No specific integrity metrics reported).
    Flow Rate Measurements: Achieve specified flow rates."Verification testing for the subject device included… flow rate measurements… Subject product testing is believed to have yielded acceptable results." (No specific flow rates reported).
    Burst Strength: Withstand internal pressure without bursting."Verification testing for the subject device included… burst strength… Subject product testing is believed to have yielded acceptable results." (No specific burst pressure reported).
    Tensile Strength: Withstand pulling forces without breaking."Verification testing for the subject device included… tensile strength… Subject product testing is believed to have yielded acceptable results." (No specific tensile strength reported).
    Guidewire Compatibility: Function correctly with guidewires."Verification testing for the subject device included… guidewire compatibility testing and performance under simulated conditions. Subject product testing is believed to have yielded acceptable results." (No specific compatibility metrics reported).
    Torsional Strength: Withstand twisting forces."In addition, torsional strength… tests also yielded acceptable results." (No specific torsional strength reported).
    Torqueability: Ability to transmit torque for steering."In addition, …torqueability… tests also yielded acceptable results." (No specific torqueability metrics reported. The device is described as having "the ability to be steered like a guide wire as needed," which implies this aspect was evaluated and found acceptable).
    Corrosion Resistance: Resist degradation from corrosion."In addition… corrosion resistance tests also yielded acceptable results." (No specific corrosion resistance metrics reported).
    Substantial Equivalence: Similar in design, materials, sterilization, principles of operation, and indications for use to predicate devices, raising no new questions of safety and efficacy."Technological similarities between the PVS 1700 Steerable Delivery System®-10 and predicate devices include the basal tubular design and dimensions, polymeric materials and construction, and hydrophilic coating. In instances where the technological characteristics may differ slightly, it has been demonstrated that there are no new questions raised regarding safety and efficacy of the PVS 1700 Steerable Delivery System®-10." "Based on the indications for use, technological characteristics, and safety and performance testing, the subject PVS 1700 Steerable Delivery System®-10 meets the minimum requirements that are considered adequate for its intended use and is substantially equivalent in design, materials, sterilization, principles of operation and indications for use to current commercially available catheters/cited predicates."

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify a "test set" in the context of patient data for clinical studies, as the evaluation primarily involved bench testing (performance testing) and biocompatibility testing. Therefore, there is no information on data provenance (country of origin, retrospective/prospective).

    For bench tests, typically multiple units of the device are tested, but the exact sample sizes for each specific test (e.g., how many catheters were burst tested) are not provided in this summary.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. The ground truth for this device's evaluation (bench testing) is based on engineering specifications, material science standards (ISO), and direct physical measurements/observations, not on expert consensus from clinical cases.

    4. Adjudication Method for the Test Set

    Not applicable, as this device was evaluated via bench testing rather than through a clinical study requiring human interpretation or adjudication of patient cases.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No. This type of study is not relevant for the regulatory pathway (510(k) for a microcatheter) and the type of evidence presented (bench testing for substantial equivalence).

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical microcatheter, not an AI algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for proving the device meets acceptance criteria is primarily:

    • Engineering specifications and design requirements: For dimensional inspection, flow rate, tensile strength, burst strength, torqueability, torsional strength, and guidewire compatibility.
    • International Standards (ISO 10993-1 and ISO 10555-1): For biocompatibility and performance testing of intravascular catheters.
    • Comparison to predicate devices: To establish substantial equivalence in terms of technological characteristics and indications for use.

    8. The Sample Size for the Training Set

    Not applicable. This isn't an AI/machine learning device, so there is no concept of a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032146

    Validate with FDA (Live)

    Date Cleared
    2003-08-12

    (29 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PVS 1300 Synchro® 0.014" & 1600 Synchro® 0.010" Neuro Guidewire series of products are intended for general intravascular use, which includes peripheral and neurovascular use. The device can be used to selectively introduce and position catheters and other interventional devices within the peripheral & neurovasculature. The device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.

    Device Description

    products having a 0.014" & 0.010" outside diameter respectively, being sterile, single use/disposable products, with shapeable tip which is used to gain intravascular access to and facilitate the positioning and exchange of interventional devices in small diameter, tortuous vasculature for peripheral, neuro diagnostic and interventional procedures. The guidewire can be torqued to facilitate navigation through the vasculature. A torque device, (Merit Medical Systems (K936032)) is supplied with the wire to facilitate deployment & positioning. A guidewire introducer (B. Braun (K760389)) is also supplied and may be used to aid introduction of the guidewire into the catheter hub and/or hemostasis valve and to gently shape the guidewire's distal flexible tip, if desired, according to standard practice. Neither the guidewire introducer or the torque device are intended to enter the body. The products are projected to be provided in a 180cm - 300cm length range, with 200cm being nominal. The Nitinol tip length is projected to be presented in a 25cm - 65cm range, with 35-45cm being nominal. A traditional range of flexibility profiles shall also be provided, ranging from support (stiff) to flex (soft). The device is coated on the outer diameter with a lubricious coating over the distal segment of the device. The marker coil is platinum wire at the distal tip of the device to aid visualization under fluoroscopy. The subject device has the ability to access distal, tortuous vasculature, with steerability and torque transmission properties.

    AI/ML Overview

    The provided text describes a 510(k) submission for guidewires, which are physical medical devices, not software or AI-powered devices. Therefore, the questions related to AI/algorithm performance metrics, expert reviews, ground truth, and training sets are not applicable to this document. The submission focuses on demonstrating substantial equivalence to predicate guidewires through material properties and performance testing.

    Here's the relevant information that can be extracted from the provided text, structured to best fit your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    BiocompatibilityVerified in accordance with ISO 10993-1 (Biological Evaluation of Medical Devices - Part 1). Materials test results confirmed biocompatibility as an external communicating, blood contact, short duration (<24 hours) device.
    Dimensional InspectionYielded acceptable safety & performance outcomes (as part of overall performance testing). Specific criteria not detailed in this summary.
    Fatigue AssessmentYielded acceptable safety & performance outcomes (as part of overall performance testing). Specific criteria not detailed in this summary.
    Tip FlexibilityYielded acceptable safety & performance outcomes (as part of overall performance testing). Specific criteria not detailed in this summary.
    Tip ShapingYielded acceptable safety & performance outcomes (as part of overall performance testing). Specific criteria not detailed in this summary.
    Tensile StrengthYielded acceptable safety & performance outcomes (as part of overall performance testing). Specific criteria not detailed in this summary.
    Guidewire Compatibility TestingYielded acceptable safety & performance outcomes (as part of overall performance testing). Specific criteria not detailed in this summary.
    Performance Under Simulated ConditionsYielded acceptable safety & performance outcomes (as part of overall performance testing). Specific criteria not detailed in this summary.
    Torsional StrengthYielded acceptable results. Specific criteria not detailed in this summary.
    TorqueabilityYielded acceptable results. Specific criteria not detailed in this summary.
    Corrosion ResistanceYielded acceptable results. Specific criteria not detailed in this summary.

    Study Proving Device Meets Acceptance Criteria:

    The study conducted was a series of "Safety and Performance Tests" on the materials comprising the PVS 1300 Synchro® 0.014" & PVS 1600 Synchro® 0.010" guidewires. These tests were performed in accordance with ISO 11070:1998, Sterile, Single-Use Intravascular Catheter Introducers.

    2. Sample Size for the Test Set and Data Provenance

    The document does not specify the exact sample sizes used for each individual performance test (e.g., how many guidewires were tested for tensile strength or fatigue). It generally states "Subject product testing has yielded acceptable safety & performance outcomes."

    The data provenance is from laboratory testing of the manufactured guidewires by Precision Vascular. It is prospective in the sense that these tests were conducted specifically for this 510(k) submission to demonstrate the device's performance. The country of origin of the data is not explicitly stated, but the manufacturer is based in West Valley City, UT, USA, suggesting the testing was likely conducted in the USA or by a contracted lab.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This question is not applicable. The "ground truth" for a physical medical device like a guidewire is its adherence to predefined engineering specifications and performance standards through physical testing, not expert interpretation of data.

    4. Adjudication Method for the Test Set

    This question is not applicable, as there was no expert review or "ground truth" establishment in the context of interpretation that would require an adjudication method. The results are from objective physical tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    This question is not applicable. This is a physical medical device, not an AI or imaging device that would involve human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by engineering and material specifications and internationally recognized standards (e.g., ISO 10993-1, ISO 11070:1998). The performance tests directly assess whether the device meets these objective, quantifiable standards.

    8. The Sample Size for the Training Set

    This question is not applicable. There is no algorithm or AI model being "trained" for this device.

    9. How the Ground Truth for the Training Set was Established

    This question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K023700

    Validate with FDA (Live)

    Date Cleared
    2002-12-04

    (30 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PVS 1600 Synchro® 0.010" Neuro Guidewire series of products is intended for neurovascular use. It can be used to selectively introduce and position catheters and other interventional devices within the neurovasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.

    Device Description

    The PVS 1600 is a member of the Synchro® Neuro Guidewire family of products having a 0.010" outside diameter, being a sterile, single use/disposable product, with a shapeable tip which is used to gain intravascular access to and facilitate the positioning and exchange of interventional devices in small diameter, tortuous vasculature for neuro diagnostic and interventional procedures. The guidewire can be torqued to facilitate navigation through the vasculature. A torque device, (Merit Medical Systems (K936032)) is supplied with the wire to facilitate deployment & positioning. A guidewire introducer (B. Braun (K760389)) is also supplied and may be used to aid introduction of the guidewire into the catheter hub and/or hemostasis valve and to gently shape the guidewire's distal flexible tip, if desired, according to standard practice. Neither the guidewire introducer or the torque device are intended to enter the body. The product is projected to be provided in a 180cm - 300cm length range, with 200cm being nominal. The Nitinol tip length is projected to be presented in a 35cm - 65cm range, with 55cm being nominal. A traditional range of flexibility profiles shall also be provided, ranging from standard (stiff) to flex (soft). The device is coated on the outer diameter with a lubricious coating over the distal segment of the device. The marker coil is platinum wire at the distal tip of the device to aid visualization under fluoroscopy. The subject device has the ability to access distal, tortuous vasculature, with steerability and torque transmission properties.

    AI/ML Overview

    The provided document describes the PVS 1600 Synchro® 0.010" Neuro Guidewire and its demonstration of substantial equivalence to predicate devices, rather than a study with specific acceptance criteria and detailed performance metrics as would typically be presented for a new or significantly modified device under a comprehensive clinical trial.

    However, based on the information provided, I can extract the relevant details regarding the safety and performance tests conducted to support its substantial equivalence.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not specify quantitative acceptance criteria with corresponding performance statistics in a tabular format as one would expect from a controlled study with predefined thresholds. Instead, it states that "Subject product testing has yielded acceptable safety & performance outcomes" and that tests "also yielded acceptable results." The criteria are implied to be met if the PVS 1600 Synchro® 0.010" Neuro Guidewire is comparable to its predicate devices and conforms to relevant standards.

    Therefore, the table below provides the performance tests conducted and the general outcome stated in the document. The "Acceptance Criteria" are inferred from the context of demonstrating substantial equivalence to predicate devices and adherence to relevant standards (ISO 10993-1, ISO 11070:1998).

    Acceptance Criteria (Implied)Reported Device Performance
    Biocompatibility in accordance with ISO 10993-1Materials test results confirmed biocompatibility of the subject device when tested as an external communicating, blood contact, short duration (<24 hours) device.
    Conformance to ISO 11070:1998 for Intravascular Catheter Introducers (Verfification Testing)
    - Dimensional InspectionSubject product testing yielded acceptable safety & performance outcomes.
    - Fatigue AssessmentSubject product testing yielded acceptable safety & performance outcomes.
    - Tip FlexibilitySubject product testing yielded acceptable safety & performance outcomes.
    - Tip ShapingSubject product testing yielded acceptable safety & performance outcomes.
    - Tensile StrengthSubject product testing yielded acceptable safety & performance outcomes.
    - Guidewire Compatibility TestingSubject product testing yielded acceptable safety & performance outcomes.
    - Performance under Simulated ConditionsSubject product testing yielded acceptable safety & performance outcomes.
    Torsional Strength (Comparable to Predicates)Yielded acceptable results.
    Torqueability (Comparable to Predicates)Yielded acceptable results.
    Corrosion Resistance (Comparable to Predicates)Yielded acceptable results.
    Technological similarities to predicate devicesTechnological similarities between the PVS 1600 Synchro® 0.010" Neuro Guidewire and the PVS 1300 Synchro™ predicate remain identical. Comparative truth for competitive predicate device features including basal design and dimensions, generic materials & construction, and hydrophilic coating. Where characteristics differ, no new questions raised regarding safety or efficacy.

    2. Sample Size and Data Provenance for the Test Set:

    • Sample Size: The document does not specify numerical sample sizes for the performance tests conducted. It refers to "Performance testing of materials" and "Subject product testing."
    • Data Provenance: The document does not provide details about the country of origin of the data or whether the tests were retrospective or prospective. These appear to be laboratory-based engineering performance tests rather than clinical studies on patient data.

    3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set):

    Not applicable. The document describes engineering and materials performance tests, not clinical studies involving expert interpretation or ground truth establishment based on human review.

    4. Adjudication Method (Test Set):

    Not applicable, as no adjudication by experts for ground truth is mentioned. The tests appear to be objective measurements against engineering standards and comparison to predicate devices.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. The document describes a 510(k) submission for substantial equivalence based on technological characteristics and performance testing, not a clinical trial evaluating human reader performance with or without AI assistance.

    6. Standalone (Algorithm Only) Performance Study:

    Not applicable. The device is a physical medical instrument (guidewire), not an AI algorithm or software device that would perform a standalone analysis. The performance studies focus on the physical and material properties of the guidewire.

    7. Type of Ground Truth Used:

    For the performance tests, the "ground truth" is implied to be:

    • Compliance with established international standards (ISO 10993-1 for biocompatibility, ISO 11070:1998 for catheter introducers).
    • Comparative performance against existing legally marketed predicate devices, demonstrating similar safety and efficacy profiles.
    • Objective measurements of physical and mechanical properties.

    8. Sample Size for the Training Set:

    Not applicable. The guidewire is a physical device, not an AI, machine learning, or software device that requires a training set.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable. There is no training set mentioned or implied as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020733

    Validate with FDA (Live)

    Device Name
    PVS 1500 SDS
    Date Cleared
    2002-03-21

    (15 days)

    Product Code
    Regulation Number
    870.1210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PVS 1500 SDS is intended to be used to access tortuous vasculature for infusion of diagnostic, embolic, and therapeutic agents into the distal, peripheral, coronary, and neurovasculature, and for guide wire exchange/support during diagnostic Or interventional procedures.

    The PVS 1500 SDS is intended for use in accessing distal peripheral, coronary, and neurovasculature for selective, controlled, regional delivery of diagnostic, embolic, and therapeutic agents into selected vessels.

    Device Description

    The PVS 1500 SDS is a 2.4 F tubular device, 155 cm in length, with a lumen to be used for delivery of contrast, drugs, or embolics. The lumen is constructed from a polyment material and has an inside diameter of 0.021". The device is coated on the outer diameter with a lubricious coating over the distal segment of the device. Two radiopaque markers are positioned 31 mm apart at the distal tip of the device to aid visualization under fluoroscopy. The proximal end of the device has a standard luer adapter for attachment of accessories and can be used to flush the 1500 SDS. The subject device has the ability to access distal, tortuous vasculature over a guide wire, deliver embolics and agents, and has the ability to be steered like a guide wire as needed.

    AI/ML Overview

    The provided text describes a medical device, the PVS 1500 SDS (an infusion catheter), and its testing to demonstrate substantial equivalence to predicate devices, rather than a study evaluating an AI/ML powered device. Therefore, many of the requested categories for AI/ML device studies are not applicable.

    Here's an analysis based on the provided text:

    Acceptance Criteria and Device Performance for PVS 1500 SDS

    The study primarily focused on demonstrating the PVS 1500 SDS's substantial equivalence to existing predicate devices by meeting established performance standards for infusion catheters.

    Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Tests PerformedReported Device Performance
    BiocompatibilityISO 10993-1, Biological Evaluation of Medical Devices - Part 1Confirmed biocompatibility as an external communicating, blood contact, short duration (<24 hours) device.
    Safety and PerformanceISO 10555-1, Sterile, Single-Use Intravascular Catheters - Part 1Yielded acceptable results.
    - Dimensional inspectionAcceptable results.
    - Hub integrityAcceptable results.
    - Flow rate measurementsAcceptable results.
    - Burst strengthAcceptable results.
    - Tensile strengthAcceptable results.
    - Guide wire compatibility testingAcceptable results.
    - Performance under simulated conditionsAcceptable results.
    - Torsional strengthAcceptable results.
    - TorqueabilityAcceptable results.
    - Corrosion resistanceAcceptable results.

    Information Not Applicable or Not Provided in the Text (Primarily relevant for AI/ML studies):

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): The document doesn't specify sample sizes for each performance test, nor does it mention data provenance in the context of an AI/ML test set. The tests described are laboratory-based engineering and material performance tests.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not applicable. "Ground truth" in this context refers to the physical and chemical properties of the device, established through standardized testing procedures, not expert interpretation of data.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. Performance was measured against established engineering standards.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for this device's performance is derived from standardized engineering and biocompatibility testing against recognized industry standards (ISO 10993-1, ISO 10555-1).
    7. The sample size for the training set: Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.

    Summary of Study Approach:

    The study for the PVS 1500 SDS was a series of benchtop and laboratory-based engineering and biocompatibility tests. The objective was to demonstrate that the device's physical, mechanical, and biological properties met established standards for infusion catheters and were substantially equivalent to legally marketed predicate devices. The "acceptance criteria" were the thresholds and specifications defined by relevant ISO standards (ISO 10993-1 and ISO 10555-1) for medical devices of this type. The device "met the acceptance criteria" by yielding "acceptable results" in all conducted tests.

    Ask a Question

    Ask a specific question about this device

    K Number
    K002907

    Validate with FDA (Live)

    Date Cleared
    2001-03-08

    (171 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K002707

    Validate with FDA (Live)

    Date Cleared
    2000-11-08

    (70 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K990823

    Validate with FDA (Live)

    Date Cleared
    1999-07-23

    (133 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These guidewires are intended for use in percutaneous procedures to introduce and position catheters and other interventional devices within the coronary and peripheral vasculature.

    Device Description

    The PVS 1400 Guidewire is a 0.014" outside diameter, single-use guidewire, which is used to gain intravascular access to and to facilitate the positioning and exchange of catheters in small diameter, tortuous vasculature for coronary and peripheral diagnostic or interventional procedures. The wire can be torqued to facilitate navigation through the vasculature. The proximal wire material is stainless steel. The guidewire is coated to provide lubricity. The tip material of the guidewire is nitinol.

    AI/ML Overview

    Here's an analysis of the provided text regarding the PVS 1400 Guidewire, focusing on acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes performance criteria for various physical characteristics of the guidewire. Acceptance for these criteria is demonstrated by achieving "similar," "better," or meeting specific numerical thresholds compared to predicate devices.

    Acceptance Criteria (Characteristic)Reported Device Performance (PVS 1400 Guidewire)
    Tip FlexibilitySimilar values compared to three predicate devices; similar to two predicates but less flexible than one.
    Tensile StrengthSimilar values compared to two predicate devices; similar to one predicate device and lower tensile strength than the other.
    TorqueabilityMuch better torqueability (much less difference between proximal input angle and distal tip output angle in a simulated tortuous anatomy) than the predicate device against which it was compared.
    Torsional StiffnessApproximately 7.5 times higher than the ACS predicate device. Higher than Microvena UltraSelect® Standard Tip and Boston Scientific/Medi-Tech Target Stubbie® predicate devices.
    Torsional StrengthApproximately 3 times higher than the ACS predicate device. Higher than Microvena UltraSelect® Standard Tip and Boston Scientific/Medi-Tech Target Stubbie® predicate devices.
    Turns to Failure (Torsional)Approximately the same or more turns to failure as Microvena UltraSelect® Standard Tip and Boston Scientific/Medi-Tech Target Stubbie® predicate devices.
    Catheter CompatibilitySimilar to that of the three predicate devices against which it was compared.
    Coating Flake ResistanceSimilar to that of the predicate device against which it was compared.
    Biocompatibility (Cytotoxicity, Systemic Toxicity, Intracutaneous Reactivity, Sensitization, Hemocompatibility)Established as biocompatible based on testing per FDA guidance for an external communicating device in contact with circulating blood for less than 24 hours.
    Sterilization (Ethylene Oxide Residuals)Ethylene Oxide: ≤ 25 ppm
    Ethylene Chlorohydrin: ≤ 25 ppm
    Ethylene Glycol: ≤ 250 ppm

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document does not explicitly state the sample size for each performance test. It refers to "testing" that was performed on the PVS 1400 Guidewire, implying a set of units were tested.
    • Data Provenance: The data appears to be prospective in nature, as it describes tests specifically conducted on the PVS 1400 Guidewire to compare it to predicate devices. The country of origin for the data is not specified, but the submission is to the US FDA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    Not applicable. This device, being a guidewire, relies on objective physical performance tests rather than expert opinion for its "ground truth" or performance evaluation. The "experts" would be the engineers and technicians conducting the physical tests according to established protocols.

    4. Adjudication Method for the Test Set:

    Not applicable. The performance evaluation is based on objective measurements and comparisons to predicate devices, not on expert adjudication of subjective assessments.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No. This type of study (MRMC) is typically used for diagnostic devices (e.g., imaging AI) where human readers interpret cases. The PVS 1400 Guidewire is a physical medical device (guidewire) whose performance is evaluated through bench testing.

    6. If a Standalone Study (algorithm only without human-in-the-loop performance) was done:

    Yes, in essence. The performance data presented (tip flexibility, tensile strength, torqueability, etc.) are all standalone evaluations of the PVS 1400 Guidewire's physical characteristics, often compared to predicate devices. There is no "algorithm" in the sense of a software-based AI, but the device's inherent performance is tested independently of human intervention during its function in the context of these tests.

    7. The Type of Ground Truth Used:

    The ground truth used for performance validation is primarily objective physical measurements and comparisons to established predicate devices. For biocompatibility, the ground truth is established by standardized biological tests and their respective pass/fail criteria. For sterilization, the ground truth is the measured residual levels of sterilants compared to specified limits.

    8. The Sample Size for the Training Set:

    Not applicable. As a physical medical device, there is no "training set" in the context of machine learning or AI. Its design and manufacturing process are refined based on engineering principles and material science, not data training.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable for the reasons stated above (no training set for a physical device). The "ground truth" for the device's fundamental design aspects would come from engineering specifications, material properties, and clinical requirements for guidewires.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1