Search Results
Found 7 results
510(k) Data Aggregation
(28 days)
Philips Lumea IPL is indicated for the removal of unwanted hair for a single user. It is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Philips Lumea IPL is an Intense Pulsed Light (IPL) device intended for removal of unwanted hair. The device is to be used by persons in the age range of 18-65 years with fair up to medium brown skin tones, and naturally dark blond, brown, dark brown and black hair.
The Philips Lumea IPL applies spectrally adjusted light to the user skin, which is absorbed by the melanin and after some applications hair sheds naturally and growth is delayed or inhibited.
The Philips Lumea IPL is an Over-the-counter hand-held device powered by a wall mount adapter. It is provided in a non-sterile state and intended to be used by single person for multiple uses in homecare environment.
N/A
Ask a specific question about this device
(95 days)
Philips Lumea IPL is indicated for the removal of unwanted hair for a single user. It is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Not Found
This document is a 510(k) clearance letter from the FDA for the Philips Lumea IPL device. It does not contain any information about acceptance criteria or a study proving that an AI/software device meets acceptance criteria.
The letter primarily covers:
- The FDA's review and determination of substantial equivalence for the Philips Lumea IPL.
- Regulatory information and requirements for marketing the device.
- The device's indications for use: "removal of unwanted hair for a single user" and "permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime."
- The device is for Over-The-Counter Use.
Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving a device meets them, as the provided text doesn't concern an AI/software device or its performance study.
Ask a specific question about this device
(28 days)
The Philips Avent double electric breast pump Advanced is intended to express and collect milk from the breast of a lactating woman, and to alleviate engorgement of the breast. The device is intended for a single user.
The Philips Avent Double electric breast pump Advanced is intended to express and collect milk from the breast of a lactating woman. The device is intended for a single user. The subject devices are for repeated use by a single user and are for use in home and professional healthcare environments. The devices are not sterile. The subject devices are electrically powered double (Double Advanced) breast pumps consisting of the following key components: a motor unit with a press-button user interface, two expression kits with two milk collection containers connected with the motor unit via silicone tubes. The motor unit operates on embedded software. Software updates by end-users are not supported. The breast pump extracts milk from the breasts by creating a seal around the nipple and applying and releasing suction to the nipple. The milk is collection container, which can be used for storage. To prevent milk from flowing into the vacuum system, the expression kit includes a silicone diaphragm that physically separates the milk-contacting pathway from the vacuum system. The motor unit operates on a rechargeable battery or on an external mains adapter that is provided with the system via a micro-USB type connection. The rechargeable battery can be charged from the external mains adapter if the motor unit is not in operation. The breast pump is provided with the same accessories as the predicate device (K201381), with the addition of a breast pump belt. The breast pump belt intended to carry the electrical breast pump while the device is used to express and collect milk. The belt is intended to be worn over clothes. The expression kit is provided in two types: cushion-type (SCF394/61) and flange-type (SCF394/62). The cushion-type expression kit has a silicone cushion which is mounted on the expression kit pump body. The cushion comes into contact with the breast for the duration of the expression session. In the flange-type expression kit, the silicone cushion is integrated in a rigid frame. This integrated part (flange) can be attached to the pump body with a screw connection. The silicone part comes into contact with the breast for the duration of the expression. The rigid frame is not in contact with the breast. All other components of the subject devices are not in contact with the breast.
The provided document is a 510(k) summary for the Philips Avent Double electric breast pump Advanced. This type of regulatory submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving that the device meets specific acceptance criteria through a clinical study with performance metrics in the way one might expect for a novel diagnostic or therapeutic device.
Therefore, the requested information regarding acceptance criteria, reported device performance, sample sizes, expert involvement, multi-reader studies, standalone performance, and ground truth establishment is not directly available or applicable in the context of this 510(k) summary.
The summary focuses on:
- Comparison to a Predicate Device: The core of the 510(k) is to show that the new device (Philips Avent Double electric breast pump Advanced SCF394/61 and SCF394/62) is substantially equivalent to a previously cleared predicate device (Philips Avent Double electric breast pump Advanced K201381). This means demonstrating that it has the same intended use and similar technological characteristics, and that any differences do not raise new questions of safety or effectiveness.
- Non-Clinical Performance Testing: The provided "performance testing" is limited to biocompatibility testing, which is a common requirement for devices that come into contact with the body.
Here's how the requested information maps to the provided document:
1. Table of acceptance criteria and the reported device performance
- Acceptance Criteria for Device Performance (Clinical/Effectiveness): Not explicitly defined in terms of specific clinical efficacy metrics (e.g., milk volume expressed, comfort levels, duration of engorgement relief). The "acceptance criteria" here is primarily substantial equivalence to the predicate device.
- Reported Device Performance (Clinical/Effectiveness): No clinical performance data (e.g., how much milk is expressed, how effectively it alleviates engorgement in users) is reported in this summary. The summary states that the devices have "similar technological features, including design, user interface, vacuum pressure range, cycle speeds, and power source" and that any differences "do not raise different types of safety and effectiveness questions."
The only specific performance data mentioned are related to the biocompatibility studies:
| Test Type | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Cytotoxicity | Pass | Passed |
| Skin Sensitization | Pass | Passed |
| Skin Irritation | Pass | Passed |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Clinical Test Set: Not applicable. No clinical test set for efficacy or effectiveness is described.
- Biocompatibility Test Set: The sample sizes for the in vitro and in vivo biocompatibility tests are not specified in the document. Data provenance (e.g., country of origin, retrospective/prospective) is also not specified for these tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Clinical/Effectiveness Ground Truth: Not applicable. No clinical ground truth was established by experts for efficacy or effectiveness. The device's performance is assumed to be substantially equivalent to the predicate.
- Biocompatibility Ground Truth: Not applicable in the sense of expert consensus for a "ground truth." Biocompatibility testing follows established protocols and standards (ISO 10993) where results are interpreted against predefined pass/fail criteria.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. There was no clinical test set requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is a breast pump, not an AI-powered diagnostic or decision support tool. Therefore, MRMC studies and AI assistance comparisons are not relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not applicable. This device is a mechanical/electrical breast pump, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For demonstrating substantial equivalence: The "ground truth" is the performance and safety profile of the legally marketed predicate device (K201381). The comparison is based on technical specifications and adherence to recognized standards.
- For biocompatibility: The ground truth is established by standardized testing protocols (ISO 10993) which define acceptable biological responses.
8. The sample size for the training set
- Not applicable. This device does not use machine learning or AI models that require a training set.
9. How the ground truth for the training set was established
- Not applicable. As there is no training set mentioned, there is no ground truth establishment for it.
In summary: The provided 510(k) premarket notification is for a powered breast pump, which is typically cleared based on substantial equivalence to an existing product and adherence to general controls and performance standards, rather than complex clinical trial data demonstrating specific performance metrics against a defined acceptance criterion. The "study" referenced is primarily a comparison of technological characteristics to a predicate device and non-clinical biocompatibility testing.
Ask a specific question about this device
(128 days)
The Philips Avent single/double electric breast pump Advanced is intended to express and collect milk from the breast of a lactating woman, and to alleviate engorgement of the breast. The device is intended for a single user.
The Philips Avent Single/Double electric breast pump Advanced is intended to express and collect milk from the breast of a lactating woman. The device is intended for a single user.
The subject devices are for repeated use by a single user and are for use in home and professional healthcare environments. The devices are not sterile.
The subject devices are electrically powered single (Single Advanced) and double (Double Advanced) breast pumps consisting of the following key components: a motor unit with a press-button user interface, and expression kit with milk collection container (double pump: two kits and containers) connected with the motor unit via a silicone tube. The motor unit operates on embedded software. Software updates by end-users are not supported.
The breast pump extracts milk from the breasts by creating a seal around the nipple and applying and releasing suction to the nipple. The milk is collected in a milk collection container, which can be used for storage. To prevent milk from flowing into the vacuum system, the expression kit includes a silicone diaphragm that physically separates the milk-contacting pathway from the vacuum system.
The motor unit operates on an external mains adapter that is provided with the system via a micro-USB type connection. Models SCF392 and SCF394 include a rechargeable battery that can be used for device operation. The rechargeable battery can be charged from the external mains adapter if the motor unit is not in operation.
The expression kit is provided in two types: cushion-type and flange-type.
The cushion-type expression kit has a silicone cushion which is mounted on the expression kit pump body. The cushion comes into contact with the breast for the expression session.
In the flange-type expression kit, the silicone cushion is integrated in a rigid frame. This integrated part (flange) can be attached to the pump body with a screw connection. The silicone part comes into contact with the breast for the duration of the expression. The rigid frame is not in contact with the breast.
All other components of the subject devices are not in contact with the breast.
The provided document is a 510(k) summary for the Philips Avent Single/Double electric breast pump Advanced. This type of document is a premarket notification to the FDA to demonstrate that a device is substantially equivalent to a legally marketed predicate device.
The questions you are asking typically relate to the performance of AI/ML-driven medical devices in diagnostic or screening applications, where metrics like sensitivity, specificity, and reader performance are critical.
This document, however, describes a powered breast pump, which is a physical device, not an AI/ML diagnostic tool. Therefore, many of your specific questions regarding acceptance criteria and studies (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC studies, standalone performance, training set size, ground truth for training set) are not applicable to the information provided for this type of medical device.
I can, however, extract the information that is relevant to the document for a medical device such as a breast pump, which focuses on safety and performance in a different context.
Here's the relevant information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document doesn't provide a direct "acceptance criteria table" with numerical thresholds for clinical performance metrics (like accuracy for an AI device). Instead, it lists various engineering and safety standards that the device complies with, along with technical specifications that define its functional performance. The "reported device performance" is compliance with these standards and the specified operating ranges.
| Category | Acceptance Criteria / Standard | Reported Device Performance |
|---|---|---|
| General Compliance | Substantial equivalence to predicate device (K161532) in terms of indications for use, safety, and effectiveness. | "The performance testing described above demonstrate that the subject devices are as safe and effective as the predicate device and supports a determination of substantial equivalence." |
| Electrical Safety | IEC 60601-1:2005 (3rd Edition) with US deviations per AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009(R)2012 and A2:2010/(R)2012 (Consolidated Text) IEC 60601-1-11 Edition 2.0 2015-01 (for use in the home) | Complies with IEC 60601-1:2005 and IEC 60601-1-11 Edition 2.0 2015-01. |
| Electromagnetic | IEC 60601-1-2:2014 | Complies with IEC 60601-1-2:2014. |
| Software | FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (May 11, 2005) | Software Verification and Validation Testing performed as recommended in the FDA Guidance Document. |
| Biocompatibility | ISO-10993 (including cytotoxicity, sensitization, and irritation tests: ISO 10993-5:2009, ISO 10993-10:2010) | Complies with ISO-10993 (cytotoxicity, sensitization, irritation). "All milk contacting components are compliant with 21 CFR 174-179, 21 CFR 177.1520 and 21 CFR 177.2600." |
| Lithium-Ion Battery | IEC 62133-2 Edition1.0 2017-02 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems. | Lithium-Ion Battery Testing according to IEC 62133-2 Edition1.0 2017-02. |
| Backflow Protection | N/A (implicit: effective separation of milk-contacting pathway from vacuum system) | Backflow protection testing performed. Device features a silicone diaphragm on the expression kit that separates the vacuum system of the motor unit from milk-contacting parts (similar to predicate). |
| Vacuum Performance | Stimulation Mode: Vacuum range for predicate: 127 ± 15 mmHg. Expression Mode: Vacuum range for predicate: -168 to -250 mmHg. (Subject device aims to fulfill comparable function within safe limits.) | Stimulation Mode: -45 to -150 mmHg Expression Mode: -45 to -270 mmHg (Note: These ranges are comparable to the predicate. The subject device's range for expression is wider, extending to a higher suction level, but this is presented as a difference not necessarily an exceedance of criteria.) |
| Cycle Speed | Stimulation Mode: Cycle speed for predicate: 85 to 120 cycles/min. Expression Mode: Cycle speed for predicate: 42 to 53 cycles/min. (Subject device aims to fulfill comparable function within safe limits.) | Stimulation Mode: 105 to 120 cycles/min Expression Mode: 53 to 85 cycles/min (Note: The subject device's cycle speeds are generally within or overlapping the predicate device's ranges, or considered functionally equivalent by the FDA.) |
| Use-life | N/A (implicit: device maintains performance over its expected lifetime) | Use-life testing performed. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document. The document describes compliance with engineering and electrical standards through testing, not a clinical study with human subjects for performance evaluation in the way an AI diagnostic device would be tested. The tests are focused on the device's hardware and software functionality, and material safety.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable/not provided. Ground truth established by medical experts is paramount for diagnostic AI/ML devices. For a breast pump, "ground truth" relates to physical and electrical performance, and biocompatibility, which are assessed through engineering and materials science tests, not expert clinical consensus on interpretations.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable/not provided. Adjudication methods are used in studies involving subjective interpretation, typically by multiple human readers, which is not the case for device safety and performance testing for a breast pump.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC study is not applicable/not provided. This type of study is relevant for AI-assisted diagnostic tools. The breast pump is a therapeutic/supportive device, not a diagnostic one.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable for this device in the context of typical AI/ML evaluation. While the device contains software, its "performance" is mechanical and electrical, not an AI algorithm making independent diagnostic decisions. The software verification and validation are for the embedded control software.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device's performance relies on objective measurements and adherence to established engineering, electrical, and biocompatibility standards. For example:
- Electrical safety tests verify voltage, current, and insulation against specified limits.
- EMC tests verify emission and immunity levels against regulatory standards.
- Vacuum/cycle performance tests measure actual vacuum levels and cycles per minute against the device's specified operating ranges.
- Biocompatibility tests use standardized laboratory assays to detect cytotoxicity, sensitization, and irritation.
8. The sample size for the training set
This information is not applicable/not provided. There is no "training set" in the context of an AI/ML algorithm for this breast pump. The software is likely deterministic control logic, not a machine learning model that requires training data.
9. How the ground truth for the training set was established
This information is not applicable/not provided due to the absence of a machine learning training set.
Ask a specific question about this device
(88 days)
The Philips Avent Comfort Single/Twin Electric Breast Pump is intended to express and collect milk from the breast of a lactating woman. The device is intended for a single user.
The subject devices are electrically powered single (Comfort Single) and double (Comfort Double) breast pumps consisting of a motor and pump body, press-button user interface, and collection kit. A vacuum pump creates a vacuum at the breast of the lactating woman. This vacuum is released after a short period of time. By repeating this behavior, a cyclic vacuum generation pattern is created which enables expressing of milk from the breast. The expressed milk will be collected in the container which can be used for storage. The breast pump supports different operating modes: stimulation, expression 1, expression 2 and expression 3. The stimulation mode is entered upon start of the device when suction is felt on the breast. Once the milk starts flowing an expression mode may be chosen that has the most comfort, e.g. the combination of rhythm and suction level.
The provided text describes the 510(k) summary for the Philips Avent Comfort Single/Twin Electric Breast Pump (K161532). It outlines performance testing, but it does not contain information about acceptance criteria or a study proving that a "device" (in the context of an AI/ML medical device) meets specific performance criteria like sensitivity, specificity, or accuracy. This document is for a traditional medical device (a breast pump), not an AI/ML algorithm. Therefore, many of the requested fields are not applicable.
However, I can provide information based on the performance testing that was conducted for this breast pump, mapping it to the closest relevant categories where possible.
Acceptance Criteria and Device Performance (Breast Pump)
Since this is a physical medical device and not an AI/ML algorithm, the "acceptance criteria" are related to safety, electrical performance, software validation, usability, biocompatibility, and vacuum performance, rather than diagnostic accuracy metrics.
1. Table of Acceptance Criteria and the Reported Device Performance
| Category / Test Type | Acceptance Criteria (Implied by Standards Compliance) | Reported Device Performance |
|---|---|---|
| Electrical Safety and EMC | Compliance with AAMI / ANSI ES60601-1:2005 (Medical electrical equipment – Part 1: General requirements for basic safety and essential performance) and AAMI / ANSI / IEC 60601-1-2:2007 (Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests). | The device complies with the stated standards. Specific values are not provided, but compliance implies meeting safety thresholds for electrical hazards and electromagnetic interference/emission. |
| Software Verification and Validation | Compliance with AAMI / ANSI / IEC 62304:2006 (Medical device software - Software life cycle processes) and FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". | Software V&V testing was conducted complying to the stated standard and guidance. This implies the software functions as intended and is safe. |
| Usability Testing | Adherence to FDA Guidance 1497, Medical Device Use-Safety Incorporating Human Factors into Risk Management. The implicit criterion is that the device is safe and effective for its intended user population, preventing use errors, and is at least substantially equivalent in usability to the predicate device. | Testing demonstrated that the usability of the subject device is substantially equivalent to the predicate device. |
| Biocompatibility Testing | Compliance with ISO 10993-1:2009 (Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process), ANSI / AMI / ISO 10993-5:2009 (Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity), and ISO 10993-10:2010 (Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization). The implicit criterion is that patient-contacting materials do not pose unacceptable biological risks (e.g., cytotoxicity, irritation, sensitization). | The testing demonstrated that the patient contacting materials are biocompatible. |
| Vacuum Performance Test (Key Performance Specs) | - Suction levels (stimulation): within acceptable ranges and comparable to predicate (e.g., 127mmHg +/- 15mmHg compared to 40 - 140mmHg for predicate).- Cycles per second (stimulation): within acceptable ranges and comparable to predicate (e.g., 1.42 - 2.0 compared to 1.7 - 1.93 for predicate).- Suction levels (expression): within acceptable ranges and comparable to predicate (e.g., 168 – 250mmHg compared to 45 - 245mmHg for predicate).- Cycles per second (expression): within acceptable ranges and comparable to predicate (e.g., 0.70 - 0.87 compared to 0.83 – 1.36 for predicate).- Maximum Vacuum: comparable to predicate (e.g., 250mmHg compared to 270mmHg for predicate).- Vacuum stability, durability, pump temperatures during operation. | - Suction levels (stimulation): 127mmHg +/- 15mmHg- Cycles per second (stimulation): 1.42 - 2.0- Suction levels (expression): 168 – 250mmHg- Cycles per second (expression): 0.70 - 0.87- Maximum Vacuum: 250mmHgThe testing demonstrated the mechanical performance and performance specifications of the subject device and validated that the performance of the subject device is substantially equivalent to the predicate device. |
Regarding AI/ML specific questions:
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not applicable. This device is a breast pump, not an AI/ML algorithm. Performance testing was conducted on hardware, software functionality, and physical characteristics.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. Ground truth in the context of diagnostic accuracy for AI/ML does not apply here. Experts involved in testing would be engineers, usability specialists, and potentially clinicians for biocompatibility or usability assessments, but not to establish "ground truth" for diagnostic purposes.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. No adjudication method for diagnostic outcomes. Testing involved validation against standards and comparison to a predicate device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. No MRMC study was conducted as this is not an AI-assisted diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is a physical breast pump; there is no standalone algorithm in the sense of an AI/ML diagnostic or assistive tool.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not applicable in the AI/ML diagnostic sense. For this device, "ground truth" equates to established engineering specifications, safety standards (e.g., electrical safety, EMC, biocompatibility), and the performance characteristics of the legally marketed predicate device.
8. The sample size for the training set
- Not applicable. There is no "training set" in the context of AI/ML for this device. Software verification and validation would use test cases, not a training set.
9. How the ground truth for the training set was established
- Not applicable. See point 8.
Ask a specific question about this device
(102 days)
The Philips Self-Adhesive Electrodes are intended for use with the PulseRelief device for applying Transcutaneous Electrical Nerve Stimulation (TENS) and Electrical Muscle Stimulation (EMS)
The Philips Self-Adhesive Electrodes is used as a transcutaneous electrical nerve stimulation electrode in conjunction with the Philips PulseRelief device which is an electrical stimulator for TENS or EMS. Electrodes are passive devices serving as an interface between a user's skin and a neurostimulation device. It is composed of: 1. Metal target, 2. Rivet, 3. Textile adhesive layer, 4. Carbon layer, 5. Hydrogel layer, 6. Backing sheet. Proper current distribution is delivered via use of a printed silver pattern.
The provided text describes the acceptance criteria and performance data for the Philips Self-Adhesive Electrodes in the context of demonstrating substantial equivalence to a predicate device for FDA 510(k) clearance. This is not a study proving a device meets acceptance criteria for an AI/ML algorithm or a new medical device that requires clinical performance studies based on efficacy or accuracy.
The "acceptance criteria" here refers to the parameters assessed for substantial equivalence with a legally marketed predicate device (Axelgaard ValuTrode® Neurostimulation Electrodes K130987). The "study" refers to the performance data gathered to support this claim, primarily focusing on physical and electrical characteristics as well as biocompatibility, rather than clinical efficacy studies often associated with AI/ML devices.
Therefore, for most of the requested information points (especially those related to AI/MRMC studies, ground truth, experts, and data provenance related to clinical performance), the document does not provide the relevant data because this is a 510(k) submission for a simple medical device (electrodes), not an AI/ML diagnostic tool.
I will populate the table and answer the questions based only on the information available in the provided document, acknowledging where the information is not applicable or not present for this type of device submission.
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (from predicate or general standards for electrodes) | Reported Device Performance (Philips Self-Adhesive Electrodes) |
|---|---|
| Biocompatibility: | |
| - Cytotoxicity test | Tested in accordance with FDA Blue Book Memorandum #G95-1 and ISO 10993-1. (Results implicitly positive, as the conclusion states the device is safe and effective). |
| - Sensitization test | Tested in accordance with FDA Blue Book Memorandum #G95-1 and ISO 10993-1. (Results implicitly positive, as the conclusion states the device is safe and effective). |
| - Irritation test | Tested in accordance with FDA Blue Book Memorandum #G95-1 and ISO 10993-1. (Results implicitly positive, as the conclusion states the device is safe and effective). |
| Electrical Performance: | |
| - Current distribution (equal) | Tested to assure the current is equally distributed. |
| - Impedance (proper electrical connection) | Tested for Impedance to assure proper electrical connection to the skin. |
| - Average power density (max 0.25 W/cm²) | Calculated average power density: 0.0026 W/cm². This is below the maximum value of 0.25 W/cm². (Calculation: Max current 60mA, Max voltage 60V, Surface area 25cm² (50x50mm). Worst case r.m.s. current in program 1: (2*90µs/10ms)*60mA = 1.08mA. Current density = 1.08mA/25cm² = 43µA r.m.s. Power = 43µA * 60V = 0.0026 W/cm².) |
| Mechanical Performance: | |
| - Initial adhesion force | Tested. (Results implicitly positive, as the conclusion states the device is safe and effective). |
| - Adhesion force after multiple uses | Tested. (Results implicitly positive, as the conclusion states the device is safe and effective). |
| - Connector/Magnetic clamping reliability (multiple use) | Tested to assure multiple use and to assure the TENS PulseRelief stays fixed to the Philips Self-Adhesive Electrode during use. (Results implicitly positive, as the conclusion states the device is safe and effective). |
| Packaging/Storage: | |
| - Storage conditions (41°F to 80.6°F / +5°C to +27°C) | Tested for storage conditions. (Results imply compliance with specified range). |
| - Accelerated aging (function as expected within expiration date) | Tested for accelerated aging in worst storage condition to assure function within the expiration date. (Results implicitly positive, as the conclusion states the device is safe and effective). |
2. Sample size used for the test set and the data provenance
The document does not specify the explicit sample sizes for the various performance tests (e.g., number of electrodes tested for electrical properties, adhesion, accelerated aging).
The data provenance is from tests conducted on the Philips Self-Adhesive Electrodes, which are manufactured by Axelgaard (the same manufacturer as the predicate device). The nature of the testing described (biocompatibility, electrical, mechanical, packaging) suggests it is prospective testing specifically for this device's submission. No geographical origin of the testing data is specified beyond the parent company being Philips Consumer Lifestyle, located in Stamford, Connecticut, USA.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable to this type of device submission. The tests performed are engineering, material, and safety tests, not clinical evaluations requiring expert interpretation or ground truth establishment in a diagnostic sense.
4. Adjudication method for the test set
This information is not applicable. Adjudication methods are typically used in clinical studies involving human interpretation or multi-reader scenarios, not for the direct physical/electrical testing described for this electrode.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This device is an electrode, not an AI-assisted diagnostic tool. Therefore, this section is not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
No, a standalone algorithm performance study was not done. This device does not involve an algorithm.
7. The type of ground truth used
The "ground truth" for the performance tests outlined here is derived from engineering and material standards, electrical measurement principles, and established biocompatibility testing protocols (e.g., ISO 10993-1). It's based on objective quantifiable results against predefined acceptance limits, not expert consensus or direct clinical outcomes in the diagnostic sense.
8. The sample size for the training set
This information is not applicable. This device is not an AI/ML algorithm and therefore does not have a "training set."
9. How the ground truth for the training set was established
This information is not applicable, as there is no training set for this device.
Ask a specific question about this device
(95 days)
The OTC TENS/EMS stimulator PulseRelief is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities(arm) and lower extremities (leg) due to strain from exercise or normal household work activities. It should be applied to normal, healthy, dry and clean skin of adult patients, and is to be used for stimulate healthy muscles in order to improve and facilitate muscle performance.
PulseRelief is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities and can be used to stimulate healthy muscles in order to improve and facilitate muscle performance. Pulse Relief is designed to be used at home, by adults of all genders.
For Transcutaneous Electrical Nerve Stimulation (TENS) Self-adhesive electrodes are placed on the skin close to the area of pain. The device delivers pulses across the skin through these electrodes. This activates the underlying sensory nerves. The user can choose 15 pre-set TENS programs with different pulse settings. In each program, the intensity of the pulse can be adjusted.
For Electrical Muscle Stimulation (EMS). the electrodes are placed near the muscle to be stimulated. The device generates electrical pulses and passes these pulses across the surface of the skin to activate the underlying motor nerves. This makes the muscle contract. Contraction helps to condition the muscle in in order to facilitate performance.
The PulseRelief system consists of the following elements:
- Control unit with cable: generates the TENS/EMS pulses; connects to the 1st electrode; communicates with the mobile device; connects to the battery unit. The control unit has an on-off button.
- Battery unit: contains a battery and charging circuit; connects the 200 electrode to the control unit.
- Electrodes: Self-adhesive re-usable hydrogel electrodes with metal target plate for magnetic clamping to the control unit and battery unit.
- App on mobile device. The app controls the functions of the PulseRelief device via a Bluetooth Low Energy connection between the mobile device and the PulseRelief device.
- Adapter: for charging the battery unit via a micro-USB socket. To connect the USB-cable to the battery unit, the control unit needs to be disconnected.
This document, K151035, is a 510(k) submission for the PulseRelief device, a Transcutaneous Electrical Nerve Stimulator (TENS) and Electrical Muscle Stimulator (EMS). It makes a claim of substantial equivalence to a predicate device (K130802). The document discusses performance data related to safety and electromagnetic compatibility (EMC), software verification, usability, and risk management. However, it does not contain information about acceptance criteria or a study that proves the device meets specific performance acceptance criteria related to its intended therapeutic efficacy (e.g., pain relief or muscle performance improvement).
The document focuses on demonstrating that the PulseRelief device is as safe and effective as the predicate device by meeting recognized standards for medical device development and manufacturing, rather than presenting a clinical study with performance-based acceptance criteria for its therapeutic claims.
Therefore, the following points will reflect the information available in the provided text.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
Based on the provided document, the "performance" described is primarily related to safety, electrical specifications, and adherence to regulatory standards, rather than direct therapeutic efficacy. Specific numerical acceptance criteria for clinical outcomes (e.g., percentage reduction in pain, improvement in muscle strength) are not provided.
The document indicates compliance with various standards, which serve as performance/acceptance criteria in the context of regulatory submission for substantial equivalence.
| Acceptance Criteria Category | Specific Criteria (Standard/Requirement) | Reported Device Performance |
|---|---|---|
| Biocompatibility | FDA Blue Book Memorandum #G95-1 & ISO 10993-1 (Cytotoxicity, Sensitization, Irritation for intact skin contact) | "The biocompatibility evaluation for the Brand X device was conducted in accordance with the FDA Blue Book Memorandum #G95-1... and International Standard ISO 10993-1... As dictated by the application and duration of contact with the intact skin, the battery of testing included the following tests: Cytotoxicity, Sensitization, Irritation." (Implies compliance, as no adverse findings are reported) |
| Electrical Safety | IEC 60601-1, IEC 60601-1-6, IEC 60601-1-11, IEC 60601-2-10 | "The system complies with the IEC 60601-1, IEC 60601-1-6, IEC 60601-1-11 and IEC 60601-2-10 standards for safety." |
| Electromagnetic Compatibility (EMC) | IEC 60601-1-2, FCC Part 15 (Subpart B - Unintentional Radiators, Subpart C - Intentional Radiators) | "The system complies with... the IEC 60601-1-2 standard for EMC, and IEC 62133 for battery safety. For FCC part 15 RADIO FREQUENCY DEVICES, Subpart B-Unintentional Radiators, Subpart C-Intentional Radiators." |
| Battery Safety | IEC 62133 | "The system complies with... IEC 62133 for battery safety." |
| Software Verification & Validation | FDA's Guidance for Industry and FDA Staff: "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," IEC 62304 (for Moderate level of concern) | "Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance... The software for this device was considered as a 'Moderate' level of concern. IEC 62304 was followed." (Implies compliance) |
| Usability | IEC 62366, FDA Guidance 1757 ("Applying Human Factors and Usability Engineering to Optimize Medical Device Design") | "Usability testing according to IEC 62366 and following FDA Guidance 1757, Applying Human Factors and Usability Engineering to Optimize Medical Device Design, was conducted." (Implies compliance/satisfaction of guidance, as no issues are reported) |
| Risk Management | ISO 14971:2007 Medical Devices - Application of risk management to medical devices (Second Edition) | "The risks to people, property and the environment associated with the use of the PulseRelief was managed according to ISO 14971:2007..." (Implies compliance) |
| Symbols | ISO 15223-1:2012 Medical Devices - Symbols To Be Used With Medical Device Labels, Labelling, And Information To Be Supplied - Part 1: General Requirements (Second Edition) | "The symbols used on the PulseRelief itself, on its packaging or in the associated documentation is conform [sic] ISO 15223-1:2012..." (Implies compliance) |
| Output Specifications (TENS mode) | Adherence to ranges for voltage, current, duration, frequency, net charge, current density, power density. For example, max average power density below 250mW/cm2 as per FDA guidance. | E.g., Maximum Output Voltage: 31 @500Ω, 69 @ 2 kΩ, 70 @ 10 kΩ. Pulse Duration: 60 ~ 350 μs. Frequency: 1 ~ 100 Hz. Max Average Power Density: 0.24 ~ 1.69 @500Ω (stated as "well below 250mW/cm2 as required by the FDA-guidance"). (See tables on pages 9-11 for full details compared to predicate.) |
| Output Specifications (EMS mode) | Adherence to ranges for voltage, current, duration, frequency, net charge, current density, power density. | E.g., Maximum Output Voltage: 31 @500Ω, 69 @ 2 kΩ, 70 @10 kΩ. Pulse Duration: 150 ~ 350 µs. Frequency: 40 ~ 65 Hz. Max Average Power Density: 0.62~1.15 @500Ω. (See tables on pages 9-11 for full details compared to predicate.) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document primarily focuses on engineering testing, regulatory compliance, and comparison to a predicate device rather than clinical trials for validating therapeutic efficacy. Therefore, there is no information provided regarding a clinical "test set" sample size or data provenance (country of origin, retrospective/prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Given the nature of the submission (safety, electrical performance, and regulatory compliance), there is no mention of a "ground truth" derived from experts related to clinical outcomes (e.g., pain assessment or muscle performance). The "ground truth" here is adherence to technical standards and specifications.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
No adjudication method is mentioned as the submission is not focused on clinical diagnostic or treatment performance that would require such a process.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done or mentioned. This type of study is not applicable to a TENS/EMS device submission focused on substantial equivalence based on technical and safety specifications. There is no AI component described in the device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The PulseRelief device itself is the "standalone" device providing stimulation. The "performance" described here is the device's ability to meet its technical specifications and safety standards, as detailed in Table 1 above. There is no separate algorithm-only performance study mentioned distinct from the device's main function.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" in this submission relates to:
- Compliance with recognized electrical, safety, software, and biocompatibility standards: This is implicitly defined by the standards themselves (e.g., IEC 60601-1, ISO 10993-1).
- Device specifications: The engineering measurements of the device's electrical output parameters (voltage, current, frequency, pulse duration, etc.) are compared against defined design specifications and generally against the predicate device's reported parameters.
- Risk assessment: Compliance with ISO 14971 for risk management.
There is no ground truth related to clinical outcomes data for therapeutic efficacy presented in this document.
8. The sample size for the training set
No training set is mentioned or applicable in the context of this device and submission type. The device is not based on machine learning or AI that would require a training set.
9. How the ground truth for the training set was established
As there is no training set, this question is not applicable.
Ask a specific question about this device